Trek PD04002A User Manual

Model 04002A PD
Charged Plate Monitor Controller
! Utilizes Standard,
Custom and Wide­Temperature Range Charged Plates
!
! High-Temperature
Charged Plate (Model PD04002AP) 25 mm x 25
Trek’s PD04002A versatile microprocessor based charged plate monitor is ideally suited to monitor the performance of air ionizers that are used in critical wide temperature operations associated with semiconductor, LCD, electronic assembly, and other processes where static charge accumulations pose a threat to production yields and/or product quality.
The Model PD04002A utilizes a new technique which requires less than 0.1 picoampere of ion field current to achieve full accuracy and stability. Other product designs require up to 2000 times more ion current to be drawn from the ion field to achieve stability.
The ultra sensitivity of the Model PD04002A makes it the ideal choice in air ionizer monitor applications where small area ion collecting sensors are used to meet the requirement for high accuracy ion balance measurement to levels of less than one volt. The voltage of the charged plate is initialized prior to discharge time testing. Balance tests, positive discharge tests, and negative discharge tests can be initialized from a remote location through digital command signals.
The charged plate voltage can be remotely monitored through a rear panel connection. The voltage output at the Output “HI” terminal in reference to the Output “LO” terminal (ground) is a proportional buffered representation of the charged plate voltage at a scale factor of 1 to 22 (other scale factors are available). The Model PD04002A wide temperature range charged plate monitor functions with standard, custom, or high temperature charged-plates. The monitor and high­temperature charged-plate can be used on a table top or securely mounted by fasteners to the bottom of the unit.
mm (1² x 1²) Cable Length: 3 meters Capacitance: 20 pF
±5 pF
!
! Operating Range
-60° to 160° C
!
! Relative Humidity
To 85%, noncondensing
!
! Voltage Monitor Output
for Remote Monitoring
!
! Exceptional Accuracy
and Stability
!
! Compact Design
!
! Compliant
Measurement and Power Solutions
www.trekinc.com
TM
Model 10/40A Specifications
Model Charged Plate Monitor PD04002A
Performance Features (cont.) General
Charged Plate Voltage Range Dimensions
0 to ±55 V DC or peak AC, nominal.
Digital Command Terminals
Plate (O) discharges the floating
plate to zero volts.
122 mm W x 43 mm H x 153 mm D (4.8² W x1.7² H x 6² D).
Plate (+) precharges the floating
Charged Plate Voltage Measurement Accuracy
0.2% of full scale.
Large Signal Bandwidth (-3 dB)
DC to greater than 200 Hz.
plate to +55 V.
Plate (-) precharges the floating
plate to -55V.
Positive discharge tests, Negative discharge tests, and Balance tests can be initialized
Weight
0.3 kg (0.6 lb).
Power
The Model PD04002A operates from a +24 V ±10% supply @ 0.21 amps.
from a remote location through
Small Signal Bandwidth (-3 dB)
DC to 2.5 kHz.
Charged Plate Self-Discharge Rate (no incident ion flow)
Less than 2 V per minute at 55 V for relative humidity up to 85%.
Charged Plate Capacitance
20 pF ± 4 pF. (The capacitance is independent of charged plate connecting cable length.)
these digital command signals. The voltage of the charged plate is initialized to +55 V or -55 V for discharge time testing or zero for balance testing.
All the above digital command inputs are at a 10 V DC High Level. A minimum of 6 mA of current sink capability is required to pull the digital command inputs to ground (0 V).
High-Temperature Charged Plate (Trek Model PD04002AP)
Connection Block
Phoenix Contact Connector. Connections include:
+24 V DC power input (+), +24 V DC return (-), Voltage Monitor (Hi), Voltage Monitor (Lo), Digital Reset (plate 0), Digital Reset (plate +), and Digital Reset (plate-)
Charged Plate Assembly Input Connector
BNC connector.
25 mm x 25 mm (1² x 1²).
Features
Voltage Monitor Output V
0 to +5 V DC represents a charged plate voltage of
-55 V to +55 V (other scale factors are available).
When the charged plate is at
-55 volts, the voltage at the Voltage Monitor Output V HI is 0 V. When the charged plate is at 0 volts, the voltage at the voltage monitor is +2.5 volts. When the charged plate is at +55 volts, the voltage at the voltage monitor is +5 volts.
Accuracy
0.2% of full scale.
Cable Length
3 meters.
Capacitance
20 pF 5 pF.
±
Operating Range
-60° to 160° C.
Relative Humidity
To 85%, noncondensing.
Mounting
Model PD04002A Monitor
Four (4) M4 mounting screws can be used to secure the monitor assembly from the bottom.
Charged Plate
One (1) M4 mounting screw can be used to secure the charged-plate monitor assembly from the bottom.
User Connections
Connection block and
Unit Operating Conditions
Temperature
5 °C to 35 °C.
Relative Humidity
To 85%, noncondensing.
Certification
TREK, INC. certifies that each Model PD04002A Charged Plate Monitor is tested and calibrated to specifications using measurement equipment traceable to the National Institute of Standards and Technology or traceable to consensus standards. CE compliant.
Caution: These instruments are designed to make electrostatic voltage measurements only. Do not use this instrument to make measurements of “hard” voltage sources or voltage sources which can deliver currents high enough to cause harmful shocks or personal injury.
M4 ground lug.
Copyright ã 2011 TREK, INC. All specifications are subject to change. 1128/DEC