Trek 800 User Manual

Trek Model 800 Infinitron
Ultra-High Impedance Voltmeter
The ±100 V Trek Model 800 Infinitron Voltmeter may be used in either contacting or non-contacting mode to acquire precision surface voltage measurements. It is especially beneficial when used with applications that demand infinitely high loading impedance levels far beyond the reach of currently available high impedance voltmeter instruments. The Model 800 comes with a guarantee of virtually no modification of the object being measured. This allows the instrument to indicate, with high precision, the voltage level of both conductive and insulative objects and surfaces.
Key Specifications
! Measurement Range: 0 to ±100 V DC or peak AC
! Measurement Accuracy: Better than ±0.1% of full scale (voltage monitor output) ! Speed of Response: Less than 3.5 ms for a 100 V input step ! Input Characteristics
Resistance: Greater than 1 x 10 W Capacitance: Less than 1 x 10 F
16
-15
®
®
Typical Applications Include
! Accurate reading of electrostatic voltage levels associated with ESD sensitive components, circuits and surfaces
Features and Benefits
! Two voltage monitor outputs scale factors at 20:1 and 1:1 ! Probe electrode may be easily replaced with other sensor tips
Monitor provides a low voltage replica of the measured electrostatic potential for monitoring purposes or for use as
!
a feedback signal in a closed loop system Easy-to-read LED display
!
CE compliant
!
Designed to be operated on a bench top
!
Available Probes
! Model 800P Probe
- Sensor
1.27 mm gold plated rounded tip electrode. The sensors may be easily replaced, dependent on the measurement requirements
- Orientation
Pencil probe structure with end contact sensor.
- Probe Dimensions
192.2 mm L x 33.9 mm Diameter (7.57” L x 1.34” Diameter)
- Probe Cable Length
3 m ±76 mm (10 ft ±3 in.)
TREK, INC. • 190 Walnut Street • Lockport, NY 14094 • USA • 800-FOR TREK
716-438-7555 • 716-201-1804 (fax) • www.trekinc.com • sales@trekinc.com
Model 800 Specifications
Performance
Measurement Range
0 to ±100 V DC or peak AC
Measurement Accuracy
Voltage Monitor Output
Better than ±0.1% of full scale
Voltage Display
Better than ±0.1% of reading, ±1 digit (referred to measured input)
Speed of Response (10% to 90%)
Less than 3.5 ms for 100 V step
Large Signal Bandwidth (-3 db)
DC to greater than 100 Hz for 200 V pp Input Resistance
Greater than 1 X 10
16
W
Input Capacitance
-15
Less than 1 X 10 F
Voltage Monitor
Output
Two BNC outputs provide buffered low­voltage replica of the measured voltage
Ratio
20:1 of the measured to output voltage and 1:1 representation
Output Voltage
Less than ±5 mV
Output Noise
Less than 10 mV rms
Output Impedance
Less than 0.5 W
Features
LCD Display Screen with LED back-light
127 mm x 38 mm (5" x 1.5") screen displays prompts and voltage measurements with date and time
Display Resolution
240 x 64 pixels
Hold Mode
Controlled by local pushbuttons or remotely through HOLD BNC Connector
Zero
Controlled by local pushbuttons or remotely by applying a momentary (100 ms minimum) TTL low through a designated BNC Connector
Null Offsets (Auto Null Adjustment)
Nulling of the input voltage and current is necessary when the unit is first powered on, or when the display prompts that an auto null adjustment is necessary. Placing the probe in a holster on the side of the unit and pressing the NULL OFFSETS pushbutton provides nullification of the voltage offsets and the “leakage” currents, which are dependent on environmental conditions. An audio tone will signal when the null offsets process is completed.
TEMP/RH Meter Connector
A connector receives an input from an optional thermo hygrometer (Omega Model HH311) and enables the Model 800 to display environmental data on the LCD screen.
USB Communications Port
4-pin Type B connector
Mechanical
Dimensions
10.2 cm H x 31.6 cm W x 24.1 cm D (4” H x 12.5” W x 9.5” D)
Weight
2.5 kg (5.5 lb)
BNC Connectors
Two Voltage Monitors ( Hold Reset
20:1, 1:1)
Ground Receptacle
Threaded ground stud
Power ON/OFF
A momentary push-button
Probe Connector Location
Front panel
Operating Conditions
Temperature
10°C to 35°C (50°F to 95°F)
Relative Humidity
0 to 80%, noncondensing
Altitude
To 2000 m (6561.68 ft.)
Electrical
AC Line Cord Receptacle
Selected per geographic destination
Line Voltage
12 V AC, ±10% @ 1 A
DC Connector
2.1 mm DC power plug
Accessories
Operator’s Manual with software
PN: 23338
Ground Cord
PN: N9044
AC Line Cord
Selected per geographic destination
USB Cable
1.8 m (6 ft) PN: BA103 Probe
Model 800P PN: 17491
Probe Tips
1.27 mm gold plated rounded tip electrode
Additional tips can be customized in various other sizes to cater to a wide field of applications; please contact the factory for more information
Carrying Case
PN: 47032
Note: All specifications measured with a 5-minute warmup time.