Trek Model 542A Non-Contacting
Electrostatic Voltmeter for EOS/ESD
The Trek Model 542A Electrostatic Voltmeter (pictured at left
with the Model 542P-S probe) provides accurate noncontacting measurements of the electrostatic surface voltage
associated with EOS/ESD processes. The instrument is
configured with a miniature electrostatic field chopper probe
that can be remotely located and easily positioned within
process equipment to provide highly accurate, non-contacting,
DC-stable, spacing-independent voltage measurements in
either ionized or non-ionized environments. A 20x4
alphanumeric LCD screen displays the present measured
voltage, the positive peak voltage value, the negative peak
voltage value and additional menu information.
Key Specifications
!
Measurement Range
Model 542A-1: ±10 kV DC or peak AC
Model 542A-2: ±20 kV DC or peak AC
!
Measurement Accuracy -
Better than ±5% of the reading ±0.2% of full scale over a probe-tosurface separation of
Model 542A-1: 15 mm to 30 mm
Model 542A-2: 30 mm to 60 mm
Alphanumeric LCD Display: 20 x 4 characters
!
Typical Applications Include
! Semiconductor
! LCD
! Electronic assembly
! ESD-sensitive processes
Walking Test Adapter
Optional walking test adapter kit
[CN 1K040] for the model 542A
allows analysis of charge levels
on the human body in support of
EN 1815, Assessment of Static
Electrical Propensity on Resilient
and Textile Floor Coverings.
Features and Benefits
! Chopper probe is DC-stable with or without incident air flow
! Drift-free measurements
! LCD screen displays present voltage and holds the most positive and negative values
! Visual and audible alarms activate when the preset voltage threshold levels are reached
! Analog voltage monitor output
! USB and RS-232 serial ports
! NIST-traceable Certificate of Calibration provided with each unit
! compliant
! Optional Walking Test Adapter kit available
TREK, INC. • 190 Walnut Street • Lockport, NY 14094 • USA • 800-FOR TREK
716-438-7555 • 716-201-1804 (fax) • www.trekinc.com • sales@trekinc.com
Model 542A Specifications
0 to ±10 kV DC or peak AC
0 to ±20 kV DC or peak AC
Speed of Response (10%
to 90%)
Less than 50 ms for a ±1 kV step
Better than ±5% of the reading ±0.2% of
full scale over a probe-to-surface
separation of:
30 mm to 60 mm
Drift with Time
Less than ±1% full scale, non-cumulative
1/1000th of the measured voltage
1/2000th of the measured voltage
Activated if measured voltage exceeds
preset threshold limits; positive/negative
limits may be programmed separately
Front-panel LED illuminated at threshold
Programmable pulsating or continuous
tone. (+) and (-) alarms have different tone
rates for the pulsating tone selection
Form C relay contact rated at 175 V
maximum, 5 W
TTL output with a TTL low as the alarm
“ON” status
Resets Alarms and Peak Hold to zero
Adjustable to produce zero volts when
probe coupled to a known zero voltage
source
20 character by four line (20x4) LCD
displays the present voltage and holds the
most positive and most negative
measured values
Provides control of specific functions and
acquires sensor data utilizing Trek
software and a PC connected to the RS232 serial port or the USB Type B port
(connectors for each are on the back
panel)
Provides a current of 4 mA to 20 mA
representing -10 kV to +10 kV (542A-1) and
-20 kV to +20 kV (542A-2)
Used to select and program menu options the
↑↓ set the alarm threshold voltages, alarm
conditions and alarm reset type
97 mm H x 152 mm W x 204 mm D
(3.8” H x 6” W x8” D)
15°C to 35°C (59°F to 95°F)
5% to 85% RH, non-condensing
15 V DC ±20%, 800 mA, power bus or
AC/DC adapter with a 2.1 mm DC plug.
Positive polarity connected to the center
contact
PN: 24004 (with PN 45833 software CD)
Model No: 542P-45D. Aperture 3.8 mm (0.15”)
diameter. Dimensions 11 mm dia x 79 mm L
(0.43” dia. x 3.1” L)
Model No: 542P-S. Aperture 4 mm (0.156”)
diameter. Dimensions 13 mm dia x 75 mm L
(0.51” dia x 3” L)
5 meters (16 ft.) nominal
Probe Aperture
Probe Aperture
11 mm
13 mm
79 mm
9.7 mm
75 mm
9.7 mm
*Measured using the true rms feature of the HP Model 34401A digital multimeter
Measurement and Power Solutions
**Vacuum application probes are also available
Copyright ã 2013 TREK, INC. All specifications are subject to change. 1339/JRB