Transmille EA022 Specifications

PROCESS CONTROL CALIBRATION ADAPTER MODEL EA022
The EA022 adapter expands the functionality of the 1000 series calibrator to enable the calibration of process control calibrators & pressure transducers. This compact adapter also features a 24V loop supply and is the ideal addition to expand workload capabilities for the 1000 Series ultra portable calibrator.
MEASUREMENT CAPABILITIES
The EA022 process control calibration adapter allows the measurement of up to 30V DC, ideal for verification of output functions of process control meters
The EA022 also features the ability to measure up to 30mA directly with the 1000 series calibrator, making it a useful on-site compact tool for measuring the full range of process control calibrator outputs
The EA0022 has is the ability to measure resistance up to 50kOhms using a 4 wire connection, enabling measurements of resistance and PRT outputs of process control calibrators
PRESSURE TRANSDUCER SUPPORT
The EA022 also provides a direct connection on the rear of the unit for Transmille TPM pressure modules.
The EA022 also features a 24V loop supply ideal for powering transducers under calibration.
DIRECT FRONT PANEL
CONTROL
NO PC REQUIRED
USB SOFTWARE CONTROL ALSO AVAILABLE
Specifications*
Function
Voltage measurement
Current measurement
Resistance measurement
* Due to continuous development specifications are subject to change
Range
100mV : 1V : 30V
30mA
50 to 50kWW
Accuracy % Setting % Range Floor
0.02% 0.01% 5uV
0.03% 0.01% 500nA
0.05% 0.02% 10mW