![](/html/19/1933/193373b7ebb33b29f934e9346a9eb715409e1337cad4735454d18f9dce7063ba/bg1.png)
PROCESS CONTROL CALIBRATION ADAPTER
MODEL EA022
The EA022 adapter expands the functionality of the
1000 series calibrator to enable the calibration of
process control calibrators & pressure
transducers. This compact adapter also features a
24V loop supply and is the ideal addition to expand
workload capabilities for the 1000 Series ultra
portable calibrator.
MEASUREMENT CAPABILITIES
The EA022 process control calibration adapter allows the
measurement of up to 30V DC, ideal for verification of output
functions of process control meters
The EA022 also features the ability to measure up to 30mA
directly with the 1000 series calibrator, making it a useful on-site
compact tool for measuring the full range of process control
calibrator outputs
The EA0022 has is the ability to measure resistance up to
50kOhms using a 4 wire connection, enabling measurements of
resistance and PRT outputs of process control calibrators
PRESSURE TRANSDUCER SUPPORT
The EA022 also provides a direct
connection on the rear of the unit for
Transmille TPM pressure modules.
The EA022 also features a 24V loop
supply ideal for powering
transducers under calibration.
DIRECT FRONT PANEL
CONTROL
NO PC REQUIRED
USB SOFTWARE CONTROL
ALSO AVAILABLE
Specifications*
Function
Voltage measurement
Current measurement
Resistance measurement
* Due to continuous development specifications are subject to change
Range
100mV : 1V : 30V
30mA
50 to 50kWW
Accuracy
% Setting % Range Floor
0.02% 0.01% 5uV
0.03% 0.01% 500nA
0.05% 0.02% 10mW