Transmille EA015 Data sheet

EA015 Thermocouple Simulation Extended Specifications TRANSMILLE LTD
General Specifications
Adapter connection
Connects to male 'D' type adapter interface connector on 3000 Series front panel
Connections 1 x 9 way male 'D' type connector
1 x Neutral (copper) industry standard thermocouple plug Connection to Calibrator Via supplied 9 Way male to female serial lead (straight through connection) Output impedance 10 Ohms
Type Rel. (°C) Rel. (°C) Rel. (°C) Rel. (°C)
J -180°C to 150°C 0.04 0.045 0.05 0.07
150°C to 750°C 0.24 0.27 0.3 0.42
K -140°C to 200°C 0.08 0.09 0.1 0.14
200°C to 1340°C 0.28 0.315 0.35 0.49
T -250°C to 400°C 0.16 0.18 0.2 0.28 R -50°C to 500°C 0.16 0.18 0.2 0.28
500°C to 1700°C 0.8 0.9 1 1.4
S -50°C to 1200°C 0.48 0.54 0.6 0.84
1200°C to 1700°C 1.28 1.44 1.6 2.24
B 0°C to 1200°C 0.08 0.09 0.1 0.14
1200°C to 1820°C 1.04 1.17 1.3 1.82
N -270°C to 260°C 0.08 0.09 0.1 0.14
260°C to 1300°C 0.32 0.36 0.4 0.56
All thermocouple simulation specifications ± 2uV. Note 1 : Specifications apply between 17°C and 27°C. Outside this range an allowance of 0.18 x 1 Year Spec. per °C should be added.
Does not include cold junction compensation errors
EA015 Thermocouple Simulation Extended Specifications Due to continuous development specifications may be subject to change.
V4.00
EA015 30A 2-10-50 Clamp Coil Extended Specifications TRANSMILLE LTD
y
A%A%A%A
A%A%A%A
General Specifications
Adapter Connection 4x 4mm safety sockets mounted on the rear of the unit Colour Cream Connection to Calibrator Via adapter interface & adapter lead set (suppled) Coil Configuration 2 Turn (LHS) : 10 Turn (RHS) : 50 Turn (CENTRE) Coil Type High accuracy balanced configuration Min. internal jaw dimensions 10mm (2 Turn & 10 Turn) : 25mm (50 Turn) Maximum Current 30A Maximum RMS voltage 4V Frequency Range DC to 500Hz Construction Loose wound coil (for heat dissipation) in moulded ABS enclosure Durability Fully enclosed coil for maximum protection from mechanical damage Compatibilit
Designed for use with Transmille 3000 Series calibrators and ProCal Software
2 Turn Coil Accuracy (Input 0 to 30A : Freq. DC - 30Hz to 60Hz : Effective Output 0 to 60A)
90 Day Rel. 180 Day Rel. 1 Year Rel. 2 Year Rel.
%
Effective accuracy - Coil only (wound clamps) 0.35 + 0.008 0.35 + 0.008 0.35 + 0.008 0.35 + 0.008 Effective accuracy - Coil only (hall effect clamps) 0.48 + 0.07 0.48 + 0.07 0.48 + 0.07 0.48 + 0.07 Total uncertainty with 3050 (All clamps) 0.51 + 0.09 0.51 + 0.09 0.52 + 0.09 0.56 + 0.10 Total uncertainty with 3041 (wound clamps) 0.36 + 0.028 0.36 + 0.028 0.36 + 0.028 0.38 + 0.036 Total uncertainty with 3041 (hall effect clamps) 0.49 + 0.090 0.49 + 0.090 0.49 + 0.090 0.50 + 0.098 Total uncertainty with 3010 calibrator (wound clamps) 0.36 + 0.010 0.36 + 0.010 0.36 + 0.010 0.37 + 0.010 Total uncertainty with 3010 (hall effect clamps) 0.48 + 0.072 0.49 + 0.072 0.49 + 0.072 0.49 + 0.072
10 Turn Coil Accuracy (Input 0 to 30A : Freq. DC - 30Hz to 60Hz : Effective Output 0 to 300A)
90 Day Rel. 180 Day Rel. 1 Year Rel. 2 Year Rel.
Effective accuracy - Coil only (wound clamps) 0.41 + 0.01 0.41 + 0.01 0.41 + 0.01 0.41 + 0.01 Effective accuracy - Coil only (hall effect clamps) 0.59 + 0.11 0.59 + 0.11 0.59 + 0.11 0.59 + 0.11 Total uncertainty with 3050 (All clamps) 0.61 + 0.13 0.62 + 0.13 0.62 + 0.13 0.65 + 0.14 Total uncertainty with 3041 (wound clamps) 0.42 + 0.03 0.42 + 0.03 0.42 + 0.03 0.43 + 0.04 Total uncertainty with 3041 (hall effect clamps) 0.60 + 0.13 0.60 + 0.13 0.60 + 0.13 0.61 + 0.14 Total uncertainty with 3010 calibrator (wound clamps) 0.41 + 0.012 0.42 + 0.012 0.42 + 0.012 0.42 + 0.012 Total uncertainty with 3010 (hall effect clamps) 0.59 + 0.112 0.59 + 0.112 0.60 + 0.112 0.60 + 0.112
50 Turn Coil Accuracy (Input 0 to 30A : Freq. DC - 30Hz to 60Hz : Effective Output 0 to 1500A)
90 Day Rel. 180 Day Rel. 1 Year Rel. 2 Year Rel.
%
Effective accuracy - Coil only (wound clamps) 0.24 + 0.04 0.24 + 0.04 0.24 + 0.04 0.24 + 0.04 Effective accuracy - Coil only (hall effect clamps) 0.45 + 0.42 0.45 + 0.42 0.45 + 0.42 0.45 + 0.42 Total uncertainty with 3050 (All clamps) 0.48 + 0.44 0.48 + 0.44 0.49 + 0.44 0.53 + 0.45 Total uncertainty with 3041 (wound clamps) 0.25 + 0.06 0.26 + 0.06 0.26 + 0.06 0.28 + 0.07 Total uncertainty with 3041 (hall effect clamps) 0.46 + 0.44 0.46 + 0.44 0.46 + 0.44 0.47 + 0.45 Total uncertainty with 3010 calibrator (wound clamps) 0.25 + 0.042 0.25 + 0.042 0.25 + 0.042 0.26 + 0.042 Total uncertainty with 3010 (hall effect clamps) 0.45 + 0.42 0.46 + 0.42 0.46 + 0.42 0.46 + 0.42
Accuracy is dependant on proper alignment of the clamp meter within the coil Certain clamp meters have alignment marks which should be aligned with the centre of the coil. Certain types of clamp meter may have additional errors, or be outside the range which can be driven by the 3041/3010 directly
Uncertainty calculated as the square root of the square of coil accuracy + square of calibrator accuracy using empirical data obtained for both wound & hall effect instruments from a wide range of manufacturers Clamp coil adaptor built into EA015 workstation adapter
EA015 30A 2-10-50 Clamp Coil Extended Specifications Due to continuous development specifications may be subject to change.
V4.00