Transmille EA001A Specifications

EA001A Extended Specifications TRANSMILLE LTD
K
General Specifications
Adaptor connection
Connects to male 'D' type adapter interface connector on 3000A Series front panel
Indicators Incorporates 'Active', 'Measure' & 'Source' LED's mounted in the adaptor case Adaptor Dimensions 17cm x 12cm x 3.5cm Connections 1 x 9 way male 'D' type connector
1 x Neutral (copper) industry standard thermocouple plug with integral temperature senso
r
1 x Neutral (copper) industry standard thermocouple socket
with integral temperature sensor Connection to Calibrator Via supplied 9 Way male to female serial lead (straight through connection) Sensor Type
Thermistor
Sensor incorporated into thermocouple plug to eliminate any lead effects Output impedance 10 Ohms
Thermocouple Range 90 Day
1
180 Day
1
1 Year
1
2 Year
Type Source / Measure Rel. (°C) Rel. (°C) Rel. (°C) Rel. (°C)
J -210°C to -100°C 0.18 0.21 0.23 0.32
-100°C to -30°C 0.09 0.10 0.11 0.15
-30°C to 150°C 0.07 0.08 0.09 0.13
150°C to 760°C 0.11 0.13 0.14 0.20
760°C to 1200°C 0.15 0.17 0.19 0.27
-200°C to -100°C 0.22 0.24 0.27 0.38
-100°C to -25°C 0.12 0.14 0.15 0.21
-25°C to 120°C 0.09 0.10 0.11 0.15
-120°C to -1000°C 0.16 0.18 0.20 0.28 1000°C to 1370°C 0.21 0.23 0.26 0.36
T -250°C to -150°C 0.48 0.54 0.60 0.84
-150°C to 0°C 0.08 0.09 0.10 0.14 0°C to 120°C 0.07 0.08 0.09 0.13
-120°C to 400°C 0.09 0.10 0.11 0.15
R -0°C to 250°C 0.64 0.72 0.80 1.12
250°C to 1000°C 0.35 0.40 0.44 0.62
1000°C to 1760°C 0.41 0.46 0.51 0.71
S 0°C to 250°C 0.64 0.72 0.80 1.12
250°C to 1000°C 0.35 0.40 0.44 0.62
1000°C to 1760°C 0.41 0.46 0.51 0.71
1
B 600°C to 800°C 0.58 0.66 0.73 1.02
800°C to 1000°C 0.52 0.59 0.65 0.91 1000°C to 1550°C 0.43 0.49 0.54 0.76 1550°C to 1820°C 0.44 0.50 0.55 0.77
Note 1 : Does not include cold junction compensation errors Cold Junction Compensation Error = ± 0.2°C All thermocouple measurement specifications assume correct compensation cable is being used Specifications apply between 17°C and 27°C. Outside this range an allowance of 0.18 x 1 Year Spec. per °C should be added. All specifications apply for 3000A Firmware Version 12.2.3 onwards
EA001A Extended Specifications Due to continuous development specifications may be subject to change.
Page 1 of 2 - V 1.0
Loading...