Transmille EA001A User Manual

EA001A
Thermocouple Simulation
& Measurement Adapter
Operation Manual
Version 1.00 May 2013 All product names are trademarks of their respective companies
EA001A THERMOCOUPLE SIMULATION & MEASUREMENT ADAPTER OPERATION MANUAL
Guarantee and service
Transmille Ltd. guarantees this instrument to be free from defects under normal use and service for a period of 1 years from purchase. This guarantee applies only to the original purchaser and does not cover fuses, or any instrument which, in Transmille’s opinion, has been modified, misused or subjected to abnormal handling or operating conditions.
Transmille’s obligation under this guarantee is limited to replacement or repair of an instrument which is returned to Transmille within the warranty period. If Transmille determines that the fault has been caused by the purchaser, Transmille will contact the purchaser before proceeding with any repair.
To obtain repair under this guarantee the purchaser must send the instrument in its original packaging (carriage prepaid) and a description of the fault to Transmille at the address shown below. The instrument will be repaired at the factory and returned to the purchaser, carriage prepaid.
Note : TRANSMILLE ASSUMES NO RESPONSIBILITY FOR DAMAGE IN TRANSIT
THIS GUARANTEE IS THE PURCHASER’S SOLE AND EXCLUSIVE GUARANTEE AND IS IN LEIU OF ANY OTHER GUARANTEE, EXPRESSED OR IMPLIED. TRANSMILLE SHALL NOT BE LIABLE FOR ANY INCIDENTAL, INDIRECT, SPECIAL OR CONSEQUENTIAL DAMAGES OR LOSS.
Transmille Ltd. Unit 4, Select Business Centre Lodge Road Staplehurst Kent TN12 0QW United Kingdom
Tel : +44 0 1580 890700 Fax : +44 0 1580 890711
EMail : sales@transmille.com Web : www.transmille.com
Transmille Ltd. Page 2
EA001A THERMOCOUPLE SIMULATION & MEASUREMENT ADAPTER OPERATION MANUAL
TABLE OF CONTENTS
EA001A THERMOCOUPLE SIMULATION & MEASUREMENT ADAPTER.......................................4
MAIN FEATURES..................................................................................................................................5
DEDICATED EXTERNAL ADAPTER FOR THE ULTIMATE IN ACCURACY.......................................................5
DIRECT CONNECTION - NO COMPENSATION CABLES REQUIRED...............................................................5
DESIGN NOTES.....................................................................................................................................6
INTUITIVE OPERATION & CONTROL........................................................................................................6
CONNECTIONS.....................................................................................................................................7
Simulation Function Connection ...............................................................................................7
Measurement Function Connection...........................................................................................7
USING THE THERMOCOUPLE SIMULATION FUNCTION.................................................................8
CONNECTING THE UUT.........................................................................................................................8
USING THE MENUS ................................................................................................................................9
SETTING TEMPERATURE OUTPUTS ......................................................................................................10
Manual Keypad Entry ................................................................................................................10
Deviate Output ...........................................................................................................................10
USING THE THERMOCOUPLE MEASUREMENT FUNCTION.........................................................11
CONNECTING THE UUT.......................................................................................................................11
USING THE MENUS ..............................................................................................................................12
COLD JUNCTION COMPENSATION .................................................................................................14
OVERVIEW .........................................................................................................................................14
AUTOMATIC COLD JUNCTION COMPENSATION .....................................................................................15
MANUAL COLD JUNCTION COMPENSATION ..........................................................................................15
AUTOMATED CALIBRATION USING PROCAL................................................................................16
CARE & MAINTENANCE....................................................................................................................17
CLEANING THE ADAPTER ....................................................................................................................17
HANDLING PRECAUTIONS....................................................................................................................17
SERVICING INFORMATION....................................................................................................................17
Appendix A : Specifications
Transmille Ltd. Page 3
EA001A THERMOCOUPLE SIMULATION & MEASUREMENT ADAPTER OPERATION MANUAL
EA001A Thermocouple Simulation & Measurement Adapter
External Unit
Low noise & Thermally stable
Simulation and measurement capabilities
Dual Integrated temperature sensors for accurate cold junction compensation
11 thermocouple types : Type B
Type C Type E Type J Type K Type L Type N Type R Type S Type T Type U
Direct control from the 3000A Series front panel
Compatible with ProCal calibration software
Transmille Ltd. Page 4
EA001A THERMOCOUPLE SIMULATION & MEASUREMENT ADAPTER
p
OPERATION MANUAL
Main Features
Dedicated External Adapter For the Ultimate in Accuracy
Temperature gradients / thermal EMFs which can be introduced by internally mounted thermocouple simulation methods are one of the greatest sources of error with thermocouple temperature calibration.
For the ultimate in accuracy, Transmille has designed a dedicated external adapter. This keeps the electronics required to generate the low level signals used for thermocouple simulation as close as possible to the measuring input of the thermometer. This allows the signal to be as free from electrical noise as possible, and also eliminates errors caused by heat generated by surrounding electronics in the calibrator.
Direct Connection - No compensation cables required
Uses neutral (copper) thermocouple connectors
CJC sensor built in to thermocouple
plug for automatic cold junction
ensation
com
CJC sensor built in to thermocouple input socket for automatic cold junction compensation
For direct connection to any thermocouple type input without the need for compensation cables
Transmille Ltd. Page 5
EA001A THERMOCOUPLE SIMULATION & MEASUREMENT ADAPTER OPERATION MANUAL
Design Notes
Special care has been taken over the cold junction compensation - a common source of errors in thermocouple calibration. The cold function compensation (CJC) sensor is mounted in the thermocouple plug itself. By measuring the cold junction at the instruments input allows any type of thermocouple to be simulated without using compensation cables.
Intuitive Operation & Control
Easy selection of thermocouple type
Auto / Manual Cold Junction compensation
Quick and easy to use, simply plug in the temperature adapter into the 3000 Series calibrator, select thermocouple simulation mode and thermocouple type using the soft keys and enter the temperature required.