Transmille EA001 Data sheet

EA001 Thermocouple Simulation Extended Specifications TRANSMILLE LTD
General Specifications
Adaptor connection
Connects to male 'D' type adapter interface connector on 3000 Series front panel
Indicators Incorporates an 'Active' LED mounted in the adaptor case Adaptor Dimensions 12.5cm x 6.5cm x 2.5cm Connections 1 x 9 way male 'D' type connector
1 x Neutral (copper) industry standard thermocouple plug
Sensor incorporated into thermocouple plug to eliminate any lead effects Output impedance 10 Ohms
Thermocouple Range
90 Day
1
180 Year
1
1 Year
1
2 Year
Type Rel. (°C) Rel. (°C) Rel. (°C) Rel. (°C)
J -180°C to 150°C 0.04 0.05 0.05 0.07
150°C to 750°C 0.24 0.27 0.30 0.42
K -140°C to 200°C 0.08 0.09 0.10 0.14
200°C to 1340°C 0.28 0.32 0.35 0.49
T -250°C to 400°C 0.16 0.18 0.20 0.28 R -50°C to 500°C 0.16 0.18 0.20 0.28
500°C to 1700°C 0.80 0.90 1.00 1.40
S -50°C to 1200°C 0.48 0.54 0.60 0.84
1200°C to 1700°C 1.28 1.44 1.60 2.24
B 0°C to 1200°C 0.08 0.09 0.10 0.14
1200°C to 1820°C 1.04 1.17 1.30 1.82
N -270°C to 260°C 0.08 0.09 0.10 0.14
260°C to 1300°C 0.32 0.36 0.40 0.56
E 0°C to 800°C 0.64 0.72 0.80 1.12
Note 1 : Does not include cold junction compensation errors All thermocouple simulation specifications ± 2uV. Cold Junction Compensation Error = ± 0.2°C
Specifications apply between 17°C and 27°C. Outside this range an allowance of 0.18 x 1 Year Spec. per °C should be added.
1
EA001 Thermocouple Simulation Extended Specifications Due to continuous development specifications may be subject to change.
V4.20