Transmille 8500 User Manual

Low Thermal Scanner
Operation Manual
Version 1.00 Jan 2012 All product names are trademarks of their respective companies
8500 LOW THERMAL SCANNER OPERATION MANUAL
Guarantee and Service
Transmille Ltd. guarantees this instrument to be free from defects under normal use and service for a period of 1 years from purchase. This guarantee applies only to the original purchaser and does not cover fuses, or any instrument which, in Transmille’s opinion, has been modified, misused or subjected to abnormal handling or operating conditions.
Transmille’s obligation under this guarantee is limited to replacement or repair of an instrument which is returned to Transmille within the warranty period. If Transmille determines that the fault has been caused by the purchaser, Transmille will contact the purchaser before proceeding with any repair.
To obtain repair under this guarantee the purchaser must send the instrument in its original packaging (carriage prepaid) and a description of the fault to Transmille at the address shown below. The instrument will be repaired at the factory and returned to the purchaser, carriage prepaid.
Note : TRANSMILLE ASSUMES NO RESPONSIBILITY FOR DAMAGE IN TRANSIT
THIS GUARANTEE IS THE PURCHASER’S SOLE AND EXCLUSIVE GUARANTEE AND IS IN LEIU OF ANY OTHER GUARANTEE, EXPRESSED OR IMPLIED. TRANSMILLE SHALL NOT BE LIABLE FOR ANY INCIDENTAL, INDIRECT, SPECIAL OR CONSEQUENTIAL DAMAGES OR LOSS.
Transmille Ltd. Unit 4, Select Business Centre Lodge Road Staplehurst Kent TN12 0QW United Kingdom
Tel : +44 0 1580 890700 Fax : +44 0 1580 890711
EMail : sales@transmille.com Web : www.transmille.com
Transmille Ltd. Page 2
8500 LOW THERMAL SCANNER OPERATION MANUAL
TABLE OF CONTENTS
GUARANTEE AND SERVICE................................................................................................................2
8500 LOW THERMAL SCANNER .........................................................................................................4
MAIN FEATURES...................................................................................................................................5
FRONT PANEL ......................................................................................................................................6
REAR PANEL ........................................................................................................................................7
OPERATING NOTES .............................................................................................................................8
INTERFACE / POWER CONNECTIONS...............................................................................................8
CONNECTION RECOMMENDATIONS.................................................................................................9
EXAMPLE CONNECTION : RESISTANCE STANDARD..................................................................................9
EXAMPLE CONNECTION : PRT PROBE ...................................................................................................9
CHANNEL SELECT MODES ...............................................................................................................10
MANUAL CHANNEL SELECT .................................................................................................................10
REMOTE CHANNEL SELECT .................................................................................................................11
SPECIFICATIONS................................................................................................................................12
CARE & MAINTENANCE .....................................................................................................................13
Transmille Ltd. Page 3
8500 LOW THERMAL SCANNER OPERATION MANUAL
8500 Low Thermal Scanner
The model 8500 Low Thermal Scanner provides a low thermal 4 terminal switching system making it ideal for 4 wire resistance measurements in resistance thermometry and for resistance comparisons.
The use of ultra low thermal 4mm terminals make connecting the scanner to UUT's easy, allowing the measurement system to be quickly re-configured. With the output from the scanner located on the rear panel connections can be routed efficiently, providing a neat and simple cabling solution.
Applications include scanning of both electronic and standard reference cells, standard resistance comparisons measurements and PRT probe measurements. Special care has taken in the scanner designed to minimize thermal generated EMF voltages. By reducing internal self heating to almost zero by using a very low power circuit, even the power supply is external, using latching relays which only need a single pulse of power to set and using gold platted de-oxgenated copper terminals thermal voltages have been reduced to less than 150nV.
Control is from the Serial RS232 or USB interface (using converter) or via the front panel control buttons. LED indicators show the selected channel. The internal firmware ensures that two channels can never be selected at the same and that the switching is break before make.
Transmille Ltd. Page 4