Transmille 3010, 3041 User Manual

0 (0)
3000
MULTI PRODUCT CALIBRATORS
SERIES
www.TransmilleCalibration.com
THE VERSATILE SOLUTION
FOR LABORATORIES & SERVICE CENTERS
3000SERIES
Transmille : The Complete Solution
Ü
The widest workload of any multi product calibrator - over 18 types of instruments with power to drive older analogue instruments and the accuracy to calibrate modern digital instruments
Ideal for both laboratory and on-site use. Modern technology provides a fast warm up time and extremely portable design.
Automate calibration with ProCal - the worlds easiest to use multi discipline calibration software for maximum calibration throughput
The most cost effective calibration system on the market - 3 models to suit budget & accuracy requirements
The wide range of options allows the calibrator to be tailored to meet your needs ­expandable for the future
Intuitive front panel design with large LCD display and ergonomic keyboard for ease of use
Full output protection including system ­prevents output being switched on when voltage detected on terminals
2
Guaranteed reliability with 3 year warranty as standard
THE FLEXIBLE SOLUTION FOR YOUR CALIBRATION WORKLOAD
Comprehensive Instrument Calibration
Ü
Meeting Workload Requirements
The 3000 Series calibration system has been designed to help you & your business, allowing a wide range of calibration work to be performed quickly, accurately & economically. Modern calibration is a competitive market, and cost effective equipment and software are key considerations.
Meeting the needs of today's quality systems including ISO9000 / ISO17025 require calibration traceable to national standards of a wide range of instrumentation. The 3000 Series calibration system provides the ideal platform to cover these calibration requirements, whether in the laboratory or on site. Every Transmille calibrator and adapter is supplied with UKAS calibration traceable to National standards, and are ready for use out of the box.
Offering the widest workload coverage of any calibration system presently available, the 3000 series is a true class leading solution. The unique concept of external calibration adapters combined with built-in measurement capability dramatically extends the range of instruments that can be cost effectively calibrated.
Automated Calibration & Management Software
ProCal is the genuinely easy to use calibration software which complements the 3000 Series calibration system to provide a complete laboratory solution. Maximum calibration throughput is achieved with multi discipline support. ProCal-Track adds laboratory management control with dynamic job tracking from goods inwards to despatch ­streamlines laboratory operation.
25PPM MULTI PRODUCT CALIBRATOR
Ü
Ü
Ü Ü
MID RANGE MODELLABORATORY REFERENCE
CALIBRATION ADAPTERS
CALIBRATION SOFTWARE
Ü
ÜÜ
50PPM MULTI PRODUCT CALIBRATOR
ENTRY LEVEL MODEL
TRANSCONDUCTANCE AMP
MULTI FUNCTION WORKSTATION
I-GUARD
8PPM MULTI PRODUCT CALIBRATOR
Ü
AC/DC KILOVOLT AMPLIFIER
2/10/50 TURN CLAMP COIL
Ü
CALIBRATES...
UP TO 7½ DIGIT MULTIMETERS :: 3010
UP TO 5½ DIGIT MULTIMETERS :: 3041
UP TO 4½ DIGIT MULTIMETERS :: 3050
DIGITAL OSCILLOSCOPES TO 600MHz
CLAMP METERS
WATT / POWER METERS
RLC METERS / AC BRIDGES
POWER SUPPLIES
INSULATION TESTERS
PROCESS CONTROL CALIBRATORS
HIGH RESISTANCE MEASUREMENT
ELECTROMETERS
CURRENT SHUNTS
DATA LOGGERS
PRESSURE GAUGES
HV PROBES
TORQUE SCREWDRIVERS
FREQUENCY COUNTER/TIMERS
CHART RECORDERS
AND MORE...
FOR RCD • LOOP • PAT TESTERS
See the Transmille 3200 Electrical Test Equipment Calibrator Brochure.
3
3000SERIES
3000
THE FLEXIBLE SOLUTION FOR YOUR CALIBRATION WORKLOAD
8
PPM ACCURACY MODEL 3010
25
PPM ACCURACY MODEL 3041
PPM ACCURACY MODEL 3050
50
AC/DC VOLTAGE TO 1025V
V
A
AC/DC CURRENT TO 30A 2&4 WIRE RESISTANCE TO 1GW
W
F
H
Hz
RTD PRT
RS
232
GPIB
USB
W
O
Ü
PASSIVE & SIMULATED MODES
CAPACITANCE TO 10mF
PASSIVE & SIMULATED MODES
INDUCTANCE T O 10H
FREQUENCY TO 10MHz
RTD/PRT SIMULATION
RS232 INTERFACE BUILT IN
GPIB / USB OPTIONS
AC/DC POWER CALIBRATION
WITH POWER HARMONICS
OSCILLOSCOPE CALIBRATION
250MHz • 350MHz • 600MHz OPTIONS
BUILT IN ADAPTER INTERFACE
INTEGRATED HIGH ACCURACY MEASUREMENT FUNCTION
THERMOCOUPLE SIMULATION ADAPTER
POWER SUPPLY CALIBRATION ADAPTER
HUMIDITY/TEMPERATURE SENSOR ADAPTER
OSCILLOSCOPE CALIBRATION TEST HEAD ADAPTER
PLUS...HIGH RESISTANCE / pA MEASUREMENT ADAPTER PA SOURCE ADAPTER OPTICAL TACHOMETER CALIBRATION ADAPTER PRESSURE CALIBRATION ADAPTER TORQUE CALIBRATION ADAPTER
Clear presentation of calibrator outputs & status on a
large graphic LCD display for safe & easy operation.
ADVANCED TECHNOLOGY
The 3000 series calibrators use the latest technology from the temperature stabilised reference to the sub ppm linearity D/A.
DDS (Direct Digital Synthesis)
jitter free wideband AC waveforms. Custom waveforms for use in harmonics calibration are available with the 3041/3010 power options.
An ultra-linear VMOS power amplifer
capability with continuous output available up to 20A. A power efficient design with two-stage cooling has resulted in reduced internal thermal EMFs for superior short term stability whilst reducing fan noise for a more comfortable working environment.
intelligent output protection safeguards the calibrator by testing
I-GUARD
the output sockets for voltages applied accidentally when switching from standby to output on - protects your investment from accidental misuse.
4
technology is used for generating stable,
provides a ground breaking 30A
Separate outputs for
Voltage, Low Current and High Current minimise lead changing when connecting. Dedicated BNC outputs for oscilloscope (Amplitude / Timebase / Freq. Sweep) and Frequency.
Soft keys
provide direct access to the menu commands, unlocking the potential of the 3000 Series calibrators.
Large output on and standby buttons
and useful range up/down buttons also add to the ease of use.
RA
PORTABLE
The ultra portable 3000 Series calibrators begin a new era in true on-site calibration. Using the soft, over the shoulder tailored travel case, your calibration system can be taken with you to the calibration work, whether it is in an aircraft hanger, on board a ship or oil rig.
A fast warm up time ensures the 3000 Series will be ready to work wherever you are. Connect to your laptop for a complete automated calibration system.
,
Ergonomic calculator style keypad makes it
easy to enter values. Intuitive function keys and key layout makes using the function control section second nature.
Adapter interface for power,
control and measurement for the range of calibration adapters.
S
SOURCE
MEASURE
ÜÜÜ
The 3000 Series generates all the outputs required for multimeter calibration, plus options for :
Oscilloscope Calibration
base, fast rise and bandwidth up to 600MHz. A key feature is the ability to support the growing number of oscilloscopes with 50V/Div (300V Pk-Pk) ranges.
AC/DC Power Calibration
voltage to 1025V and current to 30A outputs to simulate power up to 30.6kW (1.5MW using the 50 turn coil). Phase angle control up to 359.9° is available.
Harmonics Calibration
3010 includes five preset waveforms and support for custom defined harmonics upload using a PC.
Built-in source and measurement
M
capabilities incorporating a high linearity A/D converter.
COMPLETE WORKLOAD CAPABILITY
including amplitude, time
uses simultaneous generated
option available for 3041 &
Integrated Calibration Workstation
for clamp meter, insulation testers, process control calibrator, temperature meter & optical tachometer calibration
RCL Meters
unique to the Transmille 3000 Series. The passive inductance, resistance & capacitance in the 3000 Series are suitable for measurement using any technique, including . For total flexibility simulated resistance and capacitance options are also available.
Frequency Meter
accuracy to 1ppm Using the external calibration adapters,
Process Control Calibrators, Electrometers, Insulation Testers, Pressure Gauges and Transmitters, even Torque Screwdrivers
can be calibrated using the inductance option,
AC Bridges
calibration up to 10MHz is supported with
Power Supplies,
can be calibrated quickly and efficiently.
5
3000
SERIES OSCILLOSCOPE CALIBRATION OPTION
SCP250 / SCP350 / SCP600
AMPLITUDE
Easily selected using the softkeys, the calibrator produces either a precision 1kHz square wave or a DC level covering the range from 2mV/Div to 50V/Div in a 1,2,5 sequence. Deviation up to 10% in 0.01% steps can be applied using the digital potentiometer. The calibrator's wide range output, giving up to 300V pk-pk, can be used to directly calibrate the ever increasing number of oscilloscopes with amplitude ranges up to 50V/Div placing it in a class leading position.
TIMEBASE
The timebase function of the oscilloscope option provides outputs from 2ns/Div to 5s/Div. Deviation up to 10% in 0.01% steps can be applied using the digital control. To use, simply align the time marker with the graticule display and read the deviation from the % display on the calibrator. The comb waveform used below 100ns is ideal for triggering on both analogue and digital oscilloscopes alike. For faster timebase calibration, a sine wave is produced which makes viewing on bandwidth limited oscilloscopes easier. The timebase output can be used either directly or into a 50 Ohm input.
SPECIFICATIONS
Voltage Amplitude
Range Sequence Frequency Accuracy
Time Markers
Range Sequence Waveshape
Accuracy
Fast Rise*
* 3041 / 3010 only
Leveled Sweep
Range (SCP250) Range (SCP350) Range (SCP600) Leveled Sweep Reference Level Accuracy
50kHz Reference
Frequency Accuracy Level Accuracy
See extended specifications for full details
6
2mV/Div to 50V/Div
1 • 2 • 5
1kHz
0.01% ± 5uV
2ns to 5s
1 • 2 • 5
Comb < 100ns
Sine > 100ns
5ppm
1ns typical
5MHz to 250MHz 5MHz to 350MHz 5MHz to 600MHz
600mV into 50Ohms
50kHz
0.5dbm
30ppm
0.5%
Three options are available for the 3000 Series calibrators which are suitable for calibration of both analogue and digital storage oscilloscopes.
250MHz (SCP250) : 3050
350MHz (SCP350) : 3041 / 3010
600MHz (SCP600) : 3041 / 3010
All options provide the calibration waveforms required for amplitude, timebase and bandwidth including gain and linearity of the horizontal and vertical deflection circuits. To minimise lead changing, the oscilloscope outputs are output from a single BNC connector - for automated calibration an optional 4 channel test head adapter is available.
Amplitude to 50V/Div (300V Pk-Pk) - 1,2,5 sequence
Timebase to 2ns
Separate trigger output
All calibration signals from a single BNC
FAST RISE
A fast rise output is available with a rising edge time of typically 1ns for evaluating overshoot, undershoot and ringing of oscilloscope attenuation and amplifier circuits.
LEVELED SWEEP
The leveled sweep output of the oscilloscope option provides a continuously variable sine wave from 5MHz to 620MHz. A 50kHz reference level waveform is also available to allow the oscilloscope controls to be set to give a 6 graticule height display.
OSCILLOSCOPE CALIBRATION TEST HEAD
SOURCE
EA017
For automating oscilloscope calibration a 4 channel switching test head is available. Connect to the 3000 Series adapter interface and select channel required. For accurate calibration of 50Ohm input oscilloscopes up to 1V/Div a low impedance buffered amplitude output can be selected.
SPECIFICATIONS
Channels
Input Output
Output Impedance
Direct Buffered
See extended specifications for full details
1 4
As Input
Typically 0.2
3000
SERIES POWER CALIBRATION OPTION
PWR50 / PWRSINE / PWRDDS
AFFORDABLE AC/DC POWER CALIBRATION
Three options provide cost effective calibration of power (Watts) and VA ranges on power meters, power/harmonics analysers & clamp meters. The power option for the 3000 series calibrators allows both voltage current output to be generated simultaneously (phantom load) with an adjustable phase angle relationship.
3050 : Sine power option (PWR50)
3041/3010 : 2 options PWRSINE : Sine & DC power PWRDDS : 5 waveforms+custom & DC
Up to 1.5MW (using 50 Turn Clamp Coil)
0° to 359.9° adjustable phase relationship
Any voltage up to 1025 Volts can be set using the normal output ranges and specification of the calibrator. Currents up to 30Amps are available from the output of the calibrator without the need for an external amplifier.
The power function is extremely easy to use - select 'Power' from the soft key menu, connect the power meter to both the voltage and current output terminals and enter the voltage, current and frequency. Phase angle can be adjusted using the softkeys with a resolution of 0.1
The 3000 series calibrator dynamically controls the phase angle between the current and voltage waveforms eliminating errors caused by capacitive or inductive loading experienced when using clamp coils.
TCXO
OSCILLATOR
Extended Power Range Using The 2 / 10 / 50 Clamp Coil
Combined with the optional clamp coil adapter the power calibration option allows a current of 1500Amps to be simulated, and power to1.5Megawatt (1500Amps x 1000Volts).
Power Harmonics Analyser Calibration Using Programmable Waveforms (Option PWRDDS)
Option PWRDDS for the 3041 and 3010 calibrators allows generation of waveforms with harmonic content for calibrating power meters with power harmonics functions. This includes a set of five fixed waveforms, plus the capability to upload user defined waveforms from a PC allowing waveforms with custom harmonic content to be generated.
kWatt hour meter & Energy Calibration.
°. Power output is calculated & displayed in kW.
ACV
WAVEFORM
PHASE
CONTROL
PHASE
CONTROL
PHASE ANGLE
CONTROL
GENERATION
FREQUENCY CONTROL
DIGITAL
MICROPROCESSOR
CONTROL
FREQUENCY CONTROL
ACI
WAVEFORM
GENERATION
Using the 3000 Series VFP the Voltage, current, frequency, phase angle, time and the number of coil turns can all easily be set, click the energy button and the power is turned on for a set period making calibration of kWatt/hours and energy meters easy.
VOLTAGE
O/P
PHASE COMPARATOR
CURRENT
O/P
SPECIFICATIONS
DC Power (3041/3010)
Voltage Range Current Range Resolution 300mA to 2A 2A to 30A Accuracy 300mA to 2A 2A to 30A
AC Power
Voltage Range Frequency Range Resolution 300mA to 2A 2A to 30A
Accuracy : 3041/3010
300mA to 2A 2A to 30A
Accuracy : 3050
300mA to 2A 2A to 20A Phase Angle Resolution Phase Accuracy
See extended specifications for full details
CURRENT
PHASE
VOLTS
0 to 1025V
300mA to 30A
200uA
2mA
0.05% ± 0.35mW/Volt
0.04% ± 3.5mW/Volt
0 to 1025V
40Hz to 1kHz
200uA
2mA
0.05% ± 0.35mW/Volt
0.05% ± 3.5mW/Volt
0.1% ± 0.5mW/Volt
0.1% ± 5mW/Volt
0.1°
0.1°
7
Loading...
+ 7 hidden pages