1 Description
EFM115 1
Ohm Metrics 1
Specifications
2 Operation
Battery Installation 2
Static Field Meter 2
Ion Balance 2
Measuring Two Point Resistance 3
LED Indication Guide 4
Calibration 4
3 Parts Included List 4
4 Service and Warranty 5
1
Description
Ohm Metrics Pocket Static Field Meter:
EFM115
The EFM115 is a convenient electrostatic field meter used to locate and measure static charges and test the balance of air ionizers. Its compact design and four button inter­face makes it easy to operate. The EFM115 measures static voltages within ±20kV (20,000V) at a distance of 1".
The EFM115 can be used to measure low voltages up to 200V in ion balance mode with a built in charge plate. The combination of field meter and ion balance tester makes the EFM115 the ideal meter for static audits, and periodic verification of ionizers and meets periodic test requirements per Compliance Verification ESD TR53 and conforms to ANSI/EOS/ESD (S4.1, S7.1, S12.1, S2.1).
Ohm Metrics test and measurement products, from
Transforming Technologies, are useful, reliable tools for characterizing and identifying the electrical resistance of materials and the performance of personal grounding products.
All Ohm Metrics products are designed to support
ESDA Compliance Verification TR53.
All Ohm Metrics products can be calibrated.
All Ohm Metrics test and measurement products are
warranted for 1 full year.
2
The EFM115 features:
Easy to read, easy to use operations Lightweight, compact design Range light assures accurate and repeatable measure-
ments
Automatic range switching up to ±20kV (20,000V) at
1" (25mm)
Digital and bar graph display Automatic power off after five minutes (may be disabled
if preferred)
Product Specifications
Product Number
EFM155 Electrostatic Field Meter
Measuring range Static voltage ± 0 ~ ± 1.49 kV (Low range) ± 1.0 ~ ± 20.0 kV (High range) Ion balance voltage 0 ~ ± 200 V Measuring distance 25 mm ± 0.5 Accuracy ±10% Ambient conditions 0°C ~ 40 °C, 680%RH or lower Display Large LCD display (Digital and bar graph) Power 9V, 6F22Y manganese battery: 1PCE (Life: Approx. 30hours) Weight Approx. 140g without ion balance plate (Ion balance plate : Approx. 30g)