Electromagnetic Flowmeter LF610 Series Operation Guide
Setting Parameters Flow
Setting Menu
Items Screen Display Setting or selecting value
A
B
B DISPLAY
1 Display 1 DSPL1
First Unit : Volume, Others
%, m
3
, L, mL, bbl, gal, m/s, ft/s, COUNT, RANGE, pt, qt,
CUSTOM, GRAPH (*3)
Second Unit : Time unit
/s, /min, /h, /d
Third unit : Flow direction code B (Automatic selection bi-directionflow) F (Fixed forward flow), R (Fixed reverse flow) D (Difference)
*2 *3
2 Display 2 DSPL2
3 Custom Coefficient CS VAL 0 〜 99999999 4 Custom Unit CS UNIT 7 Characters 5 Measurement Mode EXIT
C
C RANGE 1 Range Type R TYPE SINGLE, 4F-0R, 2F-2R, EXT 2F-0R, EXT 2F-2R *4
2 Range 1 R1
Volume unit m
3
, L, mL, bbl, gal, pt, qt
Time unit /s, /min, /h, /d
Velocity unit m/s, ft/s
3 Range 2 R2
4 Range 3 R3
5 Range 4 R4 6 Range Hysteresis R HYS 0.0 〜 25.0 % (0.1% each)
7 Measurement Mode EXIT
D
D FILTER 1 Damping Constant DAMPING 00.0, 00.5, 01.0 〜 60.0 s 2 Rate-of-change Limit LIM RATE 0.0 〜 30.0 % (0.1% each) 3 Control Limit Time LIM TIME 00 〜 20 s (1s each) 4 Measurement Mode EXIT
E
E LOW CUT 1 Low Cutoff CUT VAL 0.0 〜 10.0 % (0.1% each) 2 Display Low Cutoff DSPL SET LINEAR, LOW CUT 3 Measurement Mode EXIT
F
F ZERO 1 Zero Adjustment ZERO ADJ Hold down the
key 2 Manual Zero MANUAL ±10% of 10m/s-maximum range (0.1% each) *5 3 Measurement Mode EXIT
G
G 4-20mA 1 4-20mA Alam Output ALM 4-20 UNDER 3mA, 4mA, HOLD , OVER 24mA 2 Output Low Limit LOW LIM 4.0mA, 3.2mA, 2.4mA *6 3 Measurement Mode EXIT
H
H DO 1 Digital Output 1 DO1 FUNC
NO USE, HIGH ALM, HH ALM, LOW ALM, LL ALM RNG SIG1, RNG SIG2, PRESET C, CONV ALM EMPTY ALM, PLS OUT, PLS FRD, PLS REV
*7
2 Digital Output 2 (*13) DO2 FUNC 3 Digital Output 1 Active Status DO1 STAT
NormOPEN, NormCLOSE *8
4 Digital Output 2 Active Status (*13) DO2 STAT 5 Measurement Mode EXIT
I
I DI
1 Digital Input (*13) DI FUNC
NO USE, C ST/SP, C RS/ST
RNG SW, ZERO ADJ, FIX OUT 2 Digital Input Detective Level (*13) DET LVL H LEVEL, L LEVEL *9 3 Measurement Mode EXIT
J
J CNT/PLS
1 Counting Rate CNT RATE
Units : m
3
, L, mL, bbl, gal, pt, qt Value : within the range 3.6 〜 36000000 pulse/h (0.001 〜 10,000 pps)
2 Plulse Mode PLS MODE AUTO, MANUAL
3 Pulse Width PLS WID
0.3 〜 500.0ms (0.1ms each) or Less than 40% of the pulse rate for 100% flow rate output
4 Measurement Mode EXIT
K
K PRESET C 1 Preset Count PRST VAL 0 〜 99999999 count (1 count each) 2 Preset Function OUT MODE HOLD, 50ms PLS, 500ms PLS *10 3 Measurement Mode EXIT
L
L H/L ALM1 1 High Alarm Set H SET ON, OFF 2 High Alarm Value H VAL -10 〜 110 % (0.1% each) 3 Low Alarm Set L SET ON, OFF 4 Low Alarm Value L VAL -10 〜 110 % (0.1% each) 5 Measurement Mode EXIT
M
M H/L ALM2
1 High High Alarm Set HH SET ON, OFF 2 High High Alarm Value HH VAL -10 〜 110 % (0.1% each) 3 Low Low Alarm Set LL SET ON, OFF 4 Low Low Alarm Value LL VAL -10 〜 110 % (0.1% each) 5 Measurement Mode EXIT
N
N SELF CHK 1 Empty Pipe Alarm EMPTY OFF, NORMAL, SENS, SENS-H *11 2 Self-diagnosis function SELF CHK ON, OFF 3 Alarm Output Preset CONV ALM CONV ONLY, WITH EMP 4 Measurement Mode EXIT
O
O FIX OUT 1 Fixed-value Output FIX SET ON, OFF *12 2 Fixed-current Output CUR VAL 2.4 〜 24.0 mA (0.1mA each) 3 Fixed-pulse Output PLS VAL 0 〜 10000 pps (1pps each) 4 Measurement Mode EXIT
P
P OTHERS 1 Password PASSWORD 000 〜 999 (1 each) 2 LCD Density Adjustment LCD ADJ 1(Light) 〜 5 (Dark) 3 Swich Position SW POSN TOP, BOTTOM, LEFT, RIGHT 4 Measurement Mode EXIT
Q
Q COMM 1 PROFIBUS Communication PROFIBUS *13 2 MODBUS Communication MODBUS *13 3 Measurement Mode EXIT
R
R CAL 1 0% Flow Rate Calibration CAL 0% 2 50% Flow Rate Calibration CAL 50% 3 100% Flow Rate Calibration CAL 100% 4 Checking the Excitation Current Value EXC DSPL 5 Measurement Mode EXIT
*1 This value is factory adjusted when shipped.
*2
In case of choosing COUNT or RANGE COUNT : displays totalized flow counts (8 digits) without a unit. RANGE : displays the range number (1 to 4)
*3 Only display2
*4
Range type Description
1 : SINGLE Single range
2 : 4F-0R Unidirectional flw , automatic selection of multiple ranges
3 : 2F-2R Bidirectional flows , automatic selestion of multiple ranges
4 : EXT 2F-0R Unidirectional flows , multiple ranges selected by exterminal single
5 : EXT 2F-2R Bidirectional flows , multiple ranges selected by exterminal single
*5
(1) In case of the zero adjustment (Menu No.F1),Zero offset is automatically cleared to Zero. (2) Calculate the zero offset value with the follwing equation: Zero offset value (%) = {(Actual flow rate) - (LF610 measured value)} (Sample)
Measured condition Flow rate % in measured span
Actual flow rate obtained from other instrument
10.0 m3/min 50%
LF610 measured value 10.5 m3/min 52.5%
Zero offset
-
-2.5%
If zero offset is set to -2.5% , LF610 will output 50.0% flow rate instead of 52.5%
*6
4-20mA Alarm Output Output Status
0 : UNDER 3mA Under 3mA output
1 : 4mA 4mA output
2 : HOLD Measured data hold
3 : OVER 24mA Over 24mA output
*7
DO1,DO2 items Digital output function
0 : NO USE Not used
1 : HIGH ALM High limit alarm output
2 : HH ALM High high limit alarm output
3 : LOW ALM Low limit alarm output
4 : LL ALM Low low limit alarm output
5 : RNG SIG1 Multi - range output No.1
6 : RNG SIG2 Multi - range output No.2
7 : PRESET C Preset point output
8 : CONV ALM Converter failure alarm output
9 : EMPTY ALM Empty pipe alarm output
10 : PLS OUT Pulse output (bidrectional flow)
11 : PLS FRD Pulse output (fixed forward flow)
12 : PLS REV Pulse output (fixed reverse flow)
*8
DO1/DO2 Active Status Alarm - output Status
0 : NormCLOSE Normal close
1 : NormOPEN Normal open
*9
DI Detective Level Detective Level
0 : L LEVEL L level
1 : H LEVEL H level
*10
Preset Function Output Status
0 : HOLD Output Status Level Hold
1 : 50ms PLS Pulse out (pulse width 50ms)
2 : 500ms PLS Pulse out (pulse width 500ms)
*11
Empty Pipe Alarm Digital output function
0 : OFF Not used
1 : NORMAL Used and Low-sensitive
2 : SENS Used and Middle-sensitive
2 : SENS-H Used and High-sensitive
*12
When this function is set to ON , the conditions are follwing.
Items Conditions
Current output User - set current output
Pulse output Pulse output with a user - set counting rate
Digital outputs Previous status is retained (excluding pulse output).
Data indicating Instaneous flow rates and flow velocity (no totalization).
*13 These functions are option.
Size
(mm)
Flow rate
0.3m/s 1m/s 3m/s 10m/s
2.5 0.005301 0.01767 0.05304 0.1767
4 0.01357 0.04524 0.13572 0.4524
6 0.03053 0.1018 0.3054 1.018
15 0.1908 0.6361 1.908 6.361
25 0.5301 1.767 5.301 17.67
32 0.8686 2.895 8.686 28.95
40 1.357 4.523 13.57 45.23
50 2.120 7.067 21.20 70.67
65 3.583 11.95 35.83 119.5
80 5.428 18.09 54.28 180.9
100 8.482 28.27 84.82 282.7
125 13.25 44.17 132.5 441.7
150 19.08 63.61 190.8 636.1
200 33.93 113.1 229.3 1,131
250 53.01 176.7 530.1 1,767
300 76.34 254.5 763.4 2,545
350 103.9 346.4 1,039 3,464
400 135.7 452.3 1,357 4,523
450 171.7 572.5 1,717 5,725
500 212.1 706.9 2,121 7,069
600 305.4 1,018 3,054 10,180
700 415.6 1,385 4,156 13,850
750 477.1 1,590 4,771 15,900
800 542.9 1,810 5,429 18,100
900 687.1 2,290 6,871 22,900
Factory default standard value table
Code Item
Default value
Changed value
English unit SI unit
A3 Exicitation Frequency Value Value
A4 Flow Direction NORMAL NORMAL
B1 Display 1 gal/min m
3
/h
B2 Display 2 COUNT B m
3
Display digit setting (For Display 1 and Display 2)
1/1000 1/1000
B3 Custom Coefficient 0 0
B4 Custom Unit Blank space in all field Blank space in all field
C1 Range Type SINGLE SINGLE
C2 Range 1 Value Value
C6 Hysteresis 3.0% 3.0%
D1 Damping Constant 5 sec 1 sec
D2 Rate-of-change Limit 0.0% 0.0%
D3 Control Limit Time 0.0s 0.0s
E1 Low Cutoff 1.0% 1.0%
E2 Display Low Cutoff LINEAR LINEAR
F2 Manual Zero 0.0% 0.0%
G1 4-20mADC Alarm Output 4mA 4mA
G2 Output Low Limit 4mA 4mA
H1 DO1 Function PLS OUT PLS OUT
H2 DO2 Function (*) EMPTY ALM NO USE
H3 DO1 Active Status NormOPEN NormOPEN
H4 DO2 Active Status (*) NormOPEN NormOPEN
I1 Digital input (*) C RS/ST C RS/ST
I2 DI Detective Level (*) H LEVEL H LEVEL
J1 Counting Rate Value Value
J2 Pulse width setting mode AUTO AUTO
J3 Preset width 5ms 100ms
K1 Preset count 00000000 00000000
K2 Preset Function HOLD HOLD
L1,L2 High Alarm Set / Value OFF 0.0% OFF 0.0%
L3,L4 Low Alarm Set / Value OFF 0.0% OFF 0.0%
M1,M2 High High Alarm Set / Value OFF 0.0% OFF 0.0%
M3,M4 Low Low Alarm Set / Value OFF 0.0% OFF 0.0%
N1 Empty Pipe Alarm NORMAL NORMAL
N2 Self-disgnosis function ON ON
N3 Alarm Output Preset CONV ONLY CONV ONLY
P1 Password 000 000
P2 LCD display adjustment 3 3
P3 Switch position setting BOTTOM BOTTOM
* These functions are option.
Meter Size
Ex.Freq
(Hz)
Range 1
Counting Rate
inch mm gal/min ft/s m
3
/h m/s
1/10'' 3 24 0.5 21.084 0.05 2.829 1 gal 1L
1/6'' 4 24 1.5 24.707 0.1 2.211 1 gal 1L
1/4'' 6 24 4 29.283 0.2 1.965 1 gal 1L
1/2'' 15 24 25 29.283 2 3.144 1 gal 0.01m
3
1'' 25 24 75 31.625 6 3.395 1 gal 0.01m
3
1 1/4'' 32 24 125 32.171 10 3.454 1 gal 0.01m
3
1 1/2'' 40 24 175 28.826 15 3.316 1 gal 0.01m
3
2'' 50 24 300 31.625 25 3.537 10 gal 0.1m
3
2 1/2'' 65 24 475 29.629 40 3.348 10 gal 0.1m
3
3'' 80 24 650 26.766 60 3.316 10 gal 0.1m
3
4'' 100 24 1,000 26.354 100 3.537 10 gal 0.1m
3
5'' 125 24 1,750 31.625 150 3.395 10 gal 0.1m
3
6'' 150 24 2,500 29.283 200 3.144 100 gal 1m
3
8'' 200 24 4,500 29.649 300 2.653 100 gal 1m
3
10'' 250 12 7,000 29.517 600 3.395 100 gal 1m
3
12'' 300 12 10,000 28.283 900 3.537 100 gal 1m
3
14'' 350 12 12,000 25.817 1,200 3.465 100 gal 1m
3
16'' 400 12 16,000 26.354 1,600 3.537 100 gal 1m
3
18'' 450 12 20,000 26.029 2,500 4.366 100 gal 1m
3
20'' 500 6 25,000 26.354 3,000 4.244 100 gal 1m
3
24'' 600 6 40,000 29.283 4,000 3.930 100 gal 1m
3
28'' 700 6 50,000 26.892 5,000 3.609 100 gal 1m
3
30'' 750 6 60,000 28.112 6,000 3.773 100 gal 1m
3
32'' 800 6 70,000 28.825 7,000 3.868 100 gal 1m
3
36'' 900 6 80,000 26.029 8,000 3.930 100 gal 1m
3
ERROR/ALARM
Error/Alarm type Indicating message Error contents
Self-diagnosis error ROM ERROR ROM error
RAM ERROR RAM error
PARAMETER FAILURE System parameter error
EXC CUR OPEN Excitation circuit error
EXC CUR ERROR Excitation current error
ADC ERROR ADC error
INVALID TOTAL Invaid totalizer counts
Settig error HIGH OVER SPEC Setting value exceeds the allowable high limit.
LOW OVER SPEC Setting value exceeds the allowable low limit.
HIGH OVER CNT RATE Counting rate exceeds the allowable high limit.
LOW OVER CNT RATE Counting rate exceeds the allowable low limit.
MULTI RNG ERROR Span is not appropriate for multi-range configuration.
Limit alarms HIGH ALARM Flow rate reading exceeds the high limit.
HIGH HIGH ALARM Flow rate reading exceeds the high high limit.
LOW ALARM Flow rate reading goes below the low limit.
LOW LOW ALARM Flow rate reading goes below the low low limit.
OVER 125% Measurement value exceeds the 125%
UNDER -125% Measurement value goes below the 125%
Empty pipe alarm EMPTY ALARM Detector pipe is not filled with fluid.
LCD
Meter Size vs. Velocity/Rate
Size
(inch)
Flow rate
0.98ft/s 3ft/s 10ft/s 32.8ft/s
1/10'' 0.02334 0.07115 0.2372 0.7781
1/6'' 0.05975 0.1821 0.6071 1.992
1/4'' 0.1344 0.4098 1.366 4.482
1/2'' 0.8403 2.561 8.532 28.01
1'' 2.334 7.115 23.72 77.81
1 1/4'' 3.824 11.66 38.86 127.5
1 1/2'' 5.975 18.21 60.71 199.2
2'' 9.337 28.46 94.86 311.2
2 1/2'' 15.78 48.09 160.3 526.0
3'' 23.90 72.85 242.8 796.7
4'' 37.35 113.8 379.4 1,245
5'' 58.35 177.9 592.9 1,945
6'' 84.03 256.1 853.8 2,801
8'' 149.4 455.3 1,518 4,980
10'' 233.4 711.5 2,372 7,781
12'' 336.1 1,025 3,415 11,200
14'' 457.5 1,394 4,648 15,250
16'' 597.5 1,821 6,071 19,920
18'' 756.3 2,305 7,684 25,210
20'' 933.7 2,846 9,486 31,120
24'' 1,344 4,098 13,660 44,820
28'' 1,830 5,578 18,590 61,000
30'' 2,101 6,403 21,340 70,020
32'' 2,390 7,285 24,280 79,670
36'' 3,025 9,221 30,740 100,800
unit : gal/min unit : m3/h