Electromagnetic Flowmeter LF600 Series
Operation Guide
ISetting Parameters Flow!
ISetting Menu!
0 Measurement Mode MEAS MODE
1 Exicitation Current EX CURRENT O.OOOOA ~ 0.2400A (0.0001A each)
2 Meter size METER SIZE 0.1 24 inch. 2.5mm 600mm
A
3
Exicitation Freouencv EX FREQ 6.12.24 Hz
4 Flow Direction FLOWDIRCTN NORMAL. SWITCH
Password PASSWORD 000 ~ 999 (leach)
5
6 Communication Address ADDR SET{*2)
0 Measurement Mode MEAS MODE
Main Display
1
В
Sub Display SUB DSP
2
3 Custom Coefficient CUSTOM DATA 0 99999999
4 Custom Unit CUSTOM UNIT
5 LCD Density Adiustment LCD ADJUST 1(LlEht)~5(DariO
6 Swich Position SW POSITION
0 Measurement Mode MEAS MODE
1 Ranse Tvpe RANGE TYPE SINGLE. 4F - OR. 2F - 2R. EXT 2F - OR. EXT 2F - 2R
Range 1
2
3 Range 2
C
Range 3 RANGES
4
5 Range 4 RANGE4
6 Ranee Hysteresis
0 Measurement Mode MEAS MODE
1
Dampine Constant
2 Low Cutoff LOW CUT 0.0 ~ 10.0 «(О.« each)
D
3 4-20mA Alam Cutout ALM mA SET UNDER 3.0mA, 4.0mA. HOLD, OVER 24.0mA
4 Display Low Cutoff DSP LOW CUT ON. OFF
Output Low Limit LOW LIMIT
5
0 Measurement Mode MEAS MODE
E
1 Zero Adiustment ZERO ADJUST
0 Measurement Mode MEAS MODE
1 D01 Function D01 FUNCTN
2 D02 Function D02FUNCTN(*2)
F
3 Dl Function Dl FUNCTN (*2)
4 D01 Active Status D01 ALM STS
D02 Active Status D02ALMSTS(»2)
5
6 Dl Detective Level
0 Measurement Mode MEAS MODE
1 Counting Rate COUNT RATE
G
Pulse Mode
2
3 Pulse Width
0 Measurement Mode MEAS MODE
H
Preset Count
1
2 Preset Function
0 Measurement Mode MEAS MODE
1 High Alarm Set HALM SET ON. OFF
1
2 High Alarm Value H ALM VAL -10 ~ 110)1
3 Low Alarm Set L ALM SET ON. OFF
4 Low Alarm Value L ALM VAL
0 Measurement Mode MEAS MODE
1 High High Alarm Set HH ALM SET ON. OFF
J
2 High High Alarm Value HH ALM VAL
3
Low Low Alarm Set LL ALM SET ON. OFF
4 Low Low Alarm Value LL ALM VAL
0 Measurement Mode MEAS MODE
1 Empty Pipe Alarm EMPTY ALM OFF. NORMAL. SENSITIVE. SENSITIVE-H
К
2 SelNIagnosIs function SELF CHECK ON. OFF
3 Alarm Output Preset
0 Measurement Mode MEAS MODE
L
1 Rate-of-change Limit LIMIT RATE 0.0 "30.0)1(0.1)1 each)
2 Control Limit Time LIMIT TIME
0 Measurement Mode MEAS MODE
1 Fixed-value Output FIXED OUT ON. OFF
м
Fixed-current Output FIOED OURR
2
3 Fixed-pulse Output FIXED PULSE
0 Measurement Mode MEAS MODE
N
1 Manual Zero MANUAL ZERO
0 Measurement Mode MEAS MODE
1 0% Flow Rate Calibration
0
2 50% Flow Rate Calibration
3 100% Flow Rate Calibration FLOW CAL 100 fENTl switch (Keep pushing more than 2 seconds.)
Checking the Excitation Current Value EXCURRDSP
4
___
: Entering value
___
: Selecting value
I
: Entering and selecting value
Items Screen Display Setting or selecting value
000 126 (1 each)
■First Unit : Volume, Others
MAIN DSP
RANGE1
RANGE2
RANGE HYS
DAMPING 00.0.00.5.01.0 ~ 60.0 6
Dl DETLV{*2)
PLS MODE AUTO. MANUAL
PLS WIDTH
PRESET ONT
PRESET FNO HOLD. 50ms PULSE. 500ms PULSE
ALM PRESET
FLOW CALO
FLOW CAL 50
%, m3,1, ml, bbl, gal, m/s, ft/s, COUNT, RANGE, pt, qt,
CUSTOM, GRAPH И)
■Second Unit ;Time unit
■Third unit :Flow direction code
В (Automatic selection bi-directionflow)
F (Fixed forward flow), R (Fixed reverse flow)
D (Différence)
7 Characters
TOP, ВОПОМ. LEFT, RIGHT
■Volume unit
m3,1, ml, gal, bbl, pt, qt
■Time unit
/s, /min, /h, /d
■Velocity unit
m/s, ft/s
0.0 ~ 25.0)1(0.1)1 each)
4.0mA, 3.2mA, 2.4mA
Hold down the [SET] kev
NO USE, HIGH ALM, HH ALM, LOW ALM, LL ALM
EMPTY ALM. PULSE OUT. PULSE FRD. PULSE REV
NO USE. ONT STA/STP, CNT RES/STA
RANGE SW. ZERO ADJ. FIXED OUT
H LEVEL, L LEVEL
Units : m3,1, ml, bbl, gal, pt, qt
Value : within the range 3.6 ~ 36000000 pulse/h
0.3 500.0ms (0.1ms each)
or Less than half of the pulse rate for 100% flow rate output
0 99999999 count (1 count each)
-10 ~ 110)1
-10-110%
-10-110%
WITHOUT EMP. WITH EMP
00 — 20 s (1 s each)
2.4" 24.0 mA (0.1mA each)
0 — 10000 pps (Ipps each)
±10% of lOm/s-maximum range (0.1% each)
[ENTl switch (Keep pushing more than 2 seconds^
None (Displav)
None (Display)
(0.001 - 10,000 pps)
This value is factory adjusted when shipped.
These functions are option.
case of choosing COUNT or RANGE
COUN : displays totalized flow counts (8 digits) without a unit
RANGE: displays the range number (1 to 4)
*4
Only sub displey
*5
1: SINGLE
2:4F-OR
3:2F-2R
4: EXT. 2F5: EXT. 2F-
Single range
Unidirectional flw, automatic selection of multiple ranges
Bidirectional flows, automatic selestion of multiple ranges
Unidirectional flows, multiple ranges selected by exterminai single
Bidirectional flows, multiple ranges selected by exterminai single
_ _ _ _ _ _ __ _ _ _ _ _
Description
4-20mA Alarm Output Output Status
0: UNDER 3.0mA
1:4.0mA
2: HOLD
3: OVER 24.0mA
*7
D01.D02 Items Digital output function
0:NO USE
1: HIGH ALM
Under 3mA output
4mA output
Measured data hold
Over 24mA output
Not used
High limit alarm output
2:HHALM High high limit alarm output
3:L0WALM
4:LLALM
5:RNG SIG1
6:RNG SIG2
7:PRESET
Low limit alarm output
Low low limit alarm output
Multi - range output No.1
Multi - range output No.2
Preset point output
8:CONVALM Converter failure alarm output
9:EMPTYALM
10:PULSE OUT
11:PULSEFRD
12:PULSE REV
Empty pipe alarm output
Pulse output (bldrectlonal flow)
Pulse output (fixed forward flow)
Pulse output (fixed reverse flow)
D01/D02 Active Status Alarm - output Status
0:NORMAL CLOSE
1: NORMAL OPEN
*9
Dl Detective Level Detective Level
0:L LEVEL
1:H LEVEL
*10
Preset Function Output Status
0:HOLD
♦10
1:50ms PULSE
2:500ms PULSE
*11
Empty Pipe Alarm Setting and Degective Level
0:OFF
1: NORMAL
2:SENSITIVE
2:SENSITIVE-H
*12
When this function is set to ON, the conditions are foilwing.
♦11
Items Conditions
Current output
Pulse output
Digital outputs
Data Indicating
♦12
*13
(1) In case of the zero adjustment (Menu No.EI),Zero offset is automatically cleared to Zero.
Normal close
Normal open
L level
H level
Output Status Level Hold
Pulse out (pulse width 50ms)
Pulse out (pulse width 500ms)
Not used
Used and Ditective level low
Used and Ditective level middle
Used and Ditective level high
User - set current output
Pulse output with a user - set counting rate
Previous status Is retained (excluding pulse output).
Instaneous flow rates and flow velocity (no totalization).
(2) Calculate the zero offset value with the foilwing equation:
Zero offset value (U) = ((Actual flow rate) - (LF600 measured value))
♦13
Measured condition Flow rate 91 In measured span
Actual flow rate obtained
from other instrument
10.0 m3/min
5091
LF600 measured value 10.5 m3/min 52.591
Zero offset
-
-2.591
If zero offset is set to -2.5X, LF600 will output 50.091 flow rate instead of 52.591
LF600seriesGuide070207cs.indd 1-2 07.2.7 11:06:05 AM
^Factory default standard value table(
► Meter Size vs. Velocity/Rate(
Code
A3
Exicitation Freouencv
A4 Flow Direction NORMAL NORMAL
Password 000 000
A5
B1
Main Displav
B2
Sub Displav
B3
Custom Data
B4 Custom Unit Blank Blank
LCD Density Adjustment
B5
B6
Swich Position BOTTOM BOnOM
01 Range Type SINGLE SINGLE
Range 1 Value Value (*2)
C2
D1 Damping Constant 1 sec 5 sec
Low Cutoff
D2
D3 4-20mA Alam Output 4mA 4mA
D4 Displav Low Cutoff OFF OFF
Output Low Limit
D5
FI D01 Function PULSE OUT PULSE OUT
F2 D02 Funotlon NO USE EMPTY (*4)
F3 Dl Function CNT RES/STR CNT RES/STR (*4)
D01 Active Status
F4
D02 Active Status NORMAL OPEN NORMAL OPEN
F5
F6 Dl Detective Level H Level H Level
G1 Counting Rate Value (*3) Value (*3)
Pulse Mode
G2
HI Preset Count 0 0
H2 Preset Function
11,12 High Alarm Set/Value OFF 0.0% OFF 0.0%
13.14 Low Alarm Set/Value OFF 0.0% OFF 0.0%
J1.J2 High High Alarm Set/Value OFF 0.0% OFF 0.0%
J3.J4 Low Low Alarni Set/Value OFF 0.0% OFF 0.0%
K1 Empty Pipe Alarm NORMAL NORMAL
Self-diagnosis function
K2
КЗ Alarm Output Preset
LI Rate-of-change Limit 0.0% 0.0%
Control Limit Time 0 sec 0 sec
L2
*1, *1, ’l‘3 : Setting value in each size.
*4 : These function are option.
•ERROR/ALARM#
Error/Alarm type Indicating message Error contents
Self-diagnosis error
Settig error
Limit alarms
Empty pipe alarm
Item
NORMAL OPEN NORMAL OPEN
WITHOUT EMP WITHOUT EMP
*ROM ERROR*
*RAM ERROR*
PARAMETER FAILURE System parameter error
EX CURRENT OPEN Excitation circuit error
EX CURRENT ERROR Excitation currenr error
ADC ERROR
INVALID TOTAL
*HIGH OVER SPEC.* Setting value exceeds the allowable high limit.
*LOW OVER SPEC.* Setting value exceeds the allowable low limit.
*HIGH OVER CNT RATE*
*LOW OVER CNT RATE*
*MULTI RANGE ERROR*
HIGH ALARM
HIGH HIGH ALARM
LOW ALARM
LOW LOW ALARM Flow rate reading goes below the low low limit.
OVER 125%
UNDER -125%
EMPTY ALARM
Default value
SI unit
Value (*1) Value (*1)
mVh gal/min
3
m COUNT B
0 0
3 3
1% 1%
4mA 4mA
AUTO AUTO
HOLD HOLD
ON ON
English unit
Changed value
Meter Size
mm
2,5
4
6
15
25
*1 Ex.Fre(|
inch
1/10”
1/6”
(Hz)
mVh
24
24
1/4” 24 0.2
1/2” 24 2 3,144
1” 24
32 11/4” 24
40
11/2” 24
50
2” 24
21/2” 24
65
80 3”
100
125 5”
150 6”
200 8”
250 10”
300
12” 12
350
14” 12
400 16”
450 18”
24
4” 24
24
24
24
12
12
12
1,200
1,600
2,500 4,366 20,000 26.029
500 20” 6 3,000
600
24”
ROM error
RAM error
ADC error
Invaid totalizer counts
Counting rate exceeds the allowable high limit.
Counting rate exceeds the allowable low limit.
Span is not appropriate for multi-range configuration.
Flow rate reading exceeds the high limit.
Flow rate reading exceeds the high high limit.
Flow rate reading goes below the low limit.
Measurement value exceeds the 125%
Measurement value goes below the 125%
Detector pipe is not filled with fluid.
6 4,000 3.930 40,000 29.283
*2 Range 1
m/s
gal/min
2.829
0.05
0.1 2,210
1.965
6
3.395 75 31,625
10
3,454
3,316
15
3,537
25
40 3.348
60 3.316 650 26,766
100 3,537 1,000
150
3.395
200
3,144
ft/s
21,084 1L
0.5
24,707 1L
1.5
29.2831LIgal
4
29.283 0,01m^ Igal
25
0,01m^
0,01m^
32,171
125
28.826 0,01m^
175
300
475
0,1m*
31,625
29.629
0,1m*
0,1m*
26,354 0,1m*
1,750
0,1m*
31,625
2,500 29.283
300 2,653 4,500 29.649
600
900
7,000
3.395
10,000 28.283
3,537
12,000
3,465
16,000
3,537
25,000
4,244
29,517 1m*
25,817 1m*
26,354 1m*
26,354 1m*
♦3 Counting
Rate
Igal
Igal
Igal
Igal
Igal
10 gal
10 gal
10 gal
10 gal
10 gal
100 gal
1m*
100 gal
1m*
100 gal
100 gal
1m*
100 gal
100 gal
100 gal
1m*
100 gal
100 gal
1m*
unit: gal/min unit:m Vh
Size
(inch)
1/10”
1/6”
1/4” 0.1344 0.4482 4.482
1/2” 0.8401 2.801 28.01
1” 2.334
1 1/4” 3.824
1 1/2”
2” 9.334 31.12 311.2
2 1/2” 15.78 52.61 526.1
3” 23.90
4”
5” 58.34 194.5 1,945
6” 84.01 280.1 2,801
8” 149.4 498.0 4,980
10”
12” 336.1 1,121
14” 457.5 1,525 15,252
16”
18” 756.0 2,521 25,206
20” 933.9
24”
# Accuracy#
0.98ft/s 3.28ft/s 32.8ft/s
0.02334
0.05975
5.975
37.35 124.5 1,245
233.4
597.5
1,345
Flow rate
0.0778 0.7780
0.1992 1.992
7.780 77.80
12.75 127.5
19.91 199.1
79.65 796.5
778.0 7,780
1,991
3,112 31,124
4,482 44,821
11,205
19,914
Micro meter size (Diameter from 2.5 to 6mm)
Flow rate as a
Flow speed of full scale
percentage of From 0.3 to 1.0m/s From 1.0 to lOm/s
range (0.98 to 3.28 ft/s) (3.28 to 32.8 ft/s)
From Oto 50%FS
From 50 to 100%FS
Small/Middle meter
size (Diameter from 15mm to 450mm)
±0.8% FS ±0.4% FS
±0.8% FS ±0.8% of rate
Flow rate as a Flow speed of full scale
percentage of From 0.1 to 1.0m/s From 1.0 to lOm/s
range (0.328 to 3.28 ft/s) (3.28 to 32.8 ft/s)
From Oto 20%FS
From 20 to 509ÍFS
From 50 to 100%FS
±0.25% FS
±0.5% of rate
±0.1% FS
±0.5% or rate
Large meter size (Diameter 500 mm or longer)
Flow rate as a
Flow speed of full scale
percentage of From 0.3 to 1.0m/s From 1.0 to lOm/s
range (0.98 to 3.28 ft/s) (3.28 to 32.8 ft/s)
From Oto 1005ÍFS
±0.8% FS ±0.5% FS
Size
(mm)
2.5
4 0.01357 0.04524 0.4524
6 0.03053 0.1018 1.018
15
25
32
40 1.357 4.523 45.23
50 2.120
65
80 5.428 18.09 180.9
100 8.482 28.27 282.7
125 13.25 44.17 441.7
150 19.08 63.61 636.1
200 33.93 113.1 1131
250 53.01
300 76.34
350 103.9
400
450 171.7
500
600
►LCD(
Signal cable
0.3 m/s 1.0 m/s 10 m/s
0.005301
0.1908 0.6361 6.361
0.5301
0.8686
3.584
135.7
212.1
305.4
© TOSHIBA Corporation 2007
All Rights Reserved.
Feb., 2007 Rev.1
Flow rate
0.01767 0.1767
1.767 17.67
2.895 28.95
7.067 70.67
11.95 119.5
176.7 1767
254.5 2545
346.4 3464
452.3 4523
572.5 5725
706.9 7069
1018 10180
LCD display
LF600seriesGuide070207cs.indd 3-4 07.2.7 11:06:07 AM