Toshiba LF410, LF600, LF602 User Manual

Page 1
Field Intelligent Device Series
Electromagnetic Flowmeter
Introduction
The electromagnetic flowmeter uses Faraday’s Law of electromagnetic induction to measure the process flow. The device consists of two units: a detector, through which the fluid to be measured flows and in which low-level signals proportional to flow rates are obtained; and a converter, which supplies excitation current to the detector, and amplifies the signals from the detector and then processes and converts the signals into the 4–20mAdc current signal or communication signal. Thanks to the unique patented magnetic field distribution technology, the meter is highly immune for upstream flow disturbances. Combined with multi-functional converter LF600 (combined type) or LF602 (separate type) equipped with its original noise-suppression circuit and arithmetic operation capability, LF410 has high tolerance to noise, giving stable output even for slurry fluid measurement. IR (Infrared) switches enable parameter setting of the converter without removing the cover. Flow direction can be set in either way, and its 128 x 128 dot matrix LCD display allows the LCD rotated to 90, 180 and 270 degrees with a software.
The AF900 hand-held terminal (HART* communicator) can be used to communicate with the flowmeter from a remote place. PROFIBUS-PA* interface is available as option.
*1: HART protocol (Highway Addressable Remote
Transducer) is a communication protocol for industrial sensors recommended by the HCF Communication Foundation).
*2: PROFIBUS is the communication protocol for the
factory automation and process automation that PROFIBUS Organization recommends. Instead of analog control with a conventional analog signal (4-20mA), it is one kind of the fieldbus which digitized all signals. Flowmeters support PROFIBUS-PA.
I/O
Power supply
Detector
Combined type
LF410/LF600 LF414/LF600
Converter
Converter
Signal cable
Terminal box
Figure1. Configuration
1
(HART
Detector
Separate type LF410/LF602 LF414/LF602
2
Power supply
I/O
LF410/LF600 LF410 LF602 LF414/LF600 LF414
Figure2. LF410 series Flowmeters
*3
Specifications
Overall Specifications
Measurement range in terms of flow velocity:
0.3 m/s to 0 –10 m/s (0 –1.0 ft/s to 0 32.8 ft/s).
0 0 0.1 m/s to 0 0.3 m/s (0 0.3 ft/s to 0 1.0 ft/s) range is available optionally.
Accuracy: See the following graph.
Pulse output:
Vs > 0.5 m/s (1.64 ft/s): +/ Vs < 0.5 m/s (1.64 ft/s): +/
-
+/
1 mm/s (0.039 inch/s). Current output: plus +/ span)
Note: Span = Range in the magmeters.
Error
(% of rate)
5.0
4.0
3.0
2.0
1.0
0.0
0 1 2 10
0123
Note: The accuracy above is measured under standard
operating conditions using the weighing method at Toshiba's flow calibration facility.
*3: CE mark is under application.
LF410 /LF600 LF410 /LF602
15 to 200 mm (1/2" to 8")
Certification number
Z01207
-
0.5 % of rate.
-
0.3% of rate
-
8 uA (0.05 % of
Absolute figure
Flow Ve losity (m /s)
Accuracy
No. EJL-114B
Page 2
LF410/LF600 LF410/LF602
Fluid conductivity: 5µS/cm minimum
Fluid temperature:
–10 to +180°C : Ceramic type (14 to 356 °F)
Note : 120°C (248°F) above is separate type
–10 to +120°C : tefolon PFA
(14 to 248°F)
Ambient temperature:
Without cFMus Approval:
-
20 to + 60 °C (-4 to 140 °F)
With cFMus Approval:
-10 to +45°C (14 to 113 °F)
Structure: NEMA 4X (IP 67) Watertight
Power consumption:
17W(27VA) or less
19W(29VA) or less (with PROFIBUS)
Conformance to European Community Directives:
EMC directive 89/336/EEC The low voltage 93/68/EEC PED 97/23/EC (Note 1)
Note : See table 2 for detail.
Approved hazardous location certifications:
Model: LF414/LF600 and LF414/LF602
cFMus explosion proof: FM Class I, Division 2, Groups A,B,C, and D. FM Class II, Division 2, Groups E, F and G. FM Class III.
Detector and converter combination:
LF410/LF600: Combined type for standard
specification.
LF410/LF602: Separate type for standard
specification.
LF414/LF600: Combined type with Ex approval
of Class I, Division 2 (cFMus).
LF414/LF602: Separate type with Ex approval
of Class I, Division 2 (cFMus).
Model LF410 and LF414 Detectors
Mounting style: Wafer type
Fluid pressure:
-
0.1 to 2.0MPa
(-15 to 300psi, or -1.0 to 20 bar)
Note: The test pressure before shipping from the factory
is equal to twice the nominal pressure rating of the customer specified flange connection during 15 minutes.
Connection flange standards: ANSI 150, ANSI
300, BS10 and 16, DIN PN10 and PN16, JIS10K, JIS16K and JIS20K
Principal materials:
Case — 25 to 100mm (1” to 4”): stainless steel 15,
150, and 200mm (1/2”, 6”, and 8”): carbon steel
Linings — 15 to 100mm (1/2” to 4”): Ceramic
tube (std.) & teflon PFA (opt.)
150 and 200mm (6” to 8”): teflon PFA Electrodes — 316L stainless steel (std.) Grounding rings — 316 stainless steel (std.)
Note: See Table 3 for optional materials and other
related information.
Measuring tube material — 304 stainless steel
(in case of Teflon PFA lining)
Coating — 25 to 100mm (1” to 4”): no coating
(stainless steel body). 15, 150, and 200mm (1/2”, 6”, and 8”): phthalic acid resin coating with pearl-gray colored.
Heat shock resistance : for a ceramic tube detector
Heating: ΔT 150
Cooling: ΔT 100
Note: The above means that the ceramic tube detector
withstands the shock of sudden heating (temperature difference 150°C or less per
0.5seconds) and sudden cooling (temperature difference 100°C or less per 0.5seconds).
°C/0.5sec (302°F/0.5sec)
°C/0.5sec (212°F/0.5sec)
Dimensions and weights: See Figures 3 to 8.
Cable connection port: for separate type detectors.
Cable glands
LF410: without cFMus approval
Provided as standard R(PT) 1/2 male screws.
LF414: with cFMus approval
Not provided 3/4-14NPT male screws are required.
Applicable diameter — 11 to 13mm
(0.433 to 0.512 inch)
Model LF600 and LF602 converters
Input signals
Analog signal — the voltage signal from detector,
proportional to process flow rate (for LF602 separate type converter).
Digital input DI (opt.)
Signal type: 20 to 30Vdc voltage signal Input resistance: 2.7kΩ Number of inputs: one point
DI function — One of the following functions can
be assigned to the optional DI signal.
Range switching — Selects either the higher or
lower range in the unidirectional or bidirectional 2-range setting.
Totalizer control — Starts and stops the built-in
2
Page 3
LF410/LF600 LF410/LF602
totalizer.
Fixed-value outputs — Outputs fixed-values for
current and pulse outputs.
Zero adjustment — Executes zero adjustment
(on-stream at zero flow rate).
Output signals
Current output:
4–20mAdc (load resistance 0 to 750Ω)
Note: The current output cannot be used with
the PROFIBUS-PA ccommunication.
Digital outputs — One point (std.) and one more point is optionally available as follows.
Digital output DO1 (std.):
Output type: Transistor open collector Number of outputs: One point Output capacity: 30Vdc, 200mA maximum
Digital output DO2 (opt.):
Output type: Solidstate relay output (non polarity) Number of outputs: One point Output capacity: 150Vdc, 150mA maximum or 150 V ac (peak to peak), 100mA maximum
DO1 and DO2 functions — One of the following
functions can be assigned to DO1 (std.) and/or DO2 (opt.)
Pulse output (available only for DO1,DO2) Pulse rate: 3.6 to 36,000,000 pulses/hr (DO1)
3.6 to 360,000 pulses/hr (DO2)
(Over 3,600,000 pulses/hr, auto-setting) Pulse width: 0.5 to 500ms (but less than half of the period for 100% flow rate)
Note: The same and simultaneous pulse is not
available between DO1 and DO2.)
Multi-range selection outputs (Note 1)
High, High high, Low, and/or Low low alarm
outputs (Note 2)
Empty pipe alarm output
Digital Output Active Status (DO1 and DO2)
(Note 2)
Preset count output
Converter failure alarm output
Note 1: Two outputs (DO1 and DO2) are needed for
4-range switching and forward/reverse 2-range switching.
Note 2: Normal Open (default set) or Normal Close
is selected for alarm outputs when programming. The Status when power failure is kept to Normal Open.
Communications output
HART(std.) Digital signal is superimposed on 4–20mAdc
current signal as follows:
Conforms to HART protocol Load resistance: 240 to 750Ω
Load capacitance: 0.25µF maximum Load inductance: 4mH maximum
PROFIBUS (opt.) Protocol : PROFIBUS-PA
Baurate : 31.25kbps Bus voltage : 9-30Vdc
Consumption electric current of bus:less than 16mA
Manufacture Ident-No. : 093B Standard Ident-No. : 9740
HEX
HEX
Slave address : 0-126 (Default address is 126) Profile : Profile Ver.3.01 for Process Control Devices Function blocks : AI(Flow)×1 , Totalizer×1
LCD display: Full dot-matrix 128×128 dot LCD
display (back-light provided) The data on the LCD inside the converter can rotate to 90, 180, and 270 degrees by a software, without rotating the indicator itself. (Combined type only)
Parameter settings — Parameters can be set as
follows:
• IR Switches: Three key switches are provided to
set configuration parameters.
• Digital communication: The AF900 hand-held terminal or PROFIBUS is needed to set parameters.
•Zero adjustment: Zero point adjustment can be started by pressing the switch in the converter.
Damping: 0.5 to 60 seconds (selectable in 1 second
increments)
Zero and span calibration: Built-in calibration
signal source allows converter unit check.
Conditions when power fails: The outputs and
display will remain as follows when power fails. Parameter setting values are stored in non-volatile memory and the values will be restored when the power returns to normal condition.
• Current output: 0mAdc
• Digital output: OFF
• LCD display: No display
• PROFIBUS: No communication
Power supply: One of the following can be selected:
• 100 to 240Vac, 50/60Hz (std.) (allowable voltage 80 to 264Vac)
• 24Vdc (allowable voltage 18 to 36Vdc)
• 110Vdc (allowable voltage 90 to 130Vdc)
Surge protection: Arresters are installed in the
power supply, and current signal output circuit.
Case: Aluminum alloy (equivalent of IP 67)
Coating: Acrylic resin-baked coating, pearl-gray
colored
3
Page 4
LF410/LF600 LF410/LF602
Cable connection ports:
Cable glands
LF600 and LF602 without cFMus Approval:
Provided as standard OD of cableφ11~13mm Material Nylon 66 G (PF) 1/2 male screws.
Note: When PROFIBUS option is specified ,
cable grands size is φ6~8mm for signal cable, φ11~13mm for power cable.
LF600 and LF602 with cFMus Approval:
Not provided, 1/2-14NPT male screws are required.
Applicable diameter 11 to 13mm
(0.433 to 0.512 inch)
Vibration resistance:
No resonance to the following levels of vibration:
• 10 to 150Hz with acceleration of 9.8m/s No defect in putting vibration to each
direction of 30Hz with 29.4 m/s2in 4h.
Note: Avoid using the flowmeter in an
environment with constant vibration.
Converter LF602 dimensions and weights:
See Figure 9 (for Separate type)
MTBF:220,000 hours at 25 deg.C (77 deg.F) based
on MIL-HDBK-217F
2
4
Page 5
LF410/LF600 LF410/LF602
Installation
Dimensions
Unit : mm
225
131
(35)
(L2)
143
ΦD1
64
L1
BS16, DIN PN16 and JIS 10K dimensions:
Meter size
(mm)
15 70 252 49 approx. 5
Note: 1 inch = 25.4mm
L1
(mm)
L2
(mm)
D1
(mm)
Weight
(kg)
ANSI class 150 and class 300 dimensions:
Meter size
(inch)
1/2 2.76 9.92 1.93 approx. 11
L1
(inch)
L2
(inch)
D1
(inch)
Weight
(lb)
Figure 3. LF410/LF600 and LF414/LF600
flowmeters
Meter size 15mm (1/2")
The dimension of L1 is changed when the material of grounding ring is chosen Pt-Ir or Ta.
Meter size L1
15mm (1/2”) 77mm (3.03 inch)
25mm (1”) 95mm (3.74 inch)
40mm (1 1/2”) 115mm (4.53 inch)
50mm (2”) 126mm (4.96 inch)
80mm (3”) 126mm (4.96 inch) 100mm (4”) 136mm (5.35 inch) 150mm (6”) 242mm (9.53 inch) 200mm (8”) 312mm (12.28 inch)
Unit : mm
BS16, DIN PN16 and JIS 10K dimensions:
Meter size
(mm)
25 80 241 66 approx. 5
Note: 1 inch = 25.4mm
ANSI class 150 and class 300 dimensions:
Meter size
(inch)
1 3.15 9.48 2.60 approx. 11
Figure 4. LF410/LF600 and LF414/LF600
225
131
(L2)
87
L1
(mm)
L1
(inch)
L2
(mm)
L2
(inch)
flowmeters
Meter size 25mm (1")
ΦD1
D1
(mm)
D1
(inch)
133(35)
L1
Weight
(kg)
Weight
(lb)
5
Page 6
LF410/LF600 LF410/LF602
Unit : mm
225
131
(35)
133
(L2)
The dimension of L1 is changed when the material of grounding ring is chosen Pt-Ir or Ta
Meter size L1
15mm (1/2”) 77mm (3.03 inch)
25mm (1”) 95mm (3.74 inch)
40mm (1 1/2”) 115mm (4.53 inch)
50mm (2”) 126mm (4.96 inch)
80mm (3”) 126mm (4.96 inch) 100mm (4”) 136mm (5.35 inch) 150mm (6”) 242mm (9.53 inch) 200mm (8”) 312mm (12.28 inch)
ΦD1
R (PT)1/2Male Screw (Two Places)
88
L1
40
126
(35)
Unit : mm
91
Note1: Eye bolts are provided at the top for flowmeters
sized 200mm (8").
(L2)
BS16, DIN PN16 and JIS 10K dimensions
Meter size
(mm)
40 100 264 85 approx. 6 50 110 280 102 approx. 7
80 110* 306 127 approx. 8 100 120* 338 159 approx. 10 150 230 407 216 approx. 22 200 300 458 267 approx. 32
ANSI class 150 and class 300 dimensions
Meter size
(inch)
1-1/2 3.94 10.39 3.35 approx. 13
2 4.33 11.02 4.02 approx. 15 3 4.33* 12.04 5.00 approx. 18 4 4.72* 13.30 6.26 approx. 22 6 9.06 16.02 8.50 approx. 49 8 11.81 18.03 10.51 approx. 70
Note: 1 inch = 25.4mm
*When Teflon PFA lining is selected, L1 in the table
above becomes as follows:
L1: 115mm (4.53 inch) for meter size 80mm (3")
L1: 126mm (4.96 inch) for meter size 100mm (4")
Figure 5. LF410/LF600 and LF414/LF600
Meter sizes 40 (1 1/2") to 200mm (8")
L1
(mm)
L1
(inch)
L2
(mm)
L2
(inch)
flowmeters
D1
(mm)
D1
(inch)
Weight
(kg)
Weight
(lbs)
Φ49
64 L1
Note1: Cable glands are not provided for LF414 of FM
and CSA approved type. Refer to the part of Cable connection port at detector.
Note2: 1 inch = 25.4mm
BS16, DIN PN16 and JIS 10K dimensions:
Meter size
(mm)
15 70 179 49 approx. 3
L1
(mm)
L2
(mm)
D1
(mm)
ANSI class 150 and class 300 dimensions:
Meter size
(inch)
1/2 2.76 7.05 1.93 approx. 7
L1
(inch)
L2
(inch)
D1
(inch)
Figure 6. Separate type detectors
LF410 and LF414
Meter size 15mm (1/2")
The dimension of L1 is changed when the material of grounding ring is chosen Pt-Ir or Ta
Meter size L1
15mm (1/2”) 77mm (3.03 inch)
25mm (1”) 95mm (3.74 inch)
40mm (1 1/2”) 115mm (4.53 inch)
50mm (2”) 126mm (4.96 inch)
80mm (3”) 126mm (4.96 inch) 100mm (4”) 136mm (5.35 inch) 150mm (6”) 242mm (9.53 inch) 200mm (8”) 312mm (12.28 inch)
Weight
(kg)
Weight
(lb)
6
Page 7
LF410/LF600 LF410/LF602
Unit : mm
R(PT)½ Male Screw (Two places)
88 40
87
(35) 91
(L2)
Φ66
(126)
L1
Note1: Cable glands are not provided for LF414 of FM
and CSA approved type. Refer to the part Cable connection port at detector.
Note2: 1 inch = 25.4mm
BS16, DIN PN16 and JIS 10K dimensions:
Meter size
(mm)
25 80 168 66 approx. 3
L1
(mm)
L2
(mm)
D1
(mm)
Weight
(kg)
ANSI class 150 and class 300 dimensions:
Meter size
(inch)
1 3.15 6.61 2.60 approx. 7
L1
(inch)
L2
(inch)
D1
(inch)
Weight
(lb)
Figure 7. Separate type detectors
LF410 and LF414
Meter size 25mm (1")
The dimension of L1 is changed when the material of grounding ring is chosen Pt-Ir or Ta
Meter size L1
15mm (1/2”) 77mm (3.03 inch)
25mm (1”) 95mm (3.74 inch)
40mm (1 1/2”) 115mm (4.53 inch)
50mm (2”) 126mm (4.96 inch)
80mm (3”) 126mm (4.96 inch) 100mm (4”) 136mm (5.35 inch) 150mm (6”) 242mm (9.53 inch) 200mm (8”) 312mm (12.28 inch)
Unit : mm
R(PT)½Male Screw (Two places)
(35)
ΦD1
(126)
91
L1
88
40
(L2)
Note1: Cable glands are not provided for LF414 of FM
and CSA approved type. Refer to the part of Cable connection port at detector.
Note2: Eye bolts are provided at the top for flowmeters
sized 200mm (8").
Note3: 1 inch = 25.4mm
BS16, DIN PN16 and JIS 10K dimensions:
Meter size
(mm)
40 100 190 85 approx. 4 50 110 207 102 approx. 5
80 110* 232 127 approx. 6 100 120* 264 159 approx. 9 150 230 324 216 approx. 21 200 300 385 267 approx. 35
L1
(mm)
L2
(mm)
D1
(mm)
Weight
(kg)
ANSI class 150 and class 300 dimensions:
Meter size
(inch)
1-1/2 3.94 7.48 3.35 approx. 9
2 4.33 8.15 4.02 approx. 11 3 4.33* 9.13 5.00 approx. 13 4 4.72* 10.39 6.26 approx. 20 6 9.06 12.76 8.50 approx. 47 8 11.81 15.16 10.51 approx. 78
L1
(inch)
L2
(inch)
D1
(inch)
Weight
(lbs)
* When Teflon PFA lining is selected, L1 in the table
above becomes as follows:
L1: 115mm (4.53 inch) for meter size 80mm (3") L1: 126mm (4.96 inch) for meter size 100mm (4")
Figure 8. Separate type detectors
LF410 and LF414
Meter size 40 (1 1/2") to 200mm (8")
7
Page 8
LF410/LF600 LF410/LF602
t
d
4-φ11
151
151
74±0.3
74±0.3
4 – Φ11
Attachment
取付金具
IR Switch
赤外線ス
Cable ground
LCD display
LCD 表示
151
178
200
178
200
151
156 (36)
(36) 156
25
25 82
82
Figure 9. Separate type converter LF602
External Connections
Combined type LF410/LF600 and LF414/LF600 flowmeters
Ground terminal
Intermediate voltage (IV) wire,
2
or more.
5.5mm
To be grounded with 100Ω or less ground resistance.
I/O Cable (CVV-S)
13
13
Excitation cable ground
ブル グラ
Signal cable ground
信号 グラ
Unit : mm
I/O cable ground
ケーグラ
Ground terminal
接地端子
(M4)
(M4)
取付金具
Power supply cable ground
Blind screw
矢視A
電源
封止
Instrument Panel : Ordered separately
Current output (4~20mAdc) or PROFIBUS
Digital Output 1 Signal common
for DI and DO
Digital Output 2
Digital Inpu (20 to 30Vdc)
Grounde
with 100Ω or less ground resistance
Grounded with 100Ω or less ground resistance
Power supply
1
*
Power Cable (CVV) Power Switch
(External double-pole power switch)
*1 Locate an external double-pole power switch on the power line near the flowmeter within easy reach of operation.
Use the appropriate switch rating as shown below:
Switch rating: 250Vac, 6A or more In rush current: 15A or more
Figure 10. Combined type LF410/LF600 and LF414/LF600
flowmeters Wiring Diagram
8
Page 9
r
<Separate type LF410/LF602 and LF414/LF602 flowmeters>
Terminal board
LF410/LF600 LF410/LF602
Thick walled steel conduit
Signal cable (2-wire shielded hard-rubber
sheathed cable)
IV wire
5.5 mm
Grounded with 100 or less ground resistance
Power switch (External double-pole power switch )
[Instrument panel : ordered separately]
Current output (4~20mAdc) or PROFIBUS
Digital output 1
Connected detecto
Excitation cable
(3-wire shielded hard-rubber cable)
Power cable
(CVV)
Digital input cable
(CVV-S)
Signal common for DI and DO
Digital output 2
Digital input (20~30Vdc)
Symbol Description Cable
L1 (+) L2 (-)
Power supply Power cable (CVV)
GND Ground (for arrester)
FG Frame ground DI Digital Input (20~30Vdc)
DO1 Digital Output 1 DO2 Digital Output 2
COM Signal Common for DI, DO1, DO2
I/O cable (CVV-S)
Current Output (4~20mAdc)
or PROFIBUS
Shielded cable for
PROFIBUS-PA
2
or more
Power supply
X Y Excitation Output
E
A
B Signal Input
G
Excitation cable (for
LF602 only)
Signal cable (for LF602
only)
Figure 11. Separate type LF410/LF602 and LF414/LF602
flowmeters wiring Diagram
9
Page 10
LF410/LF600 LF410/LF602
Wiring Precautions
(1) Explosion proof type flowmeters are not
provided cable glands. Refer to the part Cable connection port at detector and converter.
(2) Connect the grounding wire (IV wire 5.5mm² or
more) to a good earth ground (100Ω or less ground resistance). Make the wire as short as possible. Do not use a common ground shared with other equipment where earth current may flow. An independent earth ground is recommended.
(3) The allowable cable lengths between the detector
and converter for the separate type flowmeter depend on the electrical conductivity of the object fluid. See Figure 12 below.
(4) DO1, DO2 (opt.), and DI (opt.) use the same
common terminal (COM). This COM can not connect to other equipments which have their own ground terminal. (Power supply for connecting to DI or DO, etc…) Need to wire separately.
(7) Only one PROFIBUS-PA cable goes through a
cable gland of the Electromagnetic Flowmeter. Please use the junction box at system configuration
.
300 200
100
50 30
10
5
3
Allowable cable length (m)
3 5 10 2030 50100 200
Conductivity (μS/cm)
Figure 12. Electrical Conductivity and
Cable Length
Wiring Precautions (PROFIBUS)
(1) For wiring path, avoid places near electrical
equipment that may cause electromagnetic induction or electrostatic induction interference (such as a motor, transformer and wireless transmitter).
(2) Use a PROFIBUS-PA cable for signal cable. In
addition, make sure to use a shielded cable to improve noise resistance. Furthermore, installation of signal cable in metal conduit is recommended.
(3) General PROFIBUS-PA cables are designed for
indoor use where cables are not exposed to humidity, rain, etc. When you install cables, make sure to check the operating conditions such as the operating temperature range of the cable by contacting its manufacturer.
(4) When you carry out cable end treatment of
PROFIBUS-PA cable, use a dedicated cable stripper etc. so that the core wire of the cable will not be nicked or damaged. In addition, for cables, be careful of allowable maximum bend diameter etc. (Basically, do not install cables in a way cables are twisted or bent.).
(5) Consider installing a PROFIBUS-PA arrester in
the communication path of PROFBUS-PA so that the electromagnetic flowmeter will not be affected by lightning etc.
(6) The electromagnetic flowmeter is not equipped
with terminating resistors.Use the terminating resistor unit for PROFIBUS-PA or junction box, if necessary.
10
Page 11
LF410/LF600 LF410/LF602
Piping Precautions
(1) Design piping so that the flowmeter detector pipe
is always filled with the fluid being measured, whether the fluid is flowing or not.
(2)The detector has no adjustable piping mechanism.
Install an adjustable short pipe where needed.
Upstream side
Downstream side
When using a 90°bent, Tee, diffuser, or fully opened valve L ≧ When using other types of valves When no valve plate protrudes into the detector pipe
L:Required straight pipe length
D:Meter size
Grounding of the Combined type and Separate type
Combined type
Grounding terminal
Grounding wire
Gasket
(100 ohm or less
ground resistance)
(3) The required straight pipe length should comply
with the requirements as follows.
(4) Be sure to grounding the flowmeter according to
below figure.
5D L ≧ 10D L ≧ 0
Separate type
Grounding
terminal
Grounding terminal
Grounding wire
(100 ohm or less
ground resistance)
Gasket
(100 ohm or less
ground resistance)
11
Page 12
LF410/LF600 LF410/LF602
(m
)
Meter Size
To select the meter size:
See Figure 13 and find
meter sizes within the
velocity of 0.1 to 10m/s for a specified full-scale (measuring range high limit) flow. Select one that has its full-scale velocity between 1 and 3m/s.
Note: Make sure the full-scale flow rate used for the
final planning stage stays within 10m/s in terms of flow velocity.
Flow velocity (m/s)
Flow rate
3
/h
For SI unit Unit: m
Meter size
(mm)
15 0.1908 0.6361 6.361 25 0.5301 1.767 17.67 40 1.357 4.523 45.23 50 2.120 7.067 70.67
80 5.428 18.09 180.9 100 8.482 28.27 282.7 150 19.08 63.61 636.1 200 33.93 113.1 1131
0.3 m/s 1 m/s 10 m/s
Flow rate
For English unit Unit: gal/min
Meter size
(inch)
1/2 0.8401 2.801 28.01
1
1 1/2 5.975 19.91 199.1
2 9.334 31.12 311.2 3 23.90 79.65 796.5 4 37.35 124.5 1,245 6 84.01 280.1 2,801 8 149.4 498.0 4,980
0.98ft/s 3.28ft/s 32.8ft/s
2.334 7.780 77.80
Flow rate
Calibration Range
If the calibration range is not specified, the standard range as shown below will be used. If the range is specified, we will use the specified range for calibration.
Table 1. Standard Flow Range
Meter size mm (inch)
15 (1/2) 2 3.144 25
25 (1) 6 3.395 75
40 (1 1/2) 15 3.316 175
50 (2) 25 3.537 300
80 (3) 60 3.316 650 100 (4) 100 3.537 1,000 150 (6) 200 3.144 2,500 200 (8) 300 2.653 4,500
3
/h
Note: The unit of "gal/min" is not exchanged
Flow rate
(converted) by "m
Standard flow range
(m³/h)
Flow velocity
3
/h".
(m/s)
Flow rate
(gal/min)
PED matrix in each flange connection.
The following sizes fall under the category for PED in each flange connection when the meter ships to EU. All of them had complied with it from a notified body.
Table 2 PED matrix in each flange connection
Flange standard Meter size
DIN PN 16 and BS 16
DIN PN 10 and BS 10
ANSI 150 and JIS10K
150 to 400mm
(6 to 16 inch) 250 to 400mm (10 to 16 inch)
6 to 16 inch
(150 to 400mm)
Figure 13. Flow Rate and Flow Velocity
12
Page 13
LF410/LF600 LF410/LF602
About establishment environment
Do not store or install the flowmeter in:
• Places where there is direct sunlight.
• Places where excessive vibration or mechanical shock occurs.
• Places where high temperature or high humidity conditions obtain.
• Places where corrosive atmospheres obtain.
• Places submerged under water.
• Place where there is slop floor. To put the flowmeter temporarily on the floor, place it carefully with something, such as stopper, to support it so that the flowmeter will not topple over.
In places like the following places, there is the case that infrared switches do not function correctly. (If this is unavoidable, use an appropriate cover.
•Places where gets very bright light onto operation panel. (direct sunlight, reflection light of sunlight by windowpanes, diffused reflection light of strength etc.)
•Places where smoke and steam occur near.
•Places where a snow, ice or mud that may attached.
Ordering Information
1. When ordering the LF410 series flowmeters, refer to Tables 3 and 4 (Type Specification Codes). An entry must be made for each of the columns in each of these tables.
2. Fluid characteristics: (1) Type of fluid to be measured and its characteristics (2) Fluid temperature (3) Fluid pressure (4) Electrical conductivity of the fluid
3. Measuring range
4. I/O function setting
5. Ordering scope: Flow calibration data: (required or not)
6. Other items Specifications other than standard items
Consult Toshiba representative before ordering
when choose materials of the wetting parts such as lining, electrodes, and grounding rings.
13
Page 14
LF410/LF600 LF410/LF602
Table 3. Specification Code (Wafer type detector LF410 Series)
Model Specification Code
1 2 3 4 5 6 7 8 9 10 11 12 13 14 L F 4 1 0 Normal specification type
L F 4 1 4 Hazardous location certification type (Note 1)
Meter size D 15mm ") E 25mm (1") F 40mm (") G 50mm (2") H 80mm (3") J 100mm (4") K 150mm (6") L 200mm (8")
Mounting Style L Detector/Converter combined type (LF410/LF600) M Detector/Converter separate type (LF410/LF602) P Detector/Converter combined type with PED (LF410/LF600) (Note 6) Q Detector/Converter separate type with PED (LF410/LF602) (Note 6) FM and CSA Class - I Division 2 type (Note 1) A Detector/Converter combined type (LF414/LF600) B Detector/Converter separate type (LF414/LF602)
Connection flange standard (Note 2) C ANSI 150 D ANSI 300 E BS PN 10 F BS PN 16 G DIN PN 10 H DIN PN 16 J JIS 10K K JIS 16K L JIS 20K Z other
Electrode Material (Note 6)
B 316L stainless steel C Ti (titanium) D Pt-Ir (platinum/iridium) E Ta (tantalum) F Hastelloy C (Equivalent)
Z other Lining Material (Note 6)
A Alkali-resistant ceramic (Note 3) B Acid-resistant ceramic (Note 3) C Teflon PFA Grounding Ring Material (Note 6) C 316 stainless steel D 316L stainless steel E Ti (titanium) F Ta (tantalum) G Pt-Ir (platinum/iridium) H Hastelloy C (Equivalent) Z other Flow and calibration velocity range A 0.3 to 10 m/s (standard range calibration) B 0.3 to 10 m/s (specified range calibration) C 0.1 to 10 m/s (specified range calibration) Excitation and Signal Cables A not provided B 30m cable, provided (Note 4) C other lengths, provided (Note 4) Coating A no coating B phthalic acid resin coating pearl-gray colored (Note 5) C black tar epoxy resin 0.3 mm D black tar epoxy resin 0.5 mm E Black tar epoxy resin 0.5 mm for submersible type (Note 7)
Size code explanation: :Object : Standard : Option –: Not available
Note1: Cable glands are not provided. Refer to the part of “Cable connection port” at detector and converter. Note2: Same shape of flange standard among ANSI 150, ANSI 300, BS10, BS16, DIN10, DIN16, JIS10K, JIS16K, and JIS20K. Note3: The type of material used is alumina ceramic, this is suitable for both acid and alkali liquids.
This ceramic lining and the electrodes are held with seals, there are two types, one for acids, and the other for alkalis.
Note4: Separate type detector only. Specifyng the code “C”, indicate the length of cables from 1 to 300m in 1 meter increments. Note5: Phthalic acid resin coating is standard for the flowmeter with meter size 15mm(1/2”), 150mm(6”), and 200mm(8”). Note6: Consult Toshiba before ordering when choose materials at the wetting parts. Note7: Ceramic lining only. (Teflon PFA lining is not available.) Note8: Check the Table 2 whether your chosen meter size meets this directive or not when the meter is shipped to EU. If yes, need to choose this code.
Description
Detector category
Normal type
Gr.-A Gr.-B Gr.-C Gr.-D
√ √ √ √ √ √
√ √
√ √ √ √ √ √
                 
         
           
   
   
     
Ex. type
√ √ √ √ √ √
√ √
  
√ √
– –
14
Page 15
Table 4. Specification Code for converters
Model Specification Code 1 2 3 4 5 6 7 8 9 10 11 12 13 14 L F 6 0 Electromagnetic flowmeter converter
0 Integral (Combined) type 2 Separate (Remote) type
A
F
Shape A B
Converter mounting fitting A C E
1 2
Digital input/output
Current output and Communication function 1 2
Power supply
1 2 3
A E
Purpose
Standard cFMus class I, Division 2 approved
Integral type with case Separate type with case
None Panel, Accessory for wall mounting (BNP material: SUS304) Accessory for pipe installation (BNP material: SUS304)
Digital output points 1 (DO1) Digital output points 2 (DO1+DO2) +Digital input point 1 (DI)
Current output + HART communication PROFIBUS communication (Current output is not usable)
100Vac-240Vac, 50/60Hz 24Vdc 110Vdc
Instruction manual
Japanese English
Contents
Code explanation: : Standard : Option –: Not available
LF410/LF600 LF410/LF602
LF600
type
 
– –
 
 
  
 
LF602
type
  
 
 
  
 
15
Page 16
LF410/LF600 LF410/LF602
ISO9001 and ISO14001 are certified.
Misuse of this product can result in damages to property or human injury. Read related manuals carefully before using this product.
No. EJL-114B
Specifications are subject to change without notice.
Printed in Japan 2006-10 (TDOC)
© TOSHIBA Corporation 2006
All Rights Reserved.