Toshiba Activion 16 Product Data

Product Data
No. MPDCT0290EA
Multislice HELICAL CT SCANNER
APPLICATION
ActivionTM16 is a 16-slice Helical CT system that supports whole-body scanning. The system generates a minimum of 32 slices per 1.5 sec­onds using the Selectable Slice-thickness Multirow Detector (SSMD). In addition, the high-speed rotation mechanism and the fast reconstruction unit of the system allow quick image acquisition to further improve throughput in CT examina­tions.
FEATURES
• Multislice detector
Detection elements with high-power and uniform output characteristics enable a minimum slice thickness of 0.5 mm, and accurate isotropic data can be acquired. The adoption of the SSMD method means that high­speed as well as high-resolution scanning are support­ed. The minimum slice thickness is reduced to 0.5 mm, mak­ing it possible to select the desired slice image for scan­ning from among 0.5 mm, 1 mm, 2 mm, 3 mm, 4 mm, and 5 mm, depending on purpose.
High-speed scan
Data for 16 slices can be acquired simultaneously in each scan. For example, scanning of the lung fields over a range of 30 cm at a slice thickness of 1 mm can be completed in 10 seconds or less. Since the acquisition is completed in a short period of time, this obviously alleviates the burden on the patient, but also improves throughput by eliminating the need to wait for the X-ray tube to cool down.
High-quality images
It is now possible to use thin-slice helical scanning for routine examinations. Based on high-resolution voxel data, smooth and finely-detailed 3D and MPR images can be obtained with the same size in the X, Y, and Z directions (isotropic). In CT cerebral angiography, for example, scanning over a range of 40 mm at a slice thickness of 0.5 mm can be completed in 4 seconds and image processing (such as 3D, MPR, or tomographic image processing) can be performed for a single volume acquisition dataset through simple operations. In addition, by stacking the data acquired using thin slices, images with reduced partial volume effects can be obtained.
Outstanding operability
Operability is improved as described below. – In 3D image processing or time-consuming image pro-
cessing such as for regions in which calcified areas are superimposed on contrast medium, bone elimination can be performed easily while observing reference images.
– Most 3D images can be generated using the optimal
conditions by simply selecting an appropriate preset icon.
Selectable image slice thickness
It is possible to acquire the data for routine examination, detailed examination and to generate 3D images in a sin­gle scan. For example, by performing a helical scan with a 0.5-mm slice thickness, it is possible to generate images at vari­ous slice thicknesses from the same data, such as 10­mm slice images for routine examinations, 5-mm slice images for detailed examinations, and 0.5-mm slice images for generating 3D images. It is also possible to set the image slice thickness with multiple ranges. For example, by performing a helical scan of the head with a 0.5-mm slice thickness, it is possible to generate images with optimal slice thickness for each region in a single reconstruction, such as 5.0-mm slice images for the cranial base as well as 10-mm slice images for the cerebral parenchyma.
Exposure reduction
This system incorporates the quantum denoising soft­ware (QDS) as a standard function, which is effective for reducing patient exposure. The QDS is an adaptive filter that can recognize recon­structed objects. It can perform sharp filter processing for regions where the degree of change is high, such as tissue borders; and smooth processing for regions where the degree of change is low (close to uniform). This makes it possible to further improve the quality of images acquired using normal dosages, and improves the quali­ty of images acquired with small dosages to an image quality level obtained with normal dosages. As a result, it is possible to reduce the exposure dose for the patient, since the scanning can be performed using the optimal dose for the expected image quality.
2
SURE
FluoroTM(option)
Conventional CT fluoroscopy shows only a single slice, but
SURE
Fluoro (Multislice CT fluoroscopy) permits realtime image reconstruction to display 3 images obtained by combining data from the SSMD.
SURE
Fluoro significantly improves operability in biopsy and interventional proce­dures.
Tilt Helical scanning
Helical scanning with gantry tilt, from 30° forward to 30° backward, is available because the system employs the TCOT reconstruction technique.
COMPOSITION
Standard composition (Model:TSX-031A)
Gantry ................................................................................1
Patient couch......................................................................1
Console ........................................................................1 set
Accessories Inter-unit cablesManualsSet of phantomsAcquisition supportFootswitch for the patient couch
Note: The console desk is not included in the standard con-
figuration.
Optional items
Cerebral blood-flow analysis system (CBP-study) (CSCP-002A)
System transformer (CETF-004B)
Display system for dental application (CDP-07A)*
FlyThrough software (CFT-03A)*
Vessel view (CVV-001A)
SURE
Fluoro (TSXF-003E)
LCD monitor for
SURE
Fluoro (15-inch type) (CMM-003E)
Magneto-optical disk drive unit (CMO-12A)
DICOM storage SCP (COT-30D)
DICOM MWM (COT-32D)
DICOM MPPS (COT-33D)
DICOM Q/R SCP (COT-34D)
DICOM Q/R SCU (COT-35D)
DICOM storage commitment SCU (COT-41D)
DICOM PGP profile (COT-44A)
DICOM fast transfer system (COT-45A)
Color printer interface (CCP-03A)
ECG-gated scan system (CHEG-004C)
Injector synchronization system (CKIS-004A)
Orbital synchronized scan system (CKOS-001A)*
Pediatric scanning system (CHKS-002A)
* Not available in the U.S.A.
PERFORMANCE SPECIFICATIONS
Scan parameters
Scan regions: Whole body, including head
Scan system: 360° continuous rotate/rotate
Scan plan
programming: More than 360 different
sequences can be pre-pro­grammed.
Scan timeCT scan: 0.48 s (Partial), 0.75, 1, 1.5, 2,
and 3 s
Scan cycle time (for 0.75-s scan)SCAN & SCAN mode: Min. 1.5 s (rapid sequence
scanning, couch-top movement 10 mm)
Note: The scan cycle time refers to the time between the start of
one scan and the next. SCAN & VIEW mode permits immediate viewing of images after acquisition of each individual slice.
Scan fieldCT scan:
φ180 mm (SS) φ240 mm (S) φ320 mm (M) φ390 mm (L) φ500 mm (LL)
– Scanoscopy:
*: For the short patient couch version Note: The actual range that can be viewed is less than the
couch-top movement range in scanoscopy.
Slice thickness: 0.5, 1, 2, 3, 4, and 5 mm These slice thicknesses are implemented by stacking the data acquired in one of the following acquisition modes.
Acquisition16-row 0.5 mm
× 16 rows
1 mm
× 16 rows
– 4-row 0.5 mm
× 4 rows
1 mm
× 4 rows
2 mm
× 4 rows
3 mm
× 4 rows
4 mm
× 4 rows
5 mm
× 4 rows
Gantry tilt angle: From forward 30° to backward
30° (in 0.5° increments) Remote control from the console is possible.
Tube position for scanoscopy: 0°, 90°, 180°, and 270°
Any arbitrary angle can be specified (in 5° increments).
Gantry aperture: 720 mm in diameter
Axialdirection
Up to 390 mm
200 mm to 1,750 mm (1,450 mm*)
Longitudinal
direction
Adjustable from
Helical scan
X-ray tube rotation speed: 0.75, 1, 1.5 s/360°
Continuous scan time: Max. 100 s
Scan start time delay: Min. 1 s
Setting is possible in increments of 0.1 s.
Image slice thickness:For 16-row scanning
(TCOT): The maximum image slice thick-
ness is 10 mm.
– For 4-row scanning
(MUSCOT): The maximum image slice thick-
ness is 10 mm.
Scan field in the longi­tudinal direction: Max. 1,750 mm/scan (for the
long patient couch version) Max. 1,450 mm/scan (for the short patient couch version) Up to 10 scan plans are pro­grammable in one eXam Plan. (Multiple and/or Multi-directional Helical)
Gantry tilt: Helical scan is possible in the
range from 30° forward to 30° backward (only for 16-slice acquisition).
Couch-top speed: The couch-top speed can be
specified in the range from
0.8 mm/s to 96 mm/s.
Helical pitch: For 16-row scanning (TCOT): Setting is possible in the ranges
from 10 to 16 and from 18 to 24
in increments of 0.1. For 4-row scanning (MUSCOT): Setting is possible in the range
from 2.5 to 3.5 and from 4.5 to
6.0 in increments of 0.5.
Helical pitch = Couch-top movement
(mm/rot.)/nominal scanning
slice thickness (mm)
CT pitch factor = Helical pitch/number of slices
scanned in a single rotation
Note: The CT pitch factor is defined by IEC 60601-2-44 Amd.1:
2002
SURE
Exposure
TM
Function for continuously vary-
ing the X-ray tube current to
ensure the optimal X-ray dose
during helical scanning.
Image reconstruction time: Up to 10 images/s (0.1 s/image)
Real-time helical reconstruction time: 12 images/s (0.083 s/image)
(1 slice, 512
× 512 matrix)
MPDCT0290EA
Patient couch
Vertical movement System: Hydraulically driven – Speed of vertical
movement: Up: 16 to 24 mm/s (50 Hz)
19 to 28 mm/s (60 Hz)
Down: 20 to 30 mm/s
Stroke: Approx. 588 mmMinimum couch-top
height: Approx. 310 mm
– Maximum couch-top
height: Approx. 944 mm
Couch-top movement System: Motor-driven or manual – Speed of movement: 130 mm/s (fast)
10 mm/s (slow)
– Stroke: 2,190 mm (for the long patient
couch version) 1,890 mm (for the short patient couch version)
– Scannable range: 1,800 mm (for the long patient
(with headrest) couch version)
1,500 mm (for the short patient couch version)
– Step feed pitch: 0.5 to 600 mm in 0.5-mm
increments
– Reproducibility: ±0.25 mm
Repeatable to within ±0.25 mm after 600-mm move­ment
Couch-top width: 470 mm
Remote control from the console is possible.
Note: This function allows the user to check the image on the
console and adjust to the couch-top without leaving the console. Adjustment is possible in 10-mm increments at the console.
Load limit Max. allowable load: 205 kg (450 lb)
Footswitch: Auto set/Auto home function can
be selected.
Voice-recorded instruction and scan system (VoiceLink)
Voice instructions to the patient can be recorded electroni­cally by the operator and automatically played back during scan sequences as part of the eXam Plan.
Number of messages: Max. 32 messages
Number of seconds: Max. 128 s for a total of
32 messages
Recording time: Max. 30 s per message
Delay time setting: The delay time between the end
of the message and the start of scanning can be set up to 10 s in increments of 1 s.
3
4
SURE
StartTM:
Continuous scan time: Max. 100 sRegion of interest
(ROI): Max. 3 ROIs
– CT number
measurement interval: 0.083 s (12 measurements/s)
Scan start delay time: Min. 3 sDisplay function: Mean CT number and elapsed
time
Specification of reconstruction position: By entering the couch-top posi-
tion or using the scanogram
Reconstruction method: TCOT reconstruction (applicable
to 16-row acquisition data) MUSCOT reconstruction (applic­able to 4-row acquisition data)
Reconstruction mode: Full image
Half image Detail image
The relationships between the slice thickness and the imaging area scanned in 7.5 s (0.75 s, 10 rotations) in the longitudinal direction are shown for helical pitches 10 and 20.
*: The couch-top traveling distance per rotation is shown rela-
tive to the slice thickness.
Dynamic scan
Scan time: 0.75, 1, 1.5 s/360°
Programmable time: Max. 1 hour
This refers to the maximum time within which a series of scans is performed following a predeter­mined eXam Plan.
Number of programmable scans: Max. 10
Max. time of one continuous scan is 100 s.
Scan planScan interval: Min. interval is 1 s
Setting is possible in increments of 0.1 s in a scan interval of more than 1 s.
Note: When a scanning mode with patient couch movement is
used, the minimum scan interval is limited by the time required for movement.
Scan start delay time: Min. 0.5 s
Setting is possible in increments of 0.1 s.
Scan rate: Max. 133 scans/100 s
(0.75-s scan, 133 rotations)
Image reconstructionNumber of images: Max. 4 images/scanImage interval: Reconstruction is possible in
increments of 0.1 s.
Reconstruction time: Min. 0.5 s
Real-time reconstruction
time: 12 images/s (0.083 s/image)
(1 slice, 512
× 512 matrix)
X-ray generation
X-ray exposure: Continuous
X-ray tube voltage: 80, 100, 120, and 135 kV
X-ray tube current: 10 mA to 300 mA (260 mA for
135 kV) (in increments of 10 mA)
X-ray tube heat capacity:4.0 MHU
X-ray tube cooling rate: Max. 864 kHU/min
Focal spot sizeIEC 60336: 1993
nominal : 0.9 mm
× 0.7 mm (small)
1.4 mm
× 1.4 mm (large)
X-ray detection
Detection system: Solid-state detectors
Main detector: 800 channels
× 28 elements
Data acquisition: 800 channels
× 16 rows
Reference detector: 1 set
View rate: Max. 1,200 views/s
Data processing
Reconstruction matrix: 512 × 512
Picture element (pixel) sizeCT image: Unit: mm
Scan field SS S M L LL
Pixel size
* to * to * to * to * to
0.35 0.47 0.63 0.76 0.98
*: Depending on the Vari-Area or Zoom factor
Scanogram: Unit: mm
Reconstruction filter functionsFunctions for the abdomen with BHCFunctions for the abdomen without BHCFunctions for the brain with BHCFunctions for the brain without BHCFunctions for the inner ear and boneFunctions for the lungFunctions for high-resolution modeFunctions for super-resolution mode for the inner ear,
bone, and lung
– Functions for maintenance
Standard
Enlargement ratio (area)
4.00
2.00
1.0
0.50
LL
L
M
S
Pixel size
ecilsgnitteS
ssenkciht
0102
*hctiplacileH
mm5.0mm05mm001
mm1mm001mm002
5
Image reconstruction timeCT scan: Min. 0.17 s Scanoscopy: Reconstructed and displayed
simultaneously with scanning (real-time reconstruction)
Data processing unitCentral processing
unit: 32-bit processor x 2
Memory size: 3 GbytesMagnetic disk unit: Raw data, 144 Gbytes
Image data, 73 Gbytes
Data storage
Magnetic diskRaw data: Max. 3,600 rotations
(0.75-s helical scan)
– Image data:
Max. 100,000 (per each console)
CD-R/DVD driveCD-R
· Storage capacity: 650 Mbytes (formatted)
· Media: CD-R disk (in accordance with
Orange Book part II)
· Data format: DICOM format (in conformance with DICOM standards PS3.10)
· Image data: Max. 1,000 images (assuming that images [512 Kbytes/image] are recorded on a disk at one time)
DVD
· Storage capacity: 9.4 Gbytes (double-sided)
· Media: DVD-RAM
· Image data: Max. 16,000 images (for a dou-
ble-sided disk, DICOM format)
· Raw data: Max. 120 rotations (for a single- sided disk, 0.75-s scan)
Image display
Display monitor: 19-inch color LCD Size of display area comparable to a 21-inch CRT monitor
Monitor matrix: 1,280
× 1,024
Image matrix: 1,024
× 1,024 (max.)
CT number
Display range: From -1,536 to +8,191
Note: The CT number measurement range is from -32,768 to
+32,767.
Window width/level: Continuously variable (adjustable at variable speed)
Preset window: Three types of window settings can be preset for each image.
Window types:
Linear, non-linear (6 user-pro­grammable), and double win­dows
Image retrieval
Method: On-screen menus and keyboardMode: Image, series, and patient
Autoview function: Software control, function key
Multi-frame display: Reduction/cut-off display, ROI
processing
Inset scanogram display
Selective related information display
Cine displayImage display speed: Variable
Scanogram/CT image
switching: Show scano line, zoom, hide
scano line
Slice-feed playback (CineView): High-speed image feeding
using the mouse or keyboard
Image processing
Scanogram processingSlice position display (display of planned slice,
preset slice, and last scanned slice)
– Anatomical scale (display of position, relative to select-
ed zero position)
Slice position settingEnlargement (4
× for L or LL size)
CT image processingROI setting and processing
· ROI shape: Point, rectangular, polygonal,
elliptical, irregular
· ROI processing: Mean value, standard deviation, area, number of pixels, maxi­mum value, minimum value
· ROI display: Ten ROIs can be displayed on an image.
· ROI control: Size, position, rotation
Measurement of distance and angle between two pointsProfile (oblique profile also available)HistogramCT number displayMark display (grid display, scale display)Volume calculationEnlargement, reduction, panningAddition/subtraction between imagesBand display (non-linear windowing)Comment and arrow insertionTop/bottom, right/left, black/white reversal of imageImage filteringImage rotation (arbitrary rotation)Screen saveHigh-speed axial interpolationMultiView (Auto MPR)Quantum denoising software (QDS)Boost3D
TM
Z-sharpening
Raw data processingZooming reconstructionStack reconstructionProtect/UnprotectHalf-view reconstruction of helical scan raw dataPlay/Reverse reconstruction (Helical & Dynamic scan)Priority reassignment in reconstruction queue
MPDCT0290EA
6
System managementWarm-upCalibration data acquisitionPicture retouchingOperating environment setting functionExamination record
Display of exposure
dose: CTDI
vol
(or CTDIw) /DLP/
Geometric eff.
3D color image processing
High-quality 3D images can be obtained rapidly by easy operation.
3D surface renderingClipping, texture or non-texture
3D volume renderingMaximum intensity projection (Max-IP) Minimum intensity projection (Min-IP)X-ray volume renderingIntensity volume renderingShaded volume rendering (an arbitrary opacity curve
can be set)
Display/processing function Zooming, panning, measurement (distance, angle), annotation, cutting, drilling
Cine display
MPR
3 orthogonal planes/oblique image Curved MPR
Easy accurate bone elimination function
High resolution mode
Image transfer/conversion
1000BASE-T, 100BASE-TX, 10BASE-T
Toshiba protocol
DICOM storage SCU
TIFF conversion
Filming
Ethernet: Toshiba protocol
DICOM PRINT SCU
Sheet editing function using virtual film
T-mode: Related information items such
as the patient name are dis­played in the footer area using a larger font.
Note: To use T-mode, the laser imager must support
2048 pixels x 2404 pixels for a 1 x 1 frame.
Auto filming in eXam Plans
IMAGE QUALITY
NoiseStandard deviation: Less than 0.5%Scan parameters
· Tube voltage: 120 kV
· Tube current: 260 mA
· Scan time: 1.5 s
· Reconstruction
function: FC70
· Slice thickness: 10 mm
· Scan field: S
Phantom:
φ24 cm water
Spatial resolution: 14.5 lp/cm at MTF 2%
18 lp/cm at cutoff (reference)
8.0 lp/cm at MTF 50% (MTF calculation value)
Scan parameters
· Tube voltage: 120 kV
· Tube current: 200 mA
· Scan time: 1 s
· Slice thickness: 2 mm
· Scan field: S
· Reconstruction
function: FC90
Phantom: IRIS QA phantom
High contrast resolution
X-Y plane – High-resolution
mode (FC90): 0.35 ± 0.05 mm
– Standard mode
(FC30): 0.55 ± 0.05 mm
Scan parameters
· Tube voltage: 120 kV
· Tube current: 200 mA
· Scan time: 0.75 s
· Slice thickness: 2 mm
· Scan field: S
Phantom: Toshiba standard
7
SYSTEM COMPONENTS AND THEIR FUNCTIONS
Gantry
The scanner is composed of the gantry and the patient couch. The scanner uses a fan-shaped continuous X-ray beam to scan the region to be examined. Transmitted X­rays are detected and converted into electrical signals by the SSMD. The gantry includes the main body and its support mecha­nism. The X-ray tube and the SSMD are mounted facing each other on either side of the gantry aperture, and the X­ray tube and detectors rotate continuously around the aperture of the gantry. A slipring is employed to transmit power between the gantry and the rotating X-ray high-volt­age generator assembly. The gantry can be tilted forward and backward in order to perform tilt scanning. Three-dimensional alignment lights are provided for setting slice positions. Gantry and patient couch operating controls are provided on both sides of the front of the gantry housing. The patient guide display indi­cates the scan status to the operator and the patient. The X-ray high-voltage generator is built into the gantry, and the system employs a high-frequency inverter for generat­ing and stabilizing the high voltage supplied to the X-ray tube. The generator includes electronic circuits for control­ling the speed of the rotating anode in the X-ray tube. Use of a high-frequency inverter system results in high power output combined with excellent stability. In addition, the system is compact and light weight.
X-ray generator
This unit supplies stable high voltage to the X-ray tube unit. The high-frequency inverter method is employed, resulting in a light and compact design. This unit is incorporated in the gantry.
Max. power: 42 kW
X-ray tube
This is a large-capacity, high-cooling-rate X-ray tube that is able to withstand continuous operation as in helical scan­ning.
Heat capacity: 4 MHU
Cooling rate: Max. 864 kHU/min
Patient couch
The patient couch is positioned in front of the gantry and supports the patient. The entire unit moves vertically and the top moves longitudinally. In an emergency, the couch­top can be pulled out manually with very little effort. The couch-top can also be lowered to a minimum height of approx. 310 mm from the floor, facilitating transfer of the patient from a low bed or stretcher.
MPDCT0290EA
8
Console
The console is provided with a hybrid keyboard, a monitor, and a mouse.
Functions for scanningSelection of scan parametersScanoscope controlScan controlRemote control of couch-top movementRemote control of gantry tilt
Functions for image processing Window level and window width adjustmentOther mouse-operated image processing functions
OPERATING FEATURES
Patient handling and positioning
The couch-top can be lowered to approx. 310 mm (at the center of the couch top) from the floor, making it easi­er to transfer the patient to and from a bed or stretcher.
Alignment lights are provided in the gantry aperture for fast and accurate patient positioning.
High-precision couch-top positioning is possible from the integrated console or by manual operation from the con­trol panel and clear digital readouts are provided on the gantry.
The couch-top can be pulled out manually in an emer­gency.
Scanning
Toshiba's scanoscope function provides a projection image of the patient for high-precision advance planning of the slice positions.
The longitudinal length of the scanning field for the scanogram can be adjusted up to 1,750 mm (1,450 mm for the short patient couch version). Real-time helical permits the scan to be aborted at any time. This mini­mizes the patient exposure dose.
The auto index function allows automatic incremental couch-top movement based on the slice positions deter­mined through the scanogram.
The eXam Plan function allows simple selection of pre­programmed scanning parameters for routine examina­tions, maximizing patient throughput.
Protocol comments can be saved in each eXamPlan provideng interactive onscreen instructions for all studies reducing the need to refer to a separate protocol book.
The Vari-area function allows the user to pre-select a region of interest for zooming using raw data, permitting immediate post-scan analysis. Zooming using raw data yields higher resolution than enlarging an image that has already been reconstructed.
Dynamic and rapid sequence scan modes are provided.
Multislice helical scan acquires raw data by rotating the
X-ray tube continuously while moving the patient continu­ously through the scanner. The volume data acquired can be used to reconstruct slices at any desired axial positions. This scan mode is best used for rapid patient scanning during a single breath-hold and for high-defini­tion three-dimensional and MPR imaging.
Real-time helical reconstruction mode makes it possible to observe the images being scanned in real time at a maximum at 12 frames per second. This mode shows any shift in the slice position in real time and helps the operator to check the scan field on the image, the con­trast study timing, the presence of patient body motion, etc. The patient can therefore be released immediately after scanning.
The
SURE
Start function allows the operator to start helical scanning at the timing of maximum enhancement in con­trast studies.
SURE
Start monitors the scan from the start of a contrast study at a certain slice position while measur­ing the changes in CT number on the image being dis­played in real time. When the contrast reaches the predefined threshold, helical scan automatically starts. This technique ensures optimal contrast enhancement, independent of individual differences in blood flow speed, and at the same time minimizes the dose of con­trast medium.
Data processing
Rapid reconstruction increases the reconstruction speed to 0.5 second in 0.75-second scans. (Scan & View mode)
A variety of reconstruction algorithms are available and can be selected according to the anatomical region to be examined and the clinical objective of the study. These include algorithms for the abdomen, head, bone, lung, small structures, soft tissues, etc.
Image display and processing
Reconstructed images are automatically displayed according to the window settings preset in the eXam Plan.
The window save function allows the user to store an image with window settings different from the ones set in the eXam Plan.
Filter parameters can be customized through simple on­screen menu selections. These parameters include the number of filtering passes, matrix size, and filter coeffi­cients.
Images can be rotated and reversed either right/left, top/bottom, or black/white.
The Multi-frame feature allows up to 16 images to be retrieved and displayed simultaneously on the screen.
The three-dimensional image display function allows color three-dimensional and MPR images in real-time to be generated from the volumetric scan data acquired by helical scanning. This results in higher definition and image quality than images reconstructed from conven­tional single-slice scanning. This is because helical scanning provides superior data continuity along the patient axis compared with conventional scanning.
9
Image storage and archiving
The system is provided with a 217 Gbyte magnetic hard disk as standard equipment, permitting the on-line stor­age of approximately 100,000 images and 3,600 rota­tions of raw data.
A 9.4 Gbyte DVD-RAM disk drive is provided as stan­dard equipment. The image storage capacity of the DVD-RAM disk is approximately 16,000 images per disk.
Image filming
Filming of images can be performed manually or auto­matically from the console.
Automatic filming sends an entire study to the laser cam­era. Filming is performed in background mode so that other scanner and image processing functions can be performed without interruption or delay.
When T-mode is used, related information items dis­played together with an image (surrounding the image, in a small font) are displayed in the footer area using a larg­er font, permitting not only easier reading but also sim­pler film management.
Note: To use T-mode, the laser imager must support
2048 pixels x 2404 pixels for a 1 x 1 frame.
Patient throughput
Patient throughput and cost effectiveness were major objectives in the design and production of the system.
The system incorporates a 4.0-MHU X-ray tube with a very fast cooling rate of 864 kHU/min in actual use.
High-speed scans can be performed in as little as 0.75 second per scan.
Real-time reconstruction is possible in scanoscopy.
CT images can be reconstructed in 0.5 second for 0.75-
second routine scans.
Ease of operation is ensured by incorporating a hybrid keyboard, mouse-driven menus, and a large color LCD screen.
The couch-top can be lowered very near the floor, simpli­fying patient transfer.
COMPLIANCE
Council Directive 93/42/EEC Concerning Medical Devices (Medical Device Directive)
IEC: IEC 60601-1: 1988 IEC 60601-1 Amd.1: 1991 IEC 60601-1 Amd.2: 1995 IEC 60601-1-1: 2000 IEC 60601-1-2: 2001 IEC 60601-1-3: 1994 IEC 60601-1-4 Ed.1.1: 2000 IEC 60601-2-28: 1993 IEC 60601-2-32: 1994 IEC 60601-2-44: Ed.2.1: 2002
DIMENSIONS AND MASS
SITING REQUIREMENTS
Power requirements
Phase: Three-phase
Voltage: 200 V*
Frequency: 50 Hz or 60 Hz ±0.5 Hz
Line capacity: 75 kVA
Voltage fluctuation
due to load variation: Less than 5%
Power voltage
fluctuation: Less than 10%**
* Please consult Toshiba in the case of other voltages or exces-
sive power fluctuation.
** Represents the total voltage fluctuation due to load and power
variation.
Grounding
Grounding must be provided in accordance with local regulations for medically used electrical equipment.
Unit
Dimensions L ×W × H Mass mm (in) kg (lb)
Gantry
900 × 2,070 × 1,910 1,280 (35.4 × 81.5 × 75.2) (2,822)
Patient couch
Long patient couch version
Short patient couch version
2,690 × 630 × 450 450 (105.9 × 24.8 × 17.7)
(94.1 × 24.8 × 17.7)
(992)
2,390 × 630 × 450 420
(926)
CPU cabinet 1
815 × 450 × 700 100 (32.1 × 17.7 × 27.6) (220)
CPU cabinet 2
815 × 450 × 700 85 (32.1 × 17.7 × 27.6) (187)
MPDCT0290EA
10
Power distribution board
Ambient conditions
Temperature Humidity Heat generation Scan room Gantry 20°C to 26°C 40% to 80% Approx.
Tolerance: ±2°C No 9.720 kJ/h (*1)
condensation 32,070 kJ/h (*2)
Patient couch 20°C to 26°C 40% to 80% Approx.
Tolerance: ±2°C No 1,080 kJ/h (*1)
condensation 1,800 kJ/h (*2) Operator’s room Console 16°C to 28°C 40% to 80% Approx.
No 9,720 kJ/h
condensation
*1: When scanning is not performed. *2: When scanning is performed continuously at the maximum
rated output of the system.
Minimum area for installation
For the long patient couch version: 22 m
2
For the short patient couch version: 21 m
2
NFB
150 A
NFB
100 A
In case of
3-phase, 200 V
In case of
3-phase, 400 V
System
System transformer (option)
Ground resistance: As per applicable legal requirements.
Ground bar
Room layout example
1,494
(1,194*)
1
5,530 (4,930*)
7,130 (6,530*)
1,600
1 2
*: For the short patient couch version
3,072
Gantry Patient couch
2
3
Unit: mm
Console
3
11
Installation requirements Scan room
Before installing the gantry, check the maximum permis­sible floor load.
The scanner emits radiation. X-ray shielding must be provided around the scan room and the entrance in accordance with all local requirements and regulations.
The ceiling should be at least 2,500 mm high to permit the use of a contrast medium injector.
Wiring pits and ducts are required for routing cables that connect the various units.
Operator’s room
An observation window is required for monitoring the scan room. X-ray shielding of the window glass must be provided in accordance with all local requirements and regulations, and the bottom of the window frame should be 90 cm from the floor.
Wiring pits and ducts are required for routing cables that connect the various units.
The operators room should have entrances for access to the corridor and the scan room.
Cable connections between units
Checks before bringing-in the unit
Check in advance the width of the corridor, the dimen­sions of the entrance, and the dimensions and maximum allowable load of the stairs and elevators to ensure that it is possible to bring-in the unit safely and without difficulty.
Minimum external dimensions of the entrance used for bringing-in the unit are as follows:
Width: 980 mm (38.6 in) Height: 2,070 mm (81.5 in)
The corners of corridors should be as illustrated below.
Elevator minimum load: 2,000 kg (4,400 lb)
W
W
W >1,800 mm (70.9 in)
MPDCT0290EA
Breaker box
System
transformer*
Patient couch
Gantry
Console
(CPU cabinet 2)
Console
(CPU cabinet 1)
(Scan room)
(Operator's room)
Speaker
5
12
17
17
5
17
9
*: When line voltage is 200 V ±10 V, the system transformer is not required.
Unit: m
12
OUTLINE DRAWINGS
Gantry and Patient Couch
MPDCT0290EA
13
OUTLINE DRAWINGS
344
(13.5)
556
(21.9)
1,910
(75.2)
900
(35.4)
2,070 (81.5)
(TILT ANGLE)
Gantry
14
OUTLINE DRAWINGS
1,862
(73.3)
2,690
(105.9)
588 (23.1)
1,038 (40.9)
STROKE
450
(17.7)
2,187
(86.1)
2,190
(86.2)
470
(18.5)
630
(24.8)
430*
* When the arm up holder is mounted.
(17)
Patient Couch (for the long patient couch version)
MPDCT0290EA
15
OUTLINE DRAWINGS
1,562
(61.5)
2,390 (94.1)
588 (23.1)
1,038 (40.9)
STROKE
450
(17.7)
1,887
(74.3)
1,890
(74.4)
470
(18.5)
630
(24.8)
(17)
430*
* When the arm up holder is mounted.
Patient Couch (for the short patient couch version)
700
(27.6)
700
(27.6)
574
(22.6)
414
(16.3)
450
(17.7)
500
(19.7)
450
(17.7)
815
(32.1)
TOSHIBA MEDICAL SYSTEMS CORPORATION
http://www.toshibamedicalsystems.com
©Toshiba
Medical Systems
Corporation 2007 all rights reserved.
Design and specifications subject to change without notice. "Made for Life" is a trademark of Toshiba Medical Systems Corporation. Activion,
SURE
Fluoro,
SURE
Exposure,
SURE
Start,
SURE
Plaque, and Boost3D are trademarks of Toshiba Medical Systems Corporation. This document may include trademarks and registered trademarks of other companies. Model TSX-031A 2007-04 TME/KI
Toshiba Medical Systems Corporation Nasu Works meets the Environmental Management System standard, ISO 14001.
Toshiba Medical Systems Corporation meets internationally recognized standards for Quality Management System ISO 9001, ISO 13485.
Printed in Japan
MPDCT0290EA
OUTLINE DRAWINGS
Console
Note: The console desk is not included in the standard configuration.
Some of the units shown in the photograph on the front page differ from those shown in the drawings above.
Loading...