Torbal AGC User Manual

Instruction Manual - TORBAL AGC Series Laboratory Balances
Table of Contents
Cautionary Notes and Precautions........................................................................................................................... 3
1. Specification........................................................................................................................................................ 4
2. Keys and Display Indicators................................................................................................................................ 6
3. Commands and abbreviations.............................................................................................................................. 7
4. Parts Description..................................................................................................................................................8
5. Unpacking the Scale and Getting Started............................................................................................................ 9
6. Weighing............................................................................................................................................................11
6.1.1.Zeroing the scale..................................................................................................................................11
6.1.2.Taring ................................................................................................................................................. 12
6.1.3.Clearing a Tare.................................................................................................................................... 13
7. Main Menu.........................................................................................................................................................14
7.1.1.Calibration........................................................................................................................................... 14
7.1.1.1.Start (Initializing Automatic Calibration)........................................................................................ 15
7.1.1.2. Setup (Calibration Setup)................................................................................................................ 16
7.1.1.2.1. Calibration Adjustment ............................................................................................................... 17
7.1.1.2.2.Time...............................................................................................................................................19
7.1.1.2.3.Temperature...................................................................................................................................21
7.1.2. Units of Measure Selection................................................................................................................ 22
7.1.3.Mode configuration............................................................................................................................. 24
(A) Animal Weighing......................................................................................................................... 25
(B) Parts Counting...............................................................................................................................28
(C) Check Weighing........................................................................................................................... 29
(D) Display Hold.................................................................................................................................30
(F) Totalizing...................................................................................................................................... 33
7.1.4.Automatic Zero Setting Mechanism (AZSM).....................................................................................35
7.1.5.GLP Print and Data Configuration...................................................................................................... 37
7.1.6.Printing Data and RS232 Communication Port Configurations..........................................................40
7.1.6.1. Baud and Parity............................................................................................................................... 40
7.1.6.2. Torbal Printer Configuration (RXP-4))........................................................................................... 43
7.1.6.3. Printout Receipts............................................................................................................................. 44
7.1.7.Time and Date..................................................................................................................................... 46
7.1.8.Reset (restoring default factory settings).............................................................................................48
8. Function Menu: Using the Modes and Functions.............................................................................................. 50
(A) Animal Weighing...................................................................................................................................51
(B) Parts Counting........................................................................................................................................54
(C) Check Weighing....................................................................................................................................58
(D) Display Hold.......................................................................................................................................... 61
(E) Percent Weighing................................................................................................................................... 63
(F) Totalizing................................................................................................................................................66
9. Technical Information........................................................................................................................................69
10. Common Errors and Troubleshooting............................................................................................................. 70
11. Maintenance.....................................................................................................................................................71
12. Accessories ..................................................................................................................................................... 71
13. Replacement Parts............................................................................................................................................72
14. Limited Warranty.............................................................................................................................................72
2
Instruction Manual
a
Important handling Cautions and Warnings
Always handle your scale with care. Damage caused by improper handling is not covered under the scale’s warranty.
Never drop or throw any articles onto the scale’s pan or onto any other parts of the unit!
DO NOT let the scale fall or drop from its tabletop surface!
When moving the scale do not press or apply force onto the scales pan!
DO NOT pass or handle liquids directly over the scale to avoid spillage and liquid damage!
Instruction Manual
C
CAUTIONARY NOTES AND PRECAUTIONS
The TORBAL Precision scale is a sensitive and a delicate instrument. Always handle your scale with care.
The correct location and proper environment affect the accuracy of your TORBAL Precision Scale.
The best location for your scale:
1.Stable, horizontal, vibration-free surface
2.Away from direct sunlight
3.Not exposed to high temperature variations
4.Away from direct drafts
5.The best location is on a stable bench away from drafts, doors, windows, radiators and air conditioner vents.
CAUTION:
WARNING:
o The scale is designed for indoor use only. o Do not operate the scale in hazardous areas or under dangerous conditions. o Do not use the scale in locations subject to high humidity or dust levels. o Do not connect cables in ways other than those mentioned in this manual. o Set the scale on a firm, stable, horizontal surface. o Never stand on or lean on this product. Equipment may fall or collapse, causing
breakage and possible injury.
o Before moving the product, unplug it and unplug all cables connected to it. o When storing, transporting or returning the scale for service, use the original
packaging.
a. Never attempt to repair, disassemble or modify the scale. Tampering with the scale
may result in injury and cause greater damage to the equipment.
b. Never swap the pan, pan base, or any other parts of the scale with pans or parts from other units. Pans and all components are uniquely assigned to each unit. c. Be sure to use the specified power source. d. Do not allow foreign matter to fall into the scale. e. If water or other liquids spill into the scale, unplug the power cord immediately and
contact technical support
TORBAL Precision scale is a sensitive and a delicate instrument. Always handle your scale with care.
3
CHAPTE R 1.

SPECIFICATION

AGC100 AGC200 AGC300 AGC500 AGC1000 AGC2000 AGC3000 AGC4000
Capacity
Readability (d)
Repeatability (Standard
Divination)
Linearity
Stabilization time (typical)
Tare Range
Accuracy Class
Automatic Internal
Calibration
Time Configuration
Temperature Configuration
Automatic Calibration
Adjustment Masses (g)
Pan Dimension
-100g -200g -300g -500g -1000g -2000g -3000g -4000g
20,50,
Instruction Manual - TORBAL AGC Series
Model
100g 200g 300g 500g 1000g 2000g 3000g 4000g
0.001 0.01
0.001g 0.01g
+/- 0.002g +/- 0.02g
Approx. 3 sec Approx. 3 sec
II
YES
1h, 2,h, 3h, 4h, 5h, 6h
0.5°C, 1.0°C, 1.5°C, 2.0°C
100
50, 100,
200
100, 200,
300
120mm 165mm x 165mm
100, 200,
500
200, 500,
1000
500, 1000,
2000
1000, 2000,
3000
1000, 2000,
4000
Draft Shield
Scale Dimension 216 x 345 x 90 mm
Operating Temperature
RS232 Port
Power Supply
Display Type
Display Size
Application Modes
GLP YES
Weighing Units Grams, Kilograms, Carats, Pounds, Newton, Grains, Ounces, Ounces Troy, Pennyweight
Scale’s Net Weight (lb/kg) 5kg / 11lbs
Parts
Counting
Percent
Weighing
Animal
Weighing
Totalizing
Display
Hold
Sample Size
Min Piece
Weigh
Min Reference
Weight
% Resolution
Displayed
(Reference
Weight)
Tine Intervals
Modes (Min
Weight)
Modes (Min
Weight)
Modes (Min
Weight)
Weighing, Parts Counting, Animal Weighing, Check Weighing, Display Hold, Percent Weighing,
0.1% (100mg to 7g), 0.01% (7g to 70g)
Manual, Auto (10mg), Semi-Auto (10mg) Manual, Auto (100mg), Semi-Auto (100mg)
Manual, Auto (10mg), Semi-Auto (10mg) Manual, Auto (100mg), Semi-Auto (100mg)
Manual, Auto (10mg), Semi-Auto (10mg) Manual, Auto (100mg), Semi-Auto (100mg)
Optional Not Required
+15°C to +30°C
Bidirectional
Input: 120VAC 60Hz 15W Output: 12VDC 500mA
Seven Segment LCD
30mm x 100mm
Totalizing
1 through 100
30mg 300mg
100mg 1g
0.1% (1g to 70g), 0.01% (70g to 700g)
0.001% (> 70g)
3sec.,5sec., 10sec., 15sec.
0.001% (> 700g)
4
Instruction Manual - TORBAL AGC Series

KEYS AND DISPLAY INDICATORS

Key Primary Function Secondary Function
Power On and Power Off -
T (YES)
0 (NO)
Menu
Function
P
C
Display Indicator Description Explanation
Tare – used to tare the weighing pan Enter and YES (Accept) – used to enter or accept commands
Zero – used to zero the scale NO (Reject) – used to reject commands or change values
Menu – used to access the main menu Recall – in Totalizing this key is used to recall the Grand Total
Add / Start / Lock – In Totalizing (Manual) this key is used to
Function – used to select an application
mode
Print – used to initiate printing
Clear / Cancel – used to clear or cancel
a function
add a result to the total. In Animal Weighing (Manual) this key is
used to initiate the animal weighing process. In Display Hold
(Manual) this key is used to lock the weighing result on the
display.
Decimal – used to enter a decimal when assigning limit values in
check-weighing
Exit- used to exit submenus
OFF Power Off The scale is turned OFF and in standby mode.
AUT
AZSM (Automatic Zero Setting
Mechanism)
NET Net Result
AZSM (Automatic Zero Setting Mechanism) is active
and the scale maintains a “center of zero” condition
within +/- 6d.
A tare was taken and the scale subtracted the tare
weight from the gross weight to obtain the net weight.
0 Zero The scale is maintaining a “center of zero” condition.
O
Current setting Indicates enabled functions or settings
MODE Main Menu Indicates Main Menu functions
Stability Indicator
MIN Minimum
MAX Maximum
The weighing result has stabilized and an accurate
reading may be taken.
In check weighing MIN indicates the weighing result
is below the selected under limit.
In check weighing MAX indicates that the weighing
result is above the over limit.
In check weighing OK indicates that the weighing
OK Accept
result is between the selected limits and the result is
acceptable.
TOTAL Total Result
pcs Pieces
% Percent
In totalizing this indicator signals that the displayed
result is a total sum.
In parts count this indicator shows that the result is a
piece count.
In percent weighing this indicator shows that the result
is a percentage.
5
Instruction Manual - TORBAL AGC Series
CHAPTE R 3.

COMMANDS AND ABBREVIATIONS

Command or Abbreviation Description
- - - - - ­CAL
TARING
LOAD XXX g
C
CAL FIN
AW – X
CONT SAPL ON SPL – XX
FILL
UNDER
OVER
ACCEPT
TARE
SAMPLE
READY
YES-NO
PRINT
COUNT FINISH TO START
COUNT PRESS C
The scale is taring before the calibration process.
Place a calibration weight on the pan to begin calibration.
In animal weight this command represents the average weighing time.
In parts counting, taring a container is required before continuing the counting transaction.
In parts counting or percent weighing this command indicates to place the sample on the pan.
In parts counting this command represents the current sample size setting.
In check weighing this command represents the under limit and indicates under result.
In check weighing this command represents the over limit and indicates an over result.
In check weighing this command indicates that the weighing result is within the limit range.
In percent weighing this command indicates the weight of a stored reference weight.
In percent weighing this command indicates that the scale is ready for percent weighing based
In totalizing this command indicates that a confirmation is required before the Totals are
In parts counting this command indicates that parts counting is finished.
Taring or re-zeroing in progress
The scale is in the calibration mode.
Calibration in progress
Calibration procedure is finished.
In parts counting, begin counting.
In check weighing, tare before continuing.
on a stored reference weight.
cleared.
Printing in progress
6
Instruction Manual - TORBAL AGC Series
CHAPTE R 4.

PARTS DESCRIPTION

Front View Rear View
Pan Pan Base AC Adapter
RS232 Serial Port Service Switch Level Indicator AC Adapter Socket
7
Instruction Manual - TORBAL AGC Series
CHAPTE R 5.

UNPACKING THE SCALE AND GETTING STARTED

1. Carefully remove the scale, pan, and all of its components
from packaging. Place them on a stable base where the scale
will not be affected by any mechanical vibrations or air
movements.
2. After removing the pan base and the pan from their packaging, carefully install the pan base onto the
scale by seating it on the pan support located in the middle of the scale. Once the pan base has been
installed, carefully place the pan on the base as shown below.
8
Instruction Manual - TORBAL AGC Series
3. Once the pan has been installed, level the scale by adjusting the rear feet until the level indicator shows the “air bubble” is in the center position of the sight glass. The level indicator is located on the rear left side of the scale.
4. After leveling the scale, plug the AC adapter into the AC adapter socket on the rear of the scale.
5. When the AC adapter is plugged into the wall outlet, the scale will turn “ON” automatically, initialize itself, perform automatic internal calibration and enter weighing mode.
6. To put the scale into standby mode, leave the AC adapter plugged into both the scale and the wall outlet and press the power “OFF” button ( ). OFF will light up on the upper left corner of the display, signaling the scale is in standby mode.
9
Instruction Manual - TORBAL AGC Series
CHAPTE R 6.

WEIGHING

1. To begin weighing, press the power button ( ) to turn the scale on. The scale will initialize, perform automatic internal calibration and enter the weighing mode. The scale is ready to begin weighing as soon as the stabilization ( ) and center of zero ( ) indicators appear on the display
2. When weighing, always place the mass in the middle of the pan. The weighed result may be taken when the stabilization indicator appears on the display.

6.1.1.Zeroing the scale

1. The scale may be re-zeroed manually to obtain a new “center of zero.” To re-zero the scale manually, make sure the weight and the stabilization indicator are shown on the display.
2. Re-zero the scale by pressing the zero key 0.
3. The display will show a dotted line, which indicates the scale is re-zeroing.
10
Instruction Manual - TORBAL AGC Series
4. When it has finished re-zeroing, the scale will return to weighing mode and the display will indicate
0.000g. A new center of zero has been set, and the scale is ready for use.

6.1.2.Taring

1. If a container is used for weighing, it may be tared. Taring subtracts the weight of the container from the gross weight to obtain the net weight.
2. To tare the weighing container, place it in the middle of the pan. The container’s weight will be shown on the display.
3. Once the stabilization indicator appears on the display, the container is ready to be tared. To tare the container, press the “T” button. The display will show a dotted line, indicating that the scale has begun the taring process,
4. When finished taring, the balance will return to weighing mode. The display will indicate 0.000g, and NET indicator will be shown on the display signaling the next weight taken is a net result.
11
Instruction Manual - TORBAL AGC Series

6.1.3.Clearing a Tare

1. To clear the tare, remove the tared object along with the net weight from the pan. The scale will then display a negative net tare result.
2. To clear the tare, press the “T” button.
3. The display will show dashed lines indicating the tare is clearing.
4. When it has finished clearing the tare, the scale will return to weighing mode.
12
Instruction Manual - TORBAL AGC Series
CHAPTE R 7.

MAIN MENU

Main Menu is used to configure the scale and its weighing modes. Before entering the menu, complete
and clear any active function by pressing the “C” key. To enter the Main Menu press the “MENU” key. Once
the “MENU” key is pressed the scale will display available Menu options. Commands: CALIB, UNIT, MODE,
AUTO, GLP,PRINT, SET TD and RESET will flash sequentially. To select and enter a desired menu option
press the “T” (YES) key while that menu option is displayed.
13
Note: Scrolling through the options may be accelerated by manually pressing the 0 (NO) key.

7.1.1.Calibration

When the scale is initially installed, it will automatically calibrate at startup to ensure
accuracy of weighing results. Calibration should be performed periodically or whenever the scale is
moved. You can also configure the scale to calibrate automatically with time and temperature variations.
To calibrate the scale and configure automatic calibration settings follow the steps below.
Instruction Manual - TORBAL AGC Series

7.1.1.1.Start (Initializing Automatic Calibration)

To calibrate the scale, follow the steps below:
1. Press “MENU” to enter the Main Menu.
2. When CALIB is displayed, press “T” (YES).
3. The scale will display commands Start and Setup sequentially. Press , “T” , when Start is displayed.
4. The scale will display CAL, indicating that the calibration mode has been entered.
14
Instruction Manual - TORBAL AGC Series
5. Before calibration begins, the scale will perform automatic tare. TARING will be displayed during
auto tare.
6. Once the scale tares, command “LOAD C” will be displayed, followed by “C” indicating that
automatic calibration is in progress.
7. Calibration can last from 30 to 50 seconds, when it is finished the scale will automatically return to
the weighing mode.

7.1.1.2. Setup (Calibration Setup)

To enter the automatic calibration setup menu, follow the steps below:
1. Press “Menu” to enter the Main Menu
15
Instruction Manual - TORBAL AGC Series
2. When CALIB is displayed, press “T” (YES).
3. The scale will display commands Start and Setup sequentially. Press , “T” , when Setup is
displayed.
4. Commands CALIB (Calibration Adjustment), TIME (Time Configuration), and TEMP
(Temperature Configuration) will begin to scroll sequentially.
16
Instruction Manual - TORBAL AGC Series

7.1.1.2.1. Calibration Adjustment

The Calibration Adjustment function allows the internal calibration mass value to be adjusted
with an external calibration weight that has a traceable class certification. To perform calibration
adjustments follow the steps below:
1. Press the “T” (YES) key when “CALIB” is displayed in the calibration setup menu.
2. The scale will display CAL, indicating that the calibration adjustment mode has been entered.
3. Before calibration adjustment can begin, the scale will perform automatic tare. TARING will be
displayed during auto tare.
4. Once the scale has been tared the current calibration mass setting will be displayed. To change the
17
calibration mass, press the 0 (NO) key. After making the selection press the “T” (Yes) key to
accept and confirm the setting.
Instruction Manual - TORBAL AGC Series
To Change To Accept
5. Command LOAD will be displayed. At any time during the LOAD command, place a single
calibration weight in the middle of the pan.
6. Command “C” will be displayed indicating that the scale is recording a new calibration mass value.
7. Once the new calibration value is stored the command “UNLOAD” will be displayed indicating to
remove the calibration weight from the pan.
18
Instruction Manual - TORBAL AGC Series
8. Once the weight is removed the scale will continue with calibration adjustment. CAL will be
displayed during the remainder of the adjustment process. Once finished the scale will automatically
return to the weighing mode.

7.1.1.2.2.Time

Internal calibration can be configured to calibrate automatically within a change of time. To set
a calibration time interval, follow the steps below.
1. Press the “T” (YES) key when “TIME” is displayed in the calibration setup menu.
2. “t-0” and the current time interval setting will begin to scroll on the display sequentially. To
disable time interval calibration press the “T” (YES) key when “t-0” is displayed. To enable time
calibration press the “T” (YES) key when the default time setting is displayed.
19
To Disable To Enable
Instruction Manual - TORBAL AGC Series
3. A dash will appear on the display indicating to enter a desired time interval. The time interval
can be set between 1, 2, 3, 4, 5, and 6 hours (i.e. if set to 3, calibration will be performed once
every 3 hours).
4. To set a calibration time interval press the 0 (NO) key until the desired setting is displayed.
5. When the desired setting is displayed press the “Function” key to accept and save the selection.
The scale will return to the Calibration Menu. To exit the Menu, press the C key.

7.1.1.2.3.Temperature

Internal calibration can be configured to calibrate automatically within a change of
temperature. To set a calibration temperature interval, follow the steps below
20
Instruction Manual - TORBAL AGC Series
1. Press the “T” (YES) key when “TEMP” is displayed in the calibration setup menu.
2. “°C - 0” and the current temperature interval setting will begin to scroll on the display
sequentially. To disable temperature interval calibration press the “T” (YES) key when “°C - 0”
is displayed. To enable temperature calibration press the “T” (YES) key when the temperature
setting is displayed.
To Disable To Enable
3. The temperature interval can be set to 0.5°C, 1.0 °C, 1.5 °C, 2.0 °C (i.e. if set to 2.0 °C,
calibration will be performed every time the temperature changes by 2.0 °C).
4. To set a calibration temperature interval press the 0 (NO) key until the desired setting is
displayed.
21
Instruction Manual - TORBAL AGC Series
5. When the desired setting is displayed press the “Function” key to accept and save the selection.
The scale will return to the Calibration Menu. To exit the Menu, press the C key.

7.1.2. Units of Measure Selection

TORBAL AGC series scales can operate in nine different units of measure: grams (g),
kilograms (kg), carats (ct), pounds (lb), Newton (n), grains (gn), ounces (oz), ounces troy (oz-t), and
pennyweights (dwt).
By factory default the scale is set to weigh in grams (g). To select a different unit, follow the
steps below.
1. Enter the Main Menu by pressing the “MENU” key.
22
Instruction Manual - TORBAL AGC Series
2. When command UNIT is displayed press the “T” (Yes) key.
3. The scale will display available units of measure sequentially. When the desired unit is displayed,
press the “T” (Yes) key to make the selection.
Example (Selecting Ounces)
4. To exit the Main Menu, use the “C” key.
23