Read this Operation Manual carefully before using the SP-100 for proper and safe operation.
If you come up any questions about this instrument or the Manual, ask your Tomey representative or local distributor.
yy
yy
y Do not use this instrument by any procedures other than those specified in this
Manual.
yy
yy
y Only well-trained or skilled personnel is allowed to operate this instrument.
yy
yy
y Keep the Operation Manual in a place where you can easily access while operating
the instrument.
HANDY PACHYMETER
SP-100
WARNINGS
DO NOT USE THE INSTRUMENT WITH THE OUTER COVER AS LEFT
OPEN OR REMOVED, otherwise you may be exposed to direct high voltage.
Always use the sterilized or disinfected Pachymeter Probes for measurement.
NEVER USE the Probe if it may be subjected to any visible damage on its tip.
Such use may not only cause an incorrect measurement, but also damage the
cornea.
This instrument is designed exclusively for ophthalmic use.
DO NOT USE the instrument for any purpose other than ophthalmic use.
DO NOT connect the terminals of the instrument with any cable other than
those specified in the Manual, otherwise the instrument may be damaged.
SIGNIFICANT WARNINGS
i
HOW TO USE THIS OPERATION MANUAL
THE COMPOSITION OF THE OPERATION MANUAL
The following items are provided for your efficient and effective use of
this instrument.
The Operation Manual is composed of the following parts.
1. PRIOR TO USE
Precaution and confirmation items related to the installation and us-
age of the instrument.
2. NAMES AND FUNCTIONS OF THE COMONENTS
Names and functions of the components of the instrument
3. OPERATING PROCEDURES
Vital information, required for installing and using the instrument
4. TECHNICAL INFORMATION
Useful technical information, which may be convenient for your use
5. MAINENANCE AND INSPECTION
Routine replacement, maintenance, and inspection, including replac-
ing spare parts
6. TROUBLESHOOTING
Countermeasures for troubles
7. SPARE PARTS AND OPTIONAL PARTS
Accessories and optional parts
8. SPECIFICATIONS
Specifications for the instrument
9. INDEX
ii
Refer to the index as a guide, if necessary.
HOW TO USE THIS OPERATION MANUAL
SYMBOLS USED IN THE MANUAL
The symbols, such as "DANGER", "WARNING", "CAUTION" and "NOTE" used in this
Operation Manual, represent the following messages.
Precaution for an operation state that, if unheeded, will cause a haz-
ardous situation where there is an imminent danger of serious injury
or death.
Precaution for an operation state that, if unheeded, may cause a haz-
ardous situation where there is the possibility of serious injuries or
death.
An operational precaution that, if unheeded, may cause or lead to
minor or moderate injuries or property damages.
Special precaution that is related either directly or indirectly to per-
sonal safety or to property protection.
SYMBOLS USED IN THE MANUAL
iii
Contents
1.PRIOR TO USE ............................................................................................................. 1-1
7.SPARE PARTS AND OPTIONAL PARTS ....................................................................... 7-1
7.1Spare parts ..................................................................................................................................... 7-1
7.2Optional parts ................................................................................................................................. 7-1
(with the protective cap and the case) ......................... 1
z Pachymeter test piece .................................................... 1
z AC adapter ................................................................... 1
z Printer paper roll (1 roll in the instrument) ....................... 3
z Serial communication terminal protective cap .................. 1
z Operation Manual (this manual) ...................................... 1
z Carrying Bag ................................................................. 1
z Inner Case ..................................................................... 1
z Quick Guide .................................................................. 1
1.2 Unpacking
1-3
1.3Symbols, used in this manual
The terms and symbols, used for the instrument, shall imply the following means.
PROBEPachymeter probe terminal
Connects the Probe to this terminal.
Power Switch
Turns main power ON and OFF.
Contrast adjusting volume
Adjusts the contrast of the monitor
Serial communication terminal
Various data is sent and received through the external
by using the serial communication cable.
Maintenance switch
This switch should be used for maintenance purpose only
Do not touch this switch for changing.
AC adapter terminal
Connects the AC adapter, included in the box.
1-4
1.3 Symbols used in this manual
1.4Outline of Measurement Principle
After taking measurement, the oscillator is not emitting
ultrasound waves.
The SP-100 is designed for pachymeter or corneal thickness measurement by
using ultrasound transducer, built in Pachymeter probe.
z SP-100 emits the ultrasound from the Pachymeter probe and the
probe also receives the ultrasound reflected by surface between
corneal posterior face and anterior chamber.
z SP-100 measures the period of time from emitting ultrasound to
receiving the same ultrasound, reflected by corneal posterior part.
Applying a calculation formula shown below, corneal thickness
can be calculated.
V · t
L = ———
z This instrument is powered by rechargeable battery or AC power
adapter. Even when the battery power is running short, SP-100
can perform measurements, while charging the battery, by plug-
ging to AC power adapter.
z This instrument is designed to take corneal thickness measure-
ment automatically, when the probe is applied to cornea properly.
2
Where,L: Corneal Thickness
V: Converted sound velocity
(Ultrasound Velocity through cornea)
T: Measured time (Ultrasound transmission
time between emitting and receiving)
z This instrument is capable of sending and receiving patient infor-
mation and measurement data between the external equipment
through the serial communication port. (Additional cable must be
used, when using this function)
1.4 Outline of Measurement Principle
1-5
This page is left intentionary blank.
1-6
2.NAMES AND FUNCTIONS OF THE COMPONENTS
2.1Front and Right side of Main Unit
AA
A SCREEN
AA
Measurement data and other information can be
displayed.
B B
B OPERATING PANEL
B B
The keys are located on the panel with various
functions.
CC
C PACHYMETER PROBE TERMINAL
CC
Pachymeter Probe should be connected to this
terminal.
DD
D SERIAL COMMUNICATION TERMINAL
DD
Serial Communication Cable can be connected
for data communication.
EE
E AC ADAPTER TERMINAL
EE
AC Adapter can be connected for supplying
power for operation and charging battery.
FF
F PROBE HOLDER
FF
Pachymeter Probe can be hold and kept as stable
position. Use the holder, while the probe is not
in use.
GG
G BUILT-IN THERMAL PRINTER
GG
Patient and Measurement information can be
printed by selected Standard or Simple mode.
2.2Rear and left side of Main Unit
AA
A INSTRUMENT STAND
AA
The main unit stands on desk tilted for easy operation and monitoring.
BB
B CONTRAST VOLUME DIAL
BB
LCD contrast can be adjusted.
CC
C MAINTENANCE SWITCH
CC
This Switch is for maintenance use only. DO
NOT change any setting this switch, or the unit
is damaged.
2.1 Front and Right side of Main Unit2.2 Rear and left side of Main Unit
2-1
2.3Operation Keys
General key assignments are explained in this part. However, other functions
are assigned on the same key, depending on the screen. Read each part of the
manual carefully in order to understand the key operation completely.
(Example)
AA
A FUNCTION KEYS
AA
Each key works as assigned function shown on
the bottom of the screen. Four boxes from left
to right on screen indicate each function of F1,
F2, F3 and F4 keys.
BB
B PRINT KEY
BB
Patient and Measurement Data can be printed
by built-in thermal paper. Please refer to [3.6
Printout] for details.
CC
WW
C
CC
DD
D CHARGE INDICATOR
DD
EE
E POWER KEY
EE
XX
W,
X KEYS
WW
XX
Data selection, Patient ID & Name can be typed
in by using the keys. Press both keys simultaneously to turn LED ON and OFF. (When the
unit is only powered by battry and no operation
for one minute, it automatically turns LED off)
The light keeps turning ON, while charging the
internal battery. When it turns OFF, charging
battery is completed.
The main power turns ON and OFF. When the
unit is only powered by battery and no operation for three minutes, it automatically turns OFF
to save battery power. (Auto Power OFF)
2-2
(F-keys Example) Each key, on this page, is assigned its func-
tion as follows:
x F1 ........... R/L
x F2 ........... INDEX
x F3 ........... EDIT
x F4. .......... NEW
2.3 Operation Keys
3.OPERATING PROCEDURES
3.1Safety precautions
The Probe must always be disinfected before its use.
Never use the Probe, if there is any visible damage on its tip, which may
otherwise cause an incorrect measurement or damage the cornea.
This instrument is Ophthalmic Medical Device, designed for only skilled
operators.
-Do not use this instrument other than Ophthalmic applies.
Do not use any terminal other than specified in this Operation Manual,
which may otherwise cause to damage the instrument.
Do not leave the instrument plugged with the AC Adapter when the
battery is fully charged, otherwise, this may enshorten the lifetime of
built-in rechargeable battery.
When the probe is not in use, please place the probe cap and place the
probe to the holder to protect from getting damaged.
3.1 Safety precautions
3-1
3.2Preparation before operation
3.2.1Connection of accessories
a) Connection of Pachymeter Probe
Connector of the Probe should be inserted in speci-
fied direction. Insert the connector as deep as it
sounds "click".
Insert the connector A into the connecting terminal B
marked "PROBE" on the front side of Main Unit.
b) Connection of AC Adapter
3.2.2Disinfection of the probe
Insert the connector to the terminal completely
Insert AC adapter connector A into the connecting terminal B provided on the right side of the Unit.
Disinfect the probe with proper procedure, before its use. Please
3-2
refer to [5.2.2 Disinfecting Pachymeter Probe] for details.
3.2 Preparation before operation
3.2.3Optional components
Assembling of the Disposal Tip Pachymeter Probe and alternative Probe shapes
The Disposal Type Pachymeter Probe (supplied as
an optional component) consists of the inner probe,
case, fixed ring, and disposal tip.
The Probe can be changed in two alternative shapes:
(A) with 45 shape and (B) in straight shape, de-
pending on the application convenience.
<How to assemble>
1) Cover the inner probe A with a pair of cases
B to make the inner probe clipped in between
and fix it with the rubber rings C.
2) Apply a small amount of methylcellulose (such
as Dry Eye Gel) on the tip of the inner probe
A.
3) Then securely insert the disposal tip D not to
cause air bubbles inside of the tip.
<How to assemble the Probe in alternative shape>
1) Disassemble the disposal tip D, fixed ring C,
and the case B in order.
2) Assemble the case B to be changed and next
fix it with the rings C.
3) Next apply a small amount of methylcellulose
to the tip of the inner probe.
4) Then securely insert the disposal tip D not to
cause air bubbles inside of the tip.
3.2 Preparation before operation
3-3
3.3Adjustment and Setting
3.3.1Main power on and screen adjustment
Confirm the following items before turning the power on.
xx
x
Make sure the battery is sufficiently charged. If not, con-
xx
nect the AC adapter for power supply. (With battery opera-
tion and 3 minutes of non operating time, the unit goes
into Auto Power off mode and measured data will be lost)
xx
xMake sure the Probe has been connected.
xx
xx
xMake sure the tip of the Probe is dry.
xx
Tilt the LCD screen on the proper position in order to get bet-
ter view from the front
Ԙ
(
(
(
(
24+06
*#0&;
2#%*;/'6'452
.+)*6
%*#4)'
219'4
1) Press the power key A to start operation. The
instrument will automatically calibrate the con-
nected Probe. If "PROBE ERROR!" is displayed
on the screen, check if the tip of the Probe is dry.
(Refer to [3.3.2 Startup Display] for details)
(Fig. 1)
(Fig. 2)
2) Adjust the LCD contrast, according to the
brightness of the clinical room by using contrast adjusting dial B on the left side of the
unit.
When the room is dark, press the "W " and "X"
keys C simultaneously to turn the lighting LED
on.
3)Adjust Instrument Stand D on back of the unit to
tilt it to get better screen view.
3-4
(Fig. 3)
3.3 Adjustment and Setting
3.3.2Startup Display
(Fig. 1)
1) The startup screen is displayed immediately af-
ter the power is on. The software version is seen
on the screen.
(Fig. 1)
2) The ultrasound velocity, set for measurement, is
(Fig. 2)
displayed.
(Fig. 2)
3) The unit calibrates the probe automatically.
(Fig. 3)
(Fig. 3)
3.3 Adjustment and Setting
3-5
(Fig. 4)
4) When the calibration has been completed prop-
erly, measurement Pachy screen (initial screen)
is displayed.
(Fig. 4)
5) When the Probe is not calibrated properly,
"PROBE ERROR!" will be displayed. Press F4
key A "OK" (Fig. 5) to show CALIBRATION
screen (Fig. 6).
(Fig. 5)
(Fig. 6)
A
6) Check the connection of the Probe and the
Probe tip, then press the "W " to recalibrate the
Probe.
When error appears in the screen, go to "6.
TROUBLE-SHOOTING" for the solution.
When "ERROR" repeatedly appears, contact
3-6
TOMEY representative.
3.3 Adjustment and Setting
3.3.3Screen Contents
a) Pachy screen
A The eye to be measured
Either left or right eye, to be measured, is selected on this part of the screen.
B Measurement data
The data, taken by the unit will be listed.
C Average
Average value of the listed data will be displayed.
D Battery Life Indicator
Remaining power in the built-in battery
appears. (Two Steps)
E Type of listed data
Average, Minimum or Latest options can be selected as a type of listed data
F R/L (F1 key)
Left or right eye for measurement is switched over.
G INDEX (F2 key)
Switch screen to INDEX, to input Patient ID and name.
H EDIT (F3 key)
Switch screen to EDIT, to edit measured data
I NEW (F4 key)
To measure new patient, use this key to clear Patient data and measured readings.
3.3 Adjustment and Setting
3-7
b) Measurement screen
c) Edit screen
A The eye to be measured
Either left or right eye, to be measured, is selected on this part of the screen.
B Measurement data
The data, taken by the unit will be listed.
While taking measurement, the screen automatically
changes and shows large size measuring data.
A Cursor
Selected data is highlighted on the list. Other
data is selected by using "W" or "X" keys.
B Pachy (F1 key)
Press this key to change screen to Pachy mode.
C CLR (F2 key)
Press this key and keep holding it to clear all
measurement data.
D DEL (F3 key)
Press this key and keep holding it to clear only
highlighted measurement data
E MENU (F4 key)
Press this key to change into MENU screen.
3-8
3.3 Adjustment and Setting
3.3.4Measuring conditions
a) Input Patient ID and name
The Patient ID and name are input in the "INDEX" screen. 14 letters can be
The following measuring conditions have been set, when turning
the power on.
Eye to be measured:Right eye (OD)
Corneal ultrasound velocity:Last value, when it is turned off.
Please refer to "3.9 Data Management with TOMEYLink
(Electronic Medical Record Support System)" to send Data to
TOMEYLink.
In case of using Data Communication, Patient ID must be
typed in. Otherwise the data cannot be sent.
(Fig. 1)
typed in either the Patient ID or Patient name.
1) Press F2 key A to change into INDEX Screen.
(Fig. 2)
2) Press F2 key C to change into ID input screen
to type in Patient ID. (Fig. 3)
Press F3 key D to change into Patient NAME
input screen to type in Patient Name. (Fig. 4)
(Fig. 2)
F1 key B to go back to Pachy Screen (Fig. 1).
3.3 Adjustment and Setting
3-9
3) Press "W" or "X" keys to move Cursor I to
the figures or letters, you wish to input. (Fig. 3)
Then, press SEL G to enter selected character.
(Fig. 3)
4) Press F1 key E to switch over input mode be-
tween figures and letters.
5) Press F2 key F to enter Back Space.
6)When Patient ID type in completed, press F4 key
H to go back to INDEX Screen (Fig. 1).
(Fig. 4)
3-10
7) Repeat the above procedure again to type in Pa-
tient name. (Fig.4)
3.3 Adjustment and Setting
b) The eye to be measured
c) Corneal ultrasound velocity
Select Right eye or Left eye, to be measured.
Press F1 key A to switch over Right and Left dis-
play.
d) Listed data on screen
Ultrasound Velocity can be changed by going into
setting Menu. Please refer to "3.7.3 c) corneal Ultra-
sound Velocity (VELOCITY)" for details.
One, out of three, data type A can be selected as
listed data type. The three options are as follows.
AVERAGE
The mean value of each data box is listed.
LATEST
The last measurement for each data box is listed.
MINIMUM
The minimum value of all measurements for
each data box is displayed.
Please refer to "3.7.3 d) Setting of the type of display
data (DATA SELECTION)" for details.
3.3 Adjustment and Setting
3-11
Loading...
+ 63 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.