“Teflon” is a registered trademark of E.I. du Pont de Nemours and Company.
“Watlow” and “EZ-ZONE” are registered trademarks of Watlow Electric Manufacturing Company.
“VICI is a registered trademark of Valco Instruments Co. Inc. and VICI AG.
“Kaowool” is a trademark of Morgan Thermal Ceramics, a business within the Morgan Ceramics Division of the The
Morgan Crucible Company plc.
“Dow Corning” is a registered trademark of Dow Corning Corporation.
All other trademarks are the property of Thermo Fisher Scientific Inc. and its subsidiaries.
Thermo Fisher Scientific (Thermo Fisher) makes every effort to ensure the accuracy and completeness of this manual.
However, we cannot be responsible for errors, omissions, or any loss of data as the result of errors or omissions.
Thermo Fisher reserves the right to make changes to the manual or improvements to the product at any time without
notice.
The material in this manual is proprietary and cannot be reproduced in any form without express written consent
from Thermo Fisher.
Page 4
Page 5
Revision History
Revision Level Date Comments
A 02-2004Initial release.
B 04-2005Applied standardized formatting, changes to analyzer menu,
stream setup, autocal setup menus; added “At Line Grab
Sample” menu and related items; edited Modbus register for
software v. 1.20.
C 04-2006Address change.
D 01-2007Company name change (ECO 5457).
E 04-2007Revised per ECO 5625.
F 08-2007Revised per ECO 5921.
G 10-2007Revised per ECO 6018.
H 03-2008Revised per ECO 6230.
J 04-2008Revised per ECO 6320.
K 06-2008Revised per ECO 6419.
M 04-2009Revised per ECO 6903.
N 05-2009Revised per ECO 6915.
P 06-2009Revised per ECO 6968.
R 04-2010Revised per ECO 7351.
S 11-2010Revised per ECO 7578.
T 01-2011Revised per ECO 7627.
U 04-2011Revised per ECO 7694.
V 01-2012Revised per ECO 7897.
W 02-2013Revised per ECO 8126.
X 01-2014Revised per ECO 8332.
Y 11-2014Revised per ECO 8545.
Z 06-2016Revised per ECO 8659.
AA 06-2016Revised per ECO 8667.
AB
01-2017
Revised per ECO 8995.
Thermo Fisher Scientific SOLA II User Guide v
Page 6
Page 7
Contents
Chapter 1
Chapter 2
Chapter 3
Chapter 4
Safety Information & Guidelines ................................................................... 1-1
Failure to follow appropriate safety procedures or inappropriate use of the
equipment described in this manual can lead to equipment damage or injury
to personnel.
Safety
Considerations
Safety Information & Guidelines
This chapter contains information that must be read and understood by all
persons installing, using, or maintaining this equipment.
Any person working with or on the equipment described in this manual is
required to evaluate all functions and operations for potential safety hazards
before commencing work. Appropriate precautions must be taken as
necessary to prevent potential damage to equipment or injury to personnel.
The information in this manual is designed to aid personnel to correctly and
safely install, operate, and maintain the system described; however, personnel
are still responsible for considering all actions and procedures for potential
hazards or conditions that may not have been anticipated in the written
procedures. If a procedure cannot be performed safely, it must not be
performed until appropriate actions can be taken to ensure the
safety of the equipment and personnel. The procedures in this manual
are not designed to replace or supersede required or common sense safety
practices. All safety warnings listed in any documentation applicable to
equipment and parts used in or with the system described in this manual
must be read and understood prior to working on or with any part of the
system.
Failure to perform the instructions and procedures in this manual
or other documents pertaining to this system correctly can result
in equipment malfunction, equipment damage, and/or injury to
personnel.
Thermo Fisher Scientific SOLA II User Guide 1-1
Page 14
Safety Information & Guidelines
Safety Summary
Safety Operating
Information
Safety Summary
The following admonitions are used throughout this manual to alert users to
potential hazards or important information. Failure to heed the warnings
and cautions in this manual can lead to injury or equipment
damage.
Warning Warnings notify users of procedures, practices, conditions, etc.
which may result in injury or death if not carefully observed or followed. The
triangular icons with warnings vary depending on the hazard. σ
Caution Cautions notify users of operating procedures, practices, conditions,
etc. which may result in equipment damage if not carefully observed or
followed. σ
Note Notes emphasize important or essential information or a statement of
company policy regarding an operating procedure, practice, condition, etc. σ
This section contains general safety and operating information applicable to
analytical systems, which must be understood by all persons installing, using,
or maintaining the analyzer system. This information is designed to aid
personnel in the safe installation, operation, and service of the analyzer and
sample systems. It is not designed to replace or limit appropriate safety
measures applicable to work performed by personnel. Any additional safety
and operating measures that are required must be determined by and
followed by personnel performing work on the system.
Caution Failure to heed the following information may lead to equipment
damage or injury to personnel. σ
Protective eyewear (glasses with side shields or goggles as appropriate) must
be worn when servicing any part of the analyzer or sample system. When
servicing the sample system, chemical resistant gloves appropriate for the
materials in the system must be worn. When servicing the hot analyzer oven,
internal components (e.g., detectors), or hot sample system components,
appropriate gloves must be worn. Heated components should be allowed to
cool before servicing if possible. Other appropriate equipment or clothing
must be used as required by the type of work performed.
1-2 SOLA II User GuideThermo Fisher Scientific
Page 15
Safety Information & Guidelines
Electrical Power
Electrical Power
Caution Ovens, internal components, and sample systems may be hot even
when power is not applied to the unit. Take appropriate precautions to
prevent injury resulting from contact with hot items. σ
All applicable regulations and procedures must be followed for the work
performed. Before beginning any work on the system, carefully consider all
the potential hazards and ensure that appropriate measures are taken to
prevent injury to personnel and damage to equipment.
The system uses AC power at 110 Vac (an optional step-down transformer is
available for 220 Vac). The AC power is converted to DC at several voltage
levels. Appropriate precautions must be taken to prevent sparks present in the
analyzer environment that may ignite combustible materials. Precautions must
also be taken to prevent electrical shock if the analyzer or sample system
enclosures are opened.
The AC power to the system must be free from noise, surges, sags, and spikes
for proper system operation. AC power circuit breakers and wiring must be
sized properly for the required current. All wiring installations must meet
applicable electrical codes.
The fuse is located on the analyzer terminal block.
Caution If it becomes necessary to replace the fuse, it must be replaced with
one of the same rating: Fuse, 3 A S/B (p/n TE-4510). σ
Warning Remove power prior to performing any work internal to the
instrument. An override is available for use in non-hazardous areas; however,
removal of components while the instrument is energized is not permitted. σ
Thermo Fisher Scientific SOLA II User Guide 1-3
Page 16
Page 17
Chapter 2
Function
Product Overview
The Thermo Scientific SOLA II sulfur online analyzer combines proven
detection technology, easy-to-use, menu-driven software, and advanced
diagnostics to offer unsurpassed flexibility and reliability. The instrument
offers field programmable ranges, high sensitivity, total sulfur measurement,
fast response time, linearity through all ranges, and low consumables.
Major components of the SOLA II include a sample injection valve, carrier gas
flow control system, mixing chamber, pyrolyzer, optional dryer, and a pulsed
ultraviolet fluorescence (PUVF) detector. Unlike the original SOLA instrument,
the PUVF is not a subassembly. The SOLA II is a single assembly with one
software program and one user interface.
The sample injection valve periodically transfers a small amount of sample
(approximately 1.0 µL) into an air carrier gas. The air/sample mixture passes
through the mixing chamber to ensure complete mixing and then flows to the
pyrolyzer. The pyrolyzer combusts all sample components to SO2, CO2, and
H2O at approximately 1100°C (2012°F). The optional dryer (application
dependent) removes water from the sample that is produced during
combustion. The PUVF detector accurately measures the amount of SO2
produced during combustion of the sample.
An appropriate sample conditioning system is mandatory for proper
functioning. The sample conditioning system should:
λ Regulate sample pressure and temperature.
λ Provide filtration to at least a 0.5-micron particle size. Staged filtration is
recommended (e.g. going from 10 micron to 5 micron to 0.5 micron). The
final filter should contain a hydrophobic element to remove undissolved
water.
λ Ensure that a representative sample is transported to the analyzer in the
desired time.
λ Maintain the sample in a single phase.
Thermo Fisher Scientific SOLA II User Guide 2-1
Page 18
Product Overview
Total Sulfur
Measurement
Total Sulfur Measurement
Note Liquid samples with high vapor temperatures such as naphthas and
gasolines will require sample backpressure regulation +0 25–35 psig. σ
λ Remove undissolved water.
Sample should be delivered to the sample conditioning system using a sample
probe. The sample probe should be designed and fabricated so that sample is
extracted from near the center of the process pipe, preventing the
unnecessary introduction of pipe scale and other particulate that tend to
accumulate along the process pipe walls.
Figure 2–1. Functional block diagram
Total sulfur measurement is based upon the precise measurement of the SO2
concentration produced from a wide variety of compounds containing sulfur,
such as H2S, COS, methyl mercaptan, benzothiophenes, dibenzothiophenes,
sulfides, disulfides, and thiols. Consider the measurement of a sample of
diesel, gasoline, or another common petroleum fraction such as naphtha for
liquid phase samples. For liquid phase samples, the analyzer periodically
injects a very small quantity of sample (1.0 µL) of the petroleum fraction into
a hot oven (110°C to 220°C / 230°F to 428°F) where it is vaporized and mixed
with air. Analysis of gas phase samples requires sample sizes of 0.1 to 1.0 cm3.
After thoroughly mixing with air, the sample enters the pyrolyzer, where all
components of the sample are combusted at 1100°C (2012°F) to CO2, H2O, or
SO2. The quantity of SO2 formed during the combustion process is directly
proportional to the total sulfur content of the petroleum fraction.
The analyzer is equipped with a PUVF detector that accurately measures the
quantity of SO2 formed during the combustion process. The SO2 molecules
2-2 SOLA II User GuideThermo Fisher Scientific
Page 19
Product Overview
Total Sulfur Measurement
enter the cell of the PUVF detector where they are exposed to ultraviolet (UV)
light. Absorption of UV light by SO2 molecules results in an excited state of
the SO2 molecules. The excited state SO2 molecules exist at a higher energy
state (due to absorption of energy in the form of UV light) and rapidly relax
to their original energy level or ground state through the emission of light.
This process is called fluorescence. The intensity of the light emitted by SO2
fluorescence is directly proportional to the SO2 concentration. Pulsing the UV
light allows more energy (UV light) to be delivered to the sample, increasing
the fluorescence intensity for a given SO2 concentration. This results in
increased overall sensitivity of the instrument.
At the PUVF detector, it is important to ensure that the measured light is
representative of only the SO2 concentration and not some other species. SO2
molecules emit light through fluorescence at a specific wavelength, and the
PUVF detector utilizes bandpass filters to ensure that only light from SO2 is
measured. Once the proper wavelength of light is selected, its intensity is
measure by a photomultiplier tube (PMT). The PMT converts light energy to
electrical energy through manipulation of the photoelectric effect. The
electrical signal generated by the PMT is finally processed by the analyzer
electronics and software to determine and report the concentration of total
sulfur in the liquid petroleum fraction or gas sample.
The analyzer can be configured to report total sulfur in parts per million
(ppm), parts per billion (ppb), or milligrams per liter (mg/L).
λ When using the ppm or ppb units of measure, you can choose to calibrate
the analyzer in terms of ppm (w/w), ppm (v/v), ppm (w/v), ppb (w/w),
ppb (v/v), or ppb (w/v).
λ When using density compensation and the ppm or ppb units are selected,
the display shows ppm/weight or ppb/weight.
Units of concentration calculated on a weight/weight basis are sensitive to
sample density. If the density of the calibration standard is significantly
different from the density of the sample, a density correction should be
applied.
λ If the sample density varies significantly, a density correction is also
necessary when using concentration units of ppm (w/w) or ppb (w/w).
λ If using mg/L, ppm (w/v), ppm (v/v), ppb (w/v), or ppb (v/v), a density
correction is not required.
Thermo Fisher Scientific SOLA II User Guide 2-3
Page 20
Product Overview
Principle of
Operation
Principle of Operation
The detector is based on the principle that SO2 molecules absorb UV light and
become excited at one wavelength, then decay to a lower energy state
emitting UV light at a different wavelength. Specifically,
The sample inlet bulkhead draws the sample into the analyzer. The sample is
mixed with air and passes through a pyrolyzer furnace that oxidizes the sulfur
molecules in the sample to produce SO2. The sample then flows into the
fluorescence chamber where pulsating UV light excites the SO2 molecules. The
condensing lens focuses the pulsating UV light onto a mirror assembly. The
mirror assembly contains four selective mirrors that reflect only the
wavelengths that excite SO
molecules.
2
As the excited SO
molecules decay to lower energy states, they emit UV light
2
that is proportional to the total sulfur concentration in the sample. The
bandpass filter allows only the wavelengths emitted by the excited SO
2
molecules to reach the PMT, which detects the UV light emission. The photo
detector, located at the back of the fluorescence chamber, continuously
monitors the pulsating UV light source to provide compensation for
fluctuations in the UV light source. The measured SO
concentration
2
(representing total sulfur in the sample) is processed, displayed on the front
panel display, and sent to the analog outputs.
Figure 2–2. Typical functional flow diagram
2-4 SOLA II User GuideThermo Fisher Scientific
Page 21
Product Overview
PUVF Detector
Pyrolyzer
Principle of Operation
The PUVF detector includes and controls the following:
λ UV pulsed light and associated systems
λ Reaction chamber temperature control
λ Digitizing of PMT signal
λ Smoothing of measurement signal using moving average
Measuring total sulfur with the PUVF detection method requires the
conversion of all sulfur compounds in the sample to SO2. This is typically
accomplished with the pyrolyzer, an electrically heated furnace designed by
Thermo Fisher Scientific. The pyrolyzer typically operates at a temperature of
1100°C (2012°F) to oxidize sulfur without need for a catalyst.
Figure 2–3. Pyrolyzer
Thermo Fisher Scientific SOLA II User Guide 2-5
Page 22
Product Overview
Dryer
Mixing Chamber
Principle of Operation
The optional Perma Pure dryer removes moisture from the sample prior to its
entry into the PUVF detector. A filter is positioned before the dryer to protect
it from impurities. Dryer tubing consists of multiple small tubes encased in a
large outer tube. Air circulates through the outer tube with sample passing
through the inner tube. Moisture passes from the sample through the tubing
where it is carried to the condensate drain by the airflow in the outer tube.
Figure 2–4. Optional dryer
The mixing chamber mixes the gases and permits the sample to vaporize to a
gaseous state before entering the pyrolyzer.
Figure 2–5. Typical mixing chamber
2-6 SOLA II User GuideThermo Fisher Scientific
Page 23
Product Overview
Injection Valve
Principle of Operation
The injection valve periodically injects precisely measured quantities of the
sample stream into a controlled carrier airflow. An auxiliary airflow is added
to this sample and air mixture. The sample then passes to the mixing chamber
where it vaporizes (if necessary) and is thoroughly mixed with the air. See
Appendix B for rotary injection valve service.
Figure 2–6. Injection valve
Thermo Fisher Scientific SOLA II User Guide 2-7
Page 24
Product Overview
Results may vary under different operating conditions.
Specifications
Specifications
Table 2–1. Mechanical specifications
Mechanical
specifications
Dimensions, H x W x D 40 x 24 x 18 in (102 x 61 x 46 cm)
Weight Approximately 200 lb (91 kg)
Mounting Wall or rack mount
Ambient temperature 12ºC to 40ºC (54ºF to 104ºF)
Incoming purge air
temperature
Area classification CSA:
12ºC to 40ºC (54ºF to 104ºF)
Class 1, Div. 2, Groups B, C, D, T2, or T3 (T3 optional with
back-up purge system)
Class 1, Div. 1, Groups B, C, D, T2, or T3 (optional, XPurge system; T3 optional with back-up purge system)
ATEX:
Zone 1, G Ex pxb IIC T Gb (optional, X-Purge system; T3
and T4 optional with back-up purge system)
II 3 G Ex pz IIC T3 Gc, SOLA II Flare or CV. T3 or T4
optional with back-up purge system, standard SOLA II
IECEx:
G Ex pz IIC T Gc (T3 and T4 with back-up purge system)
Zone 1, G Ex px IIC T Gb (X-Purge system); T3 and T4
with back-up purge system)
Table 2–2. Analytical specifications
Analytical specifications
2-8 SOLA II User Guide
Detector Pulsed UV fluorescence with pyrolyzer for total sulfur
measurement
Full scale range Ranges from 0–5 ppm (w/w) to 0–50,000 ppm (w/w)
(consult factory for other ranges).
Unique ranges may be assigned to Streams 1 and 2.
Trace option enables 0–2 ppm (w/w) measuring range
with LDL of 25 ppb (consult factory for other ranges).
Repeatability Calculated at 1 standard deviation
±1% of full scale at 2x injections per minute
±2% of full scale at 1x injection per minute
Standard SOLA II (not Trace level) for full scale ranges of
10ppm or lower, the repeatability is ±2% of full scale at
2x injections per minute.
Equal to repeatability, for ranges >1% full scale, consult
Linearity
factory for linearity specification.
Thermo Fisher Scientific
Page 25
Product Overview
Specifications
Response time Programmable. Analyzer is semi-continuous; initial response
occurs at each injection
Calibration Automatic or Manual, with ability for Distributed Control
System (DCS) to force an autocal (via contact inputs or
Modbus)
valve fault (lamp rate of change), chamber flow, chamber
temperature, lamp voltage, autocal fail, HIGH and HIGH
HIGH concentration
Optional: Sample flow, analog input board time-out
Alarm relays/indicators SPST, 2 A at 240 Vac or 10 A at 24 Vdc; 8 total
Analog signal output Isolated, 4–20 mA; 2 total
Analog signal load ≤ 600 ohms
Inputs Dry contact; remote suspend, remote calibration, remote
Sample tubing 316 stainless steel, cleaned and free from oils, moisture,
and debris
Sample wetted components 316 stainless steel, Teflon®, and graphite-filled Teflon;
others application dependent (Kalrez, Viton)
The following table lists the SOLA II / SOLA II Trace parts.
Note Unless otherwise specified, the parts listed below are for the SOLA II
and the SOLA II Trace units. σ
Table 2–5. SOLA II / SOLA II Trace parts list
P/N Description
75-1350-0Actuator, standard temperature, air, 36 deg, 10-port
75-1348-0Actuator, standard temperature, air, 90 deg, 6-port
89-2913-0Assembly, Lookout Software
TE-8812 Attenuator, photo cell
TE-8887 Base assembly, PMT
TE-57P713-1 Bench assembly, SOLA II standard unit
TE-57P7125-1 Bench assembly, SOLA II Trace unit
88-1216-0Cable, DB9 for RS232 Modbus
TE-6279 Cable, RS232, assembly 6'
88-1217-0Cable, TCP/IP to CPU
00-1010-SCal standard for liquid units (specify volume, S concentration, and fluid, i.e.
liter, 6 ppm wt/wt thiophene in #2 diesel)
TE-5365 Connector pins, male, bench
2-10 SOLA II User GuideThermo Fisher Scientific
Page 27
Product Overview
P/N Description
TE-4609 Connector, housing, male pin, bench
97-1592-1 Critical operation spare parts kit, Dinfa valve, liquid applications
97-1592-2 Critical operation spare parts kit, Valco valve, gas applications
97-1592-0 Critical operation spare parts kit, Valco valve, liquid applications
TE-8719 Detector filter
TE-8335 Detector lens
TE-8851 Diffuser, photo cell
75-1334-0 Dinfa valve slider 1.2 µl
19-1182-0 Discs, thermal, T 2
19-1183-0 Discs, thermal, T 3
TE-8544 Feet, rubber, shock mounts, optical bench
32-5000 Ferrule, Vespel, graphite, 1/16", for fitting to mixer
TE-8852 Filter, detector
TE-8703 Lens, bi-convex
TE-8333 Lens, condensor
TE-8076 Lens, convex plano-aci
TE-8739 Lens, plano convex- aci
TE-8850 Lens, relay
97-1626-0A Malema flow switch kit
31-1354-0 Manifold assembly, 4-position with solenoid
TE-8888 Mirror assembly, complete, SOLA II standard unit
TE-57P746 Mirror assembly, dual, complete, SOLA II Trace unit
TE-87420 Mirror, entrance
35-1527-0 Mounting bracket, TCP/IP PCB
97-1590-1 One-year spare parts kit, Dinfa valve, liquid phase applications
97-1590-2 One-year spare parts kit, Valco valve, gas phase applications
97-1590-0 One-year spare parts kit, Valco valve, liquid phase applications
TE-4811 O-ring
TE-4808 O-ring (between reaction chamber)
63-1150-0 O-ring kit, Dinfa valve actuator
63-1143-0 O-ring kit, low temperature for Valco valve actuator
TE-4829 O-ring, chamber entrance
TE-4831 O-ring, detector filter
TE-4820 O-ring, lens relay
TE-4830 O-ring, photo cell
TE-4808 O-ring, photo cell conic
63-1135-0 O-ring, pyrolyzer furnace
89-2896-0 PCB, 4-20 mA input
TE-8943 PCB, A/D
TE-8884 PCB, flasher intensity (photo detector)
TE-9681 PCB, flasher supply
TE-8951 PCB, input, PM T signal
89-2899-0 PCB, Modbus RS485, dual channel
TE-9829 PCB, motherboard
55-1228-0 PCB, Netburner TCP/IP Ethernet 10/100
TE-8949 PCB, PUVF DC power supply
2-12 SOLA II User Guide Thermo Fisher Scientific
Page 29
Product Overview
P/N Description
TE-8765 PCB, PUVF temperature control
89-2897-0 PCB, CPU
89-2803-0 PCB, stream relay
89-2898-0 PCB, digital I/O
TE-8684 Plate, filter lens
TE-8868 PMT, SOLA II standard unit
TE-8391 PMT, SOLA II Trace unit
TE-8165 Power cord, 115 V
85-1173-0 Power supply, 110/220 Vac +5 V, ±12 V, for PCB, input option
85-1164-0 Power supply, 24 VDC , DIN rail mount
TE-9901 Power supply, PMT
32-0024 Reducing union, 316 SS 1/8 tube x 1/16 tube
14-1435-0 Regulator, pressure, 0-100 psi, SS case
Parts List
30-2009 Regulator, pressure, 0-25 psi, 4-way mini
HA-100645 Relay, solid state
25-1001 RTD, used for oven heater
TE-8869 Socket assembly, PMT
97-1646-0 Quick repair kit
97-1644-0 N2, addition, field installation
97-1651-0 Solenoid manifold adapter kit
31-0030 Solenoid, 24 Vdc, 5-port, 4-way, 2-position, manifold mount
31-1367-0 Solenoid, 24 Vdc, 5-port, 4-way, 2-position, manifold mount
97-1589-1 Startup spare parts kit, Dinfa valve, liquid phase applications
97-1589-2 Startup spare parts kit, Valco valve, gas phase applications
97-1589-0 Startup spare parts kit, Valco valve, liquid phase applications
25-1189-0 Switch, differential pressure, 0.2 inH2O
25-1137-0 Switch, pressure sensor, 0-15 psi adj, oven and burner
68-2139-2 Tee, 1/4" x 1/8", Sulfinert
68-2138-2 Tee, 1/8", Sulfinert
89-2650-G Temperature controller, furnace, Watlow, complete Centalac
89-2089-G Temperature controller, oven, Watlow, complete Centalac
29-1113-0 Temperature controller, Watlow
56-1150-0 Terminal block, two pole, ceramic, SS connectors
Thermo Fisher Scientific SOLA II User Guide 2-13
Page 30
Product Overview
Parts List
P/N Description
5382 Thermistor IM-1002-A5
27-1108-0 Thermocouple, S type
TE-9934 Transducer, flow
TE-9877 Transducer, pressure assembly
TE-8774 Trigger pak for SOLA II bench, standard unit
TE-8392 Trigger pak for SOLA II bench, Trace unit
30-1025-0 Tube, quartz pyrolyzer furnace, looped
64-1304-0 Tubing, 1/16”, Sulfinert
64-1051-0 Tubing, 1/16" OD x 0.02” ID SS
40-0614 Tubing, 1/16" OD x 0.01" ID, SS, capillary
64-1305-0 Tubing, 1/4", Sulfinert
64-1303-0 Tubing, 1/8”, Sulfinert
40-0611 Tubing, black nylon, 1/4" OD x 0.035” wall
40-0609 Tubing, FEP, 0.125” OD x 0.062” ID, high temperature
40-0600 Tubing,1/8” OD x 0.085” ID, 0.020, 304SS
97-1591-1 Two-year spare parts kit, Dinfa valve, liquid phase applications
97-1591-2 Two-year spare parts kit, Valco valve, gas phase applications
97-1591-0 Two-year spare parts kit, Valco valve, liquid phase applications
75-1343-0 Valve head, Valco, 10-port
75-1333-0 Valve head, Valco, 6-port
45-1837-0 Valve, 3-way, air operated
75-1335-0 Valve, Dinfa, 8-port, 1.2 µl, complete assembly
45-1823-0 Valve, Valco rotor 1.0 µl, 6-port
45-1835-0 Valve, Valco rotor, 10-port
75-1340-0 Valve, Valco, 10-port, complete assembly
75-1332-0 Valve, Valco, 6-port, 1µl, complete assembly with rotor
45-1235-0 Valve, 3-way ball, tube to tube, 316SS, TFE seat, 1/8” port size
TE-8868 PUVF PMT
TE-8943 A/D PCB
TE-8949 PUVF DC power supply board
TE-8774 PUVF lamp trigger pak (consult factory for SOLA II Trace)
TE-8884 PUVF flash intensity board (consult factory for SOLA II Trace)
TE-9681 Flasher supply board (consult factory for SOLA II Trace)
TE-8951 Input board
TE-8765 PUVF temperature control board
TE-8666 PUVF flash lamp
89-2897-0 CPU
89-2898-0 I/O PCB
85-1164-0 Power supply, 24 Vdc, DIN rail mount
Table 2–9. Dinfa injection valve replacement parts
P/N Description
89-2803-0 Solenoid relay PCB
29-1230-0 Heater for pyrolyzer assembly
HA-101812 Ferrules for pyrolyzer tube fittings, graphite
30-1025-0 Pyrolyzer tube, looped
63-1135-0 Pyrolyzer chamber O-ring seal
47-1362-0 Inline micron filter for orifice flow
27-1108-0 Thermocouple
75-1335-0 Injection valve assembly, 8-port, 1.2 µl
75-1334-0 Injection valve slider, 1.2 µl
TE-9934 Flow transducer assembly
TE-8868 PUVF PMT
TE-8943 A/D PCB
TE-8949 PUVF DC power supply board
TE-8774 PUVF lamp trigger pak (consult factory for SOLA II Trace)
TE-8884 PUVF flash intensity board (consult factory for SOLA II Trace)
TE-9681 Flasher supply board (consult factory for SOLA II Trace)
TE-8951 Input board
TE-8765 PUVF temperature control board
2-20 SOLA II User Guide Thermo Fisher Scientific
Page 37
Product Overview
Storage
Storage
P/N Description
TE-8666 PUVF flash lamp
89-2897-0 CPU
89-2898-0 I/O PCB
85-1164-0 Power supply, 24 Vdc, DIN rail mount
If storing the instrument, the storage environment should be protected and
free from extremes of temperatures and high humidity.
Thermo Fisher Scientific SOLA II User Guide 2-21
Page 38
Page 39
Chapter 3
Requirements
Installation
λ Material and power: Consult the specifications for information required
for installation.
λ Operating environment: For optimum reliability and equipment life, we
recommend that the analyzer be installed in a location protected from
extremes in temperature and weather. The analyzer operates best in a
controlled environment. Ambient temperature and purge air must not
exceed the limits listed in the specifications.
λ Mounting requirements: The mounting site must be as close as possible to
the sampling point. The analyzer must be sheltered from extreme weather
conditions. Avoid mounting the analyzer in high vibration areas. Mount
the analyzer in an accessible location.
λ Cable glands used to supply electrical power must be IP40 rated metallic
cable glands.
λ Blanking elements or plugs used shall be in accordance with national
standards.
Caution This product is extremely heavy. Care must be taken at all times to
avoid injury. Never attempt to move this product alone or without the use of
lift gear. When handling the analyzer, ensure that all four corners are
supported.
σ
Thermo Fisher Scientific SOLA II User Guide 3-1
Page 40
Installation
Sample Line
Installation
Sample Tubing
Preparation
Sample Line Installation
Note Pressure buildup or liquid accumulation in the analyzer vents degrades
performance. Vent lines must be as short as possible and routed in a manner
that prevents accumulation of liquids. σ
There must be no back pressure on the vent or drain lines. All vents must be
referenced to atmospheric pressure. If the analyzer is installed in a pressurized
analyzer shelter, route all vents from the purge controllers and the PUVF
detector to the exterior of the shelter. All sample lines must be as short as
possible. Sample lines must use new 316 stainless steel that has been prepared
as described in the following section “Sample Tubing Preparation.”
The analyzer will produce erratic readings if backpressure is created or varied
by an obstructed or improperly routed vent.
Review the following cautions prior to performing this procedure.
Caution Isopropyl alcohol is extremely flammable, hazardous if breathed, and
dries skin on contact. When using isopropyl alcohol, avoid breathing the
vapors and contact with the skin. Appropriate measures must be taken to
prevent ignition of the isopropyl alcohol vapors. Use isopropyl alcohol only
where there is adequate ventilation and no ignition sources are present. Refer
to a Material Safety Data Sheet (MSDS) for isopropyl alcohol for additional
important information. σ
Caution Acetone is extremely flammable, hazardous if breathed, and dries
skin on contact. When using acetone, avoid breathing the vapors and contact
with the skin. Appropriate measures must be taken to prevent the acetone
vapors from igniting. Use acetone only where there is adequate ventilation and
no ignition sources are present. Refer to a MSDS for acetone for additional
important information. σ
Caution Acetone dissolves many plastics. Use caution to prevent the acetone
from contacting materials that may be marred or damaged. σ
Caution Take necessary precautions to prevent exposure to hazardous
materials when conditioning the sample tubing with sample. σ
Proper preparation of sample tubing prior to installation is very important.
Prepare the sample tubing as follows.
3-2 SOLA II User Guide Thermo Fisher Scientific
Page 41
Installation
Power source specification:
Power wiring specification:
Electrical
Connections
DCS & External
Connections
Digital Inputs
Electrical Connections
1. Thoroughly rinse the tubing inside with isopropyl alcohol (isopropanol) or
acetone to remove any oils that may be present.
2. Flush the inside of the tubing with deionized water.
3. Rinse the inside of the tubing again with isopropyl alcohol (isopropanol)
or acetone.
4. Thoroughly blow-dry the tubing with clean air (free of oil and moisture).
Refer to the following AC power information and wire information when
planning and connecting power for the analyzer.
power wire rated for at least 600 Vac and 20 A at the required length.
Warning This apparatus must be earth grounded! σ
Warning Installation of this instrument requires an external, lockable
electrical power isolation switch supplied by the customer. σ
If the optional X-Purge unit is installed, refer to Appendix G for important
installation and operation information.
Wiring to the relay contacts should be sized according to the load imposed by
the alarm systems installed by the user. Maximum current capability of the
alarm contacts is 2 A at 240 Vac or 10 A at 24 Vdc. Voltage is not supplied to
the relay contacts by the analyzer.
110 or 220 Vac, 50/60 Hz, 2000 W
Use stranded, 3-wire copper or tin-plated copper
The following inputs and outputs can be connected to a Distributed Control
System (DCS) or other devices as desired.
Thermo Fisher Scientific SOLA II User Guide 3-3
Page 42
Installation
Optional Analog Inputs
DCS & External Connections
The following inputs are available for controlling the analyzer from a DCS or
other external source. Refer to the wiring diagrams shipped with your
instrument.
λ Remote Stream Control: A manual switch or DCS output signal can be
used to select the following stream operation when the multiple stream
option is installed. For this control to function, enable the Remote Control
function the Analyzer Setup parameters (Chapter 6).
Table 3–1.
Stream
Select 0
Close* Open Analyzes only stream 1
Open Close* Analyzes only stream 2
Close* Close* Analyzer alternates between stream 1 and stream 2 at the
Open Open Analyzer operation for streams is performed as specified in
*Close = Shorted to associated ground terminal
λ Remote Measurement Range Control: To switch to high measurement
Stream
Select 1
Action
time intervals specified in theStream 1 Dwell Time and
Stream 2 Dwell Time parameters (Chapter 6)
the Stream Setup parameters (Chapter 6)
range, short the Remote Range connection to the Remote Range common
terminal. To switch to low measurement range, open the connection
across the Remote Suspend connectors. The Remote Control function
must be enabled (“Analyzer Setup”, Chapter 6).
λ Remote:
Autocal/Validation: The Autocal/Validation function can be initiated by
closing the contact.
Remote Suspend: To suspend analysis, short the Remote Suspend
connection to the Remote Suspend common terminal. To restart analysis,
open the connection across the Remote Suspend connectors. The Remote
Control function must be enabled (“Analyzer Setup”, Chapter 6).
Analog inputs are available for connecting densitometers used to correct each
stream measurement based on the density of the sample.
3-4 SOLA II User Guide Thermo Fisher Scientific
Page 43
Installation
Analog Outputs
Other Outputs
Installation
Checklist
Installation Checklist
λ Density Input Stream 1: This 4–20 mA DC input signal represents the
density of the Stream 1 sample. The measured densities at 4 mA and at 20
mA can be programmed into the analyzer using the Density
Compensation Setup menu (Chapter 6).
Density Input Stream 2: This 4–20 mA DC input signal represents the density
of the Stream 2 sample. The measured densities at 4 mA and at 20 mA
can be programmed into the analyzer using the Density Compensation
Setup menu (Chapter 6).
The following outputs are available with the analyzer. Refer to the wiring
diagrams shipped with your instrument for wiring details.
λ Analog Output 1: This 4–20 mA DC signal represents the measured
concentration of Stream 1. The zero value (4 mA DC) represents zero
measured concentration in the sample. Full scale (20 mA DC) represents
the measured concentration specified in High Range or Low Range
(depending on range selected).
λ Analog Output 2: This 4-20 mA DC signal represents the measured
concentration of Stream 2. It is available only if the dual-stream option is
installed. The zero value (4 mA DC) represents zero measured
concentration in the sample. Full scale (20 mA DC) represents the
measured concentration specified in High Range or Low Range
(depending on range selected).
The following outputs are also provided: High/High Alarm Relay, High Alarm
Relay, Off Line Status Relay (fail-safe), Malfunction Alarm Relay (fail-safe,
normally closed), Range Selected, Stream Selected, Purge Alarm (fail-safe).
The following checklist may be copied for use during system installation.
Details for the installation and various specification requirements are
contained in this manual, application specific drawings, and information
supplied with the instrument.
Thermo Fisher Scientific SOLA II User Guide 3-5
Page 44
Installation
Installation Checklist
NOT
Installation Checklist
Materials used meet specifications defined in Chapter 2.
Operating environment meets requirements defined in Chapter 3.
Mounting meets requirements defined in Chapter 3.
Analyzer condition inspected as follows:
No physical damage, broken parts, or observable defects.
All electronic boards are securely seated.
All cables and wiring connectors are in place and fully seated.
No loose parts (wires, nuts, screws, cables, debris, etc.).
All tubing is properly connected and fittings are tight.
Installation and flow connections adhere to the following:
Sample system properly designed to condition (control pressure, flow, temperature, and particulate)
sample as required by system.
Sample system fast loop panel is installed as close as possible to analyzer for fastest response time.
System gases meet quality and quantity (pressure and flow) specifications according to Chapter 2.
PUVF vent is vented to atmospheric pressure.
PUVF vent is
Inject purge vent will contain instrument air plus 5–10 cm
connected to headers with varying pressure or that may accumulate liquids.
3
of hydrocarbon liquid or gas during suspension
of analyzer, loss of combustion air, or inject valve fault. Inject purge vent is routed to accommodate
disposal of hydrocarbon.
All tubing supply lines to analyzer are sized to match or exceed connector size on the analyzer.
All internal and external tubing fittings are physically checked for tightness and tested for leaks.
Installation and electrical connections adhere to the following:
AC power wiring meets requirements.
AC power wiring is properly connected to the instrument with proper earth grounding.
Signal wiring (DC signals, communications, etc.) meets requirements.
Signal wires properly connected to the instrument.
All electrical conduit seals poured with sealing compound such as Chico A5r or equivalent.
3-6 SOLA II User Guide Thermo Fisher Scientific
Page 45
Chapter 4
Initial Startup
Startup & Shutdown
Perform this procedure when starting up a new installation or when major
service work is performed. Otherwise, perform the procedure according to the
section titled “Startup after Short-Term Shutdown.”
See the application data that shipped with the instrument, the specifications
in Chapter 2, and the installation instructions in Chapter 3 in this manual for
information necessary in performing the following procedure.
1. Verify proper electrical power and connections:
a. Ensure the instrument electrical power wiring is properly sized and
connected.
b. Ensure the power voltage and frequency matches the instrument
requirements.
c. Ensure a suitable circuit breaker and power switch is installed.
d. Ensure the instrument is properly grounded.
e. Inspect all electrical connections. Terminals must be snug; wire and
cable plugs must be fully seated. Perform a visual check for electrical
shorts.
f. Inspect the plug-in cards; ensure they are properly seated in their
connectors.
g. Ensure the signal wiring is properly sized and connected.
2. Verify proper plumbing:
a. Ensure the correct supply tubing is properly connected to the
instrument.
b. All sample lines to the analyzer MUST be cleaned and dried before
initial use.
c. Check all tubing connections to ensure they are tight and free of
leaks. Pressure test the lines to check for leaks or use a liquid leak
detector.
Thermo Fisher Scientific SOLA II User Guide 4-1
Page 46
Startup & Shutdown
Initial Startup
3. Apply instrument/purge air to the instrument for at least 15 minutes, and
set the pressure. See application data sent with instrument for proper
settings.
Warning For Zone II areas, initial purge must be carried out only when the
area is known to be non-hazardous. σ
4. Apply power to the instrument.
Note If the X-Purge option is installed, you must follow the instructions
included in Appendix G. σ
a. Verify that temperature controllers are set as specified in the
application notes shipped with the instrument or as recorded in the
system logbook. Note that the pyrolyzer and oven do not begin
heating immediately due to the safety interlocks.
Note The pyrolyzer temperature controller does not accurately read low
temperatures near ambient; however, it does read accurately at normal
operating temperatures. σ
b. Verify that the display is functioning. The screen in Figure 4–1
appears, which displays the version and whether a valid configuration
is loaded.
Figure 4–1.
c. Verify that the configuration settings in the menus match the
application data sheet or system logbook.
d. Suspend the analyzer (Chapter 5) to prevent accidental injection of
sample while setting up the instrument.
4-2 SOLA II User Guide Thermo Fisher Scientific
Page 47
Startup & Shutdown
Startup after
Term
Shutdown
Startup after Short-Term Shutdown
5. Apply carrier air to the instrument, and set the flows. See application data
sent with the instrument for proper settings.
a. Verify that the pyrolyzer and oven heater temperature come up to
temperature and stabilize at the control point.
b. Adjust the zero/carrier air pressure and flow again after the pyrolyzer
and oven stabilize at the correct temperature.
6. Turn on the sample flow to the instrument, and adjust it to the correct
pressure and flow. Note that this procedure varies depending on the
installed sampling system. Consult the application notes and drawings
provided with the instrument for more information.
7. Enable analysis of the sample by taking the instrument out of the suspend
mode (Chapter 5). The unit begins injecting and analyzing sample when
the instrument temperatures reach operating levels.
Short-
8. Allow the analyzer system to stabilize. Monitor the measured values on
the front panel for consistent analysis readings to determine when the
analyzer system has stabilized.
9. Calibrate the analyzer according to “Calibration” in Chapter 6. Allow
ample time for this initial calibration to stabilize. Subsequent calibrations
proceed more quickly than this initial calibration.
If starting the analyzer for the first time or if major service has been
performed, use the initial startup procedure detailed in the previous section
“Initial Startup.”
When starting the analyzer after a short-term shutdown, perform the
following:
1. Open the instrument/purge air and carrier air to the instrument. Refer to
the applications information, manuals, and drawings shipped with the
instrument, and adjust the pressure regulators to the appropriate settings.
2. Apply power to the analyzer.
Note If the X-Purge option is installed, you must follow the instructions
included in Appendix G. σ
Thermo Fisher Scientific SOLA II User Guide 4-3
Page 48
Startup & Shutdown
Short-Term
Shutdown
Maintenance
Shutdown
Short-Term Shutdown
3. Allow the analyzer to warm up until the instrument stabilizes.
4. Adjust the flow rates.
5. Turn on sample to the instrument.
6. If necessary, calibrate the analyzer (“Calibration,” Chapter 6).
The analyzer includes an interlock system that prevents the injection valve
from operating until the pyrolyzer and oven temperatures reach the operating
value.
Follow this procedure when temporarily shutting down the analyzer. To shut
down the analyzer for maintenance purposes or for a long-term shutdown,
refer to the following section “Maintenance Shutdown.”
1. Turn the solenoid manifold regulator to 0 psig. Doing so blocks sample
flow to the injection valve and purges sample from the injection valve.
2. Observe reported sulfur value.
3. Do not interrupt power, instrument air, carrier or auxiliary air, or open
oven doors until reported sulfur value is less than 0.5 ppm or reported
sulfur value has not changed by more than 2% for 15 minutes.
Note Sample is purged automatically by 3-way diverter valves when power is
removed from the instrument. σ
The analyzer system must be fully shut down and the sample system
decontaminated as appropriate PRIOR to performing maintenance. Follow this
procedure when shutting down the analyzer for maintenance or a long-term
outage.
1. Close the sample flow to the instrument and purge sample from the unit
using air.
2. Turn off power to the analyzer.
4-4 SOLA II User Guide Thermo Fisher Scientific
Page 49
Startup & Shutdown
Emergency
Shutdown
Emergency Shutdown
3. Allow the instrument to cool.
a. If the analyzer is used in an ATEX hazardous area with a T3 rating, do
not open the oven door for 45 minutes, 50 minutes for Zone 2 unit.
b. If the analyzer is used in an ATEX hazardous area with a T4 rating,
do not open the oven door for 140 minutes.
Caution Failure to allow adequate cooling time before opening the oven can
lead to equipment damage or injury to personnel. σ
4. Turn off all air supplies.
Caution Parts of the instrument may be hot even after power is removed.
Allow the system to cool completely before performing maintenance. σ
1. Close the sample supply to the system.
2. Turn off the main power to the system.
Thermo Fisher Scientific SOLA II User Guide 4-5
Page 50
Page 51
Chapter 5
The Interface
Operation
The analyzer utilizes a 4-line x 20-character, vacuum fluorescent display and 4
push buttons as the user interface. All programming and adjustments are
accomplished using the push buttons. Following is an example of how to use
the interface to enter the passcode.
1. Press any button, and you are prompted to enter the passcode, as seen in
Figure 5–1.
2. Press the button below Ent (Enter) and Figure 5–2 appears.
3. To increase the first digit’s value press the button below the up arrow
(∧). Decrease the value by pressing the button below the down arrow
(∨). When the correct number is displayed, press the button below the
right arrow (>) to enter the next digit.
4. Repeat the process used to enter the first digit for the remaining digits.
5. When you have completed this, press the button below Done.
You can access all program features and edit parameters by following these
steps.
Figure 5–1.
Thermo Fisher Scientific SOLA II User Guide 5-1
Page 52
Operation
Suspended to suspend the analysis. In suspended mode, the
Active
Begin Analyzing
Suspending an
Analysis
Begin Analyzing
Figure 5–2.
1. If the analyzer is shut down, start up the instrument by following the
appropriate set of instructions in Chapter 4.
2. If the system power is off, open the instrument and carrier air flows to
the analyzer, and adjust the pressures to the appropriate settings (refer to
application data that shipped with the analyzer).
3. Open sample flow to the analyzer.
4. Apply power to the instrument.
5. Analysis begins automatically when the pyrolyzer reaches operating
temperature. If you need to change the configuration, press any button,
and enter the passcode.
Follow these steps to suspend an analysis.
1. Enter the Analyzer Mode top-level menu (Figure 5–3).
2. Select
3. To resume the analysis, go to the Analyzer Mode menu, and select
analyzer stops injecting, and the sample diverter valve allows only air to
run through the burner and mixing chamber.
.
5-2 SOLA II User Guide Thermo Fisher Scientific
Page 53
Operation
Injection Valve
Failure Alarm
Injection Valve Failure Alarm
Figure 5–3.
Over time, the injection valve begins to wear, resulting in an increased
potential for port-to-port leakage of sample. Such leakage can cause erroneous
measurement and materials buildup in the system. When the analyzer detects
sample leakage through the injection valve, it triggers the Injection Valve
Failure alarm and switches the sample diverter valves to flow air through the
injection valve rather than sample.
Note The Injection Valve Failure alarm must be acknowledged after valve
repair or replacement. σ
If an injection valve begins leaking due to damage or wear, excess sample
begins flowing to the pyrolyzer. This extra load of sample in the pyrolyzer
exceeds the ability to fully combust the materials in the sample. When these
incompletely combusted materials enter the PUVF optical bench, they absorb
UV light.
The PUVF includes a system to monitor and compensate for decreases in the
UV light output as the bulb ages. However, in the case of injection valve
failure, the PUVF system begins rapidly increasing the UV lamp output in an
effort to compensate for the absorption of UV by incompletely combusted
sample products. The user can enter the rate of change that triggers this
alarm. The programmable range is from 1 to 50 V/30 seconds. If the analyzer
detects the programmed rate of change in UV lamp voltage, it triggers the
Injection Valve Failure alarm to protect the system and reduce the
stabilization time for the PUVF bench on system restart.
When the Injection Valve Failure or Chamber Flow alarm activates, a pair of
sample diverter valves switches to flow air through the injection valve instead
of sample. This prevents excess sample and incomplete combustion products
from further contaminating the system. The diverter valves switch air to the
injection valves under the following conditions:
There are five top-level menus; this chapter addresses the Configuration menu
(Figure 6–1). The Configuration menu contains 10 setup menus: Analyzer
Setup, Stream Setup, Clock Setup, Modbus Setup, Dual Range Setup, Inject
Setup, Autocal Setup, Alarms Setup, Calibration Setup, and Density
Compensation Setup.
Figure 6–1. Top-level configuration menu
1. Access the submenus within the Analyzer Setup menu by pressing
Figure 6–2.
.
2. The first submenu is Program Passcode. The factory default is 0000. If the
passcode remains 0000, you are not required to enter the passcode to
Thermo Fisher Scientific SOLA II User Guide 6-1
Page 56
Configuration
Analyzer Setup
access the menus. Change the passcode to prevent unauthorized access to
the menus.
Figure 6–3.
3. Return to the Analyzer Setup menu. Press the down arrow until the
Remote Control submenu appears (Figure 6–4), allowing you to
enable/disable the remote control function. “DCS & External Connections”
in Chapter 3 addresses which functions may be operated remotely.
Figure 6–4.
4. The Average Time submenu allows you to set the averaging time. The
average time is a period (1–240 seconds) during which the analyzer takes
SO2 measurements used to determine a moving average of the results. For
example, if you set the averaging time to 10 seconds, the average
concentration of the last 10 seconds is output at each update. If you set
the averaging time to 60 seconds, the average concentration of the last 60
seconds is output at each update. Thus, lower averaging time means faster
response to concentration changes. Longer averaging times are typically
used to smooth output data.
See the figure below.
Figure 6–5.
6-2 SOLA II User Guide Thermo Fisher Scientific
Page 57
Configuration
Analyzer Setup
5. The average override is expressed as a percentage of change and
determines when the unit switches to the fast average. As the reading
stabilizes at the new level, the analyzer progressively increases the average
used until it reaches the programmed average time.
Figure 6–6.
6. The fast average time provides the ability to respond to changes faster
without sacrificing stability achieved using a longer average time.
Note Both the regular and fast average times need to be set at multiples of
the injection rate. For example, if the rate is two injections per minute,
averages should be set to 30, 60, 90, 120, 150, 180, 210, or 240 seconds. σ
Figure 6–7.
7. There are three selections within this submenu: NONE, 4 to 20 mA, and
Flow Switch.
Thermo Fisher Scientific SOLA II User Guide 6-3
Page 58
Configuration
Enter
Stream Setup
Stream Setup
Figure 6–8.
This section steps through the menu items in the Stream Setup menu.
1. Access the submenus within the Stream Setup menu by pressing
Figure 6–9.
2. The Stream Mode submenu allows you to configure the analyzer to
continuously monitor stream 1 (Stream 1), continuously monitor stream 2
(Stream 2), or to alternate between the streams (Timed Stream).
.
Figure 6–10.
3. If you select the Timed Stream option as the mode, you must determine
the dwell time. Dwell time is the period of time the analyzer monitors one
stream before it begins monitoring the other. Figure 6–11 displays the
6-4 SOLA II User Guide Thermo Fisher Scientific
Page 59
Configuration
Stream Setup
submenu that allows you to program the dwell time for stream 1 (the
length of time the analyzer stays on stream 1 before it begins monitoring
stream 2). You can then program the dwell time for stream 2 (the length
of time the analyzer stays on stream 2 before returning to stream 1).
Dwell time may be from 1 to 1440 minutes (24 hours).
Figure 6–11.
4. When using the second stream for validation or during the calibration
cycles, it might be desirable to reduce the flow in order to conserve
calibration/validation standards. You can program each stream for high or
low flows. If equipped with the second stream option, the analyzer has
NO and NC pneumatic outputs to control the flow in the sampling
system. The calibration and validation routines always select the low flow.
Figure 6–12.
5. If you select the Timed Stream option, you must also determine how long
the analyzer purges the line before it begins analyzing the other stream.
This period is called the purge time. In respect to calibration, purge time
is how long the analyzer purges the line before it begins analyzing the
calibration stream. You can set the purge time between 1 and 9999
seconds.
See the figure below.
Thermo Fisher Scientific SOLA II User Guide 6-5
Page 60
Configuration
Clock Setup
Clock Setup
Figure 6–13.
1. Access the Clock Setup menu to change the time and date.
Figure 6–14.
2. Enter the time in hh:mm:ss, 24-hour format and the date (in the following
screen) in mm/dd/yyyy format.
Figure 6–15.
Figure 6–16.
6-6 SOLA II User Guide Thermo Fisher Scientific
Page 61
Configuration
Disable
Modbus Setup
Modbus Setup
1. The Modbus Setup menu enables to you set the baud rates and IDs for
COM 3 and COM 4.
Figure 6–17.
2. Select one of the baud rates available for RS485 Com 3 (38400, 19200,
9600, 4800, 2400, and 1200 bps), or select
.
Figure 6–18.
3. Set the ID (network address) for Com 3. The address can be between 1
and 255.
Figure 6–19.
Set the baud rate and the ID for COM 4 in the same manner using the
submenus for COM 4.
Thermo Fisher Scientific SOLA II User Guide 6-7
Page 62
Configuration
Dual Range
Setup
Dual Range Setup
This menu allows you to set the dual range mode.
1. Set the dual range mode to Remote Range Select or to Auto Range for the
4–20 mA output signal. If you select Auto Range, the 4–20 mA output
signal switches to high range when the reading goes above the low range
full scale value. When the signal drops below 90% of the low range full
scale value, the system switches back to low range.
When Remote Range Select is enabled, output scaling is controlled by a
remote contact input. When the input contacts are open, the system
operates on low range. When the input contacts are closed, the system
operates on high range.
Figure 6–20.
Figure 6–21.
2. The submenu shown below enables you to specify the full scale value (20
mA) for the high range of stream 1 as output on the 4–20 mA signal.
Figure 6–22.
6-8 SOLA II User Guide Thermo Fisher Scientific
Page 63
Configuration
Inject Setup
Inject Setup
3. The submenu shown below enables you to specify the full scale value (20
mA) for the low range of stream 1 as output on the 4–20 mA signal.
Figure 6–23.
Set the 20 mA scaling for the high and low ranges for stream 2 using the
submenus for stream 2.
1. Enter the Inject Setup menu to set injection parameters.
Figure 6–24.
2. The injection time is the amount of time that the injection valve stays in
the injection position. Set the inject time at half of the inject rate (1 to
9999 seconds).
Figure 6–25.
Thermo Fisher Scientific SOLA II User Guide 6-9
Page 64
Configuration
if a 6-port valve is
used, set the rate to twice the desired injection interval.
Autocal Setup
Autocal Setup
3. Set the injection rate (the amount of time to wait between sample
injections). The rate can be between 2 and 9999 seconds. A 6-port valve
injects at each actuator move in both positions. Thus,
Figure 6–26.
1. Enter this menu to configure the autocal function.
Figure 6–27.
2. The Autocal Mode submenu allows you to enable/disable periodic
automatic calibration of the analyzer.
Figure 6–28.
6-10 SOLA II User Guide Thermo Fisher Scientific
Page 65
Configuration
Autocal Setup
3. If the autocal function is enabled, access the Autocal Interval submenu to
specify the length of time between automatic calibrations. The interval can
be set between 1 to 9999 minutes.
Figure 6–29.
4. Allowed Deviation is the percentage of deviation from the programmed
autocal value that is acceptable without updating the calibration factor.
This function is used primarily when using calibration sample for system
validation tests. When an automatic calibration is performed, the analyzer
calculates the percentage the new calibration factor deviates from the
previous calibration factor. If the percentage deviation is less than the
allowed deviation, the system keeps the previous calibration factor. If the
deviation exceeds the allowed deviation and is less than the Recal
Deviation, the new calibration factor replaces the previous one.
Figure 6–30.
5. Recal Deviation is the percentage of deviation from the programmed
autocal value that is acceptable to compute new calibration factors. Access
this submenu to specify this percentage. Note that an Autocal Fail alarm
activates if the deviation exceeds the amount specified.
See the figure below.
Thermo Fisher Scientific SOLA II User Guide 6-11
Page 66
Configuration
Autocal Setup
Figure 6–31.
6. If set to recalibrate, the calibration line moves, maintaining the same
slope. It is similar to the manually initiated adjust reading function.
Setting both deviations to the same value disables the autocal, making it a
validation only function.
The autocal/validation functions can be used with a programmable value
other than the high calibration value.
Figure 6–32.
7. You can direct the reading during auto calibration to stream 1 or 2, 4–20
mA output, or none of the outputs.
Figure 6–33.
6-12 SOLA II User Guide Thermo Fisher Scientific
Page 67
Review the following for instructions pertaining to alarm setup.
Enter
Alarms Setup
Configuration
Alarms Setup
1. Access the menu items by pressing
.
Figure 6–34.
2. The Sample Flow alarm is available if the flow meter option is present.
The set point for the low sample flow alarm is a percentage of the 4–20
mA signal from the flow meter.
Figure 6–35.
3. To set up a Chamber Flow alarm, enter the set point at which the low
flow alarm for the flow measured at the detector chamber in the PUVF
should activate. The value can be from 0 to 1000 cc/m.
Figure 6–36.
Thermo Fisher Scientific SOLA II User Guide 6-13
Page 68
Configuration
Alarms Setup
4. To set up a Lamp Voltage alarm, enter the maximum lamp voltage at
which the alarm should activate. The value can be from 0 to 1200 V.
Figure 6–37.
5. Set up a Lamp Voltage Rate of Change alarm by entering the rate of
change in lamp voltage at which the alarm should activate. The value can
be from 1 to 50 V per 30 seconds.
Figure 6–38.
6. Set up a Chamber Temperature alarm by entering the +/- deviation from
45°C at which the alarm should activate.
Figure 6–39.
6-14 SOLA II User Guide Thermo Fisher Scientific
Page 69
Configuration
Alarms Setup
7. Select one of the TS Level Alarms modes. Select Disable, Non Latching, or
Latching mode for the concentration alarms. If you select nonlatching, the
alarm clears when the measured concentration drops below the associated
value. If you select latching, the alarm must be acknowledged before it will
clear.
Figure 6–40.
8. Set the High Alarm for stream 1 by entering the first level of high
concentration for stream 1 at which the alarm should activate.
Figure 6–41.
9. Set the High High alarm for stream 1 by entering the second level of high
concentration for stream 1 at which the alarm should activate. There are
also submenus for setting the high alarm and the high high alarm for
stream 2. If applicable, set these alarms as done for stream 1.
Figure 6–42.
Thermo Fisher Scientific SOLA II User Guide 6-15
Page 70
Configuration
Calibration Setup
Calibration
Overview
Calibration
The analyzer requires a two-point linear calibration. During the calibration
procedure the average of the raw detector signals corresponding to two
known sulfur concentrations are latched by the unit and retained in nonvolatile memory.
During normal operation, the average detector reading is interpolated on the
line to derive the sulfur concentration reported.
The graph below illustrates a SOLA II calibration.
Figure 6–43.
Prior to calibrating the instrument, the sulfur concentration expected at the
two points must be entered in the calibration setup menus (Configuration >
low calibration values.
For the graph above, the values would be:
Low cal value = 10.00 ppm Low cal = 555.600 kHz
High cal value = 100.00 ppm High cal = 15000.000 kHz
The high calibration value needs to be as close as possible to the full range
sulfur concentration expected in the process, and the low calibration value is
normally set to zero so that only one standard is required for calibration.
). In the software, these points are referred to as high and
6-16 SOLA II User Guide Thermo Fisher Scientific
Page 71
The values and a graph for a typical calibration are shown below.
Low cal value = 0.00 ppm Low cal = 555.600 kHz
High cal value = 100.00 ppm High cal = 15000.000 kHz
Figure 6–44. Typical SOLA II calibration
Configuration
Calibration
The high and low calibration can be performed independently. Changing
either one of the calibration points will change the slope of the line. This is
illustrated in the next two graphs.
Figure 6–45. High calibration
Thermo Fisher Scientific SOLA II User Guide 6-17
Page 72
Configuration
Calibration
Figure 6–46. Low calibration
If necessary, the Adjust Reading function in the Calibration Setup menu can
be used to correct the calibration by moving the line without changing the
slope (see Figure 6–47). Enabling the Autocal function will cause the software
to perform this correction automatically.
Figure 6–47. Corrected calibration
Several calibration functions can be performed. These are discussed in the
following sections.
6-18 SOLA II User Guide Thermo Fisher Scientific
Page 73
Configuration
Low Cal Value
Recalibrate Low
YES
Performing a Low
Calibration at Zero*
Calibration
*No injections
Note This is the most common form of low calibration that requires only one
standard. For SOLA II Trace, it is recommended that a low concentration
standard be used. σ
Note To prevent repetition, these sections do not include screen shots. Each
menu item within the Calibration Setup menu is shown in “Calibration Setup
Menu Items.”σ
To perform a low calibration with no injections, follow the steps below.
1. Go to Configuration > Calibration Setup >
. Set the low
calibration value to zero.
2. Exit the Calibration Setup menu.
3. Reduce the pressure at the solenoid manifold air regulator to zero (this is
the bottom regulator located in the pneumatic cabinet). Lowering the
solenoid manifold pressure will accomplish two things:
a. The diverter valve will switch to the diverter position blocking the
sample.
b. The injection valve will stop injecting, letting only the carrier gas run
through the system.
4. Let the unit run until the sulfur concentration reported is stable.
5. Go to Configuration > Calibration Setup >
. Press
to
accept the new low calibration.
At this point, the average of the detector signal is latched to correspond to a
zero sulfur concentration.
Thermo Fisher Scientific SOLA II User Guide 6-19
Page 74
Configuration
High Cal Value
Purge Time
Enter
Performing a High
Calibration
The Calibrate High
Function
The Recalibrate High
Function
Calibration
Performing a high calibration can be achieved in two ways, depending on the
sample conditioning system installed.
The high calibration standard is normally contained in a pressurized cylinder.
The pneumatic output of the analyzer can be used to switch the
corresponding selection valve in the SCS, or the high calibration standard can
be manually introduced to the analyzer. Based on this, the Calibration Setup
menu offers two functions:
λ Calibrate High: Used when the analyzer is expected to switch the valves.
λ Recalibrate High: Used when the calibration standard is introduced
manually to the analyzer.
1. Go to Configuration > Calibration Setup >
. Set the high
calibration value to match the concentration of the standard used.
2. Step down to the Average Cal Readings screen. It is recommended that
you set this value to the maximum (240).
3. Exit the Calibration Setup menu, and go to Configuration > Stream Setup
>
. Set this value. It should be long enough for the unit to
stabilize (time varies depending on the location of the switching valve in
the SCS and the flow rate of the sample). Exit this menu.
4. Go back to the Calibration Setup menu. Step down to the Calibrate High
screen, and press
to perform the calibration.
At this point, the analyzer will energize the calibration solenoid and start a
purge cycle. Once the purge cycle is completed, the detector reading will be
averaged for the programmed number of calibration readings (readings are
performed once a second). The resulting average will be latched to correspond
to the high calibration value concentration.
6-20 SOLA II User Guide Thermo Fisher Scientific
Page 75
Configuration
Cal Inlet
High Cal Value
Recalibrate High
YES
Adjust Reading
Adjusting the
Reading
Calibration
Use this function if manually introducing the high calibration standard to the
analyzer.
Note It is possible to force the calibration solenoid to turn on by going to
Configuration > Stream Setup >
and selecting ON. If you do this, you
can use the Recalibrate High function on systems where the analyzer
pneumatic output selects the calibration stream. σ
Note It is important to remember that if the calibration standard is
introduced manually and running as stream one, density compensation must
be turned off if used. σ
1. Go to Configuration > Calibration Setup >
. Set the high
calibration value to match the concentration of the standard used. Exit the
menu.
2. Connect the calibration standard to the unit, and let it flow through the
injection valve.
3. Let the reported concentration reach a new value and stabilize.
4. Go to Configuration > Calibration Setup >
. Press
to
accept the new high calibration.
At this point, the current average of the detector signal is latched to
correspond to the high calibration value concentration.
If the reported value of the SOLA II does not match a known sample value or
the standard flowing through the instrument, you can adjust the reading. To
do so, go to Configuration > Calibration Setup >
. The current
reading will be displayed in an editing screen so that you can modify the
reading to match the desired reading.
Note Do not use this function if the difference between the expected reading
and the analyzer’s reported reading is significant. The Adjust Reading function
should only be used for making minor corrections to the calibration. Using
this function improperly can momentarily hide malfunctions in the unit that
need to be corrected. σ
Thermo Fisher Scientific SOLA II User Guide 6-21
Page 76
Configuration
Calibration Setup
Menu Items
Calibration
This section contains the menu items within the Calibration Setup menu in
the order they appear. Specific instructions on the various calibration
functions are discussed in the previous sections.
Figure 6–48. Main menu heading
1. Select the engineering unit related to the concentration entered for the
calibration values: ppm, ppb, or mg/L.
Figure 6–49.
2. Enter the concentrations of the high calibration standard. See “Performing
a High Calibration” for more.
Figure 6–50.
6-22 SOLA II User Guide Thermo Fisher Scientific
Page 77
Configuration
Calibration
3. Enter the concentration of the low calibration standard. Note that if using
only one calibration standard and no injections to calibrate the
background, enter zero. See “Performing a Low Calibration at Zero” for
more.
Figure 6–51.
4. Set the number of readings the analyzer should average during calibration
(1–60 readings).
Figure 6–52.
5. Begin a calibration for the high concentration standard. See “Performing a
High Calibration” for more.
Figure 6–53.
Thermo Fisher Scientific SOLA II User Guide 6-23
Page 78
Configuration
Calibration
6. Begin a calibration for the low concentration standard or for zero
background.
Figure 6–54.
7. Perform a calibration without the purge cycle for the high concentration
standard. See “Performing a High Calibration” for more.
You can do the same for the low concentration standard by accessing the
Recalibrate Low menu item (not shown). See “Performing a Low
Calibration at Zero” for more.
Figure 6–55.
8. Adjust calibration during operation to match lab results through the
Adjust Reading menu item. Enter a concentration, and the reading is
forced to match the value entered. See “Adjusting the Reading” for more.
Figure 6–56.
6-24 SOLA II User Guide Thermo Fisher Scientific
Page 79
Configuration
Density
nsation
Setup
Density Compensation Setup
9. Review/edit the frequency stored during calibration of the high standard.
You can do the same for the low standard or zero background accessing
the Low Cal KHz menu item.
Figure 6–57.
10. Calibration Output
Compe
Figure 6–58.
Density correction is used for reporting the total sulfur in ppm w/w and can
be accomplished using a density transmitter (densitometer) for each stream
with the output connected to the analyzer’s 4–20 mA inputs or by entering a
fixed correction factor.
When using the optional densitometer inputs, the densities that correspond to
4 mA and 20 mA must be entered. The analyzer maintains a moving average
of the 4–20 mA readings and converts the readings to g/cc using a linear
interpolation. The sulfur reading is then corrected by multiplying it by the
ratio between the calibration sample density and the measured density of the
sample.
When the density of the sample is known and remains relatively constant, a
fixed correction factor can be used. In this method, the total sulfur reading is
multiplied by the value entered. If no correction is desired, set the correction
factor to 1.000.
The following provides instructions for both methods. If using a densitometer,
refer to the instructions with Figures 6–59 through 6–63. Menu items for
stream 1 are used.
Thermo Fisher Scientific SOLA II User Guide 6-25
Page 80
Configuration
Live
Density Compensation Setup
1. Enter the Density Compensation Setup menu.
Figure 6–59.
2. Select
. If you are using stream 2, be sure the menu item is for stream
2.
Figure 6–60.
3. Enter the density of the calibration standard (in units of g/cc).
Figure 6–61.
6-26 SOLA II User Guide Thermo Fisher Scientific
Page 81
Configuration
Fix
Density Compensation Setup
4. In units of g/cc, enter the density that corresponds to the 4 mA output of
the densitometer used (Figure 6–62). In the following screen (Figure 6–
63), enter the density that corresponds to the 20 mA output of the
densitometer.
Figure 6–62.
Figure 6–63.
The following provides instructions on entering a correction factor using
stream 2.
1. Select
in the Str2 Compensation menu item. If using stream 1, be sure
the menu item is for stream 1.
Figure 6–64.
Thermo Fisher Scientific SOLA II User Guide 6-27
Page 82
Configuration
Density Compensation Setup
2. Enter the correction factor. The value may be from 0 to 9.999.
Figure 6–65.
Note that when using a correction factor, you do not need to enter the Cal
Density menu.
6-28 SOLA II User Guide Thermo Fisher Scientific
Page 83
Chapter 7
Ack
Viewing Alarms
The second top-level menu is the View/Ack Alarm menu.
Figure 7–1.
Access this menu to review all active alarms. Use the up and down arrows to
scroll through the alarms individually. Press
Figure 7–2.
to acknowledge the alarm.
Thermo Fisher Scientific SOLA II User Guide 7-1
Page 84
Page 85
Chapter 8
Tgle
Input & Output
Test Menu
Set Outputs
Diagnostics
There are four submenus within the Diagnostics top-level menu: Input &
Output Test, View Parameters, Com Ports RX TX, Pulse UV Control.
Figure 8–1.
The Input & Output Test menu consists of four groups of submenus: Set
Outputs Menu, View Mux Analogs, 4-20 mA Inputs, and View Digital Inputs.
Figure 8–2.
In the Set Outputs group, press
to toggle the associated relay on and off.
Figure 8–3.
Following are the relays that can be accessed through the Set Outputs group.
Thermo Fisher Scientific SOLA II User Guide 8-1
Page 86
Diagnostics
View MUX Analogs
Input & Output Test Menu
Table 8–1.
Output Function
Malfunction Malfunction relay (fail safe).
Off Line Denotes calibration cycle or suspended analysis (fail safe).
H Alarm High alarm relay.
HH Alarm High high alarm relay.
Stream Selected Indicates the active stream.
Range Selected
RLY 7 Purge fail (fail safe).
RLY 8 Spare
Stream 1 4–20 mA Press Enter to change the 4–20 mA output for stream 1 to 4, 8, 12,
16, or 20 mA.
Stream 2 4–20 mA Press Enter to change the 4–20 mA output for stream 2 to 4, 8, 12,
16, or 20 mA.
Sample Solenoid Controls the sample diverter.
Stream 1 Solenoid Controls the valve for stream 1.
Stream 2 Solenoid Controls the valve for stream 2.
Inject Solenoid Controls the injection valve.
Cal Solenoid Controls the calibration/stream valve.
Solenoid 6 Controls high/low flow
Solenoids 7–8 See note below
Bench CNTRL ## Reserved for factory test only; the 2 pound signs (##) indicate the
control output.
Note Solenoids 7 and 8 are reserved for future applications and are not
installed in the standard unit. σ
8-2 SOLA II User Guide Thermo Fisher Scientific
Page 87
Diagnostics
View 4–20 mA Inputs
Input & Output Test Menu
You can view the MUX analogs by accessing this menu. Note that all signals
are displayed as raw frequencies.
Figure 8–4.
Following are the analogs in the order they appear within menu:
1. ADC 0 Lamp V
2. ADC 1 Chamber Flow
3. ADC 7 Pressure
4. ADC 9 PMT V
5. ADC 10 Chamber T
6. ADC 11 Ambient T
Access this submenu to view the optional 4–20 mA inputs.
Figure 8–5.
Thermo Fisher Scientific SOLA II User Guide 8-3
Page 88
Diagnostics
View Digital Inp ut s
Input & Output Test Menu
When you access the View Digital Inputs menu, the status of the inputs is
reported but the functions are ignored.
Figure 8–6.
Following is a list of inputs and their functions.
Table 8–2.
Input Function
Purge Fail Displays the status of the purge sense pressure switch.
Flow Switch Displays the status of the flow monitoring switch.
Watlow Alarm Displays the status of the Watlow alarm input1
I/O Board Displays the status of all eight inputs simultaneously. Zero (0)
denotes open. One (1) denotes closed.
CPU Digital In
Dip Switches Displays the positions of dip switches located in the motherboard (dip
switches are used to indicate options installed)
I/O Board Option
4–20 mA Input
Option
Stream Select 1 Displays the status of stream select switch #1
Stream Select 2 Displays the status of stream select switch #2
Range Control Displays the status of the range control switch
Remote Cal Displays the status of the remote calibration switch
Remote Suspend Displays the status of the remote suspend switch
8-4 SOLA II User Guide Thermo Fisher Scientific
Page 89
Diagnostics
View Parameters
View Parameters
Each menu item within the View Parameters menu displays the parameter
name, current value, minimum value, and maximum value (see Figure
8–8). When you select a parameter to view, the minimum and maximum
values are reset to the current value. Displaying the minimum, maximum, and
current values for a period of time enables you to monitor signal stability.
Figure 8–7.
Figure 8–8.
Following are the menu items within the View Parameters menu.
Table 8–3.
Menu Item Description
At line sample Access this item to view the latest at line sample results:
Sample AVG: The Average Concentration of the sample.
SD: The Standard Deviation computed during the average.
% RSD: The Relative Standard Deviation computed as a
percentage of the average.
TS Reading The concentration reading.
Response Factors The slope (kHz/ppm or kHz/ppb sensitivity) and the offset
(calculated at 0 concentration).
Working Average Sec When switching from the normal average to the fast average and
back to normal, the average time used can be observed from this
menu item.
Thermo Fisher Scientific SOLA II User Guide 8-5
Page 90
Diagnostics
View Com
Activity
View Com Activity
Menu Item Description
% of Deviation The percentage of deviation of the most current reading with
respect to the average. This is the value used to switch to the fast
average.
Rate of Change ppm/s The rate of change of the sulfur reading expressed as ppm/sec.
AVG PMT signal kHz The average of the raw detector signal.
Normalized PMT kHz The frequency proportional to the PMT output normalized to the
gain of 100; also displays the gain currently selected.
PMT Signal kHz The actual frequency proportional to the PMT output; also displays
the gain currently selected.
PMT Voltage The input voltage to the PMT.
Internal Temperature The temperature inside the electronics enclosure.
Chamber Temperature The temperature inside the reaction chamber.
Chamber Flow The flow through the reaction chamber.
Chamber Pressure The pressure sensed at the reaction chamber.
Lamp Intensity The frequency proportional to the lamp intensity.
Lamp Voltage The lamp input voltage.
Stream 1 Density The density of stream 1.
Stream 2 Density The density of stream 2.
This menu displays activity for Coms 1–4.
Figure 8–9.
8-6 SOLA II User Guide Thermo Fisher Scientific
Page 91
Diagnostics
Pulse UV Control
Pulse UV Control
Following are the menu items within the Pulse UV Control menu.
Caution This menu is intended for factory test only. Incorrect settings can
result in damage to the unit. σ
Table 8–4.
Menu Item Description
Flash Lamp ON/OFF Turn the UV lamp flashing on or off.
Auto Gain ON/OFF Turn the amplifier auto gain selection on or off.
Note Auto gain defaults to ON during normal operation.
PMT Signal Gain Select the amplifier gain (1, 2, 5, 10, 20, 50, 100).
Note The selection is only valid when auto gain is turned
σ
OFF.
Test Gains Place the gain selection under manual control to verify the
functionality of the programmable gain amplifier.
Test LED ON/OFF Test the LED by turning it on and off.
σ
Flash Frequency Change the UV lamp flash frequency (normally 10 Hz).
Flash Time Change the duration of the flashing pulse (normally 100 µs).
S&H Time (Sample &
Hold Time)
Change the duration of the sample portion of the sample and hold
analog front end (normally 100 µs). (Sample portion is the amount
of time the sample and hold device is in the SAMPLE state.)
Thermo Fisher Scientific SOLA II User Guide 8-7
Page 92
Page 93
Chapter 9
At Line-Grab Sample
This function was developed primarily for pipeline applications. The sampling
system was designed for the pipeline enclosure in order to provide an
additional pump to facilitate introduction of sample. Access this top-level
menu to enable the analyzer to analyze and report the results of a sample
previously collected from a source other than the regular process input. The
sample to be analyzed is introduced into the instrument in the same manner
that a calibration standard is.
There are three parameters to configure: starting sample ID, average sampling
time, and the relative standard deviation of the samples taken.
1. Each sample is identified with a number (starting sample ID). The
numbers increment sequentially until a new starting sample ID is
programmed.
Figure 9–1.
2. Average sampling time is the number of seconds that the sample is
averaged to compute the result. It can be between 60 and 240 seconds.
Figure 9–2.
Thermo Fisher Scientific SOLA II User Guide 9-1
Page 94
At Line-Grab Sample
3. The relative standard deviation (RSD) of the samples taken is computed
as a percentage of the computed average. If it is less than or equal to the
programmed value, the analyzer reports a successful sample. If it is
greater than the programmed value, an error is reported.
Figure 9–3.
Activate the function from the front panel or the SOLA web interface.
Figure 9–4.
Figure 9–5. At line sampling screen
9-2 SOLA II User Guide Thermo Fisher Scientific
Page 95
Chapter 10
Safety
Precautions
Maintenance
Schedule
Maintenance & Tr oubl eshooting
Caution Some internal components can be damaged by small amounts of
static electricity. Take appropriate precautions (use a properly grounded
antistatic wrist strap) when handling electronic boards and components.
To avoid damaging internal components, follow these precautions when
performing any service procedure:
λ Wear an antistatic wrist strap that is properly connected to earth ground.
If an antistatic wrist strap is not available, be sure to touch a grounded
metal object before touching any internal components.
λ Handle all printed circuit boards by the edges.
λ Carefully observe the instructions in each procedure. σ
Table 10–1. Maintenance schedule
Frequency Tasks
Monthly Calibrate the analyzer (Chapter 6).
Every six months Visually inspect and clean the instrument.
Check the instrument flow rates as explained later in this chapter.
Replace the injection valve rotor or slider according to Appendix B.
Annually Test the instrument for internal leaks as explained later in this
chapter.
Every 18 months Replace the pyrolyzer heater as explained later in this chapter.
Thermo Fisher Scientific SOLA II User Guide 10-1
Page 96
Maintenance & Troubleshooting
Visual Inspection &
Cleaning
Leak Tests
Visual Inspection & Cleaning
The analyzer should be inspected occasionally for obvious visible defects, such
as loose connectors, loose fittings, cracked or clogged Teflon lines, and
excessive dust or dirt accumulation. Dust and dirt can accumulate in the
instrument and can cause overheating or component failure. Dirt on the
components prevents efficient heat dissipation and may provide conducting
paths for electricity.
Warning Remove all instrument power before cleaning electronics. σ
The best way to clean the inside of the instrument is to first carefully vacuum
all accessible areas and then blow away the remaining dust with low-pressure
compressed air. Use a soft paintbrush or cloth to remove stubborn dirt.
Use the following leak test procedure to verify that there are no system leaks.
1.Set the sample and air regulators for zero pressure.
2.Replace the tubing from the EXHAUST bulkhead on the left side of the
instrument with a plug.
3.Adjust the auxiliary air regulator for 20 psig pressure.
4.Turn the auxiliary air regulator fully counterclockwise to close the supply
flow.
5.Watch the analyzer pressure for 15 minutes.
6. If the pressure drops significantly, perform the following to locate the
leak:
a.Adjust the sample and air regulators for 20 psig pressure.
b.Check the system fittings with a liquid or electronic leak detector.
c.Correct any leakage.
d.Repeat the leak test.
10-2 SOLA II User GuideThermo Fisher Scientific
7.When the system passes the leak test, adjust the sample and air regulators
fully counterclockwise to close supply flows.
Page 97
Maintenance & Troubleshooting
Flow Rate
Checks
Flow Rate Checks
8. Remove the plug from the EXHAUST bulkhead and reconnect the
EXHAUST vent line.
9. Adjust the sample and air regulators to the pressures specified in the
application notes shipped with the instrument or as recorded in the
instrument logbook.
10. Allow the analyzer to warm up until it stabilizes.
Figure 10–1. Three-way valve for measuring flow rates
1. Obtain a precision flow measurement instrument such as a bubble meter.
2. Ensure that all pressure regulators are set to the correct pressures as
shown on the calibration data shipped with the instrument or as recorded
in the instrument logbook.
3. Ensure that the analyzer is stabilized at normal operating temperatures
before proceeding.
4. Suspend the analyzer according to Chapter 5.
Thermo Fisher Scientific SOLA II User Guide 10-3
Page 98
Maintenance & Troubleshooting
The Rotary Valve
The Mixing
Chamber
The Rotary Valve
5.Turn the MEAS. CLEAN AIR 1 3-way valves to point down for flow
measurement.
6. Connect the flow meter to the MEAS. CLEAN AIR 1 port and adjust the
CLEAN AIR 1 regulator to obtain the required flow.
7.Turn the MEAS. CLEAN AIR 1 3-way valves to point up for normal
operation.
8.Turn the MEAS. CLEAN AIR 2 3-way valves to point down for flow
measurement.
9. Connect the flow meter to the MEAS. CLEAN AIR 2 port and adjust the
CLEAN AIR 2 regulator to obtain the required flow.
10. Turn the MEAS. CLEAN AIR 2 3-way valves to point up for normal
operation.
11. Restart the analyzer according to Chapter 5.
For detailed instructions on maintaining, removing, and replacing the rotary
valve, refer to Appendix B.
1.Shut down the system according to Chapter 4.
2.Loosen the fittings that connect the tubing to the mixing chamber.
3.Loosen the screw located in the center of the bracket holding the mixing
chamber to the bottom of the enclosure.
4. Carefully slide the mixing chamber from the bracket.
Note Install the mixing chamber by following the above steps in reverse. σ
10-4 SOLA II User GuideThermo Fisher Scientific
Page 99
Maintenance & Troubleshooting
Replacing the
Pyrolyzer Heater
Replacing the Pyrolyzer Heater
Warning The pyrolyzer can be extremely hot, even after power is turned off.
Use extreme care to prevent burns! σ
The pyrolyzer heater is enclosed in the pyrolyzer housing, which is shown in
the figure below.
Figure 10–2. Pyrolyzer housing
The following table lists the part numbers for the parts referenced in this
procedure. With the exception of the heater being replaced, the existing parts
can be reused if in good condition. The part numbers are provided here for
convenience.
1.Follow the maintenance shutdown procedure in Chapter 4 to shut down
the analyzer system. Allow the system to cool completely.
2.Carefully remove the two sample tubes from the top of the pyrolyzer
housing using a backup wrench to keep the stainless steel fittings from
turning.
3.Loosen the terminal block screws and disconnect the thermocouple wires.
Be sure to note the orientation of the wire colors.
4. If installed, disconnect the wires from the oven temperature switches; the
wire between the two sensors can be left in place.
5. Slightly loosen the two screws that attach the bottom of the housing to
the back of the oven.
6. Support the housing while removing the top two mounting screws that
hold it to the back of the oven. After both screws have been removed, lift
the pyrolyzer up and out of the oven. Do not discard the mounting
screws. You will need them later.
7. Remove the six screws and washers from the lid of the housing (Figure 10-
3). Set them aside for later use.
10-6 SOLA II User GuideThermo Fisher Scientific
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.