Thermo Scientific Nicolet Summit User Manual

Nicolet FTIR Spectrometers
Nicolet Summit
User Guide
269-334400 Revision E December 2019
© 2019 Thermo Fisher Scientific Inc. All rights reserved.
Microsoft, Windows and Excel are either trademarks or registered trademarks of Microsoft Corporation in the United States and/or other countries. All other trademarks are the property of Thermo Fisher Scientific Inc. and its subsidiaries.
Thermo Fisher Scientific Inc. provides this document to its customers with a product purchase to use in the product operation. This document is copyright protected and any reproduction of the whole or any part of this document is strictly prohibited, except with the written authorization of Thermo Fisher Scientific Inc.
The contents of this document are subject to change without notice. All technical information in this document is for reference purposes only. System configurations and specifications in this document supersede all previous information received by the purchaser.
Thermo Fisher Scientific Inc. makes no representations that this document is complete, accurate or error­free and assumes no responsibility and will not be liable for any errors, omissions, damage or loss that might result from any use of this document, even if the information in the document is followed properly.
This document is not part of any sales contract between Thermo Fisher Scientific Inc. and a purchaser. This document shall in no way govern or modify any Terms and Conditions of Sale, which Terms and Conditions of Sale shall govern all conflicting information between the two documents.
For Research Use Only. This instrument or accessory is not a medical device and is not intended to be used for the prevention, diagnosis, treatment or cure of disease.
WARNING Avoid an explosion or fire hazard. This instrument or accessory is not designed for use in an explosive atmosphere.
C

Contents

Chapter 1 Welcome. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1
Conventions Used . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Ordering Parts. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Contacting Us . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Chapter 2 Get Started with Your Nicolet Summit Spectrometer . . . . . . . . . . . . . . . . . . . . . . .3
Before Your Summit Spectrometer Arrives . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Personal Safety with the Summit Spectrometer . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Unpack and Get Started with the Spectrometer. . . . . . . . . . . . . . . . . . . . . . . . . 12
Back up and Recover Your Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Connect an External Monitor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Install a Sampling Accessory . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Understand the Spectrometer LightBar . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
View Your Data with the OMNIC Anywhere App . . . . . . . . . . . . . . . . . . . . . . 34
Chapter 3 Options and Accessories. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .35
Sampling Accessories for the Summit Spectrometer. . . . . . . . . . . . . . . . . . . . . . 36
Install and Maintain a Purge Kit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
Chapter 4 Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .46
Maintenance Schedule for the Summit Spectrometer . . . . . . . . . . . . . . . . . . . . 47
Align the Spectrometer Optics. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
Calibrate the Laser Frequency . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
Clean the Spectrometer and Touchscreen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
Replace the Desiccant . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
Replace the Source . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
Install the Touchscreen Monitor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
Replace the Sample Compartment Windows. . . . . . . . . . . . . . . . . . . . . . . . . . . 68
Thermo Scientific Nicolet Summit User Guide iii
1

Welcome

The Thermo Scientific Nicolet™ Summit Fourier transform infrared (FTIR) spectrometer lets you perform chemical analyses of sample materials by collecting data in the mid-IR spectral range using a variety of accessories. The system has integrated verification features, a powerful software suite, and many other features that make data collection easy. You can install optional hardware and perform several service and maintenance procedures yourself. This document, or other provided documentation, contains the information you need.
Be sure to read the safety guide that came with the system before operating the instrument.
Note We recommend that you maintain seal and desiccation and/or purge your instrument at all times. Equipment damage due to failure to maintain seal and desiccation and/or purge is not covered by the warranty. If you have questions about this requirement, please contact us.

Conventions Used

Safety precautions and other important information use the following format:
WARNING Indicates a hazardous situation which, if not avoided, could result in death or serious injury.
CAUTION Indicates a hazardous situation which, if not avoided, could result in minor or moderate injury.
NOTICE Follow instructions with this label to avoid damaging the system hardware or losing data.
Note Contains helpful supplementary information.
Tip Provides information that can make a task easier.
Thermo Scientific Nicolet Summit User Guide 1
1
Welcome

Ordering Parts

Ordering Parts

Contacting Us

To o r de r p ar t s , contact us.
If you need to send the instrument or an accessory to us for repair, call or e-mail us first for any shipping requirements or other instructions.
For Technical Support, please contact: www.thermofisher.com
If you have problems with your system, call Technical Support.
2 Nicolet Summit User Guide Thermo Scientific
2

Get Started with Your Nicolet Summit Spectrometer

To get started with your new Nicolet Summit Spectrometer, review the site and safety requirements before the instrument arrives, unpack and setup the instrument, and connect any additional accessories.
Contents
Before Your Summit Spectrometer Arrives
Personal Safety with the Summit Spectrometer
Unpack and Get Started with the Spectrometer
Back up and Recover Your Data
Connect an External Monitor
Install a Sampling Accessory
Understand the Spectrometer LightBar
View Your Data with the OMNIC Anywhere App
Thermo Scientific Nicolet Summit User Guide 3

Before Your Summit Spectrometer Arrives

Prepare a suitable workspace for your spectrometer before it arrives to ensure the most accurate data and the best long-term performance.
This article explains what to do when your spectrometer arrives and introduces environmental and electrical factors that can influence your spectrometer’s performance. For a more comprehensive explanation of these requirements, see the Site and Safety Information installed on your Summit spectrometer.
Contents
• Unpacking the Spectrometer
• Preparing the Workspace

Unpacking the Spectrometer

Before you open the box, there are two important steps to take as soon as the instrument arrives:
• Check the exterior of the shipping box for damage
If you notice any signs of damage, contact us or your local distributor for instructions.
• Allow the spectrometer to come to room temperature
Inside the shipping box, the spectrometer is sealed in a plastic bag to keep it dry. You
must allow 24 hours for the spectrometer to reach room temperature before you open the bag. If you open the bag before the spectrometer is warmed up, condensation
could form that could damage internal optical components and cause permanent damage.
The warranty will not cover damage due to improper moving techniques or to removing the sealed plastic bag before the instrument has come to room temperature.

Preparing the Workspace

Before your instrument arrives, make sure your workspace will be able to properly support the spectrometer. In addition to allowing enough room for the spectrometer, consider several environmental and electrical requirements.
Thermo Scientific Nicolet Summit User Guide 4
Before Your Summit Spectrometer Arrives
Spectrometer Dimensions
The Summit spectrometer requires a fairly small space, but make sure you also leave room around the instrument so that heat can dissipate from the vents and so that you can easily access the instrument’s ports, power button, and cables.
• Summit instrument weight: 10.9 kg (24 lbs)
• Summit instrument weight with touchscreen option: 12.6 Kg (27.8 lbs)
• Dimensions (W x H x D):
34 cm x 24 cm x 32 cm; (13.3 in. x 9.6 in. x 12.7 in.)
• Dimensions with touch screen display option (W x H x D):
53 cm x 43 cm x 32 cm; (20.8 in. x 17.0 in. x 12.7 in.)
Environmental Factors
The Summit spectrometer is a robust instrument designed to be used in many environments. However, for best performance, keep it in a relatively dust-free and low-humidity environment. The spectrometer operates reliably at temperatures between 15 and 35 °C, but for optimal performance should be kept between 20 and 22 °C.
Humidity can cause condensation to form inside the instrument, which could damage internal components. A few precautions can typically protect your instrument from humidity:
• Maintain the instrument’s desiccant, including while the instrument is in storage.
• Avoid rapid changes in temperature.
Keep the instrument away from sources of cool or hot air, such as near heating and air
conditioning vents or large windows.
If your spectrometer will be kept in an especially humid environment, consider installing a
purge gas kit.
Thermo Scientific Nicolet Summit User Guide 5
Before Your Summit Spectrometer Arrives
Electrical Requirements
Power supplied to the spectrometer must be from dedicated, uninterrupted sources and free from the following issues:
• Voltage dropouts
• Transient spikes
•Frequency shifts
• Other line disturbances
If you suspect problems with your power, we recommend a power quality audit. Contact us or your local electrical authority for more information.
Electrical Service Specifications
The following table lists specifications for electrical service. Contact our service representative in your area if you have questions about the requirements.
Requirements Specification
Input Current 1.6 A Max
Input Voltage 100–240 VAC
Line Frequency 47–63
Line Disturbances Sags, surges, or other line disturbances must not exceed 10% of
input voltage
Noise Less than 2 V (common mode)
Less than 20 V (normal mode)
Thermo Scientific Nicolet Summit User Guide 6

Personal Safety with the Summit Spectrometer

While the Nicolet™ Summit™ Spectrometer is designed to be a safe instrument, you should take a few precautions to protect yourself from potential hazards that can arise during normal use and maintenance.
CAUTION This guide is an introduction to potential dangers that you should be aware of, but it is not a comprehensive guide. Before using the instrument, see the Site and Safety
Information for a full description of these potential hazards.

Potential Hazards during Normal Use

During normal use, most hazards are due to following sources:
• Potentially hazardous samples and solvents
• Parts of the spectrometer that are hot or emit heat
• The instrument’s laser
You can avoid possibly harming yourself or damaging the instrument by understanding potential hazards and by taking a few precautions.
Hazardous Samples and Solvents
Take special precautions if you are using or plan to measure potentially hazardous samples or solvents, such as pressurized gases or corrosive or flammable solvents.
Use Proper Ventilation
There are no special ventilation requirements for your spectrometer, but you may need additional ventilation during certain types of analysis. Ensure that you have proper ventilation if you will be analyzing highly toxic samples, dissolving your samples in solvents that may interact with the infrared source, or sampling flammable gases.
The pyrolysis of solvents containing halogenated hydrocarbons may produce hydrochloric acid (HCl), hydroflouric acic (HF), or phosgene (COCl
WARNING Avoid toxic inhalation. Hydrochloric acid, hydrofluoric acid, and phosgene are highly toxic. If you are using solvents containing halogenated hydrocarbons, ensure your work area is properly ventilated.
Thermo Scientific Nicolet Summit User Guide 7
).
2
Personal Safety with the Summit Spectrometer
Volatile and Flammable Solvents
The infrared source inside the spectrometer can ignite flammable and volatile samples and solvents. Take the following actions when working with samples and solvents that are flammable:
• Work with the sample compartment windows installed.
• Ensure that the workspace is properly ventilated with an active venting system that is free of spark or other sources of ignition and that prevents flammable vapors from collecting in the atmosphere surrounding the instrument.
• Do not leave flammable solvents or samples near the instrument.
• Do not leave flammable solvents or samples in the sample compartment for longer than necessary.
• Purge the spectrometer with clean, dry air or nitrogen.
Corrosive Solvents
Solvents that produce HCl or HF vapors in the sample compartment may severely damage the system. If you are using halogenated solvents, purge the instrument with clean, dry air or nitrogen.
NOTICE Your warranty does not cover equipment damage that is due to a failure to purge the instrument.
HCl and HF vapors may also compromise the coating of KBr sample compartment windows. If you are planning to work with corrosive solvents regularly, consider installing ZnSe sample compartment windows, instead.
Biohazard or Radioactive Materials and Infectious Agents
Biological samples, such as tissues, body fluids, infectious agents, and blood, have the potential to transmit infectious diseases. Follow your organization’s Biosafety Program protocols for working with potentially infectious materials.
Thermo Scientific Nicolet Summit User Guide 8
Sources of Heat
Vents
Heat from infrared source (underneath the instrument)
Personal Safety with the Summit Spectrometer
Parts of the spectrometer can become quite hot during normal use. Be careful around the spectrometer’s infrared source and the vents.
Figure 1. Take care around the instrument’s infrared source and the vents
The outward facing surface of the infrared source is located on the bottom of the instrument and can become quite hot. Do not touch the underside of the instrument during operation or shortly after use.
The spectrometer’s vent is located on the left side of the instrument. During normal use, the spectrometer dissipates hot air from internal components to the area outside the spectrometer through these vents. Leave enough space around the instrument vents for hot air to dissipate.
Laser and Optical Safety
You will never be exposed to unsafe levels of laser radiation during normal use of the spectrometer. If the cover is removed during a service procedure, you may need to take special precautions, such as using protective eyewear. Your service person will notify you if this is necessary.
WARNING Avoid personal injury. Never stare into the laser beam or at its reflection. Never tamper with the laser, even if you are replacing a defective laser, or you could be exposed to laser light or to high voltage.

Potential Hazards during Maintenance

You may be exposed to different hazards while performing maintenance on the instrument than you would encounter during normal use. During maintenance, the primary hazards involve purging the instrument and working with the instrument’s internal components.
Thermo Scientific Nicolet Summit User Guide 9
Purging the Instrument
In especially humid environments, we recommend installing a source of clean, dry air or nitrogen to purge the spectrometer. Purging the instrument can help protect internal optical components from damage caused by a moist environment or corrosive solvents, and it can ensure more accurate results.
See “Install and Maintain a Purge Kit” for details on purchasing and installing a purge kit for your spectrometer.
DANGER Avoid fire and explosion hazard.
• Use only dried air or nitrogen to purge your spectrometer.
• Never use a flammable, combustible, or toxic gas to purge this instrument. The purge gas must be free of oil and other reactive materials. Heat from the source or from laser absorption may ignite flammable gases or reactive materials in the purge gas.
Working with Internal Components
Personal Safety with the Summit Spectrometer
Typically, there should be no reason for you to remove the instrument cover or attempt to replace internal components. However, if you need to remove the cover for maintenance, be aware that you risk being shocked, burned, and exposed to laser light.
CAUTION Avoid shock hazard.
Even after this instrument has been disconnected from all voltage sources, capacitors may remain charged for up to 30 seconds and can cause an electrical shock.
CAUTION Avoid burn hazard.
Internal components, especially the infrared source, can become extremely hot during normal operation. Turn off the instrument and wait at least 10 minutes before replacing any components.
WARNING Avoid personal injury.
• Never stare into the laser beam or at its reflection. Never tamper with the laser. You could be exposed to laser light or high voltage.
• If you adjust the laser or perform procedures that are not described in the user guides and manuals, you could be exposed to hazardous radiation.
Thermo Scientific Nicolet Summit User Guide 10
Personal Safety with the Summit Spectrometer
Replacing the desiccant
When you open the desiccant compartment, you must prevent flammable liquids or gases from entering the compartment. For instructions on changing the desiccant, see “Replace the
Desiccant.”
DANGER Avoid explosion hazard.
Before you open the desiccant compartment, turn off the instrument power, unplug the power cord and remove all accessories and samples from the system. Entry of flammable liquids or gases into the desiccant compartment could cause an explosion. If such entry occurs, contact us immediately and do not apply power to the instrument until the condition has been corrected.
Replacing the IR Source
The IR source becomes extremely hot during use. If you need to replace the IR source, avoid burn and explosion hazards.
CAUTION Avoid burn hazard.

Summary

The source becomes extremely hot during normal use. Always allow the source to cool for at least 10 minutes after the spectrometer is turned off before you work with the source.
DANGER Avoid explosion hazard.
Before you remove the source from the spectrometer, turn off the instrument power, unplug the power cord, disconnect any purge lines, and remove all accessories and samples from the system. Entry of flammable liquids or gases into the source compartment could cause an explosion. If such entry occurs, contact us immediately and do not apply power to the instrument until the condition has been corrected.
The Nicolet Summit spectrometer is a safe and rugged instrument, but you can be exposed to hazards during use and maintenance. During normal operation, use caution when handling potentially hazardous samples and solvents and avoid parts of the spectrometer that are hot or emit heat. During maintenance, take precautions to avoid harm or damage that could occur when purging the instrument, handling internal components, or changing the desiccant.
Thermo Scientific Nicolet Summit User Guide 11

Unpack and Get Started with the Spectrometer

Unpacking and setting up your Thermo Scientific™ Nicolet™ Summit Spectrometer requires you to place the instrument in your workspace, to connect and power on the spectrometer, and to run a short performance test.

Before You Begin

Wait 24 hours before taking the spectrometer out of its plastic shipping bag to prevent condensation from forming that can damage internal optics. The warranty will not cover damage due to opening the sealed plastic bag before the instrument has come to room temperature.
Reminder: Review the spectrometer's Site and Safety Information to ensure that the workspace is safe and suitable for the instrument.
Thermo Scientific Nicolet Summit User Guide 12

What’s in the Box?

Unpack and Get Started with the Spectrometer
The Summit spectrometer comes with several accessories and options. The following image shows what comes with your spectrometer, including many optional add-ons.
1. Summit or Summit PRO Spectrometer
2. Power cord
3. Ethernet cable
Optional
4. Everest ATR Accessory
5. Touchscreen monitor
6. ZnSe window replacement kit, including window replacement tool and nitrile gloves
7. Purge kit
8. Wireless keyboard and mouse (not shown: USB extension cable)
9. Mini DisplayPort cable (included with the optional 22” high-definition monitor)
10. Mini DisplayPort to HDMI adapter
11. WiFi access dongle (comes pre-installed if purchased)
Not shown: Optional 22” High-Definition Monitor
Thermo Scientific Nicolet Summit User Guide 13

Unpacking and Setup

The steps for unpacking and setting up your spectrometer vary somewhat depending on the options you purchased.
1. Unpack the shipping box.
Unpack and Get Started with the Spectrometer
Remove the spectrometer from the shipping box and bag and remove additional packing
materials.
NOTICE Use care when lifting the spectrometer.
• Do not grasp the touchscreen monitor to lift the spectrometer.
• Do not touch the sample compartment windows while unpacking the instrument. Touching the windows could cause permanent damage.
Thermo Scientific Nicolet Summit User Guide 14
Unpack and Get Started with the Spectrometer
1. Ethernet port
2. Power input
3. Mini DisplayPort
4. USB ports (2)
5. Power button
6. WiFi adapter port (under the top cover)
2. Set up the monitor.
• If you have the attached touchscreen monitor, grasp the monitor by the sides and position it to face the front of the instrument.
• If you are using an external monitor, connect the monitor to the spectrometer using the Mini DisplayPort. If you are also using the optional wireless mouse and keyboard, use the provided USB extension cable to position the wireless receiver farther from the USB inputs.
Note Some USB and Mini DisplayPort devices are known to interfere with the signal from wireless USB devices, such as wireless keyboards and mice. If you notice performance issues with your wireless keyboard and mouse, such as lag or delayed or missed input, use the provided USB extension cable to position the wireless USB receiver farther from the USB ports.
3. (Optional) Replace the sample compartment windows.
If you purchased the ZnSe windows option, replace the windows. For instructions on replacing the windows, see Replace the Sample Compartment Windows.
4. Turn on the spectrometer.
NOTICE You must create a new password the first time you power on the spectrometer. If this password is lost or forgotten, all data saved on the spectrometer will be unrecoverable. It is recommended that you leave the password blank during initial setup and later create a secure password and additional user accounts as needed.
a. Connect the power supply to the spectrometer and to a properly grounded AC power
source.
Thermo Scientific Nicolet Summit User Guide 15
Unpack and Get Started with the Spectrometer
b. Press the instrument’s power button to turn on the spectrometer. The spectrometer’s
LightBar will display a cycling green light while the instrument warms up and will
show a solid green bar when the instrument is ready for use.
5. Verify the spectrometer’s performance.
Performance verification and qualification workflows use a polystyrene reference material inside the spectrometer to run a series of standard tests to verify your instrument operation and to ensure the accuracy of your data.
a. Launch the Thermo Scientific™ OMNIC™ Paradigm software. The first time the
software starts, it will open with the touchscreen interface. Later, you can switch between the touchscreen and the desktop interfaces.
b. Using the touchscreen interface, navigate to the Workflows tab and select Nicolet
FTIR – Factory Qualification. To run the workflow, touch the Run icon and follow the on-screen prompts.
For more on performance verification and qualification workflows, see Qualify Your
Spectrometer.
Thermo Scientific Nicolet Summit User Guide 16
Unpack and Get Started with the Spectrometer
6. Insert a sampling accessory
To install the optional Everest ATR accessory or iD1 Transmission accessory, grasp the accessory by the front and rear handholds and lower the accessory onto the spectrometer. Sampling accessories fit over two alignment pins on the spectrometer baseplate and are held in place by strong magnets.

Next Steps

• For details on installing an accessory, see Install a Sampling Accessory.
• For details on using the Everest ATR accessory, see Sampling Accessories.
Congratulations! Your instrument is ready for use. However, before you start collecting data, back up your system and create a recovery drive to protect your system from potential problems in the future. Windows 10 includes several built-in tools for backing up and recovering your data.
For instructions on creating a recovery drive and system image and on restoring your system, see “Back up and Recover Your Data.”
Thermo Scientific Nicolet Summit User Guide 17

Back up and Recover Your Data

Before you start collecting data, you should take a few precautionary steps to protect your system from potential problems in the future. Windows 10 includes several built-in tools for backing up and recovering your data.
IMPORTANT! To ensure that you can recover any data in the event of damage to the hard drive or operating system, create recovery media, create a system image, and establish procedures for backing up your data.

Create a Recovery Drive

Create a recovery drive so that you can access advanced startup options in Windows 10—such as System Image Recovery, System Restore, and Startup Settings—even in your installation of Windows becomes damaged.
Required Materials
• FAT32 formatted flash drive with at least 8 GB of storage
NOTICE Everything on the drive will be deleted during this process. Before you proceed, make sure you have backed up any files you wish keep.
To create a recovery drive
1. Insert the flash drive in one of the USB ports on the side of the spectrometer.
2. In the Windows 10 search box in the taskbar, search for “Create Recovery Drive” and select Create Recovery Drive.
Thermo Scientific Nicolet Summit User Guide 18
Back up and Recover Your Data
3. When asked to allow changes to your device, select Yes .
4. De-select “Back up system files to the recovery drive” and click Next.
5. Select your drive from list of available drives and click Next.
6. Click Create. When the process is finished, click Finish to close the window.
7. Safely remove the flash drive and store it in a safe place. It’s a good idea to label the drive as Summit Recovery Drive.
For more on creating recovery media, see Microsoft’s online help.

Create a System Image

A system image is a snapshot of your entire system, including your data and files, installed applications, and other settings. In case of complete hard drive or operating system failure, you can use the system image to restore your device to the state it was in when the image was created. It is up to you how often you want to update the system image, but it’s a good idea to keep an image of a clean installation from the first time you set up your instrument.
Depending on how much data you copy to the system image, the process may take a while. If you are copying a large amount of data, it may take more than an hour for the process to complete.
Required Materials
• NTFS formatted SSD
Thermo Scientific Nicolet Summit User Guide 19
Loading...
+ 53 hidden pages