DYMO and LabelWriter are either trademarks or registered trademarks of Newell Rubbermaid in the United States and/or other
countries. Wi-Fi is either a trademark or a registered trademark of Wi-Fi Alliance in the United States and/or other countries. Bluetooth
is either a trademark or a registered trademark of Bluetooth Special Interest Group. Windows is either a trademark or a registered
trademark of Microsoft Corporation in the United States and/or other countries. All other trademarks are the property of Thermo
Fisher Scientific inc. and its subsidiaries.
For U.S. Technical Support, please contact:
Thermo Fisher Scientific
3411 Silverside Road
Tatnall Building, Suite 100
Wilmington, DE 19810 U.S.A.
Telephone: 302 479 7707
For International Support, please contact:
http://www.nanodrop.com/support
Contact your local distributor. For contact
information go to:
Thermo Fisher Scientific Inc. provides this document to its customers with a product purchase to use in the
product operation. This document is copyright protected and any reproduction of the whole or any part of this
document is strictly prohibited, except with the written authorization of Thermo Fisher Scientific Inc.
The contents of this document are subject to change without notice. All technical information in this
document is for reference purposes only. System configurations and specifications in this document supersede
all previous information received by the purchaser.
Thermo Fisher Scientific Inc. makes no representations that this document is complete, accurate or errorfree and assumes no responsibility and will not be liable for any errors, omissions, damage or loss that might
result from any use of this document, even if the information in the document is followed properly.
This document is not part of any sales contract between Thermo Fisher Scientific Inc. and a purchaser. This
document shall in no way govern or modify any Terms and Conditions of Sale, which Terms and Conditions of
Sale shall govern all conflicting information between the two documents.
For Research Use Only. This instrument or accessory is not a medical device and is not intended to be used
for the prevention, diagnosis, treatment or cure of disease.
WARNING Avoid an explosion or fire hazard. This instrument or accessory is not
designed for use in an explosive atmosphere.
with NanoDropQC Software User GuideThermo Scientific
About the Spectrophotometer
NanoDrop OneC Spectrophotometer
Arm
Pedestal
1
Cuvette
holder
NOTICE Locate the instrument away from air vents and exhaust fans to minimize
evaporation
The Thermo Scientific™ NanoDrop™ OneC is a compact, stand-alone
UV-Visible spectrophotometer developed for micro-volume analysis of a wide variety of
analytes. The patented sample retention system enables the measurement of highly
concentrated samples without the need for dilutions.
C
The NanoDrop One
NanoDrop QC PC Control software can be installed on a local PC and used to control the
instrument and view data. The instrument can be connected to an optional printer with a
USB cable or to a remote printer through an Ethernet connection or wireless network.
NOTICE Before operating a NanoDrop One instrument, please read the safety and
operating precautions and then follow their recommendations when using the instrument.
system comes with preloaded software and a touchscreen display.
Thermo ScientificNanoDrop OneC with NanoDropQC Software User Guide 1
1
About the Spectrophotometer
Features
Features
Touchscreen
TheNanoDrop OneC spectrophotometer features the patented micro-volume sample
C
retention system. The NanoDrop One
also features a cuvette holder for analyzing dilute
samples using standard UV-visible cuvettes.
TheNanoDrop OneC comes with a built-in, 7-inch high-resolution touchscreen preloaded
with easy-to-use instrument control software. The touchscreen can slide left or right to
accommodate personal preference, and tilt forward or back for optimal viewing
2 NanoDrop One
C
with NanoDropQC Software User GuideThermo Scientific
Cuvette Holder
Instrument light path
1
About the Spectrophotometer
Cuvette holder
Features
USB-A port
The NanoDrop OneC includes a cuvette holder for measuring dilute samples, colorimetric
assays, cell cultures and kinetic studies. The cuvette system has these features:
• extended lower detection limits
• 37 °C heater option for temperature-sensitive samples and analyses
• micro-stirring option to ensure sample homogeneity and support kinetic studies
For details, see Measure a Sample using a Cuvette.
Two more USB-A ports are located on instrument back panel
Thermo ScientificNanoDrop One
C
with NanoDropQC Software User Guide 3
1
About the Spectrophotometer
Accessories
Accessories
This section lists the accessories included for use with the NanoDrop OneC.
PR-1 Pedestal Reconditioning Kit
Specially formulated conditioning compound that
can be applied to the pedestals to restore them to a
hydrophobic state (required to achieve adequate
surface tension for accurate sample measurements).
The kit includes conditioning compound and
applicators. For more information, see
Reconditioning the Pedestals.
PV-1 Performance Verification Solution
Liquid photometric standard used to check instrument performance. For more information,
see Performance Verification.
4 NanoDrop One
C
with NanoDropQC Software User GuideThermo Scientific
Instrument Detection Limits
Measurement LocationPathlength (mm)
Pedestal1.012.5
1
About the Spectrophotometer
Instrument Detection Limits
Upper Detection Limit
(10 mm Equivalent Absorbance)
0.262.5
0.1150
0.05300
0.03550
Cuvette101.5
53
27.5
115
Thermo ScientificNanoDrop One
C
with NanoDropQC Software User Guide 5
1
About the Spectrophotometer
This page is intentionally blank.
6 NanoDrop One
C
with NanoDropQC Software User GuideThermo Scientific
Instrument Set up
Register Your Instrument
2
Register your instrument to receive e-mail updates on software and accessories for the
NanoDrop One
To register your instrument
1. Do one of the following:
On the website, locate NanoDrop One Registration and follow the instructions to register the
instrument.
Update Software
Quickly and easily download and install the latest NanoDrop One software and release notes
from our website. Follow the steps to update or upgrade the software on your local instrument
and/or install or update the NanoDrop QC software on a personal computer (PC). An
Internet connection is required to download software.
To install or update NanoDrop QC software on a PC
1. Insert the USB flash drive containing the installer software into an available USB port on
C
instrument. An Internet connection is required for registration.
–From any PC that is connected to the Internet, use any web browser to navigate to
our website.
your PC
2. Launch Start.exe. The software installer will run.
To install or update NanoDrop QC software on the instrument
1. Copy the .zip file with the new software from your computer to a USB storage device. Do
not attempt to unzip the folder.
C
2. Insert the USB device into any USB port on the NanoDrop One
3. From the instrument Home screen, tap Settings > System > Update Software and
choose the latest version of software.
Thermo ScientificNanoDrop OneC with NanoDropQC Software User Guide 7
instrument.
2
Instrument Set up
Setting Up User Account Control
Setting Up User Account Control
User account control is managed using the Security Administration application. When you
launch Security Administration, you will need to enter your Windows log-in information.
User Account Control
Launch the Security Administration application and select NanoDrop QC from the directory
on the left to reveal Access Control and System Policies.
Access control
Access control is used to grant or deny individual users or groups access to protected features
in the instrument application. Add and remove users and groups to the access list and set
access rights using the drop-down for each entity.
System policies
8 NanoDrop One
System Policies is used to set options that define the behavior of the client application. See
“Security Administration Policies.”
C
with NanoDropQC Software User GuideThermo Scientific
Security Administration Policies
System policies allow you to assign data and method creation and deletion and editing
privileges for users and groups.
Launch the Security Administration application and select NanoDrop QC -> System
Policies
You can add, delete, or edit policy groups and enable or disable the group’s users permission to
delete data. When you are finished, select Save. Changes will take effect the next time
NanoDrop QC is launched. Changes made to the Security Administration policies are
applicable to the local PC only and will not affect other computers on the network.
2
Instrument Set up
Setting Up User Account Control
Thermo ScientificNanoDrop One
C
with NanoDropQC Software User Guide 9
2
Instrument Set up
Technical Support
Technical Support
For U.S./Canada Support, please contact:
Thermo Fisher Scientific
3411 Silverside Road
Tatnall Building, Suite 100
Wilmington, DE 19810 U.S.A.
Contact your local distributor. For contact information go to:
http://www.nanodrop.com/Order.aspx
If you are experiencing an issue with your system, refer to the troubleshooting information. If
the issue persists, contact us. If you are outside the U.S.A. and Canada, please contact your
local distributor.
If your instrument requires maintenance or repair, contact us or your local distributor.
10 NanoDrop One
C
with NanoDropQC Software User GuideThermo Scientific
Applications
Use the NanoDrop OneC to perform UV-Vis, Chemometrics, or your own custom
measurements.
The UV-Vis application can be set up directly from the touchscreen and allows the
instrument to function as a conventional spectrophotometer. Up to 40 wavelengths from 190
nm to 850 nm can be monitored and reported.
The Chemometrics application allows you to use your unique chemometrics method. The
method is created using the NanoDrop QC PC Control software that is installed on your PC
and then loaded into the instrument though a USB storage device. See the NanoDrop QC PC
Control software to learn about Chemometric method features that are supported.
3
The Custom application provides additional flexibility for the method that you use with the
instrument. See the NanoDrop QC PC Control software to learn about custom method
features that are supported.
• Measure UV-Vis 12
• Measure Chemometrics 18
• Measure Custom 29
Thermo ScientificNanoDrop OneC with NanoDropQC Software User Guide 11
3
Applications
Measure UV-Vis
Measure UV-Vis
Measure UV-Vis
The UV-Vis application allows the instrument to function as a conventional
spectrophotometer. Sample absorbance is displayed on the screen from 190 nm to 850 nm.
Up to 40 wavelengths can be designated for absorbance monitoring and inclusion in the
report. Automatic pathlength adjustment and a single-point baseline correction can also be
used.
Measures the absorbance of any sample at
up to 40 wavelengths across the ultra-violet
(UV) and visible regions of the spectrum.
Measure UV-Vis
Reported Results
Settings
Detection Limits
To make UV-Vis measurements
NOTICE
• Do not use a squirt or spray bottle on or near the instrument as liquids will flow into
the instrument and may cause permanent damage.
• Do not use hydrofluoric acid (HF) on the pedestals. Fluoride ions will permanently
damage the quartz fiber optic cables.
Before you begin...
Before taking pedestal measurements with the NanoDrop One instrument, lift the instrument
arm and clean the upper and lower pedestals. At a minimum, wipe the pedestals with a new
laboratory wipe. For more information, see Cleaning the Pedestals.
To measure a sample using the UV-Vis application
1. From the Home screen, select UV-Vis.
2. Specify up to 40 wavelengths to monitor (or you can specify them later if desired) and
whether automated pathlength adjustment, analysis wavelength, and baseline correction
will be used.
12 NanoDrop One
C
with NanoDropQC Software User GuideThermo Scientific
3
Applications
Measure UV-Vis
3. Pipette 1–2 µL blanking solution onto the lower pedestal and lower the arm, or insert
the blanking cuvette into the cuvette holder.
Tip: If using a cuvette, make sure to align the cuvette light path with the instrument
light path.
4. Tap Blank and wait for the measurement to complete.
Tip: If Auto-Blank is On, the blank measurement starts automatically after you lower
the arm. (This option is not available for cuvette measurements.)
5. Lift the arm and clean both pedestals with a new laboratory wipe, or remove the
blanking cuvette.
6. Pipette 1-2 µL sample solution onto the pedestal and lower the arm, or insert the
sample cuvette into the cuvette holder.
7. Start the sample measurement:
–Pedestal: If Auto-Measure is On, lower arm; if Auto-Measure is off, lower arm and
tap Measure.
–Cuvette: Tap Measure.
When the sample measurement is completed, the spectrum and reported values are
displayed (see the next section).
8. When you are finished measuring samples, tap End Experiment.
9. Lift the arm and clean both pedestals with a new wipe, or remove the sample cuvette.
Best practices for UV-Vis measurements
• Ensure the sample absorbance is within the instrument’s absorbance detection limits.
• Blank with the same buffer solution used to re-suspend the analyte of interest. The
blanking solution should be a similar pH and ionic strength as the analyte solution.
• Run a blanking cycle to assess the absorbance contribution of your buffer solution. If the
buffer exhibits strong absorbance at or near an analysis wavelength, you may need to
choose a different buffer or application. See Choosing and Measuring a Blank for more
information.
• For micro-volume measurements:
–Ensure pedestal surfaces are properly cleaned and conditioned.
–Ensure samples are homogeneous before taking a measurement. Avoid introducing
bubbles when mixing and pipetting.
–Follow best practices for micro-volume measurements.
–Use a 1-2 µL sample volume. See Recommended Sample Volumes for more
information.
Thermo ScientificNanoDrop One
C
with NanoDropQC Software User Guide 13
3
Applications
Measure UV-Vis
• For cuvette measurements (NanoDrop OneC instruments only), use compatible cuvettes
and follow best practices for cuvette measurements.
UV-Vis Reported Results
UV-Vis measurement screen
For each measured sample, this application shows the absorbance spectrum and a summary of
the results. Here is an example as it appears on the local instrument display:
Menu of options;
tap to open
UV-Vis spectrum
Sample name;
tap to edit
Absorbance at userdefined wavelength 1
(450 nm)
Tap to edit
Absorbance at user- defined
wavelength 2 (623 nm)
Tap to edit
Tap to add
Tap row to
select sample
and update
spectrum; tap more rows to
overlay up to five
spectra. Press and hold
sample row to
view
Pinch and zoom to
adjust axes;
double-tap to reset
14 NanoDrop One
C
with NanoDropQC Software User GuideThermo Scientific
Swipe screen left to
view table with more
measurement results
Tap to end
experiment and
export data
measurement
details.
Drag tab
down/up to see
more/less
sample data
3
Applications
Measure UV-Vis
Note Micro-volume absorbance measurements and measurements taken with
nonstandard cuvettes are normalized to a 10.0 mm pathlength equivalent.
UV-Vis reported values
The initial screen that appears after each measurement (see previous image) shows a summary
of the reported values. To view all reported values, press and hold the sample row. Here is an
example:
Application
User-defined
wavelengths
Sampling
method
Baseline correction
wavelength
Sample name;
tap to edit
Absorbance
at 635 nm
Date/time
measured
Automated
pathlength
setting
Baseline
correction
absorbance
Absorbance at
450 nm
Absorbance at
623 nm
Note Scroll up to display absorbance values for any additional user-defined wavelengths.
Thermo ScientificNanoDrop One
C
with NanoDropQC Software User Guide 15
3
Applications
Measure UV-Vis
Below is an example of the measurement screen with reported values as it appears in the
NanoDrop QC PC software:
Back to Home screen
(ends experiment)
Sample name;
select to edit
Run blank
Measure sample
solution
Select to toggle
Auto-Measure ON/OFF
Tips: Auto-Measure
default is ON
Auto-Blank default
is OFF
Select to add
absorbance at userdefined wavelengths
With sample selected, click and drag an area to zoom
Right-click and select Autoscale to fit spectra to window
Tips:
Click sample row to select sample and update spectrum
Shift-click multiple sample rows to overlay up to five spectra
Click a sample and hover locations on spectra to view measurement values
End experiment
and export data
Sample rows
16 NanoDrop One
C
with NanoDropQC Software User GuideThermo Scientific
Settings for UV-Vis Measurements
To show the UV-Vis settings, from the Home screen, select UV-Vis.
SettingAvailable OptionsDescription
3
Applications
Measure UV-Vis
Monitored
wavelengths
Enter up to 40
wavelengths between
190 nm and 850 nm
Analytical WavelengthAny wavelength between
190 nm and 850 nm
Automated Pathlength On or Off
(affects pedestal
measurements only)
User-defined wavelengths to be measured and reported at run
time. Absorbance values for the first three entered wavelengths
are displayed in the measurement screen. To see absorbance
values for 8 monitored wavelengths, swipe left in the
measurement screen to show the Data table. To see all monitored
wavelengths, press and hold a sample row to show the Sample
Details screen (scroll up to display absorbance values for any
additional user-defined wavelengths).
Note: If Baseline Correction is selected, all displayed absorbance
values are the corrected values.
This is the wavelength the software will use to determine the
pathlength selection.
Optional automated pathlength selection. Allows the software
to use the optimal (shorter) pedestal pathlength for high
concentration samples to help prevent detector saturation (see
Detection Limits for details).
• When selected, the shorter pathlength is used when any
wavelength between 220 nm and 850 nm has 10 mm
equivalent absorbance value of 12.5 or higher. For
wavelengths between 190 nm and 219 nm the change to the
shorter pathlength occurs when any wavelength in this range
has a 10 mm equivalent absorbance value of 10 or higher.
• When deselected, the pedestal pathlength is restricted to
10 mm across all wavelengths.
Note: In either case, displayed absorbance values have been
normalized to a 10 mm pathlength equivalent.
Baseline CorrectionOn or off
Optional user-defined baseline correction. Can be used to
correct for any offset caused by light scattering particulates by
Enter baseline correction
wavelength in nm or use
default value (750 nm)
subtracting measured absorbance at specified baseline correction
wavelength from absorbance values at all wavelengths in sample
spectrum. As a result, absorbance of sample spectrum is zero at
specified baseline correction wavelength.
Thermo ScientificNanoDrop One
C
with NanoDropQC Software User Guide 17
3
Applications
Measure Chemometrics
Measure Chemometrics
Make chemometric measurements.
Measure Chemometrics
Create Chemometric Method
Edit Chemometric Method
Reported Results
Settings
Detection Limits
Measure Chemometrics
The NanoDrop OneC model instrument can be used to make chemometric measurements.
Note The instrument arm can be up during cuvette measurements, which allows you to
add reagents to the sample solution if desired.
To make chemometric measurements
NOTICE
• To prevent damage from spills, keep containers of liquids away from the instrument.
• Do not use a squirt or spray bottle on or near the instrument as liquids will flow into
the instrument ans may cause permanent damage.
Chemometric methods can only be created on a personal computer running the
NanoDrop QC PC Control software. If you want to run a chemometric method and store the
measurement results on the instrument, the method must also reside on the instrument. (This
is the only way to run a chemometric method if your instrument is not connected to the
computer with an Ethernet cable.)
18 NanoDrop One
C
with NanoDropQC Software User GuideThermo Scientific
3
Applications
Measure Chemometrics
Load a chemometrics method
To measure a sample using a chemometrics method, you must first load the method onto the
instrument.
To load chemometrics methods onto the instrument:
1. Export the method from the personal computer and copy the method file to the root of a
portable USB device such as a memory stick.
Method files have a “.qmethod” filename extension.
2. Connect the USB device to one of the USB ports on the instrument.
3. From the Home screen, select the Chemometric Application icon
4. Use the list box at the top of the screen to indicate the USB port used.
5. Select Load Method.
A message box shows the NanoDrop One methods available on the selected USB device.
6. Select one or more method names in the Load Method box to select the methods to load.
7. Select Load.
Thermo ScientificNanoDrop One
C
with NanoDropQC Software User Guide 19
3
Applications
Measure Chemometrics
To measure a sample using the chemometrics application
1. From the Home screen, select the Chemometrics icon.
The Chemometrics Method Selection screen is displayed. If one or more chemometric
methods exist in the currently selected Data Storage Location, they will be listed in the
Select Method box. A description of the selected method appears in the Method Details
box.
2. Select a method:
20 NanoDrop One
• select an existing method by selecting the method name in the Select Method box.
3. Select Run Method.
4. Follow the on-screen instructions to measure a sample.
C
with NanoDropQC Software User GuideThermo Scientific
3
Applications
Measure Chemometrics
To measure a sample using the chemometrics application from the PC software
1. From the Home screen, select the Chemometric Method icon.
The Chemometrics Method Selection screen is displayed. If one or more chemometric
methods exist in the currently selected Data Storage Location, they will be listed in the
method selection pane below the search feature. Details of the selected method appears in
the method details pane to the right.
2. Select a method:
• select an existing method by selecting the method name in the method selection
pane.
3. Select .
Follow the on-screen instructions to measure a sample.
Thermo ScientificNanoDrop One
C
with NanoDropQC Software User Guide 21
3
Applications
Measure Chemometrics
Create Chemometrics Method
Chemometrics methods can be created only on the NanoDrop QC PC Control software.
However, once the method is created, it can be saved in the NanoDrop One database on the
local instrument, or in the NanoDrop QC database on the PC. To create a new
Chemometrics method:
–From the NanoDrop QC software, select the Chemometric Method icon
–From the Chemometric Method Management pane, select
22 NanoDrop One
–Enter both a name and description for your chemometric method
–Select the quant method (.qnt fiile) you want to use
C
with NanoDropQC Software User GuideThermo Scientific
–Adjust the method settings as desired
3
Applications
Measure Chemometrics
–Select Save
Note The method is saved in the currently selected Data Storage Location (local
instrument or a connected PC).
–To run the method, select Run Method
Edit Chemometrics Method
Chemometrics methods can be edited only on the NanoDrop QC software. To edit an
existing chemometrics method:
–From the NanoDrop QC software, select the Chemometric Method icon
–From the Chemometric Method Management screen, select the method you would
like to edit from the list of loaded methods.
–From the drop-down menu select Edit.
–Adjust the method settings as desired. You can select the components to be displayed
on the LC.
–Select Save
Thermo ScientificNanoDrop One
C
with NanoDropQC Software User Guide 23
3
Applications
Measure Chemometrics
Chemometrics Reported Results
Chemometrics method measurement screen (shown from Data
Viewer)
For each measured sample, this application shows the absorbance spectrum and a summary of
the results. Here is an example:
Menu of options;
tap to open
UV-visible spectrum
Analyte
concentrations
Tap row to
select sample
and update
spectrum; tap more rows to
overlay up to
five spectra.
Press and
hold sample
row to view
measurement
details.
Drag tab
down/up to
see more/less
Pinch and zoom to
adjust axes;
double-tap to reset
24 NanoDrop One
C
with NanoDropQC Software User GuideThermo Scientific
sample data
Swipe screen left to
view table with more
measurement results
Note Micro-volume absorbance measurements and measurements taken with
nonstandard cuvettes are normalized to a 10.0 mm pathlength equivalent.
3
Applications
Measure Chemometrics
Chemometrics method reported values
The initial screen that appears after each measurement (see previous image) shows a summary
of the reported values. To view all reported values, press and hold the sample row. Here is an
example:
Method
name
Sampling
method
Sample name;
tap to edit
Date/time
measured
Analyte
concentration
Method
details
Thermo ScientificNanoDrop One
C
with NanoDropQC Software User Guide 25
3
Applications
Measure Chemometrics
Data Table
To see the data table, swipe the rate measurement screen (see above) to the left. Each row in
the table shows the absorbance values at all user-defined wavelengths at a given stage and time.
Scroll down to see measurement information that is out of view. The image below highlights
the available features.
Measurement
number
Sample name
Analyte concentration
Press and hold
row to view
measurement
details
26 NanoDrop One
Swipe screen right to return
to Rate measurement screen
C
with NanoDropQC Software User GuideThermo Scientific
Tap to end experiment
and export data
Loading...
+ 114 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.