Thermo Scientific LTQ XL Getting Started

Page 1
LTQ XL ®
Getting Started
97355-97042 Revision A June 2006
For Research Use Only Not for use in Diagnostic Procedures
Page 2
© 2004 Thermo Electron Corporation. All rights reserved.
®
Microsoft
E.I. du Pont de Nemours & Company. Unimetrics
and Windows® are registered trademarks of Microsoft Corporation. Teflon® is a registered trademark of
®
is a registered trademark of Unimetrics Company. Tygon® is a
registered trade mark of Saint-Gobain Performance Plastics Company. Rheodyne® and the Rheodyne logo ( ) are registered trademarks of Rheodyne, L.P.
All other trademarks are the property of Thermo Electron Corporation and its subsidiaries.
Technical information contained in this publication is for reference purposes only and is subject to change without notice. Every effort has been made to supply complete and accurate information; however, Thermo Electron Corporation assumes no responsibility and will not be liable for any errors, omissions, damage, or loss that might result from any use of this manual or the information contained therein (even if this information is properly followed and problems still arise).
This publication is not part of the Agreement of Sale between Thermo Electron Corporation and the purchaser of an LC/MS system. In the event of any conflict between the provisions of this document and those contained in Thermo Electron Corporation’s Terms and Conditions, the provisions of the Terms and Conditions shall govern.
System Configurations and Specifications supersede all previous information and are subject to change without notice.
Printing History: Revision A printed June 2006. Software Revision: LTQ 2.2, Xcalibur 2.2
Page 3
Regulatory Compliance
Thermo Electron San Jose performs complete testing and evaluation of its products to ensure full compliance with applicable domestic and international regulations. When the system is delivered to you, it meets all pertinent electromagnetic compatibility (EMC) and safety standards as described below.
EMC Directive 89/336/EEC
EMC compliance has been evaluated by TUV Rheinland of North America, Inc.
EN 55011 1999 EN 61000-4-3 2002 EN 55011 1998
IEC 61000-4-3 A1-1998
EN 61000-3-2 1995, A1; 1998,
A2; 1998, A14; 2000
EN 61000-3-3 1998 EN 61000-4-5 1995, A1; 2001 EN 61000-3-3 1998
EN 61326-1 1998, A3 EN 61000-4-6 1996, A1; 2001 EN 61326-1 1998
EN 61000-4-2 2000 EN 61000-4-11 1994, A1; 2001 EN 61000-4-2 2000
IEC 61000-4-2 2001 IEC 61000-4-11 2001-03
FCC Class A, CFR 47 Part 18
2005 CISPR 11 1999, A1; 1999, A2;
EN 61000-4-4 IEC 61000-4-4
IEC 61000-4-5 2005
IEC 61000-4-6 2004
1995, A1; 2001, A2; 2001; A2-1995
2002
EN 61000-3-2 1995, A1; 1998,
A2; 1998, A14; 2000
Low Voltage Safety Compliance
This device complies with Low Voltage Directive EN 61010-1:2001.
Changes that you make to your system may void compliance with one or more of these EMC and safety standards. Changes to your system include replacing a part or adding components, options, or peripherals not specifically authorized and qualified by Thermo Electron. To ensure continued compliance with EMC and safety standards, replacement parts and additional components, options, and peripherals must be ordered from Thermo Electron or one of its authorized representatives.
Page 4
FCC Compliance Statement
THIS DEVICE COMPLIES WITH PART 15 OF THE FCC RULES. OPERATION IS SUBJECT TO THE FOLLOWING TWO CONDITIONS: (1) THIS DEVICE MAY NOT CAUSE HARMFUL INTERFERENCE, AND (2) THIS DEVICE MUST ACCEPT ANY INTERFERENCE RECEIVED, INCLUDING INTERFERENCE THAT MAY CAUSE UNDESIRED OPERATION.
CAUTION: Read and understand the various precautionary notes, signs, and symbols contained inside this manual pertaining to the safe use and operation of this product before using the device.
Notice on Lifting and Handling of
Thermo Electron San Jose Instruments
For your safety, and in compliance with international regulations, the physical handling of this Thermo Electron San Jose instrument requires a team effort for lifting and/or moving the instrument. This instrument is too heavy and/or bulky for one person alone to handle safely.
Notice on the Proper Use of
Thermo Electron San Jose Instruments
In compliance with international regulations: If this instrument is used in a manner not specified by Thermo Electron San Jose, the protection provided by the instrument could be impaired.
Page 5
WEEE Compliance
This product is required to comply with the European Union’s Waste Electrical & Electronic Equipment (WEEE) Directive 2002/96/EC. It is marked with the following symbol:
Thermo Electron has contracted with one or more recycling/disposal companies in each EU Member State, and this product should be disposed of or recycled through them. Further information on Thermo Electron’s compliance with these Directives, the recyclers in your country, and information on Thermo Electron products which may assist the detection of substances subject to the RoHS Directive are available at
www.thermo.com/WEEERoHS.
WEEE Konformität
Dieses Produkt muss die EU Waste Electrical & Electronic Equipment (WEEE) Richtlinie 2002/96/EC erfüllen. Das Produkt ist durch folgendes Symbol gekennzeichnet:
Thermo Electron hat Vereinbarungen getroffen mit Verwertungs-/Entsorgungsanlagen in allen EU-Mitgliederstaaten und dieses Produkt muss durch diese Firmen wiederverwertet oder entsorgt werden. Mehr Informationen über die Einhaltung dieser Anweisungen durch Thermo Electron, die Verwerter und Hinweise die Ihnen nützlich sein können, die Thermo Electron Produkte zu identifizieren, die unter diese RoHS Anweisung fallen, finden Sie unter www.thermo.com/WEEERoHS
.
Page 6
Conformité DEEE
Ce produit doit être conforme à la directive européenne (2002/96/EC) des Déchets d'Equipements Electriques et Electroniques (DEEE). Il est marqué par le symbole suivant:
Thermo Electron s'est associé avec une ou plusieurs compagnies de recyclage dans chaque état membre de l’union européenne et ce produit devrait être collecté ou recyclé par celles-ci. Davantage d'informations sur la conformité de Thermo Electron à ces directives, les recycleurs dans votre pays et les informations sur les produits Thermo Electron qui peuvent aider la détection des substances sujettes à la directive RoHS sont disponibles sur www.thermo.com/WEEERoHS.
Page 7

Contents

Preface .............................................................................................xi
About This Guide ...................................................................... xi
Related Documentation .............................................................xi
Safety and Special Notices..........................................................xi
Contacting Us...........................................................................xii
Assistance...............................................................................xii
Changes to the Manual and Online Help...............................xii
Chapter 1 Introduction..........................................................................................1
Why Use the LTQ XL MS Detector?..........................................2
Which MS Detector Technique—ESI or APCI—Is Better for
Analyzing My Samples?............................................................4
Using ESI/MS..........................................................................4
Using APCI/MS.......................................................................5
Should I Use Sheath, Auxiliary, and/or Sweep Gases?..................7
How Can I Introduce My Samples into the MS Detector?..........8
What Types of Buffers Should I Use? What Types Should I
Avoid?....................................................................................10
How Should I Set Up the MS Detector for Various LC Flow
Rates?.....................................................................................11
What is Tuning and Calibration of the MS Detector All About?...
13 What Types of Experiments Can I Perform with the LTQ XL MS
Detector?................................................................................16
General MS or MSn Experiments ..........................................16
Data-Dependent Experiments................................................17
Ion Mapping Experiments .....................................................20
Ion Tree Experiments.............................................................23
Chapter 2 Setting Up the Ion Source for Tuning and Calibrating the MS
Detector27
Placing the LC/MS System in Standby......................................28
Removing the APCI Probe........................................................29
Removing the Ion Max Ion Source Housing .............................33
Installing the Ion Sweep Cone...................................................34
Installing the Ion Max Ion Source Housing ..............................36
Installing the ESI Probe ............................................................39
Thermo Electron Corporation LTQ XL Getting Started vii
Page 8
Contents
Chapter 3 Tuning and Calibrating Automatically in the ESI/MS Mode.... 43
Setting Up the Syringe Pump for Tuning and Calibration ........45
Setting Up the MS Detector in the Xcalibur Data System for
Tuning and Calibration .........................................................47
Testing the Operation of the MS Detector in the ESI/MS Mode..
52
Tuning the MS Detector Automatically in the ESI/MS Mode ..56
Saving Your ESI/MS Tune Method ..........................................60
Calibrating the MS Detector Automatically ..............................62
Cleaning the MS Detector after Tuning and Calibrating...........65
Chapter 4 Tuning with Your Analyte in LC/ESI/MS Mode........................... 69
Setting Up to Introduce Sample by Syringe Pump into Solvent
Flow from an LC ...................................................................71
Setting Up to Tune the MS Detector with Your Analyte...........74
Optimizing the MS Detector Tune Automatically with Your
Analyte...................................................................................77
Saving the ESI/MS Tune Method.............................................81
Chapter 5 Acquiring ESI Sample Data Using the Tune Plus Window ...... 83
Setting Up to Acquire MS/MS Data in the Full Scan Type.......84
Optimizing the Isolation Width and Setting Up to Optimize
the Collision Energy............................................................84
Optimizing the Collision Energy Automatically for an MS/MS
Experiment .........................................................................89
Setting Up to Introduce Sample by Loop Injection into Solvent
Flow from an LC ...................................................................92
Acquiring MS Data in the SIM Scan.........................................94
Chapter 6 Setting Up the Ion Source for Acquiring Data in APCI/MS/MS
Mode101
Removing the ESI Probe .........................................................102
Removing the Ion Max Ion Source Housing ...........................105
Removing the Ion Sweep Cone ...............................................106
Installing the Corona Needle ..................................................107
Installing the Ion Max Ion Source Housing ............................108
Installing the APCI Probe .......................................................111
Chapter 7 Optimizing the MS Detector with Your Analyte in APCI/MS
Mode115
Setting Up the Inlet for Tuning Using High-Flow Infusion....116
Setting Up the MS Detector for APCI/MS Operation ............119
viii LTQ XL Getting Started Thermo Electron Corporation
Page 9
Contents
Optimizing the Tune of the MS Detector Automatically in
APCI/MS Mode...................................................................123
Saving the APCI/MS Tune Method........................................126
Cleaning the MS Detector after Tuning in APCI Mode..........128
Chapter 8 Acquiring APCI Sample Data Using the Tune Plus Window..131
Setting Up to Introduce Sample by Loop Injection into Solvent
Flow from an LC .................................................................132
Acquiring APCI Data in the SIM Scan Mode .........................134
Appendix A Sample Formulations......................................................................139
Caffeine, MRFA, and Ultramark 1621 Stock Solutions ..........140
Stock Solution: Caffeine.......................................................141
Stock Solution: MRFA.........................................................142
Stock Solution: Ultramark 1621 ..........................................142
ESI Calibration Solution: Caffeine, MRFA, Ultramark 1621..143
Reserpine ................................................................................144
Stock Solution: Reserpine ....................................................144
Reserpine Tuning Solution and Reserpine APCI Sample
Solution ............................................................................144
Appendix B LTQ XL High Mass Range Calibration .........................................147
High Mass Range Calibration Solution...................................148
Normal Mass Range Calibration .............................................149
Enter Normal Mass Range Data into Tune Plus .....................150
Tune on m/z 524.3 .................................................................152
High Mass Range Calibration Procedure.................................154
Notes ......................................................................................159
Index..................................................................................................161
Thermo Electron Corporation LTQ XL Getting Started ix
Page 10
Page 11

Preface

About This Guide Welcome to the Thermo Electron, LTQ XL™ system. The LTQ XL is a

member of the Thermo family of MS detectors.
This LTQ XL Getting Started manual provides information on how to set up, calibrate, and tune the LTQ XL MS detector, and how to acquire LC/MS data. All of these procedures can be performed from the Xcalibur® Tune Plus window.
To perform analyses in ESI mode, see Chapters 2, 3, 4, and 5. To perform analyses in APCI mode, see Chapters 2, 3, 6, 7, and 8.
Related
Documentation
Safety and Special
Notices
In addition to this guide, Thermo Electron provides the following documents for the LTQ XL system:
• LTQ XL Preinstallation Guide
• LTQ XL Getting Connected
• LTQ XL Hardware Manual
Help is also available from within the software.
Make sure you follow the precautionary statements presented in this guide. The safety and other special notices appear in boxes.
Safety and special notices include the following:
CAUTION Highlights laser-related hazards to human beings. It includes information specific to the class of laser involved. Each DANGER notice is accompanied by the international laser radiation symbol.
CAUTION Highlights hazards to humans, property, or the environment. Each CAUTION notice is accompanied by an appropriate CAUTION symbol.
Thermo Electron Corporation LTQ XL Getting Started xi
Page 12
Preface

Contacting Us

IMPORTANT Highlights information necessary to avoid damage to software, loss of data, invalid test results, or information critical for optimal performance of the system.
Note Highlights information of general interest.
Note Helpful information that can make a task easier.
Contacting Us There are several ways to contact Thermo Electron.

Assistance For new product updates, technical support, and ordering information,

contact us in one of the following ways:
Visit Us on the Web
Changes to the Manual
and Online Help
www.thermo.com/finnigan
Contact Technical Support
Phone: 1-800-685-9535 Fax: 1-561-688-8736
techsupport.finnigan@thermo.com
Contact Customer Service
In the US and Canada for ordering information: Phone: 1-800-532-4752 Fax: 1-561-688-8731
International contacts for ordering information: Visit www.thermo.com/finnigan for the current listing,
To suggest changes to this guide or to the Help, use either of the following methods:
• Fill out a reader survey online at www.thermo.com/lcms-techpubs
• Send an e-mail message to the Technical Publications Editor at
techpubs.finnigan-lcms@thermo.com
xii LTQ XL Getting Started Thermo Electron Corporation
Page 13

Chapter 1 Introduction

The LTQ XL™ is a member of the Thermo family of MS detectors. The LTQ XL MS detector is an advanced analytical instrument that includes a syringe pump, a divert/inject valve, an atmospheric pressure ionization (API) source, an MS detector, and the Xcalibur data system. In a typical analysis, a sample can be introduced in any one of the following ways:
• Using the syringe pump (direct infusion)
• Using the inject valve fitted with a loop and an LC pump (flow injection analysis)
• Using a valve and an LC system fitted with a column (LC/MS)
In analysis by LC/MS, a sample is injected onto an LC column. The sample is then separated into its various components. The components elute from the LC column and pass into the MS detector where they are analyzed. Analysis by direct infusion or flow injection provides no chromatographic separation of components in the sample before it passes into the MS detector. The data from the MS detector is then stored and processed by the Xcalibur data system.
This introduction answers the following questions:
Why Use the LTQ XL MS Detector?
Which MS Detector Technique—ESI or APCI—Is Better for Analyzing
My Samples?
How Can I Introduce My Samples into the MS Detector?
What Types of Buffers Should I Use? What Types Should I Avoid?
How Should I Set Up the MS Detector for Various LC Flow Rates?
What is Tuning and Calibration of the MS Detector All About?
What Types of Experiments Can I Perform with the LTQ XL MS
Detector?
Thermo Electron Corporation LTQ XL Getting Started 1
Page 14
1
Introduction

Why Use the LTQ XL MS Detector?

Why Use the LTQ XL
MS Detector?
The attribute that sets the LTQ XL MS detector apart from other LC detectors is the high level of analytical specificity that it provides. The LTQ XL MS detector can provide multiple levels of analysis. Each level of analysis adds a new dimension of specificity for positive compound identification. The various levels of analysis are as follows:
• Chromatographic separation and compound detection (non MS technique utilizing chromatographic retention time)
• Mass analysis (molecular mass information)
• Two-stage mass analysis, MS/MS (structural information)
• Multiple MS-MS mass analysis, MSn (structural information)
• ZoomScan™ analysis (charge state information)
Chromatographic separation and compound detection can be obtained by all LC/detector systems. Retention time alone, however, does not positively identify a compound because many compounds can have the same retention time under the same experimental conditions. In addition, even if a compound is identified correctly by retention time, quantitation results can be in error because other compounds in the sample might co-elute with the compound of interest.
Single stage mass analysis allows for the identification of analytes of interest. Atmospheric pressure ionization typically produces mass spectra that provide molecular mass information.
Two-stage mass analysis allows for even more positive compound identification. MS/MS analysis monitors how a parent ion fragments when exposed to an additional stage of ionization. There are two types of MS/MS analysis: Full Scan MS/MS and Selective Reaction Monitoring (SRM). Full Scan MS/MS monitors the production of all product ion from a specific parent ion. SRM MS/MS analysis monitors a specific reaction path: the production of a specific product ion from a specific parent ion. Using MS/MS analysis, you can easily quantitate target analytes in complex matrices such as plant or animal tissue, plasma, urine, groundwater, or soil. Because of the specificity of MS/MS measurements and the ability to eliminate interferences by an initial mass selection stage, quantitative target compound analysis is easily accomplished using the LTQ XL MS detector.
Multiple MS/MS mass analysis provides a unique capability to obtain structural information that can be useful in structure elucidation of
n
metabolites, natural products, and sugars. MS
techniques on the LTQ XL MS detector allow for stepwise fragmentation pathways, making interpretation of MSn spectra relatively straightforward. The LTQ XL MS
2 LTQ XL Getting Started Thermo Electron Corporation
Page 15
1
Introduction
Why Use the LTQ XL MS Detector?
detector has several advanced features that make its MSn capabilities extremely powerful for qualitative analysis. (See “What Types of
Experiments Can I Perform with the LTQ XL MS Detector?” on page 16.)
ZoomScan analysis provides information about the charge state of one or more mass ions of interest. ZoomScan data is collected by using slower scans at higher resolution. This allows for unambiguous determination of charge state, which in turn allows for the correct determination of molecular mass.
In addition to the aforementioned levels of analysis, there is an additional technique called Wideband Activation. The Wideband Activation option allows the LTQ XL MS detector to apply collision energy to ions during MS/MS fragmentation over a fixed mass range of 20 u. This option allows the LTQ XL MS detector to apply collision energy to both the parent ion, as well as to product ions created as a result of non-specific losses of water (18 u) or ammonia (17 u), for example, or to product ions formed from the loss of fragments less than 20 u. When you want enhanced structural information and you do not want to perform MS3 analysis with the LTQ XL MS detector, choose the Wideband Activation option for qualitative MS/MS. Because the collision energy is applied to a broad mass range, signal sensitivity is somewhat reduced when you choose this option. Therefore, increase the value of the collision energy (Activation Amplitude) to compensate somewhat for the reduction of sensitivity.
Thermo Electron Corporation LTQ XL Getting Started 3
Page 16
1
Introduction
Which MS Detector Technique—ESI or APCI—Is Better for Analyzing My Samples?
Which MS Detector
Technique—ESI or
APCI—Is Better for
Analyzing My
Samples?

Using ESI/MS The ESI mode typically produces mass spectra consisting of multiply

The LTQ XL MS detector includes two standard atmospheric pressure ionization source probes:
• Electrospray ionization (ESI) probe
• Atmospheric pressure chemical ionization (APCI) probe
Typically, more polar compounds such as amines, peptides, and proteins are best analyzed by ESI, and nonpolar compounds such as steroids are best analyzed by APCI.
Sample ions can carry a single charge or multiple charges. The number of charges carried by the sample ions depends on the structure of the analyte of interest, the mobile phase, and the ionization mode.
charged ions (for proteins and peptides) depending on the structure of the analyte and the solvent. For example, the resulting mass spectrum of a higher molecular mass protein or peptide typically consists of a distribution of multiply charged analyte ions. The resulting mass spectrum can be mathematically manipulated to determine the molecular mass of the sample.
1
The ESI mode transfers ions in solution into the gas phase. Many samples that previously were not suitable for mass analysis (for example, heat-labile compounds or high molecular mass compounds) can be analyzed by ESI. ESI can be used to analyze any polar compound that makes a preformed ion in solution. The term preformed ion can include adduct ions. For example, polyethylene glycols can be analyzed from a solution containing ammonium acetate, because of adduct formation between the NH and oxygen atoms in the polymer. With ESI, the range of molecular masses that can be analyzed by the LTQ XL MS detector is greater than 100,000 u, due to multiple charging. ESI is especially useful for the mass analysis of polar compounds, which include: biological polymers (for example, proteins, peptides, glycoproteins, and nucleotides); pharmaceuticals and their metabolites; and industrial polymers.
You can use the ESI mode in either positive or negative ion polarity mode. The ion polarity mode is determined by the polarity of the preformed ions in solution: Acidic molecules form negative ions in high pH solution, and basic molecules form positive ions in low pH solution. A positively charged ESI needle is used to generate positive ions and a negatively charged needle is used to generate negative ions.
+
ions in the solution
4
1
Optional ionization sources [atmospheric photo ionization (APPI), atmospheric pressure matrix
assisted laser desorption ionization (AP MALDI), and nanospray] are also available.
4 LTQ XL Getting Started Thermo Electron Corporation
Page 17
1
Introduction
Which MS Detector Technique—ESI or APCI—Is Better for Analyzing My Samples?
You can vary the flow rate from the LC into the MS detector over a range from 1 μL/min to 1000 μL/min. See Ta bl e 3 . (In ESI, the buffer and the buffer strength both have a noticeable effect on sensitivity. Therefore, it is important to choose these variables correctly.) In the case of higher molecular mass proteins or peptides, the resulting mass spectrum consists typically of a series of peaks corresponding to a distribution of multiply charged analyte ions.
The ESI process is affected by droplet size, surface charge, liquid surface tension, solvent volatility, and ion solvation strength. Large droplets with high surface tension, low volatility, strong ion solvation, low surface charge, and high conductivity prevent good electrospray.
Mixed organic/aqueous solvent systems that include organic solvents such as methanol, acetonitrile, and isopropyl alcohol are superior to water alone for ESI. Volatile acids and bases are good, but salts above 10 mM are not recommended. Strong mineral acids and bases are extremely detrimental to the instrument.
The rules for a good electrospray are as follows:
• Keep non-volatile salts and buffers out of the solvent system. For example, avoid the use of salts containing sodium or potassium and avoid the use of phosphates. If necessary, use ammonium salts instead.
• Use organic/aqueous solvent systems and volatile acids and bases.
• If possible, optimize the pH of the solvent system for your analyte of interest. For example, if your analyte of interest contains a primary or secondary amine, your mobile phase should be slightly acidic (pH 2 to 5). The acid pH tends to keep positive ions in solution.

Using APCI/MS Like ESI, APCI is a soft ionization technique. APCI provides molecular

mass information for compounds of medium polarity that have some volatility. APCI is typically used to analyze small molecules with molecular masses up to about 2000 Da.
APCI is a gas phase ionization technique. Therefore, the gas phase acidities and basicities of the analyte and solvent vapor play an important role in the APCI process.
APCI is a very robust ionization technique. It is not affected by minor changes in most variables such as changes in buffer or buffer strength. The rate of solvent flowing from the LC into the MS detector in APCI mode is typically high (between 0.2 and 2 mL/min). See Ta bl e 3 .
Thermo Electron Corporation LTQ XL Getting Started 5
Page 18
1
Introduction
Which MS Detector Technique—ESI or APCI—Is Better for Analyzing My Samples?
You can use APCI in positive or negative ion polarity mode. For most molecules, the positive-ion mode produces a stronger ion current. This is especially true for molecules with one or more basic nitrogen (or other basic) atoms. Molecules which generally produce strong negative ions, with acidic sites such as carboxylic acids and acid alcohols, are an exception to this general rule.
Although, in general, fewer negative ions are produced than positive ions, negative ion polarity can be more specific. This is because the negative ion polarity mode sometimes generates less chemical noise than does the positive mode. Thus, the signal-to-noise ratio might be better in the negative ion mode than in the positive ion mode.
6 LTQ XL Getting Started Thermo Electron Corporation
Page 19
1
Introduction

Should I Use Sheath, Auxiliary, and/or Sweep Gases?

Should I Use Sheath,
Auxiliary, and/or
Sweep Gases?
Nitrogen gas can be applied to the system using any combination of the three gas sources: Auxiliary gas, Sweep gas, and/or Sheath gas. When Sheath gas is used, nitrogen is applied as an inner coaxial gas (when used in tandem with Auxiliary gas), helping to nebulize the sample solution into a fine mist as the sample solution exits the ESI or APCI nozzle. (Sheath gas is not used with the NSI source.) When Auxiliary gas is being used, nitrogen flows through the ion source nozzle, the vapor plume is affected; the spray is focused and desolvation is improved. When Sweep gas is used, the nitrogen flows out from behind the sweep cone and can result in solvent declustering and adduct reduction.
When you are analyzing complex matrices such as plasma or nonvolatile salt buffers, Sweep gas is required for ruggedness. In full-scan MS or data dependent scan experiments, the signal-to-noise ratio can be improved by application of Sweep gas. In some cases, signal intensity can be increased by using Auxiliary gas, particularly for higher LC flow rates.
All analyses are analyte dependent and require separate optimization with Sheath gas, Sweep gas, and Auxiliary gas to determine which combination will yield optimum performance. It is especially important to optimize with each gas independently before you perform experiments using MS
techniques and before you perform any quantitative analysis experiments
because optimum results could be achieved with any combination of Sheath, Sweep, and/or Auxiliary gas. See Ta bl e 2 and Tab le 3 for additional information on using supplemental gas flows.
n
Thermo Electron Corporation LTQ XL Getting Started 7
Page 20
1
Introduction

How Can I Introduce My Samples into the MS Detector?

How Can I Introduce My Samples into the
MS Detector?
You can introduce your samples into the MS detector in a variety of ways. Refer to Ta bl e 1 .
The syringe pump is often used to introduce calibration solution for automatic tuning and calibrating in ESI mode. You can also use this technique to introduce a solution of pure analyte at a steady rate in ESI mode, for example, for determining the structure of an unknown compound.
You can also use a Tee union to direct samples from the syringe pump into an LC flow (without a column), which then enters the MS detector. This technique is used to introduce sample at a steady rate and at higher solvent flow rates; it is used especially for tuning in ESI or APCI on an analyte of interest. You can also use this technique to introduce a solution of pure analyte at a steady rate in ESI or APCI.
You can introduce samples from a syringe into the loop of the injector valve. You can then use the divert valve to introduce the sample into an LC flow, which then enters the MS detector. This technique is used in ESI or APCI to introduce pure analytes into the MS detector in a slug. It is useful when you have a limited quantity of pure analyte.
You can also use an LC autosampler to introduce samples into an LC flow. This technique is also used in ESI or APCI to introduce a slug of pure analyte into the LC flow and then into the MS detector.
Finally, you can perform LC/MS experiments by using an LC autosampler to introduce a mixture onto an LC column. This technique is used with ESI or APCI to separate the analytes before they are introduced sequentially into the MS detector.
You can refer to subsequent chapters in this manual and to LTQ XL Getti n g Connected for plumbing diagrams and methods of sample introduction.
Table 1. Sample introduction techniques
Syringe Pump Flow (no LC Flow)
Sample Introduction Technique
Syringe pump* ESI automatic tuning
Analytical Technique
and calibrating
ESI analysis of a pure analyte solution
Figure Reference
LTQ XL Getting Started
Figure 2-5
8 LTQ XL Getting Started Thermo Electron Corporation
Page 21
How Can I Introduce My Samples into the MS Detector?
Table 1. Sample introduction techniques
1
Introduction
LC Flow Without Chromatographic Separation (no column)
LC Flow With Chromatographic Separation
Sample Introduction Technique
Syringe pump into LC flow (connected by Tee union)*
Loop injection into LC flow ESI or APCI analysis
Autosampler injection into LC flow (one or multiple injections)
Autosampler injections into LC column via LC flow (one or multiple injections
Analytical Technique
ESI or APCI automatic optimization of tuning on analyte of interest
ESI or APCI analysis of a pure analyte solution
of a pure analyte solution
ESI or APCI analysis of a pure analyte solution
ESI or APCI analysis of mixtures
Figure Reference
LTQ XL Getting Started
Figure 4-1 (ESI) Figure 6-1 (APCI)
LTQ XL Getting Started
Figure 5-6 (ESI) Figure 8-1 (APCI)
LTQ XL Getting Connected
Figure 11-5 (ESI) Figure 11-8 (APCI)
*Provides steady state introduction of sample (direct infusion)
Thermo Electron Corporation LTQ XL Getting Started 9
Page 22
1
Introduction

What Types of Buffers Should I Use? What Types Should I Avoid?

What Types of Buffers
Should I Use? What
Types Should I Avoid?
Many LC applications use nonvolatile buffers such as phosphate and borate buffers. It is best to avoid the use of nonvolatile buffers with the MS detector because they can cause the following problems:
• Blocking the capillary in the probe
• Causing salt buildup on the spray head and thus compromising the integrity of the spray
Use volatile buffers when you use the MS detector. Many volatile buffer solutions are available that can be used instead of nonvolatile ones. Volatile buffer solutions can include the following:
• Acetic acid
• Ammonium acetate
• Ammonium formate
• Ammonium hydroxide
• Triethylamine (TEA)
• Trifluoroacetic acid
10 LTQ XL Getting Started Thermo Electron Corporation
Page 23
1
Introduction

How Should I Set Up the MS Detector for Various LC Flow Rates?

How Should I Set Up
the MS Detector for
Various LC Flow
Rates?
The ESI probe can generate ions from liquid flows2 of 1 μL/min to
1.0 mL/min. This flow rate range allows you to use a wide range of separation techniques: CE, CEC, capillary LC, microbore LC, and analytical LC.
The APCI probe can generate ions from liquid flows3 of 200 μL/min to
2.0 mL/min. This flow range allows you to use microbore LC, analytical LC, and semi-preparative LC.
As you change the rate of flow of solvents entering the MS detector, you need to adjust several of the MS detector parameters, as follows:
For ESI, you need to adjust the capillary temperature and adjust the gas flow rates for the Sheath, Auxiliary, and/or Sweep gas.
For APCI, you need to adjust the capillary temperature and vaporizer temperature and adjust the gas flow rates for the Sheath, Auxiliary, and/or Sweep gas.
In general, an increase in the rate of liquid flowing into the MS detector, requires a higher temperature of the ion transfer capillary (and vaporizer) and the higher gas flow rate.
Ta bl e 2 provides guidelines for ESI operation for ion transfer capillary
temperatures and gas flow rates for various LC solvent flow rates.
Ta bl e 3 provides guidelines for APCI operation for the ion transfer capillary
temperature, vaporizer temperature, and gas flow rate for a range of LC solvent flow rates.
2
The ESI probe can generate ions from liquid flows of as low as 1 μL/min. However, flows below 5 μL/min require more care, especially with the position of the fused silica sample tube within the ESI probe.
3
For the APCI probe, flows below 200 μL/min require more care to maintain a stable spray.
Thermo Electron Corporation LTQ XL Getting Started 11
Page 24
1
Introduction
How Should I Set Up the MS Detector for Various LC Flow Rates?
Table 2. Guidelines for setting operating parameters for LC/ESI/MS
Ion Transfer Capillary Temperatu re
setting: 150 to 200
°C
setting: 200 to 275
°C
setting: 250 to 350
°C
Sheath Gas
Not required
Typica l setting: 5 to 15 units
Required
Typica l setting: 20 to 40 units
Required
Typica l setting: 40 to 60 units
Auxiliary and/or Sweep Gas
Not required
Typical setting: 0 units
Not required, but might help depending on conditions
Typical setting: 0 to 20 units
Not required, but usually helps to reduce solvent background ions
Typical setting: 0 to 20 units
LC Flow Rates
Infusion or LC at flow rates of < 10 μL/min
LC at flow rates from 50 to 200 μL/min
LC at flow rates from 100 to 500 μL/min
Suggeste d Column Size
Capillary Typ ic al
1 mm ID Ty pi ca l
2 to 3 mm IDTypi ca l
*
LC at flow rates from 0.4 to 1mL/min
*
Note: Be sure to choose either Auxiliary gas and/or Sweep gas according to the hints in
Should I Use Sheath, Auxiliary, and/or Sweep Gases?
Table 3. Guidelines for setting operating parameters for LC/APCI/MS
LC Flow Rate
LC at flow rates from
0.2 to 2mL/min
*
Note: Be sure to choose either Auxiliary gas and/or Sweep gas according to the hints in Should
I Use Sheath, Auxiliary, and/or Sweep Gases?
4.6 mm ID Typ ic al
Ion Transfer Capillary Temperatur e
Typi ca l setting: 150 to 225
°C
setting: 300 to 400
°C
Vaporizer Temperatur e
Typi cal setting: 400 to 550
°C
Required
Typica l setting: 60 to 100 units
Sheath Gas
Required
Typi ca l setting: 40 to 100 units
Required
Typical setting: 10 to 40 units
*
Auxiliary and/or Sweep Gas
Not required, but usually helps to reduce solvent background ions
Typical setting: 0 to 20 units
12 LTQ XL Getting Started Thermo Electron Corporation
Page 25
1
Introduction

What is Tuning and Calibration of the MS Detector All About?

What is Tuning and
Calibration of the MS
Detector All About?
To optimize the performance of data acquisition on the LTQ XL MS detector, tune and calibrate in four steps:
• In ESI mode, you infuse a calibration solution into the MS detector at a steady rate of 5 μL/min for several minutes. In Tune Plus, you observe the signal at m/z 195, the mass-to-charge ratio of caffeine in the calibration solution. Then, while observing the signal at m/z 195, you adjust probe positions and gas flows to achieve the greatest signal strength while still maintaining a stable spray of ions into the MS detector.
• Once you have established a stable spray of ions into the MS detector, tune the MS detector. In this step, you use the automatic tuning procedure in Tune Plus to ensure that the transmission of ions into the MS detector is optimum. You observe the Tune Plus window as the Xcalibur data system tunes your LTQ XL MS detector automatically.
• After your tune method is optimized, calibrate the MS detector. In this step, you want to ensure that the calibration parameters complete automatic calibration successfully. The Calibrate dialog box in Tune Plus provides a readback of the status of the calibration parameters, both during the automatic calibration and when calibration is complete.
• Lastly, if you want to maximize the detection of one or more particular ions, you can optimize the tune of the MS detector with your analyte of interest in the ionization mode that you are going to use to analyze your samples. You choose a mass-to-charge ratio of your analyte of interest. Alternatively, you can choose an ion in the calibration solution that is closest to the mass-to-charge ratio for your ion of interest. (It is sometimes possible to acquire qualitative data without optimizing the parameters, but detection sensitivity might be compromised.)
Calibration parameters are instrument parameters whose values do not vary with the type of experiment. It is recommended that you calibrate the MS detector at least once every three months and that you check the calibration about once a week.
Automatic and semi-automatic calibration (including checking the calibration) require that you introduce calibration solution into the MS detector at a steady flow rate while the procedure is running. You introduce the solution directly from the syringe pump into the MS detector in the ESI/MS mode.
Thermo Electron Corporation LTQ XL Getting Started 13
Page 26
1
Introduction
What is Tuning and Calibration of the MS Detector All About?
Tune parameters are instrument parameters whose values can vary with the type of experiment. For example, if your experiment requires quantitative data on one or more particular ions, you need to tune the MS detector with your analyte if you change any one of the parameters specific to the experiment or analyte.
Automatic and semi-automatic tuning procedures (including optimizing the collision energy) require that you introduce calibration solution, or a tuning solution of your analyte of interest, into the MS detector at a steady rate in either of two ways:
• Introduce the solution directly from the syringe pump. See Setting Up
• Introduce the sample from the syringe pump into the effluent of the LC
the Syringe Pump for Tuning and Calibration in Chapter 3: “Tuning and Calibrating Automatically in the ESI/MS Mode”.
by using a Tee union. See Setting Up to Introduce Sample by Syringe
Pump into Solvent Flow from an LC in Chapter 4: “Tuning with Your Analyte in LC/ESI/MS Mode”.
Use the first method for tuning if you intend to use an experiment type at a low flow rate involving the syringe pump. The second method is useful if you intend to use an experiment type at a higher flow rate involving the LC. However, the second method of introduction puts a comparatively large amount of analyte into the MS detector. Therefore, before you can perform an analytical run to analyze for the analyte, you might need to clean the API spray shield.
CAUTION Do not use the calibration solution at flow rates above 10 μL/min. Ultramark 1621 can contaminate your system at high concentrations.
In most cases, you can use the tune you obtain from the automatic or semi-automatic tuning procedures for your analytical experiments. However, for some applications, you might need to tune several MS detector parameters. In that case, you would tune manually. With the manual tuning process, you introduce a tuning solution at a steady flow rate.
Note The most important parameters that affect the signal quality during ESI/MS operation are the ion transfer capillary temperature, tube lens voltage, gases, and solution flow rate. For optimum sensitivity, tune with the instrument in the same operational mode as the mode you use for the analytical experiment.
14 LTQ XL Getting Started Thermo Electron Corporation
Page 27
What is Tuning and Calibration of the MS Detector All About?
Ta bl e 4 summarizes methods of sample introduction for each of the
calibration and tuning procedures.
Table 4. Summary of methods of sampleintroduction for calibration and tuning
Calibrating Tuni ng
1
Introduction
Sample/ Sample Intro
Calibration solution/ Syringe pump
Your tune solution/ Syringe pump
Your tune solution/ Syringe pump into LC flow by using Tee union
Check Auto
Semi­auto
Auto
Semi­auto
Manual
9 9 9 9 9 9 9
9 9 9 9
9 9 9 9
Collision Energy
Thermo Electron Corporation LTQ XL Getting Started 15
Page 28
1
Introduction

What Types of Experiments Can I Perform with the LTQ XL MS Detector?

What Types of
Experiments Can I
Perform with the LTQ
XL MS Detector?
General MS or MSn
Experiments
This topic describes several types of experiments that you can perform with the LTQ XL MS detector. The experiments can be grouped into the following categories:
• General MS or MS
• Data-Dependent™
• Ion Mapping™
• Ion Tree
You can specify which type of experiment you want to perform in the Instrument Setup window, and then save it in an Instrument Method (.METH) file.
Note Procedures for these experiments are beyond the scope of this LTQ XL Getting Started manual. If you need more information, refer to online
Help.
You can use a General experiment to collect qualitative data for structural analysis. The Xcalibur data system includes an Instrument Method template in Instrument Setup so you can get started with a General MS or MSn experiment. For an example of a General MS or MSn experiment template, see Figure 1
n
In a General MS experiment, you need to specify the mass range of your analyte(s) of interest. In a General MS/MS experiment, you need to specify a parent (precursor ion) that fragments into distinctive product ions. In a
n
General MS the parent ions of interest. The LTQ XL MS detector can then collect data on the ions in the range or on the product ions of the parent ion(s) that you specify.
If you use a General experiment to collect data for qualitative (structural) analysis, you specify the scan mode (MS through MS data in the Scan Event Settings group box. If you specify MS/MS or MS you then choose the parent ion(s) for which you want data in the MSn Settings table. The LTQ XL MS detector can then collect distinct qualitative information for structural analysis or for spectral reference.
experiment, you need to specify the mass-to-charge ratios of all
n
) for which you want
n
,
16 LTQ XL Getting Started Thermo Electron Corporation
Page 29
1
Introduction
What Types of Experiments Can I Perform with the LTQ XL MS Detector?
Figure 1. MS Detector Setup page in Instrument Setup, showing a template for a General MS experiment
The LTQ XL MS detector can generate reproducible, analyte-specific spectra, even from laboratory to laboratory. Consequently, reference spectra that are generated with the LTQ XL MS detector can be used to confirm structures of compounds generated with other LTQ XL systems.
Data-Dependent
Experiments
A Data-Dependent experiment is best used for the qualitative analysis of unknown compounds for structure elucidation or confirmation. The LTQ XL MS detector uses the information in a data-dependent experiment to make decisions about the next step of the experiment automatically—without input from a user. Instrument Setup contains the Instrument Method templates that you need to get started with data-dependent experiments. For an example of a data-dependent Triple Play experiment template, see Figure 2.
Thermo Electron Corporation LTQ XL Getting Started 17
Page 30
1
Introduction
What Types of Experiments Can I Perform with the LTQ XL MS Detector?
A data-dependent experiment produces a great deal of data from a single sample analysis. You can run a data-dependent experiment even if you know very little about your sample, and even if you are unfamiliar with the variables of mass spectroscopy. In a data-dependent experiment, you can specify parent ions for fragmentation or you can let the LTQ XL MS detector automatically select the ions for fragmentation. The LTQ XL MS detector can collect the structural information for every parent ion in the sample automatically, even if the sample is a mixture of compounds.
A data-dependent experiment requires minimal input from a user about how the experiment should best proceed. The user specifies that one or more scan events of an experiment segment are to be run as data-dependent. Then, the LTQ XL MS detector collects MS/MS or MSn data and makes decisions about what the next step in the experiment should be to collect even more data. For example, in a data-dependent Triple Play experiment for a mixture of compounds, the LTQ XL MS detector can decide which parent ion to isolate, the charge state of the parent ion, and the molecular mass of the compound.
Ion Mapping experiments can be data-dependent. (The Total Ion Map, Neutral Loss Ion Map, and Parent Ion Map experiments are not data-dependent.) The Data-Dependent Zoom Map experiment collects ZoomScan data on every scan interval in a specified mass range.
Ion Tree experiments are types of data-dependent experiments. These experiments provide methods for automatically interpreting MSn data and arranging the data in formats that are easy to manipulate.
You can approach the setup of data-dependent experiments in either of two ways:
• If you have some idea of the parent ion, or if you expect a certain kind of parent, you can set up a list of possible parent ions. Then, when one of the parent ions you specified is detected, you can acquire product spectra and analyze the information. Conversely, you can also set up a list of ions that you do not want to be selected for fragmentation.
18 LTQ XL Getting Started Thermo Electron Corporation
Page 31
1
Introduction
What Types of Experiments Can I Perform with the LTQ XL MS Detector?
Figure 2. MS Detector Setup page in Instrument Setup, showing a template for a Data-Dependent Triple Play
experiment. (To select a scan event that makes active the Dependent Scan check box, click either the Scan Event 2 or Scan Event 3 button.
• If you have little information about your compound, you can set up the parameters of a data-dependent experiment so that if the intensity of the ion signal is above a specified threshold, the LTQ XL MS detector generates product spectra. Parameters that you might specify, for
n
example, include threshold values for the intensity of the MS or MS
ion signal. Whatever threshold values you choose should accomplish the isolation of your parent ions of interest.
You can find useful structural information about your compound automatically with the simplest data-dependent experiment, Data-Dependent MS/MS. You specify the MS scan range, and you do not
Thermo Electron Corporation LTQ XL Getting Started 19
Page 32
1
Introduction
What Types of Experiments Can I Perform with the LTQ XL MS Detector?
even need to specify a parent ion. The LTQ XL MS detector can then collect full scan MS data, pick the most intense parent ion in the spectrum, and fragment the ion to generate product ions.
A Data-Dependent Triple-Play experiment is the same as Data-Dependent MS/MS, but includes the identification of the charge state of the parent with the LTQ XL ZoomScan feature. A Data-Dependent Triple-Play experiment collects full scan MS data, and then uses ZoomScan to determine the charge state of the parent ion and calculate the molecular mass. The parent ion is then fragmented into product ions (MS/MS). For example, if the LTQ XL MS detector determines a charge state equal to 2, and if the mass-to-charge ratio of the parent ion is m/z 500, then the mass-to-charge ratios of the product ions can be up to m/z 1000 (or 2 ×
500).
Use a data-dependent experiment (from templates in Instrument Setup) to do the following:
• Identify low-level impurities in high-purity compounds (Data-Dependent MS/MS)
• Identify metabolites in a complex mixture (Chromatographic Separation with Data-Dependent MS/MS)
• Build a custom library of composite MSn spectra (Ion Tree)
You can use a Data-Dependent MSn experiment to identify process impurities. In the quality assurance process for aspirin, for example, the LTQ XL MS detector can identify impurities of less than 0.1%.
A Data-Dependent MS/MS experiment of a complex mixture of drug metabolites can provide highly specific structural information. Characteristic masses along the metabolic pathways of a drug, for example, can produce MS/MS spectra that are specific to the structure of the drug. These spectra are essential in metabolite identification.
A data-dependent experiment can produce a composite spectrum of, for example, MS2, MS3, and MS4 data. The LTQ XL MS detector can store the MSn fingerprint data in a custom MSn library spectrum. The data is valuable for use in process control, quality assurance, or research.

Ion Mapping Experiments An Ion Mapping experiment is best used to get full structural

characterization of unknown molecules in complex mixtures. In an Ion Mapping experiment, you can get product ion scans on every parent ion over a specified mass range. An Ion Mapping experiment can help to
20 LTQ XL Getting Started Thermo Electron Corporation
Page 33
1
Introduction
What Types of Experiments Can I Perform with the LTQ XL MS Detector?
identify automatically which parent ions were fragmented to yield a specified product ion. The experiment “maps” one or more parent ions by using the information from product ion scans.
The LTQ XL MS detector includes the following Ion Mapping templates in Instrument Setup so you can get started with an Ion Mapping experiment:
• Total (or full scan) Ion Map
• Neutral Loss Ion Map
• Parent Ion Map
These Ion Mapping experiments, in general, require that sample solution enter the MS Detector at a composition that is constant throughout. Therefore, you use infusion to introduce your sample for these Ion Mapping experiments. See Figure 3 for an example of an Ion Mapping experiment template.
Thermo Electron Corporation LTQ XL Getting Started 21
Page 34
1
Introduction
What Types of Experiments Can I Perform with the LTQ XL MS Detector?
Figure 3. Total Ion Map page in Instrument Setup, showing a template that contains parameters for an Ion Mapping
experiment
In a Total (or full scan) Ion Mapping experiment, you get product ion scans for each parent ion, so you can determine which parent ions lost a particular fragment to yield a particular product ion. Furthermore, you can determine which parent ions are related to specific product ions. For example, you can map the spectral peaks in a mass range from m/z 400 to m/z 2000 and specify to scan for MS/MS product ions in incremental steps of every mass-to-charge ratio, every fifth mass-to-charge ratio, or every tenth mass-to-charge ratio.
A Neutral Loss Ion Mapping Experiment collects scans for masses that have lost neutral fragments. As with Full Scan Ion Mapping, you can get product ion scans on every parent ion. However, a Neutral Loss Ion Map identifies which parent ions lost a neutral fragment of a particular mass. For example, you can specify a neutral loss of 80 u (as in the case of a phosphorylated
22 LTQ XL Getting Started Thermo Electron Corporation
Page 35
1
Introduction
What Types of Experiments Can I Perform with the LTQ XL MS Detector?
peptide in a tryptic digest). A Neutral Loss Ion Mapping experiment can step through each product mass in the mixture. The experiment searches for evidence of the loss of a neutral moiety of mass 80 u.
A Parent Ion Mapping experiment identifies all the ions that produce a particular molecular ion that you specify. For example, if you specify a product ion mass of m/z 50, a Parent Ion Map includes all the parent ions that yielded the specified product ion, m/z 50.
A Data-Dependent Zoom Map is an Ion Mapping experiment that collects ZoomScan data on every scan interval in a mass range that you specify, as well as Data-Dependent MS/MS product spectra on every mass above an intensity threshold.
The results of any of the Ion Mapping experiments can be viewed in the Xcalibur Qual Browser window.

Ion Tree Experiments In an Ion Tree experiment, the LTQ XL MS detector can collect MS

automatically. You can specify a particular parent ion for fragmentation, or you can let the LTQ XL MS detector find the parent ions automatically and fragment them to any level between MS2 and MS10. The LTQ XL MS detector automates the collection of data by deciding what actions need to occur next for the experiment to progress. See Figure 4 for an example of an Ion Tree experiment template.
n
data
Thermo Electron Corporation LTQ XL Getting Started 23
Page 36
1
Introduction
What Types of Experiments Can I Perform with the LTQ XL MS Detector?
Figure 4. Data-Dependent Ion Tree page in Instrument Setup, showing a template for an Ion Tree experiment
In an Ion Tree experiment, you can specify either of two options that prioritize how the LTQ XL MS detector gathers information: depth focus and breadth focus.
• Depth focus characterizes an ion by performing a series of MSn-level fragmentations (for example, MS/MS, MS3, MS4, etc.) before characterizing the next most intense ion in the MSn series.
• Breadth focus characterizes all ions to the same MSn level before
n
advancing to the next MS
level.
For example, if you specify a Maximum Depth of 3 and a Maximum Breadth of 2 in an Ion Tree experiment, the following occurs.
24 LTQ XL Getting Started Thermo Electron Corporation
Page 37
1
Introduction
What Types of Experiments Can I Perform with the LTQ XL MS Detector?
First, with either depth or breadth focus, the LTQ XL MS detector scans for parent ions (MS) over the specified mass range. The most intense ion of the MS spectrum is selected for fragmentation (MS/MS).
• Second, if you chose the depth focus, after the most intense ion of the MS spectrum is fragmented—producing an MS/MS spectrum—the LTQ XL MS detector selects and fragments the most intense ion of the MS/MS spectrum. This results in an MS3 spectrum, the level specified as the maximum depth for this example. The LTQ XL MS detector then backs up one level and fragments the second most intense ion of the MS/MS spectrum, creating more product ions on the level of MS3 from this parent ion. This process is then repeated for the second most intense ion in the MS spectrum.
• If you chose the breadth focus, after the most intense ion of the MS spectrum is fragmented—producing an MS/MS spectrum—the LTQ XL MS detector selects and fragments the second-most intense ion of the same MS spectrum. The fragmentation of parent ions continues to the Max Breadth level that you specified (2, for this example). After the two most intense peaks on the MS level are fragmented, the LTQ XL MS detector scans the first MS/MS spectrum to select and fragment the two most intense ions. This results in product ions on the level of MS3, the level specified as the maximum depth for this example. This process is then repeated for the second most intense ion in the MS spectrum.
The results of a Data-Dependent Ion Tree experiment can be viewed in the Xcalibur Qual Browser window. The results are displayed as a structure tree that originates from a particular parent ion.
Thermo Electron Corporation LTQ XL Getting Started 25
Page 38
Page 39
Chapter 2 Setting Up the Ion Source for
Tuning and Calibrating the MS Detector
This chapter provides information on setting up the hardware for tuning and calibrating your LTQ XL MS detector. You tune and calibrate the MS detector in the ESI mode before you acquire data in either the ESI or APCI mode.
This chapter contains the following topics:
Placing the LC/MS System in Standby
Removing the APCI Probe
Removing the Ion Max Ion Source Housing (optional)
Installing the Ion Sweep Cone (optional)
Installing the Ion Max Ion Source Housing
Installing the ESI Probe
Thermo Electron Corporation LTQ XL Getting Started 27
Page 40
2
Setting Up the Ion Source for Tuning and Calibrating the MS Detector

Placing the LC/MS System in Standby

Placing the LC/MS
System in Standby
The LC/MS system needs to be placed in Standby condition before you can remove the ion source.
Place the LC/MS system in Standby
1. If necessary, stop the flow of solvent to the API source as follows:
a. If the Xcalibur data system is not already open, choose Start >
Programs > Xcalibur > Xcalibur from the Windows® taskbar to open the Xcalibur window.
b. In the Xcalibur Home Page window – Roadmap view, choose
GoTo > Instrument Setup to open the Instrument Setup window.
c. Click the Surveyor® MS Pump button on the view bar in the
Instrument Setup window to display the Surveyor MS Pump view.
d. Choose Surveyor MS Pump > Direct Control to open the Surveyor
MS Pump Direct Control dialog box.
e. In the Direct Control dialog box, click the Pump Off button to stop
the MS pump.
On Off Standby
2. If Tune Plus is not already open, choose Start > Programs > Xcalibur > LTQ XLTune from the taskbar to open Tune Plus.
You can determine the state of the MS detector by observing the state of the On/Standby button on the Control / Scan Mode toolbar. (The three different states of the On/Standby button are shown at the left.)
3. If the MS detector is On, click the On/Standby button to place the MS detector in the Standby mode. When the MS detector is in Standby, the LTQ XL MS detector turns off the ion source sheath gas, auxiliary gas, and high voltage.
The LC/MS system is now in Standby and it is safe to remove the ion source.
If the ESI probe is already installed in the Ion Max™ ion source housing, leave the LC/MS system in Standby and go to Chapter 3: “Tuning and
Calibrating Automatically in the ESI/MS Mode”.
If the ESI probe is not already installed in the Ion Max ion source housing, go to the next section, Removing the APCI Probe.
28 LTQ XL Getting Started Thermo Electron Corporation
Page 41
2
Setting Up the Ion Source for Tuning and Calibrating the MS Detector

Removing the APCI Probe

Removing the APCI
Probe
This topic describes how to remove the APCI probe from the Ion Max ion source housing.
Note The following procedures assume that you are familiar with your instrument and software. If you need additional guidance, refer to LTQ XL online Help, LTQ XL Getting Connected, Ion Max API Source Hardware Manual, or the LTQ XL Hardware Manual.
CAUTION AVOID BURNS. At operating temperatures, the APCI vaporizer can severely burn you! The APCI vaporizer typically operates between 400 and 600 °C. Always allow the heated vaporizer to cool to
room temperature (for approximately 20 min) before you touch or remove this component.
Remove the APCI probe
1. Unplug the vaporizer heater cable from the vaporizer heater cable socket on the APCI probe. See Figure 5.
2. Disconnect the sample transfer line from the APCI probe. (See
Figure 5.)
3. Remove the auxiliary gas line (green-colored fitting) from the APCI probe. (Figure 5)
4. Remove the sheath gas line (blue-colored fitting) from the APCI probe.
Thermo Electron Corporation LTQ XL Getting Started 29
Page 42
2
Setting Up the Ion Source for Tuning and Calibrating the MS Detector
Removing the APCI Probe
Vaporizer Heater Cable
Sheath Gas Fitting
Sample Transfer Line
Auxiliary Gas Fitting
8 kV Cable
Corona Needle High Voltage Receptacle
Figure 5. Ion Max ion source housing with APCI probe installed
CAUTION AVOID BURNS. At operating temperatures, the APCI
vaporizer can severely burn you! The APCI vaporizer typically operates between 400 and 600 °C. Always allow the heated vaporizer to cool to
room temperature (for approximately 20 min) before you touch or remove this component.
5. Remove the APCI probe as follows:
a. Connect the vaporizer heater cable to the ESI interlock socket on
the ion source housing. See Figure 6.
b. Release the probe locking lever to loosen the probe collar. You might
need to unscrew the lever a few turns to permit probe movement.
c. Carefully pull the probe straight back in the port in the housing
until it meets with the slot in the API interlock block. The guide pin on the probe manifold will prevent you from rotating the probe until the pin is aligned with the slot in the API interlock block. Once the probe is all the way back and aligned with the slot, turn the probe 45 degrees counter-clockwise to free the probe from the alignment notch.
d. Pull the probe straight out to remove it from the ion source housing.
e. Store the APCI probe in its original shipping container.
30 LTQ XL Getting Started Thermo Electron Corporation
Page 43
2
Setting Up the Ion Source for Tuning and Calibrating the MS Detector
Removing the APCI Probe
ESI Interlock Socket
API Interlock Block
Probe Collar
Probe Locking Lever
Door Locks
Ion Source Housing Door
Figure 6. Ion Max ion source housing, detail of components
6. Remove the 8 kV cable from the corona needle high voltage receptacle as follows:
a. Unlock the cable by rotating the locking ring counter-clockwise.
b. Unplug the 8 kV cable from the corona needle high voltage
receptacle.
CAUTION AVOID INJURY. The corona discharge needle is very sharp and can puncture your skin. Handle it with care.
7. Remove the corona needle as follows:
a. Unlock the ion source housing door by turning the locks 90 degrees
so that the knobs are horizontal.
b. Open the ion source housing door.
c. Using pliers, grasp the corona needle and pull it straight out of the
corona needle contact. See Figure 7.
Thermo Electron Corporation LTQ XL Getting Started 31
Page 44
2
Setting Up the Ion Source for Tuning and Calibrating the MS Detector
Removing the APCI Probe
d. Close and lock the ion source housing door.
8. Store the corona needle in its original shipping container.
The APCI probe and the corona needle are now properly removed from the Ion Max ion source housing.
If you want to install the optional ion sweep cone, go to the next section,
Removing the Ion Max Ion Source Housing.
If you do not want to install the ion sweep cone, go to “Installing the ESI
Probe” on page 39.
Figure 7. Corona needle, view from rear
Corona Needle Contact
Corona Needle (grasp this with pliers to remove it)
32 LTQ XL Getting Started Thermo Electron Corporation
Page 45
2
Setting Up the Ion Source for Tuning and Calibrating the MS Detector

Removing the Ion Max Ion Source Housing

Removing the Ion Max
Ion Source Housing
The Ion Max ion source housing is removed to access the ion sweep cone.
Note If an ion source probe is still installed in the ion source housing, the external liquid lines should first be disconnected before removing the ion source housing.
Remove the ion source housing
1. Remove the drain tube from the ion source housing drain. See Figure 8.
2. Rotate the ion source housing locking levers 90 degrees to release the ion source housing from the ion source mount assembly.
3. Remove the ion source housing by pulling the housing straight off of the ion source mount assembly
4. Place the ion source housing in a safe location for temporary storage.
The Ion Max ion source housing is now properly removed.
Ion Source Housing Locking Levers
Ion Source Housing Drain
Figure 8. Ion Max ion source housing, detail of components
Thermo Electron Corporation LTQ XL Getting Started 33
Page 46
2
Setting Up the Ion Source for Tuning and Calibrating the MS Detector

Installing the Ion Sweep Cone

Installing the Ion
Sweep Cone
The ion sweep cone is a metallic cone that is installed over the ion transfer tube. The ion sweep cone channels the sweep gas towards the entrance of the capillary. This helps to keep the entrance of the ion transfer tube free of contaminants. The net result is a significant increase in the number of samples that can be analyzed without a loss of signal intensity. In addition, keeping the ion transfer tube entrance cleaner reduces the need for frequent MS detector maintenance.
Install the ion sweep cone
1. Remove the ion sweep cone from its storage container. Inspect and clean it if necessary.
2. Note the location of the sweep gas supply port in the API cone seal. The gas inlet on the ion sweep cone is placed in this port. See Figure 9 and
Figure 10.
Sweep Gas Supply Port
Figure 9. Sweep gas supply port in the API cone seal
34 LTQ XL Getting Started Thermo Electron Corporation
Page 47
2
Setting Up the Ion Source for Tuning and Calibrating the MS Detector
Figure 10. Ion sweep cone, showing the gas inlet
Installing the Ion Sweep Cone
Gas Inlet
CAUTION AVOID BURNS. At operating temperatures, the ion transfer tube can severely burn you! The ion transfer tube typically operates between 200 and 400 °C. Always allow the ion transfer
capillary to cool to room temperature (for approximately 20 min) before you install the ion sweep cone. Always be careful not to touch
the entrance end of the ion transfer tube when it is exposed.
3. After the ion transfer tube has cooled to room temperature, carefully align the gas inlet on the ion sweep cone with the sweep gas supply port in the API cone seal. Firmly press the ion sweep cone into position.
4. If necessary to achieve a proper ion sweep cone installation, you might adjust the set screws around the perimeter of the ion sweep cone.
The ion sweep cone is now properly installed on the MS detector.
Thermo Electron Corporation LTQ XL Getting Started 35
Page 48
2
Setting Up the Ion Source for Tuning and Calibrating the MS Detector

Installing the Ion Max Ion Source Housing

Installing the Ion Max
Ion Source Housing
Reinstall the Ion Max ion source housing
1. Carefully align the two guide pin holes on the rear of the ion source housing with the ion source housing guide pins on the MS detector, and carefully press the ion source housing onto the ion source mount. See
Figure 11 and Figure 12.
Ion Source Housing Locking Levers
Guide Pin Holes
Ion Source Housing Drain
Figure 11. Rear view of the Ion Max ion source housing
36 LTQ XL Getting Started Thermo Electron Corporation
Page 49
2
Setting Up the Ion Source for Tuning and Calibrating the MS Detector
Installing the Ion Max Ion Source Housing
Ion Source Housing Guide Pins
Figure 12. Ion source mount showing ion source housing guide pins
2. Rotate the ion source housing locking levers 90 degrees to lock the ion source housing onto the ion source mount assembly.
CAUTION Prevent solvent waste from backing up into the ion source and MS detector. Always ensure that liquid in the drain tube is able to drain to a waste container.
3. Reinstall the ion source drain tube as follows:
Thermo Electron Corporation LTQ XL Getting Started 37
Page 50
2
Setting Up the Ion Source for Tuning and Calibrating the MS Detector
Installing the Ion Max Ion Source Housing
CAUTION Do not vent the API source drain tube (or any vent tubing connected to the waste container) to the same fume exhaust system to which you have connected the forepumps. The analyzer optics can become contaminated if the API source drain tube and the (blue) forepump exhaust tubing are connected to the same fume exhaust system.
CAUTION Your laboratory must be equipped with at least two fume exhaust systems. Route the (blue) forepump exhaust tubing to a dedicated fume exhaust system. Route the drain tube from the API source to a waste container. Vent the waste container to a dedicated fume exhaust system.
a. Connect the 1-in. ID Tygon® tubing to the ion source housing drain.
b. Attach the free end of the hose to a dedicated drain system. Ideally,
the drain system should be vented to a fume exhaust system.
The Ion Max ion source housing is now properly installed on the MS detector.
38 LTQ XL Getting Started Thermo Electron Corporation
Page 51
2
Setting Up the Ion Source for Tuning and Calibrating the MS Detector

Installing the ESI Probe

Installing the ESI
Probe
Install the ESI probe
1. Remove the ESI probe from its storage container. Inspect and clean it if necessary.
Note If your ESI probe does not already have a sample tube (fused-silica capillary or metal needle) and safety sleeve attached, follow the procedure for installing a sample tube and PEEK safety sleeve that is outlined in Installing a New Fused-Silica Sample
Tube and PEEK Safety Sleeve in the Ion Max API Source Hardware Manual.
2. Ensure that the probe locking lever on the ion source housing is unlocked (opened to its widest position). See Figure 13.
3. Insert the ESI probe into the port in the ion source housing, align the guide pin on the probe body at a minus 45 degree angle from the API interlock block. See Figure 14
ESI Interlock Plug
API Interlock Block
Probe Port
Grounding Bar
Probe Collar
APCI Vaporizer Heater Cable
Aux Gas Fitting
8kV Cable
Sheath Gas Fitting
Probe Locking Lever (open position)
Figure 13. Ion Max ion source housing probe locking lever open
Thermo Electron Corporation LTQ XL Getting Started 39
Page 52
2
Setting Up the Ion Source for Tuning and Calibrating the MS Detector
Installing the ESI Probe
ESI Needle High Voltage
Connector Receptacle
Sheath Gas
Inlet (S)
Guide
Pin
Auxiliary Gas
Inlet (A)
Sample
Inlet Fitting
Sheath Liquid /
Calibrant Inlet (C)
Figure 14. ESI probe, side view
4. Push the probe into the port until the guide pin meets with the probe collar on the ion source housing.
5. Turn the probe 45 degrees clockwise and align the guide pin with the slot in the API interlock block (you might need to pull the probe towards you slightly to properly align the pin with the notch). Once you have turned the probe far enough to align the pin with the alignment notch at the rear of the port, push the probe straight in until the guide pin stops at the bottom of the alignment notch.
6. Lock the probe in place by rotating the probe locking lever towards the front of the housing; closing the probe locking lever towards the rear of the ion source housing might make it difficult to unlock. You might first need to tighten the locking lever threaded shaft by rotating it clockwise a few turns if rotating the lever does not tighten the probe collar enough.
7. Insert the APCI vaporizer heater cable into the API interlock socket.
40 LTQ XL Getting Started Thermo Electron Corporation
Page 53
2
Setting Up the Ion Source for Tuning and Calibrating the MS Detector
Installing the ESI Probe
8. Insert the stainless steel ZDV fitting (grounding union) into the grounding bar on the ion source housing. See Figure 15.
ESI Interlock Socket
Sheath Gas Inlet (S)
Grounding Bar
Grounding Union
PEEK Safety Sleeve
8 kV Cable
Auxiliary Gas Inlet (A)
Sample Inlet
Figure 15. Ion Max ion source housing with ESI probe installed
9. Connect the sheath gas fitting (blue) to the sheath gas inlet (S) on the probe. (See Figure 15.)
10. Connect the auxiliary gas fitting (green) to the auxiliary gas inlet (A) on the probe. (See Figure 15.)
11. Connect the 8 kV cable to the ESI needle high voltage receptacle on the ESI probe. Tighten the locking ring on the 8 kV connector.
12. Connect the sample transfer tubing to the grounding union.
The ESI probe is now properly installed in the Ion Max ion source housing.
Leave the LC/MS system in Standby and go to Chapter 3: “Tuning and
Calibrating Automatically in the ESI/MS Mode”.
Thermo Electron Corporation LTQ XL Getting Started 41
Page 54
Page 55
Chapter 3 Tuning and Calibrating
Automatically in the ESI/MS Mode
This chapter provides information on how to tune and calibrate the LTQ XL MS detector in the ESI/MS mode. For most applications, you tune and calibrate in the ESI mode through automatic procedures. The procedures use a calibration solution that is introduced into the MS detector in low flow mode. The procedures properly tune and calibrate the MS detector for ESI operation (refer to Table 2 on page 2 for ESI operating parameter guidelines). You need to calibrate the MS detector every one to three months of operation for optimum performance over the entire mass range of the detector.
To tune and calibrate your MS detector automatically in the ESI/MS mode, you do the following:
• Infuse a low concentration calibration solution containing caffeine, MRFA, and Ultramark 1621 into the ESI source by using the syringe pump. (Refer to the section, Setting Up the Syringe Pump for Tuning
and Calibration.)
• Test the efficiency and stability of the spray of calibration solution into the MS detector. You can observe the following singly-charged, positive ions for caffeine, MRFA, and Ultramark 1621 in the Tune Plus window: m/z 195, 524, 1222, 1522, and 1822.
• Tune the MS detector from the Tune Plus window to optimize automatically the lenses.
• Calibrate the MS detector to adjust automatically the voltages of the linear trap.
This chapter contains the following sections:
Setting Up the Syringe Pump for Tuning and Calibration
Setting Up the MS Detector in the Xcalibur Data System for Tuning
and Calibration
Testing the Operation of the MS Detector in the ESI/MS Mode
Thermo Electron Corporation LTQ XL Getting Started 43
Page 56
3
Tuning and Calibrating Automatically in the ESI/MS Mode
Tuning the MS Detector Automatically in the ESI/MS Mode
Saving Your ESI/MS Tune Method
Calibrating the MS Detector Automatically
Cleaning the MS Detector after Tuning and Calibrating
44 LTQ XL Getting Started Thermo Electron Corporation
Page 57
3
Tuning and Calibrating Automatically in the ESI/MS Mode

Setting Up the Syringe Pump for Tuning and Calibration

Setting Up the Syringe
Pump for Tuning and
Calibration
You introduce tuning and calibration solution into the API source with a syringe infusion pump. A syringe pump allows you to infuse a sample solution into the API source for extended periods of time.
The syringe pump and syringe are located on the front panel of your LTQ XL MS detector. To infuse solution for tuning and calibration, you install on the pump a 500-mL Unimetrics® syringe containing the calibration solution.
Note To minimize the possibility of cross-contamination, use a different syringe and section of fused silica tubing for the calibration solution than you do for your sample solution.
Set up the syringe pump for infusion
1. Connect a 4 cm (1.5 in.) segment of Teflon® tube with a (brown) fingertight fitting and a (brown) ferrule to the (black) LC union. See
Figure 16.
LC Union
(
P/N
00101-18202)
Fingertight fitting
P/N
00101-18081)
(
Ferrule
(
P/N
00101-18196)
Teflon Tube
(
P/N
00301-22803)
Figure 16. Plumbing connections for the syringe
2. Load a clean, 500-μL Unimetrics syringe with 450 μL of the calibration solution. (Refer to Appendix A: “Sample Formulations” on page 139 for a procedure for making the calibration solution.)
3. Insert the syringe needle into the segment of Teflon tube.
4. Place the syringe into the syringe holder of the syringe pump.
5. While squeezing the blue release button on the syringe pump handle, push the handle forward until it just contacts the syringe plunger.
6. Connect a fused-silica infusion line from the LC union to the (stainless steel) grounding union as follows. See Figure 17.
Thermo Electron Corporation LTQ XL Getting Started 45
Page 58
3
Tuning and Calibrating Automatically in the ESI/MS Mode
Setting Up the Syringe Pump for Tuning and Calibration
a. Connect the infusion line with a (brown) fingertight fitting and a
b. Connect the other end of the infusion line with a (red) fingertight
The syringe pump is now properly set up for infusing solution into the MS detector.
Go to the next section, Setting Up the MS Detector in the Xcalibur Data
System for Tuning and Calibration.
(brown) ferrule to the free end of the LC union.
fitting and a (brown) ferrule to the grounding union.
Figure 17. ESI/MS plumbing connections for the fused-silica infusion line
46 LTQ XL Getting Started Thermo Electron Corporation
Page 59
3
Tuning and Calibrating Automatically in the ESI/MS Mode

Setting Up the MS Detector in the Xcalibur Data System for Tuning and Calibration

Setting Up the MS
Detector in the
Xcalibur Data System
for Tuning and
Calibration
You first tune manually with calibration solution to establish a stable spray of solution and to ensure that enough ions are detected to calibrate the MS detector. You then calibrate the MS detector automatically to optimize the parameters that affect ion detection. With the optimized MS detector, the Xcalibur data system can isolate and fragment ions and determine their mass-to-charge ratios. Perform a calibration periodically, every one to three months, for optimum performance of the MS detector.
Note The following procedures assume that you are familiar with your LTQ XL instrument and the Tune Plus window. If you need additional guidance, see LTQ XL online Help, LTQ XL Getting Connected, and/or the LTQ XL Hardware Manual.
CAUTION Before you begin normal operation each day, ensure that you have sufficient nitrogen for your API source. If you run out of nitrogen, the LTQ XL MS detector automatically turns Off to prevent the possibility of atmospheric oxygen from entering the ion source. The presence of oxygen in the ion source when the MS detector is On could be unsafe. (In addition, if the LTQ XL MS detector automatically turns Off during an analytical run, you could lose data.)
On Off Standby
Set up the MS detector in the Xcalibur data system for tuning and calibration in the ESI/MS mode
1. If you have not already done so, open the Tune Plus window from the Start button on your Windows XP task bar, as follows:
a. Choose Start > Programs > Xcalibur > Xcalibur to display the
Xcalibur Home Page – Roadmap view.
b. Click the Instrument Setup button to display the Instrument Setup
window.
c. Click the Finnigan LTQ XL button to display the New Method
page.
d. Click the Tune Plus button to display the Tune Plus window. See
Figure 18.
2. In the Tune Plus window, on the Control/Scan Mode toolbar, click the On/Off/Standby button to take the MS detector out of the Standby (or
Thermo Electron Corporation LTQ XL Getting Started 47
Page 60
3
Tuning and Calibrating Automatically in the ESI/MS Mode
Setting Up the MS Detector in the Xcalibur Data System for Tuning and Calibration
Off) mode and turn it On. When you turn the MS detector to On, you initiate the following events:
• The MS detector begins scanning.
• Nitrogen flows into the ESI probe.
• The LTQ XL MS detector applies high voltage to the ESI probe.
• The Xcalibur data system shows a real-time display in the Spectrum view.
Figure 18. Tune Plus window, showing the MS detector in the Standby mode
48 LTQ XL Getting Started Thermo Electron Corporation
Page 61
3
Tuning and Calibrating Automatically in the ESI/MS Mode
Setting Up the MS Detector in the Xcalibur Data System for Tuning and Calibration
Note The Xcalibur data system contains customized tune files for different applications in the folder C:\Xcalibur\methods, including one for low flow LC/ESI/MS operation.
3. Open the Tune Method file that stores the factory default tune settings for low-flow ESI operation, as follows:
a. Choose File>Open to display the Open dialog box.
b. Browse for the folder C:\Xcalibur\methods. Select the file
AutoTune.LTQTune.
c. Click Open to open the file. Tune Plus downloads the Tune Method
parameters to the MS detector.
4. Examine the pre-tune ESI source settings as follows:
a. From the Instrument Setup toolbar, click the API Source button to
open the ESI Source dialog box. Verify that the settings in your dialog box are the same as those shown in Figure 19.
b. Click OK to return to the Tune Plus window.
Figure 19. ESI Source dialog box, showing the settings to start a typical low
flow experiment
5. Set the scan parameters for tuning and calibration, as follows:
Thermo Electron Corporation LTQ XL Getting Started 49
Page 62
3
Tuning and Calibrating Automatically in the ESI/MS Mode
Setting Up the MS Detector in the Xcalibur Data System for Tuning and Calibration
a. On the Control/Scan Mode toolbar, click the Define Scan button
to open the Define Scan dialog box. See Figure 20. (If your dialog box appears different from the one shown in the figure, it is probably because the advanced settings are not displayed. You can turn on the advanced settings as follows: In Tune Plus, choose ScanMode, and then click Advanced Scan Features to select the option.)
b. In the Scan Description group box, in the Mass Range list box,
select Normal to allow for a selection of mass ranges between m/z 150 to 2000.
c. In the Scan Rate list box, select Normal to specify a normal scan
rate.
d. In the Scan Type list box, select Full specify a full scan.
e. In the Scan Time group box, in the Microscans spin box, enter 1 to
set the total number of microscans to 1.
f. In the Max. Inject Time spin box, enter 200.000 to specify a 200 ms
maximum injection time.
g. In the Source Fragmentation group box, confirm that the On check
box is not selected ( ) to specify that the ion source fragmentation option is turned off.
h. In the Scan Ranges group box, in the Input list box, select From/To
to make available the First Mass and Last Mass text boxes in the Scan Ranges table.
50 LTQ XL Getting Started Thermo Electron Corporation
Page 63
3
Tuning and Calibrating Automatically in the ESI/MS Mode
Setting Up the MS Detector in the Xcalibur Data System for Tuning and Calibration
Figure 20. Define Scan dialog box, showing the default settings for ESI/MS operation
i. In the Scan Ranges group box, in the Scan Ranges table, in the First
Mass text box, enter 150 to set the first mass for the scan range to m/z 150.
j. In the Last Mass text box, enter 2000 to set the last mass for the scan
range to m/z 2000.
k. Ensure that the settings in your Define Scan dialog box are the same
as those shown in Figure 20.
l. Click OK to apply the MS detector scan parameters and to close the
Define Scan dialog box.
6. On the Control/Scan Mode toolbar, click the Centroid/Profile button to toggle the data type to profile. (The picture on the button should be the same as that shown here.)
7. Click the Positive/Negative button to toggle the ion polarity mode to positive. (The picture on the button should be the same as that shown here).
The MS detector is now properly set up in the Xcalibur data system for tuning and calibration in the ESI/MS mode.
Thermo Electron Corporation LTQ XL Getting Started 51
Page 64
3
Tuning and Calibrating Automatically in the ESI/MS Mode

Testing the Operation of the MS Detector in the ESI/MS Mode

Testing the Operation
of the MS Detector in
the ESI/MS Mode
You are now ready to test whether your MS detector is operating properly. To test for proper operation, you infuse the calibration solution into the ESI source, and then you monitor the real-time display of the mass spectrum of calibration solution. You want to ensure that a stable spray of solution enters the MS detector.
Test the operation of the MS detector in the ESI/MS mode
1. Click the Syringe Pump button to display the Syringe Pump dialog box. See Figure 21.
Figure 21. Syringe Pump dialog box
2. Turn on the syringe pump and set an infusion flow rate of 5 μL/min, as follows:
a. In the Flow Control group box, click the On option button to make
active the Flow Rate spin box.
b. Type 5 in the Flow Rate spin box to specify a rate of 5 μL/min.
Note This procedure assumes that you are using the 500-μL Unimetrics syringe that is provided with your LTQ XL system. If you are using another type of syringe, select the option button corresponding to your syringe.
c. If you are using a standard Unimetrics (or Hamilton) syringe, set up
the syringe parameters as follows:
52 LTQ XL Getting Started Thermo Electron Corporation
Page 65
3
Tuning and Calibrating Automatically in the ESI/MS Mode
Testing the Operation of the MS Detector in the ESI/MS Mode
i. In the Type group box, click the Unimetrics (or Hamilton)
option button to specify the proper syringe type.
ii. Click the Vo lu m e list box arrow to display the list of available
volumes, and then select 500 (or your syringe size) from the list to set the proper syringe volume. Note that, if you are using a Unimetrics syringe, the LTQ XL MS detector automatically sets the syringe ID to its proper value of 3.260 mm.
d. If you are not using a Unimetrics (or Hamilton) syringe, set up the
syringe parameters as follows:
i. In the Type group box, click the Other option button to make
active the syringe ID spin box.
ii. Type the inner diameter of your syringe in the Syringe ID spin
box.
e. Click OK to apply the syringe parameters, start the syringe pump,
and close the Syringe Pump dialog box.
3. On the File/Display toolbar, click the Display Spectrum View button to ensure that the Spectrum view is displayed.
4. Monitor the data for the calibration solution, as follows:
a. In the Spectrum view of the Tune Plus window, observe the mass
spectra of the singly-charged ions of calibration solution. The ions are as follows. See Figure 22.
• Caffeine: m/z 195
• MRFA: m/z 524
• Ultramark 1621: m/z 1022, 1122, 1222, 1322, 1422, 1522, 1622,
1722, 1822
Note Based on the LC flow rate of your experiment, you can specify the value of each of the following tuning parameters on the LTQ XL MS detector: sheath, Auxiliary, and Sweep gas pressures, ESI needle (or “spray”) voltage, ion transfer capillary temperature, and probe position. Automatic tuning sets the values of the other parameters.
b. At the top of the Spectrum view, notice the values for the ionization
time (IT) and normalization level (NL). See Figure 22.
c. Click the API Source button to open the ESI Source dialog box.
(See the Spray Current readback shown in Figure 19.)
Thermo Electron Corporation LTQ XL Getting Started 53
Page 66
3
Tuning and Calibrating Automatically in the ESI/MS Mode
Testing the Operation of the MS Detector in the ESI/MS Mode
d. Observe the value for the Spray Current readback and the values for
If you answered “yes” to the questions in step 4.d, then your MS detector is operating properly.
If you answered “no” to either of these questions, try the following troubleshooting measures:
• Ensure that the fused-silica sample tube does not extend beyond the tip of the ESI needle.
NL and IT in the Spectrum view. As calibration solution infuses, and the readback values fluctuate, ask yourself the following questions about the ion current signal:
• Is the signal present?
• Is the signal stable, varying by less than about 15% from scan to scan?
• Ensure that the entrance to the ion transfer capillary is clean, and is not covered with a piece of septum.
• Ensure that the solution entering the probe is free of air bubbles and that the tubing and connectors are free of leaks.
Congratulations! You have demonstrated that your MS detector is operating properly in the ESI mode. You are now ready to tune and calibrate the MS detector. Leave your LTQ XL MS detector as it is, and go to the next section, Tuning the MS Detector Automatically in the ESI/MS Mode.
54 LTQ XL Getting Started Thermo Electron Corporation
Page 67
3
Tuning and Calibrating Automatically in the ESI/MS Mode
Testing the Operation of the MS Detector in the ESI/MS Mode
IT=ionization time
NL=normalization level
Figure 22. Spectrum view of the Tune Plus window, showing ionization time (IT) and normalization level (NL) of the
calibration solution
Thermo Electron Corporation LTQ XL Getting Started 55
Page 68
3
Tuning and Calibrating Automatically in the ESI/MS Mode

Tuning the MS Detector Automatically in the ESI/MS Mode

Tuning the MS
Detector
Automatically in the
ESI/MS Mode
You tune the MS detector automatically in the ESI/MS mode to optimize important parameters, including heated capillary voltage and tube lens voltage.
Tune the MS detector automatically
1. On the Control/Scan Mode toolbar, click the Tu n e button to display the Tune dialog box.
2. If necessary, click the Automatic tab to display the Automatic tuning page. See Figure 23.
3. In the What to Optimize On group box, select the Mass option button to make active the Mass spin box.
Figure 23. Tune dialog box, showing the Automatic tuning page
56 LTQ XL Getting Started Thermo Electron Corporation
Page 69
3
Tuning and Calibrating Automatically in the ESI/MS Mode
Tuning the MS Detector Automatically in the ESI/MS Mode
4. In the Mass spin box, enter 195.1 to specify that the LTQ XL MS
detector optimize your Tune Method on the peak at m/z 195.1.
Note In this example, you use the mass peak at m/z 195.1 to optimize the Tune Method. However, you can optimize the tune on any mass peak of the calibration solution.
5. Start the automatic tuning procedure, as follows:
a. Click Start. A message box displays the following message:
Please ensure that the 500 microliter syringe is full.
Ensure that the syringe contains at least 450 μL calibration solution.
b. Click OK to close the message box, and return to the Tune dialog
box.
6. On the File/Display toolbar, click the Graph View button to display the Graph view. See Figure 24.
7. Observe the Tune Plus window and the Tune dialog box. While automatic tuning is in progress, the LTQ XL MS detector displays various tests in the Spectrum and Graph views in Tune Plus and displays various messages in the Status group box in the Tune dialog box. Your Tune Plus window should now look similar to the one shown in
Figure 24.
8. Click the ESI Source dialog box to examine the ESI source parameters after tuning. Compare the settings shown in Figure 25 with the pre-tune settings shown in Figure 19 on page 3-49.
9. Click the Ion Optics toolbar button to display the Ion Optics dialog box. The parameters in the Ion Optics dialog box are optimized automatically by the LTQ XL MS detector. See Figure 26.
You have now successfully tuned the MS detector in ESI/MS mode using the calibration solution. Go to the next section, Saving Your ESI/MS Tune
Method.
Thermo Electron Corporation LTQ XL Getting Started 57
Page 70
3
Tuning and Calibrating Automatically in the ESI/MS Mode
Tuning the MS Detector Automatically in the ESI/MS Mode
Figure 24. Tune Plus window, showing the results of a typical automatic tune procedure
58 LTQ XL Getting Started Thermo Electron Corporation
Page 71
3
Tuning and Calibrating Automatically in the ESI/MS Mode
Tuning the MS Detector Automatically in the ESI/MS Mode
Figure 25. ESI Source dialog box, showing typical parameters after automatic
tuning
Figure 26. Ion Optics dialog box, showing examples of voltages of lenses and
Intermultipoles, which are optimized by the LTQ XL automatic tuning procedure
Thermo Electron Corporation LTQ XL Getting Started 59
Page 72
3
Tuning and Calibrating Automatically in the ESI/MS Mode

Saving Your ESI/MS Tune Method

Saving Your ESI/MS
Tune Method
You can save the parameters you just set in a Tune Method specific to your particular analyte and solvent flow rate. (In this case, you save settings obtained using calibration solution.) You can recall the Tune Method and use it as a starting point for optimizing the MS detector on a different analyte of interest or at a different flow rate.
Note You must save the Tune Method while the MS detector is On.
Save your ESI/MS Tune Method (for low-flow operation) when automatic tuning is complete
1. Choose File > Save As to display the Save As dialog box. See Figure 27.
Figure 27. Save As dialog box, showing files in the folder C:\Xcalibur\methods
2. Select the C:\Xcalibur\methods folder.
60 LTQ XL Getting Started Thermo Electron Corporation
Page 73
3
Tuning and Calibrating Automatically in the ESI/MS Mode
Saving Your ESI/MS Tune Method
3. Click the File Name text box, and type ESImyTune to name the Tune Method ESImyTune.LTQTune.
4. Click Save to save the Tune Method, and return to the Tune Plus window. Note that the Tune Method is named ESImyTune.LTQTune.
Once you have tuned the MS detector, you are now ready to calibrate.
Thermo Electron Corporation LTQ XL Getting Started 61
Page 74
3
Tuning and Calibrating Automatically in the ESI/MS Mode

Calibrating the MS Detector Automatically

Calibrating the MS
Detector
Automatically
Calibrate the MS detector automatically from the Tune Plus window
1. Choose Control > Calibrate to display the Calibrate dialog box.
2. If necessary, click the Automatic tab to display the Automatic calibration page. See Figure 28.
Figure 28. Calibrate dialog box, showing the Automatic calibration page
3. Start the automatic calibration procedure, as follows:
a. Click Start. A message box displays the following message:
Please ensure that the 500 microliter syringe is full.
62 LTQ XL Getting Started Thermo Electron Corporation
Page 75
3
Tuning and Calibrating Automatically in the ESI/MS Mode
Calibrating the MS Detector Automatically
Ensure the syringe contains at least 450 μL calibration solution.
b. Click OK to close the message box, and return to the Calibrate
dialog box.
4. Observe the Tune Plus window and the Calibrate dialog box. While the automatic calibration is in progress, the LTQ XL MS detector displays a variety of test results in the Spectrum and Graph views and displays a variety of messages in the Status box of the Calibrate dialog box.
The automatic calibration procedure typically takes about 40 min.
When the LTQ XL MS detector completes the calibration procedure it restores the full scan ESI mass spectrum in the Spectrum view. The Instrument Messages dialog box is displayed, which indicates whether or not the calibration procedure for an item is successful.
• If a calibration item is successful, the LTQ XL MS detector saves the new calibration parameter automatically to the hard disk.
• If a calibration item fails, you can try calibrating on that item again after you ensure the following: the spray is stable, the solution flow rate is sufficient, and all the ions in the calibration solution are present with adequate signal-to-noise ratios. If the sensitivity of the ions is low, increase the solution flow rate somewhat, and then use the semi-automatic calibration procedure to calibrate the specific parameter that failed. See Figure 29. Consider deselecting the ZoomScan Mode option if repeated failures occur.
When all calibration items are successful, your MS detector is properly tuned and calibrated for low-flow experiments. A successful calibration exhibits adequate intensities of the following calibrant ions: m/z 195, 524, 1222, 1522, and 1822. In many cases, fine tuning on your particular analyte is not necessary if the intensity of these ions is sufficient. You are ready to analyze samples if you do not need to maximize the intensity of the ion signals for your analyte.
The procedures for changing the solution flow rate and optimization of the MS detector parameters for reserpine, or your particular analyte, are explained in Chapter 4: “Tuning with Your Analyte in LC/ESI/MS Mode”. Before you tune with your analyte, go to the next section, Cleaning the MS
Detector after Tuning and Calibrating.
Thermo Electron Corporation LTQ XL Getting Started 63
Page 76
3
Tuning and Calibrating Automatically in the ESI/MS Mode
Calibrating the MS Detector Automatically
Figure 29. Tune Plus window with Calibrate dialog box, showing the results of a successful semi-automatic calibration
procedure
64 LTQ XL Getting Started Thermo Electron Corporation
Page 77
3
Tuning and Calibrating Automatically in the ESI/MS Mode

Cleaning the MS Detector after Tuning and Calibrating

Cleaning the MS
Detector after Tuning
and Calibrating
On Standby
This topic describes how to clean your MS detector after using the calibration solution, in preparation for acquiring data on your analyte of interest.
Clean the MS detector after calibrating
1. Click On/Standby to put the MS detector in Standby mode. When the MS detector is in Standby, the LTQ XL MS detector turns off the sheath gas, Auxiliary gas, Sweep gas, ESI high voltage, and syringe pump. The MS detector stops scanning, and the LTQ XL MS detector freezes the displays for the Spectrum and Graph views.
CAUTION Always place the MS detector in Standby (or Off) before you open the API source to atmospheric oxygen. The presence of oxygen in the ion source when the MS detector is On could be unsafe. (The LTQ XL MS detector automatically turns the MS detector Off when you open the API source, however, it is best to take this added precaution.
2. Remove the syringe from the syringe pump holder, as follows:
a. Squeeze the blue buttons, and pull back on the syringe pump handle
to free the syringe.
b. Remove the syringe from the holder.
c. Disconnect the tip of the syringe needle from the Teflon tubing.
3. Clean the syringe thoroughly, as follows:
a. Clean the syringe with a solution of 5% formic acid in water.
b. Rinse the syringe with a solution of 50:50 methanol:water.
c. Use acetone to rinse the syringe. (Repeat this step several times.)
4. To gain access to the ion transfer capillary, the Ion Max ion source housing and the ion sweep cone need to be removed. Refer to the topic
“Removing the Ion Max Ion Source Housing” on page 33 for
instructions for removing the Ion Max ion source housing.
5. Remove the ion sweep cone as follows:
a. Put on a pair of talc-free gloves.
Thermo Electron Corporation LTQ XL Getting Started 65
Page 78
3
Tuning and Calibrating Automatically in the ESI/MS Mode
Cleaning the MS Detector after Tuning and Calibrating
CAUTION AVOID BURNS. At operating temperatures, the ion transfer tube can severely burn you! The ion transfer tube typically operates between 200 and 400 °C. Always allow the ion sweep cone to
cool to room temperature (for approximately 20 min) before you touch or remove this component. Always be careful not to touch the entrance
end of the ion transfer tube when it is exposed.
b. Grasp the outer ridges of the ion sweep cone and pull the cone
6. Remove the ion transfer capillary by using the custom tool provided.
7. Clean the ion sweep cone and the ion transfer capillary as follows:
a. Place the ion sweep cone and the ion capillary tube in a beaker of
b. Sonicate the components for 15 min.
straight off of the API cone seal. Note, you might need to loosen the set screws on the ion sweep cone in order to remove it.
50:50 methanol/water.
8. Reinstall the ion transfer capillary.
9. Reinstall the ion sweep cone as described in “Installing the Ion Sweep
Cone” on page 34.
10. Place a small Teflon-coated septum over the entrance end of the ion transfer capillary to seal the vacuum chamber of the MS detector.
11. Flush the sample transfer line, sample tube, and ESI probe thoroughly with a solution of 5% formic acid in water (or with another appropriate solvent), as follows:
Note The solvent that you use to flush the sample transfer line, sample tube, and ESI probe assembly depends on the solvent system you use to dissolve your samples. For example, if you are using a buffered solution of a high concentration, an acidic solution is appropriate.
a. Fill a clean, 250 μL Unimetrics syringe with a solution an
appropriate solvent.
b. While holding the plunger of the syringe in place, carefully insert
the needle of the syringe into the free end of the Teflon tube.
c. Flush the sample transfer line, sample tube, and ESI probe with the
solution by slowly depressing the syringe plunger. Visually confirm that the solution is exiting the tip of the ESI probe on the inside of
66 LTQ XL Getting Started Thermo Electron Corporation
Page 79
3
Tuning and Calibrating Automatically in the ESI/MS Mode
Cleaning the MS Detector after Tuning and Calibrating
the probe assembly. Use a lint-free tissue to gently remove the excess solution as it exits the probe.
d. Remove the needle of the syringe from the Teflon tube.
12. Repeat step 11 with a solution of 50:50 methanol:water.
13. Repeat step 11 with acetone.
14. Clean the spray shield as follows:
a. Fill a spray bottle with solvent solution.
b. Temporarily place a large Kimwipe (or other lint-free tissue) beneath
of the spray shield. (The Kimwipe is required to absorb the solution used to flush the ion transfer capillary and spray shield.)
c. Use the spray bottle to flush contaminants from the exterior surface
of the spray shield.
d. Remove the Kimwipe you used to absorb the solution. Swab the
surface of the spray shield with a dry Kimwipe.
e. Repeat step 14.a through step 14.d with acetone to remove the
(high molecular weight) Ultramark 1621.
15. Being careful not to touch the ion transfer capillary with your hand, remove the septum from the entrance end of the ion transfer capillary.
16. Reinstall the Ion Max ion source housing as described in “Installing the
Ion Max Ion Source Housing” on page 36.
The MS detector is now clean and ready for acquiring data on your analyte of interest.
If you plan to run analytical samples in high-flow ESI mode (using flow rates between 50 and 1000 μL/min), the procedures in Chapter 4: “Tuning
with Your Analyte in LC/ESI/MS Mode” explain how to optimize the tune
for this situation.
Thermo Electron Corporation LTQ XL Getting Started 67
Page 80
Page 81
Chapter 4 Tuning with Your Analyte in
LC/ESI/MS Mode
This chapter provides information on tuning the MS detector in the LC/ESI/MS mode using your analyte. You optimize the sensitivity of the MS detector to your analyte through an automatic procedure.
The customized Tune Methods contained in your LTQ XL data system are optimized for a wide range of applications, and they can be used without further tuning of your MS detector. However, for certain applications you might need to tune and optimize several MS detector parameters.
For instance, the most important parameters that interact with the ESI interface and signal quality are as follows:
• Electrospray voltage
• Heated capillary temperature (voltage)
• Tube lens voltage
• Capillary voltage
• Sheath gas flow rate
• Auxiliary gas flow rate
• Sweep gas flow rate
The settings for these parameters depend on the solvent flow rate and target analyte composition. In general, you should fine tune your MS detector whenever you change the solvent flow rate conditions of your particular application. In this procedure, you use the ESI low-flow Tune Method ESImyTune.LTQTune as a starting point, then further optimize the MS detector parameters using an automatic procedure. The automatic procedure adjusts the tube lens voltage, capillary voltage, and voltages applied to the ion optics until the ion transmission of your analyte is maximized.
Thermo Electron Corporation LTQ XL Getting Started 69
Page 82
4
Tuning with Your Analyte in LC/ESI/MS Mode
The capillary is heated to maximize the ion transmission to the MS detector. For ESI only, you set the ion transfer capillary temperature proportional to the flow rate of your solution. Refer to Table 1-2 for guidelines for setting operating parameters for LC/ESI/MS. For this procedure, the ion transfer capillary temperature is set to 350 °C, and the sheath gas is set to 30.
Note
1. If your experiment is performed at a flow rate below 10 μL/min, and the results you want can be obtained without optimizing the MS detector on your particular analyte, go to Chapter 5: “Acquiring ESI
Sample Data Using the Tune Plus Window” to acquire sample data.
2. Before you optimize the tune for your analyte of interest, ensure that the LTQ XL MS detector has been calibrated within the previous three months. If the system needs to be calibrated, refer to the procedures in Chapter 3: “Tuning and Calibrating Automatically in
the ESI/MS Mode”.
To tune the MS detector in the ESI/MS (high-flow) mode using your analyte, perform the following tasks:
• Set up the MS detector for your specific analyte from the Tune Plus window.
• Infuse your analyte into the MS detector using a syringe pump connected to the LC with a Tee union.
• Optimize the MS detector parameters for your analyte while the solution flows into the MS detector.
This chapter contains the following topics:
Setting Up to Introduce Sample by Syringe Pump into Solvent Flow
from an LC
Setting Up to Tune the MS Detector with Your Analyte
Optimizing the MS Detector Tune Automatically with Your Analyte
Saving the ESI/MS Tune Method
70 LTQ XL Getting Started Thermo Electron Corporation
Page 83
4
Tuning with Your Analyte in LC/ESI/MS Mode

Setting Up to Introduce Sample by Syringe Pump into Solvent Flow from an LC

Setting Up to
Introduce Sample by
Syringe Pump into
Solvent Flow from an
LC
This topic describes setting up the MS detector to introduce your analyte by syringe pump into solvent flow from an LC.
Plumbing connections for ESI/MS sample introduction into LC solvent flow from the syringe pump
1. Connect a 4 cm (1.5 in.) segment of Teflon tubing with a (brown) fingertight fitting and a (brown) ferrule to the (black) LC union. (See
Figure 17 on page 46.)
2. Fill a clean, 500-μL Unimetrics syringe with the 125 fg/μL solution of reserpine or your analyte of interest. (See Appendix A: “Sample
Formulations” for a procedure for making the reserpine tuning
solution.)
3. Insert the needle of the syringe into the segment of Teflon tube. Check that the needle tip of the syringe fits readily into the opening in the free end of the Teflon tubing. If necessary, you can enlarge the opening in the end of the tubing slightly.
4. Place the syringe into the syringe holder of the syringe pump.
5. While squeezing the blue release buttons on the syringe pump handle, push the handle forward until it just contacts the syringe plunger.
6. Connect the fused-silica infusion line from the (black) LC union to the (black) LC Tee union, as follows. See Figure 30.
a. Connect the infusion line with a (brown) fingertight fitting and a
(brown) ferrule to the free end of the LC union.
b. Connect the other end of the infusion line with a (red) fingertight
fitting and a (brown) ferrule to the side arm of the LC Tee union.
7. Connect an appropriate length of (red) PEEK tubing from the (stainless steel) grounding union to the (black) LC Tee union, as follows:
a. Use a PEEK tubing cutter to cut a 4 cm (1.5 in.) length of the PEEK
tubing.
b. Connect the PEEK tubing with a (brown) fingertight fitting and a
(brown) ferrule to the grounding union.
c. Connect the other end of the PEEK tubing with a (brown)
fingertight fitting and a (brown) ferrule to the LC Tee union.
Thermo Electron Corporation LTQ XL Getting Started 71
Page 84
4
Tuning with Your Analyte in LC/ESI/MS Mode
Setting Up to Introduce Sample by Syringe Pump into Solvent Flow from an LC
From Divert /
Inject Valve
Stainless Steel
Ferrule
Ferrules
(
P/N
00101-18120)
Fingertight
Fitting
P/N
00101-18195)
(
PEEK Tubing
(
P/N
00301-22912)
Stainless Steel Nut
LC Tee Union
(
P/N
00101-18204)
From LC Union
Figure 30. ESI/MS plumbing connections for the LC Tee union
PEEK Tubing
(
P/N
00301-22912)
Ferrules
P/N
00101-18196)
(
Infusion Line
Fused-Silica Capillary
(
P/N
00106-10504)
Grounding Union
(
P/N
00101-18182)
Fingertight
Fittings
P/N
00101-18081)
(
8. If you have not already done so, connect the PEEK safety sleeve and fused-silica sample tube from the grounding union to the ESI probe sample inlet as described in the section, Installing a New Fused-Silica
Sample Tube and PEEK Safety Sleeve in the Ion Max API Source Hardware Manual.
If you have installed the stainless steel needle in the ESI probe, connect the PEEK safety sleeve and fused-silica capillary tube from the grounding union to the ESI probe sample inlet as described in the section, Installing a New Stainless Steel Needle in the ESI Probe and
Installing a New Fused-Silica Sample Tube and PEEK Safety Sleeve in the Ion Max API Source Hardware Manual.
9. Connect an appropriate length of PEEK tubing (transfer line from the divert/inject valve) from the divert/inject valve to the free end of the (black) LC Tee union, as follows.
a. Connect a length of PEEK tubing with a (stainless steel) nut and a
(stainless steel) ferrule to port 3 of the divert/inject valve. See Figure 31.
72 LTQ XL Getting Started Thermo Electron Corporation
Page 85
4
Tuning with Your Analyte in LC/ESI/MS Mode
Setting Up to Introduce Sample by Syringe Pump into Solvent Flow from an LC
b. Connect the other end of the PEEK tubing with a (brown)
fingertight fitting and a (brown) ferrule to the free end of the LC Tee union. (See Figure 30.)
To LC Tee
Union
Plug
(optional)
3
4
5
6
Detector Position
2
1
To Wast e
From
LC
To LC Tee
Union
Plug
(optional)
3
4
5
6
To Wast e
Waste Position
From
LC
2
1
Figure 31. Divert/Inject valve, showing the correct setup for tuning by syringe
infusion and showing the flow of liquid through the valve in the Detector and Waste positions
10. Connect an appropriate length of PEEK tubing (transfer line from the LC) from the divert/inject valve to the LC, as follows:
a. Connect a length of PEEK tubing with a (stainless steel) nut and a
(stainless steel) ferrule to port 2 of the divert/inject valve.
b. Connect the other end of the PEEK tubing with a proper fitting and
a ferrule to the outlet of the LC.
11. Connect an appropriate length of PEEK tubing (waste line) from the divert/inject valve to a waste container, as follows:
a. Connect a length of PEEK tubing with a (stainless steel) nut and a
(stainless steel) ferrule to port 1 of the divert/inject valve.
b. Insert the other end of the PEEK tubing in a suitable waste
container.
The MS detector is now properly set up to introduce your analyte by syringe pump into solvent flow from an LC.
Thermo Electron Corporation LTQ XL Getting Started 73
Page 86
4
Tuning with Your Analyte in LC/ESI/MS Mode

Setting Up to Tune the MS Detector with Your Analyte

Setting Up to Tune the
MS Detector with
Your Analyte
Set up the MS detector to tune automatically on your analyte in ESI/MS mode
In this procedure, you can use the reserpine solution described in Appendix
A: “Sample Formulations”, or you can use a solution of an analyte of interest
to you.
CAUTION Do not infuse calibration solution at flow rates above 10 μL/min. Ultramark 1621 can contaminate your system at high concentrations.
Note The following procedures assume that you are familiar with your LTQ XL instrument and the Tune Plus window. If you need additional guidance, see LTQ XL online Help and/or the LTQ XL Hardware
Manual.
1. Open the Tune Plus window from the Start button on your Windows XP Desktop, as follows:
a. Choose Start > Programs > Xcalibur > Xcalibur to display the
Xcalibur Home Page – Roadmap view.
On Standby
b. Click the Instrument Setup button to display the Instrument Setup
window.
c. Click the LTQ XL button to display the New Method page.
d. Click the Tune Plus button to display the Tune Plus window.
2. In Tune Plus, click the On/Standby button to take the MS detector out of Standby mode and turn it On. The MS detector begins scanning, the LTQ XL MS detector applies high voltage to the ESI probe, and the LTQ XL MS detector shows a real-time display in the Spectrum view.
3. Open the ESImyTune.LTQTune Tune Method, the Tune Method you saved in Chapter 3, as follows:
a. On the File/Display toolbar, click the Open File icon to display the
Open dialog box.
b. Browse to the folder C:\Xcalibur\methods. Then, select the file
ESImyTune.LTQTune.
c. Click Open to open the file. Tune Plus downloads the Tune Method
parameters to the MS detector, and the title bar in the Tune Plus window should read as follows: C:\Xcalibur\methods\ESImyTune.LTQTune – Tune Plus
74 LTQ XL Getting Started Thermo Electron Corporation
Page 87
4
Tuning with Your Analyte in LC/ESI/MS Mode
Setting Up to Tune the MS Detector with Your Analyte
4. Define the scan parameters for tuning with your analyte in the ESI/MS mode, as follows:
a. In the Instrument Control toolbar, click Define Scan to open the
Define Scan dialog box. See Figure 32.
b. In the Scan Description group box, in the Mass Range list box,
select Normal to allow for a selection of mass ranges between m/z 150 to 2000.
c. In the Scan Rate list box, select Normal to specify a normal scan rate.
d. In the Scan Type list box, select SIM specify a Selected Ion
Monitoring experiment.
e. In the Scan Time group box, in the Number of Microscans spin box,
type 1 to set the total number of microscans to 1.
f. In the Max. Inject Time spin box, type 200.000 to specify a 200 ms
maximum injection time.
g. In the Scan Ranges group box, in the Input list box, select
Center/Width to make available the Center Mass and Width text boxes in the Scan Ranges table.
h. In the Source Fragmentation group box, confirm that the On check
box is not selected ( ) to specify that the ion source fragmentation option is turned off.
i. In the Scan Ranges group box (Scan Ranges table), in the Center
Mass text box, type the mass of your analyte to set the center of mass for the scan range. If reserpine is the analyte you would enter 609.20 to set the center mass for the scan range to m/z 609.20. If you are using another analyte enter its mass value into the Center Mass text box (Figure 32)
j. In the Width text box, type 2.0 to set the width of the scan range to
2.0 daltons.
k. Verify that the settings in your Define Scan dialog box are the same
as those shown in Figure 32.
l. Click OK to apply the MS detector scan parameters and to close the
Define Scan dialog box.
Thermo Electron Corporation LTQ XL Getting Started 75
Page 88
4
Tuning with Your Analyte in LC/ESI/MS Mode
Setting Up to Tune the MS Detector with Your Analyte
Figure 32. Define Scan dialog box, showing typical settings for acquiring reserpine data of the SIM type
5. On the Control/Scan Mode toolbar, click Centroid/Profile to toggle the data type to profile. (The picture on the button should be the same as that shown here.)
6. Click Positive/Negative to toggle the ion polarity mode to positive. (The picture on the button should be the same as that shown here).
You have completed setting up to tune your MS detector with your analyte in ESI/MS mode.
76 LTQ XL Getting Started Thermo Electron Corporation
Page 89
4
Tuning with Your Analyte in LC/ESI/MS Mode

Optimizing the MS Detector Tune Automatically with Your Analyte

Optimizing the MS
Detector Tune
Automatically with
Your Analyte
Optimize the MS detector tune automatically to maximize the ion transmission of reserpine (or your analyte of interest) for a high-flow experiment. Thermo Electron recommends that you begin optimizing after you have successfully passed an automatic tuning procedure and an automatic calibration procedure with the calibration solution infused at 5 μL/min.
The following procedure describes how to optimize the MS detector Tune Method with the reserpine m/z 609.2 peak at an LC flow rate of 400 μL/min. You can also carry out this procedure with your analyte of interest and at your particular LC flow rate. (Refer to Tab le 2 for guidelines about setting flow rates and temperatures.)
Optimize the MS detector Tune Method
1. On the Control/Scan Mode toolbar, click the Tu n e button to display the Tune dialog box.
2. If necessary, click the Automatic tab to display the Automatic tuning page. See Figure 33.
3. In the What to Optimize On group box, select the Mass option button to make active the Mass spin box.
4. In the Mass spin box, enter 609.2 (or the appropriate mass of your analyte of interest) to specify that the LTQ XL MS detector is to use the peak at m/z 609.2 (or the appropriate mass of your analyte of interest) to optimize your tune.
5. Ensure that the Divert/Inject valve is in the Detector position, as follows:
a. Click the Divert/Inject button to open the Divert/Inject Valve
dialog box. See Figure 34.
b. Click the Detector option button.
c. Click Close.
6. Start the automatic tuning procedure from the Tune dialog box, as follows:
a. Click Start. A message box displays the following message:
Please ensure that the 500 microliter syringe is full.
Thermo Electron Corporation LTQ XL Getting Started 77
Page 90
4
Tuning with Your Analyte in LC/ESI/MS Mode
Optimizing the MS Detector Tune Automatically with Your Analyte
b. Click OK to close the message box, and return to the Tune Plus
7. In the File/Display toolbar, click the Graph View button to display the Graph view.
Ensure the syringe pump contains at least 450 μL of the 125 fg/μL reserpine tuning solution.
window.
Figure 33. Tune dialog box, showing the Automatic tuning page
78 LTQ XL Getting Started Thermo Electron Corporation
Page 91
4
Tuning with Your Analyte in LC/ESI/MS Mode
Optimizing the MS Detector Tune Automatically with Your Analyte
Figure 34. Divert/Inject Valve dialog box
8. Observe the Tune Plus window and the Tune dialog box. While automatic tuning is in progress, the LTQ XL MS detector displays various tests in the Spectrum and Graph views in the Tune Plus window and displays various messages in the Status group box in the Tune dialog box. Your Tune Plus window should now look similar to the one shown in Figure 35.
Note The most important parameters that affect the signal quality during ESI/MS operation are electrospray voltage, ion transfer capillary temperature, heated capillary voltage, tube lens voltage, gases, and solution flow rate. If any one of these parameters is changed, you need to reoptimize MS detector parameters. You can use the Semi-Automatic tune procedure to tune the MS detector on individual parameters.
You have now successfully tuned the MS detector in ESI/MS mode for the compound reserpine (or your analyte of interest).
Thermo Electron Corporation LTQ XL Getting Started 79
Page 92
4
Tuning with Your Analyte in LC/ESI/MS Mode
Optimizing the MS Detector Tune Automatically with Your Analyte
Figure 35. Tune Plus window with the Tune dialog box, showing the Automatic tuning page
80 LTQ XL Getting Started Thermo Electron Corporation
Page 93
4
Tuning with Your Analyte in LC/ESI/MS Mode

Saving the ESI/MS Tune Method

Saving the ESI/MS
Tune Method
Save your ESI/MS Tune Method (for a high-flow experiment using your analyte) when automatic tuning is complete
Note Save the Tune Method while the MS detector is On, if any of the
ion source parameters are different from those with which you started.
1. Choose File > Save As to display the Save As dialog box. See Figure 36.
Figure 36. Save As dialog box, showing files in the folder C:\Xcalibur\methods
2. Select the C:\Xcalibur\methods folder.
3. Click the File Name text box, and enter reserpine (or the name of your analyte of interest).
4. Click Save to save the Tune Method, and return to the Tune Plus window. Note that the Tune Method is named reserpine.LTQTune.
Thermo Electron Corporation LTQ XL Getting Started 81
Page 94
The Tune Method is now properly saved and you are ready to acquire data on your analyte of interest.
Page 95
Chapter 5 Acquiring ESI Sample Data
Using the Tune Plus Window
This chapter describes how to acquire LC/ESI/MS sample data using the Tune Plus window. This experiment uses reserpine but you can use the same procedure with your analyte of interest.
Note The following procedures assume that you are familiar with your LTQ XL instrument and the Tune Plus window. If you need information, refer to the LTQ XL online Help, LTQ XL Getting Connected, and/or the LTQ XL Hardware Manual.
Ensure that you have completed the procedures in Chapter 3: “Tuning
and Calibrating Automatically in the ESI/MS Mode” and Chapter 4: “Tuning with Your Analyte in LC/ESI/MS Mode”.
This chapter contains the following sections:
Setting Up to Acquire MS/MS Data in the Full Scan Type
Setting Up to Introduce Sample by Loop Injection into Solvent Flow
from an LC
Acquiring MS Data in the SIM Scan
Thermo Electron Corporation LTQ XL Getting Started 83
Page 96
5
Acquiring ESI Sample Data Using the Tune Plus Window

Setting Up to Acquire MS/MS Data in the Full Scan Type

Setting Up to Acquire
MS/MS Data in the
Full Scan Type
Optimizing the Isolation Width and Setting Up to
Prepare to acquire MS/MS data in the Full scan type on reserpine (or on your analyte of interest). You need to optimize the isolation width and the relative collision energy parameters before you acquire MS/MS data.
You first optimize the isolation width to ensure that the ion of interest is isolated effectively, and then you optimize the collision energy to ensure that fragmentation of the parent ion is efficient. The relative collision energy for a particular analysis depends on the type of sample you are analyzing.
The information in this topic applies to operation of the LTQ XL MS detector in either the ESI or the APCI mode.
This topic contains the following subtopics:
Optimizing the Isolation Width and Setting Up to Optimize the
Collision Energy
Optimizing the Collision Energy Automatically for an MS/MS
Experiment
Optimize the isolation width and set up to optimize the collision energy for an MS/MS experiment
Optimize the Collision
Energy
On Standby
Note The collision energy is optimized on the LTQ XL MS detector by changing the values for the parameter Normalized Collision Energy in the MSn Settings group box of the Define Scan dialog box. For this experiment, and for most applications, leave the parameters Activation Q and Activation Time set to their default values. For more information about these parameters, refer to the online Help.
1. If you have not already done so, from the Tune Plus window, click the On/Standby button to take the MS detector out of Standby mode and turn it On.
2. Ensure that the Centroid data type is selected. (The picture on the button should be the same as that shown here.)
3. Ensure that the scan parameters are defined to acquire MS/MS Full scan data for reserpine (or your analyte of interest), as follows:
a. Click the Define Scan button to open the Define Scan dialog box.
See Figure 37.
84 LTQ XL Getting Started Thermo Electron Corporation
Page 97
5
Acquiring ESI Sample Data Using the Tune Plus Window
Setting Up to Acquire MS/MS Data in the Full Scan Type
Figure 37. Define Scan dialog box, showing initial settings to optimize the isolation width of an MS/MS experiment for
reserpine
b. Verify that the values in your dialog box are the same as those shown
in Figure 37. Start with a relatively wide Isolation Width. Leave the Define Scan dialog box open.
4. At this time you might want to turn on your LC pump and specify a flow rate of 0.4 mL/min, for example, to ensure that your system does not leak.
5. Click the Syringe Pump button to display the Syringe Pump dialog box. See Figure 38.
6. Turn on the syringe pump and set an infusion flow rate of 5 μL/min, as follows:
a. In the Flow Control group box, click the On option button to make
active the Flow Rate spin box.
b. Type 5 in the Flow Rate spin box to specify a rate of 5 μL/min.
Note This procedure assumes that you are using the 500-μL Unimetrics syringe that is provided with your LTQ XL system. If you are using another type of syringe, select the option button corresponding to your syringe.
Thermo Electron Corporation LTQ XL Getting Started 85
Page 98
5
Acquiring ESI Sample Data Using the Tune Plus Window
Setting Up to Acquire MS/MS Data in the Full Scan Type
Figure 38. Syringe Pump dialog box
c. If you are using a standard Unimetrics or Hamilton syringe, set up
the syringe parameters as follows:
i. In the Type group box, click the Unimetrics or Hamilton
option button to specify the proper syringe type.
ii. Click the Vo lu m e list box arrow to display the list of available
volumes, and then select 500 (or your syringe size) from the list to set the proper syringe volume. Note that, if you are using a Unimetrics syringe, the LTQ XL MS detector automatically sets the syringe ID to its proper value of 3.260 mm.
d. If you are not using a Unimetrics or Hamilton syringe, set up the
syringe parameters as follows:
i. In the Type group box, click the Other option button to make
active the syringe ID spin box.
ii. Enter the inner diameter of your syringe in the Syringe ID spin
box.
e. Click Apply to apply the syringe parameters and start the syringe
pump.
f. Finally, move the Syringe Pump dialog box out of the way, to the top
of the monitor screen.
7. In the Tune Plus window, observe the mass spectrum of reserpine (or
your analyte of interest). Also observe the values for NL and IT
86 LTQ XL Getting Started Thermo Electron Corporation
Page 99
5
Acquiring ESI Sample Data Using the Tune Plus Window
Setting Up to Acquire MS/MS Data in the Full Scan Type
(Normalization Level and Ion Time), while you optimize the value of the Isolation Width in the Define Scan dialog box, as follows:
a. In the Define Scan dialog box, in the MSn Settings group box, in the
Isolation Width box, type 3 to specify an isolation width of m/z 3. Then, click Apply.
b. In the Tune Plus window, observe the mass spectrum for the parent
ion of reserpine, m/z 609.2. Ensure that the readback values for NL and IT are relatively stable.
c. Again, in the Define Scan dialog box, in the MSn Settings group
box, in the Isolation Width box, type 2.8 to specify an isolation width of m/z 2.8. Then, click Apply.
Note The optimum value for the Isolation Width is the smallest m/z width (instrument minimum width =m/z 0.4) that gives a mass spectrum of maximum intensity for only the ions of interest. When the optimum Isolation Width is obtained the values for NL and IT are stable and the mass peak for the parent ion is at its maximum intensity and appears symmetrical. An Isolation Width value that is less than the optimum value causes a substaitial drop in the NL reading. A significant drop in sensitivity indicates that the ions of interest are not effectively isolated.
d. Repeat steps b and c above, entering successively smaller values for
Isolation Width. Continue to observe the intensity of the mass spectrum of the parent ion, and ensure that the values for NL and IT are stable with each change you make to the Isolation Width.
Note After the Isolation Width is optimized, you can compensate for minor changes in tune stability by increasing the Isolation Width value a small amount. This adjustment should be no larger than m/z=1.
8. In the Define Scan dialog box, in the MSn Settings group box, in the Normalized Collision Energy box, type 20 to specify an initial value of 20 for the collision energy. Click Apply.
9. In the Tune Plus window, observe the mass spectrum of the product ions of reserpine (or your analyte of interest). If necessary, increase the value for the Normalized Collision Energy in increments of 5% to cause the clear display of product ion mass spectrum. (After each change in value, click Apply to implement the change.) See Figure 39.
Thermo Electron Corporation LTQ XL Getting Started 87
Page 100
5
Acquiring ESI Sample Data Using the Tune Plus Window
Setting Up to Acquire MS/MS Data in the Full Scan Type
10. When you have clearly identified a product ion mass-to-charge ratio for
11. In the Tune dialog box, click the Collision Energy tab to display the
reserpine (or your analyte of interest), click the Tu n e button to display the Tune dialog box.
page. See Figure 40.
Figure 39. Define Scan dialog box, showing typical settings for acquiring MS/MS data in the Full scan type on reserpine
12. Click the Product Ion Mass option button to make active the spin box.
Ty pe 397.2 to specify the product ion at m/z 397.2 for reserpine. The LTQ XL MS detector can optimize collision energy automatically by using this product ion of reserpine.
88 LTQ XL Getting Started Thermo Electron Corporation
Loading...