ThermoProbe TP7-D, TP9-A User Manual

Thermoprobe Models TP7D and TP9A
English | Sp
ANUAL
TP7-D TP9-A
SER
M
anish | French | Russian | Chinese
EUDeclarationofConformityaccordingtodirective2014/34/EU(ATEX)
Thermoprobe,Inc.herebydeclarestheTP7‐DandTP9‐Aproductstobeinaccordancewiththefollowingstandards anddirectives:
Nameandad
DescriptionofDevices
dressofManufacturer ThermoProbe,Inc.
112AJetportDr. Pearl,MS39208USA
 TP7DandTP9A
PortableElectronicThermometers
Ex-Designation
ECTypeExaminationCertificateNotifiedBod
Auditingbod
AppliedHarmo
EN600790:2012+A112013 Explosiveatmospheres–Part0:Equipment–Generalrequirements EN6007911:2012 ExplosiveAtmospheres–Part11:EquipmentProtectionbyIntrinsic
AppliedEuropeanDire
2014/34/EU‐Equipmentandprotectivesystemsintendedforuseinpotentiallyexplosiveatmospheres
ThermoProbe,Inc.
LukeBartkiewicz President
II 1 G Ex ia IIB T4 Ga
yIntertekTesting&CertificationLimited
y(QAN) IntertekTesting&CertificationLimited
nizedStandards
ctives
 ITS17ATEX201515X
IntertekHouse CleeveRoad Leatherhead, SurreyKT227SB,UK
IntertekHouse CleeveRoad Leatherhead, SurreyKT227SB,UK IdentificationNumber:0359
Safety‘i’
10/10/2017,JK
USER INSTRUCTIONS - TP7-D & TP9-A
INTRODUCTION
This manual describes the basic function, use and safety instructions for a model TP7-D and TP9-A portable di gital thermometer instrument.
REPLACING BATTERY
When the battery voltage is low, the low battery icon will show on the display. When the battery voltage is very low, the backlight will no longer operate and the low battery icon will ‘blink’ on the
display. Replace batteries as soon as possible in a safe location after the low battery is noticed. This will ensure backlight
operation, and avoid possible malfuncti oni ng. Do not at t em pt t o cal i brat e the i nst rument if the low battery indicator is displayed.
WARNING:
Batteries must be changed in Non-hazardous area. Batteries must be of correct approved type. Batteries must be installed with correct polarity making sure the (+) end of the battery is aligned with (+)
symbol embossed in the b attery c ase.
Batteries must not be installed with polarity reversed where one cell could charge another cell. New batteries must not be mixed with old batteries. Batteries must not be mixed with batteries of other
models or manufacturers.
a) Ensure the instrument is in a non-hazardous area & powered off. b) Use a #2 Phillips drive to remove the 3 screws holding the front cover on the TP7-D or the 2 screws holding the front cover on the TP9-A. c) Use a #1 Phillips drive to remove the single screw from the battery cover. Remove the battery cover, push one battery towards the spring contact and lift battery up from the holder, and then remove the r e maining bat tery. d) Install each new battery making sure the (+) end of the battery is aligned with (+) symbol embossed in the battery case. e) Replace the retaining device and reinstall the cover.
CERTIFIED Batteries for the TP9-A and TP7-D are as follows: Manufacturer Type Part Number
Duracell AA (LR6) Alkaline MN1500 Panasonic AA (LR6) Alkaline LR6XWA GP (Gold Peak) AA (LR6) Alkaline GP15A
AUTHORIZED REPAIR
It is recommended that service beyond the scope of this m anual be performed by ThermoProbe, Inc. or one of its authorized distributors.
See www.thermoprobe.net for video on proper use of this instrument. Refer to Am erican Pet roleum Institut e measurement standard Chapter 7.2.
10/2017, JK
g
r
USER INTERFACE
Power/Select Button
Up/Down Menu Buttons Icons for Memory
readin
s
Up/Down Menu Buttons
Temperature indicating arrows
Indicators for calibration points
Low battery indicato
Power Button:
Pressin
g the Power button once will turn on the device. (Note: The instrument will automatically power off 20 mi n ut e s
after the last button push.) Pressing and holding the “Power” button until it displays ‘OFF’ will turn off the instrument.
Selection MENU:
Use the “Up/Down” Menu buttons
Saves current stable temperature up to 4 readings
LOG:
C-F: Select Celsius or Fahrenheit temperature display
DECIMAL: Select 0.1 or 0.01 display resolution
MEMORY - Use “up/down” buttons to display:
Lowest reading Average reading Highest reading Saved (Logged) readings 1-10
(Use CLR LOG to erase saved readings)
CALIBRAT: Enter calibration/adjustment mode (Requires 3 digit password: 112) - See Calibration Procedure
for the following selections:
10/2017, JK
Backlight
When the instrument is operating in low-light conditions a photocell will detect this situation and turn the backlight on.
Temperature Logging
It is necessary for the temperature to be stable before logging temperature. The display arrows will flash 3 times when the temperature reading has stabilized.
If you wish to log the temperature, Press the “down” button once until LOG is displayed. Press the “Select” button once to save a reading. An acknowle dgment of a sa ved reading will occur with a display of “LOG 1-10”. This can be repeated for up to 10 saved readings.
The logged temperatures can be accessed from the MEMORY menu. Logged temperatures will display with the LOG icon and a prefix (1,2,3,4,5,6,7,8,9,0) corresponding to the previously logged readings. The up/down b uttons can be used to advance through the readings. Select EXIT to return to the temperature display.
Logged readings will be retained even after the unit is powered off. Logged readings can be cleared by going to the Memory section of the Menu and selecting CLR LOG. New readings cannot be taken until the readings are cleared.
USB Memory
A micro-usb connection is available on the circuit board allowing access to calibration & logged data. Warning: Do not access the circuit board in a hazardous location.
Lowest, Highest & Average readings
The lowest, highest and average readings can be accessed through the MEMORY menu. The readings are indicated by the LOW, AVG or HIGH icons on the display. These readi ngs are not related to the logged readings, but are determined from the temperature when the unit is powered on. These readings are deleted after the unit is powered off.
Display Codes
OPEN CKT - Indicates the sensor is operating above its tem perature l imit, the Probe Assembly is open circuited
from a cut or broken section, or the cable is not properly inserted at the circuit board t ermi nal. The most common cause is a damaged cable.
SHORT CKT - Indicates the sensor is operating below its temperature limit or the Probe Assem bly is short circuite d
due to a smashed or cut section. The most common cause is a damaged cable.
NONE – There are no logged readings saved in memory. 2 WIRE – A 2 wire probe connection has been detected. 3 WIRE – A 3 wire probe connection has been detected. NO CAL – The device does not have stored calibration data for the temperature probe. Perform an
adjustment/calibration before use.
10/2017, JK
Probe Types
The TP7-D or TP9-A can use either 2-wire or 3-wire probe assemblies.
ThermoProbe replacement probe assemblies are availa bl e i n different configurations. The cable length is available in lengths up to 50 meters or 165 feet. Standard brass marki ngs are available applied in 5 fe et or 1 meter increments. The sensors are available with 4 weight types. The probe assembly part confi gur at i ons are as fo l l ows:
REPLACING THE PROBE ASSEMBLY
NOTES:
1) Replacement of the probe assembly requires re-calibration of the device. Replacement should only be done by experienced personnel and if calibration equipment is available.
2) Please refer to IEC/EN 60079-19 (Explosive atmospheres - Part 19: Equipment repair, overhaul and reclamation) when making the repair.
3) Only use replacement probe assemblies obtained from ThermoProbe, Inc. or one of its authorized distributors.
a) First fo b) On the circuit board push the terminals clamps down and remove the wires noting the wire lead color code arrangement. See Figure 1.
c) Set the cover and circuit board aside and remove the strain relief knot in cable assembly. d) Unwrap the cable from the assembly and pull the cable free of the rubber grommet. e) Insert the new cable wire through the rubber grommet and then pull several inches of cable past the grommet. f) Tie a simple overhand knot in the cable at the grommet for strain relief and pull the knot up to the grommet. g) On the circuit board, push the terminal clamp levers down and insert the new wire leads according to the terminal color codes. The label indicates how to connect a 2 wire probe vs. a 3 wire probe (includes green wire). See Figure 1. h) Reinstall the batteries and cover and re-spool the cable assembly. i) Perform a calibration (see calibration procedure).
Figure 1: Probe Assembly Lead Attachment
llow REPLACING BATTERY instructions a through c to remove batteries.
White – positive sensor wire Green – Silver – negative sensor wire & shield wire
cable compensating wire (not used on 2 wire models)
10/2017, JK
CALIBRATION PROCEDURE
The calibration mode should only be accessed by qualified personnel with proper equipment; otherwise
calibration integrity may be compromised. Read the following instructions carefully.
At least 2 points are required to make an adjustment. (2-Point calibration). Additional points can be taken
(3-Point or 4-Point calibration) to calibrate a large range of temperatures (e.g. 0°F to 300°F) or if you want to match specific points in your range. You should include points at the bottom and top of the range. You must have the proper equipment for every point of calibration.
Do not attempt to calibrate the instrument if the low battery indicator has been displayed since the new
calibration values may not be properly s tored to memory.
Refer to API 7.2 or another recognized standard for routine calibration verification recommendations. Calibration must not be performed in any environment considered to be hazardous.
Equipment needed:
Ice Bath or other low temperature bath with reference thermometer. Warm to hot fluid bath between 20°C (approx 68°F), or higher up to 90°C (approx 194°F) with reference
thermometer. (see Note)*
Optional high temperature oil bath at about 150°C/300°F and reference thermometer.
*Note for limited calibration: If entire range capability of instrument is not required, the 2 point high adjustment can be made at a temperature relatively close to the common temperature of the liquid measured and accuracy will be maintained within the limited range. For example: If liquid product to be measured is commonly less than 38°C (approx. 100°F), then a “high point” calib ra t io n c a n be made near that temperature. Temperature accuracy above this calibration point cannot be assured.
To calibrate proceed with the following steps:
1. Enter calibration mode by going to CALIBRAT in the selection Menu. When this is selected, the user will be
prompted to enter a 3 digit password. Using the “up/down” buttons allows each digit to be adjusted. Once the correct value is set, advance to the next digit by pressing the “select” button. After the 3 “select” button to enter Calibration mode. If an incorrect password is entered, FAIL will display and the unit will return to temperature mode. Once in Calibration mode the ‘CAL’ icon on the display will blink. The calibration mode password is: 112
2. Select ADJUS
2PT CAL, 3PT CAL or 4PT CAL using the select button.
T from the calibration menu and then select the desired number of calibration points. Pick
NOTE: If the user is not ready to enter the calibration mode, the EXIT option can be chosen.
rd
digit is set press the
10/2017, JK
3. T
he device is now in adj ustm en t m ode . The di spl ay will show ADD PT1. The “up/down” buttons can be used
to select a different point to adjust or CANCEL. If the CANCEL option is chosen then the calibration procedure is exited and the prior calibration values are re-activated.
ADD PT1 = Lowest temperature point ADD PT2 = the next higher temperature point ADD PT3 = the next higher temperature point (only used in 3-point calibrati on or 4-point calibrati on mode) ADD PT4 = the highest temperature point (only used in 4-point calibrati on mode) ADJ DONE = Save and exit calibration mode (all points adjusted & valid) CANCEL = Exit calibration mode without saving
Pick which point to adjust and press the select button to begin. The current temperature will display and the CAL icon and the number of the point to adjust will now be blinki ng.
NOTES:
o Calibration can be performed to hundredths of a degree. o The up/down buttons can be used to increase or decrease the display reading. o Holding the up/down buttons adjusts 0.1 degrees increments. o Momentary Presses of the up/down buttons for less than 0.5 seconds adjusts 0.01 degrees for every
press.
o The display arrows will flash 3 times when the temperature reading has stabilized. o All points must be saved before selecting ADJ DONE. A DATA ERR message will display if all
points have not been saved or are not in increasing temperatures. (Ex: PT1 = 32F, PT2 = 120F, PT3 = 250F)
o While in calibration mode the temperature will display based on the prior calibration curve. o The RESETCAL feature can be used to return the unit to a factory calibration curve.
(calibration/adjustment will still be required ).
o The RESETCAL (reset to factory calibration) feature should be used i f th e pro be is repl a ced usi ng
a 2-wire assembly instead of a 3-wire assembly or if prior calibration has caused the unit to read excessively out of tolerance.
4. Once the temperature has stabilized, using a reference device check the actual temperature in the bath and use
the “up/down” buttons to adjust the device to the a ctual temperature. Once the device temperature matches the actual temperature, press the select button to save the set ti ng. The display will show SAVE or EXIT. If
EXIT is selected it will return to the ADD PT1 menu. Once SAVE is
selected the display will advance to the
next temperature point for adjustment showing ADD PT…. The up/down buttons can be used to select a different point to adjust or CANCEL. If the CANCEL option is chosen then the calibration procedure is exited and the prior calibration values are re-activated.
o NOTE: Once SAVE is selected the temperature display will return to the previous reading until the
calibration is completed.
o NOTE: If SAVE is selected before the temperature has stabilized, the display will show NOT
STABLE. Wait for the temperature to stabilize before saving.
5. Move the probe to the next bath and repeat step 4. After you save the highest temperature point the display
will flash READY and the new calibration settings will be in effect. The buttons will now resume their normal operating functions. The calibration setti ngs are saved to flash memory when the device is turned off. The unit will not turn off automatically. Manually turning the unit off saves the calibration data.
ThermoProbe, Inc.
112A JETPORT DR.
PEARL, MS 39208
Tel: +1 601.939.1831
Fax: +1 601.355.1831
sales@thermoprobe.net
www.thermoprobe.net
10/2017, JK
SAFETY INSTRUCTIONS - TP7-D & TP9-A REV 102017
These ThermoProbe instruments are intended for use in both hazardous (potentially flammable or explosive) and non­hazardous areas under dry conditions at ambient temperatures between -20 to 40°C.
The instruments are not intended for use in permanent outdoor installations and are not intended or tested for i cing conditions. stresses (e.g. vibration, heat, impact, etc.). requirements.
a) The user must have a thorough knowledge of the products to be measured and must know of the safety precautions to be taken when working with the material to be measured. b) The instrument shall be checked concerning severe defects; check that instrument is com plete (i ncl uding
grounding/bonding cable), has good batteries, etc. If necessary, check measurement accuracy. If any defects are found, the instrument should not be used until repairs have b een made. c) The instrument, especially cable and probe, should be clean for safety and ease of use. d) The physical measurement location should be evaluated for primary and secondary risks. e) Power source must be removed before performing any maint enance. f) Exchange of components other than the batte ri es may compromise ATEX/IECEx or other certifications and shall only be undertaken by ThermoProbe or one of its qualified service providers. See also “Authorized Repair” section. g) To reduce the risk of fire or explosion, this device must be bonded to the vessel according to clause 6.3.2 e), IEC/EN 60079-14 before and during introduction into the vessel and shall remain bonded until the sensor probe is completely withdrawn from the vessel. h) The device must remain bonded to ground/earth using the provided connection whenever a hazardous atm osphere could be present as well as d uri n g s i t uat i o ns wh ere electrostatic charging can occur such as the the thermometer cable or filling or emptying of the tank.
CAUTION: In the event that any part of the instrument should become electrostatically charged in a potentially hazardous location, follow company policies for testing and clearing the area of any hazardous gases before attempting to bond the instrument to earth ground. If this is not possible allow sufficient time for the instrument to naturally dissipate any charges before attempting to bond to earth ground. Given the atmosphere, this could take several hours.
Additional means of protection should be used where the equipment may be exposed to excessive external
The user must have a working knowledge of appropriate safety
unwinding/winding of
GUIDANCE NOTE
Problems with aggressive substances and environments: Be aware of aggressive substances and that ext ra protection may be needed. Caustic soda, highly basic and acidic substances will erode aluminum and copper ground clip and wire. The Sensor-
Cable assembly has external surfaces of stainless steel and fluoropolymer material. Exposure to Excessive heat can melt the plastic components of the instrument.
10/2017, JK
SAFETY APPROVALS FOR TP7-D AND TP9-A:
AEx ia IIB T4 Ga
Applicable Standards are: Agency or Safety Designation
IEC 60079-0:Ed 6, IEC 60079-11:Ed 6 IECEx EN 60079-0:2012 + A11 3013, EN60079-11:2012 Europe: ATEX
INTRINSIC SAFETY
Intrinsically safe equipment is defined as "equipment and wiring which is incapable of releasing sufficient electrical or thermal energy under normal or abnormal conditions to cause ignition of a specific hazardous atmospheric mixture in its most easily ignited concentration." (ISA-R P1 2.6 ) This is achieved by limiting the amount of powe r avai l a bl e t o the electrical equipment in the hazardous area to a level below that whi ch will ignit e the gas es.
In order to have a fire or explosion, fuel, oxygen and a source of ignition must be present. An intrinsically safe system assumes the fuel and oxygen is present in the atmosphere, but the system is designed so the electrical energy or thermal energy of a particular instrument loop can never be great enough to cause ignition.
BATTERIES
WARNING:
Batteries must be changed in Non-hazardous area. Batteries must be of correct approved type. Batteries must be installed with correct polarity making sure the (+) end of the battery is aligned with (+)
symbol embossed in the b attery c ase.
New batteries must not be mixed with old batteries. Batteries must not be mixed with batteries of other models or manufacturers. Batteries must not be installed with polarity reversed where one cell could charge another cell.
CERTIFIED Batteries for the TP9-A and TP7-D are as follows: Manufacturer Type Part Number
Duracell AA (LR6) Alkaline MN1500 Panasonic AA (LR6) Alkaline LR6XWA GP (Gold Peak) AA (LR6) Alkaline GP15A
10/2017, JK
INSTRUCCIONES PARA EL USUARIO - TP7-D Y TP9-A
INTRODUCCIÓN
Este manual describe las instrucciones básicas de funcionamiento, uso y seguridad, para los modelos de instrumento de termómetro digital portátil TP7-D y TP9-A.
CAMBIO DE LA BATERÍA
Cuando la batería está baja, en la pantalla aparece el ícono de batería baja. Cuando la batería está muybaja, la retroiluminación ya no funciona y el ícono de batería baja "parpadea" en la
pantalla. Al reconocer que la carga de las baterías está baja, cámbielas tan pronto como sea posible y en un lugar seguro. Esto
garantiza el funcionamiento de la retroiluminación y evita el posible mal funcionamiento. No intente calibrar el instrumento si el indicador de batería baja está encendido.
ADVERTENCIA:
Se deben cambiar las baterías en un área sin riesgos. Las baterías deben ser del tipo aprobado correcto. Se deben instalar las baterías con la polaridad correcta, asegurándose de que el lado (+) de la batería esté
alineado con el símbolo (+) realzado en el espacio de la batería.
No se deben instalar las baterías con la polaridad invertida, donde una celda podría cargar otra celda
equivocada.
No se deben mezclar baterías nuevas con baterías viejas. No se deben mezclar las baterías de diferentes
modelos o fabricantes.
a) Asegúrese de que el instrumento esté en un área no peligrosa y que esté apagado. b) Utilice un desatornillador Phillips n.º 2 para retirar los 3 tornillos que sostienen la tapa delantera en el TP7-D o los 2 tornillos que sostienen la tapa delantera en el TP9-A. c) Utilice un desatornillador Phillips n.º 1 para retirar el tornillo de la cubierta de la batería. Retire la cubierta de la batería, empuje una batería hacia el contacto de resorte y saque la batería del espacio, luego retire la batería restante. d) Instale cada batería nueva asegurándose de que el lado (+) de la batería esté alineado con el símbolo (+) realzado en el espacio de la batería. e) Vuelva a colocar el dispositivo de retención y reinstale la cubierta.
Las baterías CERTIFICADAS para el TP9-A y para el TP7-D son las siguientes:
Fabricante Tipo Número de parte
Duracell AA (LR6) Alcalina MN1500 Panasonic AA (LR6) Alcalina LR6XWA GP (Gold Peak) AA (LR6) Alcalina GP15A
REPARACIÓN AUTORIZADA
Se recomienda que el mantenimiento que no esté incluido en el ámbito de este manual sea llevado a cabo por ThermoProbe, Inc. o por uno de sus distribuido res aut o ri zad os.
10/2017, JK
Vea en www.thermoprobe.net un video sobre el uso adecuado de este instrumento. Consulte el Capítulo 7.2 de la norma de medición del Instituto Americano del Petróleo (American Petroleum Institute).
INTERFAZ DEL USUARIO
Botón de Encendido/Selección
Botones Arriba/Abajo Íconos para lecturas
de memoria
Botones Arriba/Abajo
Flechas indicadoras de temperatura
Indicadores de puntos de calibración
Indicador de batería
Botón de encendido:
Al
presionar una vez el botón de Encendido se enciende el dispositivo. (Nota: El in strumento se apaga automáticamente 20 minutos después de la última presión en el botón). hasta que la unidad muestra "OFF", se apaga el instrumento.
Al presionar y sostener el botón “Power”
10/2017, JK
MENÚ de selección:
Utilice los botones de menú “Arriba/Abajo”
LOG: Guarda la temperatura estable actual por hasta 4 lecturas.
C-F: Selecciona grados Celsius o Fahrenheit para la temperatura que aparecerá en pantalla.
DECIMAL: Selecciona la resolución de 0.1 o 0.01 para la pantalla.
MEMORY - Utilice los botones “Arriba/Abajo” para mostrar:
Lectura más baja Lectura promedio Lectura más alta Lecturas guardadas (Logged) 1-10
(Use CLR LOG para eliminar las lecturas guardadas)
CALIBRAT: Ingresa al modo de calibración/ajuste (Requiere una contraseña de 3 dígitos: 112) - Cons ul t ar el Proce di mie nto de calibración
para las siguientes selecciones:
Retroiluminación
Cuando el instrumento opera en condiciones de baja iluminación, una fotocelda detecta la situación y enciende la retroiluminación.
Registro de temperatura
Es necesario que la temperatura esté estable antes de registrarla. La flechas de la pantalla parpadean 3 veces cuando la lectura de la temperatura se estabiliza.
Si desea registrar la temperatura, presione el botón “down” hasta que aparezca LOG. Presione el botón “Select” una vez para guardar una lectura. Se presenta una confirmación de la lectura guardada mostrando “LOG 1-10”. Se puede repetir esto para guardar hasta 10 lecturas.
Se puede acceder a las temperaturas guardadas desde el menú MEMORY. Las temperaturas guardadas se muestran con el ícono LOG y un prefijo (1, 2, 3, 4, 5, 6, 7, 8, 9, 0) correspondiente a las lecturas guardadas previamente. Se pueden usar los botones arriba/abajo para desplazarse entre las lecturas. Seleccione EXIT para regresar a la pantalla de temperatura.
10/2017, JK
ecturas guardadas se conservan incluso después de apagar la unidad. Se pueden eliminar las lecturas guardadas
Las l al ingresar a la sección Memory y seleccionar CLR LOG. No se pueden tomar nuevas lecturas hasta que se eliminan las lecturas.
Memoria USB
Hay una conexión micro-USB disponible en la tarjeta del circuito que permite acceder a la calibración y los datos guardados. Advertencia: No acceda a la tarjeta del circuito en un área peligrosa.
Lecturas mínima, máxima y promedio
Se puede acceder a las lecturas mínima, máxima y promedio a través del menú MEMORY. Las lecturas tienen una indicación de ícono LOW (mínima), AVG (promedio) o HIGH (máxima) en la pantalla. Estas lecturas no están relacionadas con las lecturas guardadas, se determinan a partir de la temperatura cuando se enciende la unidad. Dichas lecturas se borran cuando se apaga la unidad.
Códigos de pantalla
OPEN CKT - Indica que el sensor está funcionando por encima de su límite de temperatura, se abre el circuito del
Ensamble del sensor de una sección cortada o rota o el cable no está insertado adecuadamente en la terminal de la tarjeta de circuito. La causa más común es un cable dañado.
SHORT CKT - Indica que el sensor está funcionando por debajo de su límite de temperatura o que el Ensamble del
sensor está en cortocircuito debido a una sección rota o cortada. La causa más común es un cable dañado.
NONE – No hay lecturas guardadas en la memoria. 2 WIRE – Se detectó la conexión de una sonda de 2 cables. 3 WIRE – Se detectó la conexión de una sonda de 3 cables. NO CAL – El dispositivo no tiene datos de calibración almacenados para la sonda de temperatura. Realice un
ajuste/una calibración antes del uso.
Tipos de sonda
El TP7-D o el TP9-A pueden utilizar ensambles de sonda de 2 o de 3 cables.
Los ensambles de sonda de reemplazo ThermoProbe están disponibles en diferentes configuraciones. Hay cables disponibles en longitudes de hasta 50 met ros o 16 5 pi es. Hay marcas estándar de latón disponibles aplicadas en incrementos de 5 pies o 1 metro. Los sensores están disponibles en 4 pesos diferentes. Las configuraciones de la pieza del ensamble de sonda son las siguientes:
10/2017, JK
REEMPLAZO DEL ENSAMBLE DE SONDA
NOTAS:
1) El reemplazo del ensamble de sonda requiere de una nueva calibración del dispositivo. El remplazo solo debe hacerlo personal experimentado y solo si hay equipo de calibración disponible.
2) Consulte IEC/EN 60079-19 (Atmósferas explosivas - Parte 19: Reparación transformación o recuperación de un equipo) al hacer la reparación.
3) Utilice únicamente ensambles de sonda de reemplazo obtenidos de ThermoProbe, Inc. o de uno de sus distribuidores autorizados.
a) Siga primero las instrucciones de REEMPLAZO DE BATERÍA a hasta c para retirar las baterías. b) En la tarjeta del circuito presione las abrazaderas de las terminales hacia abajo y retire los cables teniendo en cuenta el arreglo por colores de las terminales de los cables conductores. Vea la Figura 1.
Blanco – cable positivo del sensor Ver
de – cable de compensación (no usado en modelos de 2 cables)
Plateado – cable negativo del sensor y con blindaje c) Coloque aparte la cubierta y la tarjeta de circuito y deshaga el nudo de alivio de tensión en el ensamble del cable. d) Desenrolle el cable del ensamble y jale el cable para liberarlo del ojal de hule. e) Inserte el nuevo cable a través del ojal de hule y luego jálelo varias pulgadas más allá del ojal. f) Haga un nudo simple en el cable a la altura del ojal como alivio de tensión y jale el nudo contra el ojal. g) Presione hacia abajo las palancas de las abrazaderas de las terminales e inserte las nuevas terminales de acuerdo con el código de colores. La etiqueta indica la manera de conectar la sonda de 2 cables o la sonda de 3 cables (incluye cable verde). Vea la Figura 1. h) Vuelva a instalar las baterías y la cubierta y enrolle el ensamble del cable. i) Realice una calibración (consulte el procedimiento de calibración).
Figura 1: Sujeción del cable del ensamble de sonda
PROCEDIMIENTO DE CALIBRACIÓN
Únicamente personal calificado y con el equipo adecuado debe acceder al modo de calibración, de otra
forma puede verse comprometida la integridad de la calibración. Lea cuidadosamente las siguientes instrucciones.
Se requieren por lo menos 2 puntos para hacer un ajuste. (C al i bració n de 2 pu nt o s ). Se puede n t omar
puntos adicionales (calibración de 3 o 4 punt os) pa ra calib r a r un rang o amplio de temperaturas (p.ej. 0 °F a 300 °F) o si se desea hacer coincidir puntos específicos en el rango. Se deben incluir puntos en la parte inferior y en la parte superior del rango. Se debe contar con el equipo adecuado para cada punto de calibración.
No intente calibrar el instrumento si el indicador de batería baja está encendido, ya que los valores de la
nueva calibración pueden no ser guardados adecuadamente en la memoria.
10/2017, JK
Loading...
+ 32 hidden pages