ThermoProbe TL1-A User Manual

0 (0)
ThermoProbe TL1-A User Manual

ThermoProbe

Report No.: 2013-06-04 - 1-11923

Model: TL1-A

Calibration Data As Found

New Unit or no "As Found" data available

Report of Test

This report is to certify that the instrument listed below has been calibrated by ThermoProbe, Inc. to NIST traceable criteria.

Unit SN: 1-11923

Calibration Date: 6/4/2013

Ambient Temp: 75 °F +/- 2º

Calibrated By: MS

Report of Test

Test Method: The calibration procedures used were ThermoProbe, Inc. Calibration Procedures based on ASTM E- 644-06. This probe was immersed in a constant temperature bath with a reference thermometer which determined the actual test temperature. The readings were compared and correction factors for the probe were calculated. The As Left readings reflect the TL1's readings after calibration

Nominal Temp

Bath

Fluid

Reference

Calibration Date

Next Calibration

Due

 

 

 

 

 

 

(-)20.0º C

(-)4.0º F

Fluke 7340

water/glycol

TL1R-6068

3/27/2013

3/27/2014

 

 

 

 

 

 

 

0.0º C

32.0º F

Fluke 7340

water/glycol

TL2-0016

3/27/2013

3/27/2014

 

 

 

 

 

 

 

48.9º C

120.0º F

PolyScience 8101

mineral oil

TL2 - 0008

7/20/2012

7/20/2013

 

 

 

 

 

 

 

92.8º C

199.0º F

Fluke 6330

mineral oil

Fluke 1502A - ASP WSP500

3/27/2013

3/27/2014

 

 

 

 

 

 

 

149.0º C

300.2º F

Fluke 6330

Silicon oil

Fluke 1502A - ASP WSP500

7/20/2012

7/20/2013

 

 

 

 

 

 

 

290.0º C

554.0º F

Fluke 9172

metrology well

Fluke 1502A - ASP WSP500

7/20/2012

7/20/2013

 

 

 

 

 

 

 

Calibration Data As Left

This device has been adjusted to read as closely as possible to actual temperature.

Tested temperatures and corrections are as follows:

Nominal Value

Actual Test Temp.

Reading of TL1

Correction

Tolerance

In

Measurement

Tolerance

Uncertainty

 

 

 

 

 

 

 

 

 

 

º F

º C

º F

º C

º F

º C

º F

º C

º F

º C

 

º F

º C

32

0

32.011

0.006

32.03

0.02

-0.02

-0.01

0.10

0.06

Yes

0.030

0.017

120

49

120.015

48.897

120.02

48.90

0.00

0.00

0.10

0.06

Yes

0.030

0.017

199

93

199.022

92.790

199.05

92.81

-0.03

-0.02

0.10

0.06

Yes

0.030

0.017

300

149

300.230

149.017

300.23

149.02

0.00

0.00

0.10

0.06

Yes

0.030

0.017

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Callendar Van Dusen Coefficients:

R0: 100.00

A: 3.91400E-03

B: -6.03510E-07

C: -4.17100E-11

Traceability: This calibration is traceable to NIST through an unbroken chain of comparisons. IsoTech calibrated the references shown above using transfer standards which in turn were calibrated by their primary reference thermometer.

Uncertainty Statement: Uncertainties were computed using the concepts, methods and techniques of the ISO Guide to the Expression of Uncertainty in Measurement (the GUM). The calculated uncertainty is an expanded uncertainty (k=2). It does not consider errors due to possible damage to the TL1 from shipping, temperature drift, or thermal hysteresis effect. To maintain the accuracy of the TL1, users should take care to protect it during shipping, avoid using it to measure temperatures significantly above the highest calibrated temperature, and have the TL1 recalibrated annually.

Calibrator's Signature: ___________________________________________

Test Results Approved by: _______________________________________

Date: 6/4/2013

The results stated on this report relate only to the items specifically identified. This test report or calibration certificate shall not be reproduced except in full, without written approval of the laboratory.