The TL2-A is an affordable high precision digital thermometer. It has been designed for use in both laboratory
and industrial applications. Additional video tutorials and information are included on the software USB flash
drive. Features of the TL2-A are as follows:
Configurable RTD and SPRT combinations (100Ω or 200Ω)
• Available 2 channel temperature measurement
• 0.001 degree resolution (0.0001 for Ω)
• Implements ITS-90 or Callendar-Van Dusen calibration coefficients
• Displays temperature in Celsius, Fahrenheit, Kelvin or Ohms
• Backlit, Large LCD character display
• USB Type ‘A’ PC interface
• USB Type ‘B’ flash memory interface
• User programmable relay output interface (Temperature triggered)
• Detachable AC power cord
• IP50 enclosure rating
• Adjustable Zero Offset correction
Cautions
The TL2-A and associated temperature probes are sensitive devices and should be handled with care. Care
should be taken not to submit the temperature probes to mechanical shock, over-heating or exposure to fluids
at the probe to cable interface. Overheating the probes beyond their calibrated temperature range, damage
and/or contamination may cause sensor drift and instability.
Authorized Service
Service and Calibration should only be performed by ThermoProbe, Inc. or an authorized distributor.
Contact ThermoProbe for questions about setting up this software or using the TL2-A.
Power Button (AC Unit) – Use the power button to save your settings (temperature units, display decimals,
zero adjustments, relay settings, etc.), and to place the TL2-A in ‘Standby’ mode. If you remove AC the power
without putting the TL2-A in Standby mode, the user settings will revert to the previous settings.
Select Button - Activates the user settings menu and selects menu items.
Up and Down Buttons - Selects the display for channel 1 and channel 2, navigates the menu (display
next/previous menu item). Also, adjusts the zero correction when in the Zero Adjust Mode.
Channel Indication – Marker indicates which channel is currently being displayed. (For active display and
menu functions)
USER INTERFACE REAR
Channel 1 Channel 2
Fuse
On/Off Switch
Channel 1 &2 – Female sockets for probe connections.
USB1 - type ‘A’ socket for connecting flash USB devices
USB2 – type ‘B’ socket for PC interface cable
Relay Header – User configurable normally open or normally closed relay control
• All displayed values and changes are for the current selected channel only.
• To save settings, the Power Button must be pressed. This will set the unit to ‘Standby’.
• The low, average and high readings are cumulative from the time the TL2-A is powered on. They are
not saved when the TL2-A is shut down or put in standby and will re-start when the unit is turned on.
The low, average, and high readings will be displayed once the TL2-A's readings have stabilized (about
22 seconds.)
USB Flash Drive Logging (AC powered unit only):
1. Insert a flash drive into the USB1 type ‘A’ socket on the back of the TL2-A.
2. From the record menu, select ‘FLASH’ and ‘StArt’.
3. The TL2-A display will flash ‘ACCESS’ until the flash drive recording has started. (A ‘no DSC’ message will appear if the
flash drive is not detected).
4. The up/down arrow indicator will blink on the LCD display to indicate that logging is in process.
5. To stop the data logging, select ‘FL StoP’ from the menu. Wait for the ‘FL DonE’ message to display before removing the
flash drive from the TL2-A. (If the drive is removed before the logging has been stopped, the data will be lost.)
Once the flash drive data logging has been completed, the data can be opened on a standard PC.
Note: Each time the TL2-A records to the flash drive it will create a new file numbered sequentially higher than the previous file.
(ex: 0001.csv, 0002.csv, etc.)
1. Remove the flash drive from the TL2-A.
2. Insert the flash drive into a USB port on the PC.
3. Browse to the folder containing the flash drive contents.
4. The TL2-A data will be contained in a .csv file that can be opened with either Microsoft Excel or a text editor such as
Notepad.
5. The column format is as follows: Date, Time, Channel 1 reading, Channel 1 units, Channel 2 reading and Channel 2 units.
You can set the TL2-A to send data on command. This is useful for people who write data collection software that will request data
only when it is needed. To place the TL2-A in polling mode, select “PoLL” as the sample rate to be used under the recording section
of the TL2-A menu. (See Interface Menu Diagram) When the PC sends a question mark character (?) to the TL2-A, the TL2-A will
respond with the temperature. When the TL2-A is set to flash record mode, the polling feature will not work.
Temperature controlled Relay (AC powered unit only):
The TL2-A comes equipped with a switchable relay. This relay will toggle whenever the temperature crosses the set point threshold
set through the user menu. The provided screw-terminal header can be wired in two configurations:
• Terminals 1-2 are normally closed (NC) and will turn OFF (open) when the relay is set.
• Terminals 1-3 are normally open (NO) and will turn ON (close) when the relay is set.
Figure 1: Relay screw-terminal header
The relay function can be implemented from the user menu. For the relay function to be active, the user must select ‘EnAbLE’ from
the relay menu. Also from the relay menu, the temperature set point can be configured. (‘rELAY’ -> ‘SEt’) Use the Up & Down
buttons to select the desired setpoint. Use the select button to select each digit to modify. Ensure this setpoint is correct before
selecting the ‘EnAbLE’ function. The relay set point will remain the same regardless of the unit selection (C, F or K).
Once the relay has been toggled, it will remain in that state until the user disables it using the Select button. If desired, the user will
then need to re-enable the relay from the user menu.
Over time, a sensor may experience a shift in temperature at different ranges of the instrument. Zero correction allows the user to
make an adjustment to the current channel temperature display that effectively corrects for this shift. This offset or ‘zero correction’
is applied to the entire effective range of the instrument. Each channel has its own zero correction.
NOTE: A zero correction made at a particular point may not correct the temperature at another point. It is recommended that the
instrument be fully calibrated annually to maintain acceptable temperature tolerances for the range of the instrument.
To view the zero correction for the current channel, select the “Corr” option from the menu.
To set a zero correction:
1. Use the function and arrow buttons to navigate to the “Corr” item in the menu. Press and hold the power button. While holding the
power button, press the function button until the "Adj" menu item appears. Release both buttons.
2. Press and release the function button again. The double arrows will flash indicating that you are in zero correction mode. Use the
up and down arrows to adjust the display temperature as desired. The adjustment will only apply to the channel on the display.
3. When you've adjusted the display, press and release the function button. Use the arrow buttons and the function button to save or
cancel the zero correction and exit zero correction mode.
4. To view the zero correction, use the function button and arrow buttons to navigate to the Corr menu item as before. This time,
press and release the function button. The display will show the zero correction for the selected channel and then return to
temperature display mode.
5. To save the zero correction, the Power Button must be pressed. This will set the unit to ‘Standby’.
PC Software setup:
There is a PC driver file that is needed to connect the TL2-A to your PC. We have included a disc with the
driver software for Windows XP, Windows Server 2003, Windows Vista, Windows Server 2008, Windows 7,
Windows Server 2008 R2. Other operating systems are supported and driver software is available.
Use the Windows Add Hardware Wizard and the driver files to set up a USB Com port to connect to the TL2A. Connect using HyperTerminal or other software that can connect interface with the selected Com port.
** The USB driver software web site is (http://www.ftdichip.com/Drivers/VCP.htm)
.Note These instructions are taken from a document that is available online from Future Technology Devices International.
(http://www.ftdichip.com/Support/Documents/AppNotes/AN_104_FTDI_Drivers_Installation_Guide_for_WindowsXP%28FT_000093%29
.pdf). Detailed technical information is available from that website. Technical assistance with installing the driver software is available from
I
nstructions to install the TL2-A PC driver (Windows XP shown)
International driver) installs a Virtual Com Port (VCP) on your PC. The steps to install this driver are:
1. Place the USB flash drive provided with your TL2-A in the USB port of the PC.
2. Connect the TL2-A to the PC with the provided USB cable.
3. Make sure the TL2-A is plugged into the wall with the AC power cord.
4. Make sure the TL2-A is powered on.
5. A Found New Hardware Wizard window should appear.
The driver for the TL2-A PC (Future Technology Devices
8. On the next screen click "Install from a list or specific location (Advanced)" and click "Next >".
9. On the next screen select "Search for the best driver in these locations.", Check the "Search removable media (floppy, CD-ROM…):", and
then click "Next”.
The Sensor Data software included with the TL2-A USB flash drive allows the user to input sensor
coefficients. Install the software by browsing to the SensorData folder on the drive and run the setup.exe file.
1. Connect the TL2-A to the PC with a USB cable and power it on.
Dusen or ITS
-
90
ITS-90 only
NOTE: See ‘PC Software setup’ section to install the COM port drivers.
2. Make sure the TL2-A has been set to ‘Update’ mode from the user menu. Press the ‘Select’ button and
then up or down until you see ‘UPdAtE’. Press the select button again to allow PC connection. The
display will flash when connecting.
3. From the main software screen select swap PRT.
4. Select the appropriate COM port and set baud rate to 256000.
NOTE: See ‘Finding Com Port Section’ if you don’t know which com port to select.
5. Select start from the Connection screen. A window will open a message stating “TL2-A data
successfully transferred”. Select ‘OK’ to clear this window. Once completed the Calibration Data
screen will appear.
6. Enter the unit serial number, the probe serial number and calibration date.
7. Enter the curve data using the coefficient information provided with the PRT connected to each
channel.
8. After all the data has been entered for both channels select OK to upload the data to the TL2-A. The
Connection screen will display again. Select start to load the sensor data. A window will open a
message stating “TL2-A data successfully transferred”. Select ‘OK’ to clear this window. The TL2-A
will exit Update mode and resume temperature display.
You can use Hyperterminal to receive temperature data from the TL2-A.
Note: Hyperterminal is no longer included with Windows 7 systems. A copy of Hyperterminal is included with
the TL2-A software USB flash drive and can be run from that location.
The TL2-A will output the temperature for both channels. There are 3 columns in the interface display. The
first column is a date and timestamp. The second column is the first channel output and the third column is the
second channel output. The displayed units are the units selected from the faceplate (F, C, K or Ohms). The
displayed precision is four decimals.
TL2-A HyperTerminal Command Set:
Toggle Prompt (on)
Toggle Prompt (off)
Set Date
Set Time
Set Rate
Poll Temperature
Toggle Checksum
(on)
Toggle Checksum
(off)
Send Version
Note 1: The carriage return, <cr>, is hex digit 0x0D. The line feed (new line), <lf>, is hex digit 0x0A.
Note 2: All input commands are echoed back to the sender.
<cr>
<cr>
D YY-MM-DD<cr>
T HH:MM:SS<cr>
R [0,1,10,30,60,3600,poll]<cr>
?<cr>
C<cr>
C<cr>
V<cr>
Stop sending temperatures, show a ">", and wait
for a command.
Resume sending temperatures.
Set the TL2-A's real time clock's date.
Set the TL2-A's real time clock's time.
Set the rate the TL2-A sends temperatures (in
seconds).
Send a temperature now.
Add a checksum hash to the temperature.
Don't send a checksum hash with the
temperature.
Show the firmware version for the TL2-A (there are
2 boards and each has a firmware version).
Remarks: No temperatures are sent while prompt mode is on. Temperature sending will resume once the
prompt is toggled off. If other functions are used while the prompt is on, the responses are preceded
by a prompt <cr><lf>>.
Set Date
Command: D YY-MM-DD<cr>
Response (success): New Date is: YYYY-MM-DD<cr><lf> Response (failure): Error – use format "> D(ate) (YY)YY-MM-DD"<cr><lf>
Remarks: A lower case "d" or the word "Date" can be used.
Set Time
Command: T HH:MM:SS<cr>
Response (success): New Time is: HH:MM:SS<cr><lf> Response (failure): Error – use format "> T(ime) HH:MM:SS"<cr><lf> Remarks: A lower case "t" or the word "Time" can be used.
Set Rate
Command: R [0,1,30,60,3600,poll]<cr> (e.g. R 1<cr>, R 30<cr>, r poll<cr>
Response (success): Send Rate: X sec.<cr><lf>
Send Rate: Poll (enter ? For a temp.)<cr><lf>
Response (failure): Rate Format">R X" (X =1,10,30,60,3600,Poll)<cr><lf>.
Remarks: The brackets contain the options, separated by commas. (e.g. R 10<cr>, R poll<cr>). A lower case r
or the word Rate (or rate) can be used. Entering 0 is the same as entering poll.
Poll Temperature
Command: ?<cr>Response (success):
Response (failure): The TL2-A sends the temperature now, regardless of the temperature send rate. (When in
polling mode, the TL2-A only sends temperatures when it is polled.) Polling for a temperature resets
the timer for the 10-3600 second rate (the next temperature will be sent <rate> seconds from the
polled temperature.
Remarks: Polling can be used when not in polling mode to synchronize the send rate. It can be used in polling
mode to get the temperature exactly when it is needed.
Version
Command: V<cr>
Response (success): A lower case v can be used. Sends the TL2-A firmware version for the main and accessory
boards.
Response (failure): none
Remarks: The version information is useful when verifying that the correct firmware has been loaded on the
You can find the Com Port that was installed by clicking on "start", choosing Settings, choosing "Control Panel" and then opening the
"System" application.
Click the "Hardware" tab and then click the "Device Manager" button.