/8 DIN Total Dissolved Solids
Controller Transmitter
ppm/ppt/ºC
PREFACE
This manual serves to explain the use of the TDS 200 controller/transmitter.
This manual functions in two ways: first, as a step-by-step guide to help you
operate the meter; second, it serves as a handy reference guide.
This manual is written to cover as many anticipated applications of the TDS
200 controller/transmitter as possible. If there are doubts in the use of the TDS
200 controller/transmitter, do not hesitate to contact the nearest Authorized
Distributor.
Thermo Scientific will not accept any responsibility for damage or malfunction
to the controller/transmitter caused by improper use of the instrument.
The information presented in this manual is subjected to change without notice
as improvements are made, and does not represent a commitment on the part
of Thermo Scientific.
8.8 VIEWING TDS / TEMPERATURE ELECTRODE DATA – P6.0 22
8.9 SET TDS FACTOR – P7.0 23
8.10 CONTROLLER RESET – P8.0 23
9 RELAYS 25
10 TRANSMITTER FUNCTION 25
11 SPECIFICATIONS 26
12 ACCESSORIES 27
13 GENERAL INFORMATION 30
1 INTRODUCTION
Thank you for purchasing a ⅛ DIN TDS 200 Controller. This controller is part of a series
of quality process controllers available from Thermo Scientific / Eutech Instruments.
These sturdy, economical TDS controllers are designed with the features and reliability of
a much more expensive instrument.
Your controller includes:
• Removable terminal blocks for easy connections;
• Two mounting brackets for easy panel mounting;
Some features of this controller are:
• Two set point, two SPDT relay operation
• Scrolling, 14-segment LED guides user easily through setup functions
• Reliable power supply from 85 to 260 V AC, 50/60 Hz or DC withstands voltage
fluctuations
• Push-button operation from the front panel
• Single-point calibration for each individual range
• Adjustable hysteresis band prevents rapid contact switching around set-point
• Selectable automatic or manual temperature compensation
• Two-level password protection
• Removable terminal strips for quick and easy connections
• Built-in memory backup retains setup even if power fails, and lets you configure
unit before installation
• Isolated 4-20 mA output for remote monitoring or hard copy recording
1
2 SAFETY INFORMATION
This Controller/Transmitter shall be installed and operated only in the manner specified in
the Instruction manual. Only skilled, trained or authorized person should carry out
installation, setup and operation of the instrument.
Before powering up the unit, make sure that power source it is connected to, is as
specified in the top label. Failure to do so may result in a permanent damage to the unit.
The unit has live and exposed parts inside. If it has to be opened, make sure that the
power to the unit is off and disconnected.
The unit is Fuse protected. In the event the fuse has to be replaced, use only those as
specified in the manual.
The degree of protection against electric shock will be achieved only by
observance of the corresponding installation rules.
2
3 OVERVIEW
3.1 Front Panel
The front panel consists of a 4-digit LED display, 8 LED annunciators and 4 keys.
Annunciators
1. REL 1 Displayed when Relay 1 is activated
2. REL 2 Displayed when Relay 2 is activated
3. MEAS Displayed in measurement mode
4. CAL Displayed in calibration mode
5. SETUP Displayed in setup mode
6. ppm Displayed when measurement range in parts-per-million
7. oC Unit of the displayed parameter (temperature)
8. ppt Displayed when measurement range in parts-per-thousands
Keys
9. MODE
10. ENTER
11. ▲(increment)
11. ▼(decrement)
12. LED Display 14 segment LED display
Toggle between TDS and temperature display
Confirm changes or to enter into further levels of sub-menu
Use during calibration and setup modes to increment values.
▲/▼keys together to escape to Measurement mode
Press
Use during calibration and setup modes to decrease values.
▲/▼keys together to escape to Measurement mode
Press
8.8.8.8.
3
3.2 Back Panel
The back panel consists of three different connectors that can be used with removable
terminal blocks (included):
19 18 17 16 15 14 13
12 11 10 9 8 7 6 5 4 3 2 1
1. VAC live wire
2. VAC neutral wire
3. VAC protective ground wire
4. unused
5. Relay 2 deactivated position (normally closed)
6. Relay 2 center pole
7. Relay 2 activated position (normally open)
8. Relay 1 deactivated position (normally closed)
9. Relay 1 center pole
10. Relay 1 activated position (normally open)
11. 4-20 mA connection, negative
12. 4-20 mA connection, positive
13.
Pt 100 connection: sense (jumper to terminal 14 if using 2-wire RTD)
14. Pt 100 connection: input
15. Pt 100 connection: ground
16. TDS positive terminal
17. TDS negative terminal
18. unused
19. unused
4
3.3 Wiring
Caution: Ensure electrical mains are disconnected before proceeding.
1. Connect the power supply to the three-pin terminal block
• VAC live wire = 1
• VAC neutral wire = 2
• VAC protective ground wire = 3
TDS 200 controller accepts voltages from 85 to 260 VAC, 50/60 Hz or DC.
2. Connect the Pt 100 leads to terminals 13 to 15 of the seven-pin terminal block. Either
wire can be connected to either terminal. Terminals 13 and 14 must be shunted
unless using a 3-wire RTD.
NOTE: TDS 200 is factory set for Automatic temperature compensation. MTC can be
selected in Program P4.0.
3. Power on the controller. The display automatically shows the TDS reading, the ppm
and ‘MEAS’ annunciators lights.
NOTE: In the event Pt 100 is not connected or the connection is broken in the ATC
mode, the display flashes in TDS mode and display OR in temperature mode.
5
3.4 Panel-mounting the controller
a
d
The supplied mounting hardware allows surface mounting to all panels and protective
enclosures. Mounting cut-out size is 91 x 45 mm.
To attach the mounting to the controller:
1. Align the catch to the side of the controller, and insert threaded rods through catch.
2. Screw the threaded-rod through the catch in a clock-wise direction. Tighten until the
catch holds the controller firmly against the back of the panel or protective housing.
Repeat on the other side.
h
c
t
a
C
h
c
t
a
C
.
2
0
TOP VIEW
e
d
a
e
r
h
T
0
0
1
.
x
o
r
p
p
a
d
o
r
d
e
d
a
e
r
h
T
1
5
4
a
o
p
p
r
x
.
p
l
l
a
W
SIDE VIEW
6
4 MEASUREMENT MODE
Press MODE to toggle between:
• TDS measurement mode and
• Temperature measurement mode
Total Dissolved Solids Measurement Mode
The controller displays the selected TDS range number, R X (X ranges from 1 to 8), for 2
seconds before displaying the TDS measurement.
Press MODE key once to view the temperature measurement. The display shows ATC
(Automatic Temperature Compensation) or MTC (Manual Temperature Compensation),
then the current measured temperature (for ATC) or the set temperature (for MTC). The
o
C annunciator lights when you are measuring temperature.
NOTE: After pressing the MODE key to display Temperature, if there is no further keypress, the Controller will automatically revert to TDS Measurement mode after about 30
seconds.
See Setup program P4.0 for further instructions.
7
Loading...
+ 25 hidden pages
You need points to download manuals.