For Research Use Only. Not for use in diagnostic procedures.
Thermo Fisher Scientific | 6055 Sunol Blvd | Pleasanton, CA 94566 | USA
For descriptions of symbols on product labels or product documents, go to thermofisher.com/symbols-definition.
The information in this guide is subject to change without notice.
DISCLAIMER: T
O THE EXTENT ALLOWED BY LAW, THERMO FISHER SCIENTIFIC INC. AND/OR ITS AFFILIATE(S) WILL NOT BE
LIABLE FOR SPECIAL, INCIDENTAL, INDIRECT, PUNITIVE, MULTIPLE, OR CONSEQUENTIAL DAMAGES IN CONNECTION WITH OR
ARISING FROM THIS DOCUMENT, INCLUDING YOUR USE OF IT.
Revision history: Pub. No. MAN0019745
RevisionDateDescription
A.027 January 2021New document.
Important Licensing Information: These pr
oducts may be covered by one or more Limited Use Label Licenses. By use of these
products, you accept the terms and conditions of all applicable Limited Use Label Licenses.
TRADEMARKS: All trademarks are the property of Thermo Fisher Scientific and its subsidiaries unless otherwise specified. Tween is a
registered trademark of Croda Americas, Inc.
Contents and storage .................................................................. 5
■
Required materials not supplied ......................................................... 5
■
IMPORTANT! Before using this product, read and understand the information in the “Safety” appendix
in this document.
Product description
The TaqCheck™ SARS-CoV-2 Fast PCR Assay is a multiplex real‑time RT‑PCR assay for the detection
of SARS-CoV-2 viral RNA in human raw saliva samples. The assay has a multi-target design that
compensates for emerging SARS-CoV-2 variants and mutations to provide confidence in results. The
assay contains primer and probe sets specific to the following targets:
Product information
Table 1 Assay targets, dyes, and quenchers
TargetDyeQuencher
SARS-CoV-2 N gene
VIC™ dyeQSY™ quencher
SARS-CoV-2 S gene
Human RNase P RPP30 gene
[1]
S
erves as an internal positive control to monitor sample quality.
The assay requires the following components:
•
TaqCheck™ SARS‑CoV‑2 Control—RNA control that contains SARS-CoV-2 N protein and S protein
target regions
•
TaqCheck™ SARS‑CoV‑2 Control Dilution Buer—Dilution buer for the control
•
TaqPath™ 1-Step RT-qPCR Master Mix, CG
For catalog numbers and storage conditions, see “Contents and storage”.
IMPORTANT! It is the r
Assay to design and validate their own experimental design and analysis parameters.
[1]
esponsibility of the laboratories using the TaqCheck™ SARS-CoV-2 Fast PCR
FAM™ dyeQSY™ quencher
4
TaqCheck™ SARS-CoV-2 Fast PCR Assay User Guide
Chapter 1 Product information
Contents and storage
Contents and storage
The items listed in the following table are required for the TaqCheck™ SARS-CoV-2 Fast PCR Assay.
The items listed are sucient for 1,200 reactions.
Kit or productCat. No.AmountStorage
TaqCheck™ SARS-CoV-2 Fast PCR AssayA47693690 µL–30°C to –10°C
Unless otherwise indicated, all materials are available through thermofisher.com. "MLS" indicates that
the material is available from fisherscientific.com or another major laboratory supplier.
Catalog numbers that appear as links open the web pages for those products.
ItemSource
Real-time PCR instrument and software
An Applied Biosystems™ real-time PCR instrument compatible with
the dyes listed in Table 1 on page 4.
The assay was tested with the following instruments:
•
Applied Biosystems™ 7500 Fast Real‑Time PCR Instrument
96‑well plate (for preparing saliva samples, not for RT-PCR)AB0796, or equivalent
terile tube with leak‑proof, screw‑top lid for
[3]
eservative.
6
•
AM9932 (1 x 1,000 mL)
One of the following, or equivalent:
•
AM12500
•
339650
•
14-959-49B (fisherscientific.com)
TaqCheck™ SARS-CoV-2 Fast PCR Assay User Guide
Chapter 1 P
Required materials not supplied
roduct information
(continued)
ItemSource
•
4346906 (with barcode)
•
MicroAmp™ Fast Optical 96-Well Reaction Plate, 0.1 mL
MicroAmp™ Optical 96-Well Reaction Plate, 0.2 mL
MicroAmp™ Optical 384-Well Reaction Plate
4366932 (with barcode)
•
4346907 (without barcode)
•
4306737 (with barcode)
•
4326659 (with barcode)
•
N8010560 (without barcode)
•
4316813 (without barcode)
•
4309849 (with barcode)
•
4326270 (with barcode)
•
4343814 (with barcode)
•
4343370 (without barcode)
MicroAmp™ Clear Adhesive Film4306311
1
MicroAmp™ Optical Adhesive Film4311971, 4360954
MicroAmp™ Adhesive Film Applicator4333183
Nonstick, RNase-free microcentrifuge tubes (1.5 mL and 2.0 mL)thermofisher.com/microtubes
DNase and RNase‑free tubes for mixing reagents (capable of mixing
5 mL and 50 mL)
Nunc™ 1.8‑mL Ext
ernally-Threaded Universal Tubes374502
thermofisher.com
Nalgene™ General Long-Term Storage Cryogenic Tubes, 0.2 mL5000-0020
Nunc™ Biobanking and Cell Culture Cryogenic Tubes, 4.5 mL337516
•
Sterilin™ Certified Universal Containers – RNase, DNase, human
DNA and Pyrogen Free, 30 mL
[1]
U
se of QuantStudio™ Design and Analysis Software v2.5 is recommended, but not required. It is the responsibility of the laboratories using the
assay to design and validate their own experimental design and analysis parameters.
[2]
Follow the guidelines provided by the manufacturer.
[3]
The use of the tubes listed in the table is recommended, but not required. Laboratories are responsible for validating their sample collection
and preparation procedures for use with the assay.
30APPRN (Unlabelled)
•
30BPPRN (Graduated label)
TaqCheck™ SARS-CoV-2 Fast PCR Assay User Guide
7
2
General laboratory recommendations .................................................... 8
■
Guidelines for sample collection and storage ............................................. 8
Set up and run the real‑time PCR ....................................................... 15
■
Analyze data ......................................................................... 16
■
General laboratory recommendations
•
Implement standard operating procedures in your laboratory to prevent contamination, such as the
following:
–
Frequent glove changes
–
Frequent decontamination of surfaces, equipment, and pipettes with 10% bleach or
decontamination solution, followed by 70% ethanol
–
Use of ultraviolet light during biosafety cabinet decontamination (when available)
•
Saliva samples should always be treated as if infectious and/or biohazardous in accordance with
safe laboratory procedures.
•
To prevent degradation, keep master mixes, assays, and controls on ice or in cold blocks while in
use. Limit freeze-thaw cycles.
•
Aliquot reagents to prevent stock contamination and reduce the number of freeze-thaw cycles.
•
To ensure reliable performance of the real‑time PCR instrument, perform preventive maintenance
according to the instructions provided by the manufacturer in the instrument documentation (see
“Related documentation” on page 22).
Methods
Guidelines for sample collection and storage
•
Collect saliva sample in a collection device with a leak‑proof, screw‑top lid.
IMPORTANT! Do not collect saliva using a device that contains preservative solution.
•
Collect a minimum of 1 mL saliva.
8
TaqCheck™ SARS-CoV-2 Fast PCR Assay User Guide