Thermo Fisher Scientific’s SX40 dedicated Doppler flowmeter generates
two independent ultrasonic signals at different frequencies. By correlating
these frequencies, the instrument automatically identifies and eliminates
noise errors from sources such as variable frequency drives.
In addition, operation of the instrument is enhanced by an Expert System
that allows the flowmeter to automatically ‘‘learn’’ the application
parameters. As a result, the flowmeter can be easily commissioned in a
fraction of the time required to configure competitive ultrasonic
flowmeters.
Ordering
Information
Refer to the following list of part numbers to order spare parts for your
instrument.
Table 1–1.
Part Number
Description
22767-0001 Poly CD, specification sheet, meter installation video CD
1-0563-006 User guide
1-0561-005 HydraScan software manual
01008-0006 RS232 cable, standard serial, 6-ft, female to male, DB9
10241-0003 Hazardous area relay, 3 A, SPDT
10261-0013 Fuse, 1 A, time delay, AC input
10396-0194 Power cord connector, CGB, for 6 ft. AC power cord
10605-0001 Pipe strap, SS, 32-inch with 3/8 SS screw
10605-0003 Pipe strap, SS, 67-inch with 3/8 SS screw
10705-0003 Nut driver, 3/8-inch
10808-0002 Coupling compound, silicone RTV108, 2.8 oz.
10823-0005 Coupling compound, Sil-Glyde®, 4 oz.
20192-0001 SS tag
22061-0001 Heater assembly suitable for -40°F to 140°F (-40°C to 60°C) operation,
100- -- 120 Vac
SX40 User Guide
1-2
Thermo Fisher Scientific
Product Overview
Specifications
Part Number
22061-0002 Heater assembly suitable for -40°F to 140°F (-40°C to 60°C) operation,
22204-0003 Mounting feet, set of 4
22262-0006 PCB assembly, 4- -- 20 mA output (applicable with firmware v. 2.03)
Fluids: Liquids containing particulate or entrained gas bubbles
Pipe Size: 0.5- -- 200 in. (12- -- 5000 mm)
Physical Transmitter: NEMA 4X (IP65), flame retardant, fiberglass reinforced
polyester
Transducers
Two encapsulated dual frequency sensor heads suitable for
submersible/underground service
Encased in stainless steel shrouds, with stainless steel straps and quick
clamps
SX40 User Guide
Cable length: 30 ft. (9 m)
Weight: 12 lb. (5.4 kg)
Dimensions: Refer to Figure 1- -- 1
Thermo Fisher Scientific
1-3
Product Overview
Specifications
Figure 1–1.
Functional Outputs: 4- -- 20 mA (into 750 ohms), 12-bit, 5 kV opto-isolated; loop or
self-powered
Optional Relay: 5 A, SPDT, programmable relay; maximum of 4 Power
Supply
DC Power: 11- -- 28 Vdc 1.0A Max
AC Power: 90- -- 132 Vac or 190- -- 250 Vac (switch selectable), 50/60 Hz 0.14A Max
Keypad: 19-key with tactile action
Display: Backlit, 240 x 60 dot, high resolution graphics display
Data Logger: 90000 points, programmable in intervals of 30 s, 1, 5, 15, 30, and
60 min
Programming: Via HydraScan software or integral keypad
Serial Interface:
RS232
SX40 User Guide
1-4
Thermo Fisher Scientific
Product Overview
Specifications
Approvals
Temperature Range
Transducers: -40°F to 250°F (-40°C to 121°C); coax cable limited to
176°F (80°C)
Electronics: -20°F to 140°F (-29°C to 60°C)
With integral heater: -40°F to 140°F (-40°C to 60°C)
Hazardous Area Certification
Optional: CSA Class I, Div. 2
Optional: CSA Class I, Div. 2 provides Intrinsically Safe Class I,
Div. 1; Class II, Div. 1
Optional: Non-hazardous area provide Intrinsically Safe Class I, Div. 1;
Class II, Div. 1
Conditions of Acceptability: The SX40 must not be powered by more
than 120Vac or be connected to any equipment which generates more
than 120Vac when installed in a Class, Division 2 hazardous area.
SX40 User Guide
Thermo Fisher Scientific
2-1
Chapter 2
Installation & Wiring
Inssttaalllliinngg the
Transducers
1. Select an installation site:
Keep the following questions in mind when choosing a proper
installation site:
●
Is the chosen section of pipe always full of liquid?
●
Are there at least 5 pipe diameters upstream and 3 pipe diameters
downstream from any directional changes, pipe joints, or
narrowing/widening of the pipe?
Figure 2–1. Recommended installation locations
Conditions at locations A, B, C, D, and E can interfere with the
transmission of the ultrasonic wave and yield inaccurate or unreliable
flow readings:
●
A: Pipe may not be full
●
B: Down flow
●
C: Too close to elbow
●
D: Air collects at the top of the horizontal pipe
●
E: Sediment collects at the bottom of the horizontal pipe
Note Selecting the proper installation location is essential to flow
measurement reliability. Each application is unique and may require a
variation of installation locations. ▲
SX40 User Guide
2-2
Thermo Fisher Scientific
Installation & Wiring
Installing the Transducers
2. Select a transducer orientation:
Refer to Figure 2- -- 2. If your application allows, install the transducers at
the 3 and 9 o’clock locations. Ensure the transducers are parallel to
each other along the pipe circumference.
Figure 2–2. Recommended transducer orientations
3. Attach the transducers to the pipe strap:
a. Use a screwdriver to loosen the mounting screw and slide the lock
back on each transducer.
Figure 2–3.
SX40 User Guide
Thermo Fisher Scientific
2-3
Installation & Wiring
Installing the Transducers
b. Place the pipe strap across the mounting, and slide the lock back
into place.
Figure 2–4.
c. Tighten the screws.
Figure 2–5.
4. Mount the transducers, ensuring the transducers face the same
direction (as shown in Figure 2- -- 6):
a. Wrap the pipe strap around the pipe, and slide the end of the pipe
strap through the tension nut assembly.
b. Pull the strap tightly, and clamp the tension nut down to secure the
strap to the pipe.
c. Loosen the mounting screw of a transducer, and move the
transducer to the 3 o’clock position on the pipe. Tighten the
mounting screw to secure the transducer.
d. Loosen the mounting screw of the second transducer, and move the
transducer to the 9 o’clock position. Tighten the mounting screw
to secure the transducer.
e. Remove the strap from the pipe, and apply sonic coupling
compound to each transducer surface.
f. Reposition the strap on the pipe, placing the transducers in the
proper positions.
SX40 User Guide
2-4
Thermo Fisher Scientific
Installation & Wiring
Mounting the Enclosure
M
oouunnttiinngg the
Enclosure
g. Tighten the tension nut assembly with a 3/8-inch hex nut driver to
secure the strap.
Note Increasing the pipe strap tension improves signal transmission. ▲
Figure 2–6.
Warning To prevent the possibility of electrical shock and/or damage to
the meter, do not mount the enclosure where it can be submerged or
partially submerged. ▲
SX40 User Guide
1. Locate a sturdy, vertical mounting surface.
2. Remove the meter access cover, and locate the 4 mounting holes (1 in
each corner of the cover).
3. Attach the enclosure to the wall using the mounting screws.
4. Replace the meter access cover.
Note An optional mounting ear package is also available. ▲
Thermo Fisher Scientific
2-5
Wiring
ACC
Power
Installation & Wiring
Warning Prevent possible electrical shock and/or damage to the meter:
Disconnect power to the meter PRIOR TO wiring. ▲
Caution Connecting AC power to a DC power meter may damage the
meter. ▲
Note CE certifications require that a switch or circuit breaker be installed
in close proximity to the instrument and be labeled as the instrument
disconnect. ▲
Refer to the following instructions and Figure 2- -- 7 when wiring the
instrument for AC power.
Wiring
1. Remove the meter access cover.
2. Locate the AC input voltage selector, and turn the switch to 110 or 220
V to match the power line voltage.
3. Locate the L1, L2/N, and GND connections within power input block.
4. Turn the terminal screws on top of the input block counterclockwise to
fully open the wire connections.
5. Make the following connections:
a. Ground wire to GND (CE certification requires less than 1 ohm
between terminal and ground)
b. Hot wire to L1
c. If the input voltage is 110 V, neutral wire to L2/N
d. If the input voltage is 220 V, second hot wire to L2/N
6. Turn the terminal screws clockwise to close them connections.
7. Ensure the wires are securely connected to the input block, and replace
the meter access cover.
SX40 User Guide
2-6
Thermo Fisher Scientific
Installation & Wiring
Wiring
Figure 2–7. AC power connections
DC Power
Later versions of the instrument (those with P/N 22302-0002 Rev. G and
later) have DC operation built into the instrument. Refer to the following
instructions and Figure 2- -- 8.
1. Remove the meter access cover.
2. Locate the power input block, and turn the terminal screws on top of
the input block counterclockwise to fully open the wire connections.
3. Make the following connections:
a. +24VDC to P2 +24
b. 24 VDC return to either of the P2 GND connections
4. Turn the terminal screws clockwise to close them connections.