Thermo Fisher Scientific PrepFiler User Manual

PrepFiler™ Automated Forensic DNA Extraction Kit
USER BULLETIN
Automated DNA Purification on the HID EVOlution™ Systems
for use with: HID EVOlution™—Extraction System HID EVOlution™—Combination System
Publication Number MAN0019298
Revision A.0
For Research, Forensic, or Paternity Use Only. Not for use in diagnostic procedures.
Life Technologies Ltd | 7 Kingsland Grange | Woolston, Warrington WA1 4SR | United Kingdom For descriptions of symbols on product labels or product documents, go to thermofisher.com/symbols-definition.
The information in this guide is subject to change without notice.
DISCLAIMER: TO THE EXTENT ALLOWED BY LAW, THERMO FISHER SCIENTIFIC INC. AND/OR ITS AFFILIATE(S) WILL NOT BE LIABLE FOR SPECIAL, INCIDENTAL, INDIRECT, PUNITIVE, MULTIPLE, OR CONSEQUENTIAL DAMAGES IN CONNECTION WITH OR ARISING FROM THIS DOCUMENT, INCLUDING YOUR USE OF IT.
Revision history: Pub. No. MAN0019298
Revision Date Description
A.0
22 March 2021 New document.
Important Licensing Information: These products may be covered by one or more Limited Use Label Licenses. By use of these products, you accept the terms and conditions of all applicable Limited Use Label Licenses.
Trademarks: All trademarks are the property of Thermo Fisher Scientific and its subsidiaries unless otherwise specified. Freedom EVO, EVOware, Te-Shake, and HID EVOlution are trademarks of Tecan Group AG. Microsoft and Windows are trademarks of Microsoft Corporation.
©2021 Thermo Fisher Scientific Inc. All rights reserved.

Contents

CHAPTER 1 Product information .................................................. 6
Product description ............................................................. 6
About the HID EVOlution™ systems ................................................ 6
Purification run times ........................................................ 7
System components ........................................................ 8
Plate/tube configurations .................................................... 8
Contents and storage ............................................................ 8
Required materials not supplied ................................................... 9
CHAPTER 2 Prepare for the automated purification run ...................... 11
Perform lysis .................................................................. 11
Before first use: Prepare the wash buers ........................................ 11
Before each use: Prepare the magnetic particles .................................. 11
CHAPTER 3 Set up the robotic workstation .................................... 12
For more information ........................................................... 12
Workflow ..................................................................... 13
Perform routine maintenance .................................................... 14
Prepare the system liquid carboy ............................................ 14
Empty the waste carboy .................................................... 14
Tighten the DiTi adapter gold cones .......................................... 14
Run maintenance scripts ................................................... 15
(HID EVOlution™—Combination System only) Set up the carriers and labware .......... 16
Set up the disposable pipette tips ................................................ 19
Terms for pipette tips used on the robotic workstation ......................... 19
Fill DiTi carriers and racks ................................................... 20
Set up reagents on the workstation ............................................... 20
Procedural guidelines ...................................................... 20
Set up the reagents ........................................................ 21
Set up lysate, processing, and elution plates and/or tubes .......................... 23
Select a plate/tube configuration ............................................ 23
Set up the PrepFiler™ Processing Plate ....................................... 24
(If needed) Place barcodes .................................................. 25
PrepFiler Bulletin
Automated Forensic DNA Extraction Kit: Automated DNA Purification on the HID EVOlution™ Systems User
3
Contents
Set up lysate and/or eluate plates ............................................ 25
Set up lysate and/or eluate tubes ............................................ 27
Workstation layouts ............................................................ 30
CHAPTER 4 Perform the automated DNA purification run .................... 34
For more information ........................................................... 34
Workflow ..................................................................... 35
Before you begin ............................................................... 36
Set up and run a script ......................................................... 36
About script files .......................................................... 36
Select a script ............................................................. 36
Set up sample and reagent information ....................................... 38
Confirm workstation setup and start the run .................................. 44
(If needed) Re-cap the magnetic particles tubes ............................... 46
Record file information and exit the script ..................................... 47
Complete the run .............................................................. 48
View the qPCR/STR Sample Input and Report files ................................ 49
About the files ............................................................ 49
View the files .............................................................. 50
CHAPTER 5 Experiments and results ........................................... 51
Validation of the PrepFiler™ Automated Forensic DNA Extraction Kit on the
HID EVOlution™—Extraction System ............................................ 52
Overview of experiments and results ......................................... 52
Materials and methods ..................................................... 52
Precision studies (SWGDAM standard 2.9) .................................... 54
Reproducibility studies (SWGDAM standard 2.5) ............................... 59
Correlation studies ......................................................... 61
Cross-contamination studies (SWGDAM standard 3.6) ......................... 64
STR studies ............................................................... 65
Verification studies for remaining scripts ..................................... 69
Additional cross-contamination studies ...................................... 71
Conclusions .............................................................. 72
Validation of PrepFiler™ Wash Buer B and the related modifications to the
workstation layout and scripts ................................................. 73
Overview of experiments and results ......................................... 73
Materials and methods ..................................................... 74
Sample lysis method ....................................................... 77
Script validation ........................................................... 78
Precision and sensitivity studies ............................................. 80
Cross-contamination studies ................................................ 86
Case-type sample studies .................................................. 89
Comparative analysis studies ............................................... 93
Conclusions .............................................................. 99
4
PrepFiler™ Automated Forensic DNA Extraction Kit: Automated DNA Purification on the HID EVOlution™ Systems User
Bulletin
APPENDIX A Troubleshooting .................................................. 100
APPENDIX B Plate setup requirements for <96 samples .................... 104
Place samples in the plate (<96 samples) ........................................ 104
APPENDIX C (One-time procedure) Document TeShake™ plate
adapter temperatures ............................................................ 107
Required materials not supplied ................................................ 108
Measure the plate and adapter temperatures .................................... 108
APPENDIX D Safety ............................................................. 111
Safety information for instruments not manufactured by Thermo Fisher Scientific ..... 112
Chemical safety .............................................................. 112
Biological hazard safety ....................................................... 114
Contents
Documentation and support ...................................................... 115
Related documentation ........................................................ 115
Customer and technical support ................................................ 115
Limited product warranty ...................................................... 115
PrepFiler™ Automated Forensic DNA Extraction Kit: Automated DNA Purification on the HID EVOlution™ Systems User Bulletin
5
1
Product description .................................................................... 6
About the HID EVOlution™ systems ...................................................... 6
Contents and storage .................................................................. 8
Required materials not supplied ......................................................... 9
IMPORTANT! Before using this product, read and understand the safety information in the
manufacturer's documentation.
IMPORTANT! Before using this product, read and understand the information in the “Safety” appendix
in this document.

Product description

Product information
The PrepFiler™ Automated Forensic DNA Extraction Kit is part of an integrated solution that includes proven PrepFiler™ reagents for use in semi-automated, high-throughput workflows to provide reliable, simplified, and time-saving DNA extraction and purification.
The kit uses magnetic particles with an optimized multi-component surface chemistry to deliver robust and reliable DNA yield from tested routine forensic sample types, including:
Body fluids (blood, saliva, semen)
Stains and swabs of body fluids
Hair roots
Touch/trace samples

About the HID EVOlution™ systems

The HID EVOlution™—Extraction System and HID EVOlution™—Combination System perform automated DNA purification using a Freedom EVO™ robotic workstation with the PrepFiler™ Automated Forensic DNA Extraction Kit.
The robotic workstation automates liquid and magnetic particle handling, as shown in Figure 1. The purified DNA is collected in a 96-well plate or 1.5-mL tubes, depending on the Freedom EVOware software script that you select.
6
PrepFiler™ Automated Forensic DNA Extraction Kit: Automated DNA Purification on the HID EVOlution™ Systems User
Bulletin
Chapter 1
About the HID EVOlution™ systems
Product information
1
Figure 1 Automated DNA purification steps performed by the HID EVOlution™—Extraction System in plates (shown) or tubes (not shown)
Lysate is transferred from a spin plate or 1.5mL microfuge tubes into the PrepFiler™ 96-Well Processing Plate (shown) for automated DNA purification.
Purification
The automated purification run time is ~3–4 hours for 96 samples, depending on the type of HID EVOlution™ system and configuration that you are using.
PrepFiler™ Automated Forensic DNA Extraction Kit: Automated DNA Purification on the HID EVOlution™ Systems User Bulletin
run times
7
Chapter 1 Product information
1

Contents and storage

System components

The HID EVOlution™—Extraction System and HID EVOlution™—Combination System consist of the following components.
A Freedom EVO™ 150 robotic workstation or Freedom EVO™ 200 robotic workstation
Freedom EVOware™ software v2.1 SP1 or later, configured with the HID EVOlution™—Extraction System application
Note: Contact Technical Support for more information on verified configurations. See “Customer and technical support” on page 115.
8-channel liquid-handling arm (LiHa)
Robotic Manipulator arm (RoMa)
TeShake™ plate adapter with heating block and adapter
Plate/tube configurations
The HID EVOlution™ systems support the following plate/tube configurations. You can select one configuration per purification run.
Plate-to-plate—Process lysate from a 96-well plate and collect eluate in a 96-well plate
Plate-to-tubes—Process lysate from a 96-well plate and collect eluate in 1.5mL tubes
Tubes-to-tubes—Process lysate from 1.5mL tubes and collect eluate in 1.5mL tubes
Tubes-to-plate—Process lysate from 1.5mL tubes and collect eluate in a 96-well plate
Contents and storage
The PrepFiler™ Automated Forensic DNA Extraction Kit is intended for semi-automated workflows, and contains the reagents required for the following procedures:
Manual sample lysate preparation (DNA extraction)
Automated DNA purification
The kit is sucient for ≤960 samples, depending on the batch size.
Table 1 PrepFiler™ Automated Forensic DNA Extraction Kit (Cat. No. 4463353)
Contents
PrepFiler™ Lysis Buer 1 × 500 mL 18–25°C
PrepFiler™ Magnetic Particles 13 × 1.5 mL
PrepFiler™ Wash Buer A Concentrate 1 × 500 mL
PrepFiler™ Wash Buer B Concentrate 1 × 250 mL
PrepFiler™ Elution Buer 1 × 200 mL
Amount Storage
8
PrepFiler™ Automated Forensic DNA Extraction Kit: Automated DNA Purification on the HID EVOlution™ Systems User
Bulletin

Required materials not supplied

Unless otherwise indicated, all materials are available through thermofisher.com. "MLS" indicates that the material is available from fisherscientific.com or another major laboratory supplier. Catalog numbers that appear as links open the web pages for those products.
Table 2 Reagent preparation
Item Source
Chapter 1 Product information
Required materials not supplied
1
Ethanol (Molecular biology grade; 95% or 190 proof)
Note: Open a new bottle when preparing the PrepFiler™ Wash Buer A and Wash Buer B solutions.
Clean containers to store the prepared Wash Buer A and Wash Buer B solutions; we use:
Nalgene™ Square PETG Media Bottles with Closure: Sterile, Shrink­Wrapped Trays
342020-0500 or 342020-1000
MLS
Table 3 Automated purification
Item Source
Isopropanol (2Propanol, ACS reagent grade, ≥99.5% )
Note: Purchase isopropanol in small bottles and open fresh bottles frequently to maintain a high-quality grade reagent.
(If needed) DNA Suspension Buer (low-TE buer)
Note: DNA Suspension Buer is only needed if you run out of PrepFiler™ Elution Buer.
Magnetic-Ring Stand (96 well) AM10050
MLS
MLS
[1]
Disposable Tips (DiTi), Tecan™ Pure, Filtered, 1,000µL (30 000 631); six trays (each containing 96 DiTis)
Disposable Tips (DiTi), Tecan™ Pure, Filtered, 200µL (30 000 629); three trays (each containing 96 DiTis)
100mL disposable troughs for reagents (5 troughs) Tecan™ (10613048)
(Optional) Barcodes See the Tecan™ Freedom EVO
PrepFiler™ 96-Well Processing Plate (10 plates) A47010
PrepFiler™ Automated Forensic DNA Extraction Kit: Automated DNA Purification on the HID EVOlution™ Systems User Bulletin
Tecan™ (30000631)
https://www.tecan.com/
Tecan™ (30000629)
https://www.tecan.com/
https://www.tecan.com/
Operating Manual, Section 3.5.6
“Positive Identification (PosID)”, for
barcode requirements
[2]
9
Chapter 1 Product information
1
Required materials not supplied
Table 3 Automated purification (continued)
[1]
If collecting eluate in plates:
Item Source
N8010560 or 4306737
MicroAmp™ Optical 96-Well Reaction Plate (without barcode) or MicroAmp™ Optical 96-Well Reaction Plate with Barcode
If collecting eluate in tubes:
AM12450, or equivalent
Nonstick, RNase-free Microfuge Tubes, 1.5 mL; certified DNase- and RNase-free (250 tubes)
[1]
Recommended sources. Unless otherwise indicated, equivalent materials from other suppliers can be used after appropriate validation studies by the user laboratory.
[2]
Disposable tips that have not been certified by Tecan™ may not yield the same liquid-handling performance.
10
PrepFiler™ Automated Forensic DNA Extraction Kit: Automated DNA Purification on the HID EVOlution™ Systems User
Bulletin
Prepare for the automated
2
purification run
Perform lysis ......................................................................... 11
Before first use: Prepare the wash buers ............................................... 11
Before each use: Prepare the magnetic particles ......................................... 11

Perform lysis

Manually perform lysis according to the PrepFiler™ and PrepFiler™ BTA Automated Forensic DNA Extraction Kits User Guide (Pub. No. 4463349).
Before first use: Prepare the wash buers
1.
Mix 260 mL of PrepFiler™ Wash Buer A Concentrate with 740 mL of freshly-opened 95% ethanol in a separate, clean container to prepare a 1X solution.
2.
Mix 200 mL of PrepFiler™ Wash Buer B Concentrate with 300 mL of freshly-opened 95% ethanol in a separate, clean container to prepare a 1X solution.
If the containers are kept closed when not in use, the prepared wash buers have a shelf life of 6 months or the kit expiration date, whichever is earlier.

Before each use: Prepare the magnetic particles

1.
Incubate the PrepFiler™ Magnetic Particles tubes at 37 for 10 minutes.
2.
Vortex at medium speed until the particles are completely resuspended and homogenous, then briefly centrifuge.
3.
Use one of the following methods to remove any air bubbles:
Draw o bubbles with a disposable bulb pipette.
Use a clean pipette tip to break up the bubbles.
Use a lint-free wipe to absorb the bubbles.
IMPORTANT! Bubbles can interfere with automated liquid detection and aspiration.
PrepFiler™ Automated Forensic DNA Extraction Kit: Automated DNA Purification on the HID EVOlution™ Systems User Bulletin
11
Set up the robotic workstation
3
For more information .................................................................. 12
Workflow ............................................................................ 13
Perform routine maintenance .......................................................... 14
(HID EVOlution™—Combination System only) Set up the carriers and labware ............... 16
Set up the disposable pipette tips ...................................................... 19
Set up reagents on the workstation ..................................................... 20
Set up lysate, processing, and elution plates and/or tubes ................................. 23
Workstation layouts ................................................................... 30

For more information

This chapter provides general procedures for setting up the Freedom EVO™ 150 robotic workstation or Freedom EVO™ 200 robotic workstation for the HID EVOlution™ system.
For more information, see the appropriate manufacturer's documentation.
Document
Tecan™ HID EVOlution™—Extraction Application Manual
(395372, v2.0, June 2010)
Pre-run preparation step 4.3.2 “Prepare the Instrument”
Maintenance schedules 7.2, “Maintenance Schedule”
Maintenance procedures 7.3, “Maintenance Tasks”
Maintenance scripts 5.2, “Running Maintenance”
Setting up disposable pipette tips 4.3.5, “Setup Plasticware and
Setting up reagents 4.3.4, “Setup Reagents on
Place the plates and/or tubes 4.3.5, “Setup Plasticware and
Workstation layouts 4.3, "Preparing the Instrument"
Description Section
7.5, “Maintenance Scripts”
Samples on the Workstation”
Workstation”
Samples on the Workstation”
4.4, "Worktable Layouts"
12
PrepFiler™ Automated Forensic DNA Extraction Kit: Automated DNA Purification on the HID EVOlution™ Systems User
Bulletin
(continued)
Chapter 3 Set up the robotic workstation
Document Description Section
Workflow
3
Tecan™ HID EVOlution™— Combination Application Manual
(395967, v2.0, June 2010)
Tecan™ Freedom EVO™ Operating Manual
Pre-run preparation step 5.3.2 “Prepare the Instrument”
Maintenance schedules 12.2, “Maintenance Schedule”
Maintenance procedures 12.3, “Maintenance Tasks”
Maintenance scripts 6.2 and 10.2, “Running Maintenance”
12.5, “Maintenance Scripts”
Setting up disposable pipette tips 5.3.5, “Set Up Plasticware and
Samples on the Workstation”
Setting up reagents 5.3.4, “Set Up Reagents on
Workstation”
Place the plates and/or tubes 5.3.5, “Set Up Plasticware and
Samples on the Workstation”
Workstation layouts 5.3, "Preparing the Instrument"
5.4, "Worktable Layouts"
Barcode specifications 3.5.6, “Positive Identification (PosID)”
Workflow
Prepare the reagents and samples
Prepare the system liquid carboy
(HID EVOlution™—Combination System only) Set up the carriers and labware
Set up the disposable pipette tips
Set up reagents on the workstation
Set up lysate, processing, and elution plates and/or tubes
PrepFiler™ Automated Forensic DNA Extraction Kit: Automated DNA Purification on the HID EVOlution™ Systems User Bulletin
13
Chapter 3 Set up the robotic workstation
3

Perform routine maintenance

Perform routine maintenance
Before placing the samples, reagents, and labware (DiTis, troughs, plates, and tubes), prepare the robotic workstation.

Prepare the system liquid carboy

Ensure that the carboy next to the workstation contains enough system liquid (degassed deionized water) to complete the experiment.
1.
Degas the deionized water overnight or longer before using it on the system.
Note: The time needed for complete degassing varies, depending on the climate in each laboratory and geographical location. In some situations, it may take up to 3 days to fully degas the deionized water. We recommend that each laboratory maintain an additional carboy of fully degassed deionized water to use for replenishment.
2.
To avoid introducing air into the system liquid tubing, follow these guidelines:
Place the system liquid carboy at the same height as the worktable.
Replenish the system liquid as needed before each run to avoid liquid levels dropping below one-quarter carboy during the run.
3.
Run the routine maintenance script each time that you change the system liquid carboy.

Empty the waste carboy

Check the waste carboy, and empty if needed.

Tighten the DiTi adapter gold cones

Note: If the cones are loose, the instrument may fail to pick up pipette tips during the run, and liquid
delivery will be inconsistent.
Use your fingers to gently tighten the DiTi adapter gold cones on the LiHa and the syringe assembly
fittings.
For details, see the Tecan™ Freedom EVO™ Operating Manual.
14
PrepFiler™ Automated Forensic DNA Extraction Kit: Automated DNA Purification on the HID EVOlution™ Systems User
Bulletin

Run maintenance scripts

Before starting the run, run the appropriate maintenance scripts.
IMPORTANT! Watch for air bubbles in the syringes and tubing. Repeat system flushing as needed to
remove the air bubbles.
If Then run
It is the first run of the day PrepFiler_DailyStartUp or Combo_DailyStartUp
It is not the first run of the day PrepFiler_Flush or Combo_Flush
Chapter 3 Set up the robotic workstation
Perform routine maintenance
3
When you run DailyStartUp or Flush, you see:
Air bubbles in the lines
and/or
Intermittent flow from a DiTi cone
There are one or more DiTis on the LiHa PrepFiler_Drop_DiTis or Combo_Drop_DiTis
PrepFiler_Flush or Combo_Flush one or more times until:
There are no visible air bubbles
and
Flow from the DiTi cones is constant
PrepFiler™ Automated Forensic DNA Extraction Kit: Automated DNA Purification on the HID EVOlution™ Systems User Bulletin
15
7
8
1
Chapter 3 Set up the robotic workstation
3

(HID EVOlution™—Combination System only) Set up the carriers and labware

(HID EVOlution™—Combination System only) Set up the carriers and labware
If the HID EVOlution™—Combination System was last run for qPCR/STR, you will need to set up the carriers and labware for a purification run.
1.
If the DNA lysate is in tubes, remove the 3-position microplate carrier from Grid 7, then place six tube racks on grids 7 through 12.
Note: If the DNA lysate is in a plate, you do not need to remove the 3-position microplate carrier.
16
3-position microplate carrier
2.
Set up carriers for the DNA eluate.
If you want the DNA eluate placed in tubes, position six tube racks on grids 1 through 6.
If you want the DNA eluate placed in a 96-well plate, position the metal plate adaptor on grid 13, position 1.
PrepFiler™ Automated Forensic DNA Extraction Kit: Automated DNA Purification on the HID EVOlution™ Systems User
Bulletin
1
2
Chapter 3
(HID EVOlution™—Combination System only) Set up the carriers and labware
3.
Remove the 3-position disposable tips (DiTi) tray carrier from grid 35, then replace it with a flat
Set up the robotic workstation
carrier and three 1,000-μL DiTi boxes as shown.
3
200-µL DiTi trays (blue or white trays)
1,000-µL DiTi trays (yellow or white trays)
4.
Place the magnetic particle tube block on grid 13, position 2.
IMPORTANT! Ensure that the tubes and the block containing the tubes are positioned as shown.
Incorrect positioning may result in failure to pipet magnetic particles and/or collision of the Liquid Handling (LiHa) arm with the block.
5.
Ensure that the 96-well magnetic ring stand is on grid 19, position 2.
PrepFiler™ Automated Forensic DNA Extraction Kit: Automated DNA Purification on the HID EVOlution™ Systems User Bulletin
17
Chapter 3
3
(HID EVOlution™—Combination System only) Set up the carriers and labware
6.
Set up the robotic workstation
Set up the reagent troughs.
a.
Remove the reagent troughs from previous runs and correctly dispose of the reagents according to “Complete the run” on page 48.
b.
Place new 100-mL reagent troughs on the worktable for PrepFiler™ Wash Buer B and other reagents as shown.
Note: The trough layout shown is dierent from the originally validated layout. The validation of the new trough layout is described in “Validation of PrepFiler™ Wash Buer B and the related modifications to the workstation layout and scripts” on page 73.
18
Elution buer trough (grid 27, position 1)
·
Prepared Wash Buer B trough (grid 27, position 2)
·
Prepared Wash Buer A trough (grid 27, position 3)
·
Isopropanol trough (grid 25, position 1)
·
Lysate waste trough (grid 25, position 3)
·
The workstation should now match the setup shown in“Workstation layouts” on page 30.
PrepFiler™ Automated Forensic DNA Extraction Kit: Automated DNA Purification on the HID EVOlution™ Systems User
Bulletin
1
2
3
5
4
Chapter 3 Set up the robotic workstation

Set up the disposable pipette tips

Terms for pipette tips used on the robotic workstation

DiTis—Disposable Tips (DiTi), Tecan™ Pure, Filtered, 200- and 1,000µL
DiTi tray—Plastic tray containing 96 DiTis
DiTi rack—Aluminum holder for a single tray of 1,000µL DiTis
DiTi carrier—Aluminum holder for three trays of 200µL DiTis
Orientation nose—Pin on a DiTi rack to hold the tray in place
Set up the disposable pipette tips
3
Figure 2 DiTi terms.
Racks that contain 1,000-μL DiTis on the rear shelf positions 5, 6, and 7
1
Notch in the DiTi tray
2
Carrier for 200-μL DiTi trays (trays are blue or white)
3
Orientation nose
4
Racks that contain 1,000-μL DiTi trays (trays are yellow or white)
5
PrepFiler™ Automated Forensic DNA Extraction Kit: Automated DNA Purification on the HID EVOlution™ Systems User Bulletin
19
Chapter 3 Set up the robotic workstation
3

Set up reagents on the workstation

Fill DiTi carriers and racks

Set up the pipette tips for an automated purification run.
IMPORTANT! If nine full DiTi trays are not correctly set up on the Freedom EVO
the workstation repeatedly searches for the missing DiTi tips, during which time the samples may become unusable.
1.
Place three full trays of 1,000µL DiTis into the DiTi racks on the rear shelf (shelf positions 5, 6, and
7). For each tray:
a.
Insert the tray into a rack: Ensure that the notch in the tray is aligned with the orientation nose on the rack, snap the tray into the rack, then confirm that the tray fits snugly.
b.
Place the rack on the shelf: Ensure that the orientation pin is positioned toward the back of the shelf, then push the rack all the way to the back of the shelf.
robotic workstation,
IMPORTANT! Ensure that there are no objects placed on shelf positions 1–4 or position 8.
2.
Place three full trays of 1,000µL DiTis into the DiTi racks on grid 35, positions 1–3, as described in substep 1a. Ensure that the orientation pin is positioned in the top-left corner.
3.
Place three full trays of 200µL DiTis into the carrier on grid 29, positions 1–3. Ensure that the notch in the tray is positioned in the top-left corner.
IMPORTANT! Ensure that the 3-position DiTi carrier on grid 29 contains three 200-μL DiTi trays.
Set up reagents on the workstation

Procedural guidelines

Calculate the reagent volumes needed based on the number of samples you will process plus the specified overfill and dead volumes.
Note: The dead volume is independent of the number of samples you run.
Do not reuse isopropanol, PrepFiler™ Wash Buer A, PrepFiler™ Wash Buer B, or PrepFiler Elution Buer from previous runs; always properly dispose of used reagents after each run.
Do not use water instead of PrepFiler™ Elution Buer. Instead of PrepFiler™ Elution Buer, you can prepare low-TE buer (10 mM Tris-HCl, 0.1 mM EDTA, pH 8.0) or purchase DNA Suspension Buer (low-TE Buer) from Teknova™.
Use new reagent troughs each day.
20
PrepFiler™ Automated Forensic DNA Extraction Kit: Automated DNA Purification on the HID EVOlution™ Systems User
Bulletin

Set up the reagents

1.
Place empty troughs on the workstation according to the following table.
PrepFiler™ Elution Buer Grid 27, position 1
PrepFiler™ Wash Buer B Grid 27, position 2
PrepFiler™ Wash Buer A Grid 27, position 3
Isopropanol Grid 25, position 1
Chapter 3 Set up the robotic workstation
Set up reagents on the workstation
Empty trough Location
3
Lysate waste
IMPORTANT! Do not add acids or bases to
any wastes that contain PrepFiler™ Lysis Buer (guanidine thiocyanate).
2.
Calculate the required PrepFiler™ reagent volumes.
Reagent
Lysis
protocol
Isopropanol Standard,
300µL
Large­sample,
[4]
Prepared Wash
500µL
Up to 96 900 µL 15% 5 mL 105 mL
Buer A
Prepared Wash
Up to 96 300 µL 15% 5 mL 40 mL
Buer B
Grid 25, position 3
No. of
reactions
Reagent
volume per
reaction
Overfill
volume per
[1]
run
Dead
volume per
[2]
run
Minimum required
volume for
96 samples
(A×B)+(A×B×C)
A B C D
+D
Up to 96 180 µL 15% 5 mL 25 mL
Up to 96 300 µL 15% 5 mL 40 mL
[3]
Elution Buer Up to 96 50 µL 15% 5 mL 11 mL
[1]
Overfill (excess volume) is needed to compensate for evaporation and pipetting losses during the run.
[2]
An extra 5 mL per trough is needed to ensure that the pipette tips remain submerged during aspiration so that liquid, not air, enters the tips.
[3]
Includes overfill and dead volume. For example, the required volume of isopropanol for 96 samples when using the standard lysis protocol is (96 × 180 µL) + (96 × 180 µL × 0.15) + 5 mL = 17.28 mL + 2.59 mL + 5 mL = 24.87 mL, rounded up to 25 mL.
[4]
The large-sample (500‑µL) protocols were not tested as part of our full validation studies. If your laboratory intends to use the large-sample protocols, perform the appropriate validation studies.
PrepFiler™ Automated Forensic DNA Extraction Kit: Automated DNA Purification on the HID EVOlution™ Systems User Bulletin
21
Chapter 3
3
Set up reagents on the workstation
3.
Set up the robotic workstation
Add the amounts of PrepFiler™ reagents that you calculated in step 2 to the appropriate trough.
22
Elution buer trough
1
Prepared Wash Buer B trough
2
Prepared Wash Buer A trough
3
Isopropanol trough
4
Lysate waste trough, empty
5
4.
Gently invert two tubes of prepared PrepFiler™ Magnetic Particles to remove large air bubbles, briefly centrifuge the tubes at low speed to collect the contents at the bottom of the tubes, then open the tubes.
If a thin film or bubble (caused by surfactants) stretches across the top of the tube, gently break the surface with a clean pipette tip.
If there is foam (air bubbles) on the surface of the magnetic particles, remove the foam by pipetting. Surface foam may interfere with liquid level detection during the automated purification run.
PrepFiler™ Automated Forensic DNA Extraction Kit: Automated DNA Purification on the HID EVOlution™ Systems User
Bulletin
Chapter 3
Set up lysate, processing, and elution plates and/or tubes
5.
Place the two tubes of PrepFiler™ Magnetic Particles on the workstation in the first two slots of the
Set up the robotic workstation
metal rack on grid 13, position 2.
IMPORTANT! Ensure that the tubes and the block containing the tubes are positioned as shown.
Incorrect positioning may result in failure to pipette magnetic particles and/or collision of the Liquid Handling (LiHa) arm with the block.
3
Set up lysate, processing, and elution plates and/or tubes
Select a plate/tube configuration
The HID EVOlution™ systems support four plate/tube configurations for the Freedom EVO™ 150 robotic workstation or Freedom EVO™ 200 robotic workstation.
Select a configuration for each automated purification run.
Starting labware: Plate or tube
Configuration
Plate-to-plate Process lysate from a 96-well
plate and collect eluate in a 96­well plate
Plate-to-tubes Process lysate from a 96-well
plate and collect eluate in
1.5mL tubes
Tubes-to-tubes Process lysate from 1.5mL
tubes and collect eluate in
1.5mL tubes
Tubes-to-plate Process lysate from 1.5mL
tubes and collect eluate in a 96­well plate
[1]
Your choice is independent of whether the sample lysate is contained in a plate or in tubes
Description
that contains the lysate from
the sample lysis step
PrepFiler™ Spin Plate MicroAmp™ Optical 96-Well
PrepFiler™ Spin Plate Nonstick RNase-free
Nonstick RNase-free Microfuge Tubes (1.5mL)
Nonstick RNase-free Microfuge Tubes (1.5mL)
Ending labware: Plate or tube
to collect DNA eluate at the
end of the run
Reaction Plate
Microfuge Tubes (1.5mL)
Nonstick RNase-free Microfuge Tubes (1.5mL)
MicroAmp™ Optical 96-Well Reaction Plate
[1]
PrepFiler™ Automated Forensic DNA Extraction Kit: Automated DNA Purification on the HID EVOlution™ Systems User Bulletin
23
Chapter 3
3
Set up lysate, processing, and elution plates and/or tubes
“Workstation layouts” on page 30 shows the placement of plates and tubes for each configuration.
Set up the robotic workstation

Set up the PrepFiler™ Processing Plate

The PrepFiler™ Processing Plate is a square-well plate that is required to process reactions for all four automated purification run configurations.
During the washing and elution steps, the Robotic Manipulator arm (RoMa) moves the processing plate to the 96-Well Magnetic Ring Stand or TeShake™ plate adapter.
1.
If needed to ensure that the RoMa grips the plate tightly, place a strip of laboratory labeling tape on each side of the PrepFiler™ Processing Plate as shown.
2.
Place the PrepFiler™ Processing Plate on the TeShake™ plate adapter with well A1 in the top-left position (grid 19, position 3).
3.
To ensure that samples are transferred to the correct wells, confirm that:
The processing plate is placed on the TeShake™ plate adapter with well A1 positioned in the top-left corner
The plate wells are aligned with the holes in the TeShake™ plate adapter
24
PrepFiler™ Automated Forensic DNA Extraction Kit: Automated DNA Purification on the HID EVOlution™ Systems User
Bulletin

(If needed) Place barcodes

Perform this procedure if you use barcodes on the plates and/or tubes to track the samples in the HID EVOlution™ software. The system scans the barcodes to automatically capture sample information. For more information, see the Tecan™ HID EVOlution™ —Extraction Application Manual, Section 4.5, “Barcodes”.
IMPORTANT!
If the lysate is in spin/filter plates: Before the plate barcode is scanned during a run, you must
·
manually enter or import the sample information for each well in the plate. (See “Set up sample and reagent information” on page 40, step 1.) If the lysate is in tubes: The sample name (barcode) and sample position for each tube are
·
automatically updated in the HID EVOlution™ software when the barcodes are scanned.
1.
Select barcodes that are compatible with the PosID-3.
2.
Before placing items on the robotic workstation, ensure that the barcodes are correctly placed on the appropriate labware.
Chapter 3 Set up the robotic workstation
Set up lysate, processing, and elution plates and/or tubes
3
Component
Sample lysate PrepFiler™ Spin Plates Nonstick RNase-free Microfuge
DNA eluate MicroAmp™ Optical 96-Well Reaction
Plate
Set up lysate and/or eluate plates
IMPORTANT! To ensure that samples are transferred to the correct wells, confirm the following for
each lysate or eluate plate:
The plate is placed in the metal plate adapter with well A1 positioned in the upper left corner
·
The plate wells are aligned with the holes in the metal plate adapter
·
Plate Tube
Tubes (1.5mL)
Nonstick RNase-free Microfuge Tubes (1.5mL)
PrepFiler™ Automated Forensic DNA Extraction Kit: Automated DNA Purification on the HID EVOlution™ Systems User Bulletin
25
Chapter 3 Set up the robotic workstation
3
Set up lysate, processing, and elution plates and/or tubes
1.
If you want DNA eluate to be collected in a plate, place a MicroAmp™ Optical 96-Well Reaction Plate with well A1 in the top-left position (grid 13, position 1).
Note: The DNA eluate corresponding to the first sample is always placed in the first well (A1) of the elution plate. The Report file (PDF) that is generated at the end of the purification run lists the starting position of each sample lysate and the final position of the corresponding DNA eluate.
Note: Using 96-well plates from other manufacturers may result in liquid handling errors if the instrument is not recalibrated for use with the alternate plates.
2.
If the lysate is in a PrepFiler™ Spin Plate, place the spin plate with well A1 in the top-left position (grid 13, position 3).
26
PrepFiler™ Automated Forensic DNA Extraction Kit: Automated DNA Purification on the HID EVOlution™ Systems User
Bulletin
Set up lysate and/or eluate tubes
Set up eluate tubes in tube racks
1.
Ensure that:
You have new, labeled 1.5mL microfuge tubes equal to the number of DNA samples to be processed.
The tube racks S1–S6 are correctly positioned at grid positions 1–6.
2.
Place the first empty 1.5mL microfuge tube in the tube racks in rack S1, position 1.
Note: The DNA eluate corresponding to the first sample is always placed in the first tube (1) in the first tube rack (S1). The Report file (PDF) that is generated at the end of the purification run lists the starting position of each sample lysate and the final position of the corresponding DNA eluate.
3.
Continue placing empty tubes from back to front in vertical columns as shown. Place one empty tube for each sample to be processed. Do not leave empty positions between sample tubes.
IMPORTANT! The tubes must be contiguously loaded. Do not leave empty tube positions
between tubes.
Chapter 3 Set up the robotic workstation
Set up lysate, processing, and elution plates and/or tubes
3
Example of correct setup
4.
Ensure that the barcodes are in a readable position.
5.
Open each tube, securing the tube caps in a fixed upright position as shown.
IMPORTANT! Open tube caps carefully to prevent cross-contamination and splatter.
PrepFiler™ Automated Forensic DNA Extraction Kit: Automated DNA Purification on the HID EVOlution™ Systems User Bulletin
27
Chapter 3
3
Set up lysate, processing, and elution plates and/or tubes
Set up the robotic workstation
Set up lysate tubes in tube racks
1.
Ensure that:
You have ≤96 labeled 1.5mL microfuge tubes that contain DNA sample lysate.
The tube racks L1–L6 are correctly positioned at grid positions 7–12.
2.
Place the first sample tube in the tube racks. (Unlike the first eluate tube, which must be placed in rack S1, position 1, the first lysate tube may be placed in any position; for example, you can begin with rack L1, position 8.)
3.
Continue placing sample tubes from back to front in vertical columns as shown. Do not leave empty positions between sample tubes.
IMPORTANT! The tubes must be contiguously loaded. Do not leave empty tube positions
between tubes.
Examples of correct setup
28
4.
Ensure that the barcodes are in a readable position.
PrepFiler™ Automated Forensic DNA Extraction Kit: Automated DNA Purification on the HID EVOlution™ Systems User
Bulletin
Chapter 3 Set up the robotic workstation
Set up lysate, processing, and elution plates and/or tubes
5.
Open each tube, securing the tube caps in a fixed upright position as shown.
IMPORTANT! Open tube caps carefully to prevent cross-contamination and splatter.
3
PrepFiler™ Automated Forensic DNA Extraction Kit: Automated DNA Purification on the HID EVOlution™ Systems User Bulletin
29
Chapter 3 Set up the robotic workstation
3

Workstation layouts

Workstation layouts
The following figures show the available workstation layouts for the automated purification run configurations.
Figure 3 Plate-to-plate workstation layout
96-Well Elution Plate
1
Block for PrepFiler™ Magnetic Particles
2
PrepFiler™ Spin Plate
3
Magnetic Ring Stand
4
PrepFiler™ Processing Plate on TeShake™ plate adapter
5
Isopropanol trough
6
Lysate waste trough
7
Elution Buer trough
8
Wash Buer B trough
9
Wash Buer A trough
10
DiTi waste unit
11
200µL disposable pipette tips (DiTis)
12
200µL disposable pipette tips (DiTis)
13
200µL disposable pipette tips (DiTis)
14
1,000µL DiTis
15
1,000µL DiTis
16
1,000µL DiTis
17
Not shown: Rear shelf with 1,000µL DiTis in shelf positions 5, 6, and 7
30
PrepFiler™ Automated Forensic DNA Extraction Kit: Automated DNA Purification on the HID EVOlution™ Systems User
Bulletin
Loading...
+ 86 hidden pages