Thermo Fisher Scientific pHTestr 10-20-30-10 BNC-Spear User Manual

This Instruction Manual is also available for download on our Web-site: eutechinst.com or 4oakton.com
3. Press the CAL button to enter calibration mode. The ‘CAL’ indicator will be shown. The upper display will show the measured reading based on the last calibration while the lower display will indicate the pH standard buffer solution.
Part of Thermo Fisher Scientific
INSTRUCTION MANUAL
pHTestr 10, 20, 30,
10BNC, Spear
Large Screen
Waterproof pH / Temperature Testr
Double Junction
Introduction
Thank you for selecting our waterproof pH Testr. This manual serves the following (5) models:
• pHTestr 10 (Eutech PHTEST10 / 01X366901 / Oakton 35634-10)
• pHTestr 20 (Eutech PHTEST20 / 01X366902 / Oakton 35634-20)
• pHTestr 30 (Eutech PHTEST30 / 01X366903 / Oakton 35634-30)
• pHTestr 10BNC (Eutech PHTEST10BNC / 01X366904 / Oakton 35634-14)
• pH Spear (Eutech PHSPEAR / 01X366920 / Oakton 35634-40) This manual provides a step-by-step guide to operate the Testrs.
Before You Begin
Your instrument has been factory calibrated and usually works well out of the box. However, after extended periods of non-use, it’s best to soak the sensor in electr ode storage solution, pH buffer for 10 minutes or so. A brief rinse with deionized (DI) water is OK, but avoid soaking or storing in deionized water as this will shor ten the pH electrode life. Prior to taking measurements, periodic calibration with certified standards is recommended for best accuracy.
Note: For pHTestr10BNC, please refer to the pH electrode’s instruction manual.
pH Buffer Set Selection
Your Testr features USA (pH 4.01, pH 7.00 and pH 10.01) or NIST (pH 4.01, pH 6.86, and pH 9.18) standards. Select either one to suit your requirements.
1. While pressing the HOLD/ENT button, switch on the Testr by pressing the ON/OFF button.
2. Release the HOLD/ENT button. The display will flash either USA or NIST.
3. Press CAL button to toggle between the two buffer set standards.
4. Press the HOLD/ENT button to confirm the selection of the buffer set.
With meter powered off, press HOLD/ENT and ON/OFF at the same time. First release ON/OFF button, then HOLD/ ENT button.
Figure 1: Buffer Selection Sequence
pH Calibration
Calibration should be done regularly, preferably once a week. You can calibrate up to three points using either the USA or the NIST buffer set standards.
1. Press ON/OFF button to switch unit on.
2. Dip electrode about 2 to 3 cm into the pH standard buffer solution.
After buffer selection, whole LCD segment lits up and goes back to measurement mode.
Note: All Testrs have dual display during calibration mode. Note: To abort calibration, press the ‘CAL’ button.
4. Allow about 2 minutes for the Testr reading to stabilize before pressing the HOLD/ENT button to confirm the first calibration point. The upper display will be calibrated to the pH standard buffer solution and the lower display will then search for the next pH standard buffer solution.
5. Repeat with other buffers if necessary. Rinse electrode before dipping into next buffer.
Note: The calibration mode allows you to perform up to three calibration points before returning to the measurement mode automatically. However, if you opted to have only one or two calibration points, simply skip the remaining calibration points by exiting to the measurement mode by pressing the CAL button.
To do a 1 point calibration only, press CAL button at this point to exit to the measurement mode. Otherwise, proceed to second buffer for a second buffer for a second point calibration.
To do a 2 point calibration only, pr ess CAL button at this point to exit to the measurement mode. Otherwise, proceed to third buffer for a third buffer for a third point calibration.
After the third point calibration, the meter will automatically return to the measurement mode. At any point, an error message ‘Er. 1’ will be displayed momentarily if the confirmed pH value is not within the pH calibration window.
Figure 2: Example of pH Calibration Sequence
pH Measurement
1. Press the ON/OFF button to switch the Testr on.
2. Dip the electrode about 2 to 3 cm into the test solution. Stir and let the reading stabilize. For semi-solid samples using pHSpear, rotation is ideal for sample contact. CAUTION: Testing dry samples is not accurate and can lead to sensor damage or breakage. Soils must be wet and free of particulates that may scratch the glass sensor. Excessive force into dry samples can cause glass breakage.
3. Note the pH value or press HOLD/ENT button to freeze the reading. To release the reading, press HOLD/ENT again.
4. Press ON/OFF to turn off Testr. If you do not press a button for 8.5 minutes, the Testr will automatically shut off to conserve batteries.
HOLD Function
This feature lets you freeze the display for a delayed observation
1. Press HOLD/ENT button to freeze the measurement. A ‘HOLD’ indicator will be displayed and the measurement will be frozen.
2. Press HOLD/ENT again to release the measurement. The ‘HOLD’ indicator will not be displayed anymore indicating the held measurement is released.
Figure 4: Example of HOLD Function
User Reset
You can reset the pH calibration to the factory default by using the user reset function. Buffer set selection and temperature user calibration (pHTestr30) are not affected by the user reset function.
1. Switch off the Testr.
2. While pressing the ‘CAL’ button, press and release the ON/OFF button to enter the ‘User Reset’ selection menu. The screen will display ‘rSt’ on the bottom display with a flashing ‘nO’ selection.
3. Use the ‘CAL’ button to toggle between ‘nO’ and ‘YES’ selection.
• nO deactivates reset selection
• YES activates the reset selection
4. Press the HOLD/ENT button to confirm the selection made.
5. If you have selected ‘YES’, the unit will show ‘CO’ momentarily and proceed to the measurement mode with the calibration reset back to factory default value.
6. If ‘nO’ is selected, the unit will proceed to the measurement mode without any calibration reset performed.
Figure 5: User Reset Sequence
After reset, whole LCD segment lits up and goes back to measurement mode.
ATC - Automatic Temperature Compensation (Only for pHTestr 10, pHTestr 20 and pHTestr 30)
Through its in-built temperature sensor, the measurement error due to the changes in electrode sensitivity due to changes in temperature is compensated to give the actual pH reading of the sample measured.
MTC - Manual Temperature Compensation (Only for pHTestr 10BNC, pH Spear)
The MTC range is 0 to 50.0 °C (32.0 to122.0 °F). User reset will set temperature to default value 25°C or 77°F.
While in the measurement mode,
1. Press the HOLD/ENT button to bring the Testr to the ‘HOLD’ mode.
2. Press the CAL button continuously to switch to the °C or °F mode setting selection screen.
3. Release the CAL button to confirm your mode selection and the display will go to the manual temperature calibration mode with the upper display flashing. The upper display shows the adjustable temperature value and the lower display shows the last set temperature offset.
4. Press the HOLD/ENT button to set the upper display to the temperature value of your sample.
5. Once the setting is reached, release the HOLD/ENT button. The new value is automatically confirmed and returns to the measurement mode if no button is pressed after 5 seconds.
Notes: To exit this program without confirming the calibration, press the CAL button before the automatic confirmation takes place.
Temperature Calibration (Only For pHTestr 30)
From the measurement mode,
1. Press the HOLD/ENT button to bring the Testr to the ‘HOLD’ mode.
2. Press the CAL button for 3 seconds to switch to the °C or °F mode setting selection screen. Pressing the CAL button continuously for 3 seconds allows you to toggle in between the °C and °F mode setting selection screen.
3. Release the CAL button to confirm your mode selection and the display will go to the temperature calibration mode with the upper display flashing. The upper display shows the current measured temperature reading based on the last set offset and the lower display shows the current measured temperature reading based on factory default calibration.
4. Dip the Testr into a solution of known temperature and allow time for the in built temperature sensor to stabilize.
5. Press the HOLD/ENT button to set the upper display to the temperature value of the solution.
6. Once the new temperature setting is reached, the new value is automatically confirmed and returns to the measurement mode if no button is pressed after 5 seconds.
Notes: To exit this program without confirming the calibration, press the CAL button before the automatic confirmation takes place.
Continuous pressing of CAL button (about 3 seconds) to tog gle
between temperature setting of degrees Celsius and Farenheit
Release CAL button to start entering
When the
During calibration, upper display will flash.
Press HOLD/ENT to increase offset up to +5 °C (+9 °F) from default value. It will roll over to -5 °C (-9 °F) from default value and
start increasing again.
temperature offset
value is reached,
the new value is automatically confirmed if no
key is pressed
for 5 seconds.
Figure 6: Temperature Calibration Sequence (Only for pHTestr 30DJ)
the new temperature
Electrode Maintenance
1. Before measuring soil pH with the pHSpear, wet your soil sample with distilled water and ensure that the soil is free of particulates. Even though the pHSpear is ruggedized, hard surfaces such as stones and pebbles can still cause breakage.
2. Rinse the electrode with electrode storage solution after each measurement. Care has to be taken not to damage the sensor’s glass electrode especially while rinsing the pHSpear penetrating tip electrode.
3. In aggressive chemicals, dir ty or viscous solutions, and solutions with heavy metals or proteins, take readings quickly and rinse electrode immediately afterward. For the pHSpear, the remnants of the semi solid samples on the penetrating electrode can be removed by rubbing it with some table salt and then rinsing. Mild detergent can be used to wash the penetrating electrode clean.
4. If possible, keep a small piece of paper or sponge in the electrode cap – moistened with clean water or electrode storage solution (NOT de-ionized water) – and close the cap over the electrode. For pHSpear, ensure that the electrode is kept soaked in electrode storage solution via its protective cap.
Changing Batteries
1. Open battery compartment lid (with attached lanyard loop).
2. Remove old batteries; replace with fresh ones. Note polarity.
Self-Diagnostic Messages
3 Bars indicates battery is full (100%)
Low Battery
Indicator
2 Bars indicates 50% of the battery life is left
1 Bar indicates 25% of the batter y life is left
Blinking battery casing indicates the need to replace batteries with fresh ones as specified by manufacturer
Electrode is not in contact with solution or electrode is failing
Over Range / Under Range
Signal
Error
Message
Or / Ur
(Still)
ATC / Or / Ur
(Blinking)
Er.0
Er.1
Replacement sensor is not connected properly to the Testr during sensor replacement
Measured pH value or temperature value (pHTestr30) exceeds its specified maximum or minimum value
Blinking ‘ATC’, ‘Or’ or ‘Ur’ indicates that there is a shor t or open circuit at the built in temperature sensor
Temperature calibration error of attempting to calibrate Testr to a value which is out of range or under range
pH calibration error of attempting to confirm a calibration value which is not within the specified calibration window
Electrode Replacement
You can replace the electrode module at the fraction of the cost of a new Testr. When the Testr fails to calibrate or gives fluctuating readings in calibration standards, you need to change the electrode.
1. With dry hands, grip the ribbed Testr collar with electrode facing you. Twist the collar counter clockwise (see picture A). Save the ribbed Testr collar and O-ring for later use.
2. Pull the old electrode module away from the Testr.
3. Align the four tabs on the new module so that they match the four slots on the Testr (see picture B).
4. Gently push the module onto the slots to sit it in position. Push the smaller O-ring fully onto the new electrode module. Push the collar over the module and thread it into place by firmly twisting clockwise.
Note: It is necessary that you recalibrate your Testr prior to measurement after an electrode replacement.
Rotate collar
away from you
Rlbbed collar
Figure 7: Removal of collar
from tester
Insert Electrode Module
Electrode Module
Small O-ring
Figure 8: Example of electrode module
Small Tab Large Tab
fitting alignment
Large
O-ring
Applications
Water quality testing • pools • spas • aquariums • aquaculture • hydroponics • ecology studies • water and wastewater treatment • boilers • steam generators • car washes
• sanitation plants • labs • food sectors and more!
Warranty
The waterproof Testrs are warranted to be free from manufacturing defects for 1 year and electrode module for 6 months, unless otherwise specified. If r epair, adjustment or replacement is necessary and has not been the result of abuse or misuse within the time period specified, please contact your local distributor for assistance.
Return Of Items
Authorization must be obtained from your distributor before returning items for any reason. When applying for authorization, please include information regarding the reason the item(s) are to be returned.
Note: We reserve the right to make improvements in design, construction and appear ance of products without notice. Prices are subject to change without notice.
Accessories
Item
pHTestr 10,20,30 Replacement Sensor PHSENSOR03DJ 35624-38 pHSpear Replacement Sensor PHSENSOR04 35634-50 Carrying Pouch pH Testr 10, 20, 30 ECPOUCH01 35624-45
Eutech Instruments Order Code
Oakton Instruments Order Code
Testr Specifications
Large Screen Testrs pHTestr 10 pHTestr 20 pHTestr 30
pH Range -1.0 to 15.0 pH -1.00 to 15.00 pH Resolution 0.1 pH 0.01 pH Relative Accuracy 0.1 pH 0.01 pH Calibration Points Up to 3 points Up to 3 points
Buffer Set Standard Selection
Calibration Window (USA Buffer Set Standard)
Calibration Window (NIST Buffer Set Standard)
Temperature No Display 0-50.0°C or 32.0-122.0°F Temperature Compensation ATC MTC Temperature Resolution No 0.1 °C / °F Temperature Accuracy No 0.5 °C / 0.9 °F No
Temperature Calibration Window
Auto Off After 8.5 minutes from last key press User Reset Yes Non Volatile Memory Backup Yes LCD Display Dual Power Requirement 4 x 1.5V “A 76” micro alkaline batteries Battery Life More than 500 hrs Operating Temperature 0 – 50 °C
Testr Dimensions
Weight 3.25 oz (90 gm)
Penetrating Electrode Total Length
Shaft Length Not applicable 31 mm
Penetrating Electrode Upper Diameter
Shaft Diameter Not applicable 7 mm
USA -
4.0/7.0/10.0 NIST -
4.0/6.9/9.2
±1.0 pH
(pH 4.0 & pH 10.0),
±1.5 pH (pH 7.0)
±1.0 pH
(pH 4.0 & pH 9.2),
±1.2 pH (pH 6.9)
USA - 4.01/7.00/10.01
NIST - 4.01/6.86/9.18
±1.00 pH (pH 4.01 & pH 10.01),
±1.00 pH (pH 4.01 & pH 9.18),
No
±(5°C /9 °F)
from default value
6.5 “L x 1.5”dia. (165 x 38 mm)
Not applicable 91 mm
Not applicable 12 mm
pHTestr
10BNC
±1.50 pH (pH 7.00)
±1.25 pH (pH 6.86)
68X068042 Jun 14 Rev 8
pH Spear
0-50.0°C or
32.0-122.0°F (MTC)
9.75 “L x 1.5”dia.
(247 x 38 mm)