Thermo Fisher Scientific pHScan BNC User Manual

Open this booklet for calibration/testing/ maintenance instructions.
Changing Batteries:
1. Open batter y compartment lid (with attached lanyard loop).
2. Remove old batteries; replace
+
with fresh ones. Note polarity (shown in battery)
keypad
compartment and in picture at right).
3. Recalibrate tester after battery change.
Error Messages:
bAT Weak batteries – replace Er1 Wrong or bad buffer value, or electrode
is failing. Or, Ur Over range / under range signal, or electrode is failing.
Warranty:
The waterproof pHScan BNC is warranted against defects in materials and workmanship for a period of 24 months from the date of purchase. If repair, adjustment or replacement is necessary and has not been the result of abuse or misuse within the warranty period, please return the tester – freight pre-paid – and correction will be made without charge. Out of warranty products will be repaired on a charge basis.
Return of Items:
Authorization must be obtained from your Distributor before returning items for any reason. When applying for authorization, please include information regarding the reason the item(s) are to be returned. Note: We reserve the right to make impro vements in design, construction and appearance of products without notice. Prices are subject to change without notice.
68X247780 11/01 Rev 0
Microprocessor based pH Tester that floats!
Waterproof pHScan BNC Microprocessor-based Pocket pH Tester
Connecting your electrode:
To connect the tester to your electrode, align the
posts of the tester's BNC connector with the slots
on the electrode connector. Push together and twist the electrode connector 1/2 turn until it clicks into place.
Calibration:
Calibration should be done regularly, typically
every day that the tester is used. You can calibrate
up to three point. Choose buffer sets as described
above right.
1. Press ON/OFF button to switch unit on.
2. Dip electrode about 2 cm into buffer.
3. Press CAL button to enter Calibrate (CAL)
mode. 'CAL' flashes on the display. Then a
pH value close to pH buffer value will flash
repeatedly.
4. After at least 3 minutes, press the HOLD/CON button to confirm
calibration. The display will show
'CO' and then switch to the buffer
value reading.
5. Repeat with other buffers if necessary. Rinse electrode in tap
water before dipping into next buffer.
Calibration Troubleshooting:
Failure to press HOLD/CON to confirm calibration (step 4 above). Pressing the CAL button will resume measuring mode but will not enter the calibration value. Insufficient sampling time. If the
tester does not have a long enough
exposure to the buffer, a stable
calibration point will not be reached.
Wait at least 3 minutes before pressing HOLD/CON. Failure to re-hydrate the electrode. A
dry electrode will give fluctuating reading while it re-hydrates in a buffer,
causing errors.
Buffer Set selections:
To choose USA (pH 4.01, 7.00, 10.01) or NIST (pH 4.01, 6.86, 9.18) buffers:
1. With tester off, hold HOLD/CON key while pressing ON/OFF key.
2. Press HOLD/CON key again.
3. Press CAL key to toggle between USA and NIST
buffer sets.
4. Press HOLD/CON to confirm.
pH Testing:
1. Remove cap or storage bottle from the electrode and press the ON/OFF button to switch tester on.
2. Dip the electrode about 2 cm into the test solution. Stir once and let the reading stabilize.
3. Note the pH or press HOLD/CON button to freeze the reading. Press HOLD/CON again to release the reading.
4. Press ON/OFF to tur n off tester. If you do not press a button for 8.5 minutes the tester will automatically shut off to conser ve batteries.
Instrument Maintenance:
• Rinse the electrode with tap water or electrode storage solution after each measurement.
• Periodic soaks in warm pH 4 buffer will help remove contaminants.
• When finished using electrode, fill electrode cap with tap water or electrode storage solution and replace cap.