Thermo Fisher Scientific pH-Ion 510 User Manual

Instruction Manual
gyM
sy...
pH/Ion 510
Bench pH/Ion/mV Meter
68X090811
Technol o
adeEa
Rev. 8 DEC 06
PREFACE
Thank you for choosing the pH 510 pH/mV/Temperature or Ion 510 Ion/pH/mV/Temperature bench meter.
The instruction manual functions in two ways: firstly as a step-by-step guide to help you operate the 510 series meter. Secondly, it serves as a handy reference guide.
The instruction manual is written to cover as many anticipated applications of the pH/Ion 510 Bench meter series as possible. If there are doubts in the use of the instrument, do not hesitate to contact the nearest Authorized Distributor.
Eutech Instruments/ Oakton Instruments cannot accept any responsibility for damage or malfunction to the meter caused by improper use of the instrument.
The information presented in this manual is subject to change without notice as improvements are made, and does not represent a commitment on the part of Eutech Instruments Pte Ltd/ Oakton Instruments.
Note: Eutech Instruments Pte Ltd/ Oakton Instruments reserves the right to make improvements in design, construction, and appearance of our products without notice.
Copyright © 2002 All rights reserved. Eutech Instruments Pte Ltd Oakton Instruments
Rev.8 DEC 06
TABLE OF CONTENTS
1 INTRODUCTION.............................................................................................................................. 1
1.1 Introducing the 510 Bench Series.............................................................................................................1
1.2 Custom LCD.............................................................................................................................................1
1.3 Keypad.....................................................................................................................................................2
1.4 Rear Instrument Panel..............................................................................................................................3
2 STARTING UP................................................................................................................................. 4
2.1 Connecting the Sensor Electrode.............................................................................................................4
2.2 Connecting the Temperature Probe .........................................................................................................4
2.3 Connecting the A/C Adapter.....................................................................................................................4
2.4 Connecting the Chart Recorder................................................................................................................4
2.5 Connecting the Electrode Holder (Optional).............................................................................................5
2.6 Steps Prior to Calibration..........................................................................................................................6
3 CALIBRATION................................................................................................................................. 6
3.1 pH Calibration...........................................................................................................................................6
3.2 Ion Concentration Calibration (For Ion 510 only)......................................................................................9
3.3 mV Calibration........................................................................................................................................11
3.4 Temperature Calibration.........................................................................................................................12
4 MEASUREMENT ........................................................................................................................... 13
4.1 Automatic Temperature Compensation (ATC) .......................................................................................13
4.2 Manual Temperature Compensation (MTC)...........................................................................................13
4.3 Taking Measurements............................................................................................................................14
4.4 HOLD Function.......................................................................................................................................15
4.5 Storing and Recalling Data from Memory...............................................................................................15
5 SETUP FUNCTIONS ..................................................................................................................... 17
5.1 P1.0 Clear Memory................................................................................................................................17
5.2 P2.0 Viewing Electrode Diagnosis........................................................................................................18
5.3 P3.0 Selecting pH Buffer Standard and No of Calibration Points ..........................................................19
5.4 P4.0 Meter Reset................................................................................................................................... 20
6 pH Electrode Maintenance .......................................................................................................... 21
6.1 Storage...................................................................................................................................................21
6.2 After Use ................................................................................................................................................21
6.3 Electrode Cleaning.................................................................................................................................22
6.4 pH Electrode Rejuvenation.....................................................................................................................23
7 TROUBLESHOOTING GUIDE ...................................................................................................... 24
7.1 Error Messages......................................................................................................................................24
7.2 Troubleshooting......................................................................................................................................25
8 ADDITIONAL INFORMATION....................................................................................................... 26
8.1 pH and Temperature ..............................................................................................................................26
8.2 Use of Standard pH Buffers....................................................................................................................26
8.3 Standard pH Buffers...............................................................................................................................26
8.4 List of Accessories..................................................................................................................................27
8.5 Specifications .........................................................................................................................................29
9 WARRANTY .................................................................................................................................. 30
10 RETURN OF ITEMS ...................................................................................................................... 30
Instruction Manual pH 510/ Ion 510
1 INTRODUCTION
1.1 Introducing the 510 Bench Series
The pH 510 and Ion 510 bench meters are microproces sor-based which incorporates new ASIC (Application Specific Integrated Circuit). It is designed with convenience in mind, and offers many advanced user-friendly features. The meters are capable of storing and recalling up to 100 data sets in its non-volatile memory. In addition, as a space saver, an optional s wivel electrode holder can be attached at the either side of bench meter for resting the electrodes and probes during operation.
The pH 510 meter is capable of measuring pH/ORP with Temper ature; while the Ion 510 meter measures Ion Concentration/pH/ORP with temperature. It is equipped with a large cust omized LCD (Liquid Crystal Display) with simultaneous display of the measured values for easy reading. It is most ideal for routine pH/Ion Concentration measurement in indoor applications.
1.2 Custom LCD
The 510 meter is characterized by a large custom dual LCD. The display has mode annunciators for pH, mV, ppm and temperature readings. The secondary (lo wer) display shows the temperature reading s imultaneously with the primary (upper) display that sho ws the measur ed v alues of pH, ppm or mV. Spec ial ann unciator s such as units of measure, error message, graphical icons and modes of operation are arranged around b oth prima r y and secondary display to give you a comprehensive yet uncluttered display.
- The instrument is in
SETUP – The meter is in SETUP mode .
MEAS
- The instrument is in
the measurement mode.
CAL
the calib ration m ode
.
MEM - Data is stored
into memory
- The measured
READY
value has stabilized.
HOLD - The displayed
value has been frozen.
ERR
- An error in
operation has occured .
Buffer - Prompts user
to select buffer during calibration, or indicates buffer error when flashing together with the ERR annunciator.
SETUP
MEAS
READY HOLD
ELECTRODE
electrode error when it appears with ERR.
- indicates
ERR
CAL
MEM
R.mV
pH
ppm
C F pH
ATC
pH –Indicates buffer
values of respective USA or NIST buffer option set.
Shows the respective mode of operation.
ATC (Automatic Temperature Compensation)
activated. displayed, temperature probe is not connected. In this case the Manual Temperature Compensation is activated.
If ATC is not
- 1 -
Instruction Manual pH 510/ Ion 510
1.3 Keypad
A large splash proof membrane keypad with tactile feedback makes meter easy to use. Names and symbols describe the function button controls.
Key Measurement Mode Calibration / Setup Mode
ON/OFF
CAL/MEAS
MODE
MI/¿
MR/À
HOLD
ENTER
Powers the meter ON/OFF. When the meter is switched on, it automatically starts in the last mode of operation.
Toggles between the measurement and calibration modes of the meter. In SETUP mode, pressing CAL/MEAS key returns to the measurement mode.
Toggles between the different measurement modes available pH/Temperature/mV/Ion. (Ion mode only applies with the ION 510.)
MI inputs measurements into memory and scrolls through memory values.
MR recalls measurements from memory and scrolls through memory values.
Freezes a measurement on the display. Press again to get current readings.
Enter functions in the memory mode.
Switches from pH to Temperature in pH calibration mode
¿ scrolls values in mV calibration mode and selects temperature values and ion options in calibration mode. Scrolls to next program in the SETUP mode.
À scrolls values in mV calibration mode and selects temperature values and ion options in calibration Scrolls to next program in the SETUP mode.
Confirms and enters the value selected for calibration.
~
~
~
You can also configure functions like Meter Reset, Memory Clear and Selection of pH Buffer Standard via the keypad. Please refer to Section 5 for more information on the above functions.
- 2 -
Instruction Manual pH 510/ Ion 510
1.4 Rear Instrument Panel
This 510 series meter provides a complete set of connectors for the various accessories commonly used. Listed in the table are the details of the connections that you can make.
Rear View of Meter
Connection Function
ATC
INPUT
REF
REC
GND
DC
For phono jack connection from the temperature probe for Automatic Temperature compensation. The probe should be a 30KΩ thermistor type.
For connection to sensor electrodes with BNC type connectors. The meter accepts any pH, ORP or ISE with BNC connector. Always ensure that the connector is clean and dry.
For connection to pin type reference electrode normally used with half cell (mono) type pH electrodes or ISEs.
For connection to strip chart recorders. Use subminiature plug with positive tip. For connection to the ground earth jack (standard tip connectors). For connection for the AC/DC power adapter
- 3 -
Instruction Manual pH 510/ Ion 510
2 STARTING UP
Connect the accessory connectors at the rear of the instrument panel. During operati on, it is important that water does not get onto the BNC connector. Also avoid touching the connector with soiled or wet hands.
2.1 Connecting the Sensor Electrode
Slide the electrode connector of the electrode over the socket of the meter’s BNC c onnector marked ‘INPUT’. The meter can accept any standard pH, ORP, or Ion Selective Electrode with a BNC connector. Align the slot of the electrode connector with the posts of the socket. Rotate the connector clockwise until it locks. For separate reference electrodes, push the electrode pin into jack marked ‘REF’.
2.2 Connecting the Temperature Probe
Insert the temperature probe (provided) into the connector marked ‘ATC’.
2.3 Connecting the A/C Adapter
Slide in the adapter jack of the A/C adapter into the back panel marked ‘DC’ until it is firml y seated. Ensure that the power is switched off before installation and the correct voltage adapter is used.
2.4 Connecting the Chart Recorder
You can connect chart recorders or external output devices for continuous printout. Plug in the input connector and ground pin of the chart recorder into the ports marked ‘REC’ and ‘GND’ respectively.
Sensor electrode
Temp. probe
Separate ref. electrode
Chart recorder
Ground
AC/DC Adapter
- 4 -
Instruction Manual pH 510/ Ion 510
r
2.5 Connecting the Electrode Holder (Optional)
The integral electrode holder serves as a handy holder for mounting a few pH/Ion electrodes or a temperature probe during measurement or when idle.
This meter’s base plate has a side metal bar to which you can attach an integral swivel electrode holder. You can mount the electrode holder on either right or left side of the meter.
To position the electrode arm:
Use a Phillips screwdriver to remove the screw holding the electrode holder. Slide the side metal bar until the second screw slot lines up with the original screw hole. Use the screw removed earlier to secure the electrode holder into position.
NOTE: The side metal bar is reversible. If desired, remove screw holding electrode holder base and slide out of brackets, Slide base into brackets on opposite side and tighten screw.
Bottom of Meter
Body of Electrode Holder
Side Metal Bar
Base of Electrode Holde
To install the electrode arm to the meter:
To mount the electrode arm into the metal rod on the side bar, align the slot with the metal rod and base of electrode arm. Push it downwards until it fully sits into position. Avoid using excessive force when fixing or removing. The electrode arm is ready for use.
NOTE: Move the base of the electrode holder if you wish to swing the electrode holder about. To prevent the meter from toppling over causing accidental spills, DO NOT swing the body of the electrode holder.
- 5 -
Instruction Manual pH 510/ Ion 510
2.6 Steps Prior to Calibration
2.6.1 Switching ‘ON’ and ‘OFF’ the meter
Press ON key to switch meter on. All the LCD segments will display momentarily for a few seconds and the LCD then switches to the last measurement mode. To switch meter off, press OFF key.
2.6.2 Selecting Measurement Mode
The measurement modes available for the pH 510 meter are pH, mV (for ORP) and T emperature, while the Ion 510 meter has Ion, pH, mV and Temperature measurement modes. Select the correct mode of o peration by pressing the MODE key, and each press takes you to next measurement mode. When selecting a p articular parameter the respective mode annunciator displa ys at the right hand side of primary display; these inclu de [pH] for pH, [oC] for Temperature, [ppm] for Ion Concentration and [mV] for Millivolt.
2.6.3 Meter Reset
On first use, it is not necessary to reset meter before calibration as it has been factor y-calibrated. However if you wish to reset the meter to factory default refer to Section 5.4.
NOTE: Re-calibration of meter is required before meas urement after the meter is reset. Likewise if you wish to prepare a new set of measurements, you may reset all the last calibrated pH values simultaneo usly for added convenience. Caution: All calibration values and memory data will be erased upon activation. Re-
calibration is required.
2.6.4 Selecting pH Buffer Standard
The meter is capable of calibrating based on either USA or NIST pH buffer standards. Pleas e refer to Section
5.3 P3.0 for details on how to change the USA buffer standard to the NIST standard on the meter itself. Refer to Section 8.3 for more details on pH buffer standards available.
NOTE: Remove protective rubber cap or storage bottle (if any) from the electrode before proceeding with measurement or calibration. Take care not to exert too much force as this may cause damage to the electrode. Whenever the electrode is not in use, put it into storage bottle or protect the electro de with the rubber cap provided. Refer to the Section 6 on Probe Maintenance for more details.
3 CALIBRATION
3.1 pH Calibration
The 510 series bench meter is capable of multi-point c alibration (up to 5 points) for high accuracy acro ss the measurement range. The 5-point calibration offers flexibility of calibrating using either USA or NIST buffer standards with their 5 internationally accepted calibration points namely:
USA standard buffer options: pH1.68, 4.01, 7.00, 10.01, 12.45 NIST standard buffer options: pH1.68, 4.01, 6.86, 9.18, 12.45
For ATC measurements, attach the temperature probe to the meter. The ATC mode annunciator sho ws on the display. Insert the probe into the solution to be measur ed so that the s ample temper ature can b e recorded an d pH readings automatically temperature compensated. If ma nual temperatu re compens ation is preferre d, do not plug a temperature probe into the meter.
DO NOT REUSE SOLUTIONS AFTER CALIBRATION. Contaminants in the solution can affect the calibration, and eventually the accuracy of the measurements.
- 6 -
Instruction Manual pH 510/ Ion 510
One point calibration (Offset calibration) should only be carried out with pH 7.00 or pH 6.86; otherwise calibration will not be accepted. For Slope to Slope calibration (t wo points calibratio n), the first calibration point will only be successfully registered in the meter after the second point calibration has been completed.
It is recommended that you perform at least a 2-point calibration using buffers that bracket (one above and one below) the expected sample range, starting with pH 7 buffer first. Select a pH 1.68, 4, 7, 10 or 12.45 buffer in sachets or bottles from standard buffer solutions.
All new calibrations will replace the existing stored calibration data.
1. Press ON/OFF key. All the LCD segments will be displayed momentarily. The LCD switches to pH measurement mode. If necessary press MODE key to select pH mode.
NOTE: The meter starts in the same measurement mode that it was switched off.
2. The customized LCD display will indicate the following:
Display Remarks MEAS Measurement mode is selected pH Unit of measurement
o
C Measured temperature reading ATC Automatic Temperature Compensation (If temperature probe or “All-in-One” probe is
connected)”
3. Rinse the pH electrode well with deionized water or rinse solution. If using the ATC function with a separate temperature probe, thoroughly rinse the temperature probe as well. (Do not wipe the
electrode or the temperature probe as this may cause a b uild­up of electrostatic charge and will create calibration and measurement instability!).
One Point Calibration
4. Select the first pH buffer, pH 7.00 (or pH 6.86) and pour some into a clean container.
5. Dip both probes into the standard calibration buffer. The end of the probe must be completely immersed into the buffer. Stir the probes gently to create a homogenous solution.
6. Press CAL/MEAS key to enter the pH calibration mode. The annunciator [CAL] appears on top of the LCD to indicate the meter is in pH calibration mode. The upper display will show the measured reading while the lower display indicates appropriate pH standard buffer solution being used.
NOTE: The meter automatically recognizes the buffers in the buffer standard you have set in the SETUP mode, i.e. eith er USA (pH1.68, 4.01, 7.00, 10.01 or 12.45) or NIST (pH1.68, 4.01, 6.86
9.18 or 12.45) buffers.
Figure 1: 1st-point calibration
- 7 -
Loading...
+ 25 hidden pages