Thermo Fisher Scientific pH 800 User Manual

Page 1
Alpha Q)
%*/Q)031 $POUSPMMFS
Page 2
R
AQUAfas pHISA, pHuture, Pure Water, Sage, Sensing the Future, SensorLink, ROSS, ROSS Ultra, Sure-Flow, Titrator PLUS and TURBO2 are registered trademarks of Thermo Fisher.
1-888-pHAX-ION, A+, All in One, Aplus, AQUAsnap, AssuredAccuracy, AUTO-BAR, AUTO-CAL, AUTO DISPENSER, Auto-ID, AUTO-LOG, AUTO-READ, AUTO-STIR, Auto­Test, BOD AutoEZ, Cable-Free, CERTI-CAL, CISA, DataCOLLECT, DataPLUS, digital LogR, DirectCal, DuraProbe, Environmental Product Authority, Extra Easy/Extra Value, FAST QC, GAP, GLPcal, GLPcheck, GLPdoc, ISEasy, KAP, LabConnect, LogR, Low Maintenance Triode, Minimum Stir Requirement, MSR, NISS, One-Touch, One-Touch Calibration, One-Touch Measurement, Optimum Results, Orion Star, Pentrode, pHuture MMS, pHuture Pentrode, pHuture Quatrode, pHuture Triode, Quatrode, QuiKcheK, rf link, ROSS Resolution, SAOB, SMART AVERAGING, Smart CheK, SMART STABILITY, Stacked, Star Navigator 21, Stat Face, The Enhanced Lab, ThermaSense, Triode, TRIUMpH, Unbreakable pH, Universal Access are trademarks of Thermo Fisher.
Guaranteed Success and The Technical Edge are service marks of Thermo Fisher.
t, Cahn, ionplus, KNIpHE, No Cal, ORION, perpHect, PerpHecT, PerpHecTion,
Page 3
Preface
This manual serves to explain the use of the Alpha pH 800 process controller/transmitter. The manual functions in two ways, firstly as a step by step guide to help the user operate the instrument. Secondly, it serves as a handy reference guide. This instruction manual is written to cover as many anticipated applications of the Alpha pH 800 process controller/transmitter. If you have doubts in the use of the instrument, please do not hesitate to contact the nearest Authorized Distributor.
Thermo Scientific will not accept any responsibility for damage or malfunction of the unit due to improper use of the instrument.
The information presented in this manual is subject to change without notice as improvements are made, and does not represent a commitment on part of Thermo Scientific.
Copyright ©2009 Thermo Fisher Scientific
Page 4
Safety Information
This Controller/ Transmitter shall be installed and operated only in the manner specified in the Instruction manual. Only skilled, trained or authorized person should carry out installation, setup and operation of the instrument.
Before powering up the unit, make sure that power source it is connected to, is as specified in the top label. Failure to do so may result in a permanent damage to the unit.
The unit has live and exposed parts inside. If it has to be opened, make sure that the power to the unit is off and disconnected.
The unit is Fuse protected. In the event the fuse has to be replaced, use only those as specified in the manual.
Page 5
TABLE OF CONTENTS
1 INTRODUCTION 1
1.1 Description of Unit ................................................................................................................ 1
1.2 Applications .......................................................................................................................... 2
2 ASSEMBLY AND INSTALLATIONS 3
2.1 Measurement and Control System ....................................................................................... 3
2.2 Unit Dimensions ................................................................................................................... 4
3 ELECTRICAL CONNECTIONS 5
3.1 Connection Diagram ............................................................................................................. 5
3.2 Back Panel ........................................................................................................................... 7
4 OVERVIEW 9
4.1 Keypad and Display .............................................................................................................. 9
4.1.1 Keypad ........................................................................................................................ 9
4.2 Function Groups ................................................................................................................. 11
4.3 Control Concept .................................................................................................................. 11
5 MEASUREMENT 12
5.1 Display in Measure ment mode ........................................................................................... 12
6 CALIBRATION 13
6.1 pH Calibration ..................................................................................................................... 13
6.2 ORP mV Calibration ........................................................................................................ 15
7 ADVANCED SET UP MODE 16
7.1 Electrode Offset (OFS) sub-function .................................................................................. 16
7.2 Setting temperature (Set
7.2.1 Selecting automatic or manual temperature compensation ...................................... 17
7.2.2 Temperature calibration (ATC mode only) ................................................................ 18
7.2.3 Setting manual temperature compensation ............................................................... 19
7.3 Control Relay A/Control Relay B (SP1/SP2) sub-function ................................................. 20
7.3.1 Entering Set point 1 (Set point 2) sub-function .......................................................... 20
7.3.2 Selecting set point values .......................................................................................... 20
7.3.3 Choosing High or Low set points ............................................................................... 21
7.3.4 Selecting a hysteresis (dead band) value ................................................................. 22
7.3.5 Setting an on-delay time lag ...................................................................................... 23
7.3.6 Setting an off-delay time lag ...................................................................................... 24
7.4 Configuration (ConF) sub-function ..................................................................................... 25
7.4.1 Entering Configuration sub-function .......................................................................... 25
7.4.2 Selecting pH or ORP mV measurement .................................................................... 26
7.4.3 Reverting to factory default settings .......................................................................... 26
7.5 Calibration (CAL) sub-function ........................................................................................... 27
7.5.1 Entering Calibration mode from Advanced Set-up mode .......................................... 27
8 TECHNICAL SPECIFICATIONS 28 9 ACCESSORIES 29 10 GENERAL INFORMATION 30
10.1 Warranty ............................................................................................................................. 30
10.2 Packaging ........................................................................................................................... 30
10.3 Return of Goods ................................................................................................................. 30
10.4 Guidelines for Returning Unit for Repair ............................................................................ 30
11 APPENDICES 31
11.1 Appendix 1 .......................................................................................................................... 31
11.2 Appendix 2 .......................................................................................................................... 32
11.3 Appendix 3 .......................................................................................................................... 33
o
C) sub-function .......................................................................... 17
Page 6
Page 7
Page 8
1 INTRODUCTION
1.1 Description of Unit
Thank you for purchasing ¼ DIN 800 series pH/ORP process controllers. This unit is used for measuring either pH or ORP parameter one at a time, and the operational mode is switchable from the menu. You can use this unit to measure pH or ORP with limit control. This controller has many user­friendly and safety features which include:
Menu-driven program that simplifies set-up
Built-in non-volatile memory backup to ensure that calibration and
other information are not erased if power supply fails
Push-button two-point calibration and electrode offset adjustment from the keypad
Automatic temperature compensation (ATC)
Manual temperature compensation setting without the ATC probe, with
independent setting for calibration and process temperature
0 to 2000 second time delay adjustment on all relays – minimise false alarms
Separately adjustable high and low set point hysteresis (dead bands) prevent chattering of relays around the set points
Asymmetrical/symmetrical input for pH/ORP operation
Large dual display LCD for easy reading with clear multiple
annunciators, alarm status, operational and error messages
Two switching contacts as set-point triggering relays
Hold function freezes output current (4/20mA) and control relays
LED indicators signal control activities to monitor controller status from a
distance
1
Page 9
Protection against electromagnetic interference – Input and output
are free from external inference
1.2 Applications
Use this controller in panel mounted enclosures for applications such as water treatment and monitoring, galvanic-decontamination, chemical processing food processing, clean or waste water control and neutralisation
process
.
2
Page 10
2 ASSEMBLY AND INSTALLATIONS
2.1 Measurement and Control System
A typical measurement system consists of:
a pH/ORP process controller
a pH/ORP combination electrode with integrated or separate temperature
sensor Pt 100/1000,
an immersion, flow or process assembly with or without a potential matching pin (PMP)
a final control element such as pump or valve
a recorder
Flow Assembly
PH Controller
Chart Recorder
Process Assembly
with Electrode
Measurement Cable
3
Power Mains
(220/110 VAC)
Page 11
(
)
(
)
2.2 Unit Dimensions
MountingCut-Out
1mm
96
max. 45
92 + 0.5
Flat Gasket
To be InsertedByCustomer
Note: TheTaped Corners Have to BeOn Top
92 +0.5
92
56
32
max. 175
The field-tested control panel housing is 96 x 96 mm; with protection class IP 54 (front).
4
Page 12
A
A
A
reforcur
r
V
3 ELECTRICAL CONNECTIONS
3.1 Connection Diagram
S/
S/S /
pH
BNC
12 13
14
P E / S
151715
Si g na l Ou tputHold Input
p H
-
m
+
6
1
P ow e r M a i n s
C:
2 5
1
P E
NL
1 8 1 920 2 12 2
P
R elay
oten
P t1 0 0
36
4
P E / S
ti a l M atc h ingPin (PMP)
Relay B
7 89
Signal InputpH/ORP
10
11
* )ind i c ated c o ntactpositions a
entless condition s
ENSURE that the power cable is physically separated from the power supply.
5
Page 13
Temp.
pH
PE
N
S
L
1
3
2
0/PA
21
20
22
19 18
S
BNC
Symmetrical high-impedance
connections
Measuring Cable
Potential
Matching Pin (PM)
pH Combination
Electrode
Temp.
pH
NLS
PE
0/PA
21
22
1
3
2
20
19 18
BNC
Asymmetrical connections
Measuring Cable
Potential
Matching Pin (PM)
pH Combination
Electrode
6
Page 14
3.2 Back Panel
The back panel consists of two connectors. The first connector is the 17-way PCB edge connector and the other is the 5-way connector.
ENSURE that the power cable is physically separated from the power supply.
Connection for the 17-way screw terminals (from left to right):
1.AC mains live wire 10.No connection
2.AC mains neutral wire 11.No connection
3.AC mains protective earth wire 12.No connection
4.Low set relay resting position (NC) 13.Hold function switch terminal 1
5.Low set relay common 14.Hold function switch terminal 2
6.Low set relay working position (NO) 15.No connection
7.High set relay resting position (NC) 16.4 - 20 mA for -ve connection
8.High set relay common 17.4 - 20 mA for +ve connection
9.High set relay working position (NO)
Connections for the 5-way screw terminals:
18. Pt100/Pt1000 lead 1 terminal
19. Pt100/Pt1000 sense lead terminal
20. Pt100/Pt1000 lead 2 terminal
Note: If using a two-wire RTD, short terminal 19 to terminal 18. Pt 100/Pt 1000 is selectable via an internal jumper. Factory default is Pt100. See Appendix 1 for directions on switching the RTD type.
21. pH/ORP (potential matching pin)
21. pH/ORP (shield)
7
Page 15
/
/
A
A
/
FUSE 250VAC
100mA
LNPE
(F)
RELAY
Pt100 Pt1000
REL
YB
0V PAL
S
NC NCNC
HOLD
NC
+
-
8
Page 16
4 OVERVIEW
4.1 Keypad and Display
4.1.1 Keypad
Perform rapid 2-point calibration and view electrode status
Allows entry to Set up mode
Select individual functions within the function group of Set up
mode
Store input data in the Set up mode
Start calibration in the calibration mode
Select various function groups in the Set up mode.
Set parameters and numerical values in sub functions of Set
up mode
If pressed continuously, the setting speed increases
Return to the Measurement mode when both keys are pressed together
9
Page 17
4.1.2 Display
The LCD display features two numerical displays that show status messages and measured values for easy, quick reference. The display provides short-text information for setting parameters and configuration.
SETUP
READY
HOLD
MEAS
CAL
7.00
25.0 ATC
CON
pH
o
C
MEAS: Measurement mode
SETUP: Set-up mode of function groups
CAL: Calibration mode of pH/ORP
READY: Comes on after a successful
calibration
HOLD: Relay position and current output are frozen
ATC: Comes on in the ATC mode. Disappears in the Manual temperature Compensation mode.
“ATC” flashes if the temperature probe is
faulty in its ATC mode
Display for RELAY A/B. Green LED indicates measured value within limit while RED LED indicates measured value outside limit.
Alarm display if limit value overshoot or the ATC probe fails.
10
Page 18
4.2 Function Groups
The main function and sub-function groups are organised in a matrix format for configuration and selection of parameters. The main function groups are:
1) Offset adjustment (OFS)
2) Temperature Measurement / compensation settings (Set
o
C)
3) Control relay 1 configuration (SP1)
4) Control relay 2 configuration (SP2)
5) Configuration (ConF)
6) Calibration (CAL pH)
4.3 Control Concept
The main function and sub-function groups are organised in a matrix format as shown below. These functions can be accessed via the front keypad for configuration and selection of parameters.
11
Page 19
5 MEASUREMENT
5.1 Display in Measurement mode
When the controller is initially powered on, it automatically enters into the Measurement mode after the large dual LCD displays all segments briefly.
The upper display shows the measured pH or ORP value, while lower display shows either temperature value if controller is set for pH measurement or “OrP” if it is set for ORP measurement.
Annunciators at right side of display indicate whether controller is set for pH or mV measurement.
Similarly annunciators or icons at top or left side of display shows current status of controller, e.g. “MEAS”, “SETUP”, “CAL”, “READY”, etc. In addition, error messages also guide user in time of any faulty conditions as indicated by “ERR” or icons flashing.
5.1.1 Check electrode performance
To read current electrode slope and offset values:
1) Press CAL key followed by ENTER key. The upper display shows electrode slope. The lower display reading shows the pH reading at 0 mV.
2) Press Δ and key s to return to Measurement mode.
12
Page 20
6 CALIBRATION
Direct Calibration from Measurement mode is possible via the CAL key. Calibration procedure can also be accessed from Advanced Setup mode.
6.1 pH Calibration
This unit features five preset buffer values (1.00, 4.00, 6.86 or 7.00, 9.00, 9.18 and 10.00) for quick, two-point auto calibration. When calibrating this instrument, standard pH buffer solutions must be used.
1) Enter Calibration mode. While in Measurement mode, push CAL key. Upper and lower display reads “CAL” & “pH”. Push ENTER or CAL key again. The controller displays electrode status based on previous calibration. Electrode slope and offset information is displayed. NOTE: If display reads “CAL OrP”, see section 7.4 for procedures on how to switch from ORP to pH readings.
2) Press ENTER key again to begin calibration. The “CAL” indicator appears at top of display and
SETUP
HOLD
READY HOLD
HOLD
CAL
6.88
6.96
7.00
CAL
57.2
CAL
mV
pH
pH
pH
buffer indicator appears in bottom left corner of display. Upper display is on-line pH value as measured by electrode. Lower display indicates the
first compulsory calibration value, pH 7.00 (or
6.86). Immerse clean electrode in buffer solution. In ATC, temperature probe must also be immersed in buffer solution. Likewise for symmetrical mode, immerse potential matching pin in buffer.
13
SETUP
HOLD
READY HOLD
CAL
4.02
CAL
59.0
7.00
pH
pH
mV
pH
Page 21
3) Press ENTER key to start calibration at pH 7 (or pH 6.86). The electrode indicator and CAL indicator both flash. The controller automatically adjusts reading to match the buffer value.
4) The lower display will now show its next lower ‘pH’ buffer. Use Δ and keys to select second buffer from one of 5 remaining preset values. Remove electrode from first buffer, wash and then immerse into second buffer. Once upper display stabilises press ENTER key, electrode and CAL indicator will flash. The controller adjusts reading to the buffer.
5) After calibrating to a second buffer value, controller automatically displays electrode slope and offset. The zero point and slope are re-determined after each calibration.
6) Press ENTER key. If calibration mode was accessed using CAL key, controller will return to Measurement mode. If calibration mode was accessed from Advanced Set-up mode, controller will return to sub-function menu. Note: If there is a calibration error, controller displays “ERR”. If this happens, push both
Δ and keys (escape) to restart calibration beginning from step 1.
Note: When calibrating with manual temperature compensation, controller
automatically changes from preset process temperature to calibration temperature. After leaving Calibration mode, controller switches back to process temperature (for setting calibration temperature and process temperature, see section 7.2.3).
14
Page 22
6.2 ORP – mV Calibration
This mode allows one-point calibration.
1) Enter Calibration mode. While in Measurement mode, push CAL key. Press ENTER key. The upper and lower display reads “CAL OrP”. Note: If upper and lower display read “CAL PH”, see section 7.4 for procedures on how to switch from pH to ORP mV readings.
2) Place sensor in ORP solution.
3) Press ENTER key to begin calibration. “CAL” indicator appears on display. Upper display shows current mV output of electrode without any offset adjustment.
SETUP
HOLD
CAL
OrP
CAL
HOLD
226
mV
4) Press Δ or keys to adjust mV value to that of standard solution.
5) To confirm, press ENTER key. The controller calculates the correction factor and returns to Measurement mode.
15
Page 23
7 ADVANCED SET UP MODE
7.1 Electrode Offset (OFS) sub-function
This option is available only in the pH mode. An allowance of ±120mV is allowed to correct for electrode deviations. This feature is useful for prolonged on-line applications.
The controller will add or subtract the value from the measured pH and display the correct value. Should the offset value be too high, consider performing maintenance on the electrode followed by calibration. If the calibration is unsuccessful, please replace the electrode.
SETUP
HOLD
Ofs
SETUP
HOLD
5) Press ENTER key. The SETUP annunciator appears at the top of display.
-0.05
7.00
1) Take a sample from the system and note pH reading displayed on controller.
2) Measure pH of sample using a calibrated pH tester, hand-held meter, or bench meter. Record the correct pH value.
3) Enter Advanced set-up mode. Press ENTER.
4) Press Δ or keys to scroll until upper display shows OFS.
Lower display shows on-line measured pH value. Upper display shows current offset value.
6) Press Δ or keys until pH value coincides with correct pH value noted in Step 2. As the Δ or keys is pressed to adjust pH value, the ‘offset’ value in upper display also changes.
7) Press ENTER key to enter offset value.
16
Page 24
Note: Offset value is reset during full calibration. See section 6 for full calibration instructions. Note: The offset parameter is blanked out during ORP operation.
7.2 Setting temperature (Set oC) sub-function
7.2.1 Selecting automatic or manual temperature compensation
1) Enter Advanced Set-up mode by pressing ENTER key.
2) Press Δ or keys to scroll until display shows “Set
o
C”.
3) Press ENTER key. The lower display shows “Atc”; upper display shows “on” or “oFF” depending on whether or not ATC is selected.
4) Press Δ or keys to toggle between ATC on and off.
5) Proceed with additional Advanced Set-up procedures (press ENTER again) or return to Measurement mode by pressing Δ and keys
SETUP
HOLD
SETUP
SEt
o
on
HOLD
Atc
SETUP
HOLD
oFF
Atc
(escape) simultaneously.
17
Page 25
7.2.2 Temperature calibration (ATC mode only)
1) Select “ATC on” as described above in Section 7.2.1.
SETUP
2) Press ENTER key. The upper display
HOLD
on
indicates current temperature offset. The current measured temperature is shown in the lower display.
3) Compare current measured temperature
SETUP
HOLD
0.0
25.0
ATC
o
C
on the controller display to a thermometer known to be accurate. Note down the correct temperature value.
4) Press Δ or keys to scroll the lower display to match the correct value. Upper display will now show the offset value. You can offset temperature up to ±
o
5
C.
5) Press ENTER key to confirm your selection.
6) Continue with additional Advanced Set-up procedures, or return to the Measurement mode by pressing Δ and keys (escape) simultaneously.
18
Page 26
7.2.3 Setting manual temperature compensation
SETUP
HOLD
SETUP
HOLD
SETUP
HOLD
oFF
Atc
25.0
P.C
25.0
C.C
Note: This option is not available when the controller is set for ATC operation.
For manual temperature compensation, two different temperatures: process and calibration, can be input independently. This allows calibration at a temperature other than the process temperature. Example: setting a calibration temperature of 25 using standard buffer solutions at 25
o
C allows calibration
o
C, even if
the process temperature is different.
1) Select “ATC oFF” as described above in section 7.2.1
2) Press ENTER key. Upper display shows current process temperature and lower display shows “P.
o
C” to indicate process temperature.
3) Press Δ or keys to adjust the process temperature value, between –9.9
o
and 125
C.
4) After setting the process temperature value, press ENTER key. Upper display shows current calibration temperature and lower display shows “C.
o
C”.
5) Press Δ or keys to adjust the calibration temperature value, between –9.9 and 125
o
C.
6) Press ENTER key to confirm.
7) Continue with additional Advanced Set-up procedures, or return to Measurement mode by pressing Δ and keys (escape) simultaneously.
19
Page 27
7.3 Control Relay A/Control Relay B (SP1/SP2) sub-function
The SP1 option sets operating parameters for Relay A; the SP2 option sets operating parameters for relay B. Since these groups have the same set-up parameters, they are described together.
7.3.1 Entering Set point 1 (Set point 2) sub-function
SETUP
1) Enter Advanced Set-up mode. Push the
HOLD
SP1
ENTER key.
2) Press Δ or keys to scroll until upper display
SETUP
HOLD
4.00
SP1
pH
shows SP1 (SP2).
7.3.2 Selecting set point values
This lets you choose the value that will cause your
SETUP
HOLD
Lo
SP1
controller to activate the respective relays.
1) Follow directions in 7.3.1 to enter Control Relay mode. If you are in this mode, skip to step 2.
2) Press ENTER key. The upper display shows current set point value and lower display shows SP1 (SP2).
3) Press Δ or keys to enter value for Set point 1 (Set point 2).
4) Press ENTER key to confirm.
5) Proceed to 7.3.3 step 3, or return to Measurement mode by pressing Δ and keys simultaneously (escape).
20
Page 28
7.3.3 Choosing High or Low set points
Select a “lo” to activate relay when the pH/ORP value undershoots the set point; select a “Hi” to activate relay when the value overshoots the set point. Using both SP1 and SP2, the following configuration for the set points can be selected: lo/lo, lo/hi, hi/lo, or hi/hi.
1) Follow directions in 7.3.1 to enter Control Relay mode.
2) Press ENTER key until upper display shows Lo or Hi and lower display shows SP1 (SP2).
3) Press Δ or keys to select low (lo) or high (hi) set point for SP1 (SP2).
4) Press ENTER key to confirm.
5) Proceed to 7.3.4 step 3, or return Measurement mode by pressing Δ and keys simultaneously (escape).
21
Page 29
7.3.4 Selecting a hysteresis (dead band) value
Hysteresis prevents rapid contact switching if your value is fluctuating near set point. It does this by overshooting the set point value to a specified hysteresis value. You can set the hysteresis value from 0.1 to 1.0 pH, 10 to 100mV. Example: You have set your high set point at pH 9.0 and your hysteresis limit value at pH 0.5. If your measured value overshoots high set point pH 9.0, the controller’s relay activates, which in turn activates external devices such as a pump or valve. The actions of the external device will cause the solution’s pH to drop. When the pH drops to 8.5 pH, the relay and hence the pump will switch off.
1) Follow directions in 7.3.1 to enter Control Relay mode.
2) Press ENTER key. Scroll with Δ or keys
SETUP
HOLD
0.50
pH
until upper display shows the hysteresis (dead band) value and lower display shows “HYS”.
3) Press Δ or keys to enter your hysteresis value for Set point 1 (Set point
2). Your controller will activate at the value you select.
4) Press ENTER key to confirm your selection.
5) Proceed to 7.3.5 step 3, or return to Measurement mode by pressing Δ and
keys simultaneously (escape). NOTE: Please refer to Appendix 3 for a graphical representation of the
Hysteresis.
HYS
22
Page 30
7.3.5 Setting an on-delay time lag
You can set as time delay for each relay, which stops the relay from switching on the moment the set point is exceeded. This controller lets you set a 0 to 2000 second time delay before the relay activates.
SETUP
HOLD
0
1) Follow directions in 7.3.1 to enter Control
On.d
Relay mode.
2) Press ENTER key. Scroll with Δ or keys until upper display shows “on delay” time and lower display shows “On.d”.
3) Press Δ and keys to enter on-delay time for Set point 1 (Set point 2). The controller will delay activation for the number of seconds (0 to 2000) you select.
4) Press ENTER key to confirm your selection.
5) Proceed to 7.3.6 step 3, or return to Measurement mode by pressing Δ and keys simultaneously (escape).
23
Page 31
7.3.6 Setting an off-delay time lag
You can set a time delay for each relay, which delays relay from switching off the moment the value reaches set point and hysteresis. Controller allows a timer delay from 0 to 2000 seconds time delay before deactivating.
SETUP
HOLD
0
1) Follow directions in 7.3.1 to enter Control Relay mode.
2) Press ENTER key. Scroll with Δ or keys until
OF.d
upper display shows “off delay” time and lower display shows “OF.d”.
3) Press Δ or keys to enter off-delay time for Set point 1 (Set point 2). The controller will delay activation for the number of seconds (0 to 2000) you select.
4) Press ENTER key to confirm your selection.
5) Continue with Advanced Set-up mode procedures, or return to Measurement mode by pressing Δ and keys simultaneously (escape).
24
Page 32
7.4 Configuration (ConF) sub-function
The options available in this sub-function allows the controller to be configured as a pH or ORP controller.
7.4.1 Entering Configuration sub-function
1) Enter Advanced Set-up mode. Press the ENTER key.
2) Press Δ or keys to scroll until upper display shows “ConF”.
SETUP
HOLD
ConF
SETUP
HOLD
PH
SY
SETUP
HOLD
PH
ASY
SETUP
HOLD
SETUP
HOLD
OrP
mV
SY
OrP mV
ASY
25
Page 33
7.4.2 Selecting pH or ORP mV measurement
Besides selecting the controller to function as a pH or ORP (mV) controller, it also allows the selection of the type of input: symmetrical or asymmetrical. Symmetrical mode is used in conjunction with a potential matching pin – in an electrically noisy environment. Please refer to section 3, page 6 for a graphical representation.
1) Follow directions in 7.4.1 to enter Configuration mode.
2) Press ENTER key. Scroll with Δ or keys until upper display shows mode of operation (pH or ORP mV) and lower display shows symmetrical (SY) or asymmetrical (ASY) input type.
3) Proceed to 7.4.3, or return to Measurement mode by pressing Δ and keys simultaneously (escape).
7.4.3 Reverting to factory default settings
Use this parameter to reset all settings to factory default. Changing from “no” to “YES” and pressing ENTER key resets all settings to factory default.
WARNING: If “Yes” is selected, all settings input will be overwritten!
1) Follow directions in 7.4.1 to enter Configuration
SETUP
HOLD
no
dEF
mode.
2) Press ENTER key. Scroll with Δ or keys until upper display shows “no” or” YES”, and lower display shows “deF” (default).
3) Press Δ or keys to select no or yes. Selecting “YES” and pressing ENTER key will overwrite all user-settings and automatically return to Measurement mode.
4) Press ENTER key to confirm your selection and return to Advanced Set-up mode, or return to Measurement mode by pressing Δ and keys simultaneously (escape).
26
Page 34
7.5 Calibration (CAL) sub-function
The calibration procedure in Advanced Set-up mode is identical to the procedure in the Calibration mode. The only difference is that controller will revert back to Set-up mode (instead of Measurement mode) after calibration is completed.
7.5.1 Entering Calibration mode from Advanced Set-up mode
1) Enter Advanced Set-up mode. Press ENTER key.
2) Press Δ or keys to scroll until upper display shows “CAL”.
3) See section 6 for complete calibration procedures.
27
Page 35
±
±
±
Ω
8 TECHNICAL SPECIFICATIONS
pH Range -2.00 to 16.00 pH Resolution 0.01 pH Relative Accuracy
mV Range - 1000 to +1000 mV
Resolution 1 mV Relative Accuracy
Temperature - 9.9 to + 125.0 oC
Resolution 0.1 oC Relative Accuracy
Sensor Pt 100 /Pt 1000 (jumper selectable) Temperature Compensation Auto / manual (reference at 25.0 oC) Set-point and Controller Functions Function (switchable) Set-point controller Controller characteristics Limit controller Pickup / Dropout delay 0 to 2000 sec. Switching pH hysteresis 0.1 to 1 pH Switching ORP hysteresis 10 to 100 mV Contact outputs, controller 2 potential-free change-over contacts Switching voltage / current / power max. 250 VAC / max. 3A / max. 600 VA Electrical Data and Connections Power Requirements 110 / 220 VAC (jumper selectable) Frequency 48 to 62 Hz Power Consumption max. 0.75 VA Signal Output / Load
pH / ORP input BNC Connection terminal Terminal blocks 5-pole / 17-pole, removable Mains fuse / fine wire fuse slow-blow 250 V / 100 mA EMC Specifications Emissions According to EN 50081-1 Susceptibility According to EN 50082-1
Environmental Conditions (For Indoor Use) Altitude Up to 2000 m Maximum Relative Humidity 80% up to 31 oC decreasing linearly to 50% at 40 oC Power Supply Mains supply voltage fluctuations not exceeding + 10% of
Transient Overvoltages Category II Pollution Degree Category II Ambient temp. operating range 0 to +50 oC
Mechanical Specifications Dimensions (control panel housing - L x H x W) 175 x 96 x 96 mm Weights (control panel housing) max. 0.7 kg Material ABS with polycarbonate (front housing) Insulation (Front / Housing) IP 54 / IP 40
0.01 pH
1 mV
0.5 oC
4 to 20 mA, galvanically isolated / max. 500
the nominal voltage
28
Page 36
9 ACCESSORIES
Eutech Instruments
Order Code
ECARTSO05B ECARHTTSO05B ECARGTSO05B
EC100GTSO05B
ECARTSOHF05B ECHTPTTSO05B ECHTAUTSO05B
EC100GTSO20B
EC100GTSO10B
Product Description
Ryton-body pH Combi Electrode with 5 m cable with BNC Connector (No ATC).
Ryton-body pH Combi Electrode with 5 m cable with BNC Connector (No ATC).
Ryton-body pH Combi Electrode with 5 m cable with BNC & connector for PMP (No ATC).
Ryton-body pH Combi Electrode with Pt 100 RTD (ATC) & 5 m cable with BNC & connector for PMP & Pt 100.
Ryton-body pH Combi Electrode without ATC & 5 m cable with BNC connector. HF Resistant Glass.
Ryton-body ORP Platinum Electrode with 5 m cable with BNC & connector for PMP (No ATC).
Ryton-body ORP Gold Electrode with 5 m cable with BNC & connector for PMP (No ATC).
Ryton-body pH Combi Electrode with Pt 100 RTD (ATC) & 20 m cable with BNC & connector for PMP & Pt 100.
Ryton-body pH Combi Electrode with Pt 100 RTD (ATC) & 10 m cable with BNC & connector for PMP & Pt 100.
Note: Above pH/ORP electrodes can withstand a pressure up to 6 bar
(except for ECARHTTSO05B – 9 bar). These electrodes have integral 5 meter measuring cable. Please ask your authorised distributor or dealer for the prices, extension cables, tee joints, electrode assembly and buffer solutions.
29
Page 37
10 GENERAL INFORMATION
10.1 Warranty
Thermo Scientific warrants this product to be free from significant deviations in material and workmanship for a period of one year from the date of purchase. If repair is necessary and has not been the result of abuse or misuse within the warranty period, please return by freight pre-paid and amendment will be made without any charge. Our Customer Service Dept. will determine if product problem is due to deviations or customer abuse. Out of warranty products will be repaired on a charge basis.
10.2 Packaging
The instrument is packaged in a corrugated box with a warranty card, instruction manual and the following accessories:
17-way and 5-way (right-angled) terminal block [1 unit each]
side threaded rod with catch [2 units]
receptacle cable lug [1 unit]
rubber gasket [1 unit]
10.3 Return of Goods
Authorisation must be obtained from our Customer Service Dept. to issue a RMA (Return of Material Authorisation) number before returning items for any reason. When applying for authorisation, please include data requiring the reason of return. Items must be carefully packed to prevent damage in shipment and insured against possible damage or loss. Thermo Scientific will not be responsible for any damage resulting from careless or insufficient packing.
Warning: Shipping damage as a result of inadequate packaging i s the
user/distributor’s responsibility, whoever applicable. Please follow the guidelines below before shipment.
10.4 Guidelines for Returning Unit for Repair
Use the original packaging material, if possible when shipping the unit for repair. Otherwise wrap it with bubble pack and use a corrugated box for better protection. Include a brief description of any faults suspected for the convenience of Customer Service Dept., if possible.
30
Page 38
11 APPENDICES
11.1 Appendix 1
Before opening the unit, ENSURE that the power cable is physically separated from the power supply.
Jumper Positions - Internal to the controller
JP 1 Selects the input voltage 220 VAC. JP 2 Selects the input voltage 110 VAC. JP 3 Solder bridge selects between Pt100 and Pt1000.
Fuse Note that there is a fuse (slow-blow 100mA) internal to the
controller. Replace fuse with the recommended type only.
31
Page 39
11.2 Appendix 2
The following table shows the various pH values at different temperature of the solution during calibration.
Temperature
o
C)
(
0 1.67 4.01 6.98 7.12 9.47 10.32
5 1.67 4.01 6.95 7.09 9.38 10.25 10 1.67 4.00 6.92 7.06 9.32 10.18 15 1.67 4.00 6.90 7.04 9.27 10.12 20 1.68 4.00 6.88 7.02 9.22 10.06 25 1.68 4.01 6.86 7.00 9.18 10.01 30 1.69 4.01 6.85 6.99 9.14 9.97 35 1.69 4.02 6.84 6.98 9.10 9.93 40 1.70 4.03 6.84 6.97 9.07 9.89 45 1.70 4.04 6.83 6.97 9.04 9.86 50 1.71 4.06 6.83 6.97 9.01 9.83 55 1.72 4.08 6.83 6.97 8.99 9.81 60 1.73 4.10 6.84 6.98 8.96 9.79 70 1.74 4.12 6.85 6.99 8.92 9.76 80 1.77 4.16 6.86 7.00 8.89 9.74 90 1.80 4.20 6.88 7.02 8.85 9.73
pH 1.68
(oxalate)
pH 4.01
(phthalate)
pH 6.86
(tetraoxalate)
pH 7.00
(neutral
phosphate)
pH 9.18
(tetra borate )
pH 10.01
(carbonate)
32
Page 40
11.3 Appendix 3
Simple Explanation on the Function of Hysteresis
The controller relay activates when the set-point is reached. In the reverse direction, it does not de-activate when the value reaches the set-point. Instead, it continues to be active till the value reaches the amount set by the Hysteresis band.
S P 1 S e t t o L O S P 2 S e t t o H I
4 . 0 4 .5 9.5 1 0 . 0
S P 1
F O R W A R D D I R E C T I O N
R E V E R S E D I R E C T I O N
R E L A Y O N
R E L A Y O F F
7.0
S P 2
H Y S T E R E S I S B A N D ( D E F A U L T = 0 . 5 p H )
33
Page 41
CALIBRATION RECORD
Calibration Date Electrode Slope Electrode Offset
Page 42
Page 43
Page 44
Water Analysis Instruments
North America
166 Cummings Center Beverly, MA 01915 USA Toll Free: 1-800-225-1480 Tel: 1-978-232-6000 Dom. Fax: 1-978-232-6015 Int’l Fax: 978-232-6031
Europe
P.O. Box 254, 3860 AG Nijkerk Wallerstraat 125K, 3862 CN Nijkerk, Netherlands Tel: (31) 033-2463887 Fax: (31) 033-2460832
Asia Pacific
Blk 55, Ayer Rajah Crescent #04-16/24, Singapore 139949 Tel: 65-6778-6876 Fax: 65-6773-0836
www.thermo.com/process
© 2009 Thermo Fisher Scientific Inc.
All rights reserved. Thermo Fisher Scientific Inc.
68X216813 Rev 5
Loading...