Thermo Fisher Scientific pH 600-RF User Manual

Page 1
Instruction Manual
Alpha pH 600
RF Transmitter System
pH / ORP / ºC / ºF
68X450201 | Rev.0 | Oct 2008
Page 2
Page 3
This manual serves to explain the use of the Alpha pH 600 Series RF transmitter. It functions in two ways, firstly as a step by step guide to help you to operate the transmitter. Secondly, it serves as a handy reference guide. It is written to cover as many anticipated applications of the transmitter as possible. If there are doubts in the use of the transmitter, please do not hesitate to contact the nearest Authorized Distributor.
Thermo Scientific will not accept any responsibility for damage or malfunction to the transmitter caused by improper use of the instrument.
Remember to fill in the guarantee card and mail it back to your authorized distributor.
The information presented in this manual is subject to change without notice as improvements are made, and does not represent a commitment on the part of Thermo Scientific.
Copyright © 2008
All rights reserved
Page 4
Page 5
TABLE OF CONTENTS
1 INTRODUCTION 1
1.1 Before You Begin ........................................................................................... 1
1.2 Intended Use.................................................................................................. 1
1.3 Safety Instructions.......................................................................................... 2
1.4 Taking Out of Service / Correct Disposal of the Unit .....................................2
2 GETTING STARTED 3
2.1 Description of Instrument ............................................................................... 3
2.2 Measurement System .................................................................................... 4
2.3 Connecting Peripherals ............................................................................. 5
2.4 Instal lati on .................................................................................................... 9
2.5 Di spla y & Ke y pad ....................................................................................... 10
3 OPERATION 13
3.1 Measurement mode ....................................................................................... 13
3.2 Menu Overview ..............................................................................................14
4 CALIBRATION MODE 15
4.1 Preparing the Transmitter & Electrode for Calibration................................... 15
4.2 Entering pH/ORP Calibration Mode ............................................................... 15
4.3 pH Calibration ................................................................................................ 16
4.4 Check Calibration........................................................................................... 21
4.5 ORP Calibration .............................................................................................21
4.6 Temperature Calibration ................................................................................22
5 SETUP MODE 23
5.1 Enter Setup mode ..........................................................................................23
5.2 Electrode Offset Settings ...............................................................................24
5.3 Temperature Settings..................................................................................... 25
5.4 Buffer Selection Settings................................................................................ 27
5.5 Configuration Settings.................................................................................... 28
5.6 Viewing Electrode Properties......................................................................... 31
6 TECHNICAL SPECIFICATIONS 32
7 LIST OF ACCESSORIES 33
7.1 Thermo Scientific ........................................................................................... 33
7.2 Eutech Instruments ........................................................................................ 34
8 TROUBLESHOOTING 35
9 GENERAL INFORMATION 36
9.1 Warranty......................................................................................................... 36
9.2 Return of Goods............................................................................................. 36
10 APPENDICES 37
10.1 Appendix 1– pH Buffer Values at Various Temperatures.............................. 37
10.2 Appendix 2 – Abbreviations Used in LCD...................................................... 38
Page 6
Page 7

1 INTRODUCTION

1.1 Before You Begin

Thank you for choosing the Alpha pH 600 Series RF Transmitter.
The construction of Alpha pH 600 Series employs leading edge technology and complies with safety regulations currently in force. Notwithstanding this, improper use could lead to hazards for the user or a third-party, and/or adverse effects on the plant or other equipment. Therefore, the operating instructions must be read and understood by the persons involved before working with the pH Transmitter.
The instruction manual must always be stored close at hand, in a place accessible to all people working with the transmitter.
If you have questions, which are not or insufficiently answered in this instruction manual, please contact your authorized supplier. They will be glad to assist you.

1.2 Intended Use

The Alpha pH 600 Series RF Transmitters are intended solely for pH or ORP and temperature measurements, as described in this instruction manual. This transmitter system is currently designed for use in 923 to 928 MHz license-free ISM radio band transmission, which is licensed free in the following countries only.:
1. Argentina
2. Canada
3. Chile
4. Costa Rica
5. Dominican Republic
Users are strongly encouraged to check with their local agencies on the license status of this radio band transmission before installing this instrument.
Any other use, or use not mentioned here, that is incompatible with the technical specifications is deemed inappropriate. The operator is solely responsible for any damage arising from such use.
Other prerequisites for appropriate use include complying with:
the instructions, notes and requirements set out in this instruction manual.
all local safety regulations concerning safety at work.
6. Mexico
7. Peru
8. Puerto Rico
9. United States
10. Uruguay
11. Australia
12. Brazil
13. China
14. Hong Kong
15. Malaysia
16. Singapore
17. Taiwan
18. Thailand
19. Vietnam
- 1 -
Page 8
all information and warnings in the documentation dealing with the
products used together with this pH / ORP Transmitter (housing, sensors, etc.).
the local environmental and operational conditions.

1.3 Safety Instructions

The Alpha pH 600 RF Transmitter should be installed
and operated only by personnel familiar with the instrument and who are qualified for such work.
A defective pH 600 RF Transmitter must neither be
installed nor put into service.
The Alpha pH 600 RF Transmitter must only be
operated under the specified operating conditions (see section 6).
The Alpha pH 600 RF Transmitter must not be repaired
by the customer.
No modifications to the Alpha pH 600 RF Transmitter
are allowed. The manufacturer/supplier accepts no responsibility for damage caused by unauthorized modifications. The risk is borne entirely by the user.

1.4 Taking Out of Service / Correct Disposal of the Unit

Taking out of Service
First disconnect the unit from the power supply and then undo all electrical
connections.
Remove the unit from the wall.
Correct Disposal of the Instrument
When the Alpha pH 600 Transmitter is permanently taken out of service, obey the local environmental regulations for correct disposal or send the instrument to your local distributor, they will take care of proper disposal.
- 2 -
Page 9

2 GETTING STARTED

2.1 Description of Instrument

The Alpha pH 600 RF Transmitter is used for measuring pH or ORP and temperature values. The pH values can be measured using industrial combination pH sensors; the ORP values can be measured using industrial ORP sensor; the temperature values can be measured using 3-wire Pt100 / Pt1000 sensors. The Transmitter can be used for applications such as water treatment and monitoring, galvanic-decontamination, chemical processing, food processing, clean or wastewater control and neutralization processes.
This pH Transmitter has many user-friendly and safety features which include:
Easy to calibrate and set up
Built-in non-volatile memory to ensure that calibration and other
information are not erased if power supply fails
Menu-driven program that simplifies setup
1 or 2 point calibration with selectable USA or NIST standard pH buffer set
Automatic temperature compensation (ATC)
Manual temperature compensation setting without the ATC probe, with
independent setting for calibration and process temperature
Large dual display LCD for easy reading with clear multiple annunciators,
operational mode indicators and error indicators.
- 3 -
Page 10

2.2 Measurement System

A typical measurement system consists of:
Up to 10 Alpha pH 600 Transmitters
Alpha Link 600 Data Acquisition Software (Installation instructions included
separately)
A Wireless RF Transceiver
One pH sensor or ORP sensor with integrated or separate Pt100/Pt1000
temperature sensor per Transmitter
Appropriate measurement cables
An immersion, flow or process assembly
A computer, installed with the Alpha RF Transmitter System Software
Wireless
Housing and Sensors
Measurement
Cable
50m
50m
Power Adaptor (9 V DC)
- 4 -
100m
100m
100m
100m
Page 11

2.3 Connecting Peripherals

2.3.1 Connection Terminals

Remove Back Cover:
Remove the screws from the four corners at the back of the pH Transmitter.
Remove the back cover. The connectors are exposed on the back PCBA as shown in the Figure 1 below.
Connectors:
J202 – 9V DC power
J204 - pH electrode & Temperature probe connections (wiring has to
be done in the detachable connector:
1. Pin 1 : pH Sensor
2. Pin 2 : pH Reference
3. Pin 3 : No Connection
4. Pin 4 : No Connection
5. Temp I/P : Pt 100 Compensate
6. Temp Sens : Pt 100 Sense
7. Temp Gnd : Pt 100 GND
8. PMP : Potential Matching Pin
J202
Pin 1
Pin 2
Pin 3
Pin 4
Temp I/P
Temp Sens
Temp Gnd
PMP
J204
Figure 1: Outer Side of Back PCBA
- 5 -
Page 12
p

2.3.2 Switching Between PT100 & PT1000 Temperature Sensors

The Alpha pH 600 transmitter supports both Pt100 & Pt1000 (2-wire or 3-wire) temperature sensors. The default factory setting is Pt100. If you need to use Pt1000 temperature sensor, you have to change the jumper setting (J201) as described below.
Remove Back Cover:
Remove screws from the four corners at the back of the pH Transmitter. Remove the back cover.
Remove Back PCBA:
Remove the screw located center of the back PCBA (Figure 1). Detach the back PCBA from the Transmitter. Turn over the back PCBA. Locate J201 jumper on the inner side of the back PCBA as shown in Figure 2 below.
Figure 2: Inner Side of Back PCBA
Set Jum
For PT1000,short the top two pins
For PT100,short the bottom two pins
er J201: Set the J201 jumper to required sensor (Pt100 or Pt1000) type
- 6 -
Page 13

2.3.3 Connecting pH/ORP Electrode

1. If the pH/ORP electrode has a BNC connector, remove the BNC connector from the cable.
NOTE: Oakton Instruments offers an optional ‘BNC to Spade Lug adapter’ (Order code: 05994-90) that can be used with pH/ORP electrode without removing the BNC connector.
2. Strip the insulation of the cable so that the bare wires are exposed enough for connection as shown in Figure 3.
Insulation
Figure 3 : pH/ORP Electrode Cable
NOTE: Make sure to strip inner black layer (screen) to expose the
clear sheath.
3. Connect pH sensing wire to Pin 1 of J204 connector
4. Connect pH reference wire to Pin 2 of J204 connector
pH Referenc
Clear sheath
pH Sensing W
Screen
e Wire
ire
- 7 -
Page 14

2.3.4 Connecting Temperature Probe

For Automatic Temperature Compensated (ATC) pH readings, a 100Ω Pt RTD temperature probe (2-wire or 3-wire) can be connected to the Controller.
3-Wire Probe:
1. Connect PT100 compensate wire to Pin 5 of J204 connector
2. Connect PT100 sense wire to Pin 6 of J204 connector
3. Connect PT100 GND wire to Pin 7 of J204 connector
2-Wire Probe:
1. Short Pin 5 & 6 of J204 connector using a small piece of wire
2. Connect PT100 sense wire to Pin 6 of J204 connector
3. Connect PT100 GND wire to Pin 7 of J204 connector

2.3.5 Connect Potential Matching Pin (PMP)

If using an electrode with a PMP, connect the additional wire (PMP) from the pH electrode to Pin 8 of J204 connector. This liquid electrical reference wire is only used when the pH Controller is configured to ‘Symmetrical mode of operation’. Refer section 5.5 for details on symmetrical mode of operation.
- 8 -
Page 15

2.4 Installation

2.4.1 Mechanical Dimensions

CAL

2.4.2 Wall Mount

2
PH/ORP RF Transmitter
Alpha pH 600
ESC
ENT
Cover the catch slots at both sides with overlays
Pierce through holes at both
11
sides
- 9 -
Page 16
A

2.5 Display & Keypad

2.5.1 Display Overview

The Liquid Crystal Display (LCD) of Alpha pH 600 Transmitter has two alpha­numerical displays (Upper and a Lower).
Upper display: Measured pH, mV or relative mV value are displayed when the Transmitter is in normal operation (measurement) mode.
Lower display: Measured temperature value is displayed when the Transmitter is in normal operation (measurement) mode. In pH Calibration mode, standard pH buffer values are displayed here: while measured mV values are displayed in the ORP Calibration mode.
The two displays indicate function names, options & settings in Setup mode. Refer ‘Appendix 2 – Abbreviations Used in LCD’ for more details.
The LCD also consists of various mode indicators, status annunciators and unit of measurement indicators.
SETUP MEAS CAL
READY
ERR
-8.8.8.8
-8.8.8.
- 10 -
8
R.mV pH
°C
pH
TC
°F
Page 17
Mode Indicators
MEAS
SETUP
CAL
Measurement mode; blinks in Symmetric mode (Refer Section 3.1& 5.5 )
Setup mode (Refer Section 5)
Calibration mode (Refer Section 4)
Status Annunciators
READY
ATC
Appears when the reading is stable
Appears when Automatic Temperature Compensation (ATC) is enabled. Not visible when Manual Temperature Compensation (MTC) is enabled. Flashes if the temperature probe is faulty in its ATC mode. (Refer Section 5.3)
ERR
Appears when an error occurs (Eg. wrong key pressed)
Electrode annunciator. Appears when viewing electrode properties or during calibration error
Buffer annunciator. Appears in calibration mode
Units of Measurement Indicators
mV
Millivolt. Appears in ORP measurement / calibration modes (Refer Section 3.1, 4.5)
R.mV
pH
Relative Millivolt
Appears in pH measurement/calibration modes (Refer Section 3.1,4.3)
ºC
ºF
Temperature in Celsius (Refer Section 5.3)
Temperature in Fahrenheit (Refer Section 5.3)
- 11 -
Page 18

2.5.2 Key Functions

Alpha pH 600
ESC
Key Description
Brings directly into the Calibration entry mode. The security code must be selected to enter the calibration mode or view only mode.
CAL
While in pH Measurement mode, press CAL to enter pH Calibration entry mode.
While in mV Measurement mode, press CAL to enter mV Calibration entry mode.
To confirm your calibration values in Calibration mode.
To confirm selections in SETUP mode.
ENT
While in MEAS mode, pressing ENT takes you directly to SETUP entry mode. The security code must be selected to enter the
calibration mode or view only mode.
To scroll through the various submenus
To increment/decrement values or toggle between options
When pressed together, serves as escape function to return to
(in the SETUP/CAL modes)
MEAS mode from any point (CAL or SETUP modes).
CAL
PH/ORP RF Transmitter
ENT
- 12 -
Page 19

3 OPERATION

3.1 Measurement mode

When the pH Transmitter is powered on, the display shows all the LCD segments briefly, and then automatically enters into the Measurement mode.
MEAS
READY
pH
7.00
°C
25.0
The mode indicator ‘MEAS’ at the top of the display indicates that the pH Transmitter is in Measurement mode; ‘MEAS’ blinks when in Symmetric mode. The upper alpha-numerical display shows the measured pH or mV value, while the lower display shows the temperature value. The indicator “pH” or “mV” at the upper right side of the display indicates the current measurement mode. (Refer Section
5.5 for switching measurement modes)
NOTE:
From measurement mode you can access:
For more details, refer section 4 for Calibration mode and section 5 for Setup mode.)
To guarantee accurate readings, the measuring system (the pH Transmitter and the sensor) must be calibrated regularly.
Calibration mode (by pressing CAL key)
Setup mode (by pressing ENT key)
ATC
- 13 -
Page 20
V

3.2 Menu Overview

3.2.1 Alpha pH 600

MEAS
CAL
ENT
CAL
pH
ENT
pH
°C
ATC
SETUP
SETUP
SETUP
SETUP
ENT
ENT
ENT
ENT
Buffer Selectio
Calibrat
See Sectio
Offset Set
See Sect
Temperatur
See Sect
iew Electrode
See Secti
See Sectio
ion
n 4
tings
ion 5.2
e Settings
ion 5.3
Properties
on 5.6
n Settings
n 5.4
ENT
ENT
ENT
ENT
ENT
- 14 - - 15 -
Page 21

4 CALIBRATION MODE

4.1 Preparing the Transmitter & Electrode for Calibration

Before starting calibration, make sure that the pH Transmitter is in appropriate measurement mode (pH or ORP). When the Transmitter is powered on, it starts up with the measurement mode last used. For example, if the pH Transmitter is powered off in ORP measurement mode, it starts up in ORP mode when it is powered on. (Refer Section 5.5 for switching measurement modes)
Be sure to remove the protective electrode storage bottle or rubber cap of the electrode before calibration or measurement. If the electrode has been stored dry, wet the electrode in tap water for 10 minutes before calibrating or taking readings to saturate the pH electrode surface and minimize drift.
Rinse the electrode in de-ionized water after use, and store in electrode storage solution. If storage solution is not available, use pH 4.01 or 7.00 buffer solution.
Do not reuse buffer solutions after calibration. Contaminants in the solution can affect the calibration, and the accuracy of the measurements.
The pH Transmitter features two internationally recognized pH buffer standards. Select the buffer standard you require in the setup mode. (Refer Section 5.4) Available buffer options are:
USA buffers --- pH 4.01, 7.00 and 10.01
NIST buffers --- pH 4.01, 6.86 and 9.18

4.2 Entering pH/ORP Calibration Mode

While in Measurement mode press the CAL key.
1. The display prompts to enter the calibration security code. Press the or key to set the calibration security code to “11”, and then press the ENT key to confirm the calibration security code.
2. The display shows “CAL pH” and “CAL” indicator appears at the top of display and after 2 seconds delay it automatically leads to the cali on.
brati
Page 22
CAL
READY
pH
°C
CAL
ATC
To exit Calibration mode at any time press ▲ and
keys simultaneously (Escape).
The calibration is always carried out in the units of
measurement (pH or mV), selected in setup mode
HOLD
ENT
CAL
pH

4.3 pH Calibration

The pH Transmitter is capable of calibration of up to 2 points using USA or NIST pH buffer standards. All new calibration values will automatically override the existing calibration data. pH buffer values are referenced to 25
o
C. Ensure the measurement mode is set to pH. (Refer Section 5.5 for switching measurement modes). All pH buffer values have window of up to ± 1 pH tolerance during calibration. Calibration error can be occurred if the measured pH value exceeds this window. If you wish to abort the pH calibration, press and keys simultaneously and the pH Transmitter reverts to pH measurement mode.

4.3.1 For USA Standard Buffer

Make sure that the pH Transmitter is set to accept USA standard buffer in the Setup mode. (Refer Section 5.4 for Buffer Selection Settings) The factory default is USA standard. It is recommended that you perform a 2-point calibration at room temperature (25
o
C), starting with the first buffer at pH 7.00 followed by any other
buffer value (pH 4.01 or 10.01).
1-Point Calibration:
11 2
HOLD
CAL
ENT
HOLD
CAL
YREAD
3
pH
CAL
HOLD
pH
CAL
mV
ENT
°C
pH
MEAS
4
YREAD
pH
°C
ATC
- 16 -
Page 23
From pH measurement mode press CAL key to enter calibration mode
11
as described in section 4.2. The LCD shows ‘CAL PH’. Press ENT key to begin calibration.
Place the electrode in pH 7.00 buffer. Immerse the temperature probe in
2
the buffer solution if ATC mode is enabled. Immerse the potential matching pin in the buffer if symmetrical mode is enabled. The buffer
annunciator appears in LCD. Lower display shows pH 7.00 (USA standard buffer). Upper display shows the current uncalibrated pH reading. Allow the reading to stabilize. LCD shows ‘READY’ annunciator when the reading is stable. Press CAL key to confirm the reading.
The calibration is completed. The pH Transmitter re-calculates electrode
3
properties based on the calibration. The new slope (in mV) is shown in the upper display and normalized temperature is shown in lower display.
Press ENT key to exit from the calibration.
The pH Transmitter reverts to pH measurement mode.
4
NOTE:
Refer notes at the end of this section for additional information and possible error indicators of calibration process
2-Point Calibration:
11
CAL CAL
2 3
READY
ENT ENT ENT ENT
pH
pH
CAL
READY
4
pH
pH
CAL
55
mV
pH
MEAS
READY
pH
°C
ATC
3
CAL
READY
pH
ENT
CAL
pH
From pH measurement mode press CAL key to enter calibration mode
1
1
as described in section 4.2. The LCD shows ‘CAL PH’. Press ENT key to begin first calibration point.
Place the electrode in pH 7.00 buffer. Immerse the temperature probe in
2
the buffer solution if ATC mode is enabled. Immerse the potential matching pin in the buffer if symmetrical mode is enabled. The buffer
annunciator
appears in LCD. Lower display shows pH 7.00 (USA standard buffer). Upper display shows the current uncalibrated pH reading. Allow the reading to stabilize. LCD shows ‘READY’ annunciator
- 17 -
Page 24
3
4
55
NOTE:
when the reading is stable. Press ENT key to confirm the reading.
The pH Transmitter moves to the second calibration point. The lower display shows next standard buffer value (pH 4.01). Use and keys to select your second buffer from one of the preset values: pH 4.01 or
10.01. Remove the electrode from the first buffer, rinse and then immerse it into the second buffer. Allow the reading to stabilize. LCD shows ‘READY’ annunciator when the reading is stable.
Press ENT key (or CAL key) to confirm the reading.
The calibration is completed. The pH Transmitter re-calculates electrode properties based on the calibration. The new slope (in mV) is shown in the upper display and normalized temperature is shown in the lower display.
Press ENT key to exit from the calibration.
The pH Transmitter reverts to pH measurement mode.
To exit from pH calibration mode without confirming calibration, press and keys together.
When confirming the buffer measurement, if measured pH value is not within ±1.00pH from selected buffer value, the electrode
annunciator
blinks and ERR indicator appears in the display. This error can also occur if non-standard buffers are used or the electrode has worn out. If this happens, press both ▲and ▼ keys together to restart the calibration beginning from Step1.
When calibrating with manual temperature compensation, the meter automatically changes from the preset ‘process temperature’ to the ‘calibration temperature’. After leaving the calibration mode, the pH Transmitter reverts back to ‘process temperature’.

4.3.2 For NIST Buffer

Make sure that the pH Transmitter is set to accept NIST standard buffer in the Setup mode. (Refer Section 5.4 for Buffer Selection Settings) The factory default is USA standard. It is recommended that you perform 2-point calibration at room temperature (25
o
C), starting with the first buffer at pH 6.86 followed by any other
buffer value. (pH 4.01 or 9.18)
- 18 -
Page 25
1-Point Calibration:
141 2
HOLD
CAL
ENT ENT
From pH measurement mode press CAL key to enter calibration mode
11
as described in section 4.2. The LCD shows ‘CAL PH’. Press ENT key to begin calibration.
Place the electrode in pH 6.86 buffer. Immerse the temperature probe in
2
the buffer solution if ATC mode is enabled. Immerse the potential matching pin in the buffer if symmetrical mode is enabled. The buffer
annunciator appears in LCD. Lower display shows pH 6.86 (NIST standard buffer). Upper display shows the current uncalibrated pH reading. Allow the reading to stabilize. LCD shows ‘READY’ annunciator when the reading is stable. Press CAL key to confirm the reading.
The calibration is completed. The pH Transmitter re-calculates electrode
3
properties based on the calibration. The new slope (in mV) is shown in the upper display and normalized temperature is shown in the lower display.
Press ENT key to exit from the calibration.
The pH Transmitter reverts to pH measurement mode.
4
2-Point Calibration:
11
CAL CAL
2 3
READY
ENT ENT ENT
READY HOLD
CAL
pH
READY
pH
3
pH
CAL
HOLD
pH
CAL
CAL
mV pH
°C
4
pH
pH
CAL
MEAS
READY
55
mV
pH
MEAS
READY
ENT
pH
°C
ATC
pH
°C
ATC
3
READY
CAL
pH
ENT
CAL
pH
From pH measurement mode press CAL key to enter calibration mode as
11
described in section 4.2. The LCD shows ‘CAL PH’. Press ENT key to begin first calibration point.
Place the electrode in pH 6.86 buffer. Immerse the temperature probe in
2
the buffer solution if ATC mode is enabled. Immerse the potential matching
- 19 -
Page 26
3
4
55
NOTE:
pin in the buffer if symmetrical mode is enabled. The buffer annunciator appears in LCD. Lower display shows pH 6.86 (NIST standard buffer). Upper display shows the current uncalibrated pH reading. Allow the reading to stabilize. LCD shows ‘READY’ annunciator when the reading is stable. Press ENT key to confirm the reading.
The pH Transmitter moves to the second calibration point. The lower display shows next standard buffer value (pH 4.01). Use and keys to select your second buffer from one of the preset values: pH 4.01 or 9.18. Remove the electrode from the first buffer, wash and then immerse it into the second buffer. Allow the reading to stabilize. LCD shows ‘READY’ annunciator when the reading is stable. Press ENT key (or CAL key) to confirm the reading.
The calibration is completed. The pH Transmitter re-calculates electrode properties based on the calibration. The new slope (in mV) is shown in the upper display and normalized temperature is shown in the lower display.
Press ENT key to exit from the calibration.
The pH Transmitter reverts to pH measurement mode.
To exit from pH calibration mode without confirming calibration, press and keys together
When confirming the buffer measurement, if the measured pH value is not within ±1.00pH from selected buffer value, the
electro blinks and ERR indicator appears in the
de annunciator display. This error can also occur if non-standard buffers are used or the electrode has worn out. If this happens, press both ▲and ▼ keys together to restart the calibration beginning from Step1
When calibrating with manual temperature compensation, the meter automatically changes from the preset process temperature to the calibration temperature. After leaving the calibration mode, the pH Transmitter reverts back to process temperature.
- 20 -
Page 27

4.4 Check Calibration

MEAS
READY
pH
CAL
ENT
CAL
mV
pH
ATC
ENT
pH
1. While in Measurement mode press the CAL key. The display will prompt you to enter a security code. Leave the security code at “0” (view only mode).
2. Press the ENT key.
The display shows the slope in mV (Main Reading) and the normalized temperature is shown on the Auxiliary reading of the display.
NOTE:
To exit from any intermediate steps, press and keys (Escape) together. It will return to measurement mode.

4.5 ORP Calibration

If the unit of measurement is in mV, then the calibration is also carried out in mV.
CAL
mV
HOLD
HOLD
CAL
mV
HOLD
CAL
mV
READY
ENTCAL
mV
°C
1. Enter Calibration mode as described in section 4.2. The display shows “CAL ORP” (for mV calibration) and the “CAL” indicator appears on the display.
2. Place sensor in the solution.
3. Press the ENT key to start calibration. The main display shows the current mV output of the electrode with offset adjustment and auxiliary display shows the mV output with out any offset adjustments.
4. Press the ▲ or ▼ key to match the mV value to exact reading.
- 21 - - 22 -
Page 28
5. Press the ENT key to confirm. The transmitter calculates the correction factor and returns to the Measurement mode.

4.6 Temperature Calibration

Calibrate temperature probe only if temperature value displayed on the pH Transmitter is different from that of a calibrated thermometer. Refer to section 5.3 for further information temperature settings.
Page 29

5 SETUP MODE

5.1 Enter Setup mode

In the Setup mode the transmitter can be configured to the individual requirements.
MEAS
READY
pH
ENT ENT
°C
ATC
“User-setting” Mode
(User can change
parameter settings)
ENT
“User-Setting” Mode
view parameters)
(User can only
1. While in measurement mode, press the ENT key.
2. The display prompts to enter the security code. Set the security code with or key to:
“SCD 22” to change parameter settings
“SCD 0” to view parameter settings only (view only mode)
3. Press the ENT key.
NOTE:
To exit from any intermediate steps, press and keys (Escape) together. It will return to measurement mode.
- 23 -
Page 30

5.2 Electrode Offset Settings

NOTE:
‘Electrode Offset Setting’ is not available when the pH Transmitter is configured for ORP measurement mode. (Refer Section 5.5 for switching measurement modes)
In applications where continuous pH measurement is required, it may not be convenient to remove the electrode for calibration. In such cases, an on-line offset adjustment is recommended. The pH Transmitter allows you set an offset of up to ± 2.00pH to compensate for errors in the pH electrode.
The pH Transmitter adds or subtracts the offset value from the measured pH value and displays the corrected value. However, if you need to offset the value beyond the average offset you would expect in your application type, consider a full calibration or even electrode replacement.
To Determine the Offset Value:
1. Take a sample from the liquid of pH measurement. Record the pH reading
of the pH transmitter at the time the sample was taken.
2. Measure the pH value of your sample using another calibrated pH meter
having its own electrode (pH tester, hand-held meter or bench meter)
3. The difference between these two readings is the offset
To Adjust the Offset Value:
SETUP
1
ENT
SETUP
READY
2
pH
pH
SETUP
READY
From pH measurement mode press ENT key to enter setup mode as
11
described in section 5.1. The LCD shows the first screen of setup mode (OFS). Press ENT key to access electrode offset setting (OFS).
The upper display shows the last configured offset value (if any), otherwise
2
zero. The lower display shows currently measured pH reading (including last configured offset value). Allow the reading to stabilize. LCD shows
- 24 -
3
pH
pH
ENT
Page 31
A
READY’ annunciator when the reading is stable.
Press or key to set the upper display to the newly calculated offset
value.
As the upper display value (offset) changes, the pH Transmitter adjust the
3
lower reading automatically to suit the new offset value. Up to ± 2.00 pH offset is allowed.
Press ENT key to confirm the value. The pH Transmitter reverts to OFS screen.
NOTE:
To exit from any intermediate steps, press and keys (Escape) together. It will return to measurement mode.

5.3 Temperature Settings

Automatic Temperature Compensation (ATC):
The pH values other than pH 7.00 are affected by temperature. Use ATC feature of the pH Transmitter to compensate for pH changes when the temperature of the sample or process liquid fluctuates. Connect a separate temperature probe (Pt100/pt1000) to the pH Transmitter if an integrated pH/Temperature probe is not available.
Manual Temperature Compensation (MTC):
Set the pH Transmitter to MTC (disable ATC) when the temperature of sample or process liquid is constant and a temperature probe is not available. If you disable ATC, the pH Transmitter allows you to set your process temperature (P.ºC) and calibration temperature (C.ºC). This allows calibration at a different temperature other than the process temperature. Example: Setting calibration temperature of 25°C lets you calibrate using standard solutions at 25°C, even if your process temperature is different from 25°C.
SETUP
1
SETUP
ENT
2
SETUP
ENT
°C
3
SETUP
ENT
°C
4
SETUP
°C
TC
5
ENT
°C
ATC
SETUP
2
SETUP
ENT
°F
3
SETUP
ENT
°C
6
SETUP
ENT
7
ENT
- 25 -
Page 32
From pH or ORP measurement mode press ENT key to enter setup mode
11
as described in section 5.1. The LCD shows the first screen of setup mode (OFS). Press or key to select Temperature settings screen (SET
ºCF).
Press ENT key to access temperature settings (SET ºCF).
Selecting unit of measurement for temperature: The upper display
2
shows ‘Unit’ and the lower display shows the last configured unit of measurement for temperature.
Press or key to select the desired units for temperature (ºC or ºF). Press ENT key to confirm your selection.
Enable/disable ATC: The lower display shows ‘AtC’ and the upper display
3
shows the last configured ATC selection (‘On’ or ‘OFF’).
Press or key to enable (ATC On) or disable (ATC OFF) automatic temperature compensation. Press ENT key to confirm your selection.
If ATC enabled (ATC On):
Setting temperature offset: The upper display shows the last configured
4
temperature offset value (if any), otherwise the default is zero. The lower display shows currently measured temperature reading (including last configured offset value). LCD shows ‘ATC’ annunciator in lower-right corner.
Place a thermometer, which is known to be accurate, in your sample or process liquid. Make sure your temperature probe is placed in the same liquid. Compare the stabilized temperature reading displayed on the Transmitter with the thermometer. If there is a difference between the two readings (offset), you can adjust the reading of the Transmitter. Press or key to adjust the lower display to the correct temperature value.
As the lower display value changes, the pH Transmitter adjust the upper
55
display reading automatically to suit the new offset value. Up to ± 10 ºC/ ± 18 ºF offset is allowed.
Press ENT key to confirm the value. The pH Transmitter reverts to SET ºCF screen.
If ATC disabled (ATC OFF):
Setting process temperature: The lower display shows ‘P.ºC’ and the
6
upper display shows the last configured process temperature. Press ▲ or key to adjust the upper display to desired process temperature.
Allowable range: –10.0 to 110.0°C / 14.0 to 230°F.
Press ENT key to confirm the process temperature.
- 26 -
Page 33
Setting calibration temperature: The lower display shows ‘C.ºC’ and the
7
upper display shows the last configured calibration temperature. Press or key to adjust the upper display to desired calibration temperature.
Allowable range: –10.0 to 110.0°C / 14.0 to 230°F.
Press ENT key to confirm the calibration temperature. The pH Transmitter reverts to SET ºCF screen.
NOTE:
Press or key to access other setup screens or press and key simultaneously (escape) to return to measurement mode.
To exit from any intermediate steps, press and keys simultaneously (escape). The pH Transmitter returns to the first screen of temperature settings SET ºCF.

5.4 Buffer Selection Settings

NOTE:
The buffer selection setting is only available if the Transmitter is configured for pH measurement mode.
This allows you to set the pH Transmitter to accept either USA or NIST pH buffer standard group during pH calibration.
Buffer Group pH Buffer Points
USA pH 4.01, 7.00 & 10.01
NIST pH 4.01, 6.86 & 9.18
SETUP SETUP
ENT
SETUP
- 27 - - 28 -
ENT
ENT
Page 34
From pH measurement mode press ENT key to enter setup mode as
11
described in section 5.1. The LCD shows the first screen of setup mode (OFS). Press or key to select buffer settings screen (bUFF).
Press ENT key to access buffer settings (bUFF).
Selecting buffer group: The lower display shows the last configured
2
buffer group. Press or key to select the required buffer group (‘USA’ or ‘nST’).
Press ENT key to confirm your selection. The pH Transmitter reverts to bUFF screen.
NOTE:

5.5 Configuration Settings

Configuration settings let you configure the pH Transmitter to different measurement & operation modes & reset the pH Transmitter to factory defaults.
Asymmetrical Operation Mode (ASY):
This is the default mode of operation of the pH Transmitter, which is suitable for most applications where sample (or process) liquid is not interfered by adjacent electrical noise.
Symmetrical Operation Mode (SY):
This is the mode of operation is recommended for applications where sample (or process) liquid is subjected to high electrical interference and therefore the pH reading becomes unstable. When the pH Transmitter is set to symmetrical mode, you should immerse the potential matching pin in the sample (or process) liquid or use a pH electrode with built-in potential matching pin. The ‘MEAS’ indicator blinks when the pH Transmitter is set to Symmetric mode.
To exit from any intermediate steps, press and keys (Escape) together. It will return to measurement mode.
Page 35
SETUP SETUP SETUP
ENT
ENT
SETUP SETUP
ENT
SETUP SETUP
ENT
SETUP SETUP
SETUP
HOLD
Measurement
Mode
NOTE:
To exit from any intermediate steps, press and keys (Escape) together. It will return to measurement mode.

Selecting “pH” or “mV” Mode of Operation

1. Press ENT key and use or ▼ keys to scroll till LCD displays ‘CONF’ (Configuration). Press ENT again.
2. LCD displays ‘PH’. meter is set to operate as a pH meter. Use or keys to change from ‘PH’ to ‘ORP’, if this Transmitter is to operate as an ORP Transmitter.
3. Press the ENT key to accept selection.
4. Press S and T keys together to return to the CONF main menu.

Selecting “Asymmetrical” or “Symmetrical” Mode of Operation

In some applications, there may be some electrical interference in the sample, where the electrode is installed. The sensitive electrode picks up the signal and the effect is displayed on the LCD.
If the electrical interference is from an AC source, the reading on the LCD fluctuates wildly. If the source is DC, then readings will be stable, but at a wrong value.
- 29 - - 30 -
Page 36
In such cases, switch the Transmitter to the Symmetrical mode as below:
1. Press ENT key and use or ▼ keys to scroll till LCD displays ‘CONF’ (Configuration). Press ENT again.
2. LCD displays “ASY” (asymmetrical operation). Use or keys to select for ‘”SY” (symmetrical mode of operation). Press ENT key again.
3. Press S and T keys together to return to the CONF main mode.

Selecting Back Light ON or OFF Operation

1. Press ENT key and use or ▼ keys to scroll till LCD displays ‘CONF’ (Configuration). Press ENT again.
2. Press ENT key until the LCD displays either “LitE ON” or “LitE OFF”
3. LCD displays “LitE ON” (Back light ON). Use or keys to select for “LitE OFF” (back light OFF).
4. Press ENT key again.
5. Press S and T keys together to return to the CONF main mode.

Reset the User Calibration parameters or Setup parameters

1. Press ENT key and use or ▼ keys to scroll till LCD displays ‘CONF’ (Configuration). Press ENT again.
2. Press ENT key until the LCD displays “no DEF”.
3. Not to reset any of the setup or user calibration parameters select “nO DEF” and press ENT key. It will lead to the CONF main menu.
4. To reset the user calibration parameters, select “CAL DEF” and press ENT key. It will reset the User calibration parameters and lead to measurement mode.
5. To reset the user calibration parameters and meter setup parameters to default values, select “FAC DEF” and press ENT key. It will reset the user calibration parameters and meter setup parameters and lead to measurement mode.
Page 37

5.6 Viewing Electrode Properties

Each time you calibrate your pH electrode, the pH Transmitter re-calculates slope & offset of the electrode and shown in the LCD at the end of calibration. The setup mode allows you to view slope & offset values at any time.
NOTE:
When the measurement mode is set to ORP, only the offset value of the electrode is displayed.
11
2
NOTE:
SETUP
HOLD
1
ENT
SETUP
measurement mode
For pH
SETUP
measurement mode
For ORP
2
mV
ENT
pH
2
mV
ENT
To view the calibration and parameter, press CAL key, then ENT key.
Electrode status: T de annunciator
pH measurement mode, the slope (in mV) is shown in the upper display
he electro appears in the LCD. For
and normalized temperature is shown in the lower display.
For ORP measurement mode, the offset value (in mV) of the electrode is shown in the upper display.
Press ENT key to quit from electrode properties screen.
To exit from any intermediate steps, press and keys (Escape) together. It will return to measurement mode.
- 31 -
Page 38

6 TECHNICAL SPECIFICATIONS

General Specification (a) pH
(b) mV
(c) Temperature
Calibration Point (a) pH
(c) Temperature
Input / Output
Display
Electrical data and Connections
RF Communication Specifications
Others
pH Measuring Range 0.00 to 14.00 pH Resolution 0.01 pH Accuracy ± 0.01 pH
mV Measuring Range -1000 to 1000 mV Resolution 1 mV Accuracy ± 1 mV
Temperature measuring range -10.0 to +110.0 oC or +14.0 to +230.0 oF Resolution 0.1 Relative accuracy 0.5 oC or 1.0 oF Sensor 3 wire / 2 wire – Pt100 or Pt1000 (Jumper Selectable) Compensation Automatic or Manual
Number of calibration points Number of calibration buffers
Offset Adjustment
Input Asymmetrical / Symmetrical Output UART Serial Comm. Output for the RF RX/TX Board:
LCD UV coat, Custom Dual segments display with symbols
Back light On/Off selectable
pH/mV Input 8-pin Screw Terminal (3.5mm pitch) Temperature Input 8-pin Screw Terminal (3.5mm pitch) Power Phone Jack DC Socket
Transmission Method TDMA, FSK Modulation Frequency 923 to 928 MHZ, ISM Band Region 2 Air Baud Rate 16 Kbit/S Range Single Unit 100 m (328 ft) (in open air) RF Power +5 dB Antenna Internal diversity antenna Network Model Mesh Operating Temperature -10 to 65 °C
Power Input +9 V DC Pollution Degree 2 Dimensions (W x H x D) 96mm x 96mm x 66 mm Weight (Estimated) 210g Ambient Temp. operating range 0 to 40 oC Maximum Relative Humidity 80% up to 31 oC decreasing linearly to 50% at 40oC Ingress Protection IP65
o
C or 0.1 oF
1 or 2 points USA: 4.01, 7.00, 10.01 NIST: 4.01, 6.86, 9.18
±10 ˚C/ ±18 ˚F
for Status information.
Protocol : 8-N-1
Data rate: 9,600 kbps
- 32 -
Page 39

7 LIST OF ACCESSORIES

7.1 Thermo Scientific

pH Transmitter Replacement and Accessories
Item Description
Alpha pH 600 wall-mount pH/ORP RF transmitter TSPHCTP0600
Wireless (RF) receiver and application software TSRCV0600S
Ryton®-body pH combi electrode with Pt100 RTD (ATC) & 20 m cable with BNC & PMP EC100GTSO20B
Ryton®-body pH combi electrode with Pt100 RTD (ATC) & 10 m cable with BNC & PMP EC100GTSO10B
Ryton®-body pH combi electrode with Pt100 RTD (ATC) & 5 m cable with BNC & PMP EC100GTSO05B
Ryton®-body pH combi electrode with 5 m cable with BNC & connector for PMP (no ATC) ECARGTSO05B
Ryton®-body pH combi electrode with 5 m cable with BNC connector (no ATC); measures up to 110 ºC ECARHTTSO05B
Ryton®-body pH combi electrode without ATC & 5 m cable with BNC connector. HF resistant glass ECARTSOHF05B
Ryton®-body pH combi electrode with 5 m cable with BNC connector (no ATC) ECARTSO05B
Ryton®-body ORP gold electrode with 5 m cable with BNC & PMP (no ATC) ECHTAUTSO05B
Ryton®-body ORP platinum electrode with 5 m cable with BNC & PMP (no ATC) ECHTPTTSO05B
Order Code
Calibration Solutions
Item Description
pH 4.01 buffer solution, 480 ml bottle EC-BU-4BT
pH 7.00 buffer solution, 480 ml bottle EC-BU-7BT
pH 10.01 buffer solution, 480 ml bottle EC-BU-10BT
pH 4.01 buffer sachets, 20 ml x 20 pcs. EC-BU-4BS
pH 7.00 buffer sachets, 20 ml x 20 pcs. EC-BU-7BS
pH 10.01 buffer sachets, 20 ml x 20 pcs. EC-BU-10BS
pH De-ionized water rinse sachets, 20 ml x 20 pcs EC-RIN-WT
pH sachet assortment pack – 5 each of pH 4.01, pH 7.00, pH 10.01 and de-ionized water sachets per box
Protein cleaning solution for pH electrode EC-DPC-BT
Storage solution for pH electrode EC-RE-005
Order Code
EC-AST-PK
NOTE:
pH buffer solutions (480 ml bottle) have ±0.01 pH accuracy at 25 °C
pH buffer sachets are individually sealed, single use pouch
containing 20 ml of fresh, contamination free calibration solution
pH buffer sachets have ±0.01 pH accuracy at 25°C
- 33 -
Page 40

7.2 Eutech Instruments

pH Transmitter Replacement and Accessories
Item Description
Alpha pH 600 transmitter 56706-00
Wireless RF receiver with application software 56706-50
pH/Temp electrode with PMPO and 10-ft cable 35807-20
Platinium ORP electrode with 10-ft cable 35801-21
BNC to spade lug adapter 05994-90
Order Code
Calibration Solutions
Item Description
pH 4.01 calibration buffer, 500 ml 00654-00
pH 7.01 calibration buffer, 500 ml 00654-04
pH 10.01 calibration buffer, 500 ml 00654-08
pH 4.01 calibration buffer solution pouches, 20/box 35653-01
pH 7.00 calibration buffer solution pouches, 20/box 35653-02
pH 10.00 calibration buffer solution pouches, 20/box 35653-03
pH De-ionized water rinse sachets, 20 ml x 20 pcs 35653-00
Assortment pack – 5 each of pH 4.01, pH 7.00 and pH 10.00 solution pouches. 35653-04
Electrode cleaning solution 00653-06
Electrode storage solution 00653-04
Order Code
NOTE: To order Eutech accessories, contact the nearest Oakton distributor
- 34 -
Page 41

8 TROUBLESHOOTING

Problem Solution
Power on, but no display a) Loose connections
Unstable pH reading a) Dirty electrode
Oscillating temperature readings
Slow response a) Dirty / Oily electrode a) Clean electrode
Blinking ATC a) No temperature probe
Blinking electrode annunciator
Or (pH) a) pH electrode is not
Or (Temperature) a) Temperature probe is
Cause
a) Ensure cables make good
b) Incorrect output voltage
of the power adaptor
b) Electrical noise
interference
a) Electrical noise
interference
connection during ATC mode
a) Error in calibration a) Ensure calibration standard
connected
not connected when ATC enabled
contact
b) Use an power adaptor with
specified output voltage
a) Clean electrode and
recalibrate b) Switch to Symmetric mode
a) Ensure shield wire is
properly connected to pin 7
a) Ensure temperature sensing
cable makes good contact
solution is not
contaminated. Ensure
electrode is clean.
a) Ensure electrode makes
good contact with
Transmitter
a) Ensure electrode makes
good contact with
Transmitter
- 35 -
Page 42

9 GENERAL INFORMATION

9.1 Warranty

This transmitter is supplied with a one-year warranty against significant deviations in material and workmanship from date of purchase and a six-month warranty for probe. Each instrument will have a warranty card with a specific serial number. The warranty card must be endorsed by the Authorized Distributor at the point of sale.
If repair or adjustment is necessary and has not been the result of abuse or misuse within the designated period, please return – freight pre-paid – and correction will be made without charge. Thermo Scientific/ Eutech will determine if the product problem is due to deviations or customer misuse.
Out of warranty products will be repaired on a charged basis.
Exclusions
The warranty on your instrument shall not apply to defects resulting from:
Improper or inadequate maintenance by customer
Unauthorized modification or misuse
Operation outside of the environment specifications of the products

9.2 Return of Goods

Authorization must be obtained from our Customer Service Department or authorized distributor before returning items for any reason. A “Return Goods Authorization” (RGA) form is available through our authorized distributor. Please include data regarding the reason the items are to be returned. For your protection, items must be carefully packed to prevent damage in shipment and insured against possible damage or loss. Thermo Scientific will not be responsible for damage resulting from careless or insufficient packing. A restocking charge will be made on all unauthorized returns.
NOTE: Thermo Scientific reserves the right to make improvements in design, construction, and appearance of products without notice.
- 36 -
Page 43

10 APPENDICES

10.1 Appendix 1– pH Buffer Values at Various Temperatures

The following table shows the various pH values at different temperature of the solution during calibration.
3 pH 4.01 pH 6.86 pH 7.00 pH 9.18 pH 10.01
0 4.01 6.98 7.12 9.47 10.32
5 4.01 6.95 7.09 9.38 10.25
10 4.00 6.92 7.06 9.32 10.18
15 4.00 6.90 7.04 9.27 10.12
20 4.00 6.88 7.02 9.22 10.06
25 4.01 6.86 7.00 9.18 10.01
30 4.01 6.85 6.99 9.14 9.97
35 4.02 6.84 6.98 9.10 9.93
40 4.03 6.84 6.97 9.07 9.89
45 4.04 6.83 6.97 9.04 9.86
50 4.06 6.83 6.97 9.01 9.83
55 4.08 6.83 6.97 8.99 9.81
60 4.10 6.84 6.98 8.96 9.79
70 4.12 6.85 6.99 8.92 9.76
80 4.16 6.86 7.00 8.89 9.74
90 4.20 6.88 7.02 8.85 9.73
- 37 -
Page 44

10.2 Appendix 2 – Abbreviations Used in LCD

Abbreviation Description

ASy Asymmetrical mode

AtC

bUFF Buffer

CAL Calibration

C.ºC Calibration temperature

CONF Configuration

dEF Default values

FCt Factory (defaults)

OFS Offset

Or Reading is over range

Orn Out of range

OrP Oxidation Reduction Potential

P.ºC Process temperature

PH pH

Set Setting

Sy Symmetrical mode

tyP Type

Ur Reading is under range

Automatic Temperature Compensation

- 38 -
Page 45
Page 46
Page 47
Page 48
Thermo Scientific Water Analysis Instruments
North America 166 Cummings Center
Beverly, MA 01915 USA Toll Free: 1-800-225-1480 Tel: 1-978-232-6000 Dom. Fax: 1-978-232-6015 Int’l Fax: 978-232-6031
www.thermo.com/process
Europe
Denmark House, Angel Drove Ely, Cambridgeshire CB7 4ET, UK Tel: 44-1353-666111 Fax: 44-1353-666001
Asia Pacific
Blk 55, Ayer Rajah Crescent, #04-16/24 Singapore 139949 Tel: (65) 6778 6876 Fax: (65) 6773 0836 E-mail: eutech@thermofisher.com
Oakton Instruments
P.O Box 5136, Vernon Hills, IL 60061, USA Tel (in U.S.): 888-462-5866 Tel (outside U.S.): 1-847-549-7600 Fax: (1) 847-247-2984 E-mail: info@4oakton.com Web-site: http://
www.4oakton.com
Distributed by:
Loading...