Instruction Manual pH 5+, pH 6+, Ion 6+
1. INTRODUCTION
Thank you for selecting our pH 5+, pH 6+, or Ion 6+ portable meter. This
microprocessor-based handheld instrument is both economical and easy to use.
The pH 5+ measures pH and temperature (
The pH 6+ and Ion 6+ measure pH, mV (ORP) and temperature (
The Ion 6+ can also measure direct ion concentration of various ions (mono and
divalent) using an ion selective electrodes (ISE). The mV mode is also useful for ISEs.
The 5+/6+ series meters advance our popular 5/6 series meters that were introduced in
1998.
Each meter includes alkaline “AAA” batteries, a rubber armor / stand, and instruction
manual. Please refer to Section—7 Replacements and Access ories
on additional accessories and calibration solutions.
We take great pride in every instrument we manufacture and hope this
one serves you well.
If you are viewing an electronic PDF version of this manual, look for bold and
underlined hyperlinks
in the Table of Contents and elsewhere. Clicking on
them will immediately take you to the corresponding location in the manual.
*Find other helpful tips listed in grey boxes like this one!
o
C).
o
C).
for information
1
Instruction Manual pH 5+, pH 6+, Ion 6+
2. GETTING STARTED
Description of Keypad Functions
The pH 5+ and pH 6+ have four keys, while the Ion 6+ has six keys on its splash-proof
keypad as shown here:
pH 5+ / pH 6+ Ion 6+
Powers the meter on and off. Upon power on, the meter
automatically begins in the measurement mode that was last
used.
With meter off, press and hold
(buffer), (calibration), and (electrode) setup menus
while powering on. Press
menus.
or to access the
or again to change setup
INCREMENT: (INC) for Temperature setting and calibration.
2
Instruction Manual pH 5+, pH 6+, Ion 6+
CALIBRATE: Press to begin calibration of the selected mode.
Press again during calibration to abort calibration and return to
measurement mode without confirming any values.
During setup mode, function is similar to “escape” or “go
back”, returning to the previous screen.
With meter off, press and hold to select the reset menu
() while powering on.
Increase value or scroll up in Setup or Cal modes.
Decrease value or scroll down in Setup or Cal modes.
HOLD: Freezes measured reading—indicated by “HO”. Press
again to resume live reading.
ENTER: Confirms values in calibration mode. Confirms
selections in setup mode.
Description of LCD Annunciators
The custom LCD consists of 3½-digit segments which uses annunciators for pH, mV or
o
C (Temperature). No annunciator is shown in Ion mode. Other annunciators include
“HO” (when HOLD function is activated) and “LO” (low battery condition).
3
Instruction Manual pH 5+, pH 6+, Ion 6+
Inserting & Removing the Rubber Armor / Stand
Before removing the meter from the rubber armor, disconnect any electrodes. Push out
from the bottom edges of meter until it is completely out of boot.
To insert meter into the armor, slide in from the top of meter before pushing the bottom
edges of meter into position. Tilt the stand at the back of meter for table top usage as
desired.
Inserting New Batteries
The “LO” annunciator alerts you when battery power is running low. Power off the meter
before removing the batteries. After removing the rubber armor, push the battery cover
in the direction of the arrow and lift up—no screwdriver is required. Note the polarity of
batteries before inserting into position.
4
Instruction Manual pH 5+, pH 6+, Ion 6+
Connecting Electrodes and Temperature Sensor
Align the BNC connector slots with the posts of meter’s socket and rotate connector
clockwise until it locks—do not force. To remove, rotate the connector in counterclockwise direction until it unlocks, and slide the connector off the socket. Insert the mini
phono jack of temperature sensor into the socket on the meter. Unplug the phono jack
to perform measurements without temperature compensation (25° C default).
BNC connector of
pH, ORP, or Ion
Selective Electrode
2.5 mm Phono plug
for temperature
measurement
Powering the Meter On
Press . All LCD segments will display momentarily during the self-diagnostic test,
before scrolling the model number, software revision, pH buffer group, and number of
pH calibration points selected before returning to the measurement mode:
If a temperature probe is not connected, either 25.0
calibrated temperature value is used. If a temperature probe is connected, the current
measured temperature is used.
Note: The Ion 6+ will display “ - - - ” in Ion concentration mode if the meter has not been
calibrated or if the meter has been reset. See Section 3—Ion Calibration
o
C (factory default) or the last
.
“” (Over range) and “” (Under range) indicates the reading exceeds the maximum
or minimum measurement range. See Section 6—Specifications
. However, the most
likely reason for these error messages is that the electrode is not connected or broken.
5