Thermo Fisher Scientific pH 1000 User Manual

Page 1
Alpha pH/ORP 1000
Process Controller/Transmitter
Page 2
R
Afast, Cahn, ionplus, KNIpHE, No Cal, ORION, perpHect, PerpHecT, PerpHecTion,
AQU pHISA, pHuture, Pure Water, Sage, Sensing the Future, SensorLink, ROSS, ROSS Ultra, Sure-Flow, Titrator PLUS and TURBO2 are registered trademarks of Thermo Fisher.
1-888-pHAX-ION, A+, All in One, Aplus, AQUAsnap, AssuredAccuracy, AUTO-BAR, AUTO­CAL, AUTO DISPENSER, Auto-ID, AUTO-LOG, AUTO-READ, AUTO-STIR, Auto-Test, BOD AutoEZ, Cable-Free , CERTI -CAL, CI SA, Dat aCOLLECT , DataPLU S, dig ital Log R, Di rectCa l, DuraProbe, Environmental Product Authority, Extra Easy/Extra Value, FAST QC, GAP, GLPcal, GLPcheck, GLPdoc, ISEasy, KAP, LabConnect, LogR, Low Maintenance Triode, Minimum Stir Requirement, MSR, NISS, One-Touch, One-Touch Calibration, One-Touch Measurement, Optimum Results, Orion Star, Pentrode, pHuture MMS, pHuture Pentrode, pHuture Quatrode, pHuture Triode, Quatrode, QuiKcheK, rf link, ROSS Resolution, SAOB, SMART AVERAGING, Smart CheK, SMART STABILITY, Stacked, Star Navigator 21, Stat Face, The Enhanced Lab, ThermaSense, Triode, TRIUMpH, Unbreakable pH, Universal Access are trademarks of Thermo Fisher.
Guaranteed Success and The Technical Edge are service marks of Thermo Fisher.
Page 3
Preface
This manual serves to explain the use of the Alpha pH1000 pH/ORP process controller/transmitter. The manual functions in two ways, firstly as a step by step guide to help the user operate the instrument. Secondly, it serves as a handy ref erence guide. This instruction manual is written to cover as many anticipated applications of the Alpha pH1000 pH/ORP process controller/transmitter. If you have doubts in the use of the instrument, please do no t hesitate to co ntact the nearest Authorized Distributor.
Thermo Scientific will not accept any responsibility for damage or malfunction of the unit due to improper use of the instrument.
The information presented in this manual is subject to change without notice as improvements are made, and does not represent a commitment on part of Thermo Scien tific.
Copyright All rights reserved
©
2009
Page 4
Safety Information
The Controller/ Transmitter shall be installed and operated only in the manner specified in this Instruction ma nual. Only skilled, trained or authorized person should carry out installation, setup and operation of the instrument.
Before powering up the unit, make sure that power source is connected to it as specified on the top label. Failure to do so may result in permanent damage of the unit.
The unit has live and exposed parts inside. If it has to be opened, make sure that power to the unit is off and disconnected.
The unit is Fuse protected. In the event the fuse has to be replaced, use only those as specified in the manual.
Page 5
TABLE OF CONTENTS
1 INTRODUCTION ___________________________________________________ 1
1.1 Description of Unit............................................................................................1
1.2 Applications......................................................................................................1
2 ASSEMBLY AND INSTALLATIONS ____________________________________ 2
2.1 Measurement and Control System..... ..............................................................2
2.2 Unit Dimensions...............................................................................................2
3 ELECTRICAL CONNECTIONS ________________________________________ 3
3.1 Connection Diagram ........................................................................................3
3.2 Back Panel....................................................................................................... 5
4 OVERVIEW _______________________________________________________ 6
4.1 Keypad and Display .........................................................................................6
4.2 Function Groups...............................................................................................7
4.3 Control Concept...............................................................................................8
5 MEASUREMENT ___________________________________________________ 9
5.1 Display in Measurement mode..... .... ... ........................................ ... ... ...............9
5.2 Security Codes.................................................................................................9
6 CALIBRATION____________________________________________________ 11
6.1 pH Calibration ................................................................................................11
6.2 ORP – mV Calibration...... ... ........................................ ... ... .............................12
6.3 ORP % Calibration......................................................................................13
7 ADVANCED SET UP MODE _________________________________________ 14
7.1 Electrode Offset (OFS) sub-function.......................................................... .... 14
7.2 Setting temperature (Set
7.3 Control Relay A/ Control Relay B (SP1/SP2) sub-function............................ 16
7.4 Controller (Cntr) sub-function.........................................................................18
7.5 Current Output (mg) sub-function .................................................................. 20
7.6 Configuration (ConF) sub-function.................................................................21
7.7 Calibration (CAL) sub-function....................................................................... 24
o
C) sub-function ...... ... ... ... ...................................... 14
8 ADVANCED SET UP MODE _________________________________________ 25
8.1 Auto mode (mode after switch-on)............ ... ... ........................................ ... ... . 25
8.2 Manual mode .................................................................................................25
9 TECHNICAL SPECIFICATIONS ______________________________________ 26 10 ACCESSORIES ___________________________________________________ 28 11 GENERAL INFORMATION __________________________________________ 29
11.1 Warranty.........................................................................................................29
11.2 Packaging ......................................................................................................29
11.3 Return of Goods.............................................................................................29
11.4 Guidelines for Returning Unit for Repair ........................................................29
12 APPENDICES ____________________________________________________ 30
12.1 Appendix 1 ..................................................................................................... 30
12.2 Appendix 2 ..................................................................................................... 31
12.3 Appendix 3 ..................................................................................................... 32
12.4 Appendix 4 ..................................................................................................... 33
Page 6
1 INTRODUCTION
1.1 Description of Unit
Thank you for purchasing the ¼ DIN Alpha pH1000 pH/ORP process controller/transmitter. This unit is used for measuring either pH or ORP parameter one at a time, and the operational mode is switchable from the menu. You can use this unit to measure pH or ORP with limit control. This controller has many user-friendly and safety features which include:
Menu-driven program that simplifies set-up
Built-in non-volatile memory backup to ensure that calibration and other information
are not erased if power supply fails
Press-button two-point calibration and electrode offset adjustment from the keypad
Automatic temperature compensation (ATC)
Manual temperature compen sation setting without the ATC probe, with independent
setting for calibration and pr ocess temperature
0 to 2000 second time delay adjustment on all relays – minimise false alarms
Separately adjustable high and low set point hysteresis (dead bands) prevent
chattering of relays around the set points
Asymmetrical/symmetrical input for pH/ ORP operati o n
Three control modes: limit, proportional pulse length or proportional pulse frequency
Large dual display LCD for easy reading with clear multiple annunciators, alarm
status, operational and error messages
Two switching contacts as set-point tr iggering relays
A third rela y that can be be set up to perform either Alarm or Wash function
When set to work as Alarm relay, alerts you when readings exceed set-point limits after
a set time buffer, and if the Pt100/Pt1000 wires are broken or disconnected during the ATC function
Programmable duration and frequency in the Wash mode
Hold function freezes output current (0/4…20mA) and releases control relays
LED indicators signal control activities to monitor controller status from a distance
Protection against electromagnetic interference – galvanically isolated 0/4…20mA
output provides safety for datalo
Choice of 7 buffers from USA or NIST standards
gging and control purposes
1.2 Applications
Use this controller in panel mounted enclosures for applications such as water treatment and monitoring, galvanic-decontamination, chemical processing food processing, clean or waste water control and neutralisation process.
1
Page 7
2 ASSEMBLY AND INSTALLATIONS
6
2.1 Measurement and Control System
A typical measurement system consists of:
a pH/ORP process controller
a pH/ORP combination electrode with integrated or separate temperature sensor Pt
100/1000,
an immersion, flow or process assembly with or without a potential matching pin (PMP)
a final control element such as pump or valve
a recorder
an appropriate pH or ORP measurement cable
Flow Assembly
PH Controller
Chart Recorder
Process Assembly
with Electrode
Measurement Cable
Power Mains
(220/110 VAC)
2.2 Unit Dimensions
Flat Gasket (1mm)
(To be Inserted By Customer)
Note: The Taped Corners Have to Be On Top
92 + 0.5
92
56
92 + 0.5
Mounting Cut-Out
The field-tested control panel housing is 96 x 96 mm; with protection class IP 54 (front).
32
max. 45
max. 175
9
2
Page 8
3 ELECTRICAL CONNECTIONS
3.1 Connection Diagram
ENSURE electrical mains is disconnected before proceeding.
S/
V
S/S/
pH
AC:
Power Mains
2
1
NL
PE
18 19 20
Pt 100
3
Relay A
4
21 22
PE/S
Potential Matching Pin (PMP)
Relay B
6
5
78 9
Signal Input pH/ORP
Alarm
10
11
BNC
12 13
14
PE/S
151715
Signal OutputHold Input
pH
-
mA
+
16
*)indicated contact positions are for currentless conditions
3
Page 9
NLS
PE
22
1
3
2
Potential
Matching Pin (PMP)
NLS
PE
22
1
3
2
Potential
Matching Pin (PMP)
0 / PA
21
0 / P A
21
20
20
Temp.
19 18
Temp.
19 18
pH
s
Measuring Cable
pH Combination
Electrode
pH
Measuring Cable
pH Combination
Electrode
BNC
BNC
Symmetrical high-impedanc e connections
Asymmetrical connect ions
4
Page 10
3.2 Back Panel
A
ENSURE electrical mains is disconnected before proceeding.
The back panel consists of two connectors. The first connector is the 17-way PCB edge connector and the other is the 5-way connector.
Connection for the 17-way screw terminals (from left to right):
1. AC mains live wire 10. Alarm/Wash relay working position (NO)
2. AC mains neutral wire 11. Alarm/Wash relay common
3. AC mains protective earth wire 12. Alarm/Wash relay resting position (NC)
4. Low set relay resting position (NC) 13. Hold function switch terminal 1
5. Low set relay common 14. Hold function switch terminal 2
6. Low set relay working position (NO) 15. No connection
7. High set relay resting position (NC) 16. 4 - 20 mA for -ve connection
8. High set relay common 17. 4 - 20 mA for +ve connection
9. High set relay working position (NO)
Connections for the 5-way screw terminals:
18. Pt100/Pt1000 lead 1 terminal
19. Pt100/Pt1000 sense lead terminal
20. Pt100/Pt1000 lead 2 terminal
Note: If using a two-wire RTD, short terminal 19 to terminal 18.
Pt 100/Pt 1000 is selectable via an internal jumper. Factory default is Pt100. See Appendix 1 for directions on switching the R TD type.
21. pH/ORP (potential matching pin)
22. pH/ORP (shield)
pH/mV
FUSE 250V AC
(F)
100mA
L
N
PE
RELAY A
Pt100/
Pt1000
RELAYB
0V/
S/
PAL
LARM / WASH
HOLD
+
-
NC
Important: The Alarm relay functions as an “Active Low” device i.e. it switches OFF under Alarm condition. Therefore, the Alarm display device should be connected to the ‘NC’ contacts of the relay (11 and 12). In the Wash mode, the relay works on a positive mode. Therefore, the Wash device should be connected to contacts 10 and 11.
5
Page 11
4 OVERVIEW
4.1 Keypad and Display
4.1.1 Keypad
Perform rapid 2-point ca l ibration
Allows entry to Set up mode
Select individual functions within the function group of Set up mode
Store input data in the Set up mode
Start calibration in the calibration mode
Select various function groups in the Set up mode.
Set parameters and numerical values in sub functions of Set up mode
If pressed continuously, the setting speed increases
Return to the Measurement mode when both keys are pressed together
Return to the Measurement mode when both keys are pressed together
Switch beween AUTO and MANUAL relay operation mode
Display limit set-point values for the switch contacts in AUTO relay operation
mode
Switch between RELAY A and RELAY B in MANUAL relay operation mode
4.1.2 Display
The LCD display features two numerical displays that show status messages and measured values for easy, quick reference. The display provides short-text information for setting parameters and configuration.
MEAS: Measurement mode
SETUP: Set-up mode of function groups
CAL: Calibration mode of pH/ORP and temperature
READY: Comes on after a succe ssful calibrat ion
HOLD: Relay position and current output are frozen
ATC: Comes on in the ATC mode. Disappears in the
Manual temperature Compensation mode.
“ATC” flashes if the temperature probe is faulty in its ATC mode
ERR: Error or alarm indicator
READY HOLD
7.00
ERR
25.0
Display for RELAY A/B. Green LED indicates measured value within limit
CONMEAS CALSET UP
pH
°C AT C
while RED LED indicates measured value outside limit.
Alarm display if limit value overshoot or the ATC probe fails;with ERR annunciator appears on the display. Wash display if cleaning cycle is on.
6
Page 12
4.2 Function Groups
The main function and sub-function groups are or ganis ed in a matrix format for configuration and selection of parameters. The main function groups are:
1) Offset adjustment (OFS)
2) Temperature Measurement / compensation settings (Set
3) Control relay 1 configuration (SP1)
4) Control relay 2 configuration (SP2)
5) Control type (Cntr)
6) Current output (mg)
7) Configuration (ConF)
8) Calibration (CAL pH)
o
C)
SETUP
HOLD
SETUP
HOLD
OFS
Cal
SETUP
HOLD
SETUP
HOLD
SETUP
HOLD
°
C
SETUP
HOLD
Sp1 Sp2
SETUP
HOLD
SETUP
HOLD
C
The set-up parameters can be viewed or changed by entering a security code. See section 5.2 for security code information.
4.2.1 How to view operating parameters without access to change them:
SETUP HOLD
000
S.Cd
a) Press the Enter key. The display will prompt the user to
enter a security code (S.Cd). Leave the security code at “000” (do not enter a security code).
b) Press ENTER key again. This allows you to only view (not
change) any sub-functions’ settings.
Press the or key to scroll through the sub-functions.
c) d)
Press the ENTER key at a particular sub-function to view in detail.
e)
Press the ENTER key to return to the sub-function menu. Press the or keys simultaneously (as an Escape key) at any time to return to the
f)
Measurement mode.
Note: To simplify operations, the controller will not display parameters that are not relevant to a particular subfunction. For example, if the user set the controller for limit control, it will not display pulse length/frequency settings.
7
Page 13
4.3 Control Concept
The main function and sub-function groups are organised in a matrix format as shown below. These functions can be accessed via the front keypad for configuration and selection of parameters.
MEASURE
CODE
OFFSET
O
SET C
SET-POINT 1
SET-POINT 2
CONTROLLER
MEASUREMENT
RANGE
CONFIGURATION
CALIBRATION
Set Offset
Switch On/Off ATC
Set value 1
Set value 2
Controller type
Current output
0.20 / 4.0 mA
pH/ ORP Symm.
/ Asymmetrical
Calibrate 1st value
Process temp.
(MTC)
MIN / MAX Function
MIN / MAX Function
N.O. / N.C.
Contact
pH / ORP at 0/4 mA
Alr
Cln
Calibrate
2nd value
Calib. temp.
(MTC)
Hysteresis 1
Hysteresis 2
Proportional
range XP
Alarm delay
Display zero point
and slope
Calibrate Pt100
Pick-up delay
contact 1
Pick-up delay
contact 2
Cycle time bzw.
Max. Frequency
Alarm contact
cont./wiper contact
1000
Dur
Dropout delay
contact 1
Dropout delay
contact 2
Electrode type
Glass/Antimony
Reset to factory
default settings
The controller offers two levels of password protection: (1) for direct access to calibration function and (2) for setting or editing specific controller parameters or functions in the SETUP mode to suit individual requirements.
Note: The passwords are not user-defined and have been set by factory. It is very important to
keep these passwords strictly confidential to avoid unauthorised tampering of the system at all times.
Note: If the user reads parameters only, the controller automatically reverts to Measurement mode
if no key is pressed for 30 seconds.
8
Page 14
5 MEASUREMENT
5.1 Display in Measurement mode
When the controller is initially powered on, it automatically enters into the Measurement mode after the large dual LCD displays al l segments briefly.
The upper display shows the measured pH or ORP value, while lower display shows either temperature value if controller is set for pH measurement or “OrP” if it is set for ORP measurement.
Annunciators at right side of display indi cate whether cont roller is set for pH or mV measurement. If the controller is set for OR P % measurement, neither the pH or mV indicator displays.
Similarly annunciators or icons at top or left side of display show current status of controller, e.g. “MEAS”, “SETUP”, “CAL”, “READY”, e tc. In addition, error messages also guide user in time of any faulty conditions as indicated by “ERR” or icons flashing.
5.1.1 Check electrode performance
To read current electrode slope and offset values withou t changing them:
1) Press the CAL key followed by the ENTER key without adjusting the security code (leave code at “000”). The upper display shows electrode slope. The lower display reading shows the pH reading at 0 mV.
Note: If security code is changed to a value other than “000”, press the ENTER key to return to Measurement mode, without displaying electrode information.
2) Press the ENTER key a second time to return to Measurement mode.
5.1.2 Checking set points
To read current set point values without changing them:
1) Press the RELAY Selection (Rel A/Rel B) key. The upper display shows the set point for Relay A; the lower display shows “SP1”.
2) After two seconds the upper display shows the set-p oint val ue f or R ela y B; th e lowe r di sp la y shows “SP2”.
3) After an additional two seconds, the controller returns to its Measurement mode.
5.2 Security Codes
This controller has two levels of security protection with separate security codes. The first level allows entry into the Calibration mode: security code = 11; the second allows entry into the SETUP mode: security code = 22.
The security codes protect the controller from unauthorised tampering of its current setting. The parameters cannot be changed unless the security code is entered.
9
Page 15
5.2.1 How to enter and change para meters in Calibration mode
1) Press the CAL key. The upper display shows “000” and the
SETUP
HOLD
000
display shows “C A L Or P ”.
4) Press ENTER key again to begin calibration. Refer to Section 6 for full details on calibration.
5) Press the and keys simultaneously (escape) to return to the Measurement mode.
NOTE: To view (not change) the SETUP parameters, press the ENTER key when the security code reads “000”.
5.2.1.1 Clearing the Cali bration security code from the display
The calibration security code automatically resets from “11” to “000” after you return to Measurement mode, so you do not need to clear the security code from the display.
5.2.2 How to enter and change parameters in Advanced Setup mode
1) Press the ENTER key once. The upper display shows “000” and the lower display shows “S.Cd” to prompt you to enter the Advanced Setup security code.
2) Press the or key to scroll the display to Setup security code “22”.
3) Press the ENTER key.
4) If configured for pH control, the upper display re ads “OFS”.
5) If configured for ORP control, the upper disp lay reads “SP1”.
NOTE: Pressing the ENTER key at a value other than “22” causes the controller to revert to the Measurement mode.
6) You are now in the Advanced Setup mode. See Section 7 for complete instructions. To return to Measurement mode, press the and keys simultaneously (escape).
NOTE: If you want to view (not change) set up parameters, press the ENTER key when the security code reads “000”.
5.2.2.1 Clearing the Advanced Setup security code from the display
After you have entered the security code and returned to the Measurement mode, the security code “22” still appears on the display whenever you press the ENTER key. To conceal the security code, you must manually reset the code. To clear the Advanced Setup security code from the display:
1) Press the ENTER key in the Measurement mode.
2) Set to any security code (not 11 or 22) and complete by pressing ENTER.
NOTE: When you enter the Calibration mode with code “11” or Advanced Setup mode with security code “22”, the unit automatically enters into the HOLD mode until you return back to Measurement mode. The HOLD annunciator is displayed at the upper left of the display. While on HOLD, the current output is frozen and set point relays are deactivated.
lower display shows “C.Cd” to prompt the user to enter the Calibration security code.
2) Press the or key to scroll upper display to Calibration security code “11”.
3) Press the ENTER key. If configured for pH control, the display shows “CAL pH”. If configured for ORP control, the
10
Page 16
6 CALIBRATION
You can reach the Calibration mode directly from the Measurement mode by pressing the CAL key and entering the Calibration security code. You can also reach the Calibration mode from the Advanced Setup mode.
6.1 pH Calibration
This unit features 7 preset buffer values (1.01, 4.01, 6.86, 7.00, 9.00, 9.18 and 10.01) for fast auto calibration. When you calibrate this instrument, you need a standar d pH buffer
solution that matches one of these values.
SETUP
HOLD
CAL
pH
pH
HOLD
6.96
7.00
SETUP
HOLD
CAL
4.02
4.00
CAL
READY HOLD
the ENTER key, the electrode and CAL indicator will flash. The controller adjusts the reading to the buffer.
7) After calibrating to a second buffer value, this controller will automatically display slope in the upper displa y and zero point in the lo wer display. You can view the values for zero point and slope without calibration --- see sec tion 5.1.2 for instruction The zero poi nt and slope are re­determined after each calibration.
59.0
7.00
pH
pH
pH
mV
pH
1) Enter Calibration mode. While in the Measurement
mode, press the CAL key and scroll to Calibration code “11”. Press the ENTER key again. The upper and lower display reads “CAL pH”.
NOTE: If the display reads “CAL OrP”, see section 7.6.2 for procedures on how to switch from ORP to pH readings.
2) Press the ENTER key again to begin calibration. The “CAL” indicator appears at the top of display and the buffer indicator appears in the bottom left corner of the display. The upper display shows your present uncalibrated reading. The lower display indicates one of the two buffer values used for the first point calibration. It would either be 7.00 or 6.86.
3) Press the or key to scroll the lower display to the buffer value that matches your buffer solution.
4) Make sure the electrode is in the buffer solution. In ATC, you must also immerse the temperature probe in the buffer solution. Likewise for symmetrical mode, you must also immerse the potential matching pin in the buffer.
5) Press the ENTER key to confirm the buffer value and start the calibration of its first point. The electrode indicator and CAL indicator both flash. The controller automatic al ly adjusts the read ing to match the buf fer value.
6) The lower display will now show its next lower ‘pH’ buffer. Use and keys to select its second buffer from one of the 5 preset values. Remove the electrode from the first buffer, wash and then immerse into the second buffer. Once the upper display stabilises press
11
Page 17
8) Press the ENTER key. If you entered the calibra tion mode using the CAL key, the controller will return to the Measurement mode. If you entered the calibration mode from the Advanced Set-up mode, the controller will return to the sub-function menu.
Note: If there is a calibration error, the controller displays “ERR”. If this happens, press both the and keys (escape) to restart the calibration beginning from step 1.
Note: When calibrating with manual temperature compensation, the controller automatically changes from the preset process temperature to the calibration temperature. After leaving the Calibration mode, the controller switches back to process temperature (for setting the calibration temperature and the process temperature, see section 7.2.3).
6.2 ORP mV Calibration
This mode allows one-poin t calibration.
1) Enter Calibration mode. While in the
Measurement mode, press the CAL key and scroll to Calibration code “11”. Press the
SETUP
HOLD
SETUP
HOLD
CAL
226
C
ENTER key again. The upper and lower display reads “CAL OrP”.
NOTE: If the display reads “CAL pH”, see section
7.6.2 for procedures on how to switch from pH to ORP mV readings.
2) Place sensor in the ORP solution.
3) Press the ENTER key to begin calibration. The “CAL” indicator appears on the display. The upper display shows the current mV output of the electrode without any offset adjustment.
4) Press the and keys to match the mV value to your exact reading.
5) To confirm, press the EN TER key. The controller calculates the correction factor and returns to the Measurement mode.
12
Page 18
6.3 ORP – % Calibration
The ORP % mode allows you to calibrate at two points: a low concentration sample (20%) and a high concentration sample (80%).
To calibrate the controller for ORP %:
1) Enter Calibration mode. While in the Measurement
SETUP
HOLD
SETUP
HOLD
CAL
80
C.1
SETUP
HOLD
20
C.2
indicator flashes and the controller adjusts its display to the 20% relative value and exits to Measurement mode.
Note: If there is a calibration error, the controller displays “ERR”, and it will not automatically return to Measurement mode. If this happens, press the and keys simultaneously (escape) to restart the calibration beginning from step 1.
Note: If you stop calibration after entering the first value, or if calibration is defective, the controller will use its original calibration data.
mode, press the CAL key and scroll to Calibration code “11”. Press the ENTER key again. The upper and lower display reads “CAL OrP”.
NOTE: If the display reads “CAL pH”, see section 7.6.2 for procedures on how to switch from pH to ORP mV readings.
2) Press the ENTER key to begin calibration. The “CAL” indicator appears on the display. The upper display sh ows “80”. The low er display shows “C1” (=calibrated value 80%).
3) Place sensor in the high concentration sample (relative value of 80%).
4) Press the ENTER key to confirm the value. The electrode indicator and CAL indicator flash. The unit automatically adjusts its reading to match the solution value.
5) The display will now show ‘20’ and ‘C2’. Remove the electrode from the first standard, wash it well and then immerse it in the 20% standard. When the display stabilises, press the ENTER key. The ‘CAL’
13
Page 19
7 ADVANCED SET UP MODE
7.1 Electrode Offset (OFS) sub-function
You can perform electrode offset only in the pH mode. Th is mode allows you to change the offset parameter to make reading corrections without removing the electrode from the control system. You can make adjustments of up to ±120 mV.
The controller will add or subtract the value from the measured pH and display the correct val ue . However, if you need to offset the value beyond the average offset you would expect in yo ur application type, consider a full calibration or even electrode replacement.
1) Pull a sample from the system. Record the controller’s pH reading at the time you take the sample.
2) Measure the pH of your sample using a calibrated pH tester, hand-held meter, or bench meter. Record the correct pH value.
3) Enter Advanced set-up mode. Press the ENTER key and scroll to Advanced Set-up security code “22”. Press the ENTER key again.
4) Press the or key to scroll until the upper display shows OFS.
5) Press the ENTER key. The SETUP annunciator appears at the top of the display. The lower display shows the current measured pH value. The upper display shows the current offset value.
6) Press the or key until the pH value coincides with the correct pH value noted in Ste p
2. As the and keys are pressed to adjust the pH value, you will see that the ‘offset’ value in its upper display also changes.
7) Press the ENTER key to enter the offset value.
Note: The offset value is reset during full calibration. See Section 6 for full calibration instructions. Note: The offset parameter is blanked out during ORP operation.
7.2 Setting temperature (Set °C) sub-function
7.2.1 Selecting automatic or manual temperature compensation
SETUP
HOLD
SETUP
HOLD
SETUP
HOLD
Set
1) Enter Advanced Set-up mode. Press the ENTER key and scroll to Advance Set-up security code “22”. Press the ENTER key again.
2) Press and keys to scroll until display shows “Set °C”.
3) Press ENTER key. The lower display shows “Atc”; upper display shows “on” or “oFF” depending on whether or not ATC is selected.
4) Press and keys to toggle between ATC on and off.
5) Proceed with additional Advanced Set-up procedures (press ENTER again) or return to Measurement mode by pressing and keys (escape) simultaneously.
14
Page 20
7.2.2 Temperature calibration (ATC mode only)
SETUP
HOLD
SETUP
HOLD
0.0
25.0
°C
ATC
1) Select “ATC on” as described above in Section 7.2.1.
2) Press ENTER key. The upper display indicates current temperatu re offset. The curr ent measured temperature is shown in the lower display.
3) Compare the current measured temperature on the controller display to a thermometer known to be accurate. Note down the correct temperature value.
4) Press and keys to scroll the lower display to match the correct value. Upper display will now show the offset value. You can offset temperature up to ± 5 °C.
5) Press ENTER key to confirm your selection.
6) Continue with additional Advanced Set-up procedures, or return to the Measurement mode by pressing and keys (escape) simultaneously.
7.2.3 Setting manual temperature compensation
Note: This option is not available when the controller is set for
SETUP
HOLD
SETUP
25.0
HOLD
SETUP
25.0
HOLD
5) Press and keys to adjust the calibration temperature value, between -9.9 and 125°C.
6) Press ENTER key to confirm.
7) Continue with additional Advanced Set-up procedures, or return to Measurement mode by pressing and (escape) simultaneously.
ATC operation.
For manual temperature compensation, two different temperatures: process and calibration, can be input independently. This allows calibration at a temperature other than the process temperature. Example: setting a calibration temperature of 25 solutions at 25
1) Select “ATC oFF” as described above in section 7.2.1
2) Press ENTER key. Upper display shows current process temperature and lower display shows “P.°C” to indicate process temperature.
3) Press and keys to adjust the process temperature value, between -9.9 and 125°C.
4) After setting the process temperature value, press ENTER key. Upper displ ay shows current calibration temperature and lower display shows “C.°C”.
o
o
C allows calibration using standard buffer
C, even if the process temperature is different.
15
Page 21
7.3 Control Rel ay A/ Control Relay B (SP1/SP2) sub-function
The SP1 option sets operating parameters for Relay A; the SP2 option sets operating parameters for relay B. Since these groups have the same set-up parameters, they are described together.
7.3.1 Entering Set point 1 (Set point 2) sub-function
1)
SETUP
HOLD
SETUP
HOLD
Sp1
pH
4.00
Sp1
SETUP
HOLD
Sp1
7.3.3 Choosing High or Low set points
Select a low set point to activate controller when the pH/ORP valu e goes below the low set point; select a high set poin t to activate controller when the value overshoots the high set point. Using both SP1 and SP2, y ou can select lo/lo, lo/ hi, hi/lo, or hi/hi se t points.
1) Follow directions in 7.3.1 to enter Control Relay mode.
2) Press ENTER key. Scroll with the ▲ or ▼ keys until upper display shows Lo or Hi (for low or high set poi n t ) an d the lower displ a y shows SP1 (SP2).
3) Press the or keys to select low (lo) or high (hi) set point for SP1 (SP2).
4) Press ENTER key to confirm your selection.
5) Proceed to 7.3.4 step 3, or return Measurement mode by pressing the and keys simultaneously (escape).
7.3.4 Selecting a hysteresis (dead band) value
Hysteresis prevents rapid contact switching if your value is fluctuating near the set point. It does this by overshooting the set point value to a specified hysteresis value. You can set the hysteresis value from 0.1 to 1.0 pH, 10 to 100mV or 1 to 10%.
Example: You hav e set your high set point at pH 9.0 and your hysteresis limit value at pH 0.5. If your measured value overshoots the high set point pH 9.0, the controller’s relay activates, which in turn activates an external device such as a pump or valve. The actions of the external device will cause the solut ion’s pH to drop. When t he pH drops to 8. will switch off.
Enter Advanced Set-up mode. Press the ENTER key an d
scroll to Advanced Set-up security code “22”. Press the ENTER key again.
2) Press or key to scroll until upper display shows SP1 (SP2).
7.3.2 Selecting set point values
This lets you choose the value that will cause your controller to activate (i.e. pH 9.0).
1) Follow directions in 7.3.1 to enter Control Relay mode. If you are in this mode, skip to step 2.
2) Press ENTER key. The upper display shows current set point value and lower display shows SP1 (SP2).
3) Press or key to enter value for Set point 1 (Set point 2). Your controller will activate at the value you selec t.
4) Press ENTER key to confirm your selection.
5) Proceed to 7.3.3 step 3, or return to Measurement mode by pressing the and keys simultaneously (escape).
5 pH, the relay and hence the pump
16
Page 22
1) Follow directions in 7.3.1 to enter Control Relay mode.
s
SETUP
0.50
HOLD
Hy
4) Press the ENTER key to confirm your selection.
5) Proceed to 7.3.5 step 3, or return to Measurement mode by pressing the and keys simultaneously (escape).
NOTE: Please refer to Appendix 3 for a graphical representation of the Hysteresis.
7.3.5 Setting an on-delay time lag
You can set a time delay for each relay, w hich stops the relay from switchi n g on the moment the set point is exceeded. This controller lets you set a 0 to 2000 second time delay before the relay activates.
SETUP
HOLD
4) Press the ENTER key to confirm your selection.
5) Proceed to 7.3.6 step 3, or return to Measurement mode by pressing the and keys simultaneously (escape).
7.3.6 Setting an off-delay time lag
You can set a time delay for each relay, w hich stops the relay from switching off the moment the value reached the set point and hysteresis. This controller lets you set a 0 to 2000 second time delay before your relay deactivates.
SETUP
HOLD
of seconds (0 to 2000) you select.
4) Press the ENTER key to confirm your selection.
5) Continue with Advanced Set-up mode procedures, or return to Measurement mode by pressing the and keys simultaneously (escape).
0
0
2) Press the ENTER key. Scroll with the or key until the
pH
upper display shows the hysteresis (dead band) value an d the lower display shows “HYS”.
3) Press the or keys to enter your hysteresis value for Set point 1 (Set point 2). Your controller will activate at the value you select.
1) Follow directions in 7.3.1 to enter Control Relay mode.
2) Press the ENTER key. Scroll with the or key until the upper display shows “on delay” time and the lower di splay shows “On.d”.
3) Press the or key to enter on-delay time for Set point 1 (Set point 2). The controller will delay activation for the number of seconds (0 to 2000) you select.
1) Follow directions in 7 .3.1 to enter Control Relay mode.
2) Press the ENTER key. Scroll with the or upper shows “OF.d”.
3) Press the or key to enter on-delay time for Set point 1 (Set point 2) Your controller will delay activation for the number
display shows “off delay” t ime and the lower displ ay
key until the
17
Page 23
7.4 Controller (Cntr) sub-funct ion
You can set the c ontroller’s parameters in this sub-func tion.
7.4.1 Entering Controller sub-function
1) Enter Advanced Set-up mode. Press the ENTER key and scroll to Advanced set-up security code “22”. Press the ENTER key again.
2) Press the or key to scroll until upper display shows “Cntr”.
SETUP
HOLD
SETUP
HOLD
7.4.2 Choosing the cont ro ller type (limit or proprotional)
This mode lets you choose your controller type: limit control, pulse length fre quency proportional control, or control off.
- Use limit control with pumps or values for fast response.
- Use pulse frequency proportional control to operate your pumps smoothly.
- Use pulse length proportional control for precise control of proportioning valves.
- Use control off to operate controller as a monitor only or to keep relays from switching.
1) Follow directions in 7.4.1 to enter Controller mode.
2) Press the ENTER key. The upper display shows the current controller type and the lower display shows “tyP”.
3) Press the or key to select your controller type.
- L.Ct = limit value pickup (on/off control).
- oFF = controller off.
- PLC = pulse length control.
- PFC = pulse frequency control.
4) Press the ENTER key to confirm your selection.
5) Proceed to 7.4.3 step 3, or return to Measurement mode by pressing the and keys simultaneously (escape).
7.4.3 Choosing break/m ake co ntact relay type
Note: If the controller type “oFF” is set, the parameters listed in 7.4.3, 7.4.4 and 7.4.5 are blanked out. This mode lets you determine the re lay state under Non-Alarm condition – dEEN (de-energised) or EN (energised).
SETUP
HOLD
SETUP
HOLD
Plc
SETUP
HOLD
Pfc
18
Page 24
SETUP
HOLD
SETUP
HOLD
Note: If the controller type “oFF” or “L.Ct” is set, the paramet ers listed in 7.4. 4 and 7.4.5 are blanked out.
This mode lets you set a band as a percentage of its full scale value. You can select this range from 10 to 200%, and the lower display shows “PrP”.
SETUP
HOLD
5) Proceed to 7.4.5 step 3, or return to Measurement mode by pressing the and keys simultaneously (escape).
7.4.5 Maximum Pulse Length (tPL) or Maximum Frequency (FPF)
Note: If the controller type “oFF” or “L.Ct” is set, the paramet ers listed in 7.4. 4 and 7.4.5 are blanked out.
This mode lets you set the maximum pulse length or the maximum frequency at which the relay will operate.
SETUP
HOLD
SETUP
HOLD
100
10.0
60
F.Pf
3) Press the or key to choose the period duration or maximum frequency, depending on your mode.
4) Press the ENTER key to confirm your selection and to return to Advanced Set-up mode, or return to Measurement mode by pressing the and keys simultaneously (escape).
NOTE: The alarm contact is always of the f ail-safe type: in the event of power failure the alarm is triggered.
1) Follow directions in 7.4.1 to enter Controller mode.
2) Press the ENTER key. Scroll until the lower display shows “rEL” and the upper display shows the current sele ction (de­energised = dEEN or energised = EN).
3) Press the or key to choose de-energised or energised relay state.
4) Press the ENTER key to confirm your selection.
5) Proceed to 7.4.4 step 3, or return to Measurement mode by pressing the and keys simultaneously (escape).
7.4.4 Selecting proportional range value Xp
1) Follow directions in 7.4.1 to enter Controller mode.
2) Press the ENTER key. Scroll until the upper display shows the proportional range (a number from 10 to 200%), and the lower display shows “PrP”.
3) Press the or key to choose the proportional range value Xp.
4) Press the ENTER key to confirm your selection.
1) Follow directions in 7 .4.1 to enter Controller mode.
2) Press the ENTER key. Scroll until the lower display shows “t.PL” or “F.PF”.
- In PLC (pulse length) mode: The lower display shows “t.PL”
to indicate pulse length. The upper displ ay show s your current pulse length. You can select any value from 0.5 to 20 seconds.
- In PFC (pulse frequency) mode: The lower display shows F.PF to indicate pulse frequency. The upper display shows your current maximum pulse rate. You can select any value from 60 to 120 pulses per minute. When the measured value exceeds the Proportional Band in 7.4.4, the controller will
the relay at this rate.
pulse
19
Page 25
7.5 Current Output (rng) sub-function
This sub-function lets you set the controller current output
SETUP
HOLD
SETUP
4-20
HOLD
SETUP
HOLD
SETUP
HOLD
pH mode, ORP parameters are blocked out. If you are in the ORP mode, pH parameters are blocked out.
1) Follow directions in 7.5.1 to enter Current Output mode.
2) Press the ENTER key. Scroll with the or keys until the upp e r display shows a pH or ORP value and the lower display shows “r.0(4)”.
3) Press the or key to select pH or ORP value to be equivalent to 0(4) mA.
4) Press the ENTER key to confirm your selection.
5) Press the ENTER key to return to Advanced Set-up mode, or return to Measurement mode by pressing the and keys simultaneously (escape).
7.5
.4 Selecting pH/ORP value at 20mA
This parameter lets you choose a pH or ORP value at which th e c ontroller output will be 20 mA. NOTE: If you are in the pH mode, ORP parameters are blocked out. If you are in the ORP mode, pH parameters are blocked out.
1) Follow directions in 7.5.1 to enter Current Output mode.
2) Press the ENTER key. Scroll with the or keys until the upp e r display shows a pH or ORP value and the lower display shows “r.20”.
3) Press the or key to select pH or ORP value to be equivalent to 20 mA.
4) Press the ENTER key to confirm your selection.
5) Press the ENTER key to return to Advanced Set-up mode, or return to Measurement mode by pressing the and keys simultaneously (escape).
14
pH
0
pH
range of this unit.
7.5.1 Entering current output sub-function
1) Enter Advanced Set-up mode. Press the ENTER key and scroll to Advanced Set-up security code “22”. Press the ENTER key again.
2) Press the or key to scroll until the upper display shows “rng”.
7.5.2 Choosing the output type
This parameter lets you choose between 0-20 mA or 4-20 mA output.
1) Follow directions in 7 .5.1 to enter Current Output mode.
2) Press the ENTER key. Scroll with the or key until the upper display shows the output type (0-20 or 4-20), and the lower display shows “out”.
3) Press the or key to select your output type: 0-20 or 4-20 mA.
4) Press the ENTER key to confirm your selection.
5) Proceed to 7.5.3 step 3, or return to Measurement mode by pressing the and keys simultaneously (escape).
7.5.3 Selecting pH/ORP value at 0(4)mA
This parameter lets you choose a pH or ORP valu e at which the controller output will be 0(4) mA. NOTE: If you are in the
20
Page 26
7.6 Configuration (ConF) sub-function
This group of parameters lets you configure the controller to suit your requirements.
7.6.1 Entering the Configuration s ub-f u nction
1) Enter Advanced Set-up mode. Press the ENTER key and scroll to Advanced Set-up security code “22”. Press the ENTER key again.
2) Press the or key to scroll until the upper dis play shows “ConF ”.
SETUP
HOLD
SETUP
HOLD
Ph
Sy
SETUP
HOLD
Ph
Asy ASy
7.6.2 Selecting pH, ORP% or ORP mV measurement
This parameter group lets you select pH, mV, or % readings , and input type symmetrical or asymmetrical. Symmetrical mode is used in conjunction with a potential matching pin – in an electrically noisy environment.
Please refer to section 3, page 6 for a graphical representation.
1) Follow directions in 7.6.1 to enter Configuration mode.
2) Press the ENTER key. Scroll with the or key until the upper display shows the control type (pH, ORP mV or ORP %) and the lower display shows symmetrical (SY) or asymmetrical (ASY) input type. Choose ORP without the mV indicator for ORP% readings.
3) Proceed to 7.6.3, or return to Measurement mode by pressing the and keys simultaneously (escape).
SETUP
HOLD
Sy
SETUP
HOLD
ASy
SETUP
HOLD
SETUP
HOLD
mV
Sy
mV
21
Page 27
7.6.3 Selecting Alarm or Wash function
This function allows you to use the Alarm relay as Wash contact. The Wash contact is used in combination with automatic cleaning systems. During the wash cycle, the analogue output is set on hold.
SETUP
HOLD
SETUP
HOLD
7.6.4 Selecting the alarm time lag if relay 3 has been set to Alarm mode
This parameter group lets you select a period of time before the alarm activates when your set point has been overshot. You can select from 0 to 2000 seconds.
SETUP
HOLD
5) Proceed to 7.6.4 step 3, or return to Measurement mode by pressing the and keys simultaneously (escape).
7.6.5 Selecting steady or pulse contact for the alarm relay
This parameter group lets you select whether the alarm contact wil l operate as a steady contac t or a fleeting (single pulse) contact. Pulse contact closing time is 1 second.
SETUP
HOLD
30
Al.c
SETUP
HOLD
Al.c
1) Follow direction in 7.6.1 to enter Configuration mode.
2) Press the ENTER key until upper display shows “Alr” or “CLn”.
3) Press the or key to choose the desired function.
4) Press the ENTER key to confirm your selection.
1) Follow directions in 7.6.1 to enter Configuration mode.
2) Press the ENTER key. Scroll with the or key until the upper display shows a numerical value (in seconds) and the lower display shows “AL.d”.
3) Press the or key to select how long of an alarm delay (0 to 2000 seconds) you want.
4) Press the ENTER key to confirm your selection.
1) Follow directions in 7.6.1 to enter Configuration mode.
2) Press the ENTER key. Scroll with the and keys until the upper display shows “Stdy” or “FLEt” and lower display shows “AL.C.”.
- AL.C = alarm contact
- StdY = steady contact
- FLEt = fleeting (single pulse) contact
3) Press
4) Press the ENTER key to confirm your selection.
5) Proceed to 7.6.5 step 3, or return to Measurement mode by pressing the and keys simultaneously (escape).
the or key to select steady or pulse contact.
22
Page 28
7.6.6 Configuring the Wash contact if re lay 3 has been set to the Wash mode
SETUP
HOLD
SETUP
HOLD
SETUP
HOLD
7.6.7 Selecting pH sensor type Note: This parameter is blocked out in ORP mode. Proceed to 7.6.6.
Note: After each changeover, the factory calibration data for zero point and slope are
overwritten. Be sure to re cali brate (see section 6).
This parameter group lets you configure the controller for glass or antimony electrodes. Use an antimony electrode for systems with hydrofluoric acid content.
SETUP
HOLD
10.0
10
Glas
El
SETUP
HOLD
El
1) Follow direction in 7.6.1 to enter Configuration mode.
2) Press the ENTER key until upper display shows “Alr” or “CLn”.
3) Press the or key to select the “CLn” function. Press the ENTER key.
4) Press the or key to select the wash cycle (int. 0.1 to
199.9 hours) and press ENTER.
5) Press the or key to select the wash duration (1 to 1999 seconds) and press ENTER.
Note: During wash cycle, the controller is set to HOLD. For safety reasons, the HOLD function starts 5 seconds before the start of wash cycle and will continue for 10 seconds after the wash cycle is completed.
1) Follow directions in 7.6.1 to enter Configuration mode.
2) Press the ENTER key. Scroll with the or key until the upper display shows “GLAS” or “AnTY” and lower display shows “EL”.
- EL = electrode
- GLAS = glass
- AnTY = antimony
3) Press the or key to select glass or antimony.
4) Press the ENTER key to confirm your selection.
5)
Proceed to 7.6.6 step 3, or ret
pressing the and keys simultaneously (escape).
urn to Measurement mode by
23
Page 29
7.6.8 Reverting to factory default settings
Use this parameter to reset all settings to factory default.
1) Follow directions in 7.6.1 to enter Configuration mode.
2) Press the ENTER key until the upper display shows “no”, “FCt” or “CAL”, and the lower display shows “deF” (default).
3) Press the or key to select “no” for using the old values, “FCt” to reset all settings to factory defaults or “CAL” to reset calibration values only to the factory default values.
WARNING: Changing from “no” to “FCt” or “CAL” and pressing the ENTER key resets all settings or the calibration values respectively to factory defaults .
4) Press the ENTER key to confirm your selection and to return to Advanced Set-up mode, or return to Measurement mode by pressing the and keys simultaneously (escape).
SETUP
HOLD
SETUP
HOLD
SETUP
HOLD
Cal
7.7 Calibration (CAL) sub-function
The calibration procedure in Advanced Set-up mode is identical to the procedure in the Calibration mode. The only di fference is that the controller will revert back to Set-up mode (instead of Measurement mode) after calibrat ion is completed.
7.7.1 Entering the Calibration mode from Advanced Set-up mode
1) Enter Advanced Set-up mode. Press the ENTER key and scroll to Advanced Set-up security code “22”. Press the ENTER key again.
2) Press the or key to scroll until the upper dis play shows “CAL”.
3) See Sect i on 6 for complete calibratio n procedures.
24
Page 30
8 AUTO/MANUAL MODE
Regardless of the mode, you can control devices connected to Relay A or Relay B from the front panel of this controller. In Automatic mode, the controller’s set point values act ivate the relays. In Manual mode, you have manual control of the relays so you can prime the pump or check pump status without operating the entire system.
8.1 Auto mode (mode after switch-on)
In this mode, the controller set-point values activate the relays. To view the set-point values:
1) Press the RELAY SELECTION (Rel A/Rel B) key. The upper display shows your set-point value for Relay A; the lower display show s “SP1”.
2) After two seconds the upper display shows your set-point value for Relay B; the lower display shows “SP2”.
3) After an additional two seconds the controller will return to Measurement mode.
8.2 Manual mode
In this mode, you can manually turn on and off the control devices connected to Relay A or Relay B or both.
1) Press the RELAY CON T ROL (auto/manu) key. The upper display shows “000”; the lower display shows “S.Cd” to prompt you to enter the Advanced Set-up code.
2) Press the or key to scroll the upper display until it reads “22”.
3) Press the ENTER key. The manual ind ica tor by the RELAY CONTROL key lights up.
Note: Pressing ENTER key at a value other than “22” will ca use the controller to revert to Measurement mode, and the relays will remain in automatic mode.
4) Press the RELAY SELECTIO N key to select either Relay A or Relay B. The LED next to the currently selected relay (A or B) will light.
The manual control options now available will depend on the control type (limit, pulse frequency, or pulse length) you selected set in section 7.4.2.
- If you selected Limit control: Upper display reads the current measured value. Lower
display shows “oFF” or “on” depending on the relay status of the currently selected relay.
- If you selected Pulse length control: Upper display reads the current measured value.
Lower display shows the current pulse duration.
- If you selected Pulse frequency control: Upper display reads the current measured value.
Lower display shows the current pulse frequency.
5) Press the or key to change the Relay on/off status, pulse length, or pulse frequency. LED indicators at the right of controller will change between Red and Green to indicate Relay status.
Note: If you wish to manually chang e the sta tus of both relays, press t he RELAY SELECTION key at this point and repeat step 5 for the second relay. This first rela y will remai n under manual control while you set the second relay.
6) Press the RELAY CONTROL key to return to Measurement mode. Relays are now back under automatic control.
25
Page 31
9 TECHNICAL SPECIFICATIONS
pH Range 0.00 to 14.00 pH Resolution Relative Accuracy mV Range 0 to 100.0% / -1000 to +1000 mV
Resolution Relative Accuracy
Temperature - 9.9 to + 125.0 oC
Resolution Relative Accuracy
Sensor Pt 100 /Pt 1000 (jumper selectable) Temperature Compensation Auto / manual (reference at 25.0 oC)
Set-point and Controller Functions
Controller characteristics proporational controller Adjustabl e period with pulse length controller 0.5 t o 20 sec. Adjustable period with pulse frequency controller 60 to 120 pulses/min Pickup / Dropout delay 0 to 2000 sec. Switching pH hysteresis 0.1 to 1 pH Switching ORP hysteresis 1 to 10.0% / 10 to 100 mV Contact outputs, controller 2 potential-free change-over contacts Switching voltage / current / power max. 250 VAC / max. 3A / max. 600 VA
Alarm Functions
Function (switchable) Latching / pulse Wash cycle 0.1 to 199.9 hrs Wash duration 1 to 1999 s Pickup delay 0 to 2000 s Switching voltage / current / power max. 250 VAC / max. 3A / max. 600 VA
Electrical Data and Connections
Power Requirements 110 / 220 VAC (jumper selectable) Frequency 48 to 62 Hz Power Consumption max. 7 VA Signal Output / Load 4 to 20 mA, galvanically isolated Load Max. 600 pH / ORP input BNC, screw terminals Connection terminal Terminal blocks 5-pole / 17-pole, removable Mains fuse / fine wire fuse slow-blow 250 V / 100 mA
EMC Specifications
Emissions According to EN 50081-1 Susceptibility According to EN 50082-1
Environmental Conditions (For Indoor Use)
Altitude Up to 2000 m Maximum Relative Humidity
Power Supply Transient Overvoltages Category II
Pollution Degree Category II Ambient temp. operating range -10 to +50 oC
0.01 pH ± 0.01 pH
0.1% / 1mV ± 1 mV
0.1 ± 0.5
limit controller Function (switchable) pulse length/ frequency controller
80% up to 31 at 40
Mains suppl y voltage fl uctuations not exceeding +10% of the nominal voltage
o
C
o
C
o
o
C
C decreasing linearly to 50%
26
Page 32
Mechanical Specifications
Dimensions (control panel housing - L x H x W) 175 x 96 x 96 mm Weights (control panel housing) max. 0.7 kg Material ABS with polycarbonate (front housing) Insulation (Front / Housing) IP 54 / IP 40
27
Page 33
10 ACCESSORIES
Replacement Unit
Product Description Thermo Scientific
Code no.
Alpha-pH1000 pH/ORP Controller/Transmitter with 0/4-20 mA output and 110VAC Setting
Alpha-pH1000 pH/ORP Controller/Transmitter with 0/4-20 mA output and 220VAC Setting
TSPHCTP1001 ECPHCTP1001
TSPHCTP1002 ECPHCTP1002
Assembly Accessories
Product Description Code no.
pH Combination Electrode with Pt 100 RTD and PMP EC100GTSO05B pH Combination Electrode with HF resistant glass (w/o ATC & PMP) ECARTSOHF05B pH Combination Electrode with PMP (w/o ATC) ECARGTSO05B
Eutech
Code no.
pH Combination Electrode with high temperature resistance (110 oC) and high pressure resistance (9 bar) (w/o ATC & PMP)
pH Combination Electrode (w/o ATC & PMP) ECARTSO05B ORP Gold Electrode with PMP but w/o ATC ECHTAUTSO05B ORP Platinum Electrode with PMP but w/o ATC ECHTPTTSO05B
ECARHTTSO05B
Note: Above pH/ORP electrodes withstand up to 6 bar pressure except ECARHTTSO05B.
These electrodes have integral 5m measuring cable terminating with a BNC connector. Please contact your authorised distributor or dealer for the prices of extension measuring cables and other accessories like tee joints, electrode assembly, and calibration solutions.
28
Page 34
11 GENERAL INFORMATION
11.1 Warranty
Thermo Scientific warrants this product to be free fr om sig nificant deviations in material and workmanship for a period of one year from the date of purchase. If repair is necessary and has not been the result of abuse or misuse within the warranty period, please return by freight pre­paid and amendment will be made without any charge. Our Customer Service Dept. will determine if product problem is due to devia tions or customer abuse. Out of warranty products will be repaired on a charge basis.
11.2 Packaging
The instrument is packaged in a corrugated box with a warranty card, instruction manual and the following accessories:
- 17-way and 5-way (right-angled) terminal block [1 unit each]
- side threaded rod with catch [2 units]
- receptacle cable lug [1 unit]
- rubber gasket [1 unit]
11.3 Return of Goods
Authorisation must be obtained from our Customer Service Dept. to issue a RMA ( Return of Material Authorisation) number before returning items for any reason. When applying for authorisation, please include data requir ing the reason of return. Items must be carefully packed to prevent damage in shipment and insured against possible damage or loss. Thermo Scientific will not be responsible for any damage resulting from careless or insufficient packing.
Warning: Shipping damage as a result of inadequate packaging is the user/distributor’s
responsibility, whoever applicable. Please follow the guidelines below before shipment.
11.4 Guidelines for Returning Unit for Repair
Use the original packaging material, if possible when shipping the unit for repair. Otherwise wrap it with bubble pack and use a corrugated box for better protection. Include a brief description of any faults suspected for the convenience of Customer Service Dept., if possible.
29
Page 35
12 APPENDICES
12.1 Appendix 1
Before opening the unit, ENSURE that the power c able is physically separated from the power supply.
Jumper Positions - Internal to the cont r oller
JP 1 Selects the input voltage 220 VAC. JP 2 Selects the input voltage 110 VAC. JP 3 Solder bridge selects between Pt100 and Pt1000.
Fuse Note that there is a fuse (slow-blow 100mA) internal to the
controller. Replace fuse with the recommended type only.
30
Page 36
12.2 Appendix 2
The following table shows the various pH values at different temperature of the solution during calibration.
Temperature (oC) pH 1.00 pH 4.01 pH 6.86 pH 7.00 pH 9.00 pH 9.18 pH 10.01
0 0.96 4.01 6.98 7.12 9.33 9.47 10.32 5 0.99 4.01 6.95 7.09 9.24 9.38 10.25
10 0.99 4.00 6.92 7.06 9.16 9.32 10.18 15 0.99 4.00 6.90 7.04 9.11 9.27 10.12 20 1.00 4.00 6.88 7.02 9.05 9.22 10.06 25 1.01 4.01 6.86 7.00 9.00 9.18 10.01 30 1.01 4.01 6.85 6.99 8.95 9.14 9.97 35 1.01 4.02 6.84 6.98 8.91 9.10 9.93 40 1.01 4.03 6.84 6.97 8.88 9.07 9.89 45 1.01 4.04 6.83 6.97 8.85 9.04 9.86 50 1.01 4.06 6.83 6.97 8.82 9.01 9.83 55 1.01 4.08 6.83 6.97 8.79 8.99 9.81 60 1.02 4.10 6.84 6.98 8.76 8.96 9.79 70 1.02 4.12 6.85 6.99 8.72 8.92 9.76 80 1.02 4.16 6.86 7.00 8.68 8.89 9.74 90 1.02 4.20 6.88 7.02 8.65 8.85 9.73
31
Page 37
12.3 Appendix 3
Simple Explanation on the Function of Hysteresis
S P 1 S e t t o L O S P 2 S e t t o H I
4 . 0 4 . 5 9 . 5 1 0 . 0
S P 1
F O R W A R D D I R E C T I O N
R E V E R S E D I R E C T I O N
R E L A Y O N
R E L A Y O F F
7.0
S P 2
H Y S T E R E S I S B A N D ( D E F A U L T = 0 . 5 p H )
The controller relay activates when the set-point is reached. In the reverse direction, it does not de-activate when the value reaches the set-point. In stead, it continues to be active till the value reaches the amount set by the Hysteres is band.
32
Page 38
12.4 Appendix 4
General Instructions Concerning C ontroller Setting
MIN function MAX function
-Xw Xp
Prop. band Prop. band
SP 1
100% Yh
50%
0%
SP 2
Xp
+Xw
Control characteristic of P-Controller as proportional controller
MIN function MAX function
100% Yh
50%
-Xw
Xp = 0
0%
SP 2SP 1
Xp = 0
Control characteristic of P-Controllers as limit value switch
Characteristics of Pulse Length and Pulse Frequency
Relay
Pulse Le ngth T
tON
ON
OFF
tOFF
Controller signal of Pulse len gth controller
Time [s]
+Xw
33
Page 39
The output relay of the pulse length controller is clock-timed. The switching period T re mains constant. Depending on the divergence from the limit value, the switch-on time t
is increased
ON
or decreased in accordance with the proportional range Xp.
The following applies:
t
+ t
ON
greater div e rgence greater t
Xp exceeded → t
= T (Const.)
OFF
ON
= T (relay remains picked up)
ON
Relay
Pulse Length T
tON
ON
tOFF
OFF
Tim e [s]
Controller signal of Pulse Frequency Controllers
The output relay of the pulse frequency controller is clock-timed. The pulse duration t constant at 250mS.
Depending on the divergence from the limit value, the frequency (
1
/T) is increased or decreased
in accordance with the proportional range Xp.
The following applies:
t
= Const. (250 msec.)
ON
greater div e rgence greater f (greater frequency) Xp exceeded max. frequency
remains
ON
34
Page 40
Page 41
Page 42
Water Analysis Instruments
North America
166 Cummings Center Beverly, MA 01915 USA Toll Free: 1-800-225-1480 Tel: 1-978-232-6000 Dom. Fax: 1-978-232-6015 Int’l Fax: 978-232-6031
Europe
P.O. Box 254, 3860 AG Nijkerk Wallerstraat 125K, 3862 CN Nijkerk, Netherlands Tel: (31) 033-2463887 Fax: (31) 033-2460832
Asia Pacific
Blk 55, Ayer Rajah Crescent #04-16/24, Singapore 139949 Tel: 65-6778-6876 Fax: 65-6773-0836
www.thermo.com/process
© 2009 Thermo Fisher Scientific Inc.
All rights reserved. Thermo Fisher Scientific Inc.
68X216801 Rev 3
Loading...