Thermo Fisher Scientific pH 1000 User Manual

Alpha pH/ORP 1000
Process Controller/Transmitter
R
Afast, Cahn, ionplus, KNIpHE, No Cal, ORION, perpHect, PerpHecT, PerpHecTion,
AQU pHISA, pHuture, Pure Water, Sage, Sensing the Future, SensorLink, ROSS, ROSS Ultra, Sure-Flow, Titrator PLUS and TURBO2 are registered trademarks of Thermo Fisher.
1-888-pHAX-ION, A+, All in One, Aplus, AQUAsnap, AssuredAccuracy, AUTO-BAR, AUTO­CAL, AUTO DISPENSER, Auto-ID, AUTO-LOG, AUTO-READ, AUTO-STIR, Auto-Test, BOD AutoEZ, Cable-Free , CERTI -CAL, CI SA, Dat aCOLLECT , DataPLU S, dig ital Log R, Di rectCa l, DuraProbe, Environmental Product Authority, Extra Easy/Extra Value, FAST QC, GAP, GLPcal, GLPcheck, GLPdoc, ISEasy, KAP, LabConnect, LogR, Low Maintenance Triode, Minimum Stir Requirement, MSR, NISS, One-Touch, One-Touch Calibration, One-Touch Measurement, Optimum Results, Orion Star, Pentrode, pHuture MMS, pHuture Pentrode, pHuture Quatrode, pHuture Triode, Quatrode, QuiKcheK, rf link, ROSS Resolution, SAOB, SMART AVERAGING, Smart CheK, SMART STABILITY, Stacked, Star Navigator 21, Stat Face, The Enhanced Lab, ThermaSense, Triode, TRIUMpH, Unbreakable pH, Universal Access are trademarks of Thermo Fisher.
Guaranteed Success and The Technical Edge are service marks of Thermo Fisher.
Preface
This manual serves to explain the use of the Alpha pH1000 pH/ORP process controller/transmitter. The manual functions in two ways, firstly as a step by step guide to help the user operate the instrument. Secondly, it serves as a handy ref erence guide. This instruction manual is written to cover as many anticipated applications of the Alpha pH1000 pH/ORP process controller/transmitter. If you have doubts in the use of the instrument, please do no t hesitate to co ntact the nearest Authorized Distributor.
Thermo Scientific will not accept any responsibility for damage or malfunction of the unit due to improper use of the instrument.
The information presented in this manual is subject to change without notice as improvements are made, and does not represent a commitment on part of Thermo Scien tific.
Copyright All rights reserved
©
2009
Safety Information
The Controller/ Transmitter shall be installed and operated only in the manner specified in this Instruction ma nual. Only skilled, trained or authorized person should carry out installation, setup and operation of the instrument.
Before powering up the unit, make sure that power source is connected to it as specified on the top label. Failure to do so may result in permanent damage of the unit.
The unit has live and exposed parts inside. If it has to be opened, make sure that power to the unit is off and disconnected.
The unit is Fuse protected. In the event the fuse has to be replaced, use only those as specified in the manual.
TABLE OF CONTENTS
1 INTRODUCTION ___________________________________________________ 1
1.1 Description of Unit............................................................................................1
1.2 Applications......................................................................................................1
2 ASSEMBLY AND INSTALLATIONS ____________________________________ 2
2.1 Measurement and Control System..... ..............................................................2
2.2 Unit Dimensions...............................................................................................2
3 ELECTRICAL CONNECTIONS ________________________________________ 3
3.1 Connection Diagram ........................................................................................3
3.2 Back Panel....................................................................................................... 5
4 OVERVIEW _______________________________________________________ 6
4.1 Keypad and Display .........................................................................................6
4.2 Function Groups...............................................................................................7
4.3 Control Concept...............................................................................................8
5 MEASUREMENT ___________________________________________________ 9
5.1 Display in Measurement mode..... .... ... ........................................ ... ... ...............9
5.2 Security Codes.................................................................................................9
6 CALIBRATION____________________________________________________ 11
6.1 pH Calibration ................................................................................................11
6.2 ORP – mV Calibration...... ... ........................................ ... ... .............................12
6.3 ORP % Calibration......................................................................................13
7 ADVANCED SET UP MODE _________________________________________ 14
7.1 Electrode Offset (OFS) sub-function.......................................................... .... 14
7.2 Setting temperature (Set
7.3 Control Relay A/ Control Relay B (SP1/SP2) sub-function............................ 16
7.4 Controller (Cntr) sub-function.........................................................................18
7.5 Current Output (mg) sub-function .................................................................. 20
7.6 Configuration (ConF) sub-function.................................................................21
7.7 Calibration (CAL) sub-function....................................................................... 24
o
C) sub-function ...... ... ... ... ...................................... 14
8 ADVANCED SET UP MODE _________________________________________ 25
8.1 Auto mode (mode after switch-on)............ ... ... ........................................ ... ... . 25
8.2 Manual mode .................................................................................................25
9 TECHNICAL SPECIFICATIONS ______________________________________ 26 10 ACCESSORIES ___________________________________________________ 28 11 GENERAL INFORMATION __________________________________________ 29
11.1 Warranty.........................................................................................................29
11.2 Packaging ......................................................................................................29
11.3 Return of Goods.............................................................................................29
11.4 Guidelines for Returning Unit for Repair ........................................................29
12 APPENDICES ____________________________________________________ 30
12.1 Appendix 1 ..................................................................................................... 30
12.2 Appendix 2 ..................................................................................................... 31
12.3 Appendix 3 ..................................................................................................... 32
12.4 Appendix 4 ..................................................................................................... 33
1 INTRODUCTION
1.1 Description of Unit
Thank you for purchasing the ¼ DIN Alpha pH1000 pH/ORP process controller/transmitter. This unit is used for measuring either pH or ORP parameter one at a time, and the operational mode is switchable from the menu. You can use this unit to measure pH or ORP with limit control. This controller has many user-friendly and safety features which include:
Menu-driven program that simplifies set-up
Built-in non-volatile memory backup to ensure that calibration and other information
are not erased if power supply fails
Press-button two-point calibration and electrode offset adjustment from the keypad
Automatic temperature compensation (ATC)
Manual temperature compen sation setting without the ATC probe, with independent
setting for calibration and pr ocess temperature
0 to 2000 second time delay adjustment on all relays – minimise false alarms
Separately adjustable high and low set point hysteresis (dead bands) prevent
chattering of relays around the set points
Asymmetrical/symmetrical input for pH/ ORP operati o n
Three control modes: limit, proportional pulse length or proportional pulse frequency
Large dual display LCD for easy reading with clear multiple annunciators, alarm
status, operational and error messages
Two switching contacts as set-point tr iggering relays
A third rela y that can be be set up to perform either Alarm or Wash function
When set to work as Alarm relay, alerts you when readings exceed set-point limits after
a set time buffer, and if the Pt100/Pt1000 wires are broken or disconnected during the ATC function
Programmable duration and frequency in the Wash mode
Hold function freezes output current (0/4…20mA) and releases control relays
LED indicators signal control activities to monitor controller status from a distance
Protection against electromagnetic interference – galvanically isolated 0/4…20mA
output provides safety for datalo
Choice of 7 buffers from USA or NIST standards
gging and control purposes
1.2 Applications
Use this controller in panel mounted enclosures for applications such as water treatment and monitoring, galvanic-decontamination, chemical processing food processing, clean or waste water control and neutralisation process.
1
2 ASSEMBLY AND INSTALLATIONS
6
2.1 Measurement and Control System
A typical measurement system consists of:
a pH/ORP process controller
a pH/ORP combination electrode with integrated or separate temperature sensor Pt
100/1000,
an immersion, flow or process assembly with or without a potential matching pin (PMP)
a final control element such as pump or valve
a recorder
an appropriate pH or ORP measurement cable
Flow Assembly
PH Controller
Chart Recorder
Process Assembly
with Electrode
Measurement Cable
Power Mains
(220/110 VAC)
2.2 Unit Dimensions
Flat Gasket (1mm)
(To be Inserted By Customer)
Note: The Taped Corners Have to Be On Top
92 + 0.5
92
56
92 + 0.5
Mounting Cut-Out
The field-tested control panel housing is 96 x 96 mm; with protection class IP 54 (front).
32
max. 45
max. 175
9
2
3 ELECTRICAL CONNECTIONS
3.1 Connection Diagram
ENSURE electrical mains is disconnected before proceeding.
S/
V
S/S/
pH
AC:
Power Mains
2
1
NL
PE
18 19 20
Pt 100
3
Relay A
4
21 22
PE/S
Potential Matching Pin (PMP)
Relay B
6
5
78 9
Signal Input pH/ORP
Alarm
10
11
BNC
12 13
14
PE/S
151715
Signal OutputHold Input
pH
-
mA
+
16
*)indicated contact positions are for currentless conditions
3
NLS
PE
22
1
3
2
Potential
Matching Pin (PMP)
NLS
PE
22
1
3
2
Potential
Matching Pin (PMP)
0 / PA
21
0 / P A
21
20
20
Temp.
19 18
Temp.
19 18
pH
s
Measuring Cable
pH Combination
Electrode
pH
Measuring Cable
pH Combination
Electrode
BNC
BNC
Symmetrical high-impedanc e connections
Asymmetrical connect ions
4
3.2 Back Panel
A
ENSURE electrical mains is disconnected before proceeding.
The back panel consists of two connectors. The first connector is the 17-way PCB edge connector and the other is the 5-way connector.
Connection for the 17-way screw terminals (from left to right):
1. AC mains live wire 10. Alarm/Wash relay working position (NO)
2. AC mains neutral wire 11. Alarm/Wash relay common
3. AC mains protective earth wire 12. Alarm/Wash relay resting position (NC)
4. Low set relay resting position (NC) 13. Hold function switch terminal 1
5. Low set relay common 14. Hold function switch terminal 2
6. Low set relay working position (NO) 15. No connection
7. High set relay resting position (NC) 16. 4 - 20 mA for -ve connection
8. High set relay common 17. 4 - 20 mA for +ve connection
9. High set relay working position (NO)
Connections for the 5-way screw terminals:
18. Pt100/Pt1000 lead 1 terminal
19. Pt100/Pt1000 sense lead terminal
20. Pt100/Pt1000 lead 2 terminal
Note: If using a two-wire RTD, short terminal 19 to terminal 18.
Pt 100/Pt 1000 is selectable via an internal jumper. Factory default is Pt100. See Appendix 1 for directions on switching the R TD type.
21. pH/ORP (potential matching pin)
22. pH/ORP (shield)
pH/mV
FUSE 250V AC
(F)
100mA
L
N
PE
RELAY A
Pt100/
Pt1000
RELAYB
0V/
S/
PAL
LARM / WASH
HOLD
+
-
NC
Important: The Alarm relay functions as an “Active Low” device i.e. it switches OFF under Alarm condition. Therefore, the Alarm display device should be connected to the ‘NC’ contacts of the relay (11 and 12). In the Wash mode, the relay works on a positive mode. Therefore, the Wash device should be connected to contacts 10 and 11.
5
4 OVERVIEW
4.1 Keypad and Display
4.1.1 Keypad
Perform rapid 2-point ca l ibration
Allows entry to Set up mode
Select individual functions within the function group of Set up mode
Store input data in the Set up mode
Start calibration in the calibration mode
Select various function groups in the Set up mode.
Set parameters and numerical values in sub functions of Set up mode
If pressed continuously, the setting speed increases
Return to the Measurement mode when both keys are pressed together
Return to the Measurement mode when both keys are pressed together
Switch beween AUTO and MANUAL relay operation mode
Display limit set-point values for the switch contacts in AUTO relay operation
mode
Switch between RELAY A and RELAY B in MANUAL relay operation mode
4.1.2 Display
The LCD display features two numerical displays that show status messages and measured values for easy, quick reference. The display provides short-text information for setting parameters and configuration.
MEAS: Measurement mode
SETUP: Set-up mode of function groups
CAL: Calibration mode of pH/ORP and temperature
READY: Comes on after a succe ssful calibrat ion
HOLD: Relay position and current output are frozen
ATC: Comes on in the ATC mode. Disappears in the
Manual temperature Compensation mode.
“ATC” flashes if the temperature probe is faulty in its ATC mode
ERR: Error or alarm indicator
READY HOLD
7.00
ERR
25.0
Display for RELAY A/B. Green LED indicates measured value within limit
CONMEAS CALSET UP
pH
°C AT C
while RED LED indicates measured value outside limit.
Alarm display if limit value overshoot or the ATC probe fails;with ERR annunciator appears on the display. Wash display if cleaning cycle is on.
6
4.2 Function Groups
The main function and sub-function groups are or ganis ed in a matrix format for configuration and selection of parameters. The main function groups are:
1) Offset adjustment (OFS)
2) Temperature Measurement / compensation settings (Set
3) Control relay 1 configuration (SP1)
4) Control relay 2 configuration (SP2)
5) Control type (Cntr)
6) Current output (mg)
7) Configuration (ConF)
8) Calibration (CAL pH)
o
C)
SETUP
HOLD
SETUP
HOLD
OFS
Cal
SETUP
HOLD
SETUP
HOLD
SETUP
HOLD
°
C
SETUP
HOLD
Sp1 Sp2
SETUP
HOLD
SETUP
HOLD
C
The set-up parameters can be viewed or changed by entering a security code. See section 5.2 for security code information.
4.2.1 How to view operating parameters without access to change them:
SETUP HOLD
000
S.Cd
a) Press the Enter key. The display will prompt the user to
enter a security code (S.Cd). Leave the security code at “000” (do not enter a security code).
b) Press ENTER key again. This allows you to only view (not
change) any sub-functions’ settings.
Press the or key to scroll through the sub-functions.
c) d)
Press the ENTER key at a particular sub-function to view in detail.
e)
Press the ENTER key to return to the sub-function menu. Press the or keys simultaneously (as an Escape key) at any time to return to the
f)
Measurement mode.
Note: To simplify operations, the controller will not display parameters that are not relevant to a particular subfunction. For example, if the user set the controller for limit control, it will not display pulse length/frequency settings.
7
4.3 Control Concept
The main function and sub-function groups are organised in a matrix format as shown below. These functions can be accessed via the front keypad for configuration and selection of parameters.
MEASURE
CODE
OFFSET
O
SET C
SET-POINT 1
SET-POINT 2
CONTROLLER
MEASUREMENT
RANGE
CONFIGURATION
CALIBRATION
Set Offset
Switch On/Off ATC
Set value 1
Set value 2
Controller type
Current output
0.20 / 4.0 mA
pH/ ORP Symm.
/ Asymmetrical
Calibrate 1st value
Process temp.
(MTC)
MIN / MAX Function
MIN / MAX Function
N.O. / N.C.
Contact
pH / ORP at 0/4 mA
Alr
Cln
Calibrate
2nd value
Calib. temp.
(MTC)
Hysteresis 1
Hysteresis 2
Proportional
range XP
Alarm delay
Display zero point
and slope
Calibrate Pt100
Pick-up delay
contact 1
Pick-up delay
contact 2
Cycle time bzw.
Max. Frequency
Alarm contact
cont./wiper contact
1000
Dur
Dropout delay
contact 1
Dropout delay
contact 2
Electrode type
Glass/Antimony
Reset to factory
default settings
The controller offers two levels of password protection: (1) for direct access to calibration function and (2) for setting or editing specific controller parameters or functions in the SETUP mode to suit individual requirements.
Note: The passwords are not user-defined and have been set by factory. It is very important to
keep these passwords strictly confidential to avoid unauthorised tampering of the system at all times.
Note: If the user reads parameters only, the controller automatically reverts to Measurement mode
if no key is pressed for 30 seconds.
8
Loading...
+ 29 hidden pages