The large membrane keypad makes the instrument easy to use. Each button, when pressed,
has a corresponding graphic indicator on the LCD. See Figure 1. Some buttons have several
functions depending on its mode of operation.
Key Function
ON/OFF Powers meter on and off. The meter powers on in the mode that you were last using.
HOLD Freezes the measured reading. To activate, press HOLD while in measurement mode.
MODE 1. Selects the measurement parameter. Press MODE to toggle between pH; DO %
CAL/MEAS Toggles between Calibration and Measurement mode.
ENTER 1. Press to confirm your calibration values in calibration mode.
▲/▼
SETUP Takes you into the SETUP mode. This mode lets you customise meter preference and
For example, if you shut the meter off in mg/L measurement mode, the meter will be in
mg/L measurement mode when switched on. Note: The meter shuts off automatically
after 20 minutes of the last key press.
To release, press HOLD again.
NOTE: When auto endpoint feature is switched on, it automatically holds reading after 5
seconds of stability. The HOLD indicator appears on the display. Press HOLD again to
release.
saturation; and DO mg/L.
2. While in CAL mode under pH or mg/L (parts per million – ppm) measurement status,
this key will toggle between pH or mg/L (or ppm) calibration and temperature
calibration (refer to Section 4.2 in page 8).
1. In pH measurement mode, press CAL/MEAS to enter pH calibration mode.
2. In DO % saturation measurement mode, press CAL/MEAS to enter DO %
saturation calibration mode.
3. In DO mg/L (or ppm) measurement mode, press CAL/MEAS to enter DO mg/L (or
ppm) calibration mode.
4. To abort calibration, press CAL/MEAS key again to go back to measurement
mode.
While in advanced SETUP mode: Press CAL/MEAS to return to main menu from sub
menus. Press CAL/MEAS again to return to return to measurement mode from main
menu.
2. While in SETUP, pressing the ENTER key confirms selections.
In CALibration & SETUP mode:
In DO mg/L (or ppm) and temperature calibration mode, press up or down keys to
increase or decrease the values.
In SETUP mode, press up or down keys to scroll through the setup subgroup programs.
Within the SETUP sub-menu group, press up or down keys to change values.
defaults, view calibration and electrode data.
LIGHT. Press to activate backlit display.
3
Page 9
Instruction Manual PD 300
3 PREPARATION
3.1 Inserting the Batteries
Four AAA batteries are included with your meter.
1. Use a Philips screwdriver to remove the two screws holding the battery cover. See Figure
3 below.
2. Remove battery cover to expose batteries.
3. Insert batteries. Follow the diagram inside the cover for correct polarity.
4. Replace the battery cover into its original position and secure into place using the two
screws removed earlier.
Figure 3: Back panel of meter showing meter compartment
4
Page 10
Instruction Manual PD 300
3.2 Probe Information
Your meter includes two probes:
• pH electrode with BNC connector.
• DO/temperature probe with a notched 6-pin connector
The temperature sensing element built into the DO probe will also compensate for pH
readings as long as both probes are in your solution at the same time.
You can use any standard pH electrode with a BNC connector with this meter. DO probes, “3-
in-1” pH electrodes with a built-in temperature element, and temperature probes require a
notched 6-pin connector (see Figure 4). For replacement probes, see the Accessories section.
If you want to use a “3-in-1” pH probe with a built-in temperature element, or if you want to use
a separate temperature probe, you will need to disconnect the DO probe to allow for
connection of the separate temperature sensor.
NOTE: Keep connectors dry and clean. Do not touch connector with soiled hands.
To connect the pH electrode:
1. Slide the BNC connector of the probe over the BNC connector socket on the meter.
Make sure the slots of the connector are in line with the posts of the socket. Rotate and
push the connector clockwise until it locks.
2. To remove electrode, push and rotate the connector counterclockwise. While holding
onto the metal part of the connector, pull it away from the meter.
See Figure 4.
CAUTION: Do not pull on the probe cord or the probe wires may be damaged.
5
Page 11
Instruction Manual PD 300
To connect the DO/ temperature probe:
1. Line up the notch and 6 pins on the probe connector with the holes in the connector
located on the top of the meter. Push down and turn the metal locking ring clockwise to
lock the probe connector into place. See Figure 4.
2. To remove probe, unscrew the metal locking ring by turning the ring counterclockwise
and slide up the probe connector. While holding onto metal sleeve, pull probe away from
the meter.
NOTE
Do not substitute other DO probes. For a replacement DO probe, see the “Accessories”
section. Follow the same directions to connect an optional separate temperature element.
CAUTION: Do not pull on the probe cord or the probe wires may be damaged.
Figure 4: Connecting pH and DO/temperature probes
6
Page 12
Instruction Manual PD 300
4 CALIBRATION
4.1 Important Information on Meter Calibration
When you recalibrate your meter, old calibration points are replaced on a “point by point” basis
in pH and completely for % saturation and mg/L (ppm).
For example:
•pH: if you previously calibrated your meter at pH 4.01, 7.00 and 10.01, and you
recalibrate at pH 7.00, the meter retains the old calibration data at pH 4.01 and pH 10.01.
•% saturation and mg/L (ppm): all new calibrations automatically override existing
calibration data.
To view current calibration points:
• pH: Program 1.0 in the SETUP section, page 30.
• % saturation: Program 3.0 in the SETUP section: page 41.
• mg/L (ppm): Program 2.0 in the SETUP section, page 41.
To completely recalibrate your meter, or when you use replacement probes, it is best to clear
old calibration data by resetting the meter.
To reset the meter to its factory defaults:
• pH: Program 4.0 in the SETUP section, page 36.
• % saturation: Program 7.0 in the SETUP section: page 49.
• mg/L (ppm): Program 6.0 in the SETUP section, page 49.
NOTE
Resetting the meter will set meter to factory defaults. pH, % saturation and mg/L (ppm) must
be reset separately.
For directions on how to calibrate your meter:
See section 4.2 for Temperature Calibration
See sections 4.3 for pH calibration.
See section 4.5 for % saturation calibration
See sections 4.6 for mg/L (ppm) calibration
7
Page 13
Instruction Manual PD 300
4.2 Temperature Calibration
The temperature sensor of the the DO probe is factory
calibrated. Calibrate your sensor only if you suspect
temperature errors that may have occurred over a long
period of time or if you have a replacement probe. Make sure
the DO probe is attached prior to performing the calibration.
1. Switch the meter on. Press MODE to select either pH or
mg/L (ppm) measurement mode.
2. Press the CAL/MEAS key to enter pH or mg/L (ppm)
calibration mode. The CAL indicator will appear above the
primary display.
3. While in pH or mg/L (ppm) calibration mode, press the
MODE key to enter temperature calibration mode. The
primary display shows the temperature reading and the
secondary display shows you the factory default
temperature value.
4. Compare the primary display reading to a NISTtraceable thermometer or another thermometer known
to be accurate.
5. Press the ▲ or ▼ keys to adjust the primary display
reading to agree with your temperature standard.
6. Press the ENTER key to confirm temperature calibration
and return to Measurement mode.
NOTE
•To exit from Temperature Calibration mode without
confirming calibration, DO NOT press ENTER in Step 6.
Press CAL/MEAS instead.
•If a separate temperature probe is used for pH
measurement (EC-PHWPTEMP-01W / 35618-05), then
temperature calibration done for pH is separate and
independent from mg/L measurement mode.
•Te mperature calibration is restricted to ± 5°C from the
factory default value displayed during calibration
(shown in the secondary display).
(The values shown above are for illustration purpose. Your meter may display
different values depending on the settings & your environmental conditions)
8
Figure 5: Temperature calibration
Page 14
Instruction Manual PD 300
4.3 pH calibration
Before starting calibration, make sure you are in pH measurement mode.
NOTE
We recommend calibration with at least two calibration standards that bracket (one
above and one below) the expected sample range. One point calibration is possible as
long as the chosen buffer is close to the expected sample value.
When you re-calibrate your meter, old calibration points are replaced on a “point-by-point”
basis. For example:
•if you previously calibrated your meter at pH 4.01, 7.00, and 10.01, and you re-
calibrate at pH 7.00, the meter retains the old calibration data at pH 4.01 and pH
10.01.
Preparing for pH calibration
This meter is capable of up to 5-point pH calibration to ensure accuracy across the entire pH
range of the meter. Select from the following buffer options:
•pH 1.68, 4.01, 7.00, 10.01 and 12.45.
The meter automatically recognises and calibrates to these standard buffer values, which
makes pH calibration faster and easier.
Be sure to remove the protective electrode storage bottle or rubber ca p of the electrode
before calibration or measurement. If the electrode has been stored dry, hydrate the
electrode in electrode storage solution or tap water for 10 minutes before calibrating or taking
readings to saturate the pH electrode surface and minimise drift.
Rinse your electrode in deionised water after use, and store in electrode storage solution. If
storage solution is not available, use fresh pH 4.01 or 7.00 buffer solution for short term
storage.
Do not reuse buffer solutions after calibration. Contaminants in the solution can affect the
calibration and eventually the accuracy of the measurements. See Accessories for information
on pH buffer solutions.
9
Page 15
Instruction Manual PD 300
To Calibrate pH:
1. If necessary, press the MODE key to select
pH mode. The pH indicator appears in the
upper right hand corner of the display.
2. Rinse the probe thoroughly with de-ionised
water or a rinse solution. Do not wipe the
probe; this causes a build-up of electrostatic
charge on the glass surface.
3. Dip the probe into the calibration buffer. The
end of the probe must be completely
immersed into the sample. Stir the probe
gently to create a homogeneous sample.
NOTE
The temperature element is in the DO electrode. For
temperature compensated readings, dip the DO
electrode into the calibration buffer as well.
4. Press CAL/MEAS to enter pH calibration
mode. The CAL indicator will be shown. The
primary display will show the measured
reading while the smaller secondary display
will indicate the pH standard buffer solution.
5. Wait for the measured pH value to stabilise.
See Figure 5.
6. Press ENTER to confirm calibration. The
meter is now calibrated to the current buffer.
The lower display scrolls through the
remaining buffer options.
• For multi-point calibration, go to step 7.
• For one-point calibration, go to step 9.
10
Figure 6: pH Calibration
Page 16
Instruction Manual PD 300
7. Rinse the electrode with deionised water or in
rinse solution, and place it in the next pH buffer.
8. Follow steps 5 to 8 for additional calibration
points. See Figure 7.
9. When calibration is complete, press
CAL/MEAS to return to pH measurement mode.
NOTE
To exit from pH calibration mode without confirming
calibration, DO NOT press ENTER in step 6. Press
CAL/MEAS instead.
If the selected buffer value is not within ±1.0 pH from the
measured pH value: the electrode and buffer icon blink
and the ERR annunciator appears in the lower left
Figure 7: Next point calibration for
pH 4.01
corner of the display. See Figure 8.
To limit the number of pH buffer values available during
calibration, see Setup Section 8.4
P3.2: Selecting
number of pH calibration points on page 34.
Figure 8: Err message and
electrode icon will appear if
incorrect buffer is used
11
Page 17
Instruction Manual PD 300
4.4 Dissolved Oxygen Calibration
Before calibrating your meter, make sure to rinse the probe well with de-ionised water or rinse
solution.
NOTE
Do not let membrane surface of the probe touch any
other surface. The probe guard (the piece with holes
fitted over the end of the probe) protects the membrane;
make sure this is always attached to the probe while it
is in use. See Figure 9.
You can calibrate for dissolved oxygen in either %
saturation or mg/L (ppm). All new calibrations
automatically override existing DO calibration data.
Figure 9: DO Probe guard
Before Calibration
Before starting calibration, make sure you are in the correct measurement mode. When you
switch on the meter, the meter starts up in the units last used (either pH, mg/L (ppm) or % air
saturation). For example, if you shut off the meter in “mg/L” units, the meter will read “mg/L”
units when you switch the meter on.
NOTE
Most users will calibrate to 100% saturation even when working in mg/L. When calibrating the
meter in mg/L mode, you can make fine adjustments, typically to a mid-range value between
10 and 100%. If you are calibrating to a mid-range value, you need to set the 100% saturation
value first.
12
Page 18
Instruction Manual PD 300
4.5 Dissolved Oxygen Calibration in % Saturation
You can calibrate this meter quickly and easily in air. The exact calibration value depends on
barometric pressure. The meter is set to a factory default of 760 mm Hg, which results in a
calibration value of 100% saturation in air.
NOTE
If the barometric pressure setting has been changed from 760 mm Hg, the calibration value in
air will automatically adjust to a value other than 100%. The adjusted value will be correct for
the new barometric pressure setting.
See Section 8.6.1 on page 37 to change the pressure setting. See Addendum 2 on page 67
for a table of adjusted % saturation values.
To calibrate 100% Saturation:
1. Rinse the probe well with deionised rinse water. Do not touch
the membrane.
2. Press the MODE key to select the % saturation mode.
3. Press the CAL key. The CAL indicator will appear above the
primary display. The primary display shows the current value
of the measurement and the secondary display will show
100.0. See figure on right and NOTE above.
4. Hold the probe in the air. Wait for the reading to stabilise. If
the READY indicator feature is enabled, it will appear when
the reading is stable. (See page 45 for Ready indicator
setup.)
5. Press the ENTER key. The meter automatically calibrates to
100.0% air saturation and returns to Measurement mode.
NOTE
Whenever an error occurs during calibration, the ERR indicator
appears in the lower left hand corner of the display.
You can offset your % DO calibration. See page 40 for directions.
13
Figure 10: To calibrate for
100% saturation
Page 19
Instruction Manual PD 300
To calibrate 0% Saturation:
1. Rinse the probe well with deionised rinse water. Do not
touch the membrane.
2. Press the MODE key to select the % saturation mode.
3. Place the probe in zero oxygen solution. Stir the probe
gently to homogenise the sample. Make sure that the
READY
MEAS
128 . 5
L
A
C
S
A
E
M
23.4
%
°C
ATC
sample is continuously flowing past the membrane sensor.
4. Press the CAL key. The CAL indicator will appear above
CAL
%
the primary display. The primary display shows the current
value of the measurement and the secondary display will
show 0.0.
5. Wait for the reading to stabilise. If the READY indicator
R
E
T
N
E
feature is enabled, it will appear when the reading is stable.
(See page 45 for Ready indicator setup.)
READY
MEAS
%
6. Press the ENTER key. The meter automatically calibrates
to 0.0% air saturation and returns to Measurement mode.
23.4
°C
ATC
See figure on right.
NOTE
Figure 11: To calibrate for
0% saturation
The reading in the primary display in step 3 must read at 10% or
under for the calibration to work correctly. Whenever an error occurs during calibration, the
ERR indicator appears in the lower left hand corner of the display.
You can offset your % DO Calibration; see page 40 for directions.
14
Page 20
Instruction Manual PD 300
4.6 Dissolved Oxygen Calibration in mg/L (ppm) mode
Calibrating the meter to 100% in saturation mode will also calibrate the mg/L mode at the
value in mg/L corresponding to 100% saturation. This should produce acceptable results in
most applications.
This meter also lets you make a calibration adjustment in mg/L (ppm) mode without affecting
your % saturation calibration. To select between mg/L and ppm units, see page 46.
NOTES ON MG/L (PPM) MODE CALIBRATION
•During mg/L (ppm) measurement and calibration, the meter adjusts to the
barometric pressure value that is programmed into the meter. It also adjusts to
the salinity value that is programmed into the meter.
•You can change the barometric pressure value and salinity value in the mg/L
•Whenever an error occurs during calibration, the ERR indicator appears in the
lower left hand corner of the display.
15
Page 21
Instruction Manual PD 300
To calibrate in mg/L (ppm) mode:
1. Calibrate 100% saturation as per Section 4.5.
2. Rinses the probe well with deionised rinse water. Wipe
the outside of the probe carefully (do not touch the
membrane).
3. Dip the probe into a sample of known oxygen
concentration. Wait for the reading to stabilise. If the
READY indicator feature is enabled, it will appear when
the reading is stable.
4. Press the MODE key to select the mg/L (ppm) mode.
5. Press the CAL key. The CAL indicator will appear above
the primary display. The primary display shows the
current value of the measurement and the secondary
display shows the temperature.
6. Press the ▲ and ▼ keys to adjust the reading.
7. Press the ENTER key. The meter automatically
calibrates to the value you entered and returns to
Measurement mode.
Figure 12: To calibrate in
mg/L (ppm) mode
(The values shown above
are for illustration purpose.
Your meter may display
different values depending
on the settings & your
environmental conditions)
16
Page 22
Instruction Manual PD 300
5 MEASUREMENT
5.1 Taking pH Measurements
5.1.1
Automatic Temperature Compensation only occurs when a compatible temperature sensing
element is plugged into the meter.
If there is no temperature sensor plugged into the meter, the default manual temperature
setting is 25 °C. You can manually set the temperature to match your working conditions
using a separate thermometer.
For automatic temperature compensation (ATC) simply
plug the temperature probe into the meter (see page 6
for directions). The ATC indicator will light up on the
LCD. See figure on right.
NOTE
The temperature sensing element must be in the
sample that you are calibrating or measuring.
Automatic Temperature Compensation
Figure 13: Automatic Temperature
Compensation
5.1.2
17
Page 23
Instruction Manual PD 300
Manual Temperature Compensation (pH only)
IMPORTANT: For manual compensation, you must
disconnect the temperature probe (see page 10).
1. Switch the meter on. Press the MODE key to
select pH mode.
2. Press the CAL/MEAS key to enter pH
calibration mode. The CAL indicator will
appear above the primary display.
3. While in pH calibration mode, press the
MODE key to enter temperature calibration
mode. The primary display shows the
current temperature setting and the
secondary display shows the default value
25 °C.
4. Check the temperature of your sample using
an accurate thermometer.
5. Press the ▲ or ▼ keys to set the
temperature to the measured value from step
4.
6. Press ENTER to confirm the selected
temperature and to return to the pH
measurement mode.
See Figure 14.
The meter will now compensate pH readings for the
manually set temperature.
Figure 14: Manual temperature
compensation
NOTE
To exit this program without confirming the manual temperature compensation value,
DO NOT press ENTER in step 6. Press CAL/MEAS instead.
5.1.3
18
Page 24
Instruction Manual PD 300
Taking Measurements (pH)
Be sure to remove the electrode storage bottle or
protective rubber cap on the electrode before taking
measurements. To ensure Automatic Temperature
Compensation (ATC) for pH measurements, the DO
or ATC probe must be plugged in and immersed into
the sample.
To take readings:
1. Rinse the pH electrode with deionised or distilled water before use to remove any
impurities adhering to the probe body. If the pH electrode has dehydrated, soak it
for 30 minutes in electrode storage solution (sold separately).
2. Press ON to switch on meter.
3. Press the MODE key to select pH measurement mode. The MEAS annunciator
appears on the top center of the LCD. The ATC indicator appears in the lower right-
hand corner to indicator Automatic Temperature Compensation.
NOTE
For pH manual temperature compensation, you must disconnect the DO electrode from
the 6-pin connector. The ATC indicator will disappear from the display. You also need to
set a manual temperature compensation value. See Section 5.1.2 on page 18 to set
Manual Temperature Compensation.
Figure 15: Measurement mode
4. Dip the probe into the sample. Since the DO electrode contains the temperature
sensor, make sure it is also immersed in your solution.
When dipping the probe into the sample, the sensor or the glass bulb of the electrode
must be completely immersed into the sample. Stir the probe gently in the sample to
create a homogeneous sample.
5. Allow time for the reading to stabilise. Note the reading on the display. If the Ready
indicator is selected on, it will appear when the reading is stable. See below for
more information.
19
Page 25
Instruction Manual PD 300
Taking measurements with READY indicator selected on
If the READY indicator has been activated, the READY annunciator lights when the reading is
stable*. Switch the READY indicator on or off in SETUP Section 8.4 Program P3.1.
* The READY indicator appears and the reading holds until the measured value
exceeds the tolerance. Then, the READY annunciator turns off.
Taking measurements with the auto endpoint feature selected on
When a reading is stable for more than 5 seconds, the auto endpoint feature will automatically
“HOLD” the reading. The “HOLD” indicator appears on the left side of the display. Press the
HOLD key to release the reading. Switch the Auto endpoint feature on or off in SETUP
Section 8.4 Program P 3.1.
20
Page 26
Instruction Manual PD 300
5.2 Taking DO read in gs
During measurement, the probe can be:
z Fully immersed in the solution
z Partially immersed in the solution
Do not allow the probe’s membrane surface to touch
anything! The probe guard (the piece with holes fitted
over the end of the probe) protects the membrane; you
should leave this piece attached to the probe at all
times.
IMPORTANT: Since the DO probe consumes oxygen
from the sample, the sample must constantly flow past
the membrane to achieve more accurate readings. You
can use a stirrer to accomplish this.
To take measurements:
1. Rinse the probe well with deionised rinse water or
rinse solution.
2. Select the appropriate measurement mode. Press the
MODE key to toggle between modes:
z % Saturation
z mg/L (ppm) **
z pH
3. Dip the probe into the sample. Stir the probe gently to
homogenise the sample. Make sure that the sample is
continuously flowing past the membrane sensor.
Figure 16: Change
measurement mode from %
saturation to mg/L
4. Note the reading on the display. If the READY
indicator is switched on, it will appear when the
reading is stable.
Figure 17: Taking
measurement
5. Press the “LIGHT” key to activate the backlit LCD. This
helps you read measurements in dimly lit areas.
** To select between mg/L and ppm units, see Section 8.10.2 on page 46.
The ATC indicator should appear in the lower right hand corner of the display. If it does not,
this indicates an error.
21
Page 27
Instruction Manual PD 300
Taking measurements with READY ON indicator selected on
If the READY indicator has been activated, the READY annunciator lights when the reading is
stable.
To switch the READY indicator on or off in the SETUP program – see Section 8.10.1 on page
45 for directions.
Taking measurements with READY HOLD indicator selected on
When a reading is stable for more than 5 seconds, the
auto endpoint feature will automatically “HOLD” the
reading. The “HOLD” indicator appears on the left side of
the display. Press the HOLD key to release the reading.
To switch the Auto endpoint feature on or off in the
SETUP program – see Section 8.10.1 on page 45 for
directions.
(The values shown here are for illustration purpose. Y our meter may display
different values depending on the settings & your environmental conditions)
Figure 18 – Auto endpoint
feature
22
Page 28
Instruction Manual PD 300
5.3 Taking Pressure/Salinity Compensated DO measurements
If necessary, you can adjust the pressure and salinity values of your measurements in the
SETUP mode. The DO meter will automatically compensate for salinity and pressure based on
the values entered in the SETUP functions. The meter is factory set at 760 mm Hg (101.3
Pascals) pressure adjustment and a factor of 0.0 ppt salinity adjustment. See Addendum 2 on
page 67 for a “Pressure vs Altitude” table.
Pressure Adjustment
NOTE
This mode appears in both the % saturation and mg/L (ppm) measurement modes.
1. While in the measurement mode, press “Mode” key if necessary to enter DO
measurement mode.
2. Press the SETUP key to enter SETUP mode.
3. The upper display shows “dPr”.
4. Press the ENTER key until the upper display shows a number and the lower display
shows “Hg” or “PA”.
5. Use the ▲ or ▼keys to enter the barometric pressure. The upper display will show the
value you have entered.
6. Press ENTER to confirm pressure value.
7. Press CAL/MEAS to return to Measurement mode, or continue with step 4 on page 39 to
make a salinity adjustment [available in mg/L (ppm) mode only].
8. See section 8.6.1 - “P1.1 Pressure adjustment mode” on page 37.
Salinity Adjustment
NOTE
This mode appears in ppm (mg/L) measurement mode only.
1. Press the MODE key to select mg/L (ppm) mode.
2. Press the SETUP key to enter SETUP mode.
3. The upper display shows “dPr”.
4. Press the ENTER key until the upper display shows a number and the lower display
shows “SAL”.
5. Use the ▲ or ▼ keys to enter the salinity of your solution in parts per thousand. Note
1000 ppm = 1 ppt. The upper display will show the value you have entered.
6. Press ENTER key to confirm value.
7. Press CAL/MEAS to return to Measurement mode.
8. See section 8.6.2 - P1.2: Salinity Adjustment Mode on page 39.
23
Page 29
Instruction Manual PD 300
6 HOLD FUNCTION
This feature lets you freeze the pH, dissolved oxygen,
and temperature readings for a delayed observation.
HOLD can be used any time when in MEAS mode.
To hold a measurement, press the HOLD key while in
measurement mode. “HOLD” will appear on the
display.
To release the held value, press HOLD again.
Continue to take measurements.
NOTE
The meter shuts off automatically 20 minutes after the
last key press. If the meter is shut off either
automatically or manually, the HOLD value will be lost.
NOTE
This meter has an auto endpoint feature. When this
feature is switched on, the display will automatically
“HOLD” a reading that has been stable for more than
5 seconds. The “HOLD” indicator appears. Press the
HOLD key to release the reading. To switch on or off
the auto endpoint feature, see page 45.
Figure 19: HOLD feature
(The values shown are for illustration purpose. Your meter may display differen t values depending on the settings & your
environmental conditions)
24
Page 30
Instruction Manual PD 300
7 BACKLIT LCD
7.1 Using the backlit LCD
This feature lets you view readings in low light situations.
1. To activate the backlit LCD feature, press the
key to activate.
To conserve batteries, the backlight automatically shuts off after 20 seconds. To keep the LCD
backlit, keep the backlight key pressed.
7.2 Adjusting the LCD brightness
This mode lets you adjust the brightness of the backlit
LCD. Selecting a dimmer backlighting level helps
conserve batteries. This adjustment is only available in
the % saturation or mg/L (ppm) SETUP modes.
From the measurement mode:
1. Press the MODE key to switch to % saturation or
mg/L (ppm) mode.
2. Press SETUP key to enter setup mode.
3. Press the ▲ or ▼ keys to scroll through subgroups
until you view parameters “LCd” in the upper display.
4. Press the ENTER key. A number (0-8) appears in
the upper display and “LCd” appears in the lower
display.
5. Press the ▲ or ▼ keys to select from level 0
(dimmest light) to level 8 (brightest light).
6. Press the ENTER key to confirm selection and to
return to the subgroup menu. Press the CAL/MEAS
key to return to measurement mode.
25
Figure 20: Adjusting LCD
Brightness
Page 31
Instruction Manual PD 300
8 ADVANCED SETUP FUNCTIONS
The advanced SETUP mode lets you customise your meter’s preferences and defaults. Your
waterproof meter features different sub groups that organise all set-up parameters.
Each measurement parameter (pH, % saturation, & mg/L) has its own sets of customisable
setup functions. Some of which are unique and independent while others are common for the
whole meter.
The full selection of available sub groups is shown in the following three pages.
8.1 Advanced SETUP Mode Detailed Overview
1. In either pH or DO measurement mode, press the SETUP key to enter Set up mode.
2. Press the ▲ or ▼ keys to scroll through sub groups.
3. Press ENTER key to enter a particular parameter.
See Addendum 3 on page 68 for a table of meter factory default settings.
26
Page 32
Instruction Manual PD 300
SETUP in pH Measurement Mode
P1.0: View previous calibration data
P1.1 First calibration point (pH 1.68)
P1.2 Second calibration point (pH 4.01)
P1.3 Third calibration point (pH 7.00)
P1.4 Fourth calibration point (pH 10.01)
P1.5 Fifth calibration point (pH 12.45)
Instructions on page 30.
P2.0: View electrode data
P2.1 pH electrode offset
P2.2 pH electrode slope
Instructions on page 31.
P3.0: Unit configuration
P3.1 READY indicator and auto endpoint function –
P3.2 Number of pH calibration points: 2, 3, 4 or 5
P3.3 Select °C or °F
Instructions begin on page 32.
select on or off
P4.0: Reset to factory defaults
P4.0 Reset meter to factory defaults
Instructions on page 36.
27
Page 33
Instruction Manual PD 300
% Saturation Measurement Mode
dPr: Dissolved Oxygen parameters
• Select Hg or Pa barometeric pressure units
• Select barometric pressure
Instructions on page 37.
OFS: % saturation offset adjustment
• Set % saturation offset adjustment
Instructions on page 40.
CAL: Viewing previous calibration data
• View previous calibration data
Instructions on page 41.
ELE: Viewing probe data
• View probe slope
• View % saturation offset value
• View mV value equivalent to 100% saturation
• View mV value equivalent to 0% saturation
Instructions begin on page 42.
COF: Unit configuration
•Ready indicator ON or OFF / auto endpoint ON
or OFF
•Select °C or °F
Instructions begin on page 44.
LCd: Backlit display
• Adjust brightness of backlit LCD
Instructions on page 48.
rSt: Reset to factory default
• Reset to factory default settings
Instructions on page 49.
28
Page 34
Instruction Manual PD 300
mg/L (ppm) Measurement Mode
dPr: Dissolved Oxygen parameters
• Select Hg or Pa barometric pressure units
• Select barometric pressure
• Select salinity adjustment factor
Instructions on page 37.
CAL: Viewing previous calibration data
• View previous calibration data
Instructions on page 41.
ELE: Viewing probe data
• View probe slope
• View mV value equivalent to 100% saturation
• View mV value equivalent to 0% saturation
Instructions begin on page 42.
COF: Unit configuration
•Ready indicator ON or OFF / auto endpoint ON
or OFF
• Select mg/L or ppm units
• Select °C or °F
Instructions begin on page 44.
LCd: Backlit display
• Adjust brightness of backlit LCD
Instructions on page 48.
rSt: Reset to factory default
• Reset to factory default settings
Instructions on page 49.
29
Page 35
Instruction Manual PD 300
8.2 P1.0: Viewing previous pH calibration data
This mode lets you recall previous pH calibration data,
which helps you know when to re-calibrate your meter.
This is a “view only” mode.
From measurement mode:
1. Press the MODE key to select pH measurement
mode if necessary.
2. Press the SETUP key to enter Set up mode.
3. Press the ▲ or ▼ keys to scroll through
subgroups until you view parameter P1.0.
4. Press the ENTER key repeatedly to view previous
calibration data. See figure on right.
• P1.1 = pH 1.68
• P1.2 = pH 4.01
• P1.3 = pH 7.00
• P1.4 = pH 10.01
• P1.5 = pH 12.45
5. When you have scrolled through all calibration
data, you will automatically return to the subgroup
menu. Press CAL/MEAS key to return to
measurement mode.
NOTE
If there is no previous calibration data at a particular
point, the primary display will show “ ----“.
30
Figure 21 – P1.0: View calibration
data for pH
Page 36
Instruction Manual PD 300
8.3 P2.0: Viewing pH electrode data
Program 2 has two “view only” options that let you
check the electrode parameters for diagnostic
purposes. It lets you view:
• P2.1 = Electrode offset
• P2.2 = Electrode slope
From pH measurement mode
1. Press the MODE key to select pH measurement
mode.
2. Press the SETUP key to enter Set Up mode.
3. Press the ▲ or ▼ keys to scroll through
subgroups until you view parameter P2.0.
4. Press the ENTER key to select parameter 2.1.
5. The display shows the electrode offset value. It is
the mV offset at pH 7.00. If you have not
calibrated at any buffer, the primary display shows
0.00 mV.
6. Press the ENTER key to proceed to P2.2.
7. The display shows electrode slope in percentage.
Slope displayed is the average slope based on the
pH calibrations. Default setting is 100.0.
8. At any point, you can press the CAL/MEAS key to
return to measurement mode.
See figure on right.
31
Figure 22: Viewing electrode's
offset and slope status from pH
measurement mode
(The values shown above are for
illustration purpose. Your meter may
display different values depending
on the settings & your
environmental conditions)
Page 37
Instruction Manual PD 300
8.4 P3.0: pH Measurement co nfiguration
This subgroup program allows to customise the meter
to your specific needs. You can program the meter to:
1. Select READY indicator ON or OFF
2. Select the number of pH calibration points
3. Select between °C and °F units for temperature
readings
Figure on the right shows the setup sequence for this
program subgroup.
Figure 23 – P 3.0: Unit
configuration program
32
Page 38
Instruction Manual PD 300
P3.1: READY Indicator and auto endpoint function
This program lets you select:
• “
READY indicator on” to indicate when the
reading is stable.
• “
READY indicator off” for faster meter response.
•
Auto endpoint function on. Select auto endpoint
on to “hold” the reading when it is stable for more
than 5 seconds. The display automatically freezes,
and the HOLD indicator appears on the left side of
the display. Press the HOLD key to release the
display and access other functions.
From measurement mode
1. Press the MODE key to select pH measurement
mode.
2. Press SETUP key to enter Set Up mode.
3. Press the ▲ or ▼ keys to scroll through subgroups
until you view parameter P3.0.
4. Press the ENTER key to select parameter 3.1.
5. Press the ▲ or ▼ keys to select the configuration
you require.
• OFF switches the READY indicator off.
• ON switches the READY indicator on.
• ON and HOLD together switches the auto
endpoint feature on.
6. Press the ENTER key to confirm selection and to
proceed to step 4 of P3.2. Press the CAL/MEAS key
to return to measurement mode.
Figure 24 – P 3.1: Selecting
READY function)
NOTE
Meter default is set for READY indicator on and auto endpoint function off.
You can also change the Ready indicator and auto endpoint function in the DO mode. Any
changes you make to the Ready indicator/auto endpoint function in pH mode will also change
in DO mode.
33
Page 39
Instruction Manual PD 300
P3.2: Selecting number of pH calibration points
Program P3.2 lets you select the number of calibration
points that appear in pH calibration mode: 2, 3, 4, or 5.
The meter will automatically exit calibration mode after
you have calibrated to your selected number of points.
From measurement mode.
1. Press the MODE key to select pH measurement
mode.
2. Press SETUP key to enter Set Up mode.
3. Press the ▲ or ▼ keys to scroll through subgroups
until you view parameter P3.0.
4. Press the ENTER key twice to select parameter 3.2.
5. Press the ▲ or ▼ keys to select 2, 3, 4, or 5 point
pH calibration.
6. Press the ENTER key to confirm selection and to
return to the subgroup menu. Press CAL/MEAS
key to return to measurement mode.
See Figure on right.
Figure 25 – P3.2: Select number
of pH calibration points
34
Page 40
Instruction Manual PD 300
P3.3 Selecting °C or °F
This meter lets you select between °C and °F units for
temperature readings.
From measurement mode
1. Press the MODE key to select pH measurement
mode.
2. Press SETUP key to enter Set Up mode.
3. Press the ▲ or ▼ keys to scroll through
subgroups until you view parameter P3.0.
4. Press the ENTER key three times to select
parameter 3.3.
5. Press the ▲ or ▼ keys to toggle between °C and
°F.
6. Press the ENTER key to confirm selection and to
return to the subgroup menu. Press the
CAL/MEAS key to return to measurement mode.
Figure 26 – P3.3: Select
temperature units
See Figure on right.
NOTE
You can switch between °C and °F in Program P5.2 of % saturation mode (or P4.3 in mg/L
mode). If you switch between °C and °F in pH mode, the meter will also switch in % saturation
or mg/L mode.
35
Page 41
Instruction Manual PD 300
8.5 P4.0: Resetting to factory default settings (pH)
This program lets you reset all pH parameters to factory default settings. This clears all
calibration data and any other pH setup functions you might have changed.
The following settings will remain as you have set them:
• Temperature unit of measure (°C or °F)
• The temperature offset calibration value
• All DO calibration data and parameters
There are also some other parameters that retain settings when reset is done. See Addendum
3 on page 68 for a table of factory default settings.
From measurement mode
1. Press the MODE key to select pH measurement mode.
2. Press SETUP key to enter Set Up mode.
3. Press the ▲ or ▼ keys to scroll through subgroups until
you view parameter P4.0 in the lower display.
4. Press the ENTER key to enter parameter P4.0.
5. Press the ▲ or ▼ keys to toggle between NO and YES.
See figure on right.
• NO retains current settings
• YES resets to factory default settings.
6. Press the ENTER key to confirm selection and to return
to the measurement mode. Otherwise press CAL/MEAS
key to return to measurement mode without resetting to
factory default.
Figure 27 – P4.0: Reset to factory
default values
NOTE
See page 68 for a table of factory default settings.
36
Page 42
Instruction Manual PD 300
8.6 P1.0 (dPr): Dissolved Oxygen Parameters
This sub group is available in % saturation and mg/L (ppm) mode and lets you adjust the
barometric pressure and salinity (% saturation mode only).
8.6.1 P1.1 Pressure adjustment mode
Barometric pressure is vital to correct dissolved oxygen measurements. You need to enter the
correct barometric pressure of the area you are measuring. This mode lets you perform two
functions:
• Select either mm Hg or Pascal barometric pressure units.
• Adjust the barometric pressure. See Addendum 2 on page 67 for a
“Pressure vs Altitude” table.
From measurement mode:
1. Press the MODE key to select DO measurement mode.
2. Press the SETUP key to enter SETUP mode.
3. Press the ▲ or ▼ keys to scroll through subgroups until you view parameter “dPr” in the
upper display.
4. Press the ENTER key twice. The upper display shows either Hg or PA and the lower
display shows bAr. See figure on next page.
5. Press the ▲ or ▼ keys to toggle between mm Hg and Pascal units.
6. Press the ENTER key to confirm selection and move to the next screen. The upper
display shows the barometric pressure and the lower display shows the units selected in
Step 3.
7. Press the ▲ or ▼ keys to adjust the barometric pressure. The pressure adjustment
range is 500 to 1499 mm Hg (66.6 to 199.9 kPa).
8. Press the ENTER key to confirm selection and move to Step 4 on page 39 [salinity adjustment appears in mg/L (ppm) measurement mode only]. If instead you want to
return to measurement mode, press CAL/MEAS twice.
37
Page 43
Instruction Manual PD 300
8.6.2
Figure 28: Change pressure units (mm Hg or kPa) & adjusting its values
38
Page 44
Instruction Manual PD 300
P1.2: Salinity Adjustment Mode
NOTE
This mode appears in mg/L (ppm) measurement mode
only.
Salinity correction mode lets you correct for the
variations in oxygen solubility due to salt concentration
in the sample.
From measurement mode:
1. Press the MODE key to select mg/L (ppm) mode.
2. Press the SETUP key to enter SETUP mode.
3. Press ▲ or ▼ keys to scroll through subgroups
until you view parameter “dPr” in the upper display.
4. Press the ENTER key five times. The upper
display shows the salinity value and the lower
shows SAL.
5. Press ▲ or ▼keys to enter the correct salinity
adjustment factor. The salinity adjustment factor
range is 0.0 to 50.0 ppt.
6. Press the ENTER key to confirm selection and to
move back to subgroup “dPr”. If you want to return
to measurement mode, press CAL/MEAS.
39
Figure 29: Input Salinity values for
automatic correction
(The values shown above are for
illustration purpose only)
Page 45
Instruction Manual PD 300
8.7 P2.0 (OFS): Offset for % Saturation
Measurement
NOTE
This sub group appears in % saturation measurement
mode only.
From measurement mode:
1. Press the MODE key to select % saturation mode.
2. Press the SETUP key to enter SETUP mode.
3. Press ▲ or ▼ keys to scroll through subgroups
until you view parameter “OFS” in the upper
display.
4. Press the ENTER key. The upper display shows
the current measurement in % saturation and the
lower shows “OFS”.
5. Press ▲ or ▼ keys to offset the % saturation
measurement.
6. Press the ENTER key to confirm selection and to
move back to subgroup “OFS”. If you want to
return to measurement mode, press CAL/MEAS.
(The values shown above are for illustration purpose. Your meter may
display different values depending on the settings & your environmental
conditions)
40
Figure 30: Offset for % saturation
Page 46
Instruction Manual PD 300
8.8 P3.0 [P2.0] (CAL): Previous Calibration Information
(*Program numbers in “[ ]” are for setup in mg/L (ppm) mode.
This sub group shows you the previous calibration data. This is a “view only” parameter.
In % saturation mode: Calibration information is shown in % saturation units.
In mg/L (ppm) mode: Calibration information is shown in mg/L (ppm) units.
From measurement mode:
1. Press the MODE key to select the calibration data
you want to view: % or mg/L (ppm).
2. Press the SETUP key to enter SETUP mode.
3. Press ▲ or ▼ keys to scroll through subgroups
until you view parameter “CAL” in the upper display.
4. Press the ENTER key. The upper display shows the
calibration data.
5. Press the ENTER key to move back to subgroup
“CAL”. If you want to return to measurement mode,
press CAL/MEAS.
NOTE: If you did not calibrate this meter in a particular
mode, the screen will show “ ---“. See Figure 30.
(The values shown above are for illustration purpose. Your meter may
display different values depending on the settings & your environmental
conditions)
41
Figure 31: View calibration data
Figure 32: eter is not calibrated
Page 47
Instruction Manual PD 300
8.9 P4.0 [P3.0] (ELE): Electrode Properties
These “view only” parameters show you the electrode
properties for diagnostic purposes:
A. P4.1 [P3.1] Probe Slope: Provides an indication of the
probe’s efficiency. The value displayed is the ratio of the
theoretical value to the actual value produced by the
probe. The higher the number, the lesser output from the
probe. The ratio displays from 0.5 to 1.999% (1.0 = 100%).
P4.2 % Saturation Offset: (available in % saturation
B.
mode only): Shows the value of the % saturation offset
entered in parameter “OFS” (see page 40 for instructions).
P4.3 [P3.2] 100% Saturation mV value: Lets you view
C.
the sensor’s mV output corresponding to 100% saturation.
P4.4 [P3.3] 0% Saturation mV Value: Lets you view the
D.
sensor’s millivolt output corresponding to 0% saturation.
From measurement mode:
1. Press the MODE key to select the measurement mode for
the electrode properties you want to view: % or mg/L
(ppm).
2. Press the SETUP key to enter SETUP mode.
3. Press ▲ or ▼ keys to scroll through subgroups until you
view parameter “ELE” in the upper display.
4. Press the ENTER key. The upper display shows the probe
slope.
5. Press the ENTER key. The upper display shows the %
saturation offset. See directions for setting this offset on
page 40. NOTE: This appears only in % saturation
measurement mode. If you are in mg/L (ppm) mode, the
meter skips to step 6 on page 43.
42
Figure 33: iew electrode
properties from %
saturation
Page 48
Instruction Manual PD 300
6. Press the ENTER key. The upper display shows the
100% saturation mV value.
7. Press the ENTER key. The upper display shows the 0%
saturation mV value.
8. Press the ENTER key to move back to subgroup “ELE”.
If you want to return to measurement mode, press
CAL/MEAS.
(The values shown above are for illustration purpose. Your meter may display
different values depending on the settings & your environmental conditions)
Figure 34: View electrode
properties from mg/L
(ppm) mode
43
Page 49
Instruction Manual PD 300
8.10 P5.0 [P4.0] COF: Unit Configuration
Unit configuration mode lets you select the following parameters:
A. P5.1 [P4.1] READY indicator and auto endpoint function.
B. [P4.2] mg/L or ppm units (available in mg/L or ppm mode only)
C. P5.2 [P4.3] Temperature in °C or °F.
8.10.1
Figure 35: Unit configuration
from % saturation
measurement mode
Figure 36: Unit configuration
from mg/L measurement mode
44
Page 50
Instruction Manual PD 300
P5.1 [P4.1] READY Indicator and Auto Endpoint Function
The first program lets you select:
•
“READY indicator ON” to indicate when the
reading is stable.
• “READY indicator OFF” for faster meter response.
•
Auto Endpoint Function ON. Select auto endpoint
ON to “HOLD” the reading when it is stable for more
than 5 seconds. The display automatically freezes,
and the HOLD indicator appears on the left side of
the display. Press the HOLD key to release the
display and access other functions.
From measurement mode
1. Press SETUP key to enter SETUP mode.
2. Press ▲ or ▼ keys to scroll through subgroups until
you view parameter COF in the upper display.
3. Press the ENTER key to select parameter “rdY”
(Ready).
4. Press ▲ or ▼ keys to select the configuration you
require.
• ON switches the READY indicator on.
• OFF switches the READY indicator off.
• ON and HOLD together switches the auto
endpoint feature on.
5. Press the ENTER key to confirm selection and to
proceed to:
• In % mode: step 4 on page 47.
• In mg/L (ppm) mode: step 5 on page 46.
Or press the CAL/MEAS key twice to return to
Figure 37: To configure for
READY and Auto Endpoint
functions
measurement mode.
Note: Meter default is set for Ready indicator ON, and auto endpoint function off. Any
changes made to the Ready indicator/auto endpoint function will change in the pH
mode as well.
45
Page 51
Instruction Manual PD 300
8.10.2 [P4.2] Selecting mg/L or ppm units
NOTE
This mode appears in mg/L (ppm) measurement mode only.
This mode lets you select between mg/L or ppm dissolved oxygen units.
From measurement mode
1. Press the MODE key to select mg/L (ppm) mode.
2. Press SETUP key to enter SETUP mode.
3. Press ▲ or ▼ keys to scroll through subgroups
until you view parameter “COF” in the upper
display.
4. Press the ENTER key until “dO” appears in the
upper display.
5. Press ▲ or ▼ keys to toggle between mg/L or
ppm units.
6. Press the ENTER key to confirm selection. Press
the CAL/MEAS key twice to return to
measurement mode.
Figure 38: Change from mg/L to
ppm unit
46
Page 52
Instruction Manual PD 300
8.10.3 P5.2 [P4.3] Selecting °C or °F Temperatu re Unit
This meter lets you select between °C and °F units for temperature readings.
From measurement mode
1. Press SETUP key to enter SETUP mode.
2. Press ▲ or ▼ keys to scroll through subgroups
until you view parameter “COF” in the upper
display.
3. Press the ENTER key until “C” or “F” appears in
the upper display.
4. Press ▲ or ▼ keys to toggle between °C and °F.
5. Press the ENTER key to confirm selection and to
return to the subgroup menu. Press the
CAL/MEAS key to return to measurement mode.
Note: Any changes made to the temperature unit
selection will change in the pH mode as well.
47
Figure 39: To change from units of
measurement for temperature
Page 53
Instruction Manual PD 300
8.11 P6.0 [P5.0] LCd: Adjusting LCD Brightness
NOTE
This function is only available through DO measurement mode (% saturation or mg/L)
This mode lets you adjust the brightness of the backlit LCD. Selecting a dimmer back-lighting
level helps conserve batteries.
From measurement mode
1. Press SETUP key to enter SETUP mode.
2. Press ▲ or ▼ keys to scroll through subgroups
until you view parameter “LCd” in the upper
display.
3. Press the ENTER key. A number (0 – 8) appears
in the upper display and “LCd” appears in the
lower display.
4. Press ▲ or ▼ keys to select from level 0
(dimmest light) to level 8 (brightest light).
5. Press the ENTER key to confirm selection and to
return to the subgroup menu. Press the
CAL/MEAS key to return to measurement mode.
Figure 40: Adjusting LCD
brightness
48
Page 54
Instruction Manual PD 300
8.12 P7.0 [6.0] rSt: Resetting to Factory Default Setting (DO)
This program lets you reset all DO (% saturation and mg/L) parameters to factory default
settings. This clears all calibration data and any other DO setup functions you might have
changed.
The following settings will remain as you have set them:
• Temperature unit of measure (°C or °F)
• The temperature offset calibration value
• All DO calibration data and parameters
There are also some other parameters that retain settings when reset is done. See Addendum
3 on page 68 for a table of factory default settings.
From measurement mode
1. Press SETUP key to enter SETUP mode.
2. Press ▲ or ▼ keys to scroll through subgroups
until you view parameter “rSt” in the upper display.
3. Press the ENTER key.
4. Press ▲ or ▼ keys to toggle between NO and
YES.
• NO retains current memory.
• YES clears all memory
5. Press the ENTER key to confirm selection and to
return to measurement mode.
Figure 41: Reset to factory
defaults
49
Page 55
Instruction Manual PD 300
9 PROBE CARE AND MAINTENANCE
9.1 pH Electrode car e
Your pH electrode should be kept clean and free from dirt and contamination. Probes that are
not clean may provide slow or unusual readings. With proper care and normal usage, the pH
electrode will need to be replaced every 6 to 18 months on average. NOTE: For specialty
electrode care, consult the instruction manual included with your electrode.
pH electrode storage
For best results, always keep the pH bulb wet. Use the protective electrode storage bottle or
rubber cap filled with electrode storage solution to store your electrode. Also, you can store in
a pH 4 buffer with 1/100 part of saturated KCl. Other pH buffers are OK for short term storage,
but NEVER use distilled water for storage.
After measuring
1. Rinse the pH electrode in de-ionised water.
2. Store the electrode as recommended above in “pH electrode storage,” or as
recommended by the electrode manufacturer.
3. Prior to next use, rinse the liquid junction with de-ionised water and shake dry – never
wipe the electrode.
NOTE: If this does not restore electrode to normal response, see “Reactivating the pH
electrode” section below.
pH electrode cleaning
•Salt deposits: Dissolve the deposits by immersing the electrode in warm tap water for
ten to fifteen minutes. Thoroughly rinse with distilled water.
•Oil/Grease film: wash electrode pH bulb gently in detergent and water. Rinse
electrode tip with distilled water or use a general purpose electrode cleaner (see
Accessories).
•Clogged reference junction: Heat a diluted KCl solution to 60 to 80 °C. Place the
sensing part of the electrode into the heated solution for about 10 minutes. Allow the
electrode to cool in some unheated KCl solution.
•Protein deposits: Prepare a 1% pepsin solution in 0.1 M of HCl. Set the electrode in
the solution for five to ten minutes. Rinse the electrode with distilled water.
50
Page 56
Instruction Manual PD 300
Reactivating the pH electrode
If stored and cleaned properly, your pH electrode should be ready for immediate use. However,
a dehydrated bulb may cause sluggish response. To rehydrate the bulb, immerse the
electrode in a pH 4 buffer solution for 10 to 30 minutes. If this fails, the electrode requires
activation. Never touch or rub glass bulb. Contact builds up an electro-static charge.
pH electrode activation (for glass body electrodes only)
WARNING: Only qualified persons proficient with the safe handling of dangerous chemicals
should perform the procedure below. Provide proper containers, fume hoods, ventilation, and
waste disposal. Safety goggles and protective clothing must be worn while performing this
procedure. If possible, replace with another electrode instead of performing this re-activation
procedure.
1. Dip or stir the pH electrode in alcohol for 5 minutes.
2. Leave the electrode in tap water for 15 minutes.
3. Dip and stir the electrode in concentrate acid (such as HCL or H
) for five minutes.
2SO4
4. Repeat Step 2.
5. Dip and stir in strong base (NaOH) for five minutes.
6. Leave for 15 minutes in tap water.
7. Now test with standard calibration buffer solutions to see if the electrode yields
acceptable results. You may repeat step 3 through 6 up to three times for better
response. If the response does not improve, then your electrode is no longer functioning.
Replace with a new electrode – call your distributor for information.
51
Page 57
Instruction Manual PD 300
9.2 Dissolved Oxygen Principle
The probe is a galvanic measuring element which produces an output proportional to the
oxygen present in the medium in which it is placed. The galvanic probe design lets you take
measurements immediately – without the typical 15 minute wait of other dissolved oxygen
probes.
The probe consists of two parts:
• An upper part consisting of an anode, a cathode, and cable.
• A lower part consisting of a membrane cap, membrane, and electrolyte solution.
See Figure 42.
Oxygen diffuses through the membrane onto the cathode, where it is consumed. This process
produces an electrical current which flows through the cable to the meter. The electric current
produced is proportional to the oxygen that passes through the membrane and the layer of
electrolyte. This makes it possible to measure the partial pressure of oxygen in the sample at
a given temperature.
Since the DO in the sample is consumed by the cathode it is essential that a new sample must
flow past the membrane of the probe to prevent the occurrence of false readings. The probe
uses very little oxygen for its measurement. This enables it to function correctly with liquid
movement as low as 2.5 cm/sec.
The permeability of the membrane to oxygen varies greatly with temperature. Therefore
compensation is needed for this variation. The DO probe comes with an in-built Temperature
Compensation for the membrane variation.
52
Page 58
Instruction Manual PD 300
9.3 Probe Care and Maintenance
Under typical operating conditions, the D O probe should last for several years. Proper
care and maintenance will help you receive the maximum probe life and ensure more accurate
readings.
Since any deposits on the membrane surface act as a barrier to oxygen diffusing through the
membrane, the membrane must be cleaned at regular intervals to assure maximum reliability.
After using the probe, rinse the probe in clean water and wipe it with a soft cloth or paper to
avoid any hardening of deposits. If growth develops on the probe, use a disinfecting chemical
to clean.
NOTE
Although the membrane is strong and not easily damaged, wipe it gently while cleaning it. If
the membrane is damaged or torn, the probe will no longer function.
There are no special probe storage requirements.
53
Page 59
Instruction Manual PD 300
9.4 Membrane Housing Replacement
Replacement of the membrane cap housing/membrane is
required only when you cannot calibrate the probe, or if the
membrane is damaged.
Typical membrane damages are punctures or wrinkles
caused during measurements or cleaning. For more
information see Trouble-Shooting Guide section on page 60.
Your new DO probe comes with a replacement membrane
housing. To order more replacement membrane housing, ,
see the Accessories section on page 63.
9.4.1 To replace the membrane cap (with preinstalled membrane)
Replacement is much easier with single membrane housing.
Simply unscrew the old membrane cap housing, add solution
to the new housing, and screw the new cap housing in place.
1. Fill the membrane cap housing with electrolyte solution
and inspect the bottom for leaks. If the solution drops
are leaking from the membrane, use a new cap housing.
2. If the assembly is leak-free, fill the membrane cap
housing with electrolyte to the brim.
3. Tap the side of the housing gently to remove any air
bubble that may be sticking to the membrane
4. Screw the cap onto the probe. Excess electrolyte will
drain out.
Figure 42: DO Probe, membrane
cap housing & probe guard
5. Replace probe guard.
6. Calibrate the probe (see section 4) after the % saturation readings have stabilised.
54
Page 60
Instruction Manual PD 300
9.5 Membrane/O-ri n g Replacement (Optional Procedure)
It is recommended only experienced service personnel can perform this procedure.
This procedure is
OPTIONAL, and should only be performed if you have new membrane and
O-ring. You are also required to have a membrane installation tool. These items are available
as optional accessories in the Accessories section.
1. Pull off the probe guard. See figure on right.
2. Unscrew the membrane cap from the probe.
3. Hold the probe under hot running water and brush away the
white oxide on the cylindrical anode with a stiff plastic brush –
do not use metal cleaning material.
4. If the cathode has any deposits, remove them with a light
scouring powder. Do not polish the cathode.
5. Using the installation tool, unscrew and remove the membrane
lock from the membrane cap. See Figure 44.
6. Remove the membrane and O-ring. Discard both.
7. Rinse the membrane cap and membrane lock in tap water.
8. Install a new O-ring inside the membrane cap.
9. Install a new membrane. Make sure the membrane covers the
O-ring all around its circumference. See Figure 45.
Figure 43: DO Probe & its
various parts
55
Page 61
Instruction Manual PD 300
10. Using the installation tool, screw the membrane
lock back into the cap. Tighten the lock firmly over
the membrane and O-ring, but do not over tighten.
Close
Installation tool
Open
11. Inspect the membrane for wrinkles. If wrinkles
exist, remove the membrane and repeat steps 8 –
11.
Insert installation
12. Fill the membrane cap with water and inspect the
bottom for leaks. If water drops are leaking from
the membrane, re-seal the membrane on the O-
ring (repeat steps 8 – 11, for membrane
replacement only).
tool into slots on
membrane lock.
Then unscrew
membrane lock
from membrane
cap
13. If the assembly is leak-free, empty the water and
fill the membrane cap with electrolyte to the brim.
14. Tap the side of the housing gently to remove any
air bubble that may be sticking to the membrane.
Figure 44: Use tool to take out (or
put in) membrane
15. Screw the cap onto the probe. Excess electrolyte
will drain out.
16. Replace probe guard.
17. Calibrate the probe (see section 4) after the %
saturation readings have stabilised.
NOTE
Membranes can only be used once. When a membrane
cap is screwed onto the probe, the membrane is
stretched by the cathode. If the same O-ring and the
membrane is used a second time it will not fit
perfectly onto the cathode. This will result is erratic
readings.
56
Figure 45 : Membrane cap housing
with O-ring & membrane
Page 62
Instruction Manual PD 300
9.6 Electrolyte Solution
The electrolyte solution in your probe’s cap will deplete on usage and will need to be replaced
periodically.
Your new DO probe comes with accessories of one 50-ml replacement electrolyte solution and
a spare membrane cap. The replacement electrolyte comes premixed and ready to use. To
order more electrolyte solution, see Accessories section.
57
Page 63
Instruction Manual PD 300
9.7 DO Probe Troubleshooting Table
When experiencing difficulties with the equipment, keep in mind of the following:
1. Check for the obvious, such as physical condition of the probe, any signs of damage to
the cable, power and signal connections etc.
2. Determine whether it is the probe, meter or the surrounding environment that is causing
the problem.
The following troubleshooting table identifies most of the problems likely to occur:
Problem Probable Causes Solution
1. Fluctuating readings
when probe is shaken
or bumped lightly or
when membrane is
touched.
2. With membrane cap
removed and probe
internals thoroughly
dry, the reading from
the probe is not zero
and/or is erratic.
Probe has lost electrolyte – a
sloshing noise will be heard
when the probe is shaken.
Torn or damaged membrane.
Wet connections in the wiring
or within probe. See Problem
2.
Moisture has entered the
system – either into the probe
itself or at junctions or points
in the cable. This moisture
creates a secondary galvanic
action in addition to that
produced by the probe and
results in non-zero or erratic
readings.
Determine whether it is a physical or
electronic problem with the probe.
Unscrew the membrane cap, discard
the electrolyte, membrane and Oring.
Dry the internals of the probe,
especially the cathode with a soft
cloth.
Switch the meter ON and observe the
display.
If the display reads zero, the probe
and cable circuitry are alright.
Service probe and change
membrane.
If the display does not read zero, but
some other value, then there is
probably moisture somewhere. See
Problem 2.
Locate the source of moisture by
process of elimination. If moisture
has entered a junction box or a cable
joint, thoroughly dry out the area and
take measures to prevent
reoccurrence.
58
Page 64
Instruction Manual PD 300
Problem Probable Causes Solution
3. It is not possible to
calibrate the probe in
air – the display will
not read high enough
after fully adjusting the
offset.
4. Display values are
erratic when
membrane is lightly
touched. Membrane
has bulged outwards.
Probe has dried out – no
electrolyte inside.
Probe is overdue for servicing
– excessive build-up of anode
oxide.
A deposit has built-up on the
silver cathode, which is
inhibiting the reduction of
oxygen at its surface.
The membrane has been
damaged.
The breather hole has become
plugged. This prevents
release of internal pressure
due to expansion of electrolyte
caused by the probe being
warmed up in the sun, for
example.
NOTE: The cathode must not be polished – the surface must remain dull (do not use a wire
brush).
Service probe and change
membrane. Use a stiff nylon brush to
remove the oxide built-up from the
anode. Do not use a wire brush. It is
only necessary to remove the loose
oxide layer. If it is suspected that the
anode is badly corroded, replace with
a new DO probe. Remember to
tighten the nut under the anode
before fitting a new anode.
If it is suspected that a deposit is
coating the silver cathode, clean the
cathode with 400 grit wet/dry emery
paper or with some scouring powder.
The deposit is sometimes visible as a
brownish stain on the surface of the
cathode.
If the membrane has been damaged
change it and service the probe.
If the breather hole is blocked, clear
hole with a pin or fine gauge wire.
Change the membrane and service
the probe.
59
Page 65
Instruction Manual PD 300
10 TROUBLE SHOOTING GUIDE
Problem Cause Solution
Press ‘ON’ key but no
display
Not responding to key
press
Unstable readings a) Insufficient electrolyte in probe.
“OR” or “UR” on
upper display
Temperature reading
erratic or lower
display reads “OR”
Slow response a) Dirty / Oily probe. a) Clean probe. See Probe Care
a) Batteries not in place
b) Batteries not in correct polarity
(+ and – position).
c) Weak batteries
a) HOLD mode in operation.
b) Damaged key pad.
c) Internal program error.
b) Air bubbles in probe.
c) Dirty probe.
d) Probe not deep enough in
sample.
e) External noise pickup or
induction caused by nearby
electric motor.
f) Broken probe.
a) Probe is shorted.
b) Probe is in out-of-range (OR) or
under range (UR) condition.
c) Broken probe.
a) Temperature of solution is out
of range.
a) Check that batteries are in place
and making good contact.
b) Re-insert batteries with correct
polarity.
c) Replace batteries.
a) Cancel HOLD mode by
pressing Hold key.
b) Return to dealer.
c) Reset all internal programs by
reinserting batteries.
a) Fill probe with reference
electrolyte.
b) Tap probe to remove bubbles.
c) Clean the probe and re-
calibrate.
d) Make sure sample entirely
covers the probe sensors.
e) Move or switch off interfering
motor.
f) Replace probe.
a) Test probe. Make sure probe is
fully connected to meter.
b) Use different solution.
c) Replace probe. See
Accessories section.
a) Heat or cool solution.
& Maintenance section.
b) Check membrane of DO probe.
Replace if necessary
60
Page 66
Instruction Manual PD 300
11 ERROR MESSAGES
LCD Display Indicates Cause Solution
Err annunciator Unrecognised input
CAL & Err
annunciators on /
Buffer and
electrode
indicators blink.
Battery indicator
blinks
from keypad
Calibration error. Wrong value input at
Low battery level. Need new batteries or
Wrong input in selected
mode.
calibration.
Dirty probe.
battery connection is
bad.
Release key. Select valid
operations depending on
mode.
Check your input value,
clean probe.
See Calibration or Probe
Maintenance section.
Clean battery contacts.
Replace batteries with fresh
ones, noting polarity.
If an error message appears in the primary display (the upper row of larger digits), switching off the
meter and switching it on again may eliminate the error message.
If error persists, or the meter shows incorrect values, return the meter. See Warranty and Return of
Items sections.
For a complete diagram of the display, see page 2.
61
Page 67
Instruction Manual PD 300
12 SPECIFICATIONS
pH
Dissolved Oxygen
% Saturation of
Oxygen
Temperature
Salinity Correction
Barometric
Pressure
Correction (mm Hg)
DO Probe Galvanic with integral Temperature Sensor
Response Time 40 seconds to achieve 93% of the reading
No. of DO Calibration Points Single point at 100% in saturated air or air-saturated water
Temperature Compensation Automatic or manual from 0 to 100 °C
pH Calibration Up to 5 points (pH 1.68, 4.01, 7.00, 10.01 & 12.45)
pH Slope & Offset Display Yes
Operating Temperature Range 0 to 50 °C
HOLD function Yes
Auto-Off function 20 minutes after last key press
Averaging/Stability function Selectable
Display Customised Dual LCD
Back Lit Display Yes
Inputs BNC for pH, 6-pin military type for DO & Temperature
Power Requirements four 1.5 V AAA-sized batteries (included)
Battery Life > 100 hours continuous use
Dimensions Meter: 19L x 9.5W x 5.7H cm (7.5”x3.75”x2.25”)
Shipping weight Meter: 0.92 kg (2.0 lbs)
Range
Resolution
Relative accuracy
Range
Resolution
Relative accuracy
Range
Resolution
Relative accuracy
Range
Resolution °C
Resolution °F
Relative accuracy
Range
Resolution
Method
Range
Resolution
Method
-2.00 to 16.00 pH
0.01 pH
±0.01 pH
0.00 – 20.00 mg/L or ppm
0.01 mg/L; 0.01 ppm
± 1.5% of Full Scale
0.0 – 200.0 %
0.1 %
± 1.5% of Full Scale
-10.0 to 110.0 °C, (14.0 to 230°F)
0.1°C
0.1°F (14.0°F – 199.9°F) and 1°F(200°F to 230°F)
± 0.3°C (± 0.5°F)
0.0 – 50.0 ppt
0.1 ppt
Automatic correction after manual input
500 to 1499 mm Hg or 66.6 to 199.9 kPA
1 mm Hg or 0.1 kPA
Automatic correction after manual input
Case: 23.3L x 21.6W x 7.0H cm (9.17”x8.5”x2.8”)
Probe: 17.3L x 3.2 Dia cm (6.81”x1.26”dia) , with 3-m cable
Carrying case with meter: 2.2 kg (4.85 lbs)
62
Page 68
Instruction Manual PD 300
13 ACCESSORIES
Replacement Meters and accessories
Eutech Instruments
Item Eutech Instruments
Ordering Code
Waterproof CyberScan PD 300 portable pH/DO meter (with backlit) – with pH
electrode (EC-DA93506-03B), DO probe with built-in temperature sensor (ECDOHANDYNEW, 1 assembled membrane cap housing, 50-ml refilling
electrolyte & hard plastic carrying case with pH calibration solutions (ECPHWP-KIT).
Galvanic Dissolved Oxygen probe with built-in temperature sensor and 3meter (10-ft) cable.
Submersible Gel-filled pH combination electrode with 15-cm ABS guard, single
annular ceramic junction and 3-meter cable length
Temperature probe for CyberScan PD 300 meter EC-PHWP-TEM01W
pH Carrying kit set – hard plastic carrying case comprises of 1 x pH 4.01, pH
7.00, pH 10.01 buffer solutions (60-ml), and 1 x empty rinse water bottle (480-
ml)
Soft carrying pouch for meter EC-POUCH-02
Membrane & O-ring (pack of 5) 01X241603
Assembled Membrane Cap Housing 15X241402
Membrane removal tool 15X241502
Electrode Guard Removal Tool 15X241504
DO Refilling Electrolyte (50ml) 01X211226
EC-PDWP300/03K
EC-DOHANDYNEW
EC-DA93506-03B
EC-FE73528-01W
EC-PHWP-KIT
63
Page 69
Instruction Manual PD 300
Oakton Instruments
Item Oakton Instruments
Ordering Code
PD 300 meter with DO/Temp probe (35640-50) and pH probe (35805-23) 35632-00
PD 300 meter only 35632-02
PD 300 meter with DO/Temp probe (35640-50) and pH probe (35805-23),
electrode maintenance kit (35640-80), pH calibration solutions, and hard
carrying case.
Single-junction pH electrode, BNC, ABS body, 25-mm diameter, 10 ft (3 m)
cable
Double-junction pH electrode, BNC, ABS body, 25-mm diameter, 10 ft (3 m)
cable
Submersible DO/Temp electrode with 10 ft (3 m) cable 35640-50 (also
Submersible DO/Temp electrode with 25 ft cable 35640-52 (also
Submersible DO/Temp electrode with 50 ft cable 35640-54 (also
Submersible DO/Temp electrode with 100 ft cable 35640-56 (also
Membrane cap assembly (membrane, o-ring, membrane lock, bottom cap) 35640-72
Maintenance kit for DO electrode. Two membrane cap assemblies (35640-72)
and a 30-mL bottle of electrolyte.
Membrane installation tool. Required for 35640-74 & 35640-75 35640-79
Electrolyte solution, 500 mL 35640-71
Replacement membranes, pack of 5 35640-74
Replacement membranes, pack of 25 35640-75
Hard carrying case 35632-98
Temperature probe for pH or temperature only reading with PD 300 meter.
Note: DO reading is not possible when this probe is used.
pH 4.01 buffer sachets, 20 ml x 20 pcs. EC-BU-4BS 35653-01
pH 7.00 buffer sachets, 20 ml x 20 pcs. EC-BU-7BS 35653-02
pH 10.01 buffer sachets, 20 ml x 20 pcs. EC-BU-10BS 35653-03
pH De-ionised water rinse sachets, 20 ml x 20 pcs EC-RIN-WT 35653-00
pH sachet assortment pack – 5 each of pH 4.01,
pH 7.00, pH 10.01 and de-ionised water sachets
per box.
Protein cleaning solution for pH electrode EC-DPC-BT 00653-06
Storage solution for pH electrode EC-RE-005 00653-04
EC-AST-PK 35653-04
Oakton Instruments
Order Code
Item Oakton Instruments
Ordering Code.
Zero oxygen solution, 500 mL 00653-00
Premixed electrolyte solution, 500 mL 35640-71
Electrolyte power (58.5 g) 35640-70
65
Page 71
Instruction Manual PD 300
14 ADDENDUM 1: DISSOLVED OXYGEN & METER T HEORY
Dissolved Oxygen (DO) refers to the volume of oxygen that is contained in water. There are
two main sources of DO in water: from atmosphere and photosynthesis. Waves and tumbling
water mix air into the water where oxygen readily dissolves until saturation occurs. Oxygen is
also produced by aquatic plants and algae as a by-product of photosynthesis.
The amount of DO that can be held by water depends on 3 factors: water temperature, salinity,
and atmospheric pressure.
1. Amount of DO increases with decreasing temperature (colder water holds more oxygen).
2. Amount of DO increases with decreasing salinity (freshwater holds more oxygen than
saltwater does).
3. Amount of DO decreases with decreasing atmospheric pressure (amount of DO
absorbed in water decreases as altitude increases).
The chart below shows the solubility of DO in mg/L in water at various temperatures.
Solubility of oxygen in water con t act with water saturated air at
16
14
12
10
8
6
Solubility mg/L
4
2
0
0 102030405060
standard atmospheric pressure
Temperature °C
Figure 46: DO Solubility in Water vs Temperature °C
66
Page 72
Instruction Manual PD 300
15 ADDENDUM 2: PRESSURE VS ALTITUDE TABLE
Barometric pressure affects DO readings, therefore this meter lets you enter the correct
barometric pressure at your altitude. If you do not have equipment that lets you measure the
exact barometric pressure at your altitude, you can estimate it using the chart below.
If you change the barometric pressure setting from its factory setting (760 mm Hg), the %
saturation calibration value in air will automatically adjust to a value other than 100% (see
“corrected % saturation value” column below). The adjusted value is correct for the new
barometric pressure setting.
See page 37 for information on how to adjust the barometric pressure.
Pressure vs Altitude
ALTITUDE
feet meters
0 (sea level) 0 (sea level) 760 100
500 152.4 746 98.1
1000 304.8 732 96.3
1500 457.2 720 94.7
2000 609.6 707 93.0
2500 762.0 694 91.3
3000 914.4 681 89.6
3500 1066.8 668 87.8
4000 1219.2 656 86.2
4500 1371.6 644 84.6
5000 1524.0 632 83.0
5500 1676.4 621 81.6
6000 1828.8 609 80.0
PRESSURE (mm
HG)
CORRECTED %
SATURATION
VALUE
(1 feet ~ 0.3048 meters)
67
Page 73
Instruction Manual PD 300
16 ADDENDUM 3: METER FACTO RY DEFAULT SETTINGS
Resetting the meter to factory default settings clears all calibration data and most other setup
functions you might have changed. The following settings will remain as you have set them:
• Temperature unit of measure (°C or °F)
• The temperature offset calibration value.
NOTE
DO and pH data are cleared separately from each other.
• To clear pH data, see Sub group P4.0 on page 36.
• To clear DO data, see Sub group P7.0 on page 49.
Type Parameter P No. Default Remarks
pH parameters
CAL View pH calibration data P1.1 –– Calibration data f or 1st buffer, pH 1.68
P1.2 –– Calibration data for 2nd buffer, pH 4.01
P1.3 –– Calibration data for 3rd buffer, pH 7.00
P1.4 –– Calibration data for 4th buffer, pH 10.01
P1.5 –– Calibration data for 5th buffer, pH 12.45
ELE View electrode offset P2.1 0.00 mV No offset adjustment
View electrode slope P2.2 100.0 % No slope adjustment
COF Ready indicator P3.1 Ready On Ready indicator on; auto endpoint off
# pH calibration points P3.2 3 Selects number of calibration points.
°C or °F P3.3 No default °C or °F remains as selected
rSt pH factory default P4.0 No Retains your current settings
DO parameter
dPr Barometric pressure inputs. P1.1 Hg Hg or Pa.
760 mm Hg Adjustable from 500 – 1499 mm Hg
Salinity adjustment P1.2 0 ppt Adjust from 0 – 50.00 ppt (% only)
OFS Set % saturation offset P2.0 –– Up to ±10% (% only)
CAL View calibration data P3.0 [P2.0] –– ––
ELE View probe slope P4.1 [P3.1] 1.000 ––
View % saturation offset P4.2 [P3.2] 0% offs et ––
View mV = 100% saturation P4.3 [P3.3] 37.0 mV View mV = 0% saturation P4.4 [P3.4] 0.3 mV
COF Ready indicator P5.1 [4.1] Ready On Ready indicator on; auto endpoint off
Select mg/L or ppm units [P4.2] mg/L mg/L or ppm (mg/L mode only)
°C or °F P5.2 [P4.3] Retains settings °C or °F remains as selected
LCd Adjust back lit display P6.0 [P5.0] 8 Levels 0 to 8 (brightest)
rSt Reset to factory default P7.0 [P6.0] No Yes / no
* Program numbers with “[ ]” are accessed from mg/L (ppm) mode.
Minimum 2 points.
68
Page 74
Instruction Manual PD 300
17 WARRANTY
This meter is supplied with a three -year meter warranty and six-month pH/ DO probe warranty,
against significant deviations in material and workmanship.
If repair or adjustment is necessary and has not been the result of abuse or misuse within the
designated period, please return – freight pre-paid – and correction will be made without charge.
Eutech Instruments/ Oakton Instruments will determine if the product problem is due to deviations or
customer misuse.
Out of warranty products will be repaired on a charged basis.
Exclusions
The warranty on your instrument shall not apply to defects resulting from:
Improper or inadequate maintenance by customer
Unauthorised modification or misuse
Operation outside of the environment specifications of the products
69
Page 75
Instruction Manual PD 300
18 RETURN OF ITEMS
Authorisation must be obtained from our Customer Service Department or authorised distributor
before returning items for any reason. A “Return Goods Authorisation” (RGA) form is available
through our authorised distributor. Please include data regarding the reason the items are to be
returned. For your protection, items must be carefully packed to prevent damage in shipment and
insured against possible damage or loss. Eutech Instruments/ Oakton Instruments will not be
responsible for damage resulting from careless or insufficient packing. A restocking charge will be
made on all unauthorised returns.
NOTE: Eutech Instruments Pte Ltd/ Oakton Instruments reserves the right to make improvements in
design, construction, and appearance of products without notice.
70
Page 76
For more information on Eutech Instruments/ Oakton Instruments’ products, contact
your nearest distributor or visit our website listed below:
Oakton Instruments
P.O Box 5136,
Vernon Hills, IL 60061, USA
Tel: (1) 888-462-5866
Fax: (1) 847-327-2984