Thermo Fisher Scientific PC 510 User Manual

Instruction Manual
gyM
sy...
PC 510
Bench pH/Conductivity Meter
68X090816
Technol o
adeEa
Rev. 4 07/06
Preface
Thank you for choosing the PC 510 pH and Conductivity bench meter series. This manual serves to explain the use of the PC 510 bench meter. The manual functions as a step-by-
step operational guide to help you familiarise with the meter’s features and as a handy reference guide.
This instruction manual is written to cover as many anticipated applications and uses of the PC 510 bench meter as possible. If there are doubts in the use of the meter, please do not hesitate to contact the nearest Authorised Distributors.
Eutech Instruments/ Oakton Instruments cannot accept any responsibility for damage or m alfunction to the meter caused by improper use of the instrument.
The information presented in this manual is subject to change without notice as improvements are made, and does not represent a commitment on the part of Eutech Instruments Pte Ltd/ Oakton Instruments.
Note: Eutech Instruments Pte Ltd/ Oakton Instruments reserves the right to make improvements in design, construction, and appearance of our products without notice.
Copyright © 2001 All rights reserved. Eutech Instruments Pte Ltd Oakton Instruments
TABLE OF CONTENTS
1
INTRODUCTION 1
2 METER INFORMATION 1
2.1 Meter parts 1
2.2 Customised LCD 1
2.3 Slide-out card 1
2.4 Rear instrument panel 1
2.5 AC/DC adapter 2
2.6 Electrodes 2
2.7 Electrode holder 2
3 KEYPAD FUNCTIONS 2
3.1 Keypad 2
3.2 Display 4
4 PREPARATION 5
4.1 Connecting the Sensor Electrode 5
4.1.1 To connect the pH electrode: 5
4.1.2 To connect the conductivity/temperature probe: 5
4.2 Connecting the A.C. Adapter 5
5 CALIBRATION 6
5.1 Important information on meter calibration 6
5.2 Preparing the meter for calibration 6
5.3 pH calibration 7
5.3.1 Preparing for pH calibration 7
5.3.2 Before starting 7
5.3.3 To calibrate pH 7
5.4 Conductivity/TDS calibration 9
5.4.1 Preparing for conductivity/TDS calibration 9
5.5 TDS Calibration 11
5.5.1 Calibrating for TDS directly 11
5.6 Calibration with Condu ctivity Standard and TDS factor 11
5.6 Calibration with Condu ctivity Standard and TDS factor 12
5.7 Temperature Calibration 13
6 MEASUREMENT 14
6.1 Taking pH Measurements 14
6.1.1 Automatic Temperature Compensation 14
6.1.2 Manual Temperature Compens at io n (pH ) 14
6.1.3 Taking pH Measurements 15
6.2 Taking Conductivity or TDS Measurement 15
6.2.1 Automatic Temperature Compensation 15
6.2.2 Manual Temperature Compensa t io n 16
6.2.3 Setting a manual temperature compensation value 17
6.2.4 Taking Measurements (Conductivity or TDS) 17
6.2.5 Using Auto and Manual Ranging Function (for con duct i vity & TDS) 18
7 HOLD FUNCTION 20 8 ADVANCED SETUP FUNCTIONS 20
8.1 Advanced SETUP mode Overview 22
8.2 P1.0: Viewing previous pH calibration data 25
8.3 P2.0: Viewing pH electrode da ta 26
8.4 P3.0: pH Measurement configuration 27
8.4.1 P3.1: READY Indicator and auto endpoint function 28
8.4.2 P3.2: Selecting number of pH calibration points 29
8.4.3 P3.3 Selecting USA or NIST buffer 29
8.4.4 P3.4 Selecting °C or °F 30
8.5 P4.0: Resetting to factory default settings (pH) 31
8.6 P5.0: Viewing previous conductivity calibration data 32
8.7 P6.0: Viewing conductivity probe data 33
8.8 P7.0: Conductivity or TDS measurement configuration 34
8.8.1 P7.1: READY indicator and auto endpoint function 34
8.8.2 P7.2: Selecting °C or °F 35
8.8.3 P7.3: Selecting Automatic or Manual Temperature Compensation 35
8.8.4 P7.4: Setting the TDS factor 36
8.9 P8.0: Temperature 37
8.9.1 P8.1: Selecting the temperature coefficient 37
8.9.2 P8.2: Adjusting the norma lisation temperature 38
8.9.3 P9.0: Resetting to factory default settings (conductivity) 38
9
PROBE CARE AND MAINTENANCE 39
9.1 pH Electrode care 39
9.2 Conductivity electrode 40
10 TROUBLE SHOOTING GUIDE 41 11 ERROR MESSAGES 42 12 SPECIFICATIONS 43 13 ACCESSORIES 44 14 ADDENDUM 2: CALCULATING TDS CONVERSION FACTORS 48 15 ADDENDUM 3: STANDARD pH BUFFERS 48 16 ADDENDUM 4: CALCULATING TEMPERATURE COEFFICIENTS 49 17 ADDENDUM 5: METER FACTORY DEFAULT SETTINGS 50 18 WARRANTY 51 19 RETURN OF ITEMS 51
Instruction Manual PC 510
1 INTRODUCTION
Thank you for selecting the PC 510 pH and Conductivity bench meter. This step-by-step instruct ion manual gives you a detailed description on the use and operation of features on the meter. This PC 510 pH and Conductivity bench meter is designed to be user-friendly while providing unprecedented levels of accuracy, repeatability and reliability.
The PC 510 is an advanced microprocessor-based (ASIC - Application Specific Integrated Circuit) ideal for routine measurement that best meets discerning user’s individual needs. This multi-parameter meter reads pH, mV, Conductivity, Total Dissolved Solids (TDS) and temperature (°C or °F). It has splash-proof keypad, simultaneous pH/mV/Conductivity/TDS and temperature display on a large angled custom LCD. This instruction manual is illustrated with useful hints and diagrams that show which specific key-presses to access for each function.
2 METER INFORMATION
The PC 510 meter is packaged in a corrugated box that is made of environment-friendly materials and can be re-cycled.
2.1 Meter parts
The instrument is designed to give an aesthetic look as well as ergonomic functionality. A large custom dual LCD is provided at an angle for optimum viewing. A splash-proof keypad with audible tactile response gives you a good feel of the instrument. A slide-out instruction card offers a handy reference. Listed below are the major components of the meter.
2.2 Customised LCD
The PC 510 bench meter is characterised by large dual custom LCD (Liquid Crystal Display). The display has also mode annunciators for pH, temperature, mV, conductivity and TDS readings. The secondary (lower) display shows the temperature readings simultaneously with the primary (upper) display of measured mode. Special annunciators such as graphical symbols, error messages, measurement units and modes of operation are arranged around the primary and secondary displays to give a comprehensive display. The integration of graphics and error messages into the LCD provides you a higher level of user-friendliness and easy readability.
2.3 Slide-out card
A plastic slide-out card is provided at the bottom of the PC 510 bench meter. The function of this card is to provide a quick guide to the functions of the individual keys as well as to provide a useful troubleshooting reference.
2.4 Rear instrument panel
The PC 510 bench meter provides three connectors at the rear of the meters. These connectors are labeled CON/ T E MP , pH an d DC.
CON/TEMP pH
DC
Figure 1 : View of meter rear panel
1
Instruction Manual PC 510
Connector Function
For connecting 6-pins conductivity/TDS sensor with built-in temperature sensor to the
CON/TEMP
meter or optional temperature probe for use with pH electrode (when applicable). Always make sure that the connector is clean and dry.
pH
For connecting pH sensor with a BNC connector to the meter. Always make sure that the connect or is clean and dry.
DC For connection to the AC power source to the power jack (DC).
2.5 AC/DC adapter
The AC/DC adapter converts the power mains voltage 120/220 VAC to low DC voltage for the PC 510 bench meter operation. Two basic models of adapters are available depending upon power supply specification of each country.
Description Order Code Voltage
AC Adapter 120 V EC-120-ADA / 35615-07 110-120 V, 50-60 Hz AC Adapter 220 V EC-220-ADA / 35615-08 220-240 V, 50-60 Hz
2.6 Electrodes
Your meter incl udes two probes:
pH electrode with BNC connector
conductivity probe with built-in temperature sensor with a notched 6-pin connector
The temperature sensor built into the conductivity probe will also compensate for pH readings as long as both probes are in your solution at the same time.
If you want to use a “3-in-1” pH probe with a built-in temperature sensor, or if you want to use a separate temperature probe, you will need to disconnect the conductivity probe to allow for connection of the separate temperature sensor.
2.7 Electrode holder
The integral electrode holder serves as a handy holder for mounting the pH and conductivity/temperature probes during measurement or when idle.
The bench meter’s base plate has a side metal bar to which you can attach an integral swivel electrode holder. You can mount the electrode holder on either right or left side of the meter.
To position the electrode arm:
Use a Philips screwdriver to remove the screw holding the electrode holder. Slide the side metal bar until the second screw
holder.
slot lines up with the original screw hole. Use the screw removed earlier to secure the electrode holder into position. Note the side
3 KEYPAD FUNCTIONS
metal bar is reversible. If desired, remove screw holding electrode holder base and slide out of brackets. Slide base into brackets on opposite side and tighten screws. See Figure 2.
3.1 Keypad
To install electrode arm to the meter:
The PC 510 is equipped with large tactile response keypad for
To mount the electrode arm into the metal rod on the side bar,
ease of use. All keys have primary function with some keys having
align the slot with the metal rod and base of electrode arm. Push
secondary functions.
it downwards until it fully sits into position. Avoid using excessive force when fixing or removing. The electrode arm is ready for use.
NOTE: Move the base of electrode holder if you wish to swing the electrode holder about. To prevent the meter form toppling over causing accidental spills, DO NOT swing the body of the electrode holder.
Figure 2: To position electrode arm
Body of Electrode
Holder
Base of
Electrode
Holder
Side Metal Bar
Figure 3: Installing the
electrode arm
2
Instruction Manual PC 510
ON/OFF
SET
MODE
HOLD
D
L
O
H
R
E
E
T
G
E
D
O
M
L
T
E
S
A
C
AS
E
M
Figure 4: Keypad
N
N
E
A
R
F
F
O
/
N
O
Powers the meter on or off. When meter is switched on, it starts in the mode the meter was last in when powered off.
Enters advanced setup mode. SETUP mode lets you customise meter preferences and defaults, and view calibration and probe data.
Measurement: Press MODE to toggle between pH, mV, conductivity and TDS. Calibration: In calibration mode, press MODE to access temperature calibration.
Freezes the measured reading. To activate, press HOLD while in measurement mode. To release, press HOLD again.
Note: When auto endpoint feature is switched on, meter automatically holds reading after 5 seconds of stability. The HOLD indicator appears on the display. Press HOLD to release auto endpoint feature.
CAL/MEAS
ENTER/RANGE
Toggles between Calibration and Measurement mode. Example: If you are in pH measurement mode, press CAL/MEAS to enter pH calibration mode.
Note: Temperature calibration is available from pH, TDS or Conductivity calibration mode. In Setup mode: Press CAL/MEAS to return to main menu from sub menus. Press
CAL/MEAS again to return to measurement mode from main menu. ENTER: Press to confirm values in Calibration mode and to confirm selections in Setup
mode. RANGE: Press to switch to manual ranging in Conductivity or TDS mode. Press in Setup mode to scroll up through subgroups. Also lets yo u increase the values
in the conductivity, TDS and temperature cal ibration modes. Press in Setup mode to scroll down through subgroups. Also lets you decrease the
values in the conductivity, TDS and temperature calibration modes.
3
Instruction Manual PC 510
3.2 Display
The PC 510 features a large dual display that shows the measured parameter in the primary display, plus temperature in °C or °F in the secondary display. It also features mode annunciators that describe the meter’s functions.
Primary Display
123
SETUP MEAS CAL
20
19
18
17
16
1. SETUP mode indicator
2. MEASurement mode indicator
3. CALibration indicator
4. mV indicator
READY HOLD ON OFF
K =
-8.8.8.8
ERR
-1.8.8.8
15
Figure 5: Full LCD Screen
8. micro-Siemens indicator
9. parts per thousand indicator
10. parts per million indicator
11. temperature indicator
Secondary Display
mV
pH mS
µ
ppt ppm
°C °F
pH
%
S
4 5 6 7 8 9
10
11
12
ATC
1314
14. ERRor indicator
15. probe indicator
16. calibration solution indicator
17. cell constant indicator
5. % indicator
6. pH measurement indicator
7. milli-Siemens indicator
12. pH indicator
13. Automatic Temperature Compensation (ATC) indicator
4
18. ON/OFF indicator
19. HOLD indicator
20. READY indicator
Instruction Manual PC 510
4 PREPARATION
4.1 Connecting the Sensor Electrode
4.1.1 To connect the pH electrode:
1. Slide the BNC connector of the probe over the BNC connector socket on the meter. Make sure the slots of the connector are in line with the posts of the socket. Rotate and push the connector clockwise until it locks.
2. To remove electrode, push and rotate the connector anti-clockwise. While holding onto the metal part of connector, pull it away from the meter
NOTE:
Keep connector dry and clean. Do not touch connector with soiled hands.
CAUTION
: Do not pull the probe cord or the probe wires might disconnect.
4.1.2 To connect the conductivity/temperature probe:
1. Line up the notch and 6-pins on the probe connector with the holes in the connector located on the top of the meter. Push down and screw the metal sleeve to lock the probe connector into place. See Figure 1 on page 1 for the meter rear panel view.
2. To remove probe, unscrew the metal sleeve and slide up the probe connector. While holding onto the metal sleeve, pull probe away from the meter.
NOTE
: Follow the same directions to connect an optional separate temperature element.
Keep connector dry and clean. Do not touch connector with soiled hands.
CAUTION: Do not pull on the probe cord or the probe wires might disconnect.
4.2 Connecting the A.C. Adapter
1. Before plugging in the A.C. adapter, switch off the meter and the power source of the A.C. adapter. This is a safety precaution that should be adhered to safeguard your meter.
2. The A.C. adapter should have the following settings: Output voltage: 9 V D.C.
Current: 500 mA
: Ensure that the input mains voltage (110/220/240 V) matches your adapter requirements.
NOTE
3. Insert the D.C. jack into the socket at rear panel of the meter as shown in Figure 1on page 1.
4. Switch on the power to the adapter, followed by the meter.
5
Instruction Manual PC 510
5 CALIBRATION
5.1 Important information on meter calibration
When you calibrate your meter, old calibration points are replaced on a “point by point” basis in pH, and on a “range by range” basis in conductivity or TDS.
For example:
pH
: if you previously calibrated your meter at pH 4.01, 7.00 and 10.01, and you recalibrate at pH
7.00, the meter retains the old calibration data at pH 4.01 and pH 10.01.
Conductivity
: if you previously calibrated your meter at 1413 µS in the 0 to 1999 µS range and
you recalibrate at 1500 µS (which is also in the 0 to 1999µS range), the meter will replace the old calibration data (1413 µS) in that range. The meter will retain all calibration data in other ranges.
TDS
: If you previously calibrated your conductivity meter at 300 ppm in the 0 to 999 ppm range
and you re-calibrate at 500 ppm (which is also in the 0 to 999 ppm), the meter will replace the old calibration data (300 ppm) in that range. The meter will retain all calibration data in other ranges.
To view current calibration points:
pH
: Program P1.0 in the SETUP section 8.2 page 38.
Conductivity & TDS
: Program P5.0 in the SETUP section, page 32.
To completel y recalibrate your meter , or when you use a replac ement probe, it is best to c lear old calibration data by resetting the meter.
To reset the meter to its factory defaults:
pH
: Program P4.0 in the SETUP section, page 32.
Conductivity & TDS
NOTE
: Resetting the meter will set meter to factory defaults. Conductivity and pH must be reset
: Program P9.0 in the SETUP section, page 38.
separately.
For directions on how to calibrate your meter:
See section 5.3 on page 7 for pH calibration
See section 5.4 on pages 9 for conductivity calibration
See section 5.5 on page 11 for TDS calibration
5.2 Preparing the meter for calibration
Before starting calibration, make sure you are in the correct measurement mode. When you switch on the meter, the meter starts up in the measurement mode you shut it off in. For example, if you shut the meter off in pH measurement mode, the meter will be in the pH measurement mode when you switch the meter on.
Do not re-use calibration solutions after calibration. Contaminants in the solution can affect the calibration, and eventually the accuracy of the measurements. See section 13 on “Accessories” on page 44 for information on our high quality calibration solutions.
6
Instruction Manual PC 510
5.3 pH calibration
NOTE: We recommend that you perform at least a 2-point calibration using standard buffers that bracket (one above and one below) the expected sample range.
5.3.1 Preparing for pH calibration
This meter is capable of up to 5-point pH calibration to ensure accuracy across the entire pH range of the meter. Select from the following buffer options:
USA: pH 1.68, 4.01, 7.00, 10.01 and 12.45
NIST: pH 1.68, 4.01, 6.86, 9.18 and 12.45
The meter automatically recognises and calibrates to these standard buffer values, which makes pH calibration faster and easier.
NOTE:
Selection of USA or NIST buffer standards must be done prior to calibration. Refer to Section
8.4 on P3.3 on page 29.
5.3.2 Before starting
Be sure to remove the protective electrode storage bottle or rubber cap of the probe before calibration or measurement. If the electrode has been stored dry, hydrate the probe in tap water for 10 minutes before calibrating or taking readings to saturate the pH electrode surface and minimise drift.
Wash your probe in de-ionised water after use, and store in electrode storage solution. If storage solution is not available, use pH 4.01 or
7.00 buffer for short term storage. DO NOT store electrode in distilled
or de-ionised water.
MEAS
pH
7.16
°C
22.3
L
A
C
S
A
M
E
ATC
5.3.3 To calibrate pH
1. If necessary, press the MODE key to select pH mode. The pH indicator appears in the upper right hand corner of the display.
2. Rinse the probe thoroughly with de-ionised water or rinse solution. Do not wipe the probe as this causes a build-up of electrostatic charge on the glass surface.
3. Dip the probe into the standard calibration buffer. The end of the probe must be completely immersed into the sample. Stir the probe gently to create a homogeneous sample.
NOTE
: The temperature element is in the conductivity cell. For
temperature compensated readings, dip the conductivity cell or ATC probe into the calibration buffer as well.
4. Press CAL/MEAS to enter pH calibration mode. The CAL indicator will be shown. The primary display will show the measured reading while the smaller secondary display will indicate the pH standard buffer solution.
5. Wait for the measured pH value to stabilise. If the READY indicator has been activated through the Setup, the READY appears when the reading is stable.
6. Press ENTER to confirm calibration. The meter is now calibrated to the current buffer. The lower display automatically scrolls through the remaining buffer options.
If you are performing multi-point calibration, go to step 7.
CAL
7.16
7.00
CAL
READY
7.16
7.00
E
R
T
N
E
E
G
N
A
R
CAL
READY
7.00
7.00
Figure 6: pH calibration
pH
pH
pH
pH
pH
pH
7
Instruction Manual PC 510
correct buffe
used
If you are performing one-point calibration, go to step 9.
7. Rinse the electrode with de-ionised water or rinse solution, and place it in the next pH buffer.
READY
CAL
pH
4.15
8. Follow steps 5 to 7 for additional calibration points.
9. When calibration is complete, press CAL/MEAS to return to pH measurement mode.
4.01
R
T
E
N
E
E
G
N
A
R
pH
NOTE
: To exit from pH calibration mode without confirming calibration,
DO NOT press ENTER in step 6. Press CAL/MEAS instead.
If the selected buffer value is not within ±1.0 pH from the measured pH value: the electrode and buffer icon blink and the ERR annunciator appears in the lower left corner of the display.
To limit the number of pH buffer values available during calibration, see section 8.4 Setup P3.2 on page 29.
CAL
READY
pH
4.01
4.01
Figure 7: Next point calibration
for pH 4.01
CAL
pH
pH
0.64
ERR
1.68
Figure 8: Err message and
electrode icon will appear if
in
r is
pH
8
Instruction Manual PC 510
5.4 Conductivity/TDS calibration
The PC 510 has 5 measuring ranges. You can calibrate 1 point each of the measuring ranges (up to 5 points). If you are measuring values in more than 1 range, make sure to calibrate each of the ranges you are measuring. All new calibration data will over-ride existing stored calibration data for each measuring range you calib rate.
If you are measuring in ranges near to or greater than 20 mS (10 ppt), or near to or lower than 100 µS (50 ppm), calibrate the meter at least once a week to get specified ±1% Full Scale accuracy.
If you are measuring in the mid-ranges and you washed the probe in de-ionised water and stored it dry, calibrate the meter at least once a month.
If you take measurements at extreme temperatures, calibrate the meter at least once a week.
5.4.1 Preparing for conductivity/TDS calibration
For best results, select a standard value close to the sample value you are measuring. Alternatively, use a calibration solution value that is approximately 2/3 the Full-Scale value of the measurement range you plan to use. For example, in the 0 to 1999 µS conductivity range, a 1413 µS solution is a good solution for calibration.
See the table below for recommended calibration solution ranges.
Range
Indicator
Conductivity
Range
Recommended Calibration
Solution Range
TDS Range
Recommended Calibration
Solution Range
r 1 0.00 to 19.99 µS 6.00 to 17.00 µS 0.00 to 9.99 ppm 3.00 to 8.50 ppm r 2 0.0 to 199.9 µS 60.0 to 170.0 µS 10.0 99.9 ppm 30.0 to 85.0 ppm r 3 0 to 199 9 µS 600 to 1700 µS 100 to 999 ppm 300 to 850 ppm r 4 0.00 to 19. 99 mS 6.00 to 17.00 mS 1.00 to 9.99 ppt 3.00 to 8.50 ppt r 5 0.0 to 199.9 mS 60.0 to 170.0 mS 10. 0 to 200 p pt 30.0 to 170.0 ppt
Calibration Solution Ranges
Temperature Coefficient: These meter s are fact ory set to a temper ature co effic ient of 2. 1 % per °C.
For most applications this will provide good results. See Program P8.1 on page 37 to set the temperature coefficient to different value. See Addendum 2, “Calculating Temperature Coefficients” to determine the appropriate temperature coefficient for your solution.
Normalisati on Temp eratur e: The factory default value for normalisation temperature is 25 °C. If you need to normalise to a value other than 25 °C, see Program P8.2 on page 38.
Do not reuse calibration solutions after calibration. Contaminants in the solution can affect the calibration, and eventually the accuracy of the measurements. Use fresh calibration solution each time you calibrate your meter.
All new calibration data will over-ride existing stored calibration data for each measuring range calibrated.
9
Instruction Manual PC 510
Calibrating for Conductivity:
1. If necessary, press the MODE key to select conductivity mode.
2. Rinse the probe thoroughly with de-ionised water or a rinse solution, then rinse with a small amount of calibration standard.
3. If necessary, ensure that the probe’s yellow probe guard is attached. Dip the probe into the calibration standard. Immerse the probe tip beyond the upper steel band. Stir the probe gently to create a homogeneous sample.
4. Wait for the measured conductivity value to stabilise. If the READY indicator has been activated (SETUP program P7.1 – see page
31), the READY annunciator lights when the reading is stable.
5. Press CAL/MEAS to enter conductivity or TDS calibration mode. The CAL indicator will appear in the upper right corner of the display.
6. Press the  or  key to change the value on the primary display to match the value of the calibration standard.
7. Press ENTER to confirm calibration value. The meter returns to the MEAS (measurement) mode.
8. Repeat steps 1 to 7 for other measuring ranges.
NOTES
: When entering calibration mode, the meter will display the
factory default value. If the meter was previously calibrated, the display may “jump” to the factory default / uncalibrated value when switching from measurement to calibration mode.
To exit from Conductivity calibration mode confirming calibration, DO NOT press the ENTER key in step 7. Press CAL/MEAS instead.
This will retain the meter’s old calibration data in the measuring range of the calibration.
C
E
M
READY
E R
READY
A
A
T
N
N
A
MEAS
1409
L
S
1409
1413
E
R E
G
MEAS
1413
22.3
CAL
22.3
CAL
1409
22.3
°C ATC
µ
°C ATC
µ
°C ATC
S
µ
S
µ
S
S
You can offset the conductivity reading up to ±40% from default setting. If your measured value differs by more than ±40% clean or
Figure 9: Conductivity
calibration
replace probe as needed, or use a calibration standard with a higher value as required.
A wide selection of high-quality calibration standards is avail able. See page 44 for more information.
10
Instruction Manual PC 510
5.5 TDS Calibration
5.5.1 Calibrating for TDS directly
MEAS
The factory default setting for TDS conversion factor is 0.5. If your solution has a different TDS factor, you can improve calibration accuracy by setting the TDS factor prior to calibration. See Program P7.4 on page 36 for directions.
1. If necessary, press the MODE key to select TDS mode.
2. Rinse the probe thoroughly with de-ionised water or a rinse solution, then rinse with a small amount of calibration standard.
3. If necessary, ensure that the probe’s yellow probe guard is attached. Dip the probe into the calibration standard. Immerse the probe tip beyond the upper steel band. Stir the probe gently to create a homogeneous sample. Allow time for the reading to stabilise.
4. Press the CAL/MEAS to enter TDS calibration mode. The CAL indicator will appear in the upper right corner of the display.
5. Press the  or  key to change the value on the primary display to match the value of the calibration standard.
6. Press ENTER to confirm the calibration value. The meter returns to the MEAS (measurement) mode. See Figure 10.
7. Repeat steps 1 to 6 for other measuring ranges.
NOTES
:
To exit from TDS Calibration mode without confirming calibration, DO NOT press ENTER key in step 6. Press CAL/MEAS instead. This will retain the meter’s old calibration data in the measuring range of the calibration. You can offset the TDS reading up to ±40% from the
M
READY
N
E
A
R
READY
L
C
A
E
A
T
E
R
E
G
N
S
MEAS
300
265
25.8
CAL
265
25.8
CAL
300
265
25.8
ppm
°C ATC
ppm
°C ATC
ppm
ppm
°C ATC
default setting. If your measured value differs by more than ±40%, clean or replace probe as needed, or use a calibration standard with a higher value as required.
11
Figure 10: TDS calibration
Instruction Manual PC 510
5.6 Calibration with Conductivity Standard and TDS factor
The concentration of salts dissolved in solution increases the conductivity of that solution. This relationship varies from salt to salt and is roughly linear over a given range for a given salt.
The TDS conversion factor is the number used by the meter to convert from conductivity to TDS. Instead of calibrating for TDS directly (described above), you ca n calibrate the PC 510 bench meter
by:
1. Calibrating to conductivity standards (as described above) and then
2. Entering the appropriate TDS conversion factor into the meter.
To determine the conductivity to TDS conversion factor for your solution:
Addendum 1 on page 47 lists some commonly used conversion factors.
Addendum 2 on page 48 describes how to calculate the TDS conversion factor for other
solutions.
Enter the TDS conversion factor into your meter as described under Section 8.8, in Program P7.4, Setting the TDS Factor on page 36.
12
Instruction Manual PC 510
A
5.7 Temperature Calibration
The conductivity electrode (EC-CONSEN91W / 35608-50) supplied has a built-in temperature sensor. Alternatively, a separate temperature sensing element can be used (such as temperature probe EC-WPPHTEM-01W / 35618-05), or a “3-in-1” pH/Temperature combination electrode with ATC connection.
The conductivity probe is factory calibrated. Temperature calibration is recommended only if you suspect temperature errors may have occurred over a long period of time, or if you have a replacement probe.
Temperature calibration is accessible during pH, conductivity or TDS calibration.
Temperature Calibration
1. Make sure the conductivity electrode, or temperature probe, or “3-in-1” electrode is attached to the 6-pin connector.
2. Switch the meter on.
3. Press the CAL/MEAS key to enter calibration mode (either from pH or conductivity mode). The CAL indicator will appear above the primary display.
4. While in pH (or conductivity or TDS) calibration mode, press the MODE key to enter temperature calibration mode. The primary display shows the last set temperature value and the secondary display shows the temperature reading with zero offset.
5. Dip the ATC probe into a solution of known temperature (i.e. a temperature bath). Allow time for the temperature probe to stabilise.
6. Scroll with the  or  key to set the correct temperature value (i.e. the temperature of the temperature bath). You can adjust the reading by 0.1 °C or °F increments.
7. Once you have selected the correct temperature press the ENTER key. The meter automatically returns to measurement mode.
NOTES:
Figure 12: Temperature calibration
M
C
E
M
O
E
N
A
R
READY
A
L
A
S
D
E
R
T
E G
N
MEAS
7.16
22.3
CAL
7.16
7.00
22.3
22.0
E
MEAS
7.16
22.0
in pH mode
CAL
22.3
CAL
22.3
pH
°C ATC
pH
pH
°C
TC
°C
ATC
pH
°C ATC
You can offset the temperature reading up to ±5° from default reading.
To exit this program without confirming the temperature calibration value, DO NOT press ENTER,
press CAL/MEAS in step 7 instead.
13
Loading...
+ 39 hidden pages