Wi-Fi is either a trademark or a registered trademark of Wi-Fi Alliance in the United States and/or other countries.
Bluetooth is either a trademark or a registered trademark of Bluetooth Special Interest Group. Windows is either a
trademark or a registered trademark of Microsoft Corporation in the United States and/or other countries. All other
trademarks are the property of Thermo Fisher Scientific Inc. and its subsidiaries.
For U.S. Technical Support, please contact:
Thermo Fisher Scientific
3411 Silverside Road
Tatnall Building, Suite 100
Wilmington, DE 19810 U.S.A.
Contact your local distributor. For contact
information go to:
http://www.thermofisher.com/
NanoDropDistributors
E-mail: nanodrop@thermofisher.com
Thermo Fisher Scientific Inc. provides this document to its customers with a product purchase to
use in the product operation. This document is copyright protected and any reproduction of the
whole or any part of this document is strictly prohibited, except with the written authorization of
Thermo Fisher Scientific Inc.
The contents of this document are subject to change without notice. All technical information in this
document is for reference purposes only. System configurations and specifications in this document
supersede all previous information received by the purchaser.
Thermo Fisher Scientific Inc. makes no representations that this document is complete, accurate or
error-free and assumes no responsibility and will not be liable for any errors, omissions, damage or
loss that might result from any use of this document, even if the information in the document is
followed properly.
This document is not part of any sales contract between Thermo Fisher Scientific Inc. and a
purchaser. This document shall in no way govern or modify any Terms and Conditions of Sale,
which Terms and Conditions of Sale shall govern all conflicting information between the two
documents.
For Research Use Only. This instrument or accessory is not a medical device and is not intended to
be used for the prevention, diagnosis, treatment or cure of disease.
WARNING Avoid an explosion or fire hazard. This instrument or accessory is not
designed for use in an explosive atmosphere.
Note Locate the instrument away from air vents and exhaust fans to minimize
evaporation
The Thermo Scientific™ NanoDrop™ OneC is a compact, stand-alone
UV-Visible spectrophotometer developed for micro-volume analysis of a wide variety
of analytes. The patented sample retention system enables the measurement of
highly concentrated samples without the need for dilutions.
The NanoDrop One system comes with preloaded software and a touchscreen
display. NanoDrop One PC Control software can be installed on a local PC and used
to control the instrument and view data. The instrument can be connected to an
optional printer with a USB cable or to a remote printer through an Ethernet
connection or wireless network.
Note Before operating a NanoDrop One instrument, please read the safety and
operating precautions and then follow their recommendations when using the
instrument.
Thermo ScientificNanoDrop One User Guide 9
1 About the Spectrophotometer
Features
Features
Touchscreen
TheNanoDrop OneC spectrophotometer features the patented micro-volume sample
retention system. The NanoDrop One
dilute samples using standard UV-Visible cuvettes.
TheNanoDrop OneC comes with a built-in, 7-inch high-resolution touchscreen
preloaded with easy-to-use instrument control software. The touchscreen can slide
left or right to accommodate personal preference, and tilt forward or back for optimal
viewing
C
also features a cuvette holder for analyzing
10 NanoDrop One User GuideThermo Scientific
Cuvette Holder
Instrument light path
1
About the Spectrophotometer
Cuvette holder
Features
USB-A port
The NanoDrop OneC includes a cuvette holder for measuring dilute samples,
colorimetric assays, cell cultures and kinetic studies. The cuvette system has these
features:
• extended lower detection limits
• 37 °C heater option for temperature-sensitive samples and analyses
• micro-stirring option to ensure sample homogeneity and support kinetic studies
For details, see Measure a Sample using a Cuvette.
One USB-A port is located on the front of the instrument and two more USB-A ports
are located on instrument back panel.
Thermo ScientificNanoDrop One User Guide 11
1 About the Spectrophotometer
Accessories
Accessories
This section lists the accessories included for use with the NanoDrop OneC.
DYMO™ LabelWriter™ 450 USB Label Printer
Prints two 5/16-in x 4-in self-adhesive labels for transferring sample data directly into
laboratory notebooks or posting on bulletin boards. The software allows printing of
data from each sample measurement or from a group of samples logged and
measured together.
The printer connects to the instrument (front or back) via a USB cable (included).
PR-1 Pedestal Reconditioning Kit
Specially formulated conditioning compound
that can be applied to the pedestals to restore
them to a hydrophobic state (required to
achieve adequate surface tension for accurate
sample measurements). The kit includes
conditioning compound and applicators. For
more information, see Reconditioning the
Pedestals.
PV-1 Performance Verification Solution
Liquid photometric standard used to check instrument performance. For more
information, see Performance Verification.
12 NanoDrop One User GuideThermo Scientific
Instrument Detection Limits
1
About the Spectrophotometer
Instrument Detection Limits
Measurement
LocationPathlength (mm)
Upper Detection Limit
(10 mm Equivalent Absorbance)
Pedestal1.012.5
0.262.5
0.1150
0.05300
0.03550
Cuvette101.5
53
27.5
115
Thermo ScientificNanoDrop One User Guide 13
1 About the Spectrophotometer
This page is intentionally blank.
14 NanoDrop One User GuideThermo Scientific
Instrument Set up
Register Your Instrument
Register your instrument to receive e-mail updates on software and accessories for
the NanoDrop One instrument. An Internet connection is required for registration.
To register your instrument
1. Do one of the following:
–From any PC that is connected to the Internet, use any web browser to
navigate to our website.
2
On the website, locate NanoDrop One Registration and follow the instructions to
register the instrument.
Update Software
Quickly and easily download and install the latest NanoDrop One software and
release notes from our website. Follow the steps to update or upgrade the software
on your local instrument and/or install or update the NanoDrop One software on a
personal computer (PC). An Internet connection is required to download software.
To install or update NanoDrop One software on a PC
1. Insert the USB flash drive containing the installer software into an available USB
port on your PC, or open the installation folder downloaded from the internet.
2. Launch Start.exe and click Install. The software installer will run.
To install or update NanoDrop One software on the instrument
1. Copy the .zip file with the new software from your computer to a USB storage
device. Do not attempt to unzip the folder.
2. Insert the USB device into any USB port on the NanoDrop One instrument.
3. From the instrument Home screen, tap Settings > System > Update Software and choose the latest version of software.
Thermo ScientificNanoDrop One User Guide 15
2 Instrument Set up
Setting Up User Account Control (Optional)
Setting Up User Account Control (Optional)
User account control is managed using the Security Administration application. The
Thermo Scientific Security Administration software for NanoDrop One may be
purchased for instruments used in labs requiring 21 CFR Part 11 compliance. When
you launch Security Administration, you will need to enter your Windows log-in
information.
User Account Control
Launch the Security Administration application and select NanoDrop One from the
directory on the left to reveal Access Control and System Policies.
Access control
Access control is used to grant or deny individual users or groups access to
protected features in the instrument application. Add and remove users and groups
to the access list and set access rights using the drop-down for each entity.
16 NanoDrop One User GuideThermo Scientific
System policies
System Policies is used to set options that define the behavior of the client
application. See “Security Administration Policies.”
Security Administration Policies
System policies allow you to assign data and method creation and deletion and
editing privileges for users and groups.
Launch the Security Administration application and select NanoDrop One->
System Policies
2
Setting Up User Account Control (Optional)
Instrument Set up
Thermo ScientificNanoDrop One User Guide 17
2 Instrument Set up
Setting Up User Account Control (Optional)
You can add, delete, or edit policy groups and enable or disable the group’s users
permission to delete data. Select NanoDrop One-> Access Control.
When you are finished, select Save. Changes will take effect the next time
NanoDrop One is launched.
18 NanoDrop One User GuideThermo Scientific
Technical Support
For U.S./Canada Support, please contact:
Thermo Fisher Scientific
3411 Silverside Road
Tatnall Building, Suite 100
Wilmington, DE 19810 U.S.A.
Contact your local distributor. For contact information go to:
http://www.thermofisher.com/NanoDropDistributors
If you are experiencing an issue with your system, refer to the troubleshooting
information. If the issue persists, contact us. If you are outside the U.S.A. and
Canada, please contact your local distributor.
If your instrument requires maintenance or repair, contact us or your local distributor.
Thermo ScientificNanoDrop One User Guide 19
2
Instrument Set up
This page is intentionally blank.
20 NanoDrop One User GuideThermo Scientific
Application Measurement Ranges
Detection Limits for All Applications
3
Note Detection limits provided in the tables below are approximate and apply to
micro-volume measurements only; they are based on the instrument’s
photometric absorbance range (10 mm equivalent) of 0–550 A. For
measurements with 10 mm pathlength cuvettes, the photometric absorbance
range is 0–1.5 A.
Detection limits for standard applications
Sample Type
dsDNA2.0 ng/µL (pedestal)
ssDNA1.3 ng/µL (pedestal)
Lower Detection
Limit
0.20 ng/µL (cuvette)
0.13 ng/µL (cuvette)
Upper Detection
Limit
27,500 ng/µL
(pedestal)
75 ng/µL (cuvette)
18,150 ng/µL
(pedestal)
49.5 ng/µL (cuvette)
Typical Reproducibility
±2.0 ng/µL for sample
concentrations between 2.0 and
100 ng/µL samples;
±2% for samples >100 ng/µL
±2.0 ng/µL for sample
concentrations between 2.0 and
100 ng/µL samples;
±2% for samples >100 ng/µL
a
Thermo ScientificNanoDrop One User Guide 21
3 Application Measurement Ranges
Detection Limits for All Applications
Sample Type
Lower Detection
Limit
RNA1.6 ng/µL (pedestal)
0.16 ng/µL (cuvette)
DNA Microarray
1.3 ng/µL (pedestal)
(ssDNA)
0.13 ng/µL (cuvette)
Purified BSA by
Protein A280
IgG by Protein
A280
Purified BSA by
0.06 mg/mL (pedestal)
0.006 mg/mL (cuvette)
0.03 mg/mL (pedestal)
0.003 mg/mL (cuvette)
0.06 mg/mL (pedestal)
Proteins &
Labels
0.006 mg/mL (cuvette)
Protein BCA0.2 mg/mL (20:1
reagent/sample
volume)
Upper Detection
Limit
22,000 ng/µL
(pedestal)
Typical Reproducibility
±2.0 ng/µL for sample
concentrations between 2.0 and
a
100 ng/µL samples;
60 ng/µL (cuvette)
495 ng/µL (pedestal)
±2% for samples >100 ng/µL
±2.0 ng/µL for sample
concentrations between 2.0 and
Protein Lowry0.2 mg/mL (pedestal)4.0 mg/mL (pedestal)2% over entire range
Protein Bradford 100 µg/mL (50:1
reagent/sample
volume)
15 µg/mL (1:1
reagent/sample
8000 µg/mL
100 µg/µL
±25 µg/mL for 100–500 µg/mL
samples
±5% for 500–8000 µg/mL samples
±4 µg/mL for 15–50 µg/mL samples
±5% for 50–125 µg/mL samples
volume)
Protein Pierce
660
50 µg/mL (15:1
reagent/sample
2000 µg/mL
±3 µg/mL for 50–125 µg/mL samples
±2% for samples > 125 µg/mL
volume)
±3 µg/mL for 25–125 µg/mL samples
25 µg/mL (7.5:1
1000 µg/mL
±2% for samples >125 µg/mL
reagent/sample
volume)
a
Based on five replicates (SD=ng/µL; CV=%)
22 NanoDrop One User GuideThermo Scientific
Note To minimize instrument error with highly concentrated samples, make
dilutions to ensure that measurements are made within these absorbance limits:
• For micro-volume measurements, maximum absorbance at 260 nm (for
nucleic acids) or 280 nm (for proteins) should be less than 62.5 A.
• For measurements with 10 mm pathlength cuvettes, maximum absorbance
at 260 nm (or 280 nm for proteins) should be less than 1.5 A, which is
approximately 75 ng/µL dsDNA.
Detection limits for pre-defined dyes
3
Application Measurement Ranges
Detection Limits for All Applications
Sample Type
Cy3, Cy3.5, Alexa
Fluor 555, Alexa Fluor
Lower Detection
Limit
0.2 pmol/µL
(pedestal)
660
Cy5, Cy5.5, Alexa
Fluor 647
Alexa Fluor 488, Alexa
Fluor 594
0.12 pmol/µL
(pedestal)
0.4 pmol/µL
(pedestal)
Alexa Fluor 5460.3 pmol/µL
(pedestal)
a
Values are approximate
b
Based on five replicates (SD=ng/µL; CV=%)
Upper Detection
a
Limit
100 pmol/µL
(pedestal)
60 pmol/µL
(pedestal)
215 pmol/µL
(pedestal)
145 pmol/µL
(pedestal)
Typical Reproducibility
b
±0.20 pmol/µL for sample
concentrations between 0.20 and
4.0 pmol/µL;
±2% for samples >4.0 pmol/µL
±0.12 pmol/µL for sample
concentrations between 0.12 and
2.4 pmol/µL;
±2% for samples >2.4 pmol/µL
±0.40 pmol/µL for sample
concentrations between 0.40 and
8.0 pmol/µL;
±2% for samples >8.0 pmol/µL
±0.30 pmol/µL for sample
concentrations between 0.30 and
6.0 pmol/µL;
±2% for samples >6.0 pmol/µL
Thermo ScientificNanoDrop One User Guide 23
3
Application Measurement Ranges
This page is intentionally blank.
24 NanoDrop One User GuideThermo Scientific
Nucleic Acid Applications
Measure dsDNA, ssDNA or RNA
Measures the concentration of purified
dsDNA, ssDNA or RNA samples that
absorb at 260 nm.
Measure dsDNA, ssDNA or RNA
Reported Results
Settings
4
Detection Limits
Calculations
Measure dsDNA, ssDNA or RNA
Use the dsDNA, ssDNA and RNA applications to quantify purified double-stranded
(ds) or single-stranded (ss) DNA or RNA samples. These applications report nucleic
acid concentration and two absorbance ratios (A260/A280 and A260/A230). A
single-point baseline correction can also be used.
To measure dsDNA, ssDNA or RNA samples
NOTICE
• Do not use a squirt or spray bottle on or near the instrument as liquids will
flow into the instrument and may cause permanent damage.
• Do not use hydrofluoric acid (HF) on the pedestals. Fluoride ions will
permanently damage the quartz fiber optic cables.
Thermo ScientificNanoDrop One User Guide 25
4 Nucleic Acid Applications
Measure dsDNA, ssDNA or RNA
Before you begin...
Before taking pedestal measurements with the NanoDrop One instrument, lift the
instrument arm and clean the upper and lower pedestals. At a minimum, wipe the
pedestals with a new laboratory wipe. For more information, see Cleaning the
Pedestals.
To measure nucleic acid
1. From the Home screen, select the Nucleic Acids tab and select dsDNA,
2. Specify a baseline correction if desired.
3. Pipette 1–2 µL blanking solution onto the lower pedestal and lower the arm, or
ssDNA or RNA, depending on the samples to be measured.
insert the blanking cuvette into the cuvette holder.
Tip: If using a cuvette, make sure to align the cuvette light path with the
instrument light path.
4. Tap Blank and wait for the measurement to complete.
Tip: If Auto-Blank is On, the blank measurement starts automatically after you
lower the arm. (This option is not available for cuvette measurements.)
5. Lift the arm and clean both pedestals with a new laboratory wipe, or remove the
blanking cuvette.
6. Pipette 1-2 µL sample solution onto the pedestal and lower the arm, or insert the
sample cuvette into the cuvette holder.
7. Start the sample measurement:
–Pedestal: If Auto-Measure is On, lower arm; if Auto-Measure is off, lower arm
and tap Measure.
–Cuvette: Tap Measure.
When the sample measurement is completed, the spectrum and reported values
are displayed (see the next section).
8. When you are finished measuring samples, tap End Experiment.
9. Lift the arm and clean both pedestals with a new wipe, or remove the sample
cuvette.
26 NanoDrop One User GuideThermo Scientific
Typical nucleic acid spectrum
4
Nucleic Acid Applications
Measure dsDNA, ssDNA or RNA
Comparison of nucleic acid spectra with and without two
common contaminants
Best practices for nucleic acid measurements
• Isolate and purify nucleic acid samples before measurement to remove
impurities. Depending on the sample, impurities could include DNA, RNA, free
nucleotides, proteins, some buffer components and dyes. See Preparing
Samples for more information.
Note Extraction reagents such as guanidine, phenol, and EDTA contribute
absorbance between 230 nm and 280 nm and will affect measurement
results if present in samples (even residual amounts).
Thermo ScientificNanoDrop One User Guide 27
4 Nucleic Acid Applications
Measure dsDNA, ssDNA or RNA
• Ensure the sample absorbance is within the instrument’s absorbance detection
• Blank with the same buffer solution used to resuspend the analyte of interest.
• Run a blanking cycle to assess the absorbance contribution of your buffer
• For micro-volume measurements:
limits.
The blanking solution should be a similar pH and ionic strength as the analyte
solution.
solution. If the buffer exhibits strong absorbance at or near the analysis
wavelength (typically 260 nm), you may need to choose a different buffer or
application. See Choosing and Measuring a Blank for more information.
–Ensure pedestal surfaces are properly cleaned and conditioned.
–If possible, heat highly concentrated or large molecule samples, such as
genomic or lambda DNA, to 63 °C (145 °F) and gently (but thoroughly) vortex
before taking a measurement. Avoid introducing bubbles when mixing and
pipetting.
–Follow best practices for micro-volume measurements.
–Use a 1-2 µL sample volume. See Recommended Sample Volumes for more
information.
C
• For cuvette measurements (NanoDrop One
instruments only), use compatible
cuvettes and follow best practices for cuvette measurements.
Related Topics
• Measure a Micro-Volume Sample
• Measure a Sample Using a Cuvette
• Best Practices for Micro-Volume Measurements
• Best Practices for Cuvette Measurements
• Prepare Samples and Blanks
• Basic Instrument Operations
28 NanoDrop One User GuideThermo Scientific
Nucleic Acid Reported Results
dsDNA measurement screen
For each measured sample, the dsDNA, ssDNA and RNA applications show the UV
absorbance spectrum and a summary of the results. Below is an example of the
measurement screen of the PC control software:
4
Nucleic Acid Applications
Measure dsDNA, ssDNA or RNA
Application
Sampling
method
Sample name of
next measurement;
select to edit
Measure sample
Run Blank
UV spectrum
End experiment
Menu of options;
click to open
Right click graph
area to view
options
Mammalian
display
.
Contaminant
Detection
mouse icon to
toggle on and off
Menu of table options;
click to choose which
columns to report
; select
Sample name;
select to edit
Nucleic acid
concentration
Purity ratios
Click row to select sample
and update spectrum.
Measurement screen of PC Control software
Note Micro-volume absorbance measurements and measurements taken with
nonstandard cuvettes are normalized to a 10.0 mm pathlength equivalent.
Thermo ScientificNanoDrop One User Guide 29
4 Nucleic Acid Applications
Measure dsDNA, ssDNA or RNA
The following is an example of the measurement screen of the local control:
Menu of options;
tap to open
Pinch and zoom to
adjust axes; double-tap
to reset
Sample name;
tap to edit
Nucleic acid
concentration
UV spectrum
Swipe screen left to view
table with more
measurement results
Tap to select unit
Tap to end
experiment and
export data
Purity ratios
Tap r ow to select
sample and update
spectrum; tap
more rows to
overlay up to five
spectra. Press and
hold sample row to
view measurement
details.
Drag tab down/up
to see more/less
sample data
Measurement screen of NanoDrop One local control software
30 NanoDrop One User GuideThermo Scientific
Loading...
+ 264 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.