Thermo Fisher Scientific MycoSEQ User Manual

MycoSEQ™ Mycoplasma Real-Time PCR Detection Kit
USER GUIDE
for use with the PrepSEQ™ Mycoplasma Sample Preparation Kit
Catalog Numbers 4460623, 4460626
Publication Number 4465874
Revision D.0
For Research Use Only. Not for use in diagnostic procedures.
Life Technologies Ltd | 7 Kingsland Grange | Woolston, Warrington WA1 4SR | United Kingdom For descriptions of symbols on product labels or product documents, go to thermofisher.com/symbols-definition.
The information in this guide is subject to change without notice.
DISCLAIMER: TO THE EXTENT ALLOWED BY LAW, THERMO FISHER SCIENTIFIC INC. AND/OR ITS AFFILIATE(S) WILL NOT BE LIABLE FOR SPECIAL, INCIDENTAL, INDIRECT, PUNITIVE, MULTIPLE, OR CONSEQUENTIAL DAMAGES IN CONNECTION WITH OR ARISING FROM THIS DOCUMENT, INCLUDING YOUR USE OF IT.
Revision history: Pub. No. 4465874
Revision Date Description
D.0 28 September 2020 Update to include run and analysis information for AccuSEQ™ RealTime
C.0 24 May 2018 Updated template, legal, and content information. Reorganized
PCR Software v3.1.
content. Added information about using the AccuSEQ™ Software v2.0 Mycoplasma SEQ module.
Important Licensing Information: These products may be covered by one or more Limited Use Label Licenses. By use of these products, you accept the terms and conditions of all applicable Limited Use Label Licenses.
TRADEMARKS: All trademarks are the property of Thermo Fisher Scientific and its subsidiaries unless otherwise specified.
©2020 Thermo Fisher Scientific Inc. All rights reserved.

Contents

CHAPTER 1 Product information .................................................. 6
Product description ............................................................. 6
Contents and storage ............................................................ 6
Required materials not supplied ................................................... 7
Workflow ...................................................................... 8
CHAPTER 2 Methods .............................................................. 9
Prepare the sample .............................................................. 9
Prepare the kit reagents and premix solution ........................................ 9
Prepare the PCR reactions ...................................................... 10
Setup, run, and anlayze samples with AccuSEQ™ Software v3.1 on the
QuantStudio™ 5 Instrument .................................................... 11
Sign in to the AccuSEQ™ RealTime PCR Software ............................. 11
Create a MycoSEQ™ experiment ............................................. 12
Start the run (QuantStudio™ 5 RealTime PCR Instrument) ...................... 16
Monitor the run ............................................................ 17
Analyze the results ......................................................... 18
Guidance for unknown samples ............................................. 25
Guidance for controls ...................................................... 26
Example results with AccuSEQ™ Software v3.1 or later ......................... 28
MycoSEQ
APPENDIX A Troubleshooting .................................................... 34
AccuSEQ™ software ............................................................ 34
MycoSEQ™ kit ................................................................. 36
APPENDIX B Acceptance criteria ................................................ 37
Cycle threshold value (Ct) ....................................................... 38
Melting temperature (Tm) ........................................................ 39
Derivative value (DV) ............................................................ 40
Results interpretation using the acceptance criteria ................................ 41
Other acceptance criteria considerations ......................................... 41
Mycoplasma Real-Time PCR Detection Kit User Guide
3
Contents
APPENDIX C Use the kit with 7500 Fast System AccuSEQ™ Real-Time
PCR Detection Software v2.x .................................................... 42
Required materials not supplied ................................................. 42
Prepare the sample ............................................................. 43
Prepare the kit reagents and premix solution ...................................... 43
Prepare the PCR reactions ...................................................... 44
Create a new experiment ........................................................ 44
Define the experiment properties ............................................ 45
Setup the samples and controls ............................................. 45
View the plate layout ....................................................... 46
Save the experiment ....................................................... 47
Start the run (7500 Fast RealTime PCR Instrument) ................................ 47
Analyze the results ............................................................. 47
Set the baseline and threshold values ........................................ 48
Review the Results Summary ............................................... 48
Guidance for unknown samples ............................................. 50
Guidance for controls ...................................................... 51
Guidance for inconclusive results with AccuSEQ™ software v2.0 ................ 52
Example results with AccuSEQ™ Software v. 2.1.1 ................................. 53
Positive control ............................................................ 53
Negative control ........................................................... 54
Blank extraction control .................................................... 55
Positive extraction control .................................................. 56
Inhibition control and positive control ........................................ 57
Unknown sample: Negative result ............................................ 58
Unknown sample: Positive result ............................................ 59
Unknown sample: Positive result with decreased detection of DPC .............. 60
PCR inhibition ............................................................. 61
Multicomponent plots ...................................................... 62
APPENDIX D Use the kit with 7500 System SDS Software v1.4 or later ..... 63
Prepare the kit reagents and premix solution ...................................... 63
Prepare the PCR reactions ...................................................... 63
Seal the plates ................................................................. 65
Prepare the plate document ..................................................... 66
Perform PCR .................................................................. 68
Analyze the results ............................................................. 68
Set the baseline and threshold values ........................................ 68
Guidance for unknown samples ............................................. 69
Guidance for controls ...................................................... 70
Example positive results with SDS v1.4 software .............................. 70
4
MycoSEQ™ Mycoplasma Real-Time PCR Detection Kit User Guide
Example positive control extraction results with SDS v1.4 software .............. 72
Example negative results with SDS v1.4 software .............................. 73
APPENDIX E Background information ........................................... 74
APPENDIX F Kit specificity ....................................................... 75
Sensitivity ..................................................................... 75
Kit specificity .................................................................. 75
Inclusivity – detectable species .................................................. 75
Exclusivity – undetectable organisms ............................................. 76
APPENDIX G Good PCR practices .............................................. 78
Good laboratory practices for PCR and RT-PCR ................................... 78
Plate layout suggestions ........................................................ 78
Documentation and support ....................................................... 79
Contents
Related documentation ......................................................... 79
Customer and technical support ................................................. 80
Limited product warranty ........................................................ 81
MycoSEQ™ Mycoplasma Real-Time PCR Detection Kit User Guide
5
Product information
1
IMPORTANT! Before using this product, read and understand the information in the
“Safety” appendix in this document.

Product description

The MycoSEQ™ Mycoplasma Real-Time PCR Detection Kit detects Mycoplasma species simply, reliably, and rapidly. To detect the presence of these microorganisms, the assay uses the polymerase chain reaction (PCR) to amplify a target unique to a wide variety of Mycoplasma species.

Contents and storage

Table 1 MycoSEQ™ Mycoplasma Real-Time PCR Detection Kit (Cat. No. 4460623)
Contents
Box 1: MycoSEQ™ Mycoplasma RealTime PCR Detection Kit
10 Mycoplasma RealTime PCR Primer Mix
Negative Control
2 Power SYBR™ Green PCR Master Mix
Box 2: MycoSEQ™ Discriminatory Positive/Extraction Control
MycoSEQ™ Discriminatory Positive/ Extraction Control, 1,000 copies/µL
[1]
To purchase the MycoSEQ™ Mycoplasma Real-Time PCR Detection Kit that includes the PrepSEQ™ Mycoplasma Sample Preparation Kit, use Catalog Number 4460626.
[2]
The kit contains reagents for 100 reactions.
[1]
Cap color Amount
blue
1,000 µL
white
2 × 1,000 µL –25°C to –15°C on receipt, protected
white
yellow
[2]
325 µL –25°C to –15°C on receipt
2–8°C after first use, protected from
700 µL –25°C to –15°C
Storage
2–8°C after first use
from light.
light
6
MycoSEQ™ Mycoplasma Real-Time PCR Detection Kit User Guide

Required materials not supplied

Unless otherwise indicated, all materials are available through thermofisher.com. "MLS" indicates that the material is available from fisherscientific.com or another major laboratory supplier.
Item Source
Instrument
Chapter 1 Product information
Required materials not supplied
1
QuantStudio™ 5 RealTime PCR System with AccuSEQ™ Software v3.1 or later
Generic consumables
Disposable gloves MLS
Aerosol-resistant pipette tips MLS
Pipettors:
Positive-displacement
Air-displacement
Multichannel
Repeat (1µL–1mL)
Consumables for the QuantStudio™ 5 RealTime PCR System
MicroAmp™ Fast Optical 96Well Reaction Plate with Barcode, 0.1 mL
MicroAmp™ 96-Well Base N8010531
MicroAmp™ Optical Adhesive Film 4360954
MicroAmp™ Adhesive Film Applicator 4333183
Contact your local sales representative
MLS
4346906
MycoSEQ™ Mycoplasma Real-Time PCR Detection Kit User Guide
7
Chapter 1 Product information
1
Workflow
Workflow
Prepare the sample (page 9)
Prepare the kit reagents and premix solution (page 9)
Prepare the PCR reactions (page 63)
Create a MycoSEQ™ experiment (page 12)
Start the run (QuantStudio™ 5 RealTime PCR Instrument) (page 16)
Monitor the run (page 17)
Analyze the results (page 18)
8
MycoSEQ™ Mycoplasma Real-Time PCR Detection Kit User Guide
2
IMPORTANT! This chapter describes how to prepare and run PCR samples using
AccuSEQ™ Software v3.1 or later on the QuantStudio™ 5 Instrument (0.1 mL block). If you are using AccuSEQ™ software v2.x on the 7500 Fast, see Appendix C, “Use the kit with 7500 Fast System AccuSEQ™ Real-Time PCR Detection Software v2.x”. If you are using SDS software v1.4 or later, see Appendix D, “Use the kit with 7500 System SDS Software v1.4 or later”.
IMPORTANT! For information on how to avoid PCR contamination, see Appendix G,
“Good PCR practices”.

Prepare the sample

Prepare the DNA template for the PCR reactions using the PrepSEQ™ Mycoplasma Nucleic Acid Extraction Kit.
Methods
For more information, see:
The PrepSEQ™ Sample Preparation Kits for Mycoplasma, MMV, and Vesivirus User Guide (Pub. No. 4465957)
The PrepSEQ™ Express Nucleic Acid Extraction Kit for Mycoplasma, MMV, and Vesivirus Detection User Guide (Pub. No. MAN0016799)

Prepare the kit reagents and premix solution

1.
Thaw all kit reagents completely.
2.
Vortex briefly, then spin down the reagents.
MycoSEQ™ Mycoplasma Real-Time PCR Detection Kit User Guide
9
Chapter 2
2

Prepare the PCR reactions

Methods
3.
Prepare the Premix Solution according to the following table.
Component for premix
solution
Power SYBR™ Green PCR Master Mix, 2
Mycoplasma RealTime PCR Primer Mix, 10
Total premix solution volume
[1]
Includes 10% excess to compensate for pipetting errors.
4.
Mix the Premix Solution by gently pipetting up and down, then cap the tube.
Prepare the PCR reactions
1.
Dispense the following into each well to be used, gently pipetting at the bottom of the well.
Negative control reaction
To prepare…
Volume for one 30µL
reaction
15.0 µL 66.0 µL
3.0 µL 13.2 µL
18.0 µL 79.2 µL
Add 18 µL of Premix Solution
Add 12 µL of Negative Control (water)
Volume for four 30µL
reactions
In each tube or well…
[1]
Unknown or spiked sample reaction
Inhibition-control reaction
Positive control reaction
Add 18 µL of Premix Solution
Add 10 µL of unknown sample
Add 2 µL of Negative Control (water)
Add 18 µL of Premix Solution
Add 10 µL of unknown sample
Add 2 µL of the Discriminatory Positive Control (DPC)
Add 18 µL of Premix Solution
Add 2 µL of the DPC
Add 10 µL of Negative Control (water)
Note: The MycoSEQMycoplasma Discriminatory Positive/Extraction Control can be used as a spike control that is added to the unknown sample or lysate before sample preparation.
2.
Mix each sample by gently pipetting up and down.
10
MycoSEQ™ Mycoplasma Real-Time PCR Detection Kit User Guide
Chapter 2

Setup, run, and anlayze samples with AccuSEQ™ Software v3.1 on the QuantStudio™ 5 Instrument

3.
Seal the plate with MicroAmp™ Optical Adhesive Film. See “Seal the plates” on page 65.
4.
Briefly centrifuge the reaction plate.
Methods
Setup, run, and anlayze samples with AccuSEQ™ Software v3.1 on the QuantStudio™ 5 Instrument

Sign in to the AccuSEQ™ RealTime PCR Software

Thermo Fisher Scientific recommends configuring the Windows™ 10 screen save feature to require sign in when the screensaver is activated. This prevents other users from accessing the AccuSEQ™ RealTime PCR Software and making changes.
1.
Launch the AccuSEQ™ RealTime PCR Software by double-clicking the AccuSEQ icon .
2.
Enter the Username, then Password.
(First login only) The default username is Administrator and the default password is Administrator.
2
3.
Click Sign in.
The following restrictions may be seen in the software:
Access to functions in the software is based on the permissions associated with the individual user account.
If a user account does not have permission to perform a function, the function is grayed out in the software.
If the system is configured for password expiration, you will be periodically prompted to change your password. If the system is configured to monitor failed log in attempts, you will be locked out of the software if you incorrectly enter your user name or password more than a specified number of times.
MycoSEQ™ Mycoplasma Real-Time PCR Detection Kit User Guide
11
2
1
Chapter 2 Methods
2
Setup, run, and anlayze samples with AccuSEQ™ Software v3.1 on the QuantStudio™ 5 Instrument

Create a MycoSEQ™ experiment

1.
In the Home screen, click the Factory default/Admin Defined Template tab, then select MycoSEQ.
Note: To create a copy of an existing MycoSEQ™ experiment, see chapter 6 of the AccuSEQ™ RealTime PCR Software v3.1 User Guide (Pub. No. 100094287).
Custom MycoSEQ™ experiments must be set up as Quantitation-Standard
Curve type experiments with SYBR™ chemistry.
Factory default/Admin Defined Template tab
1
MycoSEQ template
2
2.
In the Experiment Properties pane of the Setup tab:
a.
(Optional) Change the systemgenerated name of the experiment.
b.
(Optional) Enter the plate Barcode, then add Comments.
Note: Comments are not editable post analysis.
Default MycoSEQ™ settings (These settings cannot be changed in the default template. See the AccuSEQ™ RealTime PCR Software v3.1 User Guide (Pub. No. 100094287) for creating custom templates).
Experiment TypeQuantitation-Standard Curve
ChemistrySYBR™ Green Reagents
Ramp SpeedStandard - 2hrs
c.
Click Next.
12
MycoSEQ™ Mycoplasma Real-Time PCR Detection Kit User Guide
1
2
Chapter 2 Methods
Setup, run, and anlayze samples with AccuSEQ™ Software v3.1 on the QuantStudio™ 5 Instrument
3.
In the qPCR Method pane of the Setup tab, view the default volume and cycling conditions (These settings cannot be changed in the default template. See the AccuSEQ™ RealTime PCR Software v3.1 User Guide (Pub. No.
100094287) for creating custom templates).
2
Figure 1 MycoSEQ™ template default cycling conditions
4.
Click Next.
5.
In the Samples pane of the Setup tab, enter the sample Name. Add additional Samples if needed.
Note: Only the sample Name is necessary for experiments run from the factory default MycoSEQ template.
IMPORTANT! Do not change the Targets.
Samples pane
1
Add—adds additional samples
2
MycoSEQ™ Mycoplasma Real-Time PCR Detection Kit User Guide
13
108
7
1
9
64 5
2
3
Chapter 2 Methods
2
Setup, run, and anlayze samples with AccuSEQ™ Software v3.1 on the QuantStudio™ 5 Instrument
Figure 2 MycoSEQ™ template default sample plate layout
Toolbar (in order: Undo, Redo, Copy, Paste, Delete, View)
1
5 default Samples with and without an inhibition control (IC)
2
Positive control (P)
3
Selected samples. Click-drag to add additional samples.
4
No template control (N)
5
6.
(Optional) Double-click wells to add Comments. Comments can also be added post-analysis.
14
MycoSEQ™ Mycoplasma Real-Time PCR Detection Kit User Guide
Chapter 2 Methods
Setup, run, and anlayze samples with AccuSEQ™ Software v3.1 on the QuantStudio™ 5 Instrument
7.
Click Next. The Run tab is displayed.
8.
Experiments are autosaved in the software. To save additional changes made to the experiment, exit the experiment. The software prompts you to save changes. Click Yes.
Note: Clicking Save As will create a copy of the experiment.
2
MycoSEQ™ Mycoplasma Real-Time PCR Detection Kit User Guide
15
2
1
Chapter 2 Methods
2
Setup, run, and anlayze samples with AccuSEQ™ Software v3.1 on the QuantStudio™ 5 Instrument

Start the run (QuantStudio™ 5 RealTime PCR Instrument)

Start the run in the AccuSEQ™ Software v3.1 or later.
Option Description
If the experiment is open Click Start Run.
If the experiment is closed
1.
Open the experiment.
2.
Click the Run tab.
3.
Click Start Run.
1
Run tab
2
Start Run button
A message stating Run has been started successfully is displayed when the run has started.
16
MycoSEQ™ Mycoplasma Real-Time PCR Detection Kit User Guide

Monitor the run

6
3
1
2
5
4
Chapter 2 Methods
Setup, run, and anlayze samples with AccuSEQ™ Software v3.1 on the QuantStudio™ 5 Instrument
During the instrument run, you can monitor the run from the following places:
• On the instrument touchscreen.
• In the Monitor the Run pane of the AccuSEQ™ Software (Home) screen.
1
Instrument name
2
Instrument status (Ready, Running, Oine)
3
Calibration status
– Calibrated—All required calibrations are "Current". Required calibrations include:
Background, ROI/Uniformity, and Dyes. – Not calibrated—None of the required calibrations are complete. – Requires calibration—One or more dyes are not calibrated.
4
Time lapsed (if a run is in progress)
5
Total run time
6
Experiment name
• In the Open Existing Experiments pane of the AccuSEQ™ Software (Home) screen. The experiment being run is the first experiment listed. Status is Run.
2
Note: You cannot start another run while the instrument status is Running.
MycoSEQ™ Mycoplasma Real-Time PCR Detection Kit User Guide
17
2
1
Chapter 2 Methods
2
Setup, run, and anlayze samples with AccuSEQ™ Software v3.1 on the QuantStudio™ 5 Instrument
When the run is complete, the status changes to Analysis and the status bar displays as complete.
1
Open Existing Experiments pane
2
Experiment status
• In the Run tab of the AccuSEQ™ RealTime PCR Software. You can perform the following actions.
– Select wells in the plate layout to highlight respective curves in the plot.
– Hover over curves in the plot for well information.
– (Optional) Change what is displayed in the table wells, by selecting Sample
Name, Sample Color, or Target in the View dropdown list.

Analyze the results

On run completion, the Post Run Summary displays the run length, user and instrument information, and a list of any errors that occurred.
When the run is complete, unload the plate from the instrument.
IMPORTANT! The acceptance criteria that are provided in this section are based on
our current knowledge of assay performance in detection of Mycoplasma recovered from a wide variety of unknown sample matrices. We recommend that you qualify and validate the assay internally using samples that are specific to your process and manufacturing environment (raw materials, bioreactor, or cell line samples) to ensure that these criteria are appropriate.
For specific sample types, it may be necessary to make slight changes to the acceptance criteria based on specific results. We can provide you with one-on-one support during this process.
Workflow:
Review MycoSEQ™ experiments
“View the Result Summary” on page 19
“Evaluate the overall shape of the Amplification Plot curves” on page 20
“View and evaluate the Melt Curve Plot” on page 22
18
MycoSEQ™ Mycoplasma Real-Time PCR Detection Kit User Guide
1
4
3
2
Chapter 2
Setup, run, and anlayze samples with AccuSEQ™ Software v3.1 on the QuantStudio™ 5 Instrument
Methods
See the AccuSEQ™ RealTime PCR Software v3.0 User Guide (Pub. No. 100084348) for more information on analyzing results.
View the Result Summary
The Result tab is populated when the run is complete.
Calls are made based on the customizable Rule Settings within the Analysis
Settings for MycoSEQ™ experiments. See the AccuSEQ™ RealTime PCR Software v3.1 User Guide (Pub. No. 100094287) for more information.
1.
View the Plate Call.
a.
View the calls for the positive controls (POS).
b.
View the calls for the no template controls (NTC).
The number inside the circle indicates the number of samples that passed or failed. The overall Plate Status (VALID or INVALID) is determined by the POS and NTC calls.
Note: One passing POS and NTC call are required for a Plate Status to be VALID. This requirement is not editable.
2
Plate Call
1
Positive controls (POS)
2
Plate Status
3
No Template Controls (NTC)
4
2.
View the Well Calls (Unknown).
a.
View the total number of wells for each call—Present, Absent, Review, or Fail.
b.
Use the Legends to view the calls in the plate
MycoSEQ™ Mycoplasma Real-Time PCR Detection Kit User Guide
19
1
4
2
3
5
Chapter 2
2
Setup, run, and anlayze samples with AccuSEQ™ Software v3.1 on the QuantStudio™ 5 Instrument
Methods
Figure 3 Example MycoSEQ™ Results (Grid View)
Results pane
1
Well call
2
Flags—the number within the triangle indicates the number of QC flag calls in the well; review in QC Summary.
3
Grid View
4
Evaluate the overall shape of the Amplification Plot curves
You can evaluate the overall shape of the Amplification curves in the Result tab. Evaluating the amplification curve in wells with a Review call can help determine if the sample should be rerun, or whether to edit the call to Present or Absent.
Ensure that the experiment is open in the AccuSEQ™ Software.
20
1.
In the Results Analysis pane of the Result tab, click Amplification in the horizontal scroll bar.
Note: If no data are displayed in the Result Analysis pane, then click Analyze.
MycoSEQ™ Mycoplasma Real-Time PCR Detection Kit User Guide
1
Chapter 2 Methods
Setup, run, and anlayze samples with AccuSEQ™ Software v3.1 on the QuantStudio™ 5 Instrument
Amplification Plot
1
The Amplification Plot is displayed for the selected wells in the (Grid View) .
2.
Ensure that the Target selected is Mycoplasma.
3.
(Optional) Click (Settings) , then make the following selections:
Plot Type: ΔRn, Rn, or C
Graph Type: Log or Linear
Show: Legend, Cq Mark (the cycle at which the curvature of the
T
amplification curve is maximal), or Unselected
Plot Color: Target, Sample, Well, or Flag_Status
Threshold: Select Auto or specify a threshold, then decide whether to Show Threshold.
Baseline: Decide whether to Show Baseline.
2
4.
(Optional) Adjust the
a.
(Optional) In the General tab, add a Plot Title, adjust the Font and Color, then click Apply.
b.
In the X Axis or Y Axis tabs, you can:
Add a Label
Select whether you want Tick Marks
Select Auto-adjust range or enter minimum and maximum values
Note: The minimum value must be greater than 0.
(Plot Properties).
MycoSEQ™ Mycoplasma Real-Time PCR Detection Kit User Guide
21
1
2
3
4
5
6
2
1
Chapter 2
2
Setup, run, and anlayze samples with AccuSEQ™ Software v3.1 on the QuantStudio™ 5 Instrument
Methods
Figure 4 Typical Amplification Plot (4 phases)
Amplification Plot tools
1
Plateau phase
2
Linear phase
3
Exponential (geometric) phase
4
Baseline
5
Threshold
6
View and evaluate the Melt Curve Plot
You can view and evaluate the the Result tab.
Melt Curve plot
1
Melt Curve Stage dropdown
2
1.
In the Result Analysis window, select Melt Curve Plot from the horizontal scroll bar.
2.
Keep the default Melt Curve Stage.
Custom experiments with multiple data collections in the Melt Curve Stage, can select the stage that they want displayed.
Melt Curve Plot in the Result Analysis window in
22
MycoSEQ™ Mycoplasma Real-Time PCR Detection Kit User Guide
1
Chapter 2 Methods
Setup, run, and anlayze samples with AccuSEQ™ Software v3.1 on the QuantStudio™ 5 Instrument
2
Stage 3 data collection point
1
3.
Click to configure the plot, then make the following selections:
Targets: Select Mycoplasma.
Plot Settings: Select the Plot Type and Plot Color.
Plot Properties: Edit Plot Title, change fonts, colors, and labels.
Save Image: Save the image (PNG or SVG).
Use the select, pan, and zoom options to interact with the plot.
The Melt Curve Plot is displayed for data points that are selected in the plot settings. The data points for selected wells in the Grid View or Table View are highlighted in the plot.
4.
(Optional) View the default Melt Curve Stage.
5.
Confirm that amplification in the no template control and IC control wells is as expected. Use one of the following options:
Select control wells in the
Grid View or Table View, then confirm the
location of the data points in the Melt Curve Plot.
View the amplification plots for the no template controls.
MycoSEQ™ Mycoplasma Real-Time PCR Detection Kit User Guide
23
1
2
Chapter 2 Methods
2
Setup, run, and anlayze samples with AccuSEQ™ Software v3.1 on the QuantStudio™ 5 Instrument
6.
In the Melt Curve Plot, view the signal intensity and calls for the unknown samples.
Figure 5 Example Melt Curve Plot
Melting temperature (Mycoplasma)
1
Melting temperature (positive control)
2
24
MycoSEQ™ Mycoplasma Real-Time PCR Detection Kit User Guide
Setup, run, and anlayze samples with AccuSEQ™ Software v3.1 on the QuantStudio™ 5 Instrument

Guidance for unknown samples

The table shows criteria for positive and negative calls. A positive call indicates that at least one genome copy of Mycoplasma DNA was present in the unknown reaction and the sample is positive for the presence of Mycoplasma.
Note: Tm and DV assay acceptance criteria are only relevant if Ct value for present acceptance criteria are met. The AccuSEQ™ Software v3.1 flags these as "Review".
Table 2 Example acceptance criteria for unknown samples: AccuSEQ™ Software v3.1 or later
Chapter 2
Methods
2
Result C
Present < 36.2300 C
Absent ≥ 36.2300 C
t
Tm (°C) DV
t
t
75.50 < Tm < 83.00 ≥0.40
< 75.50 <0.20
Figure 6 Decision tree for unknown sample calls (with or without an inhibition control [IC])
MycoSEQ™ Mycoplasma Real-Time PCR Detection Kit User Guide
25