Thermo Fisher Scientific EC-TDS-SaltTestr User Manual

This Instruction Manual is also available for download at: www.eutechinst.com or www.4oakton.com
INSTRUCTION MANUAL
EC/TDS/SALT Testr
Large Screen Waterproof Multi Range
Conductivity/TDS/Salt Tester
with Temperature Display
Introduction
Thank you for selecting microprocessor-based waterproof EC / TDS / SALT tester with large dual line display. You have one of the following models:
Non-plus modelsNon-plus models
Non-plus models (ECTestr11, TDSTestr11 & SALTTestr11) come with user-replaceable
Non-plus modelsNon-plus models two-pin type sensor and have many user friendly features such as Dual-range measurement, the Hold function, Automatic Temperature Compensation (ATC) and Self-Diagnostic Messaging capabilities.
Plus models Plus models
Plus models (ECTestr11+ & TDSTestr11+) come with the user-replaceable cup type
Plus models Plus models sensor and have additional features such as Multi-range measurement, up to 3-point calibration and higher resolution measurement.
Before You Begin
Remove the electrode’s protective cap. Soak the electrode for a few minutes in alcohol to remove any oil stains on the electrodes which will affect the accuracy of the tester. Rinse thoroughly with de-ionized water and shake off dry.
Key Functions
KK
ee
y Functiony Function
K
e
y Function
KK
ee
y Functiony Function
- Power on and off the tester (The tester automatically switches of, if no button is pressed for 8.5 seconds)
- In measurement mode, temperature reading switches between Celsius &
°F °C/
Fahrenheit
- In calibration mode, switches the tester to temperature calibration mode
- In temperature calibration mode, exits calibration mode without confirming calibrated values
- In measurement mode, switches to hold mode freezing the display.
HOLD
- In hold mode, switches back to measurement mode
- In manual calibration and temper ature calibration modes, exits calibration mode without confirming calibrated values
- In range selection mode, selects a range
- In measurement mode, enters calibration mode
- In calibration mode, adjusts the calibration values
- In hold mode, enters TDS factor setting mode
- In TDS factor setting mode, adjusts TDS factor
Note:Note:
INC INC
Note:
Note:Note: battery compar tment
Note:Note:
Note:
Note:Note: caption of HOLD key is
DEC DEC
INC &
DEC keys are located inside the
INC INC
DEC DEC
..
.
Refer figure 1.
..
For ECTestr11 & ECTestr11+ models, the
‘HOLD/ENT’‘HOLD/ENT’
‘HOLD/ENT’
‘HOLD/ENT’‘HOLD/ENT’
-1- -2-
Figure 1: Battery compartment
+
-
Switching On
ON/OFFON/OFF
Press
ON/OFF key to switch on the tester. The LCD shows the power-up sequence
ON/OFFON/OFF
as illustrated in Figure 2.
When the tester is on, if you do not press a key for 8.5
minutes, the tester automatically switches off to conserve batteries.
MEAS
0.0
uS
26.8
Measurement
mode
ATC
°C
ON/OFF
CAL
MEAS
HOLD
S
pH% R.mV
ATC
pp m
°C °F
°C °F
Model number &
software revision number
Figure 2: Power Up Sequence
20.00 Ec.p
ms
Au o
Selected range &
upper limit of the range
Range Selection
Depending on the selected model, you can set the tester to limit its reading to a particular measuring range (PU, LO or HI) or full scale (AUTO). The default setting is AUTO. When you select a range other than AUTO, the tester can be calibrated only for that particular range. conductivity/TDS value than that of the selected measuring range, the LCD shows ‘OR’ error message. Refer selected model.
TT
o select a ro select a r
T
o select a r
TT
o select a ro select a r
angang
ang
angang
e:e:
e:
e:e:
1. Switch off the tester. Press and hold ON/OFFON/OFF
using
ON/OFF key. Release
ON/OFFON/OFF
2. The tester goes to range selection mode. The LCD shows the currently selected
Range (the default is AUTO) in the lower display. The upper display shows the maximum possible reading for the selected range. Press until you see the required range (PU, LO or HI).
Note:Note:
Note: If no key is pressed within 5 seconds, LCD shows power-up sequence and
Note:Note: tester goes to measurement mode.
3. The tester automatically confirms the last selection if no key is pressed for 5
seconds. Upper display momentarily shows ‘CO’. The LCD shows power-up sequence and the tester goes to measurement mode.
°F °C/
1
Tes t er
Off
Press & hold key and switch o n
the tester
20.00
Au o
ON/OFF
°C/°F
Figure 3: Range selection sequence from AUTO to HI for ECTestr11+
If you try to measure a sample which has a higher
SpecifSpecif
icaica
tionstions
Specif
ica
tions section for available ranges of the
SpecifSpecif
icaica
tionstions
°C/°F°C/°F
°C/°F key and then switch on the tester
°C/°F°C/°F
°C/°F °C/°F
°C/°F key.
°C/°F °C/°F
2
HOLD
ENT
ms
2 2 2
HOLD
ENT
200.0
us
2000
us
PU LO
Power-up Sequence
Auto confirmation oc curs for
the last selecte d range
if no key is presse d for 5 sec.
HOLDHOLD
HOLD key repeatedly
HOLDHOLD
HOLD
ENT
20.00
ms
HI
5 seconds
CO
HI
3
Measurement
1. Press the when the tester is in measurement mode.
2. Dip the electrode into the test solution making sure that it is fully immersed. Stir to clear any trapped air bubbles from the electrode and let the reading stabilize. For plus models, you can opt for the cup style measurement by filling the electrode cup with sample of test solution.
Note:Note:
Note: The LCD indicates ‘Or’ (over range) if the reading is outside the selected
Note:Note: range. If this occurs, select an appropriate range to suit the reading.
3. The upper display shows the main reading (conductivity/TDS/Salt) of the solution, automatically temperature compensated (ATC) to normalized temperature of 25°C. The lower display shows the temperature of the solution.
ON/OFFON/OFF
ON/OFF key to switch on the tester. The ‘MEAS’ indicators appears
ON/OFFON/OFF
HOLD Function
This feature lets you freeze the display for a delayed observation.
HOLD HOLD
1. Press
HOLD key to freeze the measurement. The tester goes to hold mode and
HOLD HOLD ‘HOLD’ indicator is displayed in LCD. The measurements are frozen and the ‘MEAS’ indicator disappears.
HOLDHOLD
2. Press
HOLD key again to release the measurement. The ‘HOLD’ indicator is no
HOLDHOLD longer displayed. The tester goes back to measurement mode.
MEAS
1500
26.8
mode
ATC
HOLD HOLD
ppm
HOLD
1500
°C
26.8
HOLD modeMeasur ement
Figure 4: HOLD Function
ATC
ppm
°C
MEAS
1535
26.2
Measurement
mode
ATC
ppm
°C
Temperature Unit of Measurement Selection
This feature lets you set the unit of measurement of temperature to either Celsius (°C) or Fahrenheit (°F).
When the tester is in the measurement mode, press display toggles between the Celsius and Fahrenheit reading.
ATC
pp
t
MEAS
6.00
°C
25.5
Figure 5: Temperature unit of measurement selection
°C/°F°C/°F
°C/°F button. The temperature
°C/°F°C/°F
MEAS
ATC
pp
t
°F
°C/
6.00
°F
78.0
-3-
About Calibration
50% tolerance
To ensure higher accuracy, the tester must be calibrated on a regular basis. Calibration can be manual or automatic (only for ECTestr11 & ECTestr11+); calibration can be 1-point or multi-point. You can choose any combination of the above two options for calibration. If you calibrate the tester for 1-point, the calibration is applied for all the measuring ranges. In applications where you need higher accuracy, and when you intend to measure values in more than one range, it is recommended to select multi-point calibration.
Selection of Automatic or Manual Calibration
ECTestr11 & ECTestr11+ models support both automatic & manual calibration while all other models have to be calibrated manually. In automatic calibration, the tester automatically detects and verifies known conductivity standard solutions (84uS, 1413uS & 12.88mS) . In manual calibration, you can use non-standard solutions which may be specific for your application.
The factory default is Automatic calibration (Auto). You can enable or disable automatic calibration as described below (for only ECTestr11 & ECTestr11+ models).
Selection of 1-point or Multi-point Calibration
The factory default is 1-point Calibration. For higher accuracy, it is recommended that you calibrate the tester for multiple ranges if you intend to measure values in multiple ranges. You can enable or disable multi-point calibration as described below.
Note:Note:
Note:
If you have selected a specific measuring range for the tester, selecting multi-
Note:Note: point calibration has no meaning, as the tester can only be calibrated for 1-point for the selected range. Set the tester to ‘AUTO’ measuring range if you wish to calibrate multi-points.
TT
o enao ena
bb
le/disale/disa
bb
b
le/disa
bb
le/disale/disa
le auto calibrle auto calibr
b
le auto calibr
bb
le auto calibrle auto calibr
T
o ena
TT
o enao ena
1. Switch off the tester. Press and hold ON/OFF ON/OFF
ON/OFF key.
ON/OFF ON/OFF
2. The tester goes to auto calibration selection mode. The lower display shows ‘A.CAL’ and the upper display blinks the current choice (‘Yes’ or ‘No’).
Note:Note:
Note:
This mode is available only for ECTestr11 & ECTestr11+ models. For other
Note:Note: models, go directly to 1-point selection mode, described in step 4 below.
INC INC
Press
DEC DEC
INC or
DEC key to select ‘Yes’ (to enable auto calibration) or ‘No’ (to disable
INC INC
DEC DEC
auto calibration)
Note:Note:
Note:
Note:Note:
NoteNote
Note: Press
NoteNote
°C/°F°C/°F
Press
°C/°F key, if you wish to skip this setting without confirming changes.
°C/°F°C/°F
°C/°F°C/°F
°C/°F key twice, if you wish to return to measurement mode without
°C/°F°C/°F
confirming changes.
HOLD/ENTHOLD/ENT
3. Press
HOLD/ENT key to confirm the selection. The display shows ‘CO’.
HOLD/ENTHOLD/ENT
4. The tester goes to 1-point calibration selection mode. The lower display shows ‘1.Pnt’ and the upper display blinks the current choice (‘Yes’ or ‘No’).
INC INC
Press
DEC DEC
INC or
DEC key to select ‘Yes’ (to enable 1-point calibration) or ‘No’ (to
INC INC
DEC DEC
disable 1-point calibration, i.e. to enable multi-point calibration).
Note:Note:
Note: Press
Note:Note:
5. Press
°C/°F°C/°F
°C/°F key if you wish to skip this setting without confirming.
°C/°F°C/°F
HOLD/ENTHOLD/ENT
HOLD/ENT key to confirm the selection. The display shows ‘CO’ for few
HOLD/ENTHOLD/ENT seconds and then shows power-up sequence The tester goes to measurement mode.
aa
tion and mtion and m
a
tion and m
aa
tion and mtion and m
ulti-point calibrulti-point calibr
ulti-point calibr
ulti-point calibrulti-point calibr
INC INC
INC key and then switch on the tester using
INC INC
-4-
aa
a
aa
tion:tion:
tion:
tion:tion:
2
HOLD
1
Tes t er
Off
Press & hold IN C key and switch on
the tester
ON/OFF
YES
A.CAL
A.CAL
Figure 6: Selecting auto calibrtaion & 1-point calibration for ECTestr11+
ENT
2
HOLD
ENT
CO
A.CAL
3 4
After
1 sec.
YES
HOLD
ENT
CO
4
HOLD
Power-up Sequence
ENT
Auto Calibration
Auto calibration feature is available only for conductivity in ECTestr11 & ECTestr11+ models. Make sure that ‘auto calibration’ is enabled as described in the previous section. Auto calibration is suitable if you use standard conductivity solutions for calibration process.
If you selected 1-point calibration, you need to choose a calibration standard corresponds to the selected measuring range of the tester listed below. If you have selected multi-point calibration & AUTO measuring range, you can choose any of the calibration standards listed below. During auto calibration, the tester recognizes the calibration standard if its value is within 50% tolerance. For multi­point calibration (with AUTO measuring range), the tester automatically scans through all possible calibration points until all of them are calibrated.
Selected Measuring Range Calibration Standard
PU (0 to 200.0 uS/cm) 84 us
LO (0 to 2000 uS/cm) 1413 uS
HI (0 to 20.00 mS/cm) 12.88 mS
AUTO 84 us, 1413 uS, 12.88 mS
TT
o pro pr
ee
parpar
e calibre calibr
aa
e calibr
e calibre calibr
tion standartion standar
a
tion standar
aa
tion standartion standar
T
TT
o pr
o pro pr
e
par
ee
parpar
Use fresh calibration standard solutions listed in the above table. Prepare each solution in two beakers - one for rinsing and the other for calibration. Rinse the electrode in de-ionized water before calibration.
TT
o beo be
T
TT
o be
o beo be
gin automagin automa
gin automa
gin automagin automa
tic calibrtic calibr
tic calibr
tic calibrtic calibr
1. Switch on the tester. Make sure the tester is in measuring mode. Press DEC DEC
DEC key to enter conductivity calibration mode.
DEC DEC
2. ‘CAL’ indicator appears in LCD. The display briefly shows ‘CAL’ and the number of points the tester will be calibrated.
3. The upper display shows the conductivity reading and the lower display sequentially shows calibration standard values 1413 uS & 12.88 mS (for ECTestr11) or 84 uS, 1413 uS & 12.88 mS (for ECTestr11+) if the measuring range of the tester is set to AUTO.
aa
a
aa
tion:tion:
tion:
tion:tion:
ds:ds:
ds:
ds:ds:
INC INC
INC or
INC INC
-5- -6-
5
Note:Note:
Note:
If you have selected a specific measuring range for the tester, the lower display
Note:Note: shows the corresponding calibration standard value that matches the selected measuring range.
4. Rinse the electrode with the calibration standard that you intend to calibrate and then dip the electrode in the other beaker with same calibration standard. Swirl gently to create a homogenous sample and allow time for the reading to stabilize.
Note:Note:
Note: For multi-point calibration, the lower display automatically locks at the
Note:Note: calibration standard value that closely matches. The tolerance range is ± 50% of the calibration standard. The tester shows error message ‘Er.1’ if you try to calibrate with a solution whose conductivity is outside the tolerance range.
Note:Note:
INCINC
Note:
Note:Note:
Press
DEC DEC
INC or
DEC key if you wish to exit from auto calibration, during any of
INCINC
DEC DEC
the above steps.
HOLD/ENT HOLD/ENT
5. Press
HOLD/ENT key to confirm the calibration. LCD shows ‘CO’ for 2 seconds. The
HOLD/ENT HOLD/ENT calibration is complete and the tester returns to measurement mode, if this is a 1­ point calibration.
6. For multi-point calibration, the tester goes to the next calibration point, lower display showing next calibration standard values. Rinse the electrode in de-ionized water and repeat step 4 & 5 to continue calibrating with next calibration standard solution.
Note:Note:
Note: The tester shows error message ‘Er.0’ and returns to measuring mode if the
Note:Note: temperature of the calibration solution is not within 0°C to 50°C.
Note:Note:
Note: The tester shows error message ‘Er.1’ if you press
Note:Note:
HOLD/ENT HOLD/ENT
HOLD/ENT key before the
HOLD/ENT HOLD/ENT
tester recognizes the calibration standard.
CAL
0.0
uS
1413
CAL
0.0
uS
3
ATC
1413 uS std. solut ion
3
and automatica lly detects
ATC
the calibration s td.
CAL
1410
uS
1413
Tester dipped in
ATC
HOLD
ENT
Confirmation o f 1st point
CAL
1410
uS
CO
displayed for
2 seconds
ATC
Measurement
Mode
2 54
OR
CAL ATC
CAL
Displayed for
2 seconds and
tester goes to
calibration mo de
1
12.88
The two calibra tion stds. values
displayed sequen tially until
tester finds a matc hing solution
within 50% tolerance.±
6
CAL AT C
1410
mS
12.88
The next calibr ation std.
displayed until te ster finds
a close match wi thin
±
Figure 7: Two-point automatic calibration sequence for ECTestr11
CAL ATC
12.80
mS
12.88
Tester dipped in
12.88 mS std. solution
& automatically detects
the calibration std.
CAL ATC
HOLD
ENT
12.80
mS
CO
Confirmation of 2nd point
displayed for
2 seconds
Measurement
Mode