Batteries
included
Specications
Range .. to 19.90 mS
Resolution 0.1 mS
Accuracy ±1 % full scale (±2 % above 10 mS)
No. Of Cal Points
Calibration Method Digital, push button
Automatic
Temperature
Compensation
Non-Volatile
Memory
Hold Function Yes
Auto-O 8.5 mins after last key press
LCD Display 3½ digit single line
Power (4) A76 / LR44 batteries included
Operating Temp. 0 to 50 ºC
Ingress Protection IP67 rating ; waterproof
Dimension
(LxWxH)/Weight
1 point; calibration standard
sold separately
Yes
Yes
Tester: 16.3 x 4.5 x 3 cm ; 90 g
Packed : 24.5 x 13.5 x 4.5 cm ; 137 g
Conductivity
.. to 19.90 mS
Solutions
For Water
Analysis
Please refer to inside for
operating instructions
Eutech Instruments
www.eutechinst.com
Oakton Instruments
www.4oakton.com
Part of Thermo Fisher Scientific
1 Year Warranty
Made in Singapore
35462-35
Applications
• Water & Wastewater Treatment • Environmental
Monitoring • Education • Hydroponics • Agriculture
• Aquaculture & Aquariums • Pools & Spas • Food & Beverage
Manufacturing • Cooling Towers • Electroplating • Printing
• Photo-Development and more!
Pocket clip >
< Transparent cap
for samples
and storage
Laboratory
• All Push-Button
Operation
• Easy To Use
• Auto Calibration
Hydroponics
• Automatic
Temperature
Compensation
• Waterproof
Environmental In-Situ Testing
68X461004 09/10 Rev B
Operating Instructions
EcoTestr EC High Waterproof Pocket Tester
Conditioning Before First Use:
1. Remove the transparent cap and add 20 mm of tap water or calibration standard
(approximately ½ full)
2. Recap the EcoTestr and soak* for at least 15 minutes
3. Rinse the cap after use
4. The sensor is now activated and ready for use
Measurement:
1. Remove cap and press to turn on the tester
2. Dip sensor into the test solution
3. Stir once and let the reading stabilize. Note the measured value
4. To hold reading, press . Screen flashes HO once, then displays measurement
with blinking unit (mS) to indicate that tester is in the hold mode. Press again to
cancel hold mode ( HC)
5. Press to shut the tester off
Note: Tester automatically shuts off after 8.5 minutes of non-use to conserve batteries.
Calibration: The EcoTestr is factory calibrated. However, to ensure accuracy, calibration
is recommended. Note: Calibration standard sold separately.
1. If the sensor was stored dry, follow “Conditioning Before First Use” steps for
best results
2. Press to turn on the tester
3. Dip sensor into calibration standard solution and wait for the value to stabilize
4. Press to begin the calibration. Display shows
CAL
momentarily and blinks the
default reading
5. Press until the blinking value matches the value of your calibration standard
at 25 ºC. Note: to set a calibration standard value that is lower than the blinking
value, continue to press past the maximum value to continue with the lowest
adjustable value
6. Release to accept the calibration value. After a few seconds, (
En
) is shown and
measurement is resumed
7. To abort calibration, press to escape (
ESC
)
Maintenance: Clean the stainless steel sensors by periodically soaking them in alcohol
for 10 - 15 minutes. Rinse sensor with clean water after each use.
Useful Notes: To avoid cross contamination, rinse between samples and calibration
standards with clean water.
* The tester cap is not leak-proof. Please keep tester upright during soaking.
Step 3
Slide out
the cover
T
IF
Battery
cover
FT
LI
F
T
Step 1
Slide in
left hand
thumb to
clear the
back
catch
Figure 1:
Removing battery cover
to change batteries
L
Step 2
Slide in
right hand
thumb to
clear the
front
catch
Changing Batteries:
1. To remove batteries, hold the tester on the left hand with the left thumb slide under
the back clip of the cover. This will release the back catch
2. Next, hold the cover with your right hand, at the same time release the front catch
with your right thumb. Slide up to remove the battery cover (Figure 1)
3. Replace old batteries with fresh ones. Note polarity as shown in battery compartment
4. No re-calibration is required after battery change. The tester’s non-volatile memory
function retains calibration settings
Your EcoTestr includes (4) pre-installed 1.5V alkaline batteries. LR44 or A76 battery types are
suitable and commonly available. Replace all (4) batteries together. Waiting too long to replace
the batteries can lead to inaccurate readings and is the most common cause of problems.
Error Messages:
1.
bA
– Batteries are weak and need replacement
2. Er – Calibration was attempted with a value that is <5 % of the full-scale range;
use a calibration standard with a higher value
– Use fresh calibration standard
– Sensor is dirty or has not been soaked
3. Or – Over Range. The measurement is above the range of the instrument
(including temperature measurement)
Reset Option: Reset option allows you to restore the calibration back to factory
default settings.
1. Press to turn off tester
2. Press and hold , then press to turn the tester on. The display shows a
flashing
rS
(reset)
3. Press again to confirm reset (
En
) or press to escape (
ESC
) reset option
Warranty: This instrument is supplied with a warranty against manufacturing defects
for a period of one year from date of purchase.
Tester & Accessories Ordering Information:
Eutech Order Code Oakton Order Code Description
ECOECTESTHIGH 35462-35 EcoTestr EC High
ECPOUCH01 35624-45 Soft carrying case with belt-loop
ECCON2764BT 00653-20 2764 µS KCl calibration solution
ECCON1288BT 00606-10 12.88 mS KCl calibration solution