Thermo Fisher Scientific DO 110 User Manual

Page 1
Instruction Manual
gyM
sy...
DO 110
Hand-held Dissolved Oxygen Meter with RS 232C
68X361305
Rev. 0 - 11/05
Technol o
adeEa
Page 2
Page 3
Preface
This manual serves to explain the use of the hand-held series meters. The model covered is the DO 110 hand-held meter.
The manual functions in two ways, firstly as a step by step guide to help the user operate the meter. Secondly, it serves as a handy reference guide.
This instruction manual is written to cover as many anticipated applications of the DO meter as possible. If there are doubts in the use of the DO110 meters, do not hesitate to contact the nearest Authorised Distributor.
It is recommended that all operators should read this manual prior to working with this instrument. Eutech Instruments / Oakton Instruments cannot accept any responsibility for damage or malfunction to the meter caused by improper use of the instrument.
The information presented in this manual is subject to change without notice as improvements are made, and does not represent a commitment on the part of Eutech Instruments Pte Ltd / Oakton Instruments.
Copyright © 2005 All rights reserved.
Eutech Instruments Pte Ltd
Oakton Instruments
Rev. 0 - 11/05
Page 4
TABLE OF CONTENTS
1 INTRODUCTION........................................................................................ ............1
2 DISPLAY & KEYPAD FUNCTIONS ...................................................................... 2
2.1 DISPLAY................................................................................................................................... 2
2.2 KEYPAD ................................................................................................................................... 3
3 PREPARATION ......................................................................................... ............4
3.1 INSERTING THE BATTERIES ........................................................................................................ 4
3.2 CONNECTING THE DO PROBE.................................................................................................... 4
3.3 CONNECTING THE AC/DC ADAPTER........................................................................................... 6
3.4 CONNECTING THE RS232C CABLE ............................................................................................ 6
3.4.1 RS232C Configuration..................................................................................................... 7
4 CALIBRATION............ ................................................ ...........................................8
4.1 IMPORTANT INFORMATION ON METER CALIBRATION..................................................................... 8
4.2 TEMPERATURE CALIBRATION ..................................................................................................... 9
4.3 DO CALIBRATION (WITH ATC) –% SATURATION MODE .............................................................. 10
4.3.1 To calibrate 100% Saturation: ........................... ..... ......... .... ..... ..... ......... .... ..... .... .......... 10
4.3.2 To calibrate 0% Saturation: .................................... .... ......... ..... ..... ......... .... ..... .... .......... 10
4.4 DO CALIBRATION IN MG/L (PPM) MODE..................................................................................... 11
5 MEASUREMENT ............................................ ..................................................... 12
5.1 AUTOMATIC TEMPERATURE COMPENSATION ............................................................................. 12
5.2 MANUAL TEMPERATURE COMPENSATION (MTC)....................................................................... 12
5.3 TAKING MEASUREMENTS......................................................................................................... 13
5.3.1 Taking measurements with READY indicator selected on... ..... ..... .... ......... ..... .... ..... ..... 13
5.3.2 Taking measurements with the AUTO HOLD feature selected on................................. 13
6 MEMORY FUNCTION..........................................................................................14
6.1 MEMORY INPUT ...................................................................................................................... 14
6.2 MEMORY RECALL ................................................................................................................... 14
7 PRINT FUNCTION ...............................................................................................15
7.1 USING DO 110 METER WITH PRINTER OR COMPUTER............................................................... 15
7.2 SENDING DATA TO COMPUTER OR PRINTER ............................................................................. 16
7.2.1 Print Current Data Manually.......................................................................................... 17
7.2.2 Print Data on Timed Interval.............................................................. ............................ 17
7.3 PRINT DATA FROM STORED MEMORY ....................................................................................... 18
8 ADVANCED SETUP FUNCTIONS......................................................................20
8.1 ADVANCED SETUP MODE OVERVIEW...................................................................................... 21
8.2 P1.0 : UNIT CONFIGURATION (COF) ........................................................................................ 23
8.2.1 P1.1: READY Indicator and Auto Hold function................................. ............................ 23
8.2.2 P1.2: Select °C or °F temperature units......................................................................... 24
8.2.3 P1.3: Selection of Automatic or Manual Temperature Compensation............................ 24
8.2.4 P1.4: Select mg/L or ppm mode.................................................................................... 25
8.3 P2.0: VIEWING PREVIOUS CALIBRATION DATA (CAL) ................................................................ 26
8.4 P3.0: VIEW ELECTRODE DATA (ELE)....................................................................................... 26
8.4.1 EL.1 Viewing the electrode Slope Factor....................................................................... 26
8.4.2 EL.2 Viewing the % Saturation Offset (Only available in % Saturation Set Up menu): .. 27
8.4.3 Viewing the 100% Saturation mV value......................................................................... 27
8.4.4 Viewing the 0% Saturation mV value............................................................................. 27
8.5 P4.0 : AUTO-OFF (ATO).......................................................................................................... 28
8.6 P5.0 : RESET TO FACTORY DEFAULT (RST) .............................................................................. 29
8.6.1 Calibration Reset.. ............................................................... .......................................... 29
Page 5
8.6.2 User Reset.................................................................................................................... 30
8.7 P 6.0: MEMORY CLEAR (CLR)................................................................................................. 31
8.8 P7.0: DISSOLVED OXYGEN PARAMETERS (DPR)....................................................................... 32
8.8.1 Pressure Setting Adjustment...... ................................................................................... 32
8.8.2 Salinity Setting Adjustment (available in mg/L or ppm Set Up mode only).................... 32
8.9 P 8.0: % SATURATION OFFSET ADJUSTMENT (OFS) ................................................................. 33
9 CYBERCOMM PORTABLE DAS........................................................................34
9.1 SYSTEM REQUIREMENTS ......................................................................................................... 34
9.2 LOADING CYBERCOMM PORTABLE DAS ................................................................................... 34
9.3 RUNNING CYBERCOMM PORTABLE .......................................................................................... 41
i. Buttons & Check-Box ........................................................................................................... 42
ii. Menu............................................................................. ....................................................... 42
iii. Communication Settings............ .... ......... ..... ..... .... ......... ..... .... ..... ......... ..... .... ......... ..... ..... 43
9.4 CAPTURING AND PRINTING DATA INTO COMPUTER USING DATA ACQUISITION............................. 44
9.5 TROUBLE-SHOOTING GUIDE ..................................................................................................... 44
10 DISSOLVED OXYGEN PROBE ..................................................... ..................45
10.1 DISSOLVED OXYGEN PRINCIPLE ........................................................................................... 45
10.2 PROBE CARE ...................................................................................................................... 45
10.3 MEMBRANE HOUSING REPLACEMENT.................................................................................... 46
10.3.1 To replace the membrane cap (with pre-installed membrane) -.............. .... ..... .... ..... ..... 46
10.4 MEMBRANE/O-RING REPLACEMENT (OPTIONAL PROCEDURE)................................................. 47
10.5 ELECTROLYTE SOLUTION ..................................................................................................... 48
11 ERROR MESSAGES........................................................................................49
12 TROUBLE-SHOOTING.....................................................................................49
13 LIST OF ACCESSORIES..................................................................................51
13.1 REPLACEMENT METER AND METER ACCESSORIES ................................................................. 51
14 SPECIFICATIONS ............................................................................................52
15 WARRANTY......................................................................................................53
16 RETURN OF ITEMS......................................... ................................................. 53
Page 6
Instruction Manual CyberScan DO 110
1 INTRODUCTION
Thank you for selecting the DO 110 meter. These meters are microprocessor-based instruments and are designed to be handy, capable of allowing one-hand operation. It is capable of measuring Dissolved Oxygen (DO) in mg/L or ppm units, % saturation and temperature. This meter has many user-friendly features – all of which are completely accessible through the water-resistant membrane keypad.
Please read this manual thoroughly before operating your meter.
1
Page 7
Instruction Manual CyberScan DO 110
2 DISPLAY & KEYPAD FUNCTIONS
2.1 Display
The LCD has a primary and secondary display.
The primary display shows the measured % Saturation or mg/L (or ppm) values.
The secondary display shows the measured temperature.
The display also shows error messages, keypad functions and program functions.
Primary Display
123
4
SETUP
17
16
15
14
1. SETUP - Setup mode
indicator
2. MEAS - Measurement mode
indicator
3. CAL - Calibration indicator
4. MEM - Memory recall mode
indicator
READY HOLD ON OFF
MEAS CAL
-1.8.8.8
MEM ERR
-1.8.8.8
Secondar y D isplay
Figure 1: Active LCD screen
7. mg/l- milligram per liter
8. °C°F - Temperature
indicator
9. ATC - Automatic
Temperature Compensation indicator
10. ERR - Error indicator 16. HOLD – Hold indicator
MEM
%
ppm
mg/l
°C °F
ATC
910111213
5
6
7
8
13.
- Electrode indicator
14. indicator
15. ON – READY/Auto HOLD set
up enable indicator.
OFF – READY set up disable indicator
- Calibration standard
5. % - Percentage Saturation
of Oxygen
6. ppm – parts per million
11. MEM - Memory location
indicator
- Low battery indicator
12
2
17. READY – Ready indicator
Page 8
Instruction Manual CyberScan DO 110
2.2 Keypad
A large membrane keypad makes the instrument easy to use. Some buttons have several functions depending on its mode of operation.
KEY
ON
OFF
CAL
MEAS
HOLD
ENTER
MI
MR
SETUP
MODE
FUNCTION
ON/OFF - Powers on and shuts off the meter. The meter will start in a measurement mode similar to the last measurement mode it is in before being switched off.
CAL - Activates the DO % Saturation or mg/l calibration mode and when used with the MODE key, it activates the temperature calibration mode.
MEAS - Allows return to measurement mode when canceling or terminating any operation.
HOLD - Activates/Deactivates freezing of the measured reading while in measurement mode.
ENTER - Confirms the calibration values in Calibration mode and the selection in SETUP and Print Function mode.
Scroll through the memory location and the stored data during memory recall.
MI (Memory Input) - Captures the measured readings with its corresponding temperature values and stores them in the memory.
MR (Memory Recall) - Retrieves the stored data from memory.
▲▼ - Sets the calibration values during mg/l or temperature calibration.
Scrolls through each SETUP and its sub group menu.
Scroll through Print Function menu and its parameters.
SETUP - Activates the parameter setting menu to allow you to customise meter configuration, view calibration point and electrode data, select auto power off, reset meter, clear memory, set % Saturation Offset and set dissolved oxygen parameters. .
MODE - Selects the measurement parameter option between % Saturation with temperature and mg/L (or ppm) with temperature.
PRINT - Allows print of current measurement or stored data to either the printer or the computer.
3
Page 9
Instruction Manual CyberScan DO 110
3 PREPARATION
3.1 Inserting the Batteries
This meter is packaged with 4 “AAA” alkaline batteries required for operation. To insert the batteries into the meter, follow the procedure outlined below.
1. To open the battery compartment, press down the catch of the battery cover. See below.
2. Note the polarity and insert the batteries into the battery compartment correctly
3. Replace the battery cover into its original position, ensuring the catch is locked in position.
Your hand-held meter is now ready for operation.
LR03 'AAA ' (AM4)
MADE IN S'PORE
Figure 3: Note battery's polarity
Figure 2: Open Battery Cover
3.2 Connecting the DO Probe
The DO 110 uses a special notched 6-pin connector to attach the probe to the meter.
4
Page 10
Instruction Manual CyberScan DO 110
NOTE: Do not substitute other probes or electrodes. For replacement probe, see the “Accessories” section.
Keep connector dry and clean. Do not touch connector with soiled hands.
1. Line up the notch and 6 pins on the meter with the holes in the 6 pin connector. Push down and screw the metal sleeve to lock the probe connector into place.
2. To remove probe, unscrew the metal sleeve and slide up the probe connector. While holding onto metal sleeve, pull probe away from the meter.
CAUTION: Do not pull on the probe cord or the probe wires might disconnect.
5
Page 11
Instruction Manual CyberScan DO 110
3.3 Connecting the AC/DC Adapter
Besides using four AAA-sized batteries as a power source, the DO 110 meter can also operate from the power mains using an AC/DC power adapter. This is extremely useful if you have an A.C. power source available (e.g. laboratory).
Before plugging in, switch off the meter and the power source of adapter. This is a safety precaution that should be adhered to safeguard your meter.
1. Switch off the meter and power sources.
2. Select the correct AC/DC Adapter which matches your input mains voltage.
3. Select the correct output voltage of the AC/DC adapter. (Output Voltage: 9 to 15 V DC, Current: >=50 mA).
4. Gently insert the power adapter D.C. jack into the meter power socket.
5. Switch on the power source of the adapter followed by the meter.
3.4 Connecting the RS232C Cable
The DO 110 meter provides a RS232C output for you to transmit your readings either to a printer or a computer via a cable. This is useful in instances where the meter is used for continuous monitoring of a certain process or experiment. Data output to the printer or the computer can then be evaluated.
The data is output in the ASCII format. This format allows the data to be imported by a wide variety of software that read ASCII data (e.g. Microsoft’s Excel, Lotus, Quattro-pro etc.). A complimentary Data Acquisition Software (DAS) is provided and it captures data transmitted into an ASCII file for later use.
Figure 4: Location of
RS232C
1. Open the printer port cover located at the bottom end of the meter. Do not use excessive force when doing this. See Figure 4.
2. Noting the orientation of the RS232C connector, plug the male connector into the RS232C port of the meter.
3. Fasten the RS232C connector by fastening the two screws at the side of the male RS232C connector.
6
Page 12
Instruction Manual CyberScan DO 110
3.4.1 RS232C Configuration
The DO 110 meter has a 9 pin female RS232C connector with the following pin out:
PIN NO. DESCRIPTION
1 -
2 Transmit Data
3 -
4 -
5 CTS (Clear to Send)
6 -
7 GND (Ground)
8 -
9 -
A one-to-one connection can be made with a 9 pin RS232C port of the computer.
5
4
3
8
9
7
1
2
6
In case DO 110 meter’s output has to be sent to a 25 pin RS232C connector, the following cable configuration may be used:
DO 110 25 pin connector
2 (TxD) (RxD) 3
5 (CTS) (RTS) 4
7 (GND) (GND) 7
7
Page 13
Instruction Manual CyberScan DO 110
4 CALIBRATION
4.1 Important Information on Meter Calibration
The amount of oxygen dissolved in water will depend on its temperature, atmospheric pressure and its salinity. While the pressure and salinity values are manually entered into the instrument, the temperature is being measured by the probe. It is therefore very important that the temperature is calibrated if necessary prior to the DO calibration.
The measurements of % Saturation of DO will linearly affect the measurement for DO in mg/L. Hence calibration in % Saturation of DO should be carried out first. This is described in the following section.
Before calibrating, press MODE key to select the correct measurement mode. There are 2 measurement modes for DO: mg/L or ppm, and % Saturation.
NOTE: All new calibrations will automatically over-ride existing calibration values.
8
Page 14
Instruction Manual CyberScan DO 110
A
4.2 Temperature Calibration
DO in mg/L is dependent on temperature, so it is first necessary to calibrate or verify the temperature reading. The temperature sensor is part of the DO probe and the “ATC” annunciator will light up on the LCD screen once the probe is connected correctly to the meter.
1. Switch the meter on. The ATC annunciator will appear at the right-hand side of the LCD.
2. Press the CAL/MEAS key to enter calibration mode.
3. While in calibration mode, press the MODE key to enter temperature calibration mode. The primary display shows the measured temperature reading according to the last set offset
1
and the secondary display shows the
factory calibrated temperature measurement.
4. Dip the DO probe into a solution of known temperature (i.e. a temperature bath). Allow time for the temperature probe to stabilise.
5. Scroll with the MI/ or MR/ keys to set the correct temperature value (i.e. the temperature of the temperature bath). You can adjust the reading in increments of 0.1°C.
6. Once you have selected the correct temperature, press the HOLD/ENTER key. The meter automatically returns to the measurement mode.
NOTES: You can offset the temperature reading up to ±5°C from original reading.
To exit this program without confirming the temperature calibration value, DO NOT press ENTER. Press CAL/MEAS instead.
READY
MEAS
98.9
CAL
MEAS
CAL
83.4
100.0
MODE
22.3
22.0
HOLD
ENTER
MEAS
98.9
22.3
CAL
22.3
CAL
22.3
22.0
%
°C ATC
%
°C
TC
MIMR
°C
ATC
%
°C ATC
1
It displays the value being measured currently, offset by the last calibration.
9
Page 15
Instruction Manual CyberScan DO 110
4.3 DO Calibration (with ATC) –% Saturation Mode
Dissolved Oxygen exact calibration value depends on barometric pressure. The meter is set to a factory default of 760 mm Hg, which results in a calibration value of 100% saturation in air.
NOTE: If the barometric pressure setting has been changed from 760 mm Hg, the calibration value in air will automatically adjust to a value other than 100%. The adjusted value will be correct for the new barometric pressure setting. See Section
8.8.1 for barometric pressure setting.
4.3.1 To calibrate 100% Saturation:
1. Press MODE key to select % Saturation mode.
READY
READY
MEAS
83.4
CAL
MEAS
CAL
25.3
%
°C ATC
83.4
2. Rinse the probe well with deionised rinse water. For
best accuracy, blot the end of the probe dry. Do not touch the membrane.
3. Hold the probe in the air gently with the sensor facing
down and press CAL key to calibrate the meter. The primary display will show the current value of measurement and the secondary display will show “100.0” (or the value corrected to its barometric pressure setting) to which the meter is going to be calibrated. Wait for the reading to stabilize.
READY
100.0
HOLD
ENTER
MEAS
100.0
25.3
%
°C ATC
%
4. Press ENTER key to confirm the calibration. The meter automatically
calibrates to 100% air saturation and returns to the measurement mode.
MEAS
4.3.2 To calibrate 0% Saturation:
1. Press MODE key to select % Saturation mode
2. Rinse the probe well with deionised rinse water.
3. Immerse the DO probe in 0% solution. Stir gently to
create a homogenous solution. Wait for the reading to stabilize.
4. Press CAL key to calibrate the meter. The primary
display will show the current value of measurement and the secondary display will show “0.0” to which the meter is going to be calibrated.
5. Press ENTER key to confirm the calibration. The meter
READY
READY
READY
3.4
CAL
MEAS
HOLD
ENTER
MEAS
25.3
CAL
3.4
0.0
0.0
automatically calibrates to 0% saturation and returns to the measurement mode.
NOTE: To exit from DO Calibration mode without confirming calibration, DO NOT press ENTER. Press CAL/MEAS instead
25.3
%
°C ATC
%
%
°C ATC
10
Page 16
Instruction Manual CyberScan DO 110
A
4.4 DO Calibration in mg/L (ppm) Mode
MEAS
The amount of oxygen dissolved in a liquid will depend on its temperature, pressure and salinity. It is therefore very important to set these parameters correctly before attempting to do a
READY
8.10
24.5
calibration.
CAL
READY
MEAS
CAL
Temperature is measured by the meter automatically. Pressure and salinity values are keyed in manually as described in Sections 8.8.1 and 8.8.2.
8.10
1. From the measurement mode, press MODE key to select mg/L (ppm).
8.10
mg/l °C ATC
mg/l
°C
ATC
2. Rinse the probe well with deionised rinse water. For best accuracy, blot the end of the probe dry. Do not touch the membrane.
3. Dip the probe into a solution whose DO value is known.
4. Wait until the reading stabilizes. Press CAL key.
5. Adjust to the required value by pressing MI/S or MR/T key.
6. Press ENTER to confirm the calibration.
NOTE: To exit from DO Calibration mode without confirming calibration, press CAL/MEAS key.
READY
READY
MR
8.24
HOLD
ENTER
MEAS
8.24
CAL
8.10
24.5
MI
mg/l °C ATC
mg/l °C
TC
11
Page 17
Instruction Manual CyberScan DO 110
5 MEASUREMENT
During measurement, care must be taken not to allow the membrane of the DO probe touch any surface. The probe can either be fully or partially immersed in the solution.
The READY indicator appears on the display when the readings stabilize. It will turn off if the readings start to fluctuate.
NOTE: It is important that the sample is stirred constantly to allow it to flow past the membrane for better readings.
5.1 Automatic Temperature Compensation
MEAS
%
98.9
The ATC indicator on the LCD indicates automatic temperature compensation (ATC).
5.2 Manual Temperature Compensation (MTC)
24.5
Figure 5: ATC indicator
°C ATC
For manual temperature compensation (MTC), you must set
CAL
the ATC mode in the SETUP menu Program P 1.3 to “NO”. See section 8.2.3. The [ATC] annunciator will disappear from
25.0
the LCD indicating manual temperature compensation mode.
25.0
1. Press the CAL/MEAS key to enter DO calibration mode.
2. While in DO calibration mode, press the MODE key to enter temperature calibration mode. The primary display act as the adjustable temperature setting and the secondary display shows the default temperature value of 25°C or indicates the last set temperature setting.
ENTER
3. Check the temperature of your sample using an accurate thermometer.
READY
4. Press the MI/ or MR/ keys to set the temperature to the measured value from step 3.
5. Press HOLD/ENTER key to confirm the selected temperature and to return to the measurement mode.
CAL
30.0
25.0
HOLD
MEAS
98.9
30.0
°C
MIMR
°C
%
°C
6. The meter will now compensate DO readings for the manually set temperature.
NOTES: To exit this program without confirming the manual temperature
compensation value, DO NOT press ENTER in step 5. Press CAL/MEAS instead.
12
Page 18
Instruction Manual CyberScan DO 110
5.3 Taking Measurements
1. Rinse the probe with deionised or distilled water before use to remove any
impurities adhering to the probe body. If it is dehydrated, soak it for 30 minutes in tap water.
2. Switch on the meter. The MEAS annunciator appears on the top of the
LCD. The ATC indicator appears in the lower right hand corner to indicate Automatic Temperature Compensation.
3. Dip the probe into the sample and allow some time for the reading to
stabilize
NOTE: When dipping the probe into the sample, make sure the tip of the probe is completely immersed. Stir the sample gently to create a homogenous sample. Be sure to tap probe very gently to remove air bubbles. Air bubbles will cause errors in the reading.
4. .When the reading is stable, the READY annunciator appears. Note the
reading on the display.
5. Press the HOLD key to freeze the value of the DO
reading for a delayed observation. Press the HOLD key again to release the held values.
6. To toggle between % Saturation and mg/l (ppm)
measurement mode, press the MODE key.
NOTE: This meter shuts off automatically after 20 minutes of last key press. If the meter is shut off either automatically or manually, the HOLD value will be lost. For longer storage, use the memory functions.
READY
MEAS
98.9
MODE
MEAS
8.24
24.5
24.5
%
°C ATC
mg/l °C ATC
5.3.1 Taking measurements with READY indicator selected on
If the READY indicator has been activated, the READY annunciator lights when the reading is stable. You can switch the READY indicator on or off in SETUP program. See 8.2.1 - P1.1: READY Indicator and Auto Hold function on page 23.
5.3.2 Taking measurements with the AUTO HOLD feature selected on
When a reading is stable for more than 5 seconds, the AUTO HOLD feature will automatically “HOLD” the reading. The “HOLD” indicator appears on the left side of the display. Press the HOLD/ENTER key to release the reading. Switch the Auto Hold feature on or off in SETUP program in section 8.2.1 - P1.1: READY Indicator and Auto Hold function on page 23.
13
Page 19
Instruction Manual CyberScan DO 110
g
y
p
6 MEMORY FUNCTION
The DO 110 meter can store up to 100 sets of data. You can store data in any combination of values such as “% Saturation and temperature” or “mg/L (or ppm) and temperature”.
6.1 Memory Input
MEAS
%
98.9
1. During any measurement function (MEAS), press MI/
key to input any data into the memory.
MI
2. MEM, “Sto” and memory number will flash. The meter
then returns to measurement mode.
3. If the memory is full, the display will flash “MEM – FUL”
MEM
momentarily (Figure 7). The first value stored will be erased to create space for the new value.
MEM
Figure 6: Data
stora
e in memor
24.5
MEM
1
°C ATC
FUL
6.2 Memory Recall
This function recalls the previous readings stored in the memory. Memory recall is in “Last In First Out” order. To recall readings:
MEM
Figure 7: Memory full
1. Press the MR/ key once to retrieve the last reading stored. The memory location screen – MEM, “Loc” and the memory number – will flash on the display.
2. Press the HOLD/ENTER key to recall the reading stored under that memory number.
loc
3. To display next stored reading, press HOLD/ENTER key. The display automatically moves to the next lower memory location screen.
4. To view a stored reading in a specific memory location, press
loc
MI/ or MR/ keys to locate it, and press HOLD/ENTER key to display the reading.
NOTE: Readings stored in memory are retained even if the unit is turned off. To erase all readings stored in memory, use the SETUP in section 8.7 – P 6.0: Memory Clear (CLr) on page 31.
14
Figure 8: Scroll up or down
ecific memory location
for s
MEM
26
MIMR
MEM
25
Page 20
Instruction Manual CyberScan DO 110
7 PRINT FUNCTION
The DO 110 meter is equipped with an RS232 communication port and a PRINT
) key that facilitate the printing of data from the meter to a printer or computer.
(
7.1 Using DO 110 Meter With Printer or Computer
1. To use the DO 110 directly with a printer, the printer should have either a 9
pin or a 25 pin RS 232C serial port.
2. Communication settings of the printer or computer’s data acquisition
software (CyberComm Portable DAS) MUST match the following parameters:
a. Baud Rate: 9600
b. Parity Bit: None
c. Stop Bit: One
d. Data Bit: 8
e. Flow Control: None
3. Please refer to your printer’s or computer’s instruction manual for more
information on its communication setup protocol.
4. Use 9-pin cable (Refer to section 13 - LIST OF ACCESSORIES on page
51) to connect the meter to the printer. If the printer has a 25 pin connector, use the 9 to 25 pin cable or make your own cable taking note of the connection parameters as described in section 3.4.1 - RS232C Configuration on page 7.
5. Turn on DO 110 meter and note the following:
a. For printer, ensure it is powered on and paper is loaded.
b. For computer, ensure that Eutech Instruments’ CyberComm
Portable Data Acquisition Software is installed and activated. See section 9.3 - on page 41.
6. To send data to the printer or computer, press the PRINT key.
15
Page 21
Instruction Manual CyberScan DO 110
7.2 Sending Data To Computer or Printer
When the PRINT key is pressed, there are two options:
dAt / CUr: Print Data on Timed Interval basis. You can print any data automatically based a pre-set timing from 5 to 30 seconds.
dAt / StO: Print Stored Data from Memory. This simply dumps all data stored in memory. You cannot print data from selected memory locations.
See Figure below.
From measurement mode.
HOLD ENTER
MR
MI
HOLD
ENTER
Figure 9: Print data based on timed interval or from memory
16
Page 22
Instruction Manual CyberScan DO 110
y
7.2.1 Print Current Data Manually
MEAS
READY
To manually print any data that is currently being measured:
1. From measurement mode, press HOLD key to freeze
HOLD
8.10
24.5
mg/l °C ATC
reading. See Figure 10.
2. Print PRINT (
) key.
3. The “Prn – CUr” annunciator will appear and flash for about 0.5 seconds. The data is transmitted to the printer or computer.
Flash for 0.5 seconds
4. The meter returns to measurement mode.
MEAS
5. Repeat steps 1 to 3 for next measurement.
7.2.2 Print Data on Timed Interval
READY
8.10
24.5
Figure 10: Print current data
manuall
mg/l °C ATC
This feature is useful to automate printing and downloading ‘live’ measurement values into computer. The time interval setting can range from 5 to 30 seconds.
1. From the measurement mode, press PRINT key.
2. Select “Dat/Cur” screen and press HOLD/ENTER key.
3. The “Sec 5” screen appears. The default value is 5 seconds time interval.
4. Press MI/ or MR/ key to set a value suitable for your application.
5. Press HOLD/ENTER key.
6. The meter goes into measurement mode with MEM annunciator flashing.
7. At a pre-set time interval, the screen flashes “Prn/CUr” momentarily (printing data) and switches back to measurement mode with MEM annunciator flashing.
8. To stop printing, press PRINT key. The MEM annunciator disappears and the meter is on measurement mode.
HOLD
ENTER
sec
5
MR
MI
sec
10
Figure 11: Print data on timed
interval
17
Page 23
Instruction Manual CyberScan DO 110
7.3 Print Data from Stored Memory
The DO 110 meter can print data that is stored in the memory. You can print out all stored data or selected data based on its memory location.
To Print All Stored Data
1. From measurement mode, press PRINT key.
2. Use MI/ or MR/ key, select “dAt – StO” option and press HOLD/ENTER key. See Figure 9.
3. At the “SEC – 0” option, use MI/ or MR/ key to select number of seconds to print each set of stored data. This setting gives flexibility to match the meter to a printer’s or computer’s ability (or buffer size) to accept data and prevent overflow.
4. Press HOLD/ENTER key. The “Prn” primary display will flash until all data is printed.
5. To stop printing, press PRINT key and exit to measurement mode. See Figure
12.
NOTE: All stored memory will remain in the meter after printing.
From measurement mode
MEAS
READY
HOLD
ENTER
sec
MR
sec
0
MI
HOLD
ENTER
Set timed interval to download stored data
10
to printer or computer
Figure 12: Print all stored data
8.10
24.5
Flash until all stored data are downloaded. Press PRINT key to exit to measurement mode
mg/l °C ATC
18
Page 24
Instruction Manual CyberScan DO 110
A
A
To Print Selected Stored Data:
1. From measurement mode, press MR/ key.
2. “MEM” indicator appears and the last memory input location is displayed.
3. Press MI/ or MR/ key to select a specific memory location, and press HOLD/ENTER key.
4. When ready, press PRINT key.
5. The screen flashes “Prn/CUr” for about 0.5 seconds and switches back to the stored data of the specific memory location.
6. Press HOLD/ENTER key. The meter is now ready to recall the next stored data in the following memory location. See Figure 13.
During measurement mode
MEAS
READY
8.10
24.5
mg/l
°C
TC
Repeat print procedure or
press CAL to exit
MR
MEM
loc
9
Moves to next memory locati on
MEM
5.69
22.4
mg/l
°C
TC
loc
20
MR
loc
10
HOLD
ENTER
5.69
22.4
MEM
Last memory input location
MI
MEM
For example,select memory locati on 10
MEM
mg/l
°C ATC
HOLD
ENTER
Flash for 0.5 seconds
Figure 13: Print based on selected stored data from memory
For more information about Eutech Instruments’ CyberComm Portable Data Acquisition Software programme, see section 9 - CYBERCOMM PORTABLE DAS on page 34.
19
Page 25
Instruction Manual CyberScan DO 110
8 ADVANCED SETUP FUNCTIONS
The advanced setup mode lets you customize your meter to suit your preference. The DO 110 meters feature different sub groups that organize all setup parameters.
The sub-groups are:
For %Saturation For mg/L
P1.0 Unit Configuration P1.0 Unit Configuration
P2.0 Calibration Points P2.0 Calibration Points
P3.0 Electrode Status P3.0 Electrode Status
P4.0 Auto Power Mode P4.0 Auto Power Mode
P5.0 Reset P5.0 Reset
P6.0 Memory Clear P6.0 Memory Clear
P7.0 % Saturation Offset Adjustment P7.0 % Saturation Offset Adjustment
P8.0 Dissolved Oxygen Parameters P8.0 Dissolved Oxygen Parameters
with Salinity Setup
20
Page 26
Instruction Manual CyberScan DO 110
8.1 Advanced SETUP Mode Overview
Press the SETUP key to enter Setup mode. Press the MI/ and MR/ keys to scroll through the sub groups.
SETUP
SETUP
SETUP
SETUP
COF
P 1.0
CAL
P 2.0
ELE
P 3.0
A
0
P1.0: Unit Configuration
P1.1 Ready indicator (On, Off or Auto Hold)
P1.2 Select °C or °F temperature units
P1.3 Select Automatic Temperature Compensation (ATC)
pH
P1.4 Select mg/L or ppm mode (from mg/L or ppm mode
only)
P2.0: View Previous pH Calibration Data (View only)
P2.1 % Saturation calibration point ( from % Saturation
setup) or mg/L(ppm) calibration point (from mg/L or ppm setup only)
P3.0: View Electrode Data (View only)
In % Saturation mode:
EL.1 Electrode Slope Factor
EL.2 % Saturation Offset
EL.3 100% Saturation mV value
EL.4 0% Saturation mV value
In mg/L (or ppm) mode:
EL.1 Electrode Slope Factor
EL.2 100% Saturation mV value
EL.3 0% Saturation mV value
P4.0: Select Auto Power On/Off
P4.0 Select auto power on or off
SETUP
SETUP
P 4.0
P 5.0
MEM
P 6.0
P5.0: Reset to Factory Default
P5.1 Reset calibrated data
P5.2 User reset back to factory default values
P6.0: Memory Clear
P6.0 Clear all stored readings
21
Page 27
Instruction Manual CyberScan DO 110
SETUP
SETUP
p 7.0
OFS
p 8.0
P7.0: Dissolved Oxygen Parameter
P7.0 Select Barometric Pressure unit
Set Barometric Pressure value
Set Salinity value (from mg/L or ppm mode only)
P8.0: % Saturation Offset Adjustment
(only in % Saturation Set up mode)
P8.0 Set offset adjustment
22
Page 28
Instruction Manual CyberScan DO 110
8.2 P1.0 : Unit Configuration (COF)
This subgroup program allows customising the meter to your specific needs. In Program 1.0, features like the ‘READY’ Indicator, Auto Hold function, °C or °F mode selection, ATC or MTC mode selection and DO mg/L or ppm mode selection are available for customization.
8.2.1 P1.1: READY Indicator and Auto Hold function
Program P1.1 lets you select “READY indicator on” to indicate when the reading is stable, or select “READY indicator off” for faster meter response.
In the DO 110 meter, program P1.1 also lets you switch the auto hold function on or off. Select auto hold on to “HOLD” the reading when it is stable for more than 5 seconds. The display automatically freezes, and the HOLD indicator appears on the left side of the display. Press the HOLD/ENTER key to release the display and access other functions. Select auto endpoint off to deactivate this feature.
From measurement mode
1. Press SETUP key to enter Set Up mode.
2. At COF – P1.0 parameter, press HOLD/ENTER
key to select parameter P1.1.
3. Press the MI/ or MR/ keys to select the
configuration you require.
OFF switches the READY indicatory off.
ON switches the READY indicator on.
ON and HOLD together switches the auto
hold feature on.
4. Press the HOLD/ENTER key to confirm
selection and to proceed to step 2 of P1.2. Press the CAL/MEAS key to return to measurement mode.
SETUP
READY
ON
SETUP
READY
OFF
SETUP
READY HOLD ON
MR
MR
p 1.1
MI
p 1.1
MI
p 1.1
NOTE: Meter default is set for READY indicator on and auto
hold function off.
Figure 14: P1.1 - Configure
READY ON, OFF or HOLD
23
Page 29
Instruction Manual CyberScan DO 110
p
8.2.2 P1.2: Select °C or °F temperature units
The DO 110 meter lets you select between °C and °F units for temperature readings.
From measurement mode
1. Press SETUP key to enter Set Up mode.
2. Press the MI/ or MR/ keys to scroll through
SETUP
MR
C
p 1.2
MI
subgroups until you view parameter P1.0.
3. Press the HOLD/ENTER key two times to select parameter 1.2.
SETUP
f
4. Press the MI/ or MR/ keys to toggle between °C and °F.
5. Press the HOLD/ENTER key to confirm selection and to return to the subgroup
Figure 15: P1.2 - Select
tem
erature units, °C or °F
p 1.2
menu. Press the CAL/MEAS key to return to measurement mode.
8.2.3 P1.3: Selection of Automatic or Manual Temperature Compensation
°C
°F
This feature lets you select between Automatic Temperature Compensation (ATC) and Manual Temperature Compensation (MTC). Meter default is ATC.
From measurement mode
1 Press SETUP key to enter Set Up menu.
2 Press the HOLD/ENTER keys until you view
parameter P1.3.
3 Press the HOLD/ENTER key again. The upper
display shows “ATC” and the lower display shows “YES” or “NO”.
4 Press MI/ or MR/ keys to select between
Automatic and Manual Temperature Compensation.
“YES” activates the Automatic Temperature
Compensation function.
“NO” activates the Manual Temperature
Compensation function.
SETUP
SETUP
SETUP
HOLD
ENTER
MI
P 1.3
YES
°C
ATC
°C
ATC
MR
°C
5 Press the HOLD/ENTER key to confirm
selection and to return to the subgroup menu.
Figure 16: P1.3-Select Automatic or Manual Temperature Compensation
6 Press the CAL/MEAS key to return to
24
Page 30
Instruction Manual CyberScan DO 110
measurement mode.
8.2.4 P1.4: Select mg/L or ppm mode
(Only in mg/L (ppm) Set Up mode)
From the mg/L (ppm) Set Up mode,
1. Press the HOLD/ENTER key four times to enter the
mg/L or (ppm) mode selection menu “dO”.
2. Use the ▲ and ▼ keys to select the desired mode
of measurement.
3. Press the HOLD/ENTER key to confirm the
selection made and to return to the subgroup menu.
4. Press the CAL key to return to the measurement
mode.
SETUP
mg/l
p 1.4
MR
SETUP
MI
ppm
p 1.4
Figure 17: P1.4- Select mg/L or ppm mode
25
Page 31
Instruction Manual CyberScan DO 110
p
p
8.3 P2.0: Viewing Previous Calibration Data (CAL)
This mode lets you recall previous calibration data, which helps you know when to re-calibrate your meter. This is a “view only” mode.
From measurement mode:
1. Press the SETUP key to enter Set up mode.
2. Press the MI/ or MR/ keys to scroll through
subgroups until you view parameter P2.0.
3. Press the HOLD/ENTER key to view previous
calibration data.
4. Press the HOLD/ENTER key to return to
subgroup menu.
8.4 P3.0: View Electrode Data (ELE)
This menu features data of the electrode properties for diagnostic purposes. The “view only” parameters such as the electrode Slope Factor (k), % Saturation Offset setting, 100% Saturation mV value and 0% Saturation mV value are very useful in determining the life efficiency of the electrode. These electrode properties can be examined through its data in % Saturation and mg/L (ppm) which is available in their Set Up menu respectively.
SETUP
cal
p 2.0
HOLD
ENTER
SETUP
CAL
100.0
p 2.1
Figure 18: P2.0 - View revious calibration data
%
8.4.1 EL.1 Viewing the electrode Slope Factor
This parameter lets you view and gives an indication of the probe’s efficiency. The value displayed is the ratio of the theoretical value and the actual value produced by the probe. The higher the number, the lower the mV output from the probe. The ratio displays from 0.5 to
1.999.
From the % Saturation menu,
1. Press the SETUP key to enter Set up mode.
2. Press the MI/ or MR/ keys to scroll through
subgroups until you view parameter P3.0.
3. Press the HOLD/ENTER key to enter the menu
and view the Slope Factor (K) of the electrode.
SETUP
ele
p 3.0
HOLD
ENTER
SETUP
K =
1.231
EL.1
Figure 19: P3.0 - View
Electrode Slo
e Factor
26
Page 32
Instruction Manual CyberScan DO 110
8.4.2 EL.2 Viewing the % Saturation Offset (Only available in % Saturation
Set Up menu):
This parameter shows you the amount of the % Saturation Offset entered in the other parameter “OFS” P7.0.
From Step 3 above,
4. Press the HOLD/ENTER key to enter the %
Saturation Offset viewing menu. The meter will display the last offset adjustment made.
8.4.3 Viewing the 10 0% S aturation mV val u e
This parameter shows the sensor’s mV output corresponding to 100% Saturation.
From Step 4 above,
5. Press the HOLD/ENTER key. The display will show
the sensor’s mV output value with respect to 100% Saturation.
8.4.4 Viewing the 0% Saturation mV value
This parameter lets you view the sensor’s mV output corresponding to 0% Saturation.
SETUP
SETUP
SETUP
HOLD
ENTER
HOLD
ENTER
HOLD
ENTER
%
10.0
EL.2
mV
30.0
EL.3
mV
From Step 5 above,
6. Press the HOLD/ENTER key. The display will show
the sensor’s mV output value with respect to 0% Saturation.
7. Press the HOLD/ENTER key again to exit to the
subgroup menu. Press CAL to return to the measurement mode.
Note: To view the electrode properties in the mg/L (ppm), repeat the whole steps above using the mg/L (ppm) Set Up menu. Viewing of the % Saturation Offset will be skipped in this menu and the viewing sequence will as follows:
EL.1 Electrode Slope Factor EL.2 100% Saturation mV value EL.3 0% Saturation mV value
0.3
EL.4
Figure 20: P3.0 - View % Saturation offset, electrode mV value in 100% and 0% saturation
27
Page 33
Instruction Manual CyberScan DO 110
8.5 P4.0 : Auto-Off (AtO)
In Program P4.0 you can activate the AUTO-OFF option to automatically power off the instrument 20 minutes after the last key selection. This feature is useful for conserving battery power. Under default conditions, the instrument has this option activated.
SETUP
1. Press the SETUP key to enter Set Up mode.
2. Press the MI/ or MR/ keys to scroll through
subgroups until you view parameter P4.0.
P 4.0
3. Press the HOLD/ENTER key.
4. Press MI/ or MR/ keys to select options
5. ‘YES” - Meter will automatically switch off 20
minutes after the last key operation.
6. ‘NO’ - Auto Off mode is switched off.
7. Press the HOLD/ENTER key to confirm and
return to P4.0.
8. At any point, you can press the CAL/MEAS key to
return to measurement mode.
HOLD
ENTER
SETUP
yes
MR
SETUP
Figure 21: P4.0 - Auto shut off
feature
MI
28
Page 34
Instruction Manual CyberScan DO 110
8.6 P5.0 : Reset to Factory Default (rSt)
This mode lets you reset all parameters to factory default settings. There are two levels of reset:
Calibration Reset: Reset only the calibration
values of % Saturation and mg/L (or ppm). This clears all previous calibrated values and reset to factory default.
User Reset: Clears all data including
calibration, memory and other customised setup functions and reset it to factory default.
8.6.1 Calibration Reset
From measurement mode:
1. Press SETUP key to enter Set Up mode.
2. Press the MI/ or MR/ keys to scroll through
subgroups until you view parameter P5.0.
3. Press the HOLD/ENTER key.
4. Press the MI/ or MR/▼ keys to toggle
between NO and YES.
SETUP
SETUP
SETUP
HOLD
ENTER
CAL
HOLD
ENTER
CAL
P 5.0
P 5.1
‘NO’ retains current settings
‘YES’ erase calibration settings.
5. Press the HOLD/ENTER key to confirm
selection.
If ‘NO’ is selected, it will retains current settings and proceed to next program P5.2 – User Reset.
If ‘YES’ is selected, all LCD segments will lit up for one second, erase calibration values, and the meter switches to the measurement mode.
6. Otherwise press CAL/MEAS key to return to measurement mode without resetting to factory default.
MR
SETUP
MI
CAL
yes
Select either "YES" or "nO", press ENTER key to confirm.
Figure 22: P5.1 - Calibration
reset
29
Page 35
Instruction Manual CyberScan DO 110
8.6.2 User Reset
This program clears all data including calibration, memory and other customised setup functions and reset it to factory default.
You can skip program P5.1 (Calibration Reset) and proceed straight to Program P5.2:
1. From Program P5.0 - rSt, press HOLD/ENTER key three times to P5.2. Press HOLD/ENTER key again to enter the User Reset menu.
2. Press the MI/ or MR/ keys to select or deselect the user reset function.
‘NO’ retains current settings
‘YES’ resets to factory default settings.
3. Press the ENTER key to confirm selection.
4. If ‘NO’ is selected, it will retains current settings and goes back to program 5.0.
5. If ‘YES’ is selected, all LCD segments will light up for one second, reset to factory default settings, and the meter will be in the measurement mode.
SETUP
SETUP
SETUP
HOLD
ENTER
MR
P 5.2
MI
yes
6. Otherwise press CAL/MEAS key to return to measurement mode without resetting to factory default.
Select either "YES" or "nO", press ENTER key to confirm.
Figure 23: P5.2 - User reset
30
Page 36
Instruction Manual CyberScan DO 110
y
8.7 P 6.0: Memory Clear (CLr)
Use this parameter to clear all memory values when you need to store a new series of values. This lets you avoid confusing the old values with the new ones.
NO is the default setting.
NOTE: Selecting YES will wipe out all memory.
From measurement mode:
1. Press the SETUP key to enter Set Up mode.
2. Press the MI/ or MR/ keys to scroll through subgroups until you view the parameter P6.0.
3. Press the HOLD/ENTER key.
4. Press the MI/ or MR/ keys to toggle between NO and YES.
NO retains current memory
YES clears all memory.
5. Press the HOLD/ENTER key to confirm selection and return to the subgroup menu.
6. Press CAL/MEAS key to return to measurement mode.
SETUP MEM
p 6.0
HOLD
ENTER
SETUP MEM
MR
SETUP MEM
MI
Figure 24: P6.0 - Clear
yes
memor
31
Page 37
Instruction Manual CyberScan DO 110
8.8 P7.0: Dissolved Oxygen Parameters (DPr)
This Set Up menu allows you to set the barometric pressure and the salinity value of the sample to be measured. You are given the option to use mmHg (Hg) or kilo Pascal (PA) barometric pressure units. The Salinity (SAL) value to be entered will be based on ppt and is available in mg/L (ppm) Set Up menu.
8.8.1 Pressure Setting Adjustment
From the mg/L or ppm measurement mode,
1. Press the SETUP to enter the mg/L or ppm Set Up mode.
2. Press the ▲ and ▼ keys to scroll through the Set Up menu till the display shows “dPr”.
3. Press the HOLD/ENTER key and the meter will enter the barometric pressure unit selection menu with the display showing either the “HG” mode or “PA” mode.
SETUP
SETUP
SETUP
HOLD
ENTER
MR
P7.0
PA
MI
H9
4. Use the ▲ and ▼ keys to toggle between the barometric pressure units and press the HOLD/ENTER key to confirm selection.
5. Use the ▲ and ▼ keys to set the pressure value and press the HOLD/ENTER to confirm the setting. The meter will show the next display of Salinity setting “SAL”.
NOTE: If the Set Up menu is derived from other than the mg/L (ppm) mode, the meter will return to the Set Up subgroup menu “dPr”.
6. Press CAL key to return to measurement mode, or continue to make a salinity setting adjustment.
8.8.2 Salinity Setting Adjustment (available in
mg/L or ppm Set Up mode only)
1. Repeat steps 1 to 5 as in the steps of Pressure Setting Adjustment.
2. Use the ▲ and ▼ keys to enter the salinity of your solution in ppt (parts per thousand).
3. Press HOLD/ENTER key to confirm the value and return to the subgroup menu “dPr”.
SETUP
SETUP
HOLD
ENTER
760
MR
H9
MI
0.0
SAL
4. Press CAL key to return to the measurement mode.
32
Page 38
Instruction Manual CyberScan DO 110
8.9 P 8.0: % Saturation Offset Adjustment (OFS)
This is a useful feature that allows you to offset meter’s value when cross referenced with another DO meter. That way, it can be standardized without you having to perform manual calculation. Your DO 110 meter allows % Saturation Offset adjustment within +/- 10.0% offset and its adjusted offset value can be viewed in the P3.0 Electrode Properties menu “EL.2”.
From the % Saturation measurement mode,
1. Dip the DO electrode in the sample solution and allow the reading to stabilize.
2. Check the reading of another DO meter being used as a reference. This reference meter should have its probe immersed in the same sample solution and at the same depth.
3. Press the SETUP to enter the % Saturation Set Up menu.
4. Scroll the Set Up menu by using the and keys until the meter displays the % Saturation Offset Adjustment P.70 menu “OFS.”.
5. Press the HOLD/ENTER key to enter the menu. The display will show the last offset adjusted value.
6. Use the ▲ and ▼ keys to enter the new value.
7. Press the HOLD/ENTER key to confirm the offset adjustment and the display returns to the subgroup menu.
SETUP
SETUP
OFS
HOLD
ENTER
MEAS
90.0
p 8.0
90.0
%
8. Press CAL key to exit the Set UP menu to return to the measurement mode.
Note: When a user calibration is done, the offset will be reset to zero.
33
Page 39
Instruction Manual CyberScan DO 110
9 CYBERCOMM PORTABLE DAS
The DAS software is designed for pH 110, CON 110 and DO 110 meters to allow you a convenient means of capturing data for future analysis using other software program such as LOTUS 123, EXCEL or DBASE in Windows©. Often one finds it cumbersome to record and transfer data from one media to another before the required processing can be done. With the DAS software, this redundant processing can be eliminated or reduced. To download this software, visit www.eutechinst.com
9.1 System Requirements
To run the DAS program, the following is required:
1. PC - IBM Compatible XT and above with CD-ROM Drive
2. EGA Monitor and above
3. Windows© Operating System ‘95 and above
4. Connecting communication RS232C cable
.
9.2 Loading CyberComm Portable DAS
Figure 25: Insert Eutech Instruments' CD-ROM containing Data Acquisition Software (DAS) into your CD-ROM drive. It will Auto Run and let you start loading the software by displaying the screen as in Figure 44. Alternatively, you can also manually load the software by clicking the START button and RUN command.
34
Page 40
Instruction Manual CyberScan DO 110
Figure 26: Click on 'Browse' button and locate CD-ROM drive
Figure 27: Locate the CyberComm Portable Setup program in the CD-ROM un der "Eutech DAS"
directory.
35
Page 41
Instruction Manual CyberScan DO 110
Figure 28: Select ‘Portable Meters’ directory to go to CyberComm Portable.
Figure 29: Open ‘CyberComm Portable’ to enter to the Setup program.
36
Page 42
Instruction Manual CyberScan DO 110
Figure 30: Select "SETUP.EXE" program and click the OPEN button
Figure 31: InstallShield Wizard dialog box appears.
Figure 32: Click on Next button
37
Page 43
Instruction Manual CyberScan DO 110
Figure 33: Key in your name and company name and click NEXT button
Figure 34: To select another Destination Directory to install the program, click on BROWS E button.
Otherwise, click NEXT button.
38
Page 44
Instruction Manual CyberScan DO 110
Figure 35: Creating a new program folder. Click on NEXT button.
Figure 36: Click on NEXT button.
39
Page 45
Instruction Manual CyberScan DO 110
Figure 37: The CyberComm Portable DAS program is fully installed. Click on FINISH button to end
installation.
40
Page 46
Instruction Manual CyberScan DO 110
9.3 Running CyberComm Portable
Before running the DAS program, please ensure that the RS232 cable is connected between the computer’s serial port and the meter’s port. See section 3.4 ­Connecting the RS232C Cable on page 6.
Figure 38: Run the CyberComm Software program
Figure 39: The opening screen will appear as above
41
Page 47
Instruction Manual CyberScan DO 110
i. Buttons & Check-Box
Enable Connection - Click this button to enable communication between meter and computer.
Clear Readings - To clear all data and start all over again.
Save Readings - To save all data displayed in either *.dat or *.txt format.
Time Stamp - To include Time and Date stamp when collecting the data. Time
and date information comes from the computer.
Figure 40: Under File Menu setting, you can change various paramete r s. Und er ABOUT menu, details of
Eutech Instruments' contact information, email add re ss and updat es are sh own.
ii. Menu
Communication Settings - To set communication port number, baud rate speed, parity and stop bits protocol.
Open - To open previously saved data file.
Save - To save current data captured.
Save As - To save current data set in another format such as *.dat or *.txt.
Exit - To exit from CyberComm Data Acquisition Software program.
42
Page 48
Instruction Manual CyberScan DO 110
Figure 41: Communication Settings for computer's Com port. It must match with COM port settings on CyberScan DO 110 meter. Please refer to Section 7.1 Using DO 110 Meter With Printer or Computer on
page 15 for the settings.
iii. Communication Settings
Connect Using - For selecting communication port. (Select ‘1’)
Baud Rate – For selecting baud rate. (Select ‘9600’ bps (bits per second)).
Parity – For selecting parity. (Select ‘None’)
Stop Bits – For selecting stop bits. (Select ‘1’).
Figure 42: Under SAVE AS menu, you can save your data as *.dat or *.txt formats
43
Page 49
Instruction Manual CyberScan DO 110
9.4 Capturing And Printing Data Into Computer Using Data Acquisition
After matching the Communication Settings between your computer using Data Acquisition program and the DO 110 meter, you can now capture data into your computer for analysis and storage purposes.
1. Ensure the 1-meter RS232 communication cable (Refer to Accessories Section) is connected between the computer and the DO 110 meter’s Com port. Refer to section 3.4 “Connecting the RS232C Cable” on page 6 for connection procedure.
2. Switch on the DO 110 meter and run the Data Acquisition software.
3. Click “ENABLE CONNECTION” button.
4. With the DO 110 meter switched on, press the PRINT key to send data to the computer.
5. You can use MODE key on the meter and change to mg/L or ppm and print data accordingly.
6. You can also check off the Time Stamp function, so as to print without the Time and Date information.
7. You can click Clear Readings button to begin another set of measurements, or click Save Readings to store readings for future retrieval.
9.5 Trouble-shooting Guide
a) Problem: Unable to PRINT
POSSIBLE CAUSES SOLUTIONS
You have not "ENABLE CONNECTION" in the Data Acquisition program.
The "Communication Settings" in the Data Acquisition program is different from meter's setup.
The COM port number in the Data Acquisition program is wrong.
Your computer's COM port setting may be wrong.
You may have used the wrong communication cable.
Click on "ENABLE CONNECTION" in the Data Acquisition program.
Match the COM port number, baud rate, parity and stop bits information between the Data Acquisition program and the meter.
Change the COM port number (1 or 2) in the Data Acquisition program.
Check your computer's hardware settings (through Windows OS, BIOS, or any other OS) and refer to computer's manual or consult with the computer's manufacturer.
Make sure you use the RS232C cable supplied together with the meter (Part No. EC-CA01M09F09).
Check the RS232C configuration as described in the meter's instruction manual.
To report any bugs, please e-mail to techsupport@eutechinst.com
44
Page 50
Instruction Manual CyberScan DO 110
10 Dissolved Oxygen Probe
10.1 Dissolved Oxygen Principle
The probe is a galvanic measuring element which produces an output proportional to the oxygen present in the medium in which it is placed. The galvanic probe design lets you take measurements immediately – without the typical 15 minute wait of other dissolved oxygen probes.
The probe consists of two parts:
An upper part consisting of an anode, a cathode, and cable.
A lower part consisting of a membrane cap, membrane, and electrolyte solution.
Oxygen diffuses through the membrane onto the cathode, where it is consumed. This process produces an electrical current which flows through the cable to the meter. The electric current produced is proportional to the oxygen that passes through the membrane and the layer of electrolyte. This makes it possible to measure the partial pressure of oxygen in the sample at a given temperature.
Since the DO in the sample is consumed by the cathode it is essential that a new sample must flow past the membrane of the probe to prevent the occurrence of false readings. The probe uses very little oxygen for its measurement. This enables it to function correctly with liquid movement as low as 2.5 cm/sec.
The permeability of the membrane to oxygen varies greatly with temperature. Therefore compensation is needed for this variation. The DO probe comes with an in­built Temperature Compensation for the membrane variation.
10.2 Probe Care
Under typical operating conditions, the probe should last for several years. Proper care and maintenance will help you receive the maximum probe life and ensure more accurate readings.
Since any deposits on the membrane surface act as a barrier to oxygen diffusing through the membrane, the membrane must be cleaned at regular intervals to assure maximum reliability.
After using the probe, rinse the probe in clean water and wipe it with a soft cloth or paper to avoid any hardening of deposits. If growth develops on the probe, use a disinfecting chemical to clean.
NOTE: Although the membrane is strong and not easily damaged, wipe it gently while cleaning it. If the membrane is damaged or torn, the probe will no longer function.
There are no special probe storage requirements.
45
Page 51
Instruction Manual CyberScan DO 110
g
10.3 Membrane Housing Replacement
Replacement of the membrane cap housing/ membrane is required only when you cannot calibrate the probe, or if the membrane is damaged.
Typical membrane damages are punctures or wrinkles caused during measurements or cleaning.
Your new DO probe comes with replacement membrane housing. To order more replacement membrane housing, see the “Accessories” section page.
10.3.1 To replace the membrane cap (with pre-installed
membrane) -
1. Replacement is much easier with single membrane housing. Unscrew the old membrane cap housing.
2. Hold the probe under hot running water and brush away the white oxide on the cylindrical anode with a stiff plastic brush – do not use metal cleaning material.
3. If the cathode has any deposits, remove them with a light scouring powder. Do not polish the cathode.
4. Fill the new membrane cap housing with electrolyte solution and inspect the bottom for leaks. If the solution
Figure 33: Positioning of O-rin
& membrane
drops are leaking from the membrane, use new cap housing.
5. If the assembly is leak-free, fill the membrane cap housing with electrolyte to the brim.
6. Tap the side of the housing gently to remove any air bubble that may be sticking to the membrane.
7. Screw the cap onto the probe. Excess electrolyte will drain out.
8. Replace probe guard.
9. Calibrate the probe (see section 4) after the % saturation readings have stabilised.
46
Page 52
Instruction Manual CyberScan DO 110
g
10.4 Membrane/O-ring Replacement (Optional Procedure)
It is recommended only experienced service personnel can perform this procedure.
This procedure is OPTIONAL, and should only be performed if you have new membrane and O-ring. You are also required to have a membrane installation tool. These items are available as optional accessories in the “Accessories” section.
Pull off the probe guard.
1. Unscrew the membrane cap from the probe.
2. Hold the probe under hot running water and brush away the white oxide on the cylindrical anode with a stiff plastic brush – do not use metal cleaning material.
3. If the cathode has any deposits, remove them with a light scouring powder. Do not polish the cathode.
4. Using the installation tool, unscrew and remove the membrane lock from the membrane cap. See Figure 35 on the following page.
5. Remove the membrane and O-ring. Discard both.
6. Rinse the membrane cap and membrane lock in tap water.
7. Install a new O-ring inside the membrane cap.
8. Install a new membrane. Make sure the membrane covers the O-ring all around its circumference.
9. Using the installation tool, screw the membrane lock back into the cap. Tighten the lock firmly over the membrane and O-ring, but do not over tighten.
10. Inspect the membrane for wrinkles. If wrinkles exist, remove the membrane and repeat steps 8 – 11.
11. Fill the membrane cap with water and inspect the bottom for leaks. If water drops are leaking from the membrane, re-seal the membrane on the O-ring (repeat steps 8 – 11, for membrane replacement only).
If the assembly is leak-free, empty the water and fill the membrane cap with electrolyte to the brim.
Figure 34: Positioning of O-rin
Close
& membrane
Installation tool
Open
Insert installation tool into slots on membrane lock. Then unscrew membrane lock from membrane cap
12. Tap the side of the housing gently to remove any air bubble that may be sticking to the membrane.
Figure 35: Use Tool to take out (or put in) membrane
13. Screw the cap onto the probe. Excess electrolyte will drain out.
47
Page 53
Instruction Manual CyberScan DO 110
14. Replace probe guard.
15. Calibrate the probe (see section 4) after the % saturation readings have stabilised.
Note: Membranes can only be used once. When a membrane cap is screwed onto the probe, the membrane is stretched by the cathode. If the same O-ring and the membrane is used a second time it will not fit perfectly onto the cathode. This will result in erratic readings.
10.5 Electrolyte Solution
The electrolyte solution in your probe’s cap will deplete on usage and will need to be replaced periodically.
Your new DO probe comes with accessories of one replacement electrolyte solution and a spare membrane cap. The replacement electrolyte comes premixed and ready to use. To order more electrolyte solution, see “Accessories” section.
48
Page 54
Instruction Manual CyberScan DO 110
11 ERROR MESSAGES
The following table provides a guideline to enable diagnosis of possible problems indicated by the messages generated by the meter. The table also provides possible solutions to the problems encountered.
ERROR MESSAGE INDICATES POSSIBLE CAUSE CORRECTIVE
ACTION
Err. Annunciator Wrong keypad input Wrong input in
selected mode.
Battery icon lights up Low Battery Battery power is low Replace batteries
Release key. Select valid operations depending on mode.
with fresh ones as soon as possible.
12 TROUBLE-SHOOTING
When experiencing difficulties with the equipment, keep in mind of the following:
1. Check for the obvious, such as physical condition of the probe, any signs of damage to the cable, power and signal connections etc.
2. Determine whether it is the probe, meter or the surrounding environment that is causing the problem.
The following troubleshooting table identifies most of the problems likely to occur:
Problem Probable Causes Solution
Nothing is displayed when the ON/OFF key is selected.
Fluctuating readings when probe is shaken or bumped lightly or when membrane is touched.
Batteries not in place.
Batteries (+ & - poles) not in correct polarity.
Weak batteries.
Probe has lost electrolyte – a sloshing noise will be heard when the probe is shaken.
Torn or damaged membrane.
Wet connections in the wiring or within probe. See Problem 2.
Insert batteries.
Re-insert batteries in the correct polarity.
Replace batteries or attach AC adapter.
Unscrew the membrane cap, discard the electrolyte, membrane and O-ring.
Dry the internals of the probe, especially the cathode with a soft cloth.
Switch the meter ON and observe the display.
If the display reads zero, the probe and cable circuitry are alright. Service probe and change membrane.
If the display does not read zero, but some other value,
49
Page 55
Instruction Manual CyberScan DO 110
then there is probably moisture somewhere. See Problem 2.
With membrane cap removed and probe internals thoroughly dry, the reading from the probe is not zero and/or is erratic.
It is not possible to calibrate the probe in air – the display will not read high enough after fully adjusting the offset.
Moisture has entered the system – either into the probe itself or at junctions or points in the cable. This moisture creates a secondary galvanic action in addition to that produced by the probe and results in non-zero or erratic readings.
Probe has dried out – no electrolyte inside.
Probe is overdue for servicing – excessive build-up of anode oxide.
A deposit has built-up on the silver cathode, which is inhibiting the reduction of oxygen at its surface.
Locate the source of moisture by process of elimination. If moisture has entered a junction box or a cable joint, thoroughly dry out the area and take measures to prevent reoccurrence.
Service probe and change membrane. Use a stiff nylon brush to remove the oxide built-up from the anode. Do not use a wire brush. It is only necessary to remove the loose oxide layer. If it is suspected that the anode is badly corroded, replace with a new DO probe. Remember to tighten the nut under the anode before fitting a new anode.
If it is suspected that a deposit is coating the silver cathode, clean the cathode with 400 grit wet/dry emery paper or with some scouring powder. The deposit is sometimes visible as a brownish stain on the surface of the cathode.
Display values are erratic when membrane is lightly touched. Membrane has bulged outwards.
NOTE: The cathode must not be polished – the surface must remain dull (do not use a wire brush).
The membrane has been damaged.
If the membrane has been damaged change it and service the probe.
50
Page 56
Instruction Manual CyberScan DO 110
13 LIST OF ACCESSORIES
13.1 Replacement Meter and Meter accessories
Eutech Instruments Items Order Code
CyberScan DO 110 Portable Dissolved Oxygen Meter complete with one Submersible Dissolved Oxygen probe with 3-meter (10-ft) cable (EC­DOHANDYNEW), DAS Software, RS232 Communication Cable, one assembled membrane cap housing (15X241402), one 60 ml refilling electrolyte (01X211226) and DO carrying kit set EC-WPDRYKIT
CyberScan DO 110 Portable Dissolved Oxygen Meter complete with one Submersible Dissolved Oxygen probe with 8-meter (25-ft) cable, DAS Software, RS232 Communication Cable, one assembled membrane cap housing (15X241402), one 60 ml refilling electrolyte (01X211226) and DO carrying kit set EC-WPDRYKIT
Galvanic Dissolved Oxygen Electrode with ATC, 3 m cable length EC-DOHANDYNEW
Membrane installation tool 15X241502
Membrane & O-ring (pack of 5) 01X241603
Replacement Membrane Housing Cap with Membrane and O-ring pre-installed 15X241402
DO Refilling Electrolyte (60ml) 01X211226
CyberScan Neutral Carrying Kit Set – Plastic Carrying Case comprises: 4 empty sample bottles (60 ml)
RS232 Communication Cable: 9-pin male to 9-pin female connector, 1m cable length EC-CA01M09F09
Serial Impact Micro-Printer: Dot-matrix, paper-roll portable printer with 25-pin female connector, with a roll of paper and 110/120 VAC power adapter provided
Serial Impact Micro-Printer: Dot-matrix, paper-roll portable printer with 25-pin female connector, with a roll of paper and 220/230 VAC power adapter provided
100/240 VAC SMPS Power Adapter (50/60 Hz) Centre +, with USA/UK/EUR plug, 9 VDC, 6 W (for standard portable meters)
Oakton Instruments
EC-DO110/01K
EC-DO110/02K
EC-WPDRYKIT
EC-MICROPRNTR01
EC-MICROPRNTR02
60X030130
Oakton Instruments Items Order Code
Oakton DO 110 Portable Dissolved Oxygen Meter Only with DAS Software 35640-22
Oakton DO 110 Portable Dissolved Oxygen Meter complete with one Submersible Dissolved Oxygen probe with 3-meter (10-ft) cable, DAS Software, one assembled membrane cap housing (15X241402) and one 60 ml refilling electrolyte (01X211226).
Oakton DO 110 Portable Dissolved Oxygen Meter complete with one Submersible Dissolved Oxygen probe with 3-meter (10-ft) cable, DAS Software, one assembled membrane cap housing (15X241402), one 60 ml refilling electrolyte (01X211226) and DO carrying kit set
RS232 Communication Cable: 9-pin male to 9-pin female connector, 1m cable length. 35615-09
Serial Impact Micro-Printer: Dot-matrix, paper-roll portable printer with 25-pin female connector, with a roll of paper and 110/120 VAC power adapter provided
Serial Impact Micro-Printer: Dot-matrix, paper-roll portable printer with 25-pin female connector, with a roll of paper and 220/230 VAC power adapter provided
Soft belt-loop carrying case 35615-75
100/240 VAC SMPS Power Adapter (50/60 Hz) Centre +, with USA/UK/EUR plug, 9 VDC, 6 W (for standard portable meters)
35640-20
35640-64
35622-00
35622-05
60X030130
51
Page 57
Instruction Manual CyberScan DO 110
14 SPECIFICATIONS
Dissolved Oxygen Range Resolution Relative accuracy
% Saturation of Oxygen Range Resolution Relative accuracy
Temperature Range (Meter) Resolution Relative accuracy
Salinity Correction Range Resolution Method
Barometric Pressure Correction (mm Hg) Range Resolution Method
0.00 – 19.99 mg/L or ppm
0.01 mg/L; 0.01 ppm ± 1.5% of Full Scale
0.0 – 199.9 %
0.1 % ± 1.5% of Full Scale
-10 to 110 °C ( 14 to 230 °F)
0.1 °C / 0.1°F (14
o
to 230 +/- 0.5
F)
o
C (+/- 0.9oF)
o
F - 199.9oF) & 1.0 (200oF
0.0 – 50.0 ppt
0.1 ppt Manual input and automatic correction
500 to 1499 mm Hg or 66.6 to 199.9 kPA 1 mm Hg or 0.1 kPA
Manual input and automatic correction Probe Galvanic Probe Temperature Operating Range 0.0 – 50.0 °C Probe Response Time
40 seconds to achieve 93% of the reading (at 25 °C with 3 meter (10-ft) cable) No. of Calibration Points Up to 2 points (0% and 100%) Data Logging 100 RS232 Yes HOLD function Yes Auto-Off function Selectable Auto Off Time 20 minutes after last key press Averaging/Stability function Selectable Display Dual LCD Power Requirements four 1.5 V AAA-sized batteries (included) or
100/240 VAC SMPS Power Adapter to 9 VDC,
6W (optional) Battery Life > 700 hours Temperature Compensation Automatic from 0 to 50 °C Operating Range 0 to 50 °C Dimensions Meter: 19.1 cm (L) x 8.9 cm (W) x 4.5 cm (H)
Boxed: 23.3 cm (L) x 21.6 cm (W) x 7.0 cm
(H)
Probe: 173 mm (L) x 32 mm (Diameter), with
3-m cable Shipping weight 920 g
52
Page 58
Instruction Manual CyberScan DO 110
15 WARRANTY
This meter is supplied with a three-year warranty, six-month warranty for probe against significant deviations in material and workmanship.
If repair or adjustment is necessary and has not been the result of abuse or misuse within the designated period, please return – freight pre-paid – and correction will be made without charge. Eutech Instruments/ Oakton Instruments will determine if the product problem is due to deviations or customer misuse.
Out of warranty products will be repaired on a charged basis.
Exclusions
The warranty on your instrument shall not apply to defects resulting from:
Improper or inadequate maintenance by customer
Unauthorised modification or misuse
Operation outside of the environment specifications of the products
16 RETURN OF ITEMS
Authorisation must be obtained from our Customer Service Department or authorised distributor before returning items for any reason. A “Return Goods Authorisation” (RGA) form is available through our Authorised Distributor. Please include data regarding the reason the items are to be returned. For your protection, items must be carefully packed to prevent damage in shipment and insured against possible damage or loss. Eutech Instruments/ Oakton Instruments will not be responsible for damage resulting from careless or insufficient packing. A restocking charge will be made on all unauthorised returns.
NOTE: Eutech Instruments Pte Ltd/ Oakton Instruments reserve the right to make
improvements in design, construction, and appearance of products without notice.
53
Page 59
Page 60
For more information on Eutech Instruments/ Oakton Instruments’ products, contact your nearest distributor or visit our website listed below:
Oakton Instruments
P.O Box 5136, Vernon Hills, IL 60061, USA Tel: (1) 888-462-5866 Fax: (1) 847-247-2984 E-mail: info@4oakton.com Web-site:
www.4oakton.com
Eutech Instruments Pte Ltd
Blk 55, Ayer Rajah Crescent, #04-16/24 Singapore 139949 Tel: (65) 6778 6876 Fax: (65) 6773 0836 E-mail:
marketing@eutechinst.com
Web-site: www.eutechinst.com
Distributed by:
Loading...