This manual serves to explain the use of the αlpha CON 2000 Series. The manual functions in two
ways, firstly as a step by step guide to help the user operate the instrument, and secondly as a handy
reference guide. This instruction manual is written to cover as many anticipated applications of the
αlpha CON 2000 as possible. If you have any doubts concerning the use of the instrument, please do
not hesitate to contact the nearest Eutech Instruments Authorised Distributor.
The information presented in this manual is subject to change without notice as improvements are made,
and does not represent any commitment whatsoever on the part of Eutech Instruments.
Eutech Instruments will not accept any responsibility for damage or malfunction of the unit due to
improper use of the instrument.
This Eutech Instruments Controller/ Transmitter shall be installed and operated only in the manner
specified in the Instruction manual. Only skilled, trained or authorised person should carry out
installation, setup and operation of the instrument.
Before powering up the unit, make sure that power source it is connected to, is as specified in the top
label. Failure to do so may result in a permanent damage to the unit.
The unit has live and exposed parts inside. If it has to be opened, make sure that the power to the unit
is off and disconnected.
The unit is Fuse protected. In the event the fuse has to be replaced, use only those as specified in the
manual.
The degree of protection against electric shock will be achieved only by observance of the
corresponding installation rules.
1.1AT THE VERY BEGINNING...................................................................................................................... 1
1.2INTENDED USE ......................................................................................................................................1
11.4.1 Control characteristic of Controllers used as limit value switch ............................................. 45
11.4.2 Control characteristic of P-Controllers as proportional controller......................................... 45
11.4.3 Control characteristic of PI-Controllers as proportional integral controller.......................... 45
11.4.4 Control signal of Pulse length Controllers............................................................................... 46
11.4.5 Control signal of Pulse Frequency Controllers........................................................................ 47
11.5APPENDIX 5–ABBREVIATIONS USED IN MENU DISPLAYS................................................................... 48
Instruction Manual αlpha CON 2000
1 INTRODUCTION
1.1 At the very beginning
We thank you for having purchased the Eutech Instruments αlpha CON 2000.
The construction of the αlpha CON 2000 employs leading edge technology and complies with
safety regulations currently in force. Notwithstanding this, improper use could lead to hazards
for the user or a third-party, and/or adverse effects on the plant or other equipment. Therefore,
the operating instructions must be read and understood by the persons involved before
work is started with the αlpha CON 2000.
Eutech Instruments do not accept any liability for damage that may arise from
neglecting information given in this manual.
This instruction manual identifies safety instructions and additional information by means of
the following symbols:
This symbol draws attention to safety instructions and warnings of potential danger which, if neglected, could result in injury to persons and/or damage to
property.
This symbol identifies additional information and instructions which, if neglected,
could lead to inefficient operation and possible loss of production.
The instruction manual must always be stored close at hand, in a place accessible to all
people working with the αlpha CON 2000.
If you have questions, which are not or insufficiently answered in this instruction manual,
please contact your Eutech Instruments supplier. He will be glad to assist you.
1.2 Intended use
Eutech Instruments αlpha CON 2000 is intended solely for conductivity and temperature
measurement, as described in this instruction manual.
Any other use, or use not mentioned here, that is incompatible with the technical specifications
is deemed inappropriate. The operator is solely responsible for any damage arising from such
use.
Other prerequisites for appropriate use include:
− observing the instructions, notes and requirements set out in this instruction manual.
− observing all local safety regulations concerning safety at work.
− observing all information and warnings in the documentation dealing with the products
used together with the transmitter (housings, sensors, etc.).
− observing the prescribed environmental and operational conditions.
1
Instruction Manual αlpha CON 2000
1.3 Safety intsructions
The αlpha CON 2000 should be installed and operated only by personnel familiar with
the transmitter and who are qualified for such work.
A defective transmitter must neither be installed nor put into service.
The αlpha CON 2000 must only be operated under the specified operating conditions
(see section 8).
The αlpha CON 2000 must not be repaired by the customer.
The αlpha CON 2000 must only be opened to replace the unit fuse or to set the
jumper for Pt100/Pt1000 temperature sensor. This work must be carried out only by
personnel familiar with the transmitter and who are qualified for such work. Make sure
the mains cable is separated from the power supply before opening the unit.
No modifications to the αlpha CON 2000 are allowed. The manufacturer/supplier
accepts no responsibility for damage caused by unauthorised modifications. The risk
is borne entirely by the user.
1.4 Putting out of service / Correct disposal of the unit
Putting out of service
• First disconnect the unit from the mains, then undo all electrical connections.
• Remove the unit from the wall / panel.
Correct disposal of the unit
When the transmitter is finally taken out of service, observe the local environmental
regulations for correct disposal or send the transmitter to your local Eutech Instruments
distributor, they will take care of proper disposal.
2
Instruction Manual αlpha CON 2000
2 PRODUCT DESCRIPTION
2.1 Description of unit
The Eutech Instruments αlpha CON 2000 is used for measuring conductivity and temperature
values. The conductivity values can be measured using limit or P/PI control. The transmitter is
available in two versions, one for panel mounting and one for wall mounting in a enclosure.
The transmitter can be used for applications such as water treatment and monitoring,
galvanic-decontamination, chemical processing, food processing, clean or wastewater control
and neutralisation processes.
This transmitter has many user-friendly and safety features which include:
• Menu-driven program that simplifies set-up.
• Built-in non-volatile memory to ensure that calibration and other information are not
erased if power supply fails.
• Push-button for calibration and sensor offset adjustment from the keypad.
• Automatic temperature compensation (ATC).
• Manual temperature compensation setting without the ATC probe, with independent
setting for calibration and process temperature.
• Two galvanically isolated current outputs 0/4...20mA.
• 0 to 2000 seconds time delay adjustment on all relays – minimises false alarms.
• Separately adjustable high and low set-point hysteresis (dead bands) prevent chattering
of relays around the set points.
• Three control modes: limit controller, P controller and PI controller (P/PI controller as
pulse length or pulse frequency).
• Large dual display LCD for easy reading with clear multiple annunciators, alarm status,
operational and error messages.
• Two switching contacts as set-point relays.
• Separate alarm relay alerting you to set point limits exceeded for a certain time and if the
Pt100/Pt1000 wires are broken or disconnected during the ATC function.
• Wash relay.
• Hold function to freeze output current (0/4...20mA) and release control relays.
• LED indicators signal control activities to visually monitor transmitter status from a
distance.
• Protection against electromagnetic interference. ( Available for panel mount only)
• Back lit and UV light protected LC display.
3
Instruction Manual αlpha CON 2000
A
2.2 Measurement and control system
A typical measurement system consists of:
• A conductivity process transmitter
• A conductivity sensor with integrated or separate temperature sensor Pt100/Pt1000.
• An appropriate measurement cable
• An immersion, flow or process assembly
• A final control element such as pump or valve
• A chart recorder
lpha CON 2000 Transmitter
Chart Recorder
Housing and
Sensors
Measurement Cable
0/4 - 20 mA
Power Mains
(80 - 250 VAC)
4
Instruction Manual αlpha CON 2000
2.3 Unit overview
Wall mounting version
Panel mounting version
5
Instruction Manual αlpha CON 2000
2.3.1 Display Overview
The LC display shows two alpha-numerical fields for parameters and measured values as well
as various mode and status indicators.
SETUP
HOLD
ERR
MEAS
4
2.3.2 Key functions
Key Description
CAL
mS
µS
°C °F
ATC
Mode indicators:
MEAS: measurement mode
SETUP: Set-up mode
CAL: Calibration mode
Status indicator:
HOLD: Unit in “HOLD” mode
ATC: Visible in ATC (Automatic Temperature
Compensation) mode. Not visible in the Manual
Temperature Compensation mode. “ATC”
flashes if the temperature probe is faulty in its
ATC mode
ERR: Error indicator
4: Measurement range number
Enter Calibration mode (requires access code)
Enter Set-up mode (requires access code)
Access sub functions (parameters) within a function group of Set-up mode
Confirm (store) set-up parameters and numerical values
Start/Confirm calibration in Calibration mode.
Select function group in the Set-up mode.
Set parameters and numerical values (if key is pressed continuously, the setting speed
increases).
Control the relays in MANUAL relay operation.
Returns to “Measurement mode” when both keys are pressed simultaneously.
Display limit values for SP1 and SP2 and settings for wash contact in AUTO relay
operation mode.
Toggle between RELAY A, RELAY B or Wash relay in MANUAL relay operation
Toggle between AUTO & MANUAL relay operation modes (requires access code to switch
to MANUAL relay operation mode)
6
Instruction Manual αlpha CON 2000
2.3.3 LED indicators
Relay indicators
If REL key is pressed the LED (A, B or W) indicates to which Relay (A, B or
Wash) the displayed limit values refer.
Relay mode indicators
Auto LED lights if relay operation is set to automatic mode. Manu LED lights
if relay operation is set to manual mode.
Relay status indicators
This LED lights if limit value is exceeded or the ATC probe fails.
2.3.4 Security codes
The access to Calibration mode, Setup mode and Manual relay operation mode is
protected with security codes. The security codes are set at the factory and cannot be
changed by the user. The following security codes are used:
Security code Mode/Description
000 View only mode/ to view actual settings
This LED lights green if measured value is within the limit for Relay A or
lights red if measured value exceeds limit.
This LED lights green if measured value is within the limit for Relay B or
lights red if measured value exceeds limit.
This LED lights if cleaning cycle is on.
11 Calibration mode/ to start calibration
22 Setup mode/ to configure parameters
22
Manual relay operation mode/ to switch relay operation mode
from AUTO to MANUAL
7
Instruction Manual αlpha CON 2000
2.3.5 Menu overview
MEAS
4
CAL
µS
°C
CCD “000” = View calibration parameters (View only mode)
ENT
1
CCD “11” = Calibration mode
2
SCD “000” = View setup parameters (View only mode)
SCD “22” = Setup mode
1
ENT
2
ENT
SETUP
HOLD
4
SETUP
HOLD
4
SETUP
HOLD
4
SETUP
HOLD
4
SETUP
HOLD
4
SETUP
HOLD
4
SETUP
HOLD
4
ENT
ENT
ENT
ENT
ENT
ENT
ENT
Temperature settings
See section 6.3
Relay A (set point 1) settings
See section 6.4
Relay B (set point 2) settings
See section 6.4
Controller settings
See section 6.5
Current output 1 settings
See section 6.6
Current output 2 settings
See section 6.7
SETUP
HOLD
4
SETUP
HOLD
4
SETUP
HOLD
4
SETUP
HOLD
4
ENT
ENT
ENT
ENT
Wash contact settings
See section 6.8
Range settings
See section 6.9
Unit settings
See section 6.10
Calibration
See section 5
8
Instruction Manual αlpha CON 2000
3 ASSEMBLY AND INSTALLATION
3.1 Mounting the unit
Wall mounting version
144 [5.67]
144 [5.67]
111.50 [4.39]
27.5 [1.08]
approx. 14 [.55]
21.5 [.85]
66.5 [2.62]
6 [.24]
Max.80
39 [1.54]
78 [3.07]
80 [3.15]
90 [3.54]
84 [3.30]
1-22 [0.03 - 0.87]
Pg13.5 (3 pcs.)
29 [1.14]
Holes for post
mounting (4X)
Holes for wall
mounting (2X)
90 [3.54]
Red Plugs for wall
mounting holes
enclosure (2X)
24 [.94]
For Pg13.5 cable glands
57.8[2.28]
175 [6.88]
Seal gasket
Threaded rod for pane l mounting
Unit:
MM [INCH]
138 [5.43]
138 [5.43]
Cut-out mounting plates for post mounting
175 [6.88]
Transmitter housing for wall mounting: protection class IP 66
9
Instruction Manual αlpha CON 2000
Panel mounting version
Flat gasket 1mm [.04]
(to be inserted by customer)
Panel cut out
Units: mm [Inch]
Transmitter housing for panel mounting: protection class IP 54 (front), IP 40 (housing)
10
Instruction Manual αlpha CON 2000
3.2 Connection Diagram
Connections for wall mounting version
Caution: Ensure electrical mains are disconnected before proceeding.
Anti-s urge
Current
250V, 315 mA,
Anti-surg e
1
L2N
Current
OP2
15
13
14
OP1
HOLD
GND
17
16
18 19 2120 23 24 25
3456
PT100/
PT1000
22
2926 2827 30 31 32 33
7 8 91011
12
Current
1
Current
OP2
13 14
OP1
15
23
17
16
HOLD
18 19 20 21 22 23 24 25
4
5
PT100/
67
PT1000
302928272631 3233
89
10
12
11
LN
1. AC mains live wire 18. 12V Power supply
2. AC mains neutral wire 19. Hold function
3. AC mains protective earth wire 20. Hold function
4. Relay A (SP 1) 21. Earth ground
5. Relay A (SP 1) 22. no connection
6. Relay B (SP 2) 23. no connection
7. Relay B (SP 2) 24. no connection
8. Wash relay 25. no connection
9. Wash relay 26. Temperature ground
10. Alarm relay (NC) 27. Temperature Sense
11. Alarm relay common 28. Temperature Input (short to terminal
12. Alarm relay (NO) 27 if using 2- wire RTD)
13. 4-20 mA temperature output, +ve terminal 29. no connection
14. 4-20 mA temperature output, -ve terminal 30. 4 Cell type Conductivity Input
15. 4-20 mA Conductivity output, +ve terminal **31. 4 Cell / 2 Cell type Conductivity Input
16. 4-20 mA Conductivity output, -ve terminal **32. 4 Cell / 2 Cell type Conductivity Input
17. 12V ground
33. 4 Cell type Conductivity Input
IMPORTANT: The Alarm relay functions as an “Active Low” device i.e. it switches
OFF under Alarm condition. Therefore the Alarm display device should be connected
to the ‘NC’ contacts of the relay (10 & 11).
11