Thermo Fisher Scientific CON 700 User Manual

Page 1
gyM
sy...
Instruction Manual
CON 700 Conductivity/ºC/ºF Bench Meter
Part of Thermo Fisher Scientific 68X541702 Rev. 3 Nov 2010
Tech nol o
adeEa
Page 2
Page 3
TABLE OF CONTENTS
1. INTRODUCTION.............................................................. 1
2. GETTING STARTED........................................................ 2
2.1 Keypad Functions ...................................................................2
2.2 LCD Annunciators...................................................................3
2.3 Meter Connections..................................................................3
3. ELECTRODE INFORMATION......................................... 4
4. CALIBRATION................................................................. 4
4.1 Automatic or Manual Calibration.............................................4
4.2 Single or Multi-Point Calibration..............................................5
4.3 General Calibration Tips .........................................................6
4.4 Automatic Conductivity Calibration Procedure........................6
4.5 Manual Conductivity & TDS Calibration Procedure.................7
4.6 Temperature Calibration .........................................................7
5. MEASUREMENT.............................................................. 7
5.1 Taking Measurements.............................................................7
5.2 Automatic and Manual Ranging..............................................8
5.3 HOLD Function .......................................................................8
5.4 Storing and Recalling Data .....................................................8
6. SETUP FUNCTIONS........................................................ 9
6.1 P1.0 CAL (Calibration)............................................................9
6.2 P2.0 ELE (Electrode Information)............................................9
6.3 P3.0 ConF (Configuration)......................................................9
6.4 P3.1 rdY (Ready / Stability Indicator)......................................9
6.5 P3.2 ºC ºF (Celcius Or Fahrenheit).........................................10
6.6 P3.3 AtC (Automatic Temperature Compensation).................10
6.7 P3.4 tdS (Total Dissolved Solids factor)..................................10
6.8 P3.5 t.CO (Temperature Coefficient).......................................10
6.9 P3.6 t.nr (Normalization Temperature in ºC)...........................11
6.10 ACAL (Automatic Conductivity Calibration).............................11
6.11 SPC (Single Point Calibration)................................................11
Page 4
6.12 CELL (Nominal Cell Constant)................................................12
6.13 P4.0 rSt (Reset) .....................................................................12
6.14 P5.0 CLr (Clear Memory).......................................................12
7. CALCULATING TDS CONVERSION FACTOR.............. 13
8. CALCULATING TEMPERATURE COEFFICIENTS........ 14
9. REPLACEMENTS AND ACCESSORIES........................ 15
10. TROUBLESHOOTING GUIDE......................................... 16
11. SPECIFICATIONS............................................................ 17
12. WARRANTY..................................................................... 18
13. RETURN OF ITEMS......................................................... 18
Page 5

1. INTRODUCTION

Thank you for purchasing our Con 700 series benchtop meter. This microprocessor-based meter is economical and simple to use. The design incorporates a large LCD for clear viewing, yet offers a small footprint to conserve space.
The CON 700 measures conductivity or TDS, and temperature (ºC or ºF) simultaneously.
Each meter includes a convenient slide-out card for quick reference.
Some configurations include an electrode arm and metal bracket which can be easily attached to the left or right side of the meter according to your preference.
The CON 700 series benchtop meter replaces our popular CON 510 series meter that was introduced in 2002.
We take great pride in every instrument we manufacture and hope this one serves you well.
1
Page 6

2. GETTING STARTED

2.1 Keypad Functions

Powers the meter on and off. Upon power on, the meter automatically begins in the mode that was last used. Calibration and memory values are retained even if meter is unplugged.
Toggle between available measurement modes— Conductivity with Temperature or TDS with Temperature. Also used to switch to Temperature calibration during Conductivity or TDS calibration mode.
Press and hold for 5 seconds to enter SETUP mode.
Toggles between measurement and calibration modes. In SETUP mode, returns user to the measurement mode.
Confirms calibration values in CAL mode. Confirms selections in SETUP mode. Changes range and resolution in MEAS mode.
View recalled values in memory mode. MI (Memory Insert) stores values into memory.
S Increase value. Scroll up in SETUP & CAL modes.
MR (Memory Recall) recalls values from memory T Decrease value. Scroll down in SETUP & CAL modes.
Freezes measured reading. Press again to resume live reading.
2
Page 7

2.2 LCD Annunciators

2.3 Meter Connections
CON/
Temp
DC
8-pin DIN connection for 2-cell Con/TDS/Temp electrode
Power supply
3
Page 8

3. ELECTRODE INFORMATION

The CON700 includes an electrode with a nominal cell constant of k = 1.0, built-in temperature sensor, and 1 meter cable. The Ultem body housing has good chemical resistant properties. The electrode design offers fast temperature response and reduces air entrapment, ensuring accurate, repeatable, and stable readings.
The wetted materials of the probe include:
Polyetherimide (Ultem) – protective probe guard
Polybutylterphalate (Valox) – sensor housing
Stainless Steel (SS 304) – 2 bands
The protective probe guard can be removed temporarily for cleaning however it must be re-attached during measurement and calibration. Erroneous results will occur while the probe guard
is removed.
Always immerse the probe beyond upper steel band for best results. Use the fill line on the outside of the probe guard for reference.

4. CALIBRATION

4.1 Automatic or Manual Calibration

The CON 700 is capable of automatic or manual calibration for conductivity, and manual calibration for TDS.
In the automatic calibration mode, the meter will automatically select one of (4) conductivity calibration standard values depending on the range and normalization temperature being used (TABLE 1).
If you only use calibration standards that are listed in TABLE 1, automatic calibration is recommended. If you intend to calibrate with one or more standards that are not listed in TABLE 1, the CON 700 must be set for manual calibration instead.
4
Page 9
The factory default is automatic conductivity calibration. See Section
6.10 to change this setting. There is no automatic calibration value
available for conductivity range 1.
TABLE 1
Automatic Calibration Values
Range # Conductivity
Range
r 1 0.00 – 20.00 μS None None r 2 20.1 – 200.0 μS 84 μS 76 μS r 3 201 – 2000 μS 1413 μS 1278 μS r 4 2.01 – 2 0.00 mS 12.88 mS 11.67 mS r 5 20.1 – 2 00.0 mS 111.8 mS 102.1 mS
Normalization Temperature
25 ºC 20 ºC
Range
#
r 1 0 – 10.00 ppm none r 2 10.1 – 100.0 ppm r 3 101 – 1000 ppm r 4 1.01 – 10.00 ppt r 5 10.1 – 100 ppt

4.2 Single or Multi-Point Calibration

Use Single-Point Calibration to apply a single calibration value across all ranges.
Use Multi-Point Calibration for individual calibration in each rang e. This will restrict an individual calibration so that it is applied to one range only. When using multi-point calibration, perform a calibration in each range that you expect to use for best results.
TDS Range
(using 0.5 TDS factor)
Automatic Calibration
Values
none none none none
5
Page 10
The factory default is Single-Point Calibration. See Section 6.11 to change this setting.

4.3 General Calibration Tips

For best results, periodic calibration with known accurate standards is recommended. A maximum of one calibration point per range can be performed. If multiple calibration points are used in the same range, the most recent one will replace the previous one.
When the electrode is replaced, it is best to clear the calibration to the factory default values (see Section 6.13).
Rinse or immerse the probe before calibration and between samples with clean water (deionized water is ideal).
The CON 700 has non-volatile memory which will retain all calibration values, as well as meter settings and memory values upon meter shut down or unexpected power loss.
To protect from erroneous calibrations, the allowable tolerance is ±40% of the factory default value.
Low conductivity standard solutions (less than 20 µS) are unstable and are very temperature dependent. As a result, reproducible calibration results are challenging in lowest measurement range #1 (0.00 to 20.0 µS).

4.4 Automatic Conductivity Calibration Procedure

1) Press
2) Dip the electrode into the conductivity standard and press Provide stirring for best results.
3) The primary display will show the factory default value, while the secondary display will lock on the appropriate automatic standard value from TABLE 1.
4) When the READY indicator appears, press to accept. The primary reading will flash briefly before returning to measurement mode upon successful calibration.
as needed to select conductivity (μS or mS).
.
6
Page 11

4.5 Manual Conductivity & TDS Calibration Procedure

1) Press (ppm or ppt) calibration.
2) Dip the electrode into the calibration standard and press . Provide stirring for best results.
3) The primary display will show the current reading, while the secondary display will be the factory default value. Adjust the
temperature reading using primary reading will flash briefly before returning to measurement mode upon successful calibration.

4.6 Temperature Calibration

The thermistor sensor used for automatic temperature compensation and measurement is accurate and stable, so frequent calibration isn’t required. Temperature calibration is recommended upon probe replacement, whenever the temperature reading is suspect, or if matching against a certified thermometer is desired.
1) Place the pr obe into a solution with a known accurate temperature
such as a constant temperature bath.
2) Press
measured temperature while the secondary displa y shows the factory default temperature.
4) Adjust the temperature using or . Press to accept or
to cancel. The meter allows an adjustable maximum value of ±
º
5
C (or ± 9ºF) from the factory default temperature.
as needed to select conductivity (μS or mS) or TDS
or . Press to accept. The
followed by . The primary display shows the

5. MEASUREMENT

5.1 Taking Measurements

1) Rinse the electrode with de-ionized or distilled water before use
to remove any impurities. Gently shake excess water droplets.
7
Page 12
2) Dip the probe into the sample beyond the upper steel band
(utilize the fill line on the outside of the probe guard for reference).
3) Allow time for the reading to stabilize. Note the reading on the display.
The clear yellow protective probe guard must be attached during measurement. Erroneous results will occur while the probe guard is removed.

5.2 Automatic and Manual Ranging

The CON 700 automatically selects the optimum range in which your readings appear. Refer to TABLE 1 for a list of the available ranges.
To turn-off automatic ranging, press
. MEAS will flash, indicating
that manual ranging is active. To manually select the next range, press again. After range 5, automatic ranging is resumed. For example, a reading of 465 μS will automatically settle in range 3.
Using manual range advancement this will read as 0.47 mS in range 4, and 0.5 mS in range 5. For best resolution, use auto-ranging.

5.3 HOLD Function

For prolonged observation of a reading, press
during measurement mode to freeze the display. The “HOLD” indicator will display when the reading is held. To release the held value and
resume measurement, press memory by pressing
.
again or insert the held value into

5.4 Storing and Recalling Data

The CON 700 can retain up to 100 data points into memory for later retrieval.
1) In the measur ement mode, press
to insert the measured value into memory. The stored memory location value (StO) is briefly displayed.
8
Page 13
2) To recall data from memory, press . The location of the most recent stored data is displayed first. Press location of the desired data, then press
3) Press to return to the stored data location. Press to return to measurement mode. To erase stored data, see Section 6.14.
or to select the
to accept.

6. SETUP FUNCTIONS

Use the set up mode to customize your instrument operation. During measurement, press and hold
mode. Press Press Press to go back one level or return to measurement mode.

6.1 P1.0 CAL (Calibration)

6.2 P2.0 ELE (Electrode Information)

6.3 P3.0 ConF (Configuration)

or to change programs or change options. to select the program or confirm selection.
Press
Press
to view stored calibration points in each range.
to view cell efficiency in each range.
for 5 seconds to enter SETUP
Press

6.4 P3.1 rdY (Ready / Stability Indicator)

Press
to access set-up programs 3.1 thru 3.9.
.
9
Page 14
Press or to choose READY “On”, READY “OFF”, or Auto HOLd.
Press
6.5 P3.2 ºC ºF (Celcius Or Fahrenheit)
Press Press Press
6.6 P3.3 AtC (Automatic Temperature Compensation)
Press Press Press
6.7 P3.4 tdS (Total Dissolved Solids factor)
Press Press or to select the desired TDS factor (.40 to
1.00). Press to confirm.
6.8 P3.5 t.CO (Temperature Coefficient)
The temperature coefficient is the amount of change in conductivity per degree temperature (% per ºC). The CON 700 is factory set to a temperature coefficient of 2.1 % per ºC. For most applications this will provide good results. The meter allows adjustment from 0.0 to 10.0.
TIP: Select 0.0% for uncompensated measurements. The temperature will be measured by the electrode and displayed in measurement mode—without compensation.
to confirm.
. or to select ºC or ºF. to confirm.
. or to select “Yes” or “No”. to confirm.
.
10
Page 15
See Section 8 – Calculating Temperature Coefficients.
Press Press Press
6.9 P3.6 t.nr (Normalization Temperature in ºC)
When Automatic Temperature Compensation is used, measurements are adjusted by the temperature coefficient to the normalization temperature. The default value is 25 ºC.
Press . Press Press
6.10 P3.7 ACAL (Automatic Conductivity Calibration)
See Section 4.1 for more details on Automatic Conductivity Calibration.
Press . Press Press
6.11 P3.8 SPC (Single Point Calibration)
See Section 4.2 for details on Single Point & Multi Point Calibration.
Press
. or to select the desired value (0.00 to 10.0). to confirm.
or to select the desired value (15.0 to 30.0). to confirm.
or to select “Yes” (Automatic) or “No” (Manual). to confirm.
. Press Press to confirm.
or to select “Yes” (Single) or “No” (Multi).
11
Page 16
6.12 P3.9 CELL (Nominal Cell Constant)
The CON 700 includes a probe with a nominal cell constant (k) of 1.0. Use probes with k = 0.1 and 10 (sold separately) for improved performance in extreme sample ranges. Use this setup function to change the cell constant if necessary. Meter default is 1.0 to match the included probe. k = 0.1 ideal for low measurements <20 µS (<10 ppm). k = 1.0 ideal for mid-range measurements k = 10 ideal for high measurements >20 mS (>10 ppt).
Press Press Press
.
or to select 0.1, 1.0, or 10.0
to confirm.
6.13 P4.0 rSt (Reset)
Press Press If “Yes”, press or to select “Cal” (calibration reset
only) or “FCt” (complete reset to factory default settings). Press to confirm.
6.14 P5.0 CLr (Clear Memory)
Press Press Press
.
or to select “Yes” (Reset) or “No” (Cancel).
.
or to select “Yes” (Erase memory) or “No”.
to confirm.
12
Page 17

7. CALCULATING TDS CONVERSION FACTOR

You can calibrate TDS using the value of the calibration standard solution at a standard temperature such as 25 ºC. To determine the conductivity­to-TDS conversion factor use the following formula:
Factor = Actual TDS ÷ Actual Conductivity @ 25 ºC
Actual TDS: Value from the solution bottle label or as a standard made using high purity water and precisely weighed salts. ppm = milligram of salt(s) per liter of water = mg/L ppt = gram of salt(s) per liter of water = g/L
Actual Conductivity: Value measured using a properly calibrated Conductivity/Temperature meter.
Both the Actual TDS and the Actual Conductivity values must be in the same magnitude of units. For example, if the TDS value is in ppm the conductivity value must be in µS; if the TDS value is in ppt the conductivity value must be in mS.
Check your factor by multiplying the conductivity reading by the factor in the above formula. The result should be in TDS value.
When the TDS factor is set to 1.0, Conductivity = TDS.
13
Page 18

8. CALCULATING TEMPERATURE COEFFICIENTS

To determine the temperature coefficient of your sample solution use this formula:
Where: tc = Temperature coefficient 25 = 25 ºC C T1 = Temp 1 T2 = Temp 2 NOTE: A controlled temperature water bath is ideal for this procedure.
1. Immerse the probe into a sample of your solution and adjust the
2. Wait for 5 minutes. Note T1 and CT1 (conductivity at T1).
3. Condition the sample solution and prob e to a temperature (T2) that is
NOTE: Record your results for future reference. Ideally T1 and T2 should
4. Calculate the temperature coefficient of yo ur solution acc ording to the
5. Enter the calculated temperature coefficient into the meter. The calculated temperature coefficient will now be applied to all the meter
readings.
= Conductivity at Temp 1 CT2 = Conductivity at Temp 2
T1
temperature coefficient to 0% (that is, no compensation) by following instructions as described in Section 6.8.
about 5 ºC to 10 ºC different from T1, and note the conductivity reading C
.
T2
bracket your measurement temperature, and should not differ by more than 5 ºC.
formula shown above.
14
Page 19

9. REPLACEMENTS AND ACCESSORIES

Item Description
CON 700 Benchtop with electrode and integral stand,100/240 VAC
Replacement electrode, k = 1.0 CONSEN9501D 35608-74 Epoxy/platinum electrode, k = 0.1 93X546101 35608-72 Glass/platinum electrode, k = 1.0 - 35608-76 Epoxy/platinum electrode, k = 10.0 93X546102 35608-78 SMPS, 100/240 VAC, 9V, 6W power adapter 60X030130 35615-07 10 μS, (20) x 20mL Sachets ECCON10BS 35653-09 84 µS, 480 mL bottle* ECCON84BT 00653-16 100 μS, 480 mL bottle ECCON100BT ­447 μS, 480 mL bottle ECCON447BS 00653-47 500 μS, 480 mL bottle ECCON500BT ­1413 μS, 480 mL bottle* ECCON1413BT 00653-18 1413 μS, (20) x 20mL Sachets* ECCON1413BS 35653-11 2764 μS, 480 mL bottle ECCON2764BT 00653-20 2764 μS, (20) x 20mL Sachets ECCON2764BS 35653-12
5.0 mS, 480 mL bottle ECCON5000BT -
12.88 mS, 480 mL bottle* ECCON1288BT 00606-10 15 mS, (20) x 20mL Sachets ECCON15000BS 35653-13
111.8 mS, 480 mL bottle* ECCON1118BT -
Part number Ordering Code
Eutech Instruments Oakton Instruments
ECCON70043S 35411-00
* Calibration Standard Value at 25 ºC
15
Page 20

10. TROUBLESHOOTING GUIDE

PROBLEM CAUSE SOLUTION
No display Main power not switched
“Ur” (Under range)
“Or” (Over range)
in primary or secondary display
Slow response
Meter not responding to key press
Secondary display continually scrolls
on. AC Adapter socket not inserted properly.
Measured value is out of range.
Electrodes not connected. Electrode clogged, dirty
or broken. Meter not calibrated. Wrong temp value.
Dirty electrode. Temperature of sample is
changing. Manual HOLD or Auto
HOLD is active. Worn keypad.
Invalid key; Button is not functional in the current operation mode.
Conductivity calibration standard is not within 40% of expected value during automatic calibration.
a) Switch on the power supply. b) Re-insert AC Adapter.
Check electrode is connected. Clean or replace electrode. Recalibrate the meter. Confirm measurement condition. Treat samples to bring within
meter measuring range. Reset meter.
Clean the electrode. Allow temperature to stabilize.
Press HOLD. See Section 6.4 to disable Auto Hold feature if enabled.
Contact Technical Service. Press alternate key.
Select valid key depending on current mode.
Ensure electrode guard has not been removed.
Use fresh or different calibration standard.
to deactivate manual
16
Page 21

11. SPECIFICATIONS

Conductivity Range
Conductivitity Resolution
TDS Range .. to 100.0 ppt @ 0.5 fact (200.0 @ 1 factor) Resolution 0.01 / 0.1 / 1 ppm ; 0.01 / 0.1 ppt Accuracy ±1% Full Scale Cell Constant 0.1, 1.0, 10.0 (selectable) TDS Factor 0.40 to 1.00 (adjustable)
Calibration points Temperature
Range (Meter)
Compensation Type & Range
Resolution Accuracy Calibration Coefficient
Auto-ranging Yes Hold Function Yes Memory 100 data sets Input 8-pin DIN Power AC/DC 9V, 6W Adapter (100/240 VAC, 50-60Hz) Dimensions (mm) meter only = 175 (L) x 155 (W) x 69 (H)
Automatic (4 points); Maximum 1 per range
Manual (5 points); Maximum 1 per range
Automatic with supplied cell or Manual.
0.0 to 80 ºC / 32.0 to 176.0 ºF with supplied cell
Offset in 0.1 º increments; Offset range: ± 5.0 ºC / ± 9.0 ºF
..µS to 200.0 mS
0.01 / 0.1 / 1 μS ; 0.01 / 0.1 mS
0.0 to 100.0 ºC / 32.0 to 212.0 ºF
0.0 to 100 ºC / 32.0 to 212.0 ºF,
0.1 ºC / 0.1 ºF
± 0.5ºC / ± 0.9ºF
0.0 to 10.00 per ºC
17
Page 22

12. WARRANTY

This meter is supplied with a warranty against significant deviations in material and workmanship for a period of THREE years from date of purchase whereas probe with a SIX month warranty.
If repair or adjustment is necessary and has not been the result of abuse or misuse within the designated period, please return – freight prepaid – and correction will be made without charge. Eutech Instruments/Oakton Instruments will determine if the product problem is due to deviations or customer misuse.
Out of warranty products will be repaired on a charged basis. The warranty on your instrument shall not apply to defects resulting from:
Improper or inadequate maintenance by customer
Unauthorized modification or misuse
Operation outside of the environment specifications of
the products

13. RETURN OF ITEMS

Authorization must be obtained from our Customer Service Department or authorized distributor before returning items for any reason. A “Return Material Authorization” (RMA) form is available through our authorized distributor. Please include data regarding the reason the items are to be returned. For your protection, items must be carefully packed to prevent damage in shipment and insured against possible damage or loss. Eutech Instruments will not be responsible for damage resulting from careless or insufficient packing. A restocking charge will be made on all unauthorized returns.
NOTE: Eutech Instruments Pte Ltd /Oakton Instruments reserve the right to make improvements in design, construction, and appearance of products without notice.
18
Page 23
NOTES
Page 24
For more information on our products, please contact our channel partner or visit our websites listed below:
Oakton Instruments
625 E Bunker Court Vernon Hills, IL 60061 USA Tel: (1) 888-462-5866 Fax: (1) 847-247-2984

info@4oakton.com www.4oakton.com

Distributed by:
Eutech Instruments Pte Ltd
Blk 55, Ayer Rajah Crescent, #04-16/24 Singapore 139949 Tel: (65) 6778 6876 Fax: (65) 6773 0836
eutech@thermofisher.com www.eutechinst.com
Part of Thermo Fisher Scientific
Loading...