7. parts per thousand indicator
(CON 410 meter only)
AM PM
12
131415
Figure 1: Full LCD Screen
410 meter only)
Compensation indicator
14. Low battery indicator
Secondary Display
ATC
1011
15. Probe indicator
17. Cell constant indicator
- 2 -
Instruction Manual CON 400/410
2.2 Keypad
The large membrane keypad makes the instrument easy to use. Each button, when pressed,
has a corresponding graphic indicator on the LCD. See Figure 2. Some buttons have several
functions depending on its mode of operation.
Key Function
ON/OFF Powers on and shuts off the meter. When you switch on the meter, the meter starts up
HOLD Freezes the measured reading. To activate, press HOLD while in measurement mode.
MODE Selects the measurement parameter.
CAL/MEAS Toggles between Calibration and Measurement mode.
ENTER /
RANGE
MI/S and
MR/T
SETUP Takes you into the SETUP mode. This mode lets you customise meter preference and
in the mode that you last switched off from. For example, if you shut the meter off in
TDS measurement mode (only in CON 410 meter), the meter will be in TDS
measurement mode when you switch the meter on.
To release, press HOLD again.
~ CON 400 meter: Toggles between conductivity and time.
~ CON 410 meter: Toggles between conductivity, TDS and time.
NOTE: Temperature calibration is available from conductivity calibration mode; see
page 9 for directions.
ENTER function: Press to confirm values in Calibration mode and to confirm selections
in SETUP mode.
RANGE function: Press to enter manual ranging function.
The MEAS indicator blinks while in manual ranging function.
In Measurement mode:
Press MI/S (memory input) to store values with its corresponding temperature values in
the memory. Press MR/T (memory recall) to retrieve data from memory.
In Calibration mode:
Press to scroll through calibration values.
In SETUP mode:
Press to scroll through the setup subgroup programmes.
defaults, and view calibration, electrode offset data and select cell constant.
Figure 2 - Keypad
- 3 -
Instruction Manual CON 400/410
3 PREPARATION
3.1 Inserting the Batteries
Four AAA batteries are included with your meter.
1. Use a Philips screwdriver to remove the two screws holding the battery cover. See Figure
3 below.
2. Remove battery cover to expose batteries.
3. Insert batteries. Follow the diagram inside the cover for correct polarity.
4. Replace the battery cover into its original position using the two screws removed earlier.
Figure 3 - Back panel of meter showing meter compartment
- 4 -
Instruction Manual CON 400/410
3.2 Conductivity Electrode Information
The CON400/410 meter uses a conductivity / TDS cell with a
sturdy 6-pin connector.
Your meter includes a conductivity probe (Part No: EC-
CONSEN91W / 35608-50) Ultem / Stainless Steel cells with a cell
constant of K = 1.0. This conductivity / TDS cell features a built-in
temperature sensor for Automatic Temperature Compensation
(ATC). It has a specially designed housing that provides fast
temperature response and reduces air bubble entrapment, which
makes it easy to obtain accurate, stable readings.
Wetted parts include:
1. Polyetherimide (Ultem™)
2. Polybutylterphalate (Valox™)
3. Stainless Steel (SS 304)
NOTE: We recommend that you do not submerge the probe above
the protective yellow cap for prolonged periods. The protective
yellow cap can be removed for cleaning, but must be attached at
all times during calibration or measurement.
See Section 8 on page 40 for “Probe Care and Maintenance”
information.
1
Figure 4 - Conductivity probe
(EC-CONSEN91W/ 35608-50)
1
Valox and Ultem are trademark names of General Electric Co.
- 5 -
Instruction Manual CON 400/410
3.3 Connecting the probe to the meter
1. Line up the notch and 6 pins on the meter with the holes in the 6-pin connector. Push
down and turn the locking ring clockwise to lock into place. See figure below.
2. To remove probe, turn the locking ring counterclockwise on the probe connector. Pull
probe away from the meter.
CAUTION: Do not pull on the probe cord or the probe wires might disconnect.
NOTE: Keep connectors clean. Do not touch connector with soiled hands.
Figure 5 - Connection for Conductivity probe (6-pin connector)
- 6 -
Instruction Manual CON 400/410
4 CALIBRATION
4.1 Important Information on Meter Calibration
Your meter has five measuring ranges. You can calibrate one point in each of the measuring
ranges (up to five points). If you are measuring values in more than one range, make sure to
calibrate each of the ranges you are measuring.
To view current calibration points, see SETUP section Program 2.0 on page 28.
The following table lists the corresponding conductivity and TDS ranges. You should calibrate
each range using a solution that falls between the values in the “recommended calibration
solution range” column.
Conductivity
Measuring
Range
0.00 Æ 19.99 μS 6.00 to 17.00 μS 0.00 Æ 9.99 ppm 3.00 to 8.50 ppm
0.0 Æ 199.9 μS 60.0 to 170.0 μS 10.0 Æ 99.9 ppm 30.0 to 85.0 ppm
0 Æ 1999 μS 600 to 1700 μS 100 Æ 999 ppm 300 to 850 ppm
0.00 Æ 19.99 mS 6.00 to 17.00 mS 1.00 Æ 9.99 ppt 3.00 to 8.50 ppt
0.0 Æ 199.9 mS 60.0 to 170.0 mS 10.0 Æ 200 ppt 30.0 to 170 ppt
When you recalibrate your meter, old calibrations are replaced on a range by range basis. For
example, if you previously calibrated your conductivity meter at 1413 μS in the 0 to 1999 μS
range and you recalibrate at 1500 μS (also in the 0 to 1999 μS range), the meter will replace
the old calibration data (1413 μS) in that range. The meter will retain all calibration data in
other ranges.
To completely recalibrate your meter, or when you use a replacement probe, it is best to clear
all calibration data in memory. To erase all the old conductivity and TDS calibration data
completely from memory, see SETUP section Program 8.0 on page 39.
For information on how to calibrate your meter:
• See section 4.3 for Temperature Calibration.
• See section 4.4 for Conductivity
• See section 4.5 for TDS Calibration (CON 410 meter only).
• See Addendum 1 for more calibration tips.
Recommended
Calibration Solution
Range
TDS Measuring
Range
Recommended
Calibration Solution
Range
- 7 -
Instruction Manual CON 400/410
4.2 Preparing the Meter for Calibration
Before starting calibration, make sure you are in the correct measurement mode. When you
switch on the meter, the meter begins with the units you shut it off in.
For best results, select a standard value close to the sample value you are measuring.
Alternatively use a calibration solution value that is approximately 2/3 the full-scale value of
the measurement range you plan to use. For example, in the 0 to 1999 μS conductivity range,
use a 1413 μS solution for calibration.
Do not reuse calibration solutions after calibration. Contaminants in the solution can affect the
calibration, and eventually the accuracy of the measurements. Use fresh calibration solution
each time you calibrate your meter.
NOTE: Your meter is factory set to a temperature coefficient of 2.1% per °C. For most
applications this will provide good results. See Program P5.1 on page 34 to set the
temperature coefficient to different value. See Addendum 3, “Calculating Temperature
Coefficients” to determine the appropriate temperature coefficient for your solution.
NOTE: The factory default value for normalization temperature is 25 °C. If you need to
normalize to a value other than 25 °C, see Program P5.2 on page 35.
- 8 -
Instruction Manual CON 400/410
μSμ
4.3 Temperature Calibration
Your probe features a built-in temperature sensor. The
temperature sensor is factory calibrated. Calibrate your
sensor only if you suspect temperature errors that may
have occurred over a long period of time or if you have a
replacement probe.
Temperature Calibration
1. Make sure the cell is attached to the 6-pin connector.
The ATC annunciator will appear at the right-hand
side of the LCD*.
2. Switch the meter on. Press the MODE key to select
conductivity or TDS mode.
3. Press the CAL/MEAS key to enter conductivity or
TDS calibration mode. The CAL indicator appears
above the primary display.
4. While in conductivity or TDS calibration mode, press
the MODE key to enter temperature calibration
mode. The primary display shows the current
temperature reading and the secondary display
shows the factory default temperature value.
5. Dip the cell into a solution of known temperature (i.e.
A temperature bath). Allow time for the built-in
temperature sensor to stabilise.
6. Scroll with the MI/S or MR/Tkeys to set the correct
temperature value (i.e. the temperature of the
temperature bath). You can adjust the reading in
increments of 0.1 °C.
7. Once you have selected the correct temperature,
press the ENTER key.
* If the ATC indicator does not light up, see Program P4.3
to switch it on.
NOTES:
•You can offset the temperature reading up to ±5°C from
the original reading (default reading).
•To exit this program without confirming the temperature
calibration value, DO NOT press ENTER. Press
CAL/MEAS instead.
•If the ATC indicator does not light, see Program P4.3,
on page 32 to switch it on.
•Since temperature readings affect the accuracy of the
conductivity/TDS measurements, it is strongly
recommended to carry out a conductivity/TDS
calibration after a temperature calibration is done.
- 9 -
MEAS
1413
L
A
C
S
A
E
M
1413
E
D
O
M
22.3
I
M
22.0
R
E
T
N
E
E
G
N
A
R
MEAS
READY
1425
Figure 10 - Temperature
Calibration
22.3
CAL
22.3
CAL
22.3
CAL
22.3
22.0
°C
ATC
S
°C
ATC
°C
ATC
R
M
°C
ATC
μ
S
°C
ATC
Instruction Manual CON 400/410
4.4 Conductivity Calibration
The CON 400 and CON 410 meters are capable of up
to 5-point conductivity calibration at one point per