Amplitude™ Solution with the TaqPath™ COVID‑19
High‑Throughput Combo Kit—Module 1
Module 1—Sample and reagent transfer to deep‑well plates for RNA extraction
Pub. No. MAN0019429 Rev. B.0
Note: For safety and biohazard guidelines, see the “Safety” appendix in the Amplitude™ Solution with the TaqPath™ COVID‑19
High‑Throughput Combo Kit Instructions for Use (Pub. No. MAN0019428). Read the Safety Data Sheets (SDSs) and follow the handling
instructions. Wear appropriate protective eyewear, clothing, and gloves.
Required samples, reagents, and consumables
For information about the location of the required materials in the deck layout, see “Load the Tecan™ Fluent™ 1080 Automation
Workstation deck” on page 6.
Required samples, reagents, and consumables
Sample input
Barcoded sample tubes in sample carrier racks376 samples per runpage 8
Tips and Plates
KingFisher™ Deep‑Well 96 Plate with Barcode (sample extraction plates)4 plates per run
For In Vitro Diagnostic Use. For Emergency Use Authorization Only | Rx Only
page 6
Workflow
Prepare and load the sample tube runners, ensure that you have all of the required materials, then prepare the Tecan™ Fluent™ 1080
Automation Workstation.
Every 6 runs, ensure that you prepare reagents before they need to be reloaded.
Before each run, confirm that it is time to load Module 1.
To start a run, select the appropriate method for your sample tube size.
Every run, follow the instrument touchscreen prompts to load the reagents and consumables as required.
Start the method on the Tecan™ Fluent™ 1080 Automation Workstation
Load the Tecan™ Fluent™ 1080 Automation Workstation deck
• Every run—Binding Bead Mix, Silicone Oil, and KingFisher™ Deep‑Well 96 Plates with Barcodes
• Every 3 runs—1,000 µL FCA Disposable Tips
• Every 6 runs—All reagents and consumables
Load the tube runners in the Tecan™ Fluent™ 1080 Automation Workstation
Every run, load the tube runners containing the sample tubes when prompted by the instrument touchscreen.
Monitor the Module 1 run
During each run, monitor the run progress on the instrument touchscreen.
Transfer sample extraction plates from Module 1 to Module 2
After a run is complete, immediately transfer the 4 Sample Extraction Plates to Module 2.
Unload the tube runners
After transferring the Sample Extraction Plates to Module 2, unload the tube runners when prompted by the instrument touchscreen.
Prepare the Tecan™ Fluent™ 1080 Automation Workstation for the next shift or to be idle
At the end of each shift, empty tip waste and perform daily Tecan™ Fluent™ 1080 Automation Workstation system care.
Before you begin
At the beginning of each shift, complete the following tasks:
• Prepare tube runners (page 3)
• Load sample tubes into tube runners (page 3)
• Ensure that you have all of the required materials to complete the desired number of runs.
• Prepare the Tecan™ Fluent™ 1080 Automation Workstation (page 4)
2Amplitude
™
Solution with TaqPath™ COVID‑19 High‑Throughput Combo Kit—Module 1 Quick Reference
Prepare tube runners
3
1
2
2
1
Note: The number of required tube runners and the control positions depend on the size of the sample tubes.
1. Ensure that you have the correct tube runners for your tube size.
Sample tube sizeTube runner typeNumber of tube runners used per run
12–14 mm13 mm, 32‑position12
15–17 mm16 mm, 26‑position16
16–18 mm17 mm, 24‑position16
2. Place plug inserts in position 1 (Negative Control) and position 2 (Positive Control) in four tube runners.
Note: We recommend labeling the tube runners that contain plugs in the control positions so that they are easily identifiable
when loading them on the Tecan™ Fluent™ Automation WorkstationTecan™ Fluent™ 1080 Automation Workstation deck. Control tube
runners must be loaded at specific grid positions on the deck (see “Load the tube runners in the Tecan™ Fluent™ 1080 Automation
Workstation” on page 8).
Figure 1 Plug insert positions in control tube runners
17 mm, 24‑position control tube runner (16–18 mm tubes)—Plugs
1
in positions 1 and 2
16 mm, 26‑position control tube runner (15–17 mm tubes)—Plugs
2
in positions 1, 2, 25, and 26
13 mm, 32‑position control tube runner (12–14 mm tubes)—Plugs
3
in positions 1 and 2
3. Place plug inserts in positions 25 and 26 in all 16 of the 26‑position (15-17 mm tubes) tube runners (including the control tube
runners).
The plug inserts block positions 25 and 26 of the 26‑position (15-17 mm tubes) to prevent insertion of sample tubes.
Figure 2 Plug insert positions in 16 mm, 26‑position tube runners (15–17 mm tubes)
16 mm, 26‑position control tube runner (15–17 mm tubes)—Plugs
1
in positions 1, 2, 25, and 26
16 mm, 26‑position tube runner (15–17 mm tubes)—Plugs in
2
positions 25 and 26
Load sample tubes into tube runners
CAUTION!
the BSL‑2 biological safety cabinet until they are all ready to be loaded into the Tecan™ Fluent™ 1080 Automation Workstation (see
“Load the tube runners in the Tecan™ Fluent™ 1080 Automation Workstation” on page 8).
Handle all samples and controls as if they are capable of transmitting infectious agents. Keep the tube runners in
Before you begin, prepare the tube runners (for more information, see “Prepare tube runners” on page 3).
1. In a BSL‑2 biological safety cabinet, uncap the sample tubes, then discard the caps.
Note: New caps are put on the sample tubes after the Module 1 run is complete.
2. Place the uncapped sample tube into the open positions in the prepared tube runners.