QUICK REFERENCE
ViralSEQ™ viral detection assays. Security, audit, and

Product description

The AccuSEQ™ RealTime PCR Software v3.1 analyzes data generated on the QuantStudio™ 5 RealTime PCR Instrument using
0.1mL 96well plates. The software analyzes data for the Applied Biosystems™ MycoSEQ™ Mycoplasma Real-Time PCR Detection Kit, the Applied Biosystems™ ProteinSEQ™ protein quantitation systems, and the Applied Biosystems™ resDNASEQ™ Quantitative DNA Kits. It also includes a custom template option to support realtime PCR assays without a pre-dened template, such as the Applied

Home screen

Sign in to the AccuSEQ™ RealTime PCR Software
1.
Launch the AccuSEQ™ RealTime PCR Software by double­clicking the AccuSEQ icon .
2.
Enter the Username, then Password.
(First login only) The default username is Administrator and the default password is Administrator.
3.
Click Sign in.
1
Open Experiment—View the last 4 experiments run on the instrument. Click Browse All to open a full list of experiments.
2
Monitor the Run—View the time remaining for the run; view the instrument calibration status. For information on instrument calibration, see the QuantStudio™ 3 and 5 Real‑Time PCR Systems Installation, Use, and Maintenance Guide (Pub. No. A43322).
3
4
5
6
Experiments list
Templates list
Instrument status
Create Experiment—Select from existing templates or create a new one.
For Research Use Only. Not for use in diagnostic procedures.
Workflow

Create an experiment from a template

Load the plate in the instrument

Start the run

Monitor the run
Review the results
Create and print a report
Create an experiment from a template
1.
Open a template by doing one of the following:
• In the Create New Experiment pane of the (Home) screen, select an existing template from the Factory default/Admin Defined Templates or My Templates tabs.
• From the (Home) screen, click (Templates) for a full list of templates. Select a template, then click Experiment).
2.
In the Experiment Properties pane of the Setup tab:
a.
(Optional) Change the systemgenerated name of the experiment.
b.
(Optional) Enter the plate Barcode.
c.
(Optional) Enter Comments.
d.
Check that the Experiment Type, Chemistry, and Ramp Speed are correct for the assay that you are running.
e.
Click Next.
3.
In the qPCR Method pane of the Setup tab:
a.
Check that the reaction volume and cycling conditions are correct for the assay that you are running.
b.
Click Next.
4.
In the Plate Setup pane of the Setup tab:
a.
Enter Samples, Targets, and standards.
b.
Click Next.
The Run tab opens.
Prepare the samples according to instructions in the user guide for the kit that you are using, then immediately load the plate into the instrument.
(Create
Load the plate in the instrument
CAUTION! Use optical at caps for tubes. Rounded caps
can damage the heated cover.
Load the plate.
1.
Touch to eject the instrument drawer.
2.
Load the plate onto the plate adapter so that:
• Well A1 of the plate is in the top-left corner of the plate adapter.
• The barcode faces the front of the instrument.
IMPORTANT! The instrument should be used by trained operators who have been warned of the moving parts hazard.
Note: Do not remove the black plate adapter before loading a plate or tube strips. If used, tube strips can t loosely in the adapter, but the heated cover applies the appropriate pressure to seat the tube strips securely in the adapter.
CAUTION! PHYSICAL INJURY HAZARD. During
instrument operation, the plate temperature can reach 100°C. Allow it to cool to room temperature before handling.
3.
Touch to close the instrument drawer.
Start the run
Start the run in the AccuSEQ™ Software.
Option
If the experiment is open Click Start Run.
If the experiment is closed
1. Open the experiment.
2. Click the Run tab.
3. Click Start Run.
Description
A message stating Run has been started successfully is displayed when the run has started.
2
AccuSEQ™ Real‑Time PCR Software v3.1 Quick Reference