Thermo Sterilemax 1277 User manual

Sterilemax
Table Top Steam Sterilizer
Operation Manual
Series 1277
LT1277X1 • 5/23/07
Model Numbers Voltage
ST75925 120V
Serial Number
Sterilemax Table Top Sterilizer
2
Table of Contents
Introduction..........................................................................................................................................................5
Safety Information ..............................................................................................................................................6
Warnings ..........................................................................................................................................................6
Specifications ......................................................................................................................................................9
Trays ................................................................................................................................................................9
Dimensions ......................................................................................................................................................9
Shipping Weight ..............................................................................................................................................9
Electrical ..........................................................................................................................................................9
Standards ........................................................................................................................................................9
Reservoir Water ..............................................................................................................................................9
Mounting ..........................................................................................................................................................9
Controls............................................................................................................................................................9
Environmental Conditions ..............................................................................................................................10
Declaration of Conformity ..............................................................................................................................10
Description ........................................................................................................................................................11
Operational Programs ....................................................................................................................................11
Program Summary Chart ..............................................................................................................................12
Controls..........................................................................................................................................................13
Sterilizer Features..........................................................................................................................................14
Installation ........................................................................................................................................................16
Location..........................................................................................................................................................16
Unpacking ......................................................................................................................................................16
Electrical Connections ..................................................................................................................................17
Optional Printer ..............................................................................................................................................17
Installation Instructions ..................................................................................................................................18
Altitude Adjustment............................................................................................................................................20
Display Parameter Setup ..................................................................................................................................21
Setting Month/Date/Year/Hour/Minute ..........................................................................................................21
Changing Temperature Units ........................................................................................................................22
Changing Pressure Units ..............................................................................................................................22
Changing Printer Mode ..................................................................................................................................23
Sterilization Procedures ....................................................................................................................................24
Materials Appropriate For Steam Sterilization ..............................................................................................24
Instrument Cleaning Prior To Sterilization......................................................................................................25
Preparation of Carbon Steel Instruments ......................................................................................................26
Preparing Items for Sterilization ....................................................................................................................26
Loading the Sterilizer Chamber ....................................................................................................................31
Handling Sterilized Materials ........................................................................................................................32
Recommended Monitoring ................................................................................................................................33
Process Monitoring ........................................................................................................................................33
Biological Indicators ......................................................................................................................................34
Operation ..........................................................................................................................................................36
Initial Startup ..................................................................................................................................................36
Outline of Operation ......................................................................................................................................37
Details of Operation ......................................................................................................................................37
Drying Sterilized Loads ..................................................................................................................................38
Optional Printer ................................................................................................................................................39
Connecting and Starting the Printer ..............................................................................................................39
3
TABLE OF CONTENTS
Control Panel ................................................................................................................................................39
Installing the Paper Roll ................................................................................................................................40
Printer Operation............................................................................................................................................40
Printer Tape Description ................................................................................................................................41
Printer Record Storage ..................................................................................................................................42
Special Operating Instructions ..........................................................................................................................44
Emergency Off (Cycle Abort) ........................................................................................................................44
Resetting Automatic OTP ..............................................................................................................................44
Erratic Controls Display or Operation ............................................................................................................45
Power Failure During Sterilizing Cycle ..........................................................................................................45
Problem Solving Chart ..................................................................................................................................46
Maintenance Schedule Chart ........................................................................................................................48
Maintenance ..................................................................................................................................................49
Daily Maintenance ........................................................................................................................................49
Weekly Maintenance ......................................................................................................................................49
Monthly Maintenance ....................................................................................................................................50
Emptying Collection Bottle ............................................................................................................................51
Cleaning Non-Recirculating Water Cooling Tank ..........................................................................................52
Changing Door Gasket ..................................................................................................................................53
Figures ..........................................................................................................................................................54
Parts List ........................................................................................................................................................57
Ordering Procedures......................................................................................................................................59
Decontamination Statement..............................................................................................................................59
One Year Limited Warranty ..............................................................................................................................60
4
Warning
Do not use this equipment for any purpose other than its intended use, described in this manual.
Do not use the sterilizer to process volatile substances or materials which could release toxic or explosive substances.
Introduction
This Thermo Scientific Sterilemax Table Top steam sterilizer has been designed for use in medical and dental offices, uni­versity laboratories, clinics, hospitals and other locations where a variety of materials require sterilizing.
Your sterilizer can be used for wrapped or unwrapped instruments, packs, liquids, test tubes, petri dishes and media.
The optional Impact printer provides a permanent record of actual exposure times and temperatures.
General Usage
Do not use this equipment for any purpose other than its intended use, described in this manual.
5
Safety Information
Alert Signals
Warning
Warnings alert you to a possibility of personal injury.
Caution
Cautions alert you to a possibility of damage to the equipment.
Note
Notes alert you to pertinent facts and conditions.
Hot Surface
Hot surfaces alert you to a possibility of personal injury if you come in contact with a surface during use or for a period of time after use.
Warnings
1. Do not use this equipment for any purpose other than its intended use, described in this manual.
2. Use only self-venting, automatic sealing stoppers when processing liquids or media. Use of other stop­pers or seals, or lack of stoppers, will create the pos­sibility of burns or injury from boiling liquids and exploding flasks.
3. Refer servicing to qualified personnel.
4. Use caution when opening the sterilizer door. Stand to the side behind the door and open door slowly and partially to protect yourself from escaping steam or boiling liquids and exploding flasks.
5. Sterilizer and trays will be hot after cycle is run. Always wear protective gloves and apron when removing a processed load. Protective gloves and apron should also be worn when reloading sterilizer following previous operation.
6. Do not attempt to open sterilizer door until the display reads "Cycle Complete". Pressure within the cham­ber could cause the door to open with extreme force, possibly causing personal injury.
7. Ensure that the sterilizer is cool before cleaning to avoid burns.
8. Pressurized steam will be discharged when the relief valve is opened. To prevent burns, place a steam barrier (such as a rolled towel) around the opening in
Consult Instructions for Use
This symbol appears on the product to refer you to the instruction manual for further information.
6
the back of the unit. To prevent burns, wear gloves or use an extension device if it becomes necessary to operate the pull ring.
SAFETY INFORMATION
9. Do not use the sterilizer to process volatile sub­stances or materials which could release toxic or explosive substances.
10. Connect only to an electrical circuit of the appropri­ate voltage and load rating.
11. Do not overfill reservoir. To avoid slippery condi­tions and possible load recontamination, immedi­ately wipe up all spillage resulting from overfilling chamber or reservoir.
12. Do not add water to reservoir while a sterilization cycle is in process. When the cycle ends and steam vents back into the reservoir, a sudden overflow of hot water could occur, possibly result­ing in burns to personnel.
13. To prevent collection of mineral deposits and corro­sion of chamber components, use distilled or low quality deionized water only. Clean chamber after each use if sterilizing saline solutions.
14. Processing goods at less than recommended time/temperature could result in unsterile goods.
15. Reprocess your load when the sterilizing cycle has been terminated prematurely. The load may not be sterile when the sterilizing cycle has been terminat­ed prematurely.
16. Open door slowly at the end of a liquid sterilization cycle. Do not allow hot bottles to be jolted. This can cause bottle explosions. Do not move bottles if any boiling or bubbling is present.
17. Allow bottles to cool before attempting to move them from sterilizer shelf or tray(s) to the storage area.
18. If media is processed, bottles and tubes should contain no more than 3/4 the total volume of the container. When processing water bottles and test
7
SAFETY INFORMATION
tubes, the bottles and tubes should contain no more than 3/4 the total volume of the container. Chamber should be cleaned daily when process­ing media.
19. Keep patients at least six (6) feet away from the sterilizer when it is in use.
20. Do not operate this sterilizer in the presence of flammable or explosive substances (such as anes­thetics). A fire or explosion could occur.
21. Never use a wire brush, steel wool, abrasive mate­rial, or chloride-containing products to clean door and chamber assembly.
22. Use a process monitor in every cycle to ensure exposure to minimum sterilizing conditions. Use a biological indicator once a week to verify your ster­ilization process.
23. Dental handpieces should be run only on optional cycle at 132°C for 10 minutes. Check manufactur­er's recommendations for sterilization parameters.
8
Specifications
Trays
Tray Dimensions
Large - 15 1/2" x 9 3/16"
(39.4 x 23.3 cm)
Small - 15 1/2" x 6"
(39.4 x 15.2 cm)
Number of Trays
The sterilizer is supplied with two trays, 1 large and 1 small, and one rack.
Dimensions
Capacity
34 Liters (Vessel) (1.2 cubic feet)
7.0 Liters (Reservoir)
Overall Dimensions
W x D x H Inches(cm) 22 7/16" x 24 1/2" x 17" (57.0 x 62.2 x 43.2) Chamber (Dia. x Depth) 12" X 18" (30.5 x 45.7)
Electrical
Heater Power
1500 W
Power requirements
120V, 50/60 Hz, 12.5 Amps 220V, 50/60 Hz, 6.5 Amps
Standards
ASME UL/cUL to UL3101-1
Reservoir Water
Distilled or low quality deionized Max. 1.5 Megohm/cm Min. .5 Megohm/cm
Mounting
Benchtop only
Controls
Fixed Cycles
- Liquid
- Packs
- Wrapped
- Unwrapped
Optional Cycle
Note
You can examine the pressure vessel rating plate for the sterilizer chamber by opening the panel on the back cover of the sterilizer, then removing the small plug of insulation on the back of the ster­ilizer chamber.
Shipping Weight
85 lbs. (38.5 kg)
9
SPECIFICATIONS
Environmental Conditions
Operating: 4°C - 40°C; 20% - 80% relative humidity, noncondensing. Installation Category II (overvoltage) in accordance with IEC 664. Pollution Degree 2 in accordance with IEC 664. Altitude limit: 3,650 meters.
Storage: -25°C - 65°C; 20% - 80% relative humidity.
Declaration of Conformity
(for 230 volt, CE models only)
We hereby declare under our sole responsibility that this product conforms with the technical requirements of the following standards:
EMC: EN 61000-3-2 Limits for harmonic current emissions
EN 61000-3-3 Limits for voltage fluctuations and flicker EN 61326-1 Electrical equipment for measurement,
Safety: EN 61010-1 Safety requirements for electrical equipment
EN/EIC 61610-2-010 Part II: Particular requirements for
EN61010-2-041 Part II: Particular requirements for autoclaves using steam
and meets the requirements of the Medical Device Directive 93/42/EEC, and its relevant transition into the national laws of the member states into which we place the device(s).
The authorized representative located within the European Community is:
Thermo Fisher Scientific 419 Sutton Road Southend On Sea Essex SS2 5PH United Kingdom
0086
control, and laboratory use; Part I: General Requirements.
for measurement, control, and laboratory use; Part I: General Requirements.
laboratory equipment for the heating of materials.
for the treatment of medical materials, and for laboratory processes.
Copies of the Declaration of Conformity are available upon request
10
Description
Operational Programs
The sterilizer has four preset cycles which have been opti­mized to provide excellent results for most applications. In addition, there is one optional cycle that allows you to set your own sterilization parameters. You access the pre-pro­grammed cycles by pushing one of the four green fixed-cycle buttons. The four preset sterilization programs are:
Unwrapped Goods
For unwrapped instruments and items made of metal,
glass or plastic. Parameters: 3 minutes at 135°C.
Wrapped Goods
For wrapped packs of clothes, bandages, pads, fabrics, rubber tubing and metal instruments. Parameters: 10 minutes at 135°C.
Note
The liquid cycle is not designed for sterilization of liquids in a clinical application. Do not prepare sterile liquids for in vivo applications with the Sterilemax system.
Dental handpieces should be run in the optional cycle at 132°C for 10 minutes. Refer to manufactur­er's recommendations for contra indication of sterilization parame­ters.
When sterilizing handpieces, refer to the manufacturer's specifica­tions on handpieces for contra indication of sterilization parame­ters.
Liquids
For distilled water, media solutions and other liquid pre­pared in closed bottles or flasks. Parameters: 15 minutes at 121°C.
Packs
For packs of instruments and towels and other packed goods. Parameters: 30 minutes at 121°C.
11
DESCRIPTION
margorPerutarepmeTemiTemiTgniyrDdeziliretSeBoTsmetI
sdiuqiL
C°121
)F°052(
setunim51enoN
,snoitulosaidem,retawdellitsiD
desolcnideraperpdiuqilrehto
retaergton(sksalfroselttob
lacinilcroftoN.)sretil2naht
.snoitacilppa
depparwnU
C°531
)F°572(
setunim3
:tluafeD
,setunim51
tesebnac
99ot1morf
setunim
latem;stnemurtsnidepparwnU
-taeH.stcejbocitsalpdnassalg
hcihwgnibutrebburtnatsiser
lacigrusnidesuebtonlliw
smetirehtoynA.snoitacilppa
setunim3rofC°531hcihwrof
.etairporppasi
depparW
C°531
)F°572(
setunim01
:tluafeD
,setunim51
tesebnac
99ot1morf
setunim
,segadnab,shtolcdepparW
depparwylesool,scirbaf,sdap
fosreyal,stnemurtsnilaudividni
ybdetarapesstnemurtsni
esoolfosyartdepparw,cirbaf
nistnemurtsni,stnemurtsni
rebburtnatsiser-taeh,sgab
rofsmetirehtoynA.gnibut
sisetunim01rofC°531hcihw
.etairporppa
skcaP
C°121
)F°052(
setunim03
:tluafeD
,setunim51
tesebnac
99ot1morf
setunim
fospuorG.slewotfoskcaP
yllaicremmocnistnemurtsni
stnemurtsnI.skcapderaperp
.egarotsdegnolorpottcejbus
rofdepparwsevolglacigruS
rofsmetirehtoynA.noitazilirets
sisetunim03rofC°121hcihw
.sdiuqiltpecxe,etairporppa
lanoitpO
denifedresU
C°001morf
C°631ot
otF°212(
)F°772
denifedresU
99ot1morf
setunim
denifedresU
99ot1morf
setunim
resuehtotetairporppasmetI latneD.sretemarapdenifed ninurebdluohsseceipdnah
01rofC°231taelcycsiht
.setunim
Note
When sterilizing handpieces, refer to the manufacturer's specifications on handpieces for contra indication of sterilization parameters.
Program Summary Chart
12
DESCRIPTION
Controls
All cycles and programming are selected and set in using the front touchpad selector buttons.
Power Switch
The Power Switch turns power to the unit on and off. When you turn the power on, the display will be set to select a steril­ization cycle. It is located under the right front corner of the sterilizer.
Cycle Buttons (Green)
These buttons allow you to run the preprogrammed cycles at the touch of a button. See Operational Programs.
Optional Cycle (Yellow)
This parameter allows you to select different time and tem­perature levels for your cycles than those provided with our preset programs. An optional cycle can be run as either a liq­uid or hard goods cycle. All material run in this cycle must be validated by the user.
Note
The "Select" button is used by ser­vice personnel to adjust the oper­ating parameters of the sterilizer.
Up/Down Arrows (Yellow)
These allow you to set the time and temperature in an option­al cycle. They are also used for scrolling through other func­tions.
Start Cycle (Dark Green)
Pressing the "Start Cycle" Switch activates the currently selected processing cycle. It is also used for scrolling through other functions.
Stop (Red)
This control is used to terminate any cycle. It is also used for scrolling through other functions.
13
DESCRIPTION
Sterilizer Features
Optional Printer
The optional printer available is an impact printer that records each cycle's performance. See page 39 for more information.
Pressure Monitor
The pressure monitor indicates the pressure inside the chamber. This monitor's readings can be found on the display as each cycle is running.
Door Security
While the chamber is under pressure, the door is held securely closed with three separate locking systems. First, there is a lock­ing handle that you must tighten to hold the door fully closed. The locking handle turns clockwise to close and counterclockwise to open. Second, there is a pressure actuated locking pin which automatically secures the door, preventing it from being inadver­tently opened while there is pressure in the chamber. Finally, there is a backup catch which will prevent the door from opening fully even if there is a small amount of pressure in the chamber. You release this catch with a lever which is located in the lower left side of the door. To fully open the door, pull the door release lever toward you.
14
Optional Non-Recirculating Tank
The optional non-recirculating tank is used to collect the remain­ing cycle water after a cycle has completed. This prevents any debris from the cycle to go back into the fresh water reservoir.
Door Switch
Actuated by the door, the door switch prevents operation of the sterilizer if the door is not fully closed.
Pressure Relief Valve
This valve protects the coded vessel from pressures exceeding the pressure rating of the vessel. The valve opens and releases steam in the event that pressure in the vessel exceeds 45 PSIG (310 kPa). The pressure line is located in the back of the unit. An
Warning
Pressurized steam will be discharged when the relief valve is opened. To prevent burns, wear gloves or use an extension device if it becomes necessary to oper­ate the pull ring.
DESCRIPTION
access hole is located in the back panel to allow you to peri­odically check relief valve operation.
Reservoir Drain
A drain assembly is provided to empty the reservoir. The reservoir drain assembly and plug are located at the lower right of the chamber opening on the front of the sterilizer (as you face the unit). The drain assembly attaches to the drain connector so a hose can be attached for draining the steriliz­er reservoir into a container.
Trays and Tray Rack
A rack for the inside of the sterilizer is provided to hold the instrument trays. The rack slides out of the chamber to allow periodic cleaning of the chamber and rack.
Three large trays and one small tray can be placed into the rack or the rack can be used with only one or two trays or no trays. Tall items can be sterilized by removing both trays and placing the objects on the bottom surface of the rack. One large and one small tray are provided with the unit. See "Loading the Sterilizer Chamber" on page 31 of this manual.
15
Installation
Caution
The surface beneath the sterilizer could reach 100°F (38°C).
Location
For optimum performance of your sterilizer, the sterilizer should be installed in conformance to the following parame­ters:
Surface
For proper operation, the sterilizer must be placed on a level, heat and water resistant surface. If the surface is not level, the water in the sterilizer will be unevenly distributed, possibly resulting in a sterilizer malfunction. Do not adjust position of the feet on the sterilizer.
Clearances
Allow at least 3" (75mm) between the sterilizer and walls on the sides and back for air circulation. Also allow at least 8" (204mm) between the top of the sterilizer and any overhang­ing cabinet above it. Overhangs should not project more than 8" (204mm) over the rear of the sterilizer.
Surroundings
If the sterilizer will be operated continuously, locate it in an area where its steam emissions will not damage equipment or materials.
Unpacking
Remove the sterilizer carefully and inspect it for any shipping damage. Do not throw away packing materials until you inspect the unit.
1. Turn the door handle counterclockwise until it is completely loose and move the locking mechanism to the left. Pull on the latch release lever built into the left side of the sterilizer door and carefully open the sterilizer door.
2. Remove all packing materials and accessories from the chamber.
3. Remove the trays and rack assembly from the chamber.
16
Warning
Connect only to an electrical cir­cuit of the appropriate voltage and load rating.
INSTALLATION
4. Check for any foreign material in the chamber. Wipe clean with a damp, lint-free cloth if neces­sary.
Electrical Connections
We recommend that you use a separate (dedicated) circuit for this sterilizer. Ensure that the circuit is rated for the load. (See Electrical Specifications.)
1. Connect power cord to power entry module on back of unit.
2. Plug power cord into a grounded outlet of the appropriate voltage and load rating.
Optional Printer
See section "Optional Printer" for installation, startup and operation.
17
Optional Non-Recirculating Tank
NOTE: If you did not purchase an optional non-recirculation tank, please proceed to Altitude Adjustment
Installation Instructions
Cap
1. To connect the non-recirculating water accessory to the back of the sterilizer.
a. Unplug the sterilizer power cord. Allow the unit to
cool to room temperature.
b. Move the sterilizer to gain access to the rear of
the unit.
c. Unscrew the cap from the nonrecirculating dis-
charge port of the sterilizer (located on the rear of the sterilizer).
d. Attach the discharge hose from the non-recircu-
lating discharge port.
Removing Cap
Tighten nut fin­ger tight plus 1/4 turn with wrench.
1/4 turn
After discharge hose assembly from cooling tank to non-recirculating discharge port.
e. Tighten the nut finger tight plus a quarter turn with
a wrench.
f. Attach the other end of the discharge hose to the
fitting on the lower left (rear) of the non-recirculat­ing cooling tank. Follow the same procedure as steps d and e above.
Caution
Be sure connection nut is started straight. Cross threading will cause leaks and dam­age the nut and fitting.
Discharge Hose Connection
18
Improper Discharge Hose Connection
Cap Replacement
OPTIONAL NON-RECIRCULATING TANK
2. Install the cap removed in step c. To the fitting located in the reservoir. See Cap Replacement
3. Fill the cooling tank with distilled or deionized water until the water begins to overflow from the cooling tank into the collection bottle (or into the drain if the overflow tube is installed).
4. Plug in the power cord.
Warning
Non-recirculating option must be installed before proceeding.
Remove Cap
Optional Overflow Drain
Filling Cooling Tank
Full mark
Fill with distilled or deionized water until it overflows into the collection bottle (approx. 3 gal.)
Collection Bottle
Caution
The non-recirculating tank (cooling tank) must be connected to the sterilizer before filling with distilled or deionized water.
Caution
Do not operate sterilizer unless the non-recirculating tank is filled with distilled or deionized water and attached to the sterilizer.
19
Altitude Adjustment
For proper operation, the sterilizer program accounts for the altitude in controlling certain functions. The sterilizer is shipped with a factory-set altitude value suitable for all alti­tudes up to an altitude of 1000 feet above sea level. If your altitude is above 1000 feet above sea level, you will need to enter a new value for the altitude adjustment. To change the altitude adjustment value:
1. At the "Select Cycle" screen, press the "Select" key and then the up arrow key. The sterilizer will display a "Configuration Info" screen, then it will automatically change to displaying the "Change Time" screen.
2. At the "Change Time" screen, press the "Select" key and then the "Liquids" key.
3. At the "Altitude Adjustment" screen, press the up arrow to select "Yes".
4. At the "Current Altitude" screen, press the up and down arrows to set the value for your location's altitude to the nearest one hundred feet.
5. After you have set your altitude value, press the "Start" key to return to the "Select Cycle" screen.
20
Display Parameter Setup
Note
Sterilemax sterilizers are shipped with the display units set as fol­lows:
120 volt units - °C, PSI
Setting Month/Date/Year/Hour/ Minute
To set the current date and time:
1. From the "Select Cycle" screen, press the "Select" key and then the up arrow key.
2. Wait for the sterilizer to display the "Change Time" Screen. Press the up arrow key to select "Yes."
220 volt units - °C, BAR
If your unit is not set to these fac­tory settings or if you require other settings, refer to "Changing Temperature Units" and "Changing Pressure Units" to set the units to your requirements.
3. At the "Set Current Month" screen, press the up or down arrow key to set the current month. Press the "Start" key when the current month is set.
4. At the "Set Current Date" screen, press the up or down arrow key to set the current date. Press the "Start" key when the current date is set.
5. At the "Set Current Year" screen, press the up or down arrow key to set the current year's final two digits. Press the "Start" key when the current year is set.
6. At the "Change Hour" screen, press the up or down arrow key to set the current twenty-four hour clock hour. Press the "Start" key when the current hour is set.
7. At the "Change Minute" screen, press the up or down arrow key to set the current minute. Press the "Start" key when the current minute is set.
8. At the "Change Temperature Units" screen, press the down arrow key to select "No."
Note
If you desire to change the tem­perature units, press the up arrow to select "Yes", then refer to "Changing Temperature Units."
9. At the "Change Pressure Units" screen, press the down arrow key to select "No".
10. At the "Change Printer Mode" screen, press DOWN ARROW key to select "NO" and return to the "Select Cycle" screen.
21
DISPLAY PARAMETER SETUP
Note
If you desire to change the pres­sure units, press the up arrow to select "Yes", then refer to "Changing Pressure Units."
Note
If you desire to change the pressure units press the UP ARROW to select "YES", then refer to "Change Printer Mode."
Changing Temperature Units
To change the units in which the chamber temperature is dis­played on the digital display and printer tape (°C or °F):
1. From the "Select Cycle" screen, press the "Select" key and then the up arrow key.
2. Wait for the sterilizer to display the "Change Time" Screen. Press the down arrow key to select "No."
3. The sterilizer will then display the "Change Temperature" screen. Press the up arrow key to select "Yes."
4. At the "Temperature Units" screen, press the up arrow key to toggle between °C and °F.
Note
If you desire to change the pres­sure units, press the up arrow to select "Yes", then refer to "Changing Pressure Units."
5. After you have selected your desired temperature units, press the "Start" key.
6. At the "Change Pressure Units" screen, press the DOWN ARROW KEY to select "NO".
7. At the "Change Printer Mode" screen, press the DOWN ARROW KEY to select "NO" and return to the "Select Cycle" screen.
Changing Pressure Units
To change the units in which the chamber pressure is displayed on the digital display and the printer tape (PSI, BAR or KPA):
1. From the "Select Cycle" screen, press the "Select" key and then the up arrow key.
2. Wait for the sterilizer to display the "Change Time" Screen. Press the down arrow key to select "No."
3. The sterilizer will then display the "Change Temperature" screen. Press the down arrow key to select "No."
22
DISPLAY PARAMETER SETUP
4. At the "Change Pressure Units" screen, press the up arrow key to select "Yes."
5. At the "Pressure Units" screen, press the up arrow to cycle between the three available units.
6. When you have selected your desired pressure units, press the "Start" key to enter units.
7. At the "Change Printer Mode" screen, press the DOWN ARROW key to select "NO" and return to the "Select Cycle" screen.
Changing Printer Modes
This display is used to change the mode serial port of use for optional printers from USED/UNUSED:
1. From the "Select Cycle" screen, press the "Select" key and then the UP ARROW key.
2. Wait for the "Change Time" screen. Press the DOWN ARROW key to select "NO."
3. The sterilizer will then display the "Change Temperature" screen. Press the DOWN ARROW key to select "NO."
4. At the "Change Pressure Units" screen, press the DOWN ARROW key to select "NO."
5. At the "Change Printer Mode" screen, press the UP ARROW to select "YES."
6. At the "Printer Mode" screen, press the UP ARROW to toggle between the available modes.
7. When you have selected your desired mode of operation, press the "Start" key to return to the "Select Cycle" screen.
23
Sterilization Procedures
Warning
Do not attempt to sterilize volatile or toxic substances. Serious injury or death could result.
Caution
Do not attempt to sterilize any object composed of materials which cannot withstand high tem­peratures (135°C or greater).
Materials Appropriate For Steam Sterilization
• Straight stainless steel instruments
• Surgical stainless steel hinged instruments
• Pouches of instruments
• Carbon steel instruments (see special preparation instructions)
• Air powered instruments made to be autoclaved
• Needles
• Syringes
• Treatment trays
• Heat resistant plastic items
• Rubber gloves
• Heat resistant rubber tubing
24
• Heat resistant rubber sheeting
• Glass slabs and beakers
• Gauze
• Sutures
• Linen packs
• Liquids, non-clinical
• Other heat-resistant, immersible items
Always check with the manufacturer of your supplies for specific instructions for sterilization
.
Warning
Thoroughly clean instruments before placing them in the steriliz­er. Sterilizing items contaminated with blood or debris will impede sterilization and may result in unsterile instruments, which may cause personal injury or death.
Warning
Always wear protective gloves when handling soiled instruments to protect yourself from the possi­bility of serious infection. Clean gloved hands with a germicidal cleaner when finished handling soiled instruments.
STERILIZATION PROCEDURES
Instrument Cleaning Prior To Sterilization
Transport soiled instruments to your cleanup area on a tray. Sort out any instruments unsuitable for sterilization (see list on page 24).
Follow manufacturer's recommendations for cleaning and preparing any particular item.
In the absence of manufacturer's specific instructions, clean instruments as follows:
1. Rinse instruments thoroughly with a hard stream of water to remove debris.
2. Sort instruments by their metallic composition. Do not mix instruments of different metals, as instru­ment damage will result.
3. Using either a fresh solution of detergent and dis­tilled water or a germicide solution, wash instru­ments in an ultrasonic cleaner for five to ten min­utes. Clean all instruments in an open position.
Caution
Sterilizing items contaminated with blood or debris may cause staining or damage to the instruments or the sterilizer.
Caution
Use of soaking solutions contain­ing phenols or quaternary ammo­nium compounds may cause cor­rosion of the instruments, trays and the stainless steel chamber.
Use a detergent or germicide with a neutral pH (7) specifically designed for use in an ultrasonic clean­er.
Use new ultrasonic cleaning solution each day.
4. After cleaning, rinse instruments thoroughly in dis­tilled or deionized water. Inspect to ensure that all debris has been removed. Repeat cleaning as necessary.
If distilled or deionized water is unavailable and instruments are rinsed in tap water instead, all instruments should be dried before sterilizing to prevent staining.
5. Refer to manufacturer's recommendations for lubri­cating instruments after ultrasonic cleaning. Use a silicon lubricant which will not interfere with steril­ization.
25
STERILIZATION PROCEDURES
Note
Do not place carbon steel instru­ments directly onto stainless steel trays. Place an unbleached towel or paper wrap on the tray and place the instruments on the towel or paper.
When wrapping instruments for sterilization, use a material which promotes drying.
Preparation of Carbon Steel Instruments
Carbon steel instruments must be carefully prepared prior to sterilization to prevent rusting. After cleaning according to the manufacturer's recommendations or the methods presented in Instrument Cleaning Prior To Sterilization, prepare car- bon steel instruments as follows:
1. Prepare a 2% solution of sodium nitrate in water (approximately one table spoon in a quart of water). Completely immerse the instruments in the solution and allow them to soak for three minutes.
2. Remove the instruments and package them for sterilization or place on trays if sterilizing unwrapped. Do not rinse or wipe instruments
prior to sterilization.
Warning
Do not overload the trays. Overloading trays may result in inadequate sterilization and unsterile instruments.
Warning
Monitor the sterilization process with process monitors and biological indi­cators as described in Recommended Monitoring (p. 33).
Preparing Items for Sterilization
General Preparation Guidelines
1. Clean items thoroughly before placing them in the sterilizer. Follow good laboratory or medical proce­dures for cleaning and decontamination of items.
2. Place absorbent paper or linen towel in the bottom of trays before use.
3. Unwrapped goods should only be processed on trays. Do not place items on the rack or the bottom of the sterilizer.
4. Liquids must not exceed 75% of capacity of flask. Use self-venting caps.
5. Open all hinged instruments.
26
Caution
Do not mix different metals in the same pack or wrapping or loose on the same tray. Instrument dam­age will result.
STERILIZATION PROCEDURES
6. Remove the covers from empty containers. Stand the containers and covers on their sides.
7. Turn curved vessels upside down so no water will collect in the vessel.
8. Place all sharps (knives, scissors, etc.) so that they do not touch during sterilization. Cotton or gauze may be used to isolate and protect the sharp edges and the smaller instruments.
9. Disassemble syringes if possible.
10. Do not plug syringes or needles.
Warning
Unwrapped items cannot be con­sidered sterile once removed from the sterilizer chamber.
Do not overload the trays. Overloading trays may result in inadequate sterilization and unsterile instruments.
Note
Unwrapped trays are used for sterilizing nonsurgical instruments, canisters and other items to prevent trans­mitting infectious disease.
Items to be Sterilized Loose and Unwrapped
Items to be sterilized loose and unwrapped should be distrib­uted in a single layer on the bottom of a lined, perforated tray. Do not overload tray; allow space around instruments for steam circulation.
Metal and glass canisters and containers should be placed upside down on a perforated tray. Make sure containers are placed so that they will drain, or they will not dry sufficiently to ensure sterilization.
Appropriate Wrapping Materials for Steam Sterilization (Partial List)
Muslin (2 layers), Nylon bagging material, Commercial autoclave paper (must be as permeable as muslin), Permeable paper and plastic bags, Cloth or paper-covered trays
Materials Inappropriate for Wrapping for Steam Sterilization (Partial List)
Canvas, Impermeable plastics, Aluminum foil, Sealed jars, canisters or tubes.
27
STERILIZATION PROCEDURES
Warning
Do not wrap packs tightly and do not over fill autoclave bags. Steam must penetrate through entire pack. Inadequate sterilization may result from improper packaging.
Do not seal with staples, pins or other methods which will puncture the wrapping material. If wrapping material is punctured, sterility can­not be assured once items are removed from the sterilizer cham­ber.
Items to be Sterilized Wrapped
Cleaned items may be wrapped individually in autoclave paper or placed in commercially prepared autoclave bags. Do not wrap tightly. Seal with autoclave tape or heat sealer.
Cleaned items may also be sterilized as a group of instru­ments wrapped together. You may desire to wrap a group of similar instruments together, or you may desire to wrap together the instruments needed for a particular procedure. Wrap the instruments loosely in 2 to 4 layers of autoclave paper or muslin or place them in autoclave bags. Seal with autoclave tape or heat sealer.
Place wrapped items on a perforated tray. Do not overload tray; allow space around wrapped items for steam circulation.
When using commercial sterilization bags, follow the manufacturer's instructions for proper use.
Do not overload the trays. Overloading trays may result in inadequate sterilization and unsterile instruments.
Note
If you desire to include a large number of instruments in a single package, see Items to be
Sterilized in Packs.
Appropriate Wrapping Materials for Steam Sterilization (Partial List)
Muslin (2 layers), Nylon bagging material, Commercial autoclave paper (must be as permeable as muslin), Permeable paper and plastic bags, Cloth or paper-covered trays
Materials Inappropriate for Wrapping for Steam Sterilization (Partial List)
28
Canvas, Impermeable plastics, Aluminum foil, Sealed jars, canisters or tubes.
STERILIZATION PROCEDURES
Warning
Do not seal with staples, pins or other methods which will puncture the wrapping material. If wrapping material is punctured, sterility can­not be assured once items are removed from the sterilizer cham­ber.
Do not wrap items tightly and do not over fill autoclave bags. Steam must penetrate through entire wrapped group. Inadequate steril­ization may result from improper wrapping.
Do not overload the trays. Overloading trays may result in inadequate sterilization and unsterile instruments.
Items to be Sterilized on Wrapped Trays
Place cleaned items loose in the bottom of a lined, perforated tray. Additional layers of instruments may be added, separat­ed by fabric. Do not overload the tray; steam must be able to penetrate through the entire tray. Wrap the tray in 2 to 4 lay­ers of muslin or other appropriate wrapping material. Seal with autoclave tape.
Items to be Sterilized in Packs
Use packs to sterilize a number of instruments together, to sterilize textile products and to sterilize surgical gloves. When
using commercial sterilization packs, follow manufactur­er's instructions for proper use.
Instruments
Loosely package cleaned instruments into packs of not more than 10 instruments each. Instruments which might nest together should be separated by a layer of absorbent material and placed in the pack in a manner that will allow water to flow out of them. Seal with autoclave tape or heat sealer.
Textile Products
Launder textile products before sterilization.
Loosely roll or fold textile products. Place in commercially prepared packages or wrap in 2 to 4 layers of muslin or other appropriate wrapping material. Seal with autoclave tape or heat sealer.
Surgical Gloves
Disposable gloves should not be sterilized.
Clean and dry gloves. Place a piece of muslin or other
Appropriate Wrapping Materials for Steam Sterilization (Partial List)
Muslin (2 layers), Nylon bagging material, Commercial autoclave paper (must be as permeable as muslin), Permeable paper and plastic bags, Cloth or paper-covered trays
Materials Inappropriate for Wrapping for Steam Sterilization (Partial List)
Canvas, Impermeable plastics, Aluminum foil, Sealed jars, canisters or tubes.
29
STERILIZATION PROCEDURES
Caution
Do not use detergents with a high chlorine or phosphate content or chlorine bleach when laundering items prior to sterilization. Damage to instruments and to the sterilizer chamber could result.
Warning
Do not overload tray; allow space around packs for steam circula­tion. Pack density should not exceed 50% of the capacity of the tray. Metal or glass should not be used to separate packs, as these will inhibit drying. Overloading the tray or failure to allow the packs to dry properly can result in non-ster­ile goods.
absorbent material into the central area of the glove. Place a strip of muslin or other absorbent material around the cuff and fold the cuff back over it. Wrap the gloves in muslin or other appropriate wrapping material. Seal with autoclave tape.
Place packs on a perforated tray. Place packs of gloves so that the glove stand on end in the sterilizer. Load packs side­by-side; do not stack packs.
Preparing Liquids for Sterilization
Place liquids in heat-proof containers. Place containers on perforated tray. Do not fill flask more than 75% full.
Items to be Sterilized in Optional Cycle
Place material to be sterilized in appropriate containers or on trays prior to placement in sterilizer. Follow instructions on display for setting cycle time, temperature and drying para­meters. If running a liquid-optional cycle, ensure that contain­ers are not more than 75% full. User must follow proper safe­ty procedures and sterilization monitoring procedures for the optional cycle. Every optional cycle should be validated by the user.
Dental handpieces should be run at 132°C for 10 minutes.
30
Warning
Do not exceed the sterilizer's rated capacity for liquid sterilization. Non-sterile liquids could result. Fill flasks no more than 75% full. User should validate material in optional cycle.
Note
User should not attempt to sterilize more than 10 handpieces per ster­ilization cycle.
STERILIZATION PROCEDURES
Warning
Use only self-venting, automatic sealing stoppers when processing liquids or media. Use of other stoppers or seals, or lack of stop­pers, will create the possibility of burns or injury from boiling liquids and exploding flasks.
If media is processed, bottles and tubes should contain no more than 3/4 the total volume of the container. When processing water bottles and test tubes, the bottles and tubes should contain no more than 3/4 the total volume of the container. Chamber should be cleaned daily when processing media.
Sterilized liquids are for non-clinical applications.
Caution
Packs must be centered in the chamber. Do not allow packs to touch the sterilizer chamber's walls. Allowing packs to touch the sterilizer chamber's walls may cre­ate an overheat condition.
Loading the Sterilizer Chamber
Loading Guidelines
1. When processing wrapped goods, fabrics and dressings, always place items in one of the three upper positions in the chamber. Processing items in the bottom position could possibly result in wet packs and unsterile goods.
2. Do not exceed the specified loads for a cycle: Unwrapped Goods - no more than 14 lbs Wrapped Goods - no more than 14 lbs in no more than 12 small cassettes or 4 large cassettes Packs - no more than 2.8 lbs and no more than 14 towels Liquids - flasks filled no more than 75% full.
3. Do not overload the sterilizer. Steam penetrates best with loosely packed loads.
4. Do not allow packs or wrappings to touch chamber walls or one another.
5. Leave adequate space between trays to allow steam circulation.
6. Do not stack trays on top of one another.
7. Use a process monitor in each load. We also rec­ommend that you use a biological indicator at weekly intervals. An indicator containing Bacillus stearothermophilus is the proper indicator for steam sterilizers.
31
STERILIZATION PROCEDURES
Warning
Sterilizer and trays will be hot after cycle is run. Always wear protective gloves and apron when removing a processed load. Protective gloves and apron should also be worn when reload­ing sterilizer following previous operation.
Open door slowly at the end of a liquid sterilization cycle. Do not allow hot bottles to be jolted. This can cause bottle explosions. Do not move bottles if any boiling or bubbling is present.
Allow bottles to cool before attempting to move them from sterilizer shelf or tray(s) to the storage area.
Handling Sterilized Materials
Loads may be removed from the sterilizer at the end of the drying cycle. (Liquids may be removed at the end of the steril­ization cycle.) Unwrapped items should be used immediately. Any unwrapped items not used immediately cannot be con­sidered sterile after removal from the sterilizer chamber.
Date all wrapped packages and packs prior to storage. Check the integrity of all wrapping materials prior to opening and using wrapped or packed items; if the wrapping materials are opened or torn, the instruments must be considered potentially unsterile and should be re-sterilized prior to use. Re-sterilize all wrapped packages and packs after one month in storage.
32
Recommended Monitoring
All functions of a sterilizer should be monitored to provide maximum assurance of proper performance.
Note
Refer to the manufacturer's recom­mendations for proper use of biologi­cal indicators.
Process Monitoring
For assurance that minimum sterilizing conditions have been achieved, we recommend the use of a process monitor with each cycle and within each package. This small process monitoring device will provide immediate indication of expo­sure to temperature and of time in the presence of humidity (steam).
Trays
Place a process monitor in the center of each wrapped or unwrapped tray of loose items.
Wrapped Instruments
Place a process monitor inside each package with an instru­ment.
Wrapped Packages of Instruments
Place a process monitor in each package.
Packs
Place a process monitor inside each pack. When sterilizing textile products, place the process monitor in the center of the stack of textile products. When sterilizing gloves, place a process monitor inside a glove inside each pack.
33
RECOMMENDED MONITORING
Note
Refer to the manufacturer's recommenda­tions for proper use of biological indica­tors.
Note
When using a biological indicator for a 135°C cycle, the user must be sure that it has been validated for use at 135°C. We recommend using the Thermo Scientific Biological Indicator (part # AY759X1) for this type of cycle. If any other Biological Indicator is used, the user must ensure that it has been vali­dated for a 135°C cycle.
Biological Indicators
Process monitors do not by themselves prove steriliza­tion has occurred. Biological indicators composed of spores of standard organisms are considered the most reliable means of verifying sterilization. A biological indi­cator functions by testing the sterilizer's ability to kill test organisms of a known high resistance. Indicators con­taining Bacillus stearothermophilus should be used for steam sterilizers. The Joint Commission on Accreditation of Hospitals recommends that biological indicators be used at least weekly to check the effective­ness of your steam sterilizer's sterilization cycle.
When testing the effectiveness of a sterilization cycle, use the appropriate "worst case" test load described for the cycle in this section:
Unwrapped Instruments
Place 100 instruments on a perforated tray lined with a 100% cotton surgical towel. Place a biological indicator in the center of the test tray. Place the test tray in the center rack of the sterilizer chamber. Prepare additional trays in the same manner without biological indicators to bring the total sterilization load to 14 lbs (6.4 kg).
Wrapped Instruments
Load a small cassette with 6 instruments. Place two bio­logical indicators inside the cassette. Wrap the cassette with two layers of appropriate wrapping. Seal with auto­clave tape. Prepare eight other small cassettes in the same manner without the biological indicators. Place the cassette with the biological monitor upright in the center of a tray. Fill the tray with the rest of the cassettes. Place the tray in the center rack of the sterilizer cham­ber.
34
RECOMMENDED MONITORING
Packs
Textiles
Prepare a pack of 14 towels (per AAMI ST-8). Place three biological indicators in the center of the pack, between the sixth and seventh towels. Wrap the pack with two strips of autoclave tape, placing the tape around the width of the pack so that the two strips divide the length of the pack into thirds.
Pouches
Place a biological indicator together with a surgical instru­ment into a self-sealing pouch. Place the pouch with the indicator in the middle of the tray. Place four other surgical instruments into pouches and place them on the tray. Place the tray in the center rack of the sterilizer chamber.
Optional Cycle: For Dental Handpieces Only
Place a biological indicator together with one dental hand­piece in a self-sealing pouch. Place the closed pouch in the front of the sterilizer tray. Place the tray in the lowest level of the chamber rack.
35
Operation
Warning
Keep patients at least six (6) feet away from the sterilizer when it is in use.
Do not operate this sterilizer in the presence of flammable or explo­sive substances (such as anes­thetics). A fire or explosion could occur.
Caution
To prevent collection of mineral deposits and corrosion of chamber components, use distilled or deion­ized water only. See Specifications. Clean chamber after each use if sterilizing saline solutions.
Connect the power cord to the unit and to a grounded outlet of appropriate voltage and rating.
Initial Startup
Power On
Push the Power Switch under the right front corner of the sterilizer to the On position. The sterilizer will not operate unless this switch is in the On position.
Filling Reservoir
Remove the cover to the water reservoir and fill the reservoir with distilled or low quality deionized water (minimum 0.5 Megohm/cm, maximum 1.5 Megohm/cm) until the water level reaches the bottom of the fill line in the reservoir. The reser­voir will hold approximately 7.0 liters (6.3 quarts) of water. The water level should never be higher than the fill line. Do not overfill.
Do not overfill reservoir. To avoid slippery conditions and possible load recontamination, immediately wipe up all spillage resulting from overfilling chamber or reservoir.
Warning
Do not add water to reservoir while a sterilization cycle is in process. When the cycle ends and steam vents back into the reser­voir, a sudden overflow of hot water could occur, possibly result­ing in burns to personnel.
Refilling Reservoir
When the reservoir runs low on water, the low level sensor will alert you of the low water condition. When the display indicates "Reservoir is empty. Fill and restart sterilization cycle," you will need to add water to the reservoir before you run another cycle. You should never need to add more than 5 liters (5.3 quarts) at a time to the reservoir. The water level should never be higher than the fill line. Do not overfill.
You may determine a refill schedule dependent upon the pattern of operation at you facility. The number of cycles you can run before the reservoir needs to be filled will vary according to your frequency of operation, duration between cycles, type of cycles, etc.
36
OPERATION
Outline of Operation
Two basic steps are required to initiate a sterilizing cycle after the load has been placed in the chamber and the door closed and latched:
1. Select the cycle you desire.
2. Follow the displayed directions for the cycle.
Cycle selections cannot be changed once "Start Cycle" has been pressed. To change a setting, the cycle must be terminated by pressing "Stop." When you select a new cycle, the cycle information will be changed automatically.
Details of Operation
Prepare Load
Follow preparation instructions on pages 27-31.
Load the Sterilizer Chamber
You may load wrapped or unwrapped instruments, linen packs, glassware and liquids. Follow loading instructions on page 33.
Close the Sterilizer Door
The door must be fully closed to run a sterilizing cycle. If the door is not fully closed, the sterilizer will stop and indicate, "The door is open. Close door to continue cycle."
Select Processing Cycle
1. Select your required cycle.
2. Set the drying cycle time for pack, wrapped and unwrapped cycles. (See "Drying Sterilized Loads.")
37
OPERATION
Warning
Processing goods at less than rec­ommended time/temperature could result in unsterile goods.
Setting Drying Time
(For Pack, Wrapped and Unwrapped cycles only.)
If drying is desired, set the drying time when you select the sterilization cycle. After you select a Pack, Wrapped or Unwrapped cycle, the cycle parameters will be displayed. After the parameters have been displayed, you will be prompted to enter a figure for drying time. You will find that the default time — displayed at the prompt — is the opti­mal setting for most loads.
If your application or load composition requires a differ­ent drying time to achieve appropriate drying, you can change the drying time setting when the display prompts you. Use the up or down arrow to adjust the displayed dry­ing time until your desired drying time is displayed.
Press "Start Cycle"
Processing will run automatically after you press "Start Cycle."
Warning
Use caution when opening the sterilizer door. Stand to the side behind the door and open door slowly and partially to protect yourself from escaping steam or boiling liquids and exploding flasks.
Do not attempt to open sterilizer door until the display reads "Cycle Complete." Pressure within the chamber could cause the door to open with extreme force, possibly causing personal injury.
Drying Sterilized Loads
At the completion of the sterilization cycle, turn the lock­ing handle counter clockwise until the door is held in place by the backup catch. The door must be opened approxi­mately 1 inch to operate the drying cycle. To prevent con-
tamination of wet packs and wrapped items, do not open the door more than 1 inch during the drying cycle. Press "Start Cycle" to start the drying cycle.
When you press "Start Cycle," the Drying Indicator will be displayed, the chamber will heat to a low temperature and the selected drying time will begin to count down. At the end of the drying cycle, a five second alarm will sound.
Optional Drying Cycles
If you desire to run a drying cycle without running a sterilizing cycle with it, you may. Press the "Optional Cycle" button and scroll the setpoint temperature below 100°C. You will enter a menu for an optional drying cycle. Set the drying time as instructed above, then press "Start
Cycle" to begin the drying cycle.
38
Optional Printer
The optional printer provides a paper tape record for future reference and verifies sterilizer performance, load control identification and quality assurance checks for each cycle.
During a sterilization cycle, the printer prints a record of the chamber temperature, pressure and status of the cycle at 1­minute intervals.
Connecting and Starting the Printer
1. Make sure the printer and sterilizer are turned OFF.
Figure A: Printer
Paper Feed Button
Paper Out Light
ON/OFF Button
Error Light
Power Light
2. Connect the enclosed data cable to the RS232 connector on the back of the steriliz­er and the power cord to the power supply.
3. Turn the printer ON.
4. Turn the sterilizer ON. The printer should print "Barnstead/Thermolyne Sterilemax," "Software Version X.XX.XX"
5. If the printer does not print heading, verify the "Printer Mode" is set to used. See "Changing Printer Mode."
6. Turn the sterilizer off and then back on to verify the printer prints out "Barnstead/Thermolyne Sterilemax," "Software Version X.XX.XX"
7. The printer is now ready to print.
Control Panel (See Figure B)
Figure B: Control Panel
Power Switch
Located on the front of the printer, this switch turns power to the printer ON and OFF. (See Figure A.) The green POWER light will illuminate when the printer is ON.
39
PRINTER
Figure C: Inserting the Paper Roll
Paper Feed Button
Press this button once to advance the tape one line, or hold the button down to advance paper continuously. When the paper roll nears the end, the red PAPER OUT light will illumi­nate.
If the red ERROR light illuminates, the printer is off line. This could be a result of the print head being too hot, in which case the printer will resume printing once it cools down; or— the paper may be jammed in the printer. Turn the printer OFF before checking for jammed paper. Remove the paper jam and turn the printer back ON. If the printer still will not print, unplug it and refer servicing to qualified personnel.
Figure D: Inserting the Paper Into the Paper Slot
Note
A colored stripe along the edge of the paper indicates that you are approaching the end of the tape. When the stripe appears, approx. 23 inches of paper remain, enough for 2-3 cycles. Replace the tape before it is completely exhausted to ensure a complete record of your sterilization procedures
Installing the Paper Roll
When the red PAPER OUT light illuminates, it is time to replace the paper roll. Replace the paper (AY669X4) roll as follows:
1. Turn the printer ON and open the cover.
2. Make sure the edge of the paper roll is straight.
3. Insert the paper roll into the printer. (See Figure C.)
4. Insert the paper straight into the paper slot (see Figure D). The paper will feed automatically.
5. Tear off a strip of paper and close the cover. (The red PAPER OUT light will be OFF if paper roll installation is successful.)
Printer Operation
The printer automatically records sterilization cycles. During sterilization cycles, the printer prints the chamber tempera­ture, pressure and status of cycle at one minute intervals.
40
Default Optional Cycle Printout
PRINTER
Printer Tape Description
The printer tape is designed to be easily read as a cycle is running, allowing you to exam at a glance the progress of the current cycle. As a result, a complete tape must be read from the bottom up. (Sample on page 41)
When you turn on the sterilizer, the printer will print "Barnstead|Thermolyne Sterilemax" and the software version number. This will print only immediately after the sterilizer has been turned on.
When you press one of the preset cycle buttons, the print­er will print the name of the cycle selected, the parameters set for that cycle and the selected pressure units. If you choose a default optional cycle, the printer will print the cur­rent default parameters. If you bypass the default optional cycle for a standard, the printer will print only the name of the cycle and the selected pressure units, since the parameters have not been set.
When you press "Start Cycle", the printer will print the number of cycles that have been run since the unit was turned on, the current date and time, the notation "Cycle Started" and column headers.
During the cycle, the printer will print the elapsed time, the current chamber temperature and the current chamber pres­sure at one minute intervals, along with a status notation. When you press "Start Cycle", the printer will print the num­ber of cycles that have been run since the unit was turned on, the current date and time, the notation "Cycle Started" and column headers.
During the cycle, the printer will print the elapsed time, the current chamber temperature and the current chamber pres­sure at one minute intervals, along with a status notation.
When the sterilizer chamber achieves the temperature set by the cycle parameters, the printer will immediately print a record line, regardless of the time since the last record was printed. The status notation will change from "Heatup" to "Exp", indicating that the load is being exposed to the temper­ature set forth in the cycle parameters. The printer will contin­ue to print record lines at one minute increments, starting one minute after the record line printed at the start of the expo­sure phase but still showing the total elapsed time for the cycle.
41
PRINTER
When the sterilizer has achieved the time condition set in the cycle parameters, the printer will print "Vent" in the cycle status column, indicating that the sterilizer is venting.
At the end of the vent phase, the printer will print "Venting Cycle Completed" and instructions for starting the drying cycle. There is no printout during the drying cycle.
Printer Record Storage
To ensure the long-term legibility of your records, it is important that you keep your printed tapes in a dry, cool, dark place.
42
Instructions for drying
End of cycle
Elapsed time increments by minutes from start of sterilization phase
Exact point at which sterilization phase begins.
Vent Phase
Exposure Phase
RINTER
P
Current status of the cycle
Chamber Pressure
Chamber Temperature
Elapsed time since cycle started
Column Headers
Date and time of cycle initiation
Cycle record number for day
Currently chosen pressure units
Parameters for this programmed cycle (default for this preset program)
Heatup Phase
Name of preset program chosen for this cycle
Software version number prints when you turn on the sterilizer.
Printer Mode is (USED) Version X.XX.XX
Sterilemax Barnstead|Thermolyne
43
Special Operating Conditions
Warning
Do not attempt to open sterilizer door until the display reads "Cycle Complete". Pressure within the chamber could cause the door to open with extreme force, possibly causing personal injury.
Reprocess your load in the event the sterilizing cycle has been ter­minated prematurely. The load may not be sterile when the steril­izing cycle has been terminated prematurely.
Failure to drain excess water from the sterilizer chamber after a cycle abort may result in subsequent unsterile loads as a result of con­tact with water in the chamber. Always drain excess water from the chamber after aborting a cycle.
Allow chamber to cool before draining excess water.
Emergency Off (Cycle Abort)
Pressing the "Stop" button will immediately terminate steriliza­tion. Allow the sterilizer to cool and press the "Stop" button again to reset the sterilizer. Drain excess water from the ster­ilizer chamber. Proceed with a new sterilization cycle normal­ly.
When terminating a load, pressing the "Stop" button will vent steam from the chamber back into the reservoir.
Resetting the Automatic OTP
Display shows: "Cycle Aborted Due to AUTO OTP Device. Press Stop Shutoff Unit Power to Reset."
1. Press the "Stop" button.
2. Wait for the display to read "Select Cycle".
3. Turn Power Switch off and then on to reset the Low Water Cutoff.
4. Open the chamber door and allow the chamber to cool.
44
Note
If display shows: "The Mechanical OTP has Tripped. Shutoff Unit Power and Reset - Caution - Unit Hot!!" Reset as described under "Resetting the Mechanical OTP Sensor" in Service Manual.
5. Select your desired cycle and run the cycle with a loaded chamber to verify operation.
6. In the event of a repetition of fault call Customer Service.
SPECIAL OPERATING SITUATIONS
Warning
Reprocess your load in the event the sterilizing cycle has been terminated prematurely. The load may not be sterile when the sterilizing cycle has been terminated prematurely.
Warning
Reprocess your load in the event the sterilizing cycle has been terminated prematurely. The load may not be sterile when the sterilizing cycle has been terminated prematurely.
Erratic Controls Display or Operation
Abnormal sterilizer operation or controls display could be
caused by an internal or external electrical voltage surge. Such a voltage surge could be cause be any of a number of events: a momentary surge in the power supply, a nearby lightening storm, or even a static electrical charge carried by operating personnel.
Abnormal operation or displays are not common occur-
rences and may never happen. No damage to the equipment will occur if such an event should happen.
If an abnormal condition should occur, terminate any pro-
cessing cycle in process. (Press "Stop.") Reset the sterilizer by turning the power switch (on the front right corner of the unit) to OFF. Wait 30 seconds and turn the power switch back on. Select your desired cycle and proceed with normal opera­tion.
Power Failure During Sterilizing Cycle
All Selector/Indicator Panel lights will be off.
Allow the unit to cool. Reset the sterilizer by turning the
power switch (under the front right corner of the unit) to OFF. Wait 30 seconds and turn the power switch back on. Drain excess water from the sterilizer chamber. Then reprocess your load normally.
45
Problem Solving Chart
sdaeryalpsid,detlahelcyC
putaeHoTeuDdetlaHelcyC"
,emiTmumixaMgnideecxE
".teseRoTpotShsuP
ehtnierutarepmetehT
dehcaertonsahlessev
tnioptesdeificepseht
.setunimnihtiw
retawetauqedarofkcehC.1
.elcycfogninnigebehttallif
/sevlavdioneloskcehC.2
.skaelrofteksagrood
tonlliwrezilirets"eeS.3
".taeh
sdaeryalpsid,detlahelcyC
revOoTeuDdeliaFelcyC"
,noitidnoCerutarepmeT
".eunitnoCoTpotSsserP
edisnierutarepmetehT
sahrebmahceht
nahteromdedeecxe
5° .tnioptesehtevobaC
,nottubPOTSehtsserP
otrebmahcwolla
niard,loocdnaezirusserped
.elcyctratser,rebmahceht
denehtgnelesahperusopxE
.emitdemmargorpdnoyeb
ehtnierutarepmeT
eromdeppordrebmahc 1naht ° eromtontubC 5naht ° ehtwoleb
ehtgnisuac,tnioptes
remitesahperusopxe
toN(.teserot
diuqilotelbacilppa
).elcyc
sidaoL.yrassecenenoN
.elcycfodnetaelirets
sdaeryalpsid,detlahelcyC
woLoTeuDdeliaFelcyC"
,noitidnoCerutarepmeT
".eunitnoCoTPOTSsserP
ehtnierutarepmeT
eromdeppordrebmahc 1naht ° ehtwolebC
diuqilanitnioptes
5nahteromro,elcyc °C
.elcycrehtoynani
,nottubPOTSehtsserP
otrebmahcwolla
niard,loocdnaezirusserped
.elcyctratser,rebmahceht
sdaeryalpsid,detlahelcyC
,denepOsaHrooDehT"
,detrobAelcyCnoitaziliretS
".potSsserP
hctiwsroodehT
ehtretfadenepo
.elcycafonoitaitini
,nottubPOTSehtsserP
otrebmahcwolla
niard,loocdnaezirusserped
,elcyctratser,rebmahceht
siroodehttahtgnirusne
.desolcyleruces
yalpsid,tratstonseodelcyC
,nepOsIrooDehT"sdaer
".eunitnoCoTrooDesolC
sawhctiwsroodehT
sawTRATSnehwnepo
.desserp
elcyC.ylerucesroodesolC
sihctiwsroodnehwtratslliw
.desolc
saHtinU",sdaeryalpsiD
eroMenOroFretaWhguonE
llifeResaelP,elcyC
".riovreseR
hctiwslevelretawehT
deneporiovreserni
gnitacidni,llifehtgnirud
.retawwol
retfA.nurotelcycwollA
llifer,dednesahelcyc
gninnurerofebriovreser
.elcycrehtona
115
Fault Indication
Fault Cause
Solution
46
PROBLEM SOLVING CHART
noitacidnItluFaseuaCtluaonituloS
yalpsid,tratstonseodelcyC
.ytpmEsiriovreseR"sdaer
noitaziliretStratseRdnalliF
".elcyC
hctiwslevelretawehT
gninnigebehttaneposi
gnitacidni,elcycafo
.retawwol
tratseR.riovreserllifeR
.elcyc
ehtdesserpsiPOTSnehW
erusserP"sdaeryalpsid
esaelP,isp1nahTretaerG
".potSoTtiaW
,deppotssahelcycehT
rebmahcehttub
ottaergootsierusserp
.roodehtnepoylefas
.pordoterusserproftiaW
tadesserpsiTRATSnehW
nigebotgnitnevfodneeht
yalpsideht,esahpgniyrdeht
nepOeBtsuMrooD"sdaer
".elcyCgniyrDnigeBoT
rooD(.desolcsirooD
rofnepoebtsum
).gniyrdetauqeda
.TRATSsserP.roodnepO
erusserprevolacinahceM
gnitnev,snepoevlavfeiler
.rebmahcmorferusserp
ehtnierusserP
dedeecxesahrebmahc
.isp54
,nottubPOTSehtsserP
otrebmahcwolla
tratseR.ezirusserped
afotneveehtnI.elcyc
llac,tluaffonoititeper
daetsnraB
.ecivreSremotsuC
detrobAelcyC"sdaeryalpsiD
.eciveDPTOotuAoTeuD
tinuffotuhs,POTSsserP
.teserotrewop
norosneserutarepmeT
dehcaersahretaeheht
rotniopffotucteserpsti
.eruliafrosnes
eeS ehtgnitteseR
PTOcitamotuA ehtnI.
noititeperafotneve
llac,tluaffo
daetsnraB
.ecivreSremotsuC
.taehtonlliwreziliretS
ehT":sdaeryalpsiD
saHPTOlacinahceM
dnAtinUffotuhS.deppirT
"!!toHtinU-noituaC.tratseR
efas-liafehT
revolacinahcem
sahhctiwserutarepmet
oteuddenepo
lamrehttnacifingis
.yawanur
rednudebircsedsateseR
lacinahceMehtgnitteseR
tneveehtnI.rosneSPTO
,tluaffonoititeperafo
cdaetsnraBlla
.ecivreSremotsuC
naftub,tuotnirproyalpsidoN
.gninnursikcabno
gnitcetorpesuF
.nwolbera
sa
esufecalpeR
rednu
debircsed
gnicalpeR
.sesuF
ehT
Fa
International
International
International
the
power supply
Power Supply
47
Maintenance Schedule Chart
Maintenance Required Person Responsible
Daily
Clean Door Gasket User
Clean Chamber* User
Weekly
Clean Chamber, Trays and Rack User
Monthly
Check Pressure Relief Valve Qualified Service Personnel
Clean Reservoir User
Yearly
Temperature Readout Verification Qualified Service Personnel
As Required
Install Paper Roll User
Changing Door Gasket Qualified Service Personnel
*When processing liquid loads
48
Warning
Refer servicing to qualified person­nel.
Maintenance
Daily Maintenance
Never use a wire brush, steel wool, abrasive material, or chlo­ride-containing products to clean door and chamber assembly.
"Caution: Hot Surface. Avoid Contact." Ensure that the sterilizer is cool before cleaning to avoid burns.
Caution
Failure to keep the interior of the stainless steel chamber free of mineral deposits and debris can cause premature failure of the sterilizer.
Cleaning Door Gasket
The door gasket and the mating surface should be wiped clean each day with a clean, damp cloth. Do not use abra­sive cleaners on the gasket or mating surface.
The gasket should be examined for cracks and dam­age that could result in a poor pressure seal. If replacement is necessary, refer to Gasket Replacement under the When Required heading of this Maintenance section.
Cleaning After Liquid Loads
Biological media tends to boil at a higher rate than other liq­uids during venting. This causes media to be spattered inside the chamber. Therefore, the chamber must be cleaned daily when you are sterilizing media. Clean as follows:
1. Allow unit to cool.
2. Wipe out chamber and door with a clean, damp cloth.
Weekly Maintenance
(More often if necessary.)
Warning
To prevent collection of mineral deposits and corrosion of chamber components, use distilled or deion­ized water only as specified. Clean chamber after each use if steriliz­ing saline solutions.
Cleaning Chamber, Trays and Rack
At least once a week, the trays and tray rack should be removed from the sterilizer chamber. The trays, tray rack and chamber should be thoroughly cleaned to remove any deposits from the surfaces. Clean the trays, rack and chamber (especially the bottom of the chamber) with appropriate anti-biological cleaners. Wipe all residue from the surfaces with a dampened, lint-free cloth.
49
MAINTENANCE
Warning
Pressurized steam will be dis­charged when the relief valve is opened. To prevent burns, place a steam barrier (such as a rolled towel) around the opening in the back of the unit. To prevent burns, wear gloves or use an extension device if it becomes necessary to operate the pull ring.
Monthly Maintenance
Pressure Relief Valve Check
To ensure that the pressure relief valve is functioning proper­ly, you must have a qualified person check the valve each month.
The pressure relief valve is located on the pressure line in the back of the unit and is accessed through a round access hole in the back panel. A suitable tool (screwdriver, rod, etc.) will be needed to test the pressure valve.
1. Run a sterilize cycle at 135°C (275°F) without a load in the chamber.
2. Allow the pressure in the chamber to reach 30 psi (206 kPa).
3. Insert a suitable tool through the access hole and push the lever on the pressure relief valve. Push the valve lever away from the valve to open the valve. Hold the valve open for 3 seconds. Steam should discharge freely.
Note
In the event that you have lost the supplied drain tubing, you can drain the reservoir as follows:
1. Tilt the front of the sterilizer up until you can access the small panel on the bottom front of the sterilizer, beneath the drain assembly.
2. Remove the screws securing the panel. Remove the panel.
3. Have a suitable container pre­pared for the water in the reser­voir. Disconnect the reservoir drain tube from the quick disconnect fix­ture on the front of the sterilizer. The reservoir will drain.
Note
Do not add more than 7 liters of water to the reservoir after cleaning.
4. Release the valve lever. The steam discharge should cease entirely.
5. After checking the valve, turn the sterilizer off by pressing "Stop" to prevent overheating.
If the valve malfunctions, the valve must be replaced.
Cleaning Reservoir
1. The reservoir drain assembly is located on the lower right front side of the sterilizer cabinet. Place a suitable container below the drain assembly. Connect the drain adapter and tube to the drain assembly and drain the reservoir into the contain­er.
2. When the reservoir is empty, wipe out the inside of the reservoir with a clean, lint-free cloth. Take care not to push debris back into the water line.
3. Refill the reservoir with distilled or low quality deionized water (minimum 0.5 megohms/cm, maxi­mum 1.5 megohms/cm) and replace the reservoir cover.
50
Warning
BURN HAZARD: Do not empty the collection bottle when the sterilizer is in operation.
MAINTENANCE
Emptying Collection Bottle
1. Empty the collection bottle when the water in the bottle reaches the FULL mark.
3. Slide the bottle away from the cooling tank to remove.
4. Pour the contents down the drain.
Optional
Overflow
Drain
Waste Water
Cooling
Tank
Note
It is important to periodically check the water level in the collection bot­tle. Failure to empty the bottle will result in water overflowing onto sur­rounding surfaces.
Collection
Bottle
From
Sterilizer
51
MAINTENANCE
Warning
BURN HAZARD: Do not clean the tank when the sterilizer is in opera­tion. Allow the non-recirculating cool­ing tank to cool to room temperature before cleaning the tank.
Cleaning Non-Recirculating Water Cooling Tank
The non-recirculating water cooling tank should be drained and rinsed out every six months. Do not clean the tank when the sterilizer is in operation.
Using the optional overflow tube (supplied), push the connec­tor fitting into the drain connection at the bottom of the tank to empty it.
Remove the drain tube when the tank is empty, and rinse the tank with a solution of household bleach (3/4 cup) and dis­tilled or deionized water (1 gal.). Drain solution and rinse with distilled or deionized water before refilling.
Refill the tank with clean distilled or deionized water to put the system back in service.
Overflow Drain
To Empty:
Push in overflow
tube until it “clicks.”
Optional Overflow
Tube
Tank
Drain
52
To Remove
Overflow Tube:
Press tab to release.
MAINTENANCE
Changing Door Gasket
1. Disconnect the sterilizer from the power supply. Ensure that the sterilizer is cool and depressur­ized.
2. Remove the old gasket by pulling it out of the door.
3. Use a warm, soapy water combination to lubricate the gasket for easier installation.
4. Working around the door from one point, insert the outer edge of the gasket under the machined lip of the gasket groove in the door.
5. Working around the door from one point, press the inner edge of the gasket into the gasket groove, ensuring that the gasket is fully seated in the groove.
6. Reconnect the sterilizer to the power supply.
53
Figures
Power Supply Board With Sheild Cover
On/Off Switch (underneath)
R
eservoir Cover
Door Release
Door Handle
Membrane Panel
Main Control Board
Pump
Power Module
OTP PC Board
Solenoid PC Board
Sterilizer Component Layout - Interior
54
FIGURES
T
rays
R
ack
D
oor Interlock
Reservoir Drain
Door Switch
Sterilizer Component Layout - Chamber and Door
55
FIGURES
Solid State  R
elay
O
TP Mechanical Relay
Pressure Relief Valve
Calibration Port
T
emperature Sensor
Fan
Vent Solenoid
Insulation
Reservoir
Pressure Transducer
Insulation Support  Bracket
Door Switch
Insulation Support Bracket
Sterilizer Component Layout - Top View
56
Parts List
Warning
Replace fuses with same type and rating
120 V 220V~
Qty. Description Part No. Part No.
1 Printer Ribbon AYX21 AYX21 1 Printer Paper Tape Roll AY669X4 AY669X4 2 Solenoid Valve RY759X1A RY759X1A 1 Power Supply TNX116 TNX116
PC Board (Solenoid) PC1277X2 PC1277X2 Pressure Relief Valve 250052401 250052401 Tank, Lower Float SW759X1A SW759X1A PC Board (Logic) PC1277X1 PC1277X1 ON/OFF Switch SWX141 SWX104 Terminal Strip TRX177 TRX177 Relay Contact RYX60 RYX60 Pump PUX16 PUX16 Door Gasket GS759X1A GS759X1A Fan FA981X1A FA759X1A Heater ELX9 ELX10 Gasket (Chamber) GS759X4 GS759X4 Relay, Solid State RYX34 RYX34 PC Board (OTP) PC1118X2 PC1118X2 Tray (Large) TY759X2A TY759X2A Tray (Small) TY759X3A TY759X3A Door Switch SW1118X1 SW1118X1 Membrane Panel SW759X3 SW759X3 Water Reservoir TY759X1 TY759X1 Safety Latch SF759X1A SF759X1A Shaft Ass'y (Main Door Latch) SF759X2A SF759X2A
Pressure Pin (Door) SP759X1 SP759X1 2 Handle (Yoke Assembly) HNX18 HNX18 1 Drain Tube Assembly TU759X9A TU759X9A
Reservoir Cover DL759X2 DL759X2
Chamber Temp. Sensor RS1277X1 RS1277X1
Automatic OTP Sensor RS759X2A RS759X2A
Mechanical OTP Sensor FZ759X1 FZ759X1 1 Pressure Transducer TDX5 TDX5
Fuse for TBX116 AZ9027 AZ9027
57
Ordering Procedures
Please refer to the Specification Plate for the complete model number, serial number, and series number when requesting service, replacement parts or in any correspondence con­cerning this unit.
All parts listed herein may be ordered from the Thermo Scientific dealer from whom you purchased this unit or can be obtained promptly from the factory. When service or replacement parts are needed we ask that you check first with your dealer. If the dealer cannot handle your request, then contact our Customer Service Department at 563-556­2241 or 800-553-0039.
Prior to returning any materials, please contact our Customer Service Department for a “Return Materials Authorization” number (RMA). Material returned without an RMA number will be refused.
58
Decontamination Statement
We cannot accept any product or component sent for repair or credit that is contaminated
with or has been exposed to potentially infectious agents or radioactive materials.
No product or component will be accepted without a "Return Materials Authorization"
(RMA) number.
59
One Year Limited Warranty
This Thermo Scientific product shall be free of defects in materials and workmanship for one (1) year from the first to occur of (i) the date the product is sold by the manufacturer or (ii) the date the product is purchased by the origi­nal retail customer (the "Commencement Date"). Except as expressly stated above, WE MAKE NO OTHER WAR­RANTY, EXPRESSED OR IMPLIED, WITH RESPECT TO THE PRODUCTS AND EXPRESSLY DISCLAIM ANY AND ALL WARRANTIES, INCLUDING BUT NOT LIMITED TO, WARRANTIES OF DESIGN, MERCHANT ABILITY AND FITNESS FOR A PARTICULAR PURPOSE.
An authorized representative of the manufacturer must perform all warranty inspections. In the event of a defect covered by this warranty, the manufacturer shall, as its sole obligation and exclusive remedy, provide free replace­ment parts to remedy the defective product. In addition, for products sold by the manufacturer within the continen­tal United States or Canada, the manufacturer shall provide free labor to repair the products with the replacement parts, but only for a period of ninety (90) days from the Commencement Date.
The warranty provided hereunder shall be null and void and without further force or effect if there is any (i) repair made to the product by a party other than the manufacturer or its duly authorized service representative, (ii) misuse (including use inconsistent with written operating instructions for the product), mishandling, contamination, over­heating, modification or alteration of the product by any customer or third party or (iii) use of replacement parts that are obtained from a party who is not an authorized dealer of Thermo Scientific products.
Heating elements, because of their susceptibility to overheating and contamination, must be returned to the factory and if, upon inspection, it is concluded that failure is due to factors other than excessive high temperature or conta­mination, the manufacturer will provide warranty replacement. As a condition to the return of any product, or any constituent part thereof, to the factory, it shall be sent prepaid and a prior written authorization from the manufactur­er assigning a Return Materials Number (RMA) to the product or part shall be obtained.
IN NO EVENT SHALL THE MANUFACTURER BE LIABLE TO ANY PARTY FOR ANY DIRECT, INDIRECT, SPE­CIAL, INCIDENTAL, OR CONSEQUENTIAL DAMAGES, OR FOR ANY DAMAGES RESULTING FROM LOSS OF USE OR PROFITS, ANTICIPATED OR OTHERWISE, ARISING OUT OF OR IN CONNECTION WITH THE SALE, USE OR PERFORMANCE OF ANY PRODUCTS, WHETHER SUCH CLAIM IS BASED ON CONTRACT, TORT (INCLUDING NEGLIGENCE), ANY THEORY OF STRICT LIABILITY OR REGULATORY ACTION.
For the name of the authorized Thermo Scientific dealer nearest you or any additional information, contact us:
2555 Kerper Blvd., Dubuque, Iowa, 52001-9918
Phone: 563-556-2241 or 1-800-553-0039
Fax: 563-589-0516
E-mail: mkt@thermofisher.com
Web: www.thermo.com
Loading...