Thermo Fisher Scientific Inc. provides this document to its customers with a product purchase to use
in the product operation. This document is copyright protected and any reproduction of the whole or
any part of this document is strictly prohibited, except with the written authorization of Thermo Fisher
Scientific Inc.
The contents of this document are subject to change without notice. All technical information in this
document is for reference purposes only. System configurations and specifications in this document
supersede all previous information received by the purchaser.
Thermo Fisher Scientific Inc. makes no representations that this document is c o mplete,
accurate or error-free and assumes no responsibility and will not be liable for any errors,
omissions, damage or loss that migh t result from any use of this document, even if the
information in the document is followed properly.
This document is not part of any sales contract between Thermo Fisher Scientific Inc. and a
purchaser. This document shall in no way govern or modify any Terms and Conditions of Sale, which
Terms and Conditions of Sale shall govern all conflicting information between the two documents
This product is subject to the regulations of the EU Waste Electrical & Electronic Equipment
(WEEE) Directive 2002/96. It is marked by the following symbol:
Thermo Fisher Scientific has entered into agreements with recycling and disposal companies in all EU
Member States for the recycling and disposal of this device. For information on recycling and disposal
companies in Germany and on the products of Thermo Fisher Scientific, which fall under the RoHS
Directive (Restriction of the use of certain hazardous substances in electrical and electronic
equipment), please visit the website www.thermo.com/WEEERoHS
Page 5
Table of Contents
Preface ...................................................................................................................................... iii
Scope of Supply ....................................................................................................................... iii
Intended Use ............................................................................................................................. iii
Mechanical Emergency Door Release ...................................................................... 7-2
Troubleshooting by User ......................................................................................... 7-3
When to contact Customer Service .......................................................................... 7-5
Chemical Compatibility Chart ..............................................................................A-1
Contact Information ................................................................................................B-1
iiST40 / 40RThermo Scientific
Page 7
Preface
P
Before starting to use the centrifuge, read through this instruction manual carefully and follow the
instructions.
The information contained in this instruction manual is the property of Thermo Fisher Scientific; it is
forbidden to copy or pass on this information without explicit approval.
Failure to follow the instructions and safety information in this instruction manual will result in the
expiration of the sellers warranty.
Scope of Supply
Article NumberQuantityCheck
50119928Instruction manual1
70009824Anti-corrosion oil1
If any parts are missing, please contact your nearest Thermo Fisher Scientific representative.
Intended Use
• This centrifuge is an IVD accessory, and therefore subject to the Directive 98/79/EC of the
European Parliament and the Council of October 27, 1998 on in vitro diagnostic medical devices.
• This centrifuge is a laboratory product designed to separate components by generation of Relative
Centrifugal Force. It separates human samples (e.g. blood, urine and other body fluids) collected
in appropriate containers, either alone or after addition of reagents or other additives.
Centrifuge Sorvall ST40 / 40R1
Power supply cable1
• As general-purpose centrifuge, it is designed to also run other containers filled with chemicals,
environmental samples and other non-human body samples.
• This centrifuge should be operated by trained specialists only.
Thermo ScientificST40 / 40Riii
Page 8
Preface
Accident Prevention
Prerequisite for the safe operation of the Sorvall ST40 / 40R is a work environment in compliance with
standards, directives and trade association safety regulations and proper instruction of the user.
The safety regulations contain the following basic recommendations:
• Maintain a radius of at least 30 cm around the centrifuge.
• Implementation of special measures which ensure that no one can approach the centrifuge for
longer than absolutely necessary while it is running.
• The mains plug must be freely accessible at all times. Pull out the power supply plug or disconnect
the power supply in an emergency.
Precautions
In order to ensure safe operation of the Sorvall ST40 / 40R, the following general safety regulations
must be followed:
• Do not manipulate the safety devices.
• The centrifuge should be operated by trained specialists only.
• The centrifuge is to be used for its intended use only.
• Plug the centrifuge only into sockets which have been properly grounded.
• Do not move the centrifuge while it is running.
• Do not lean on the centrifuge.
• Use only rotors and accessories for this centrifuge which have been approved by Thermo Fisher
Scientific. Exceptions to this rule are commercially available glass or plastic centrifuge tubes,
provided they have been approved for the speed or the RCF value of the rotor.
• Do not use rotors which show any signs of corrosion and/or cracks.
• Do not touch the mechanical components of the rotor and do not make any changes to the
mechanical components.
• Use only with rotors which have been properly installed. Follow the instructions on the Autolock
in section “Rotor Installation” on page 4-2.
• Use only with rotors which have been loaded properly. Follow the instructions given in the rotor
manual.
• Never overload the rotor. Follow the instructions given in the rotor manual.
• Never start the centrifuge when the lid is open.
• Never open the lid until the rotor has come to a complete stop and this has been confirmed in the
display.
ivST40 / 40RThermo Scientific
Page 9
• The lid emergency release may be used in emergencies only to recover the samples from the
centrifuge, e.g. during a power failure (see section “Mechanical Emergency Door Release” on
page 7-2).
• Never use the centrifuge if parts of its cover panels are damaged or missing.
• Do not touch the electronic components of the centrifuge or alter any electronic or mechanical
components.
• Please observe the safety instructions.
Please pay particular attention to the following aspects:
• Location: well-ventilated environment, set-up on a level and rigid surface with adequate
load-bearing capacity.
• Rotor installation: make sure the rotor is locked properly into place before operating the
centrifuge.
• Especially when working with corrosive samples (salt solutions, acids, bases), the accessory parts
and vessel have to be cleaned carefully.
Preface
• Always balance the samples.
Centrifuging hazardous substances:
• Do not centrifuge explosive or flammable materials or substances which could react violently with
one another.
• The centrifuge is neither inert nor protected against explosion. Never use the centrifuge in an
explosion-prone environment.
• Do not centrifuge inflammable substances.
• Do not centrifuge toxic or radioactive materials or any pathogenic micro-organisms without
suitable safety precautions.
When centrifuging microbiological samples from the Risk Group II (according to the Bio-safety
Manual" of the World Health Organization WHO), aerosol-tight biological seals have to be used.
For materials in a higher risk group, extra safety measures have to be taken.
• If toxins or pathogenic substances have gotten into the centrifuge or its parts, appropriate
disinfection measures have to be taken (see “Disinfection” on page 6-3).
• Highly corrosive substances which can cause material damage and impair the mechanical stability
of the rotor, should only be centrifuged in corresponding protective tubes.
IF A HAZARDOUS SITUATION OCCURS, TURN OFF THE POWER SUPPLY TO THE
CENTRIFUGE AND LEAVE THE AREA IMMEDIATELY.
Intended Use
The centrifuge is to be used for separating materials of different density or particle size suspended in a
liquid.
Maximum sample density at maximum speed: 1,2
Thermo ScientificST40 / 40Rv
g
-----ml
Page 10
Preface
This symbol refers to general hazards.
CAUTION means that material damage could occur.
WARNING means that injuries or material damage or contamination could occur.
This symbol refers to biological hazards.
Observe the information contained in the instruction manual to keep yourself and your
environment safe.
viST40 / 40RThermo Scientific
Page 11
Introduction and Description
Contents
• “Characteristics” on page 1-2
• “Technical Data” on page 1-3
• “Directives, Standards and Guidelines” on page 1-4
• “Functions and Features” on page 1-5
• “Mains Supply” on page 1-5
1
• “Rotor Selection” on page 1-6
Thermo ScientificST40 / 40R1-1
Page 12
1
Introduction and Description
Characteristics
Characteristics
The Sorvall ST40 / 40R is an in-vitro diagnostics device according to the In-Vitro Diagnostics
Directive 98/79/EC.
Several rotors with a wide range of tubes can be used.
The set speed is reached within seconds. The maintenance-free induction motor ensures quiet and
low-vibration operation even at high speeds, and guarantees a very long lifetime.
The user-friendly control panel makes it easy to pre-set the speed, RCF value, running time,
temperature, and running profile (acceleration and braking curves). You can choose between the
display of speed and RCF or the entry mode.
These settings can be changed even while the centrifuge is running.
With the help of the PULSE key , you can also centrifuge a sample for just a few seconds, if called
for.
The Sorvall ST40 / 40R is equipped with various safety features:
• The housing and rotor chamber consist of steel plate, the interior of armour steel, while the front
panel is made of high-impact resistant plastic.
• The lid is equipped with a view port and a lock.
• The lid of the centrifuge can only be opened while the centrifuge is switched on and the rotor has
come to a complete stop. The centrifuge cannot be started until the lid has been closed properly.
• The integrated rotor detection systems ensures that no inadmissible speed settings can be
preselected.
• Electronic imbalance recognition
• Lid emergency release: For emergencies only, e.g. during power failures (see “Mechanical
Emergency Door Release” on page 7-2)
1-2ST40 / 40RThermo Scientific
Page 13
1
Technical Data
The technical data of the Sorvall ST40 / 40R is listed in the following table.
Table 1-1 . Technical Data Sorvall ST40 / 40R
FeatureValue
Environmental conditions-Use in interior spaces
-Altitudes of up to 2,000 m above sea level
-max. relative humidity 80 % up to 31 °C; decreasing linearly up to 50 %
relative humidity at 40 °C.
permissible ambient temperature +2 °C to +35 °C
Overvoltage categoryIIII
Pollution degree22
Heat dissipationventilatedrefrigerated
230V120V230V120V
5800 BTU/h4776 BTU/h6653 BTU/h4776 BTU/h
Introduction and Description
Technical Data
IP2020
running timeunlimitedunlimited
max Speed n
min Speed n
maximum RCF value at n
maximum kinetic energy< 62.5 kJ< 62.5 kJ
noise level at maximum speed < 64 dB (A)< 64 dB (A)
temperature setting range-10 °C to +40 °C
Dimensionsventilatedrefrigerated
weight without rotor86 kg116 kg
max
min
max
Height
Width
Depth
Table top height
15200 rpm (depending on the rotor)15200 rpm (depending on the rotor)
300 rpm300 rpm
25314 xg 25314 xg
360 mm
550 mm
670 mm
310 mm
360 mm
745 mm
670 mm
310 mm
Thermo ScientificST40 / 40R1-3
Page 14
1
Introduction and Description
Directives, Standards and Guidelines
Directives, Standards and Guidelines
Table 1-2. Directives, Standards and Guidelines
Tension /
Frequency
230V 50/60Hz2006/95/EC Low Voltage Directive:
2006/42/EC & 98/37/EC Machine Directive:
2004/108/EC EMC Directive
98/79/EC In-vitro-Diagnostika (IvD)
Produced and inspected according to the following standards
and guidelines
EN 61010-1, 2
EN 61010-2-020, 2
EN 61010-2-101
EN 61326-1
EN 61326-2-6
EN 55011B
EN 61000-6-2
EN ISO 13485
230V 60Hz
120V 60Hz
100V 60Hz
UL 61010-1, 2
CAN/CSA-C22.2 No. 61010-1, 2
IEC 61010-2-20, 2
(Pollution degree 2, Overvoltage category II)
ICE 61010-2-101
Emitted interference FCC Part 15 CLASS A
NOTE: This equipment has been tested and found to comply with the
limits for a Class A digital device, pursuant to Part 15 of the FCC Rules.
These limits are designed to provide reasonable protection against
harmful interference when the equipment is operated in a commercial
environment. This equipment generates, uses, and can radiate radio
frequency energy and, if not installed and used in accordance with the
instruction manual, may cause harmful interference to radio
communications. Operation of this equipment in a residential area is
likely to cause harmful interference in which case the user will be
required to correct the interference at his own expense.
100V 50HzIEC 61010-1 2
IEC 61010-2-020 2
IEC 61010-2-101
EN 61326-1
EN 61326-2-6
EN 55011A
EN 61000-6-2
EN ISO 13485
nd
Edition
nd
Edition
nd
Edition
nd
Edition
nd
Edition
nd
Edition
nd
Edition
1-4ST40 / 40RThermo Scientific
Page 15
1
Introduction and Description
Functions and Features
Functions and Features
The following table gives an overview of the important functional and performance characteristics of
the Sorvall ST40/40R.
Table 1-3 . Functions and Features
Component / FunctionDescription / Features
Structure / HousingGalvanized steel chassis with armoured plating
ChamberStainless steel / varnished steel
DriveInduction drive without carbon brushes
Keys and displayEasy-to-clean keypad and display surface
Controls Microprocessor-controlled
Internal memoryThe most recent data is saved
FunctionsRCF-selection, temperature control, pretemp with cooling equipped
devices
Acceleration / braking profiles9 acceleration and 10 braking curves
Rotor recognitionAutomatic
Imbalance recognitionElectronic, contingent on rotor and speed
Lid lockAutomatic lid closing and locking starting from an initial hold
position
Mains Supply
The following table contains an overview of the electrical connection data for the Sorvall ST40/40R.
This data is to be taken into consideration when selecting the mains connection socket.
Table 1-4 . Electrical connection data of the Sorvall ST40/40R.
Cat.
7500 4524refrigerated230 V 50 / 60 Hz10.5 A1950 W15 AT16 AT
7500 4525refrigerated120 V 60 Hz12 A1400 W15 AT15 AT
7500 4326refrigerated100 V 50 / 60 Hz
7500 4509ventilated230 V 50 / 60 Hz11 A1700 W15 AT16 AT
7500 4510ventilated120 V 60 Hz12 A1400 W15 AT15 AT
Mains
voltage
Frequency
Rated
current
*
13 A1250 W15 AT15 AT
Power
consumpti
on
Equipment fuse Building fuse
7500 4211ventilated100 V 50 / 60 Hz13 A1250 W15 AT15 AT
*
Please contact the Thermo Fisher service to run the unit at 60Hz Line frequency.
Thermo ScientificST40 / 40R1-5
Page 16
1
Introduction and Description
Rotor Selection
Rotor Selection
The Sorvall ST40 / 40R is supplied without a rotor.
Various rotors are available to choose from.
TX-7507500 3607
with round buckets7500 3608
with rectangular buckets7500 3614
incl. Microtiter carrier7500 3617
incl. ABI-bucket7500 3618
BIOShield
BIOLiner
FIBERLite
FIBERLite
TM
TM
TM
TM
1000A7500 3603
7500 3667
F20-6 x 100 CFK7500 3698
F21-48 x 27500 3664
Microliter 48 x 27500 3602
HIGHConic
M-207500 3624
TM
6 x 1007500 3620
The technical data of the rotors and the corresponding adapters and reduction sleeves for various
commercially available containers can be found in the corresponding rotor operating manuals.
For more information visit our website at: http://www.thermo.com
1-6ST40 / 40RThermo Scientific
Page 17
Before use
Contents
• “Before Setting up” on page 2-2
• “Transporting the Centrifuge” on page 2-2
• “Location” on page 2-2
• “Aligning the Centrifuge” on page 2-3
• “Mains Connection” on page 2-3
2
• “Storage” on page 2-4
Thermo ScientificST40 / 40R2-1
Page 18
2
Before use
Before Setting up
Before Setting up
1.Check the centrifuge and the packaging for any shipping damage.
Inform the shipping company and Thermo Fisher Scientific immediately if any damage is
discovered.
2.Remove the packaging.
3.Check the order for completeness (see “Scope of Supply” on page iii).
If the order is incomplete, please contact Thermo Fisher Scientific.
Transporting the Centrifuge
• Due to its weight (see “Technical Data” on page 1-3), the centrifuge should be carried by several
people.
• Always lift the centrifuge at both sides.
Location
Figure 2-1. Lifting the centrifuge at both sides.
• The centrifuge can be damaged by impacts.
• Transport the centrifuge upright and if at all possible in its packaging.
WARNING Always lift the centrifuge on both sides. Never lift the centrifuge by its front
or the back panel.
Always remove the rotor before moving the centrifuge.
The centrifuge should only be operated indoors.
The set-up location must fulfil the following requirements:
• A safety zone of at least 30 cm must be maintained around the centrifuge.
People and hazardous substances must be kept out of the safety zone while centrifuging.
• The supporting structure must be stable and free of resonance, for example a level laboratory
bench.
2-2ST40 / 40RThermo Scientific
Page 19
• The supporting structure must be suitable for horizontal setup of the centrifuge.
• The centrifuge should not be exposed to heat and strong sunlight.
WARNING UV rays reduce the stability of plastics.
Do not subject the centrifuge, rotors and plastic accessories to direct sunlight.
• The set-up location must be well-ventilated at all times.
Aligning the Centrifuge
The horizontal alignment of the centrifuge must be checked every time after moving it to a different
location.
The supporting structure must be suitable for horizontal setup of the centrifuge.
CAUTION If the centrifuge isn’t level, imbalances can occur and the centrifuge can be
damaged.
Do not place anything under the feet to level the centrifuge.
2
Before use
Aligning the Centrifuge
Mains Connection
Figure 2-2. Mains Connection
Main Power
Mains plug
1. Turn off the power supply switch on the back (press "0").
2.Plug the centrifuge into grounded electrical sockets only.
3.Check whether the cable complies with the safety standards of your country.
Thermo ScientificST40 / 40R2-3
Page 20
2
Before use
Storage
Storage
4.Make sure that the voltage and frequency correspond to the figures on the rating plate.
5.Establish the connection to the power supply with the connecting cable.
• Before storing the centrifuge and the accesories it must be cleanedand if necessary desinfected and
decontaminated.
• Store the centrifuge in a clean, dust-free location.
• Be sure to place the centrifuge on its feet.
• Avoid direct sunlight.
2-4ST40 / 40RThermo Scientific
Page 21
Control Panel
Contents
• “Control Panel” on page 3-2
• “Keys” on page 3-2
3
Thermo ScientificST40 / 40R3-1
Page 22
3
Control Panel
Control Panel
Control Panel
The control panel contains the keys and displays of the centrifuge (only the power switch is located on
the back of the device). All parameters can be selected and changed during operation.
Program selection keys
Keys
Acceleration curve
Speed / RCF Value
Changing the display mode
Speed / RCF Value
running time
Temperature
Prewarming or Precooling
Braking curve
Bucket selection
Bucket key
The keys allow user intervention for controlling the operating mode as follows:
KeyDisplay contents
Start
Stop
Normal start of the centrifuge
End run manually
PULSE key
OPEN key
START key
STOP key
Open lid
Pulse
Change Value
Snow Symbol
Changing the
display mode
Automatic release (possible only when device is
switched on). Emergency release (see
“Mechanical Emergency Door Release” on
page 7-2)
By pressing the PULSE key the centrifuge starts
immediately and accelerates up to the end
speed. Releasing the key initiates a stopping
process at the highest braking curves.
Use the arrow keys in order to modify the
displayed value
Press the snow symbol key for prewarming or
precooling the centrifuge.
Use the Change key to change the display mode.
(Speed / RCF Value)
3-2ST40 / 40RThermo Scientific
Page 23
Operation
Contents
4
• “Switch on Centrifuge” on page 4-2
• “Lid Opening” on page 4-2
• “Close Lid” on page 4-2
• “Rotor Installation” on page 4-2
• “Entering Parameters” on page 4-4
• “Saving programs” on page 4-6
• “Centrifugation” on page 4-7
• “Temperature Adaptation during Standstill” on page 4-8
• “Short-term Centrifugation” on page 4-8
• “Removing the Rotor” on page 4-8
• “Aligning the Centrifuge” on page 4-9
Thermo ScientificST40 / 40R4-1
Page 24
4
Operation
Switch on Centrifuge
Switch on Centrifuge
1. Turn on the power switch on the back of the device.
The device performs a self-check of its software.
a.When the centrifuge lid is closed the following display shows:
The speed and time displays read 0. The temperature indicator displays the current
temperature inside the rotor chamber. The preset acceleration and braking curves and the
selected bucket are also displayed.
b. When the centrifuge lid is open the following display shows:
The speed and time displays show the preset values. The temperature indicator displays the
current temperature inside the rotor chamber. The preset acceleration and braking curves and
the selected bucket are also displayed.
Lid Opening
1. Press the OPEN key .
WARNING Do not reach into the crack between the lid and the housing. The lid is
drawn shut automatically.
Use the emergency release only for malfunctions and power failures (see “Mechanical
Emergency Door Release” on page 7-2).
Close Lid
Close the lid by pressing down on it lightly in the middle or on both sides of it. Two locks close the lid
completely.
WARNING Do not reach into the crack between the lid and the housing. The lid is
drawn shut automatically.
Note The lid should audibly click into place.
Rotor Installation
The approved rotors for the Sorvall ST40 / 40R are listed in section “Rotor Selection” on page 1-6. Use
only the rotors and accessories from this list in the centrifuge.
CAUTION Unapproved or incorrectly combined accessories can cause serious damage to
the centrifuge.
The centrifuge is equipped with an AutoLock
4-2ST40 / 40RThermo Scientific
TM
locking system.
Page 25
4
Operation
Rotor Installation
This system is used to automatically lock the rotor to the centrifuge spindle. The rotor does not have to
be bolted on to the centrifuge spindle.
Proceed as follows:
1. Open the lid of the centrifuge and if necessary remove any dust, foreign objects or residue from the
chamber.
TM
AutoLock
and o-ring must be clean and undamaged.
AutoLock
TM
O-Ring
Figure 4-1. AutoLock
TM
2. Place the rotor over the centrifuge spindle and let it slide slowly down the centrifuge spindle.
The rotor clicks automatically into place.
CAUTION Do not force the rotor onto the centrifuge spindle.
If the rotor is very light, then it may be necessary to press it onto the centrifuge spindle
with a bit of pressure.
3. Check if the rotor is properly installed by lifting it slightly on the handle. If the rotor can be pulled
up, then it must be reclamped to the centrifuge spindle.
WARNING If the rotor cannot be properly locked in place after several attempts, then the
AutoLock
TM
is defective and you are not permitted to operate the rotor.
Check for any damage to the rotor: Damaged rotors must not be used.
Keep the centrifuge spindle area of the rotor clear of objects.
CAUTION Check that the rotor is properly locked on the centrifuge spindle before each
use by pulling it at its handle.
4. If available close the rotor with the rotor lid.
Be sure to check all sealings before starting any aeorsol-tight applications.
See the information in the rotor instruction manual.
5. The display shows:
Thermo ScientificST40 / 40R4-3
Page 26
4
Operation
Entering Parameters
Entering Parameters
The Sorvall ST40 / 40R offers you a total of 9 acceleration and 10 braking curves with which samples
and gradients can be centrifuged.
After the centrifuge is turned on, the last running profile selected is shown.
Acceleration curve
1. Press the key below the ACC display in order to open the acceleration profile selection menu.
The display shows the message "Set acceleration".
2. Select the profile by pressing the key , until the desired acceleration curve shows.
Braking curve
1. Press the key below the DEC display in order to open the braking curve selection menu.
The display shows the message "Set deceleration“
2. Select the profile by pressing the key , until the desired braking curve shows.
Preselecting Speed / RCF
1. Press the key below the “SPEED” display in order to open the speed / RCF value menu.
The display shows the "RPM“ or the "RCF"-value depending on the display setting. Press the
CHANGE key to toggle between the two modes.
2. Enter the desired value by pressing the key repeatedly, until the desired value shows.
Note If an extremely low RCF value has been selected, it will be corrected automatically if the
resulting speed is less than 300 rpm.
4-4ST40 / 40RThermo Scientific
Page 27
4
Operation
Entering Parameters
Explanation of RCF Value
The relative centrifugal force is given as a multiple of the force of gravity g. It is a unitless numerical
value which is used to compare the separation or sedimentation capacity of various devices, since it
is independent of the type of device. Only the centrifuging radius and the speed come into play in
it:
2
n
RCF11 18,
r = centrifuging radius in cm
n = Rotational speed in rpm
The maximum RCF value is related to the maximum radius of the tube opening.
Remember that this value is reduced depending on the tubes and adapters used.
This can be accounted for in the calculation above if required.
-------------
〈〉
1000
r××=
Running time preselection
1. Press the key below the TIME display in order to open the runtime selection menu.
The display shows the message "Set Time".
Enter the desired runtime in H.mm.
2. Enter the desired value by pressing the key repeatedly, until the desired value shows.
Continuous operation
1. Press the START key .
During continuous operation, the centrifuge will continue running until you stop it manually.
Preselecting the temperature
You can preselect temperatures between -10 °C and +40 °C.
To set the temperature, proceed as follows:
1. Press the key below the TEMPERATURE display in order to open the temperature selection
menu.
The display shows "Set temp":
2. Enter the desired value by pressing the key repeatedly, until the desired value shows.
Thermo ScientificST40 / 40R4-5
Page 28
4
Operation
Saving programs
Prewarming or precooling the centrifuge
For setting the pretemp value for the centrifuge proceed as follows:
1. Press the key in order to open the temperature selection menu.
The display shows the message "Set PreTemp".
2. Enter the desired value by pressing the key repeatedly, until the desired value shows.
3. The display shows:
The display shows the current temperature inside the rotor chamber.
4. Press the START key .
The rotor chamber is cooled down or heated up to the preset temperature.
5. Press the STOP key .
The display shows the current temperature inside the rotor chamber.
Bucket selection
Bucket selection is only possible for swing-out rotors. The bucket code corresponds to the last four
digits of the bucket catalog number.
1. Press the BUCKET key .
The display shows the following message:
2. Press the BUCKET key
repeatedly until the bucket being used is displayed.
Saving programs
1. Enter the program parameters.
2. Press any of the program store keys for 4 seconds.
4-6ST40 / 40RThermo Scientific
Page 29
Centrifugation
Once the rotor has been properly installed, the main switch has been turned on and the lid has been
closed, you can start centrifuging.
Starting centrifuge program
Press the START key on the control panel. The centrifuge accelerates to the pre-set speed with the
time display active.
If the speed setting is higher than the maximum permissible speed or RCF-value for the particular
rotor, then the display will show the message max. 4,700 rpm once the centrifuge has been started.
Within 15 seconds you can apply this value by pressing the START key again, and the centrifuging
program will continue. Otherwise the centrifuge will stop and you will have to enter a valid number.
You cannot open the lid as long as the centrifuge is running.
Imbalance indicator
4
Operation
Centrifugation
If a load is imbalanced, this will be indicated at speed higher than approx. 300 rpm by the message
"Imbalanced load".
The run will terminate.
Check the loading and start the centrifuge once again. See the information on proper loading in the
rotor instruction manual. For information on troubleshooting, see section “Troubleshooting by User”
on page 7-3.
Stopping the centrifugation program
With preset running time
Usually the running time is preset and you only have to wait until the centrifuge stops automatically
when the preset time limit expires.
As soon as the speed drops to zero, the message END will appear in the display. By pressing the OPEN
key
You can also stop the centrifuging program manually at any time by pressing the STOP key .
Continuous operation
, you can open the lid and remove the centrifuge material.
If you selected continuous operation (see “Continuous operation” on page 4-5), you will have to stop
the centrifuge manually. Press the STOP key on the control panel. The centrifuge will be
decelerated at the designated rate. The message "END" will illuminate, and after pressing the key
OPEN , the lid will open and you can remove the centrifuged material.
Thermo ScientificST40 / 40R4-7
Page 30
4
Operation
Temperature Adaptation during Standstill
Temperature Adaptation during Standstill
The temperature cannot be adapted until the rotor has been positively identified;the speed display will
then show END.
When the rotor is not recognized (lid closed and START key not yet pressed, speed display "0"),
the centrifuge responds by ensuring that the sample cannot freeze regardless of the rotor being used.
Short-term Centrifugation
For short-term centrifuging, the Sorvall ST40 / 40R has a PULSE- function.
By holding down the PULSE key , spinning will start and continue until the key is let go.
The centrifuge accelerates and brakes at maximum power. Any rpm or RCF entered beforehand is
overridden.
Note The centrifuge accelerates to maximum speed, regardless of which rotor was installed.
Check carefully whether you have to maintain a certain speed for your application.
During the acceleration process, time is counted forwards in seconds. The reading stays displayed until
the centrifuge lid is opened.
Removing the Rotor
To remove the rotor, proceed as follows:
1. Open the centrifuge lid.
2. Grab the rotor handle with both hands and press against the green AutoLock
time, pull the rotor directly upwards with both hands and remove it from the centrifuge spindle.
Make sure not to tilt the rotor while doing this.
TM
key. At the same
4-8ST40 / 40RThermo Scientific
Page 31
Aerosol-tight Rotors
When using an aerosol tight lid the rotor can only be removed with the lid closed. This is to protect
you and the samples.
4
Operation
Aerosol-tight Rotors
CAUTION Rotors supplied with a lid for aerosol-tight applications come with a mandrill,
which belongs to the AutoLock
prevent it from being damaged.
Figure 4-2. AutoLock
WARNING Mind the AutoLockTM-mandrill inside the lid. Do not touch the mandrill.
TM
Aligning the Centrifuge
TM
. Be sure not to place the lid onto this mandrill to
-lid for aerosol-tight rotors
• To turn off the centrifuge put the mains switch to "0".
Thermo ScientificST40 / 40R4-9
Page 32
System Menu
Enter system menu?
Z
Z
Start
Start
Stop
Stop
5
Press the STOP key
to exit the system menu.
Language
V
V
End of run beep?
V
V
Keypad beep?
VU
VU
LCD powersave?
V
V
Auto lid open?
V
V
English
U
U
no
U
U
U
U
U
U
no
no
no
English
Deutsch
Francaise
Español
Italiano
Nederlands
Росс
кий
.
yes
no
Software ID:
V
V
Cycle count:
V
V
Thermo ScientificST40 / 40R5-1
0547V01
U
U
2368
U
U
0547V01
4641V05
*
Software and NVRAM identification
example values.
*
Page 33
5
System Menu
Description
Description
Enter system menu
To enter the system menu hold down any of the keys when switching the centrifuge on.
Change the selection by presing the keys UV.
Use the keys UV below the speed display in order to navigate through the system menu.
Use the keys UV below the bucket selection in order to navigate within the system menu points.
Press the STOP key to quit the system menu.
Language
Use the keys UV below the bucket selection in order to change the language in the display until the
desired language appears in the display.
Use the keys UV below the speed display in order to navigate through the system menu.
Press the STOP key to quit the system menu.
End of run beep
Use the keys UV below the bucket selection until it says YES in the display if the centrifuge should
make beep after the run. Otherwise use the keys UV below the bucket selection until it says NO.
Use the keys UV below the speed display in order to navigate through the system menu.
Press the STOP key to quit the system menu.
Keypad beep
Use the keys UV below the bucket selection until it says YES in the display if the centrifuge should
make beep when pressing any key. Otherwise use the keys UV below the bucket selection until it says
NO.
Use the keys UV below the speed display in order to navigate through the system menu.
Press the STOP key to quit the system menu.
LCD powersave
Use the keys UV below the bucket selection until it says YES in the display if the centrifuge should
enter a powdersave mode after the run. Otherwise use the keys UV below the bucket selection until it
says NO.
Use the keys UV below the speed display in order to navigate through the system menu.
Press the STOP key to quit the system menu.
5-2ST40 / 40RThermo Scientific
Page 34
5
System Menu
Description
Auto lid open
Use the keys UV below the bucket selection until it says YES in the display if the centrifuge should
open after the run. Otherwise use the keys UV below the bucket selection until it says NO.
Use the keys UV below the speed display in order to navigate through the system menu.
Press the STOP key to quit the system menu.
Software ID
Here you fine the current software version.
Use the keys UV below the speed display in order to navigate through the system menu.
Press the STOP key to quit the system menu.
Cycle count
Here you fine the current numbers of cycles.
Use the keys UV below the speed display in order to navigate through the system menu.
Press the STOP key to quit the system menu.
Thermo ScientificST40 / 40R5-3
Page 35
Maintenance and Care
Contents
• “Cleaning Intervalls” on page 6-2
• “Cleaning” on page 6-2
• “Disinfection” on page 6-3
• “Decontamination” on page 6-4
• “Autoclaving” on page 6-5
6
• “Service of Thermo Fisher Scientific” on page 6-5
Thermo ScientificST40 / 40R6-1
Page 36
6
Maintenance and Care
Cleaning Intervalls
Cleaning Intervalls
For the sake of personal, environmental, and material protection, it is your duty to clean and if
necessary disinfect the centrifuge on a regular basis.
MaintenanceRecommended interval
Clean rotor chamberdaily or when polluted
Clean rotordaily or when polluted
Accessoriesdaily or when polluted
CabinetOnce per month
Ventilation holesEvery six months
CAUTION Refrain from using any other cleaning or decontamination procedure than
those recommended here, if you are not entirely sure that the intended procedure is safe
for the equipment.
Use only approved cleansers.
If in doubt, contact Thermo Fisher Scientific.
Cleaning
When cleaning centrifug
• Use warm water with a neutral solvent.
• Never use caustic cleaning agents such as soap suds, phosphoric acid, bleaching solutions or
scrubbing powder.
• Rinse the cavities out thoroughly.
• Use a soft brush without metal bristles to remove stubborn residue.
• Afterwards rinse with distilled water.
• Place the rotors on a plastic grate with their cavities pointing down.
• If drying boxes are used, the temperature must never exceed 50 °C, since higher temperatures could
damage the material and shorten the lifetime of the parts.
• Use only disinfectants with a pH of 6-8.
• Dry aluminum parts off with a soft cloth.
• After cleaning, treat the entire surface of aluminum parts with corrosion protection oil (7000
9824). Also treat the cavities with oil.
• Store the aluminum parts at room temperature or in a cold-storage room with the cavities pointing
down.
6-2ST40 / 40RThermo Scientific
Page 37
6
Maintenance and Care
CAUTION Before using any cleaning or decontamination methods except those
recommended by the manufacturer, users should check with the manufacturer that the
proposed method will not damage the equipment.
Clean centrifuge and accessories as follows:
1. Open the centrifuge.
2. Turn off the centrifuge.
3. Pull out the power supply plug.
4. Grasp the rotor with both hands and lift it vertically off the centrifuge spindle.
5. Remove the centrifuge tubes and adaptors.
6. Use a neutral cleaning agent with a pH value between 6 and 8 for cleaning.
7. Dry all of the rotors and accessorie after cleaning with a cloth or in a warm air cabinet at a
maximum temperature of 50°C.
Disinfection
• After cleaning, treat the entire surface of aluminum parts with corrosion protection oil (7000
• Tread the bold of the swing out rotor with bold grease (75003786).
After some applications their might be ice in the rotor chamber. Let the ice melt and drain it off. Clean
the rotor chamber as described above.
Disinfection
Disinfect the centrifuge immediately whenever infectious material has spilled during centrifugation.
9824). Also treat the cavities with oil.
CAUTION When cleaning, do not allow liquids, especially organic solvents, to get on the
drive shaft or the bearings of the centrifuge.
Organic solvents break down the grease in the motor bearing. The drive shaft could freeze
up.
WARNING Infectious material can get into the centrifuge when a tube breaks or as a
result of spills. Keep in mind the risk of infection when touching the rotor and take all
necessary precautions.
In case of contamination, make sure that others are not put at risk.
Decontaminate the affected parts immediately.
Take other precautions if need be.
The rotor chamber and the rotor should be treated preferably with a neutral disinfectant.
CAUTION Before using any cleaning or decontamination methods except those
recommended by the manufacturer, users should check with the manufacturer that the
proposed method will not damage the equipment.
Observe the safety precautions and handling instructions for the cleaning agents used.
Thermo ScientificST40 / 40R6-3
Page 38
6
Maintenance and Care
Decontamination
Contact the Service Department of Thermo Fisher Scientific for questions regarding the use of other
disinfectants.
Disinfect the rotor and accessories as follows:
1. Open the centrifuge.
2. Turn off the centrifuge.
3. Pull out the power supply plug.
4. Grasp the rotor with both hands and lift it vertically off the centrifuge spindle.
5. Remove the centrifuge tubes and adaptors and dispose of them or disinfect them.
6. Treat the rotor and accessories according to the instructions for the disinfectant (soak in solution).
Adhere strictly to the given application times.
7. Be sure the disinfectant can drain off the rotor.
8. Rinse the rotor and accessories thoroughly with water.
9. Dispose of the disinfectant according to the applicable guidelines.
10. Dry all of the rotors and accessorie after cleaning with a cloth or in a warm air cabinet at a
maximum temperature of 50°C.
• After cleaning, treat the entire surface of aluminum parts with corrosion protection oil (7000
9824). Also treat the cavities with oil.
• Tread the bold of the swing out rotor with bold grease (75003786).
Decontamination
Decantaminate the centrifuge immediately whenever radioactive material has spilled during
centrifugation.
WARNING Radioactive material can get into the centrifuge when a tube breaks or as a
result of spills. Keep in mind the risk of infection when touching the rotor and take all
necessary precautions.
In case of contamination, make sure that others are not put at risk.
Decontaminate the affected parts immediately.
Take other precautions if need be.
CAUTION Before using any cleaning or decontamination methods except those
recommended by the manufacturer, users should check with the manufacturer that the
proposed method will not damage the equipment.
For general radioactive decontamination use a solution of equal parts of 70% ethanol, 10% SDS and
water.
1. Open the centrifuge.
2. Turn off the centrifuge.
3. Pull out the power supply plug.
6-4ST40 / 40RThermo Scientific
Page 39
6
Maintenance and Care
4. Grasp the rotor with both hands and lift it vertically off the centrifuge spindle.
5. Remove the centrifuge tubes and adaptors and dispose of them or disinfect them.
6. Rinse the rotor first with ethanol and then with de-ionized water.
• Adhere strictly to the given application times.
7. Be sure the decontamination solution can drain off the rotor.
8. Rinse the rotor and accessories thoroughly with water.
9. Dispose of the decontamination solution according to the applicable guidelines.
10. Dry all of the rotors and accessorie after cleaning with a cloth or in a warm air cabinet at a
maximum temperature of 50°C.
• After cleaning, treat the entire surface of aluminum parts with corrosion protection oil (7000
9824). Also treat the cavities with oil.
• Tread the bold of the swing out rotor with bold grease (75003786).
Autoclaving
Autoclaving
1. Before autoclaving clean rotor and accessories and described above.
2. Place the rotor on a flat surface.
• Rotors and adapter can beautoclaved at 121 °C.
• The maximum permissible autoclave cycle is 20 minutes at 121 °C.
Note No chemical additives are permitted in the steam.
CAUTION Never exceed the permitted temperature and duration when autoclaving.
If the rotor shows signs of corrosion or wear, it must be replaced.
Service of Thermo Fisher Scientific
Thermo Fisher Scientific recommends having the centrifuge and accessories serviced once a year by an
authorized service technician. The service technicians check the following:
• the electrical equipment
• the suitability of the set-up site
• the lid lock and the safety system
• the rotor
• the fixation of the rotor and the drive shaft
Thermo ScientificST40 / 40R6-5
Page 40
6
Maintenance and Care
Service of Thermo Fisher Scientific
Thermo Fisher Scientific offers inspection and service contracts for this work. Any necessary repairs are
performed for free during the warranty period and afterwards for a charge.
This is only valid if the centrifuge has only been mainained by a Thermo Fisher Scientific service
technician.
6-6ST40 / 40RThermo Scientific
Page 41
Trouble Shooting
Contents
• “Mechanical Emergency Door Release” on page 7-2
• “Troubleshooting by User” on page 7-3
• “When to contact Customer Service” on page 7-5
7
Thermo ScientificST40 / 40R7-1
Page 42
7
Trouble Shooting
Mechanical Emergency Door Release
Mechanical Emergency Door Release
During a power failure, you will not be able to open the centrifuge lid with the regular electric lid
release. A mechanical override is provided to allow sample recovery in the case of an emergency.
However, this should be used only in emergencies and after the rotor has come to a complete stop.
WARNING The rotor can still be spinning at high speed. If touched, it can cause serious
injuries.
Always wait a few minutes until the rotor has come to a stop without braking. The brake does not
work when there is no current. The braking process lasts much longer than usual.
Proceed as follows:
1. Make sure the rotor has stopped (view port in the lid).
WARNING Never use your hand or other tools to brake the rotor.
2. Pull out the power supply plug.
3. On the back of the housing are two black plastic plugs which you can pry out of the back plate
with a screwdriver or a knife.
Pull the release cord attached to it at the same time to trigger the mechanical lid release. The lid
will open and the samples can be removed.
Emergency Door Release
Figure 7-3. Emergency Door Release
Note You need to pull both levers in order to unlock both locks.
4. Push the cord back into the centrifuge and mount the plug.
7-2ST40 / 40RThermo Scientific
Page 43
5. Reconnect the centrifuge once the power has been restored. Switch on the centrifuge. Press the
OPEN key
to have the door locks operative again.
WARNING If you pull only one lever or you do not pressed the OPEN key to have
the door locks operative again, the door might open while the rotor os still spinning.
Troubleshooting by User
If problems occur other than those listed in this table, the authorized customer service
representative must be contacted.
Failure messageProblem with centrifugePossible causes and cures
7
Trouble Shooting
Troubleshooting by User
OvertemperatureThe centrifuge cannot be operated.
The run does not start or the centrifuge
runs down without being braked.
Incorrect bucket IDThe centrifuge cannot be operated.
The run does not start or the centrifuge
runs down without being braked.
Unapproved rotorThe centrifuge cannot be operated.
The run does not start or the centrifuge
runs down without being braked.
Rotor ID failureThe centrifuge cannot be operated.
The run does not start or the centrifuge
runs down without being braked.
Overheating in chamber.
Check the function of the refrigeration unit.
Clean the air inlet for the condenser.
Restart the centrifuge.
If an error message appears again, inform Customer Service.
Bucket code is undefined for the rotor detected, check the set
points for the given bucket code.
Is it permitted to use the current bucket in the rotor currently
mounted?
Restart the centrifuge.
If an error message appears again, inform Customer Service.
Rotor code is not in the rotor table.
Is it permitted to use the rotor currently mounted in this device?
Restart the centrifuge.
If an error message appears again, inform Customer Service.
The rotor could not be identified.
Check to see if the rotor is properly installed.
Restart the centrifuge.
If an error message appears again, inform Customer Service.
Lid blockedCentrifuge does not openRestart the centrifuge.
The emergency lid release enables you to retrieve your samples.
If an error message appears again, inform Customer Service.
Motor overtemp.The centrifuge cannot be operated.
The run does not start or the centrifuge
runs down without being braked.
PCB overtemp.The centrifuge cannot be operated.
The run does not start or the centrifuge
runs down without being braked.
Thermo ScientificST40 / 40R7-3
Restart the centrifuge.
If an error message appears again, inform Customer Service.
Restart the centrifuge.
If an error message appears again, inform Customer Service.
Page 44
7
Trouble Shooting
Troubleshooting by User
Failure messageProblem with centrifugePossible causes and cures
Emergency releaseThe centrifuge cannot be operated.
The run does not start or the centrifuge
runs down without being braked.
Imbalanced loadThe centrifuge cannot be operated.
The run does not start or the centrifuge
runs down without being braked.
Check Set SpeedThe centrifuge cannot be operated.
The run does not start or the centrifuge
runs down without being braked.
E-01 - E-12The centrifuge cannot be operated.
The run does not start or the centrifuge
runs down without being braked.
E-13The centrifuge cannot be operated.
The run does not start or the centrifuge
runs down without being braked.
The lid opens while the device is running.
Close the lid and restart the centrifuge.
If an error message appears again, inform Customer Service.
Imbalance detected.
Check the load placed in the rotor.
Check that the rotor cross bolts are well greased.
Restart the centrifuge.
If an error message appears again, inform Customer Service.
The set point speed is higher than the maximum rotor speed.
Correct the value.
Restart the centrifuge.
If an error message appears again, inform Customer Service.
Error during the self-test of the centrifuge program and the
electronics.
Restart the centrifuge.
If an error message appears again, inform Customer Service.
The check sum in the data memory is incorrect.
The software corrects errors automatically.
Check the values of the set point settings, etc.
Restart the centrifuge.
If an error message appears again, inform Customer Service.
E-15-E-16Temperature sensor broken / controller
defective
Malfunction in the temperature detection.
Restart the centrifuge.
If an error message appears again, inform Customer Service.
E-17Speed for rotor detection exceededRestart the centrifuge.
If an error message appears again, inform Customer Service.
E-21-E-22The centrifuge cannot be operated.
The run does not start or the centrifuge
runs down without being braked.
The rotor could not be identified.
Check to see if the rotor is properly installed.
Restart the centrifuge.
If an error message appears again, inform Customer Service.
E-23The centrifuge cannot be operated.
The run does not start or the centrifuge
runs down without being braked.
E-25-E-27The centrifuge cannot be operated.
The run does not start or the centrifuge
runs down without being braked.
E-28The centrifuge cannot be operated.
The run does not start or the centrifuge
runs down without being braked.
The speed control measurement returned a different result.
Restart the centrifuge.
If an error message appears again, inform Customer Service.
Time has expired for the lid lock drive while opening the lid.
Restart the centrifuge.
If an error message appears again, inform Customer Service.
Error during the self-test of the centrifuge program and the
electronics.
Restart the centrifuge.
If an error message appears again, inform Customer Service.
7-4ST40 / 40RThermo Scientific
Page 45
Failure messageProblem with centrifugePossible causes and cures
7
Trouble Shooting
When to contact Customer Service
E-29The centrifuge cannot be operated.
The run does not start.
E-30The centrifuge cannot be operated.
The run does not start or the centrifuge
runs down without being braked.
E-33The centrifuge cannot be operated.
The run does not start or the centrifuge
runs down without being braked.
E-34-E-36The centrifuge cannot be operated.
The run does not start or the centrifuge
runs down without being braked.
E-40The centrifuge cannot be operated.
The run does not start or the centrifuge
runs down without being braked.
Check wether you selected the right bucket.
Is it easy to turn the rotor when the lid is open?
Does the rotor rub against the device?
Restart the centrifuge.
If an error message appears again, inform Customer Service.
Restart the centrifuge.
If an error message appears again, inform Customer Service.
Overpressure in the refrigeration unit.
Clean the air inlet for the condenser.
Restart the centrifuge.
If an error message appears again, inform Customer Service.
Error during the self-test of the centrifuge program and the
electronics.
Restart the centrifuge.
If an error message appears again, inform Customer Service.
The centrifuge accelerates too slowly.
Check wether you selected the right bucket.
Check wether you selected the right bucket.
Is it easy to turn the rotor when the lid is open?
Does the rotor rub against the device?
Restart the centrifuge.
If an error message appears again, inform Customer Service.
E-41-E-74The centrifuge cannot be operated.
The run does not start or the centrifuge
runs down without being braked.
When to contact Customer Service
If you need to contact customer service, please provide the order no. and the serial no. of your device.
This information can be found on the back near the inlet for the power supply cable.
To identify the software version, proceed as follows:
1. Hold down any of the keys and then switch on the centrifuge.
You enter the system menu.
2. Press the START key .
3. Press the Uor V key, until the following message is displayed:
Software ID:
XXXXXXXX
Error during the self-test of the centrifuge program and the
electronics.
Restart the centrifuge.
If an error message appears again, inform Customer Service.
4. Communicate the software version to the service technician.
Thermo ScientificST40 / 40R7-5
Page 46
Chemical Compatibility Chart
A
®
CCCLEARCRIMP
®
CHEMICAL
2-mercaptoethanol
Acetaldehyde
Acetone
Acetonitrile
®
Alconox
Allyl Alcohol
Aluminum Chloride
Formic Acid (100%)
Ammonium Acetate
Ammonium Carbonate
Ammonium Hydroxide (10%)
Ammonium Hydroxide (28%)
Ammonium Hydroxide (conc.)
Ammonium Phosphate
Ammonium Sulfate
Amyl Alcohol
Aniline
Sodium Hydroxide (<1%)
Sodium Hydroxide (10%)
Barium Salts
Benzene
Benzyl Alcohol
POLYSULFONE
®
, TEFLON
®
POLYVINYL CHLORIDE
RULON A
SILICONE RUBBER
STAINLESS STEEL
TITANIUM
TYGON®VITON
,CLEARCRIMP
®
, POLYCLEAR
MATERIAL
ALUMINUM
ANODIC COATING for ALUMINIUM
BUNA N
CELLULOSE ACETATE BUTYRATE
POLYURETHANE ROTOR PAINT
COMPOSITE Carbon Fiber/Epoxy
DELRIN®ETHYLENE PROPYLENE
GLASS
NEOPRENE
NORYL®NYLON
*
PET
POLYALLOMER
POLYCARBONATE
POLYESTER, GLASS THERMOSET
POLYTHERMIDE
POLYRTHYLENE
POLYPROPYLENE
SSU- SMS- SUSSUSS- SSSSUSSSSSS
S-UU---M-U---MU U U MM- MS U - S - U
MSUUSUMS SUUSUSUUUSSUUSMMSUU
SSU- SMS- SSUSUMUU- SMUUSSSSUU
UUS- SSS- SSSSSSMSSSSSSSSSSSU
---U--S----S-SSMSSS- MS- - S- -
UUSSSSUSSSSMSSSS- SSSSSMUUSS
-SMU--U----U-SMUUSS-US-US-U
SSU- SSS- SSSSSSSU- SSSSSSSSSS
MSUSSSSSSSSSSSUU- SSSSSSMSSS
UUSUSSMSSSSS- SUMSSSSSSSSSMS
UUSUSUMSSSSSUSUMSSSSSSSSSMS
UUUUSUMS- S- SUSUUSSS- MSSSS- U
U- S- SSSSSSSS- SSM- SSSSSSMSSS
UMS- SSUSSSSSSSSS- SSSSSSUSSU
S- MU- - SS- M- S- MSSSSM---U-S-M
SSUUSUSMSUUUUUUU- SMUUSSSSUS
U- MSSS- - SMSS- SMMSSSSSSMSS- U
U- MU- - U- MMSSUSUUSSSSSSMSS- U
MUS- SSSSSSSSSSSM- SSSSSSMSSS
SSUUSUMUSUUSUUUMUMUUUSUUSUS
S- UU- - MM- M- SUUUUUUU- MSM- S- S
®
Thermo ScientificST40 / 40RA-1
Page 47
A
Chemical Compatibility Chart
®
CCCLEARCRIMP
®
CHEMICAL
Boric Acid
Cesium Acetate
Cesium Bromide
Cesium Chloride
Cesium Formate
Cesium Iodide
Cesium Sulfate
Chloroform
Chromic Acid (10%)
Chromic Acid (50%)
Cresol Mixture
Cyclohexane
Deoxycholate
Distilled Water
Dextran
Diethyl Ether
Diethyl Ketone
Diethylpyrocarbonate
Dimethylsulfoxide
Dioxane
Ferric Chloride
Acetic Acid (Glacial)
Acetic Acid (5%)
Acetic Acid (60%)
Ethyl Acetate
Ethyl Alcohol (50%)
Ethyl Alcohol (95%)
Ethylene Dichloride
Ethylene Glycol
MATERIAL
ALUMINUM
ANODIC COATING for ALUMINIUM
,CLEARCRIMP
®
, POLYCLEAR
BUNA N
CELLULOSE ACETATE BUTYRATE
POLYURETHANE ROTOR PAINT
COMPOSITE Carbon Fiber/Epoxy
DELRIN®ETHYLENE PROPYLENE
GLASS
NEOPRENE
NORYL®NYLON
*
PET
POLYALLOMER
POLYCARBONATE
POLYESTER, GLASS THERMOSET
POLYTHERMIDE
POLYRTHYLENE
POLYPROPYLENE
POLYSULFONE
®
, TEFLON
®
POLYVINYL CHLORIDE
RULON A
SILICONE RUBBER
STAINLESS STEEL
TITANIUM
TYGON®VITON
USSMSSUSSSSSSSSSUSSSSSSSSSS
M- S- SSS- SSSS- SS- - SSSSSSMSSS
MSS- SSS- SSSSSSS- - SSSSSSMSSS
MSSUSSS- SSSSSSS- - SSSSSSMSSS
MSS- SSS- SSSSSSS- - SSSSSSMSSS
MSS- SSS- SSSSSSS- - SSSSSSMSSS
MSS- SSS- SSSSSSS- - SSSSSSMSSS
UUUUSSMUSUUMUMUUUMMUUS UUUMS
U- UUSUU- SSSUSSMUMSSUMSMUSSS
U-UU-UU---SUUSMUMSSUMS-UM-S
SSU---S-SUUUUUU- - UU- USSSSUS
SSS- SSSUSUSSUUUMSMUMMSUMMUS
SSS- SSS- SSSSSSS- - SSSSSSSSSS
SSSSSSSSSSSSSSSSSSSSSSSSSSS
MSSSSSS- SSSSSSSSSSSSSSSMSSS
SSUUSSSUSUUSUUUUUUUUUSSSSMU
S- UU- - M- SU- S- MUUUMM- US- - SUU
SSU- SSS- SSUSUSU- - SSSMSSSSSS
SSUUSSS- SUSSUSUU- SSUUSSSSUU
MS U U S S MMS U U S U MU U - MMMU SSSSUU
UUS---MS-M-S-S---SS---MUS-S
SSUUSSUMSUSUUUUUMSUMUSUUS- U
SSMSSSMSSSSSMSSSSSSSMSSMSSM
S S U U S S U - S MS U U MU S MS MS MS MU S MU
MMUUSS MMSSUSUMUU- SSUUSMMSUU
SSSSSSMSSSSSUSUSSSSSSSSMSMU
SSSUSSMSSSSSUSU- SSSMSSSUSMU
S- UU- - SM- UUSUUUUUUU- USU- S- S
SSSSSSSSSSSS- SUSSSSSSSSMSMS
®
A-2ST40 / 40RThermo Scientific
Page 48
A
Chemical Compatibility Chart
®
CCCLEARCRIMP
®
CHEMICAL
Ethylene Oxide Vapor
Ficoll-Hypaque
Hydrofluoric Acid (10%)
Hydrofluoric Acid (50%)
Hydrochloric Acid (conc.)
Formaldehyde (40%)
Glutaraldehyde
Glycerol
Guanidine Hydrochloride
Haemo-Sol
Hexane
Isobutyl Alcohol
Isopropyl Alcohol
Iodoacetic Acid
Potassium Bromide
Potassium Carbonate
Potassium Chloride
Potassium Hydroxide (5%)
Potassium Hydroxide (conc.)
Potassium Permanganate
Calcium Chloride
Calcium Hypochlorite
Kerosene
Sodium Chloride (10%)
Sodium Chloride (sat'd)
Carbon Tetrachloride
Aqua Regia
Solution 555 (20%)
Magnesium Chloride
®
®
MATERIAL
ALUMINUM
ANODIC COATING for ALUMINIUM
,CLEARCRIMP
®
, POLYCLEAR
BUNA N
CELLULOSE ACETATE BUTYRATE
POLYURETHANE ROTOR PAINT
COMPOSITE Carbon Fiber/Epoxy
DELRIN®ETHYLENE PROPYLENE
GLASS
NEOPRENE
NORYL®NYLON
*
PET
POLYALLOMER
POLYCARBONATE
POLYESTER, GLASS THERMOSET
POLYTHERMIDE
POLYRTHYLENE
POLYPROPYLENE
POLYSULFONE
®
, TEFLON
®
POLYVINYL CHLORIDE
RULON A
SILICONE RUBBER
STAINLESS STEEL
TITANIUM
TYGON®VITON
S- U- - U- - SU- S- SM- - SSSUSUSSSU
MSS- SSS- SSSS- SS- SSSSSSSMSSS
UUUM- - U- - UUS- SMUSSSSMSUUU- -
UUUU- - U- - UUUUSUUUSS MMSUUU- M
UUUU- UUM- UMUUMUUU- S- USUUU- -
MMMSSSSMSSSSMSSSUSSMSSSMSMU
SSSS- - S- SSSSSSS- - SSS- - SSS- -
MSS- SSSSSSSSSSSS- SSSSSSSSSS
UUS- SSS- SSSSSSS- - SSSSSSUSSS
SSS---S-SSSSSSS- - SSSSSSSSSS
SSS- SSS- SSUSUMUSSUSSMSUSSUS
- - MU- - SS- U- SUSSMSSS- SSS- S- S
MMMU SSSSSUSSUSUMSSSSSSSMMMS
SSM- SSS- SMSSMSS- MSSSSSMSSMM
USS- SSS- SSSSSSSSSSS- SSSMSSS
MUSSSSS- SSSSSSUSSSSSSSSSSSS
USS- SSSSSSSSSSS- SSSSSSSUSSS
UUSSSSM- SSSS- SUSSSSSSSMUMSU
UUMU- - M- MSS- UMUUUSM- MU- UU- U
SSS- SSS- SSSUSSSM- SMSUSSMSUS
MUSSSSSSSSSSSSMS- SSSSSSMSSS
M- U- SMMS- M- S- SMS- SSSMSMUS- S
SSS- SSSUSMUSUMMS- MMMSSUSSUS
S- SSSSSS- - - SSSSS- SSSS- SSM- S
U-SUSSS----SSSSS- SS- S- SSM- S
U U MS S U MU S U U S U MU S S MMS MMMMU S S
U-UU--U-----UUUUUUU-----S-M
SSS---S-SSSSSSS- - SSS- SSSSSS
MSS- SSSSSSSSSSSSSSSSSSSMSSS
®
Thermo ScientificST40 / 40RA-3
Page 49
A
Chemical Compatibility Chart
®
CCCLEARCRIMP
®
CHEMICAL
Mercaptoacetic Acid
Methyl Alcohol
Methylene Chloride
Methyl Ethyl Ketone
Metrizamide
Lactic Acid (100%)
Lactic Acid (20%)
N-Butyl Alcohol
N-Butyl Phthalate
N, N-Dimethylformamide
Sodium Borate
Sodium Bromide
Sodium Carbonate (2%)
Sodium Dodecyl Sulfate
Sodium Hypochlorite (5%)
Sodium Iodide
Sodium Nitrate
Sodium Sulfate
Sodium Sulfide
Sodium Sulfite
Nickel Salts
Oils (Petroleum)
Oils (Other)
Oleic Acid
Oxalic Acid
Perchloric Acid (10%)
Perchloric Acid (70%)
Phenol (5%)
Phenol (50%)
®
MATERIAL
ALUMINUM
ANODIC COATING for ALUMINIUM
,CLEARCRIMP
®
, POLYCLEAR
BUNA N
CELLULOSE ACETATE BUTYRATE
POLYURETHANE ROTOR PAINT
COMPOSITE Carbon Fiber/Epoxy
DELRIN®ETHYLENE PROPYLENE
GLASS
NEOPRENE
NORYL®NYLON
*
PET
POLYALLOMER
POLYCARBONATE
POLYESTER, GLASS THERMOSET
POLYTHERMIDE
POLYRTHYLENE
POLYPROPYLENE
POLYSULFONE
®
, TEFLON
®
POLYVINYL CHLORIDE
RULON A
SILICONE RUBBER
STAINLESS STEEL
TITANIUM
TYGON®VITON
USU- SMS- SMSUUUU- SUUSMS USSSS
SSSUSSMSSSSSUSUMSSSSSSSMSMU
UUUUMSSUSUUSUUUUUMUUUSSMUSU
SSUUSSMSSUUSUSUUUSSUUSSSSUU
MSS- SSS- SSSS- SS- - SSSSSSMSSS
--S------MSU-SSSMSS- MSMSS- S
--SS-----MSM-SSSSSSSMSMSS- S
S- SU- - S- - SM- USMSSSSMMSM- S- S
SSU- SSS- SUUSUUUM- UUSUSMMSUS
SSSUSMS- SSUSUSUU- SSUUSMSSSU
MSSSSSSSSSSUSSSS- SSSSSSMSSS
USS- SSS- SSSSSSSS- SSSSSSMSSS
MUSSSSSSSSSSSSUSSSSSSSSSSSS
SSS- SSS- SSSSSSS- SSSSSSSSSSS
UUMSSMUSSMSSSMSSSSMSSSMUSMS
MSS- SSS- SSSSSSS- - SSSSSSMSSS
SSS- SSSSSSSSSSSS- SSSSSUSSSS
USS- SSSSSSSSSSSSSSSSSSSMSSS
S-SS---S---SSSUU--S---SSM-S
SSS- SSSSMSSSSSSM- SSSSSSSSSS
USSSSS- SSS- - SSSS- SSSSSSMSSS
SSS---SUSSSSUUMSMUUSSSUSSSS
S-S---SMSSSSUSSSSUSSSS- SSMS
S- USSSUUSUSSMSSSSSSSSSMUSMM
UUMSSSUSSSSSUSUSSSSSSSSUMSS
U- U- SUU- SMM- - MUMSMM- MSU- S- S
UUU- - UU- S UMUUMUUUMMUMSUUSUS
U S U - S MM- S U MU U S U MS MS U U S U MMMS
USU- SUM- SUMUUUUUSUMUUSUUUMS
®
A-4ST40 / 40RThermo Scientific
Page 50
A
Chemical Compatibility Chart
®
CCCLEARCRIMP
®
CHEMICAL
Phosphoric Acid (10%)
Phosphoric Acid (conc.)
Physiologic Media (Serum, Urine)
Picric Acid
Pyridine (50%)
Rubidium Bromide
Rubidium Chloride
Sucrose
Sucrose, Alkaline
Sulfosalicylic Acid
Nitric Acid (10%)
Nitric Acid (50%)
Nitric Acid (95%)
Hydrochloric Acid (10%)
Hydrochloric Acid (50%)
Sulfuric Acid (10%)
Sulfuric Acid (50%)
Sulfuric Acid (conc.)
Stearic Acid
Tetrahydrofuran
Toluene
Trichloroacetic Acid
Trichloroethane
Trichloroethylene
Trisodium Phosphate
Tris Buffer (neutral pH)
Triton X-100
Urea
Hydrogen Peroxide (10%)
®
MATERIAL
ALUMINUM
ANODIC COATING for ALUMINIUM
,CLEARCRIMP
®
, POLYCLEAR
BUNA N
CELLULOSE ACETATE BUTYRATE
POLYURETHANE ROTOR PAINT
COMPOSITE Carbon Fiber/Epoxy
DELRIN®ETHYLENE PROPYLENE
GLASS
NEOPRENE
NORYL®NYLON
*
PET
POLYALLOMER
POLYCARBONATE
POLYESTER, GLASS THERMOSET
POLYTHERMIDE
POLYRTHYLENE
POLYPROPYLENE
POLYSULFONE
®
, TEFLON
®
POLYVINYL CHLORIDE
RULON A
SILICONE RUBBER
STAINLESS STEEL
TITANIUM
TYGON®VITON
UUMSSSUSSSSU- SSSSSSSSSUMUSS
UUMM- - US- MSUUMMSSSMSMSUMU- S
MSSS- - S- SSSSSSSSSSSSSSSSSSS
SSU- SMSSSMSUSSSUSSSSUSUMSMS
USUUSUU- USSUUMUU- USMUSSUUUU
MSS- SSS- SSSSSSS- - SSSSSSMSSS
MSS- SSS- SSSSSSS- - SSSSSSMSSS
MSS- SSSSSSSSSSSSSSSSSSSSSSS
MSS- SSS- SSSSSSUSSSSSSSSMSSS
UUSSSSS- SSSUSSS- SSS- SSSUSSS
USUSSUU- SUSU- SSSSSSSSSMSSSS
USUMSUU- SUSUUMMUMMMSSSUSSMS
U- UU- UU- - UUUUMUUUUMUUSUSS- S
UUMSSSU- SSSUUSUSSSSSSSSUMSS
UUUUSUU- SMSUUMUUSSSSMSMUUMM
MUUSSUU- SSMUSSSSSSSSSSUUUSS
MUUUSUU- SSMUUSUUMSSSSSUUUMS
MUUU- UUM- - MUUSUUUMSUMSUUU- S
S-S---SMSSSS- SSSSSSSSSMMSSS
SSUUSUUMSUUSUUU- MUUUUSUSSUU
SSUUSSMUSUUSUUUSUMUUUSUSUUM
UUU- SSUMSUSUUSM- MSSUUSUUUMU
S-U---MU-U-SUUUUUUUUUSU- S- S
--UU---U-U-SUUUUUUUUUSU- U- S
---S--M------S--SSS--S--S-S
USSSSSS- SSSSSSSSSSSSSSSSSSS
SSS- SSS- SSSSSSSSSSSSSSSSSSS
S- USSSS----SSSMSSSS- SSSMS- S
UUMSSUU- SSSUSSSMUSSSSSSMSUS
®
Thermo ScientificST40 / 40RA-5
Page 51
A
Chemical Compatibility Chart
®
CCCLEARCRIMP
®
CHEMICAL
Hydrogen Peroxide (3%)
Xylene
Zinc Chloride
Zinc Sulfate
Citric Acid (10%)
*
Polyethyleneterephthalate
Key
S Satisfactory
M M = Moderate attack, may be satisfactory for use in centrifuge depending on length of exposure, speed involved,
U U = Unsatisfactory, not recommended.
-- No data available. Because no organized chemical resistance data exists for materials under the stress of centrifugation,
MATERIAL
ALUMINUM
ANODIC COATING for ALUMINIUM
,CLEARCRIMP
®
, POLYCLEAR
BUNA N
CELLULOSE ACETATE BUTYRATE
POLYURETHANE ROTOR PAINT
COMPOSITE Carbon Fiber/Epoxy
DELRIN®ETHYLENE PROPYLENE
GLASS
NEOPRENE
NORYL®NYLON
*
PET
POLYALLOMER
POLYCARBONATE
POLYESTER, GLASS THERMOSET
POLYTHERMIDE
POLYRTHYLENE
POLYPROPYLENE
POLYSULFONE
®
, TEFLON
®
POLYVINYL CHLORIDE
RULON A
SILICONE RUBBER
STAINLESS STEEL
TITANIUM
SMSSS- S- SSSSSSSSMSSSSSSSSSS
SSUSSSMUSUUUUUUMUMUUUSUMSUS
UUSSSSUSSSSSSSSSSSSSSSSUSSS
USS- SSSSSSSSSSSSSSSSSSSSSSS
MSSMSSMSSSSSSSSSMSSSSSSSSSS
etc.; suggest testing under actual conditions of use.
when in doubt we recommend pretesting sample lots. suggest testing, using sample to avoid loss of valuable material.
®
TYGON®VITON
Chemical resistance data is included only as a guide to product use.