Thermo Orion AQUAfast II Operation

Operation
User messages
RANGE
RANGERANGE
RESULT
RESULTRESULT
÷Err
E0I
-Err
LO BAT
Switch the unit on using the "power" switch
The display shows the method range.
Select measuring range using the "mode" key:
t1 t2 t3 t4 t1 .......... (Scroll)
Measuring range t1: 0 - 2 FNU Measuring range t2: 2 - 20 FNU Measuring range t3: 20 - 200 FNU Measuring range t4: 200 - 2000 FNU
The display shows the selected method range:
Fill a clean and dust-free vial with the sample up to the mark (pour the sample along the inner wall of the vial to avoid air bubbles, see Note 4). Screw the cap on and align the -mark on the vial with the -mark on the instrument. Close the sample chamber using the sample chamber cover.
Press the "zero/test" key.
The measuring range symbol flashes for approx. 9 seconds.
The display shows the result in formazin FNU turbidity units.
It is advisable to calibrate the instrument before each measurement. If the ambient temperature of the last calibration does not deviate by more than ± 3°C from the current ambient temperature, the measurements are accepted by the unit. If the temperature changes are greater than this, the unit must be re-calibrated.
If re-calibration is necessary, the display shows the following:
Performance of analysis with reduced accuracy:
Press the "zero/test" key again.
Re-calibration:
see calibration mode
Optics: LED, (λ = 875 nm)
Battery: 9 V-block battery (Life 600 tests)
Auto-OFF: Automatic switch-off occurs approx. 5 minutes
after last keypress.
Ambient conditions:5-40°C
30-90% rel. humidity (non-condensing)
Compliance:
DIN EN 55 022, 61 000-4-2, 61 000-4-8, 50 082-2, 50 081-1, DIN V ENV 50 140, 50 204 FCC Part 15 Class A ICES – 003 Issue 2
Measuring Range: 0-2 FNU 2-20 FNU 20-200 FNU 200-2000 FNU Resolution: 0.1 FNU 0.1 FNU 1 FNU 1 FNU Accuracy: ± 0.2 FNU ± 1 FNU ± 5 FNU ± 50 FNU
Technical data
Light absorption too great. Reason - e.g. dirty optics.
Measuring range exceeded.
Result below the lowest limit of the measuring range.
Replace 9 V battery immediately, no further analysis are possible.
RANGE
RANGERANGE
Method notes
Non-dissolved, finely dispersed substances in liquids (media) cause turbidity. An incident light beam is scattered unevenly in all directions by the existing turbidity. The scatter of the incident infrared light is measured in standardized manner at an angle of 90°.
Supplied turbidity standards
Measuring range-based turbidity standards are used for calibration of the meter. Reorder Orion AC201S Turbidity Standards for meter calibrations. Material safety data sheets are available.
The turbidity standards for the measuring ranges
t1 1 NTU t2 10 NTU t3 100 NTU t4 1000 NTU
are supplied in vials. These pre-filled vials with the turbidity standards generally suffice for the purpose of calibration. Changed optical properties of the measuring vials (e.g. extremely fine scratches) have a considerable effect on the measuring results, particularly in the measuring ranges t1 and t2. To ensure that the specific accuracy is achieved, it is therefore necessary to perform measurements in the ranges t1 and t2 using the vial used for calibration of the unit. To this end, the standard for calibration is poured into a clean, dry measuring vial. The measuring vial used for calibration is identified by a black dot on the white triangle. We advise you to retain this method of marking.
In contrast to formazin turbidity standards, these secondary standards are neither toxic nor potentially carcinogenic.
The turbidity standards have a shelf life of one year.
The turbidity standards have been tested and approved by:
• EPA Federal Registry, Vol. 47, No. 42, March 1982
• Standard Methods of Water and Wastewater, APHA-AWWA-WPOC F, 16th and 17th Edition
• Annual Book of ASTM Standards, Standard Test Method for Turbidity of Water, D 1889-88a, June 24, 1989
• Subcommittee Report, American Society of Brewing Chemists, 1986
• Analytica - EBC, 4th Edition, 1987, Brauerei - und Getränke-Rund­schau, Zürich
SEt
Notes
1. The vial must be clean and dry (free of dust) before starting the analysis. Clean the inside and outside of the vial using a clean lint-free cloth. Fingerprints or droplets of water as well as scratches on the sides of the vial can result in errors.
2. Once it has been filled to the mark, the vial must be positioned in the sample chamber so that the graduation with the white triangle aligns with the housing mark.
3. Tests must be carried out using closed vials. Always use black vial caps. Completely cover the sample chamber with the sample chamber cover.
4. Bubbles on the inside of the vial may also lead to errors. To prevent this, cap vials and remove bubbles by swirling before performing test.
5. Avoid spilling water in the sample chamber. If water should leak into the housing of the turbidity meter, it can damage electronic components and cause corrosion.
6. Contamination of the optical components such as the light source and photo sensor in the sample chamber can result in errors.
Check the condition of the optics at regular intervals. For cleaning use a moist cloth and cotton balls.
Re-calibrate the unit each time it has been cleaned.
7. Large temperature differentials between the turbidity meter, the sample and the operating environment can lead to incorrect measurement due to, for example, the formation of condensate in the area of the lens or on the vial.
Specified tolerances at T = 20 °C.
8. Vials and caps should be cleaned thoroughly after each analysis to prevent carry-over errors. Even minor residues can cause errors in the test results.
power
zero test
mode
Orion AQUAfast® II Turbidity Meter
User messages
E 10
E 71
E 73
E 75
Calibration factor “out of range”
No light at detector or too much light at detector, maybe due to measurement range overflow
t1:
User calibration incorrect / erased
t2:
User calibration incorrect / erased
t3:
User calibration incorrect / erased
t4:
User calibration incorrect / erased
After using the interim value calibration mode, "cAL" will appear during a new calibration to indicate the previous calibration mode used.
Calibration Mode (for supplied standards with defined values)
CAL
t1
: :
Technical changes without notice Printed in Germany.
Press and hold "mode" key.
Switch unit on using "power" key. Release "mode" key after approx. 1 second.
Press "mode" key to change the measuring range:
CAL t1 CAL t2 CAL t3 CAL t4 ..... (Scroll)
Position vial (with required standard - see “Supplied turbidity standards”) with alignment of and marks. Close the sample chamber using the sample chamber cover.
Press the "zero/test" key.
The measuring range symbol flashes for approx. 9 seconds.
The display confirms that calibration is complete.
Switch the unit off using the "power" key. The new correction factor is stored.
E 77
RANGE
RANGERANGE
Calibration Mode (for interim values)
CAL
Press and hold both "mode" and "zero/test" together.
Switch unit on using "power" key, Release "mode" key after approx. 1 second.
Press "mode" key to change the measuring range:
CAL t1 CAL t2 CAL t3 CAL t4 ..... (Scroll)
Fill a clean and dust-free vial with the standard up to the mark (pour the standard along the inner wall of the vial to avoid air bubbles, see Note 5). Screw the cap on and align the ∇-mark on the vial with the -mark on the instrument. Close the sample chamber using the sample chamber cover.
Press the "zero/test" key.
The method symbol flashes for approx. 9 seconds.
If the result corresponds with the value of the calibration standard used (within the allowed tolerance), exit calibration mode by pressing the "power" key.
Otherwise, pressing the "mode" key once increases the displayed value by 1 digit. Pressing the "zero/test" key once decreases the displayed value by 1 digit.
Press the relevant key until the displayed value equals the value of the calibration standard.
By pressing the "power" key twice, the new correction factor is calculated and stored.
Confirms calibration (3 seconds). The unit then switches itself off.
RANGE
RANGERANGE
CAL
RESULT + x
RESULT + xRESULT + x
RESULT
RESULTRESULT
: :
cAL
233312-001 Rev.D
E 24
zero test
power
mode
power
mode
zero test
power
power
mode
zero
test
zero
test
Loading...